1 Doppler Technology & Optimization Doppler From A Z Donald T. Milburn, RDCS, RVT, FSDMS Prescott, Az. San Diego local ultrasound society May 12, 2009 Donald T. Milburn, RDCS, RVT, NACT, FSDMS Director, Radiology & Vascular Education Programs 2 I d like to begin with this quote; The best interest of the patient, is the only interest to be considered. William J. Mayo What we will do 2day Doppler Overview Doppler Equation PW Doppler Color Doppler Power Doppler 3 4 Who was Mr Doppler & What is Doppler Technology? A class of technologies named after Christian Doppler, a famous Austrian physicist from the late 1800 s. He proved that any movement between a light emitting source and a receiver, will produce a detectable frequency shift. In his case, he was observing the colored lights of double star systems. His theory is the basis for all ultrasoundbased Doppler blood detection technologies. 5 Mr. Doppler was talking about light frequencies White light refracted through a prism (or atmosphere) will separate into colors In our visual world, red = the fastest frequency shifts and violet = the slowest Visible light Incidence and refraction angles (i & r) bend white light into constituent colors (through n2). Violet rays result from the highest deviations (freq. Shifts) and red from the least. In CDI, Red = positive shift, blue = negative CDI display 6
2 Why we hate Doppler physics!! The received frequency response is directly proportional to the transmitted frequency, (fo), and the transmissionreceive angle. Therefore the unconverted frequency shift estimations, f, will degrade as a function of the cosine of the angle from which the transmitted beam is displaced from the line of sight of the insonated media being tracked, (cos 0). When understanding and therefore compensating for angle limitations and the overunderestimation of frequency shifts due to the intrinsic spectral broadening inherent in multi-element electronic array probes, Spectral Doppler s Fast Fourier Transform analysis can provide increased accuracy in frequency shift estimation due to the employment of non-averaging algorithms versus the commonly accepted autocorrelation techniques used in color Doppler which preclude velocity estimation and limit CDI to a display of non-angle corrected, average frequency shifted data. 7 It s just that simple,,,,,, not! The transition between laminar and turbulent flow occurs when a critical number is exceeded. What is this # is called? Reynolds Number Laminar Flow occurs at low Reynolds numbers, where viscous forces are dominant, and is characterized by smooth, constant fluid motion, while Turbulent Flow, on the other hand, occurs at high Reynolds numbers and is dominated by inertial forces, producing random eddies, vortices and other flow fluctuations. 8 Doppler Overview Doppler Overview (Parabolic Flow) 9 Turbulent Flow will revert to Laminar Flow distal to a 10 flow reducing lesion Rouleaux flow, transverse Rouleaux flow, Sagittal Red Blood Cells (RBCs) clump together in coin-like stacks (or rouleaux formation) wherever space allows them to. They do this, actually, to reduce the viscosity of the blood and for the protection that the red cells receive by traveling in these train-like processions. So RBCs in the main arteries and veins travel, naturally, in these "rouleaux" stacks
3 Normal flow to reduced flow to high speed flow, then losing energy and returning to laminer flow Rayleigh scattering enhanced with harmonics Rayleigh scattering enhanced with harmonics Doppler Overview Flow Patterns Laminar flow: Blood traveling in parallel layers in a vessel. Central layers travel at an increased velocity relative to outer layers. Plug flow: A type of laminar flow where multiple layers of blood travel at similar speeds. Disturbed flow: Flow in which the layers of flow have been disrupted. 15 Turbulent flow: Chaotic flow in multiple directions. Arterial Vascular Hemodynamics Owen CA, Roberts M Journal of Diagnostic Medical Sonography : 16 Doppler Overview Doppler Overview Concept 1 Concept 2 Concept
4 To Perform Doppler, We Need An Angle degrees 90 degrees 0 degrees We use the Doppler theory and apply it to sound waves to detect frequency shifts ( speed ) and direction f 2 fo V C 0 The big problem with Doppler is operator vs. interpreter variability Doppler Equation Doppler Equation Cowboy s Understood Doppler Doppler Equation!!! I love hearing the lonesome wail of the train whistle as the magnitude of the frequency of the wave changes due to the Doppler effect ya ll! Lynx Lake, Prescott, Arizona 23 24
5 We witness visually and hear Doppler audibly, the Doppler Equation effect in our to day lives. I think Mr. Doppler would be proud that his theory of detecting the motion of heavenly bodies, has made stars of those whose job it is to detect the motion of blood cells within our terrestrial bodies!!! D. Milburn 25 What are the Doppler-based Blood Detection Technologies? Pulsed Wave Doppler (PW) High Pulsed Repetition Freq. Doppler (HPRF) Continuous Wave Doppler (CW) Color Doppler Imaging (CDI) Power Doppler Imaging (PDI) 26 Color Doppler Pulsed Wave Doppler PW Types of Doppler Instrumentation! High PRF Doppler HPRF Continuous Wave Doppler X-mit Receive Power Doppler CW Continuous Wave Doppler (CW) CW Doppler uses two crystals or two subsets of crystals to transmit and receive. It does not use a discrete sample volume and is not restricted by the low PRFs employed in Pulsed Doppler. CW Doppler will not alias. CW Doppler is used to detect high velocities. = CW Sample area Transmit No Aliasing Detects high velocities Receive No range location Pulsed Wave (PW) PW Doppler uses a discrete sample volume (SV) to sample blood at precise locations. The Doppler frequency is pulsed thousands of times per second, (KHz). PRF determines the speed of blood that may be detected unambiguously. f 2 f0 V Sample volume High Pulsed Repetition Frequency (HPRF) HPRF employs multiple sample volumes placed at intervals along the Doppler cursor. These sites can be sampled at a faster rate than PW Doppler and thus higher PRFs can be used. Higher PRFs will detect and correctly display higher blood flows. SV 1 SV 2 f 2 f0 V SV 3 SV 1 SV 2 Crystal 2 Accurate range location Excellent spectral quality Aliasing 29 Multiple sample volumes Equidistant spacing Good range location Adequate spectral quality Increased Nyquist limits Multiple sample volumes 30
6 It s all based on the Doppler Equation The Doppler Equation is not 2 B feared! The Doppler equation? I have to know the Doppler equation??? f = 2 f0 v cosø c Important: Frequency shift size is linear and proportional to Doppler frequency Scattering (Rayleigh) is proportional to the 4 th power of the Doppler frequency What is Spectral Doppler? PW Doppler Controls The ultrasonic evaluation of blood flow. Pulsed Wave (PW), High Pulse Repetition Frequency (HPRF) and Continuous Wave Doppler (CW) are all Spectral Doppler technologies. The differences between these technologies must be understood PW Doppler Controls SV Length The sample volume must be adjusted for every patient and every clinical situation. Sample volume length is adjusted to optimize data acquisition. Small sample volumes are used to acquire data from discrete areas and for peak velocity detection. Large sample volumes are used to analyze complex vasculature (tumor flow) and to calculate flow volumes. Use caution on HPRF (multiple sample volumes) < SV = < signalnoise > SV = > signalnoise 35 36
7 X The Sample Volume (SV) Length must be adjusted for each vessel and pathology! Z Intrinsic Spectral Broadening What happened to my spectral window? The waveform is filled in,,, is that not signal? Y Think about spatial position 37 Facts: In the beginning, (1970 s80 s),most clinical vascular spectral Doppler was performed using single crystal mechanical transducer technology. Single crystal technology had the advantage of a large, round transmit and receive surface with excellent focus capabilities. Spectral broadening in the Doppler waveforms acquired with single crystal transducers primarily was related to hemodynamics and in normal blood flow conditions, was minimal as Doppler data was received across one large collection surface However,,,,,, slow flow detection could be compromised if the blood flow was slower than the mechanical crystal could be moved through it s field of travel Facts: Phased array, electronic linear and curved array transducer technologies have the advantage of no moving parts and were therefore inherently superior in detecting slow blood flow with Doppler. However,,, multiple elements in the transducer array transmit (delay lines) & receive the Doppler reflections at different points in time in relation to their position in the receive aperture, which is dynamic (changes size). Multiple receive angles result in many slightly different Doppler frequencies being processed and included in the waveforms This creates Intrinsic spectral broadening created by the transducer technology,,, not abnormal blood flow 39 Graphic courtesy of Rick Duncan 40 Transducer elements = Aperture = Doppler cursor = Receive elements Single crystal mechanical probe Vessel = receive to one collecting surface Spectral broadening 41 42
8 So; Spectral broadening should not be relied on as an absolute diagnostic criteria when generating Doppler results with electronic array probes. That being said; If I am not so concerned about spectral broadening I can now use larger sample volumes to increase sensitivity to slowlow flow! Scale (PRF - Pulsed Repetition Frequency) Spectral Doppler aliasing is controlled using Scale (PRF). High velocity flow requires a higher Scale (PRF) Low velocity flow requires a lower Scale (PRF) Aliasing occurs when the velocity of the blood exceeds ½ of the PRF (Nyquist Limit) What is the most common Doppler artifact that Scale (PRF) will reduceeliminate? Aliasing Which way is the wheel spinning? 45 Clockwise or Counter Clockwise? 46 Baseline This control changes the spectrum baseline to accommodate higher velocity blood flow. It can be used to reduce aliasing by displaying a greater range of forward flow with respect to reverse flow and vice versa. The baseline displays as a solid line running across the spectrum. The baseline is raised and lowered in equal increments, depending on the current Doppler scale factor. Notice the baseline color? What color is a CDI baseline? 47 48
9 Invert Inverts the directional display of detected flow without affecting the baseline position. Allows negative flows to be displayed in a positive direction and viceversa. Can be used to reverse Doppler waveform presentation rather than reversing probe orientation. Useful when examining vessels from different probe positions as it keeps the waveform direction uniform. Spectral Doppler Gain Doppler Gain amplifies the overall strength of the received Doppler signal. Gain must be used with caution as it can increase or decrease spectral broadening. Spectral broadening is used as a diagnostic criteria. Too much Gain = noise Too little Gain = not enough signal Spectral Doppler Gain Receive Gain only Turn it up!!! Too much Gain = noise Too little Gain = not enough signal M&A caution! Wall Filter These are used to filter out tissue signals, motion artifacts from breathing and other patient movement. Filters will be automatically adjusted with the Scale (PRF). The higher the Scale the higher the Wall Filter and vice versa. However, if there is still motion artifact the Wall Filter can be adjusted manually. Wall Filters are electronic eraser s You will see no information within the bounds of the green bar. Filters remove noise and low flow signals. Angle Correct Optimizes the accuracy of the flow velocity. This is especially useful in vascular applications where you need to measure velocity. (converts Hz. to velocity) For optimum velocity measurements, the angle of incidence should be 60 degrees or less for vascular applications. ICAVEL requires 60 degrees or less 53 54
10 Quick Angle Quickly adjusts the angle by 60 degrees. Toggles through 60, 0 and +60 degrees. Angle Steer You can Steer the Doppler Cursor left, right or center to get more information without moving the probe. This Steer function only applies to linear probes. Linear probes can be steered left (20 degrees), center, or right (20 degrees) Fine Steer Fine Steer steers the Doppler cursor leftright between 0 and 30 degrees in 2 degree increments. Provides a Doppler cursor angle suitable for linear probe orientation. Beneficial in Vascular applications. 60 degree angle from parallel vessel 20 degrees > 20 degrees (Fine Angle Steer) Newton s 3 rd law 57 Beam steer with heeltoe maneuver 58 Sweep Speed Changes the speed at which the timeline is swept. You can speed up or slow down the timeline to see more or fewer occurrences over time. The slower the sweep speed, the closer together the spectral waveforms. The faster the sweep speed the further apart the spectral waveforms. Colorize Enhances the eye s discrimination and ability to resolve changes in shades of gray. The eye can perceive slight changes in color easier than in black and white. Useful when the spectral trace is difficult to see adequately
11 Auto Calcs Allows the selection of the automatic calculations to be either Off, Live or Frozen. Once you have a Spectral waveform you are able to turn this function on and off. Simultaneous (Duplex vs. Triplex) Simultaneous This controls if the system will use Duplex or Triplex Duplex is usually recommended as you get a cleaner Doppler trace. This results from the machine only acquiring Doppler information instead of B-mode, Color and Doppler simultaneously. Caution: Triplex can limit Scale (PRF) and high speed flow detection Spectral Doppler Frequency Doppler frequency will affect the waveform size. This can be used to increase small signals (increase sensitivity) and reduce the size of large signals. If you are scanning superficial structures, use a higher frequency as this will provide better signal to noise ratio. If scanning deeper structures and you need more penetration use a lower frequency There are lot s of automated features on new Doppler systems To this I say Midway Review Doppler Concepts? Doppler Equation? Pulsed Wave Doppler? Spectral Doppler GE Controls E9 Doppler Optimization?
12 Let s talk about Color Doppler What is Color Doppler Imaging? Color is a Pulsed Doppler technology that provides data on: - Existence of flow - Direction of flow - Flow turbulence - Flow quality Why do we need Spectral Doppler? It is the only method available to obtain reproducible, noninvasive, quantitative parameters on blood flow Why do we use Color Doppler? Color Doppler is used as a road map for accurate spectral Doppler sample volume placement. A flow disturbance can be located and Qualified with CDI and then Quantified with spectral Doppler Color Doppler saves time over blind spectral mapping. Color Doppler Processing (Autocorrelation) Autocorrelation processing looks at Doppler data from multiple sample volumes. All frequency shifts from all sample volumes are analyzed for direction, strength and mean values. All data from all sample volumes are then averaged together into a display of mean frequency shifts
13 Color Doppler menu Color Doppler menu What are the standard CDI displays? BART: Blue Away, Red Towards RABT: Red Away, Blue Towards Size - [Region of Interest (ROI)] This determines the region of color displayed on the image The larger (lateral width) the Color Box Size, the slower the frame rate To increase frame rate decrease the width of the Color Box Size. The height (axial length) of the Color Box Size does not have as much affect on the frame rate. Don t forget you can also change the B-mode Sector Width to increase frame rates!!! It s best to keep the Color Box Sized to the area of interest Color Scan Area (ROI) Color Invert Large ROI Small ROI This control allows you to reverse the displayed color of the blood flow towards and away from the transducer This may be inverted in real time or freeze frame Usually red is used for flow towards the transducer and blue for flow away from the transducer 77 78
14 How We See Color: Color blindness affects 1 in 10 men and 1 in 100 women! Disorder is passed from mother to son on the X chromosome The most common is the inability to distinguish red from green This is why stoplights are always in the same order This is why men are not allowed color to pick out their own clothes Color Maps There are several color maps available. These are velocity maps showing blue away red towards, variance maps which add green as a measure of turbulence, additive maps which add more whites to the color, and finally the symmetric maps which show positive and negative flow using the same color (useful for 3D). There are no correct maps. Use whichever map presents the most useable frequency data to your eye. Select Maps and scroll to change the map Angle Steer Allows steering (slanting) of the color box region of interest (ROI) on linear transducers This is used to improve color sensitivity without moving the probe The color ROI may be steered left, right or straight depending on the angle of the vessel Which angle steer provides maximum sensitivity? Caution!!! M&AAngle Gain artifact!! Angle Steer Angle Steer Color Gain Color Gain is used to amplify the overall strength of echo s displayed in the color ROI. To correctly set the Color Gain turn the Gain up until there is bleeding of the color. Then reduce the Color Gain until the noise disappears. This control is on the Color Flow (CF) button. Spectral Doppler Gain works the same way
15 Color Wall Filters Color Wall Filters (Thump, Wall, High Pass filter whatever) These are used to filter out motion artifacts from breathing and other patient movement. This will be automatically adjusted with the Scale (PRF). The higher the scale the higher the Wall Filter and vice versa. However if there is still motion artifact the Wall Filter can be adjusted manually. There are separate Wall Filters for Doppler and Color Doppler. Wall filters are electronic erasers NOTE!! Wall Filters can increase or decrease automatically with vertical ROI movement High Pass Filter Facts amplitud e Tissue signals are db greater in amplitude than Blood signals Tissue = High AmplitudeLow Velocity signals Blood = High VelocityLow Amplitude signals High Pass Filters remove tissue signals and preserve blood signals < HPF # = > display of low freq. blood and > tissue noise > HPF # = < display of low freq. blood signals *** Settings too high will eliminate blood signals!!! HPF # = Tissue 87 velocity Blood 88 Why do we see color here? Color Threshold There are sample volumes in both area s Color Threshold Assigns the gray scale level at which color information stops The higher the Color Threshold, the more color that will be displayed and vice versa This is generally set at about 70-90%. If there is still bleeding then you can reduce this value. **Key point some vessels are so small that they cannot be visualized. If the scanner does not see the black from blood, it will not write color. Threshold makes the system write color to the detected Doppler shift, even if the vessel is not seen. And we see no color here? 89 90
16 D:\Documents and Settings\ \Desk top\clinical Images\Lori's images\rra CLR.jpg Color Threshold Color Sample Volume Allows you to improve the color sensitivity, axial resolution and penetration of color Doppler Sample Volume size may be set at zero (0) for the highest spatial resolution, and set at four (4) for maximum penetration. Sample volume size can affect axial bleeding! > PS < PS Color Doppler Frequency Color Doppler Frequency Color Frequency may be adjusted on some transducers. If you are scanning superficial structures use a higher Frequency as this will give you better signal to noise ratio. If scanning deeper structures and you need more penetration use a lower Frequency CDI Frequency will lower automatically with ROI depth increase Color ScalePRFVelocity PRFVelocityScale (Doppler) Spectral Doppler aliasing is controlled with the Scale (PRF) High speed flow requires a higher Scale (PRF) Low speed flow requires lower Scale (PRF) Aliasing occurs when the speed of the blood exceeds ½ of the PRF Scale & Baseline control Aliasing 95 96
17 Accumulation Accumulation Color & Spectral Doppler Baseline This control changes either the Spectral Baseline or the Color Bar to accommodate higher velocity blood. It can be used to reduce aliasing by displaying a greater range of forward flow with respect to reverse flow and vice versa. This control works the same in Spectral and Color Doppler Baseline is usually set at the midpoint between forward and reverse flow. Color Baseline CDI Baseline shift SDI Baseline shift DopplerColor Baseline This control changes either the spectrum baseline or the color bar to accommodate higher velocity blood flow. It can be used to reduce aliasing by displaying a greater range of forward flow with respect to reverse flow and vice versa. This control works the same in spectral and color Doppler Baseline is usually set at the midpoint between forward and reverse flow. CDI baseline can be used to eliminate aliasing wo eliminating so flow signals CDI Baseline shift Aliasingflow reversal??? SDI Baseline shift
18 Power Doppler Power Doppler Imaging Power Doppler Imaging (PDI) is a mapping technique that looks at the strength (amplitude) of the Doppler signal instead of the frequency shift. PDI appears to be more sensitive than Color Flow when switching between the two modes. The reason is that PDI parameters are set to detect lower velocities Why Use Power Doppler? Power Doppler is used whenever extreme sensitivity to low or slow flow is needed Power Doppler is also used to eliminate aliasing and angle of insonation artifacts Power Doppler is superior to color in outlining vessel walls and depicting residual lumens in high grade stenotic lesions Color Doppler vs. Power Doppler A 3 db increase in signal strength! Weak 90 degrees are avg d to 0 i.e.: black C D I P D I Weak 90 degrees are summed to a positive number Power Doppler Transmit Frequency As in color Doppler, changing the transmit Doppler frequency in PDI can have a dramatic effect on the received signal. Flash Suppression CAUTION! Use no flash suppression with Fremitus!!!
19 To Increase Color Sensitivity Adjust Size of ROI and Angle Steer Adjust Color Focal Position Increase Color Gain and Decrease the Scale (PRF) Decrease Wall FilterIncrease Threshold Adjust Color Sample VolumeAdd Accumulation DecreaseIncrease Color Frequency Change Transducers Mr. Sensitivity Follow the Doppler Recipe s 4 Success! To Reduce Color Bleeding Decrease Color Gain Increase Color Scale (PRF) Decrease the Color Threshold Decrease Sample Volume Size Increase Color Frequency To Reduce Flash Artifact Increase the Scale (PRF) which will in turn increase the Wall Filter Increase Flash Suppression (menu page 2) Be careful when using the Fremitus technique To Eliminate Aliasing Increase the Color Scale (PRF) Increase Spectral Doppler Scale (PRF) Decrease the Color Baseline Shift the Spectral Doppler Baseline To Increase Frame RateFlow Dynamics Decrease the Color Scan Area Size (ROI) Exit Color Flow and Decrease B-mode Sector Width Decrease Depth Increase Color Scale (PRF) Change Line Density Decrease Frame Averaging High PRF may not eliminate aliasing in a 0 degree vessel Holy Cow I ve been cloned!!
20 Remember, Enough GAS Solves All Problems!! OK, time for the test! G = GAIN A = ANGLE S = STEERING Beam Steer is important for image quality and Doppler acquisition! Both images have acceptable Doppler data. Which image is steered correctly and why? B-mode image steered for IQ. CDI ROI & PW Cursor angled correctly Angle corrected to vessel wall for velocity conversion Results? QUALITY DIAGNOSTIC DATA! The optimal B- mode image interface angle is 90 degrees Optimal Doppler interface angle is 0 degrees Describe the Alternating colors Alternating colors are caused by: Look at the angle of insonation and the target interface angles 1. Shadowing degree Doppler angle 3. Refraction 4. Diffraction 5. None of the above Note the color bar orientation
21 Sound Diffraction Alternating colors are caused by 1.Shadowing 2.90-degree Doppler angle 3.Refraction 4.Diffraction 5.None of the above Why the difference in low flow sensitivity Angle and aperture Color Optimization Which CDI control has been adjusted? You have presented these two images to the interpreting physician. Heshe is confused by the displayed colors. What control change could cause the difference in the below images?
22 D:\Documents and Settings\ \Desk top\clinical Images\Lori's images\rra CLR.jpg Color bar directionorientation combined with variation in cardiac cycle What are the standard CDI displays? BART: Blue Away, Red Towards RABT: Red Away, Blue Towards Color Doppler Controls: Would Gain or scale adjustment create the difference in these displays? Color Gain Color Gain is used to amplify the overall strength of echo s displayed in the color ROI. Gain does not increase the insonation power. To correctly set the Color Gain, increase the Gain until there is bleeding of the color. Next, reduce the Color Gain until the noise disappears. This control is on the Color Flow (CF) button. Spectral Doppler Gain works the same way Color ScalePRFVelocity Review 131 We reviewed basic hemodynamics We discussed all Doppler types We learned how the Doppler Equation works and how to apply it to our advantage We showed all relevant Doppler menu s and controls and provided clinical examples I gave you our Doppler Recipes for 132
23 A thousand thank-you s for your kind attention! The E a
Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19
Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and
Ny teknologi: Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010
1 Ny teknologi: Ultralyd måler m blodstrøm Fagdagene ved St. Olavs Hospital Lasse Løvstakken Dept. Circulation and Medical Imaging 11.06.2010 2 Conventional imaging methods of blood flow using ultrasound
Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect
Objectives: PS-7.1 Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect Illustrate ways that the energy of waves is transferred by interaction with
Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality
Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL
Module 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
Scanning Acoustic Microscopy Training
Scanning Acoustic Microscopy Training This presentation and images are copyrighted by Sonix, Inc. They may not be copied, reproduced, modified, published, uploaded, posted, transmitted, or distributed
Ultrasonic Wave Propagation Review
Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic
VFI Technology to Change the Way You Work
Technology to Change the Way You Work Vascular Ultrasound Made Easier Vector Flow Imaging VFI VFI is a ground-breaking technology that can revolutionize the workflow for many Doppler ultrasound applications.
Lab 1: The Digital Oscilloscope
PHYSICS 220 Physical Electronics Lab 1: The Digital Oscilloscope Object: To become familiar with the oscilloscope, a ubiquitous instrument for observing and measuring electronic signals. Apparatus: Tektronix
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
Jitter Measurements in Serial Data Signals
Jitter Measurements in Serial Data Signals Michael Schnecker, Product Manager LeCroy Corporation Introduction The increasing speed of serial data transmission systems places greater importance on measuring
Copyright 2008 Pearson Education, Inc., publishing as Pearson Addison-Wesley.
Chapter 20. Traveling Waves You may not realize it, but you are surrounded by waves. The waviness of a water wave is readily apparent, from the ripples on a pond to ocean waves large enough to surf. It
Acoustic Terms, Definitions and General Information
Acoustic Terms, Definitions and General Information Authored by: Daniel Ziobroski Acoustic Engineer Environmental and Acoustic Engineering GE Energy Charles Powers Program Manager Environmental and Acoustic
Contents. HD11 XE Quick Guide 4535 612 62661
Contents Using the HD11 XE Help................................................3 Using the HD11 XE Ultrasound System....................................4 Imaging Modes........................................................7
Encoded Phased Array Bridge Pin Inspection
Encoded Phased Array Bridge Pin Inspection James S. Doyle Baker Testing Services, Inc. 22 Reservoir Park Dr. Rockland, MA 02370 (781) 871-4458; fax (781) 871-0123; e-mail [email protected] Product
Waves Sound and Light
Waves Sound and Light r2 c:\files\courses\1710\spr12\wavetrans.doc Ron Robertson The Nature of Waves Waves are a type of energy transmission that results from a periodic disturbance (vibration). They are
Chapter 17: Light and Image Formation
Chapter 17: Light and Image Formation 1. When light enters a medium with a higher index of refraction it is A. absorbed. B. bent away from the normal. C. bent towards from the normal. D. continues in the
Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers
Flow and Flow rate. Laminar and Turbulent flow Laminar flow: smooth, orderly and regular Mechanical sensors have inertia, which can integrate out small variations due to turbulence Turbulent flow: chaotic
FTIR Instrumentation
FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation
Various Technics of Liquids and Solids Level Measurements. (Part 3)
(Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,
Samsung Ultrasound H60
Samsung Medison is a global leading medical devices company. Founded in 1985, the company now sells cutting-edge medical devices including diagnostic ultrasound, digital X-ray and blood analyzer around
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,
Numerical Parameters Analysis of Boonton 4540 Peak Power Meter
Application Note Numerical Parameters Analysis of Boonton 4540 Peak Power Meter Mazumder Alam Product Marketing Manager, Boonton Electronics Introduction The Boonton 4540 series RF peak power meters consisting
PRINCIPLES OF DOPPLER ECHOCARDIOGRAPHY AND THE DOPPLER EXAMINATION #1
PRINCIPLES OF DOPPLER ECHOCARDIOGRAPHY AND THE DOPPLER EXAMINATION #1 Joseph A. Kisslo, MD David B. Adams, RDCS INTRODUCTION Doppler echocardiography is a method for detecting the direction and velocity
1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.
CHAPTER 3: OSCILLOSCOPE AND SIGNAL GENERATOR 3.1 Introduction to oscilloscope 1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. 2. The graph show signal change
Principles of Medical Ultrasound. Pai-Chi Li Department of Electrical Engineering National Taiwan University
Principles of Medical Ultrasound Pai-Chi Li Department of Electrical Engineering National Taiwan University What is Medical Ultrasound? Prevention: actions taken to avoid diseases. Diagnosis: the process
1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002
05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical
BIOMEDICAL ULTRASOUND
BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic
physics 1/12/2016 Chapter 20 Lecture Chapter 20 Traveling Waves
Chapter 20 Lecture physics FOR SCIENTISTS AND ENGINEERS a strategic approach THIRD EDITION randall d. knight Chapter 20 Traveling Waves Chapter Goal: To learn the basic properties of traveling waves. Slide
WAVELENGTH OF LIGHT - DIFFRACTION GRATING
PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant
Crystal Optics of Visible Light
Crystal Optics of Visible Light This can be a very helpful aspect of minerals in understanding the petrographic history of a rock. The manner by which light is transferred through a mineral is a means
Electrical Resonance
Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.
After a wave passes through a medium, how does the position of that medium compare to its original position?
Light Waves Test Question Bank Standard/Advanced Name: Question 1 (1 point) The electromagnetic waves with the highest frequencies are called A. radio waves. B. gamma rays. C. X-rays. D. visible light.
GE Healthcare. Vivid 7 Dimension Cardiovascular Ultrasound System
GE Healthcare Vivid 7 Dimension Cardiovascular Ultrasound System New dime Introducing multi-dimensional and 4D imaging. The ability to incorporate multi-dimensional and 4D imaging into your clinical routine
Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT)
Page 1 Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT) ECC RECOMMENDATION (06)01 Bandwidth measurements using FFT techniques
MestRe-C User Guide Megan Bragg 04/14/05
MestRe-C User Guide Megan Bragg 04/14/05 Some general useful features: 1. You can add Text, Titles, Temperatures, or peak labels (in any font/size/color) to the spectrum. 2. MestRe-C allows you to save
Characterizing Digital Cameras with the Photon Transfer Curve
Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently
E190Q Lecture 5 Autonomous Robot Navigation
E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator
Monitoring of Cerebral Blood Flow. Transcranial Doppler Laser Doppler Flowmetry Thermal dilution method (Hemedex)
Monitoring of Cerebral Blood Flow Transcranial Doppler Laser Doppler Flowmetry Thermal dilution method (Hemedex) Ultrasound in Tissue Some Facts: blood cell tissue probe ultrasound travels at a constant
Answer the following questions during or after your study of Wave Properties. 4. How are refraction and the speed of wave in different media related?
Wave Properties Student Worksheet Answer the following questions during or after your study of Wave Properties. 1. A person standing 385 m from a cliff claps her hands loudly, only to hear the sound return
12.1 What is Refraction pg. 515. Light travels in straight lines through air. What happens to light when it travels from one material into another?
12.1 What is Refraction pg. 515 Light travels in straight lines through air. What happens to light when it travels from one material into another? Bending Light The light traveling from an object in water
Ultrasound. Sound waves
Ultrasound Basic Idea Send waves into body which are reflected at the interfaces between tissue Return time of the waves tells us of the depth of the reflecting surface History First practical application,
Experiment 1: SOUND. The equation used to describe a simple sinusoidal function that propagates in space is given by Y = A o sin(k(x v t))
Experiment 1: SOUND Introduction Sound is classified under the topic of mechanical waves. A mechanical wave is a term which refers to a displacement of elements in a medium from their equilibrium state,
Scanners and How to Use Them
Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types
Using the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors
Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors Rudolph J. Guttosch Foveon, Inc. Santa Clara, CA Abstract The reproduction
Robot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
Reflection and Refraction
Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,
Zeiss 780 Training Notes
Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn
DIGITAL-TO-ANALOGUE AND ANALOGUE-TO-DIGITAL CONVERSION
DIGITAL-TO-ANALOGUE AND ANALOGUE-TO-DIGITAL CONVERSION Introduction The outputs from sensors and communications receivers are analogue signals that have continuously varying amplitudes. In many systems
SR2000 FREQUENCY MONITOR
SR2000 FREQUENCY MONITOR THE FFT SEARCH FUNCTION IN DETAILS FFT Search is a signal search using FFT (Fast Fourier Transform) technology. The FFT search function first appeared with the SR2000 Frequency
Section 3. Sensor to ADC Design Example
Section 3 Sensor to ADC Design Example 3-1 This section describes the design of a sensor to ADC system. The sensor measures temperature, and the measurement is interfaced into an ADC selected by the systems
K2 CW Filter Alignment Procedures Using Spectrogram 1 ver. 5 01/17/2002
K2 CW Filter Alignment Procedures Using Spectrogram 1 ver. 5 01/17/2002 It will be assumed that you have already performed the RX alignment procedures in the K2 manual, that you have already selected the
TDS5000B, TDS6000B, TDS/CSA7000B Series Acquisition Modes
TDS5000B, TDS6000B, TDS/CSA7000B Series Acquisition Modes Tektronix oscilloscopes provide several different acquisition modes. While this gives the user great flexibility and choice, each mode is optimized
Graphing Motion. Every Picture Tells A Story
Graphing Motion Every Picture Tells A Story Read and interpret motion graphs Construct and draw motion graphs Determine speed, velocity and accleration from motion graphs If you make a graph by hand it
ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1
WHAT IS AN FFT SPECTRUM ANALYZER? ANALYZER BASICS The SR760 FFT Spectrum Analyzer takes a time varying input signal, like you would see on an oscilloscope trace, and computes its frequency spectrum. Fourier's
ART 269 3D Animation Fundamental Animation Principles and Procedures in Cinema 4D
ART 269 3D Animation Fundamental Animation Principles and Procedures in Cinema 4D Components Tracks An animation track is a recording of a particular type of animation; for example, rotation. Some tracks
Duplication Images in Vascular Sonography
Article Duplication Images in Vascular Sonography Jonathan M. Rubin, MD, PhD, Jing Gao, MD, Keith Hetel, MD, Robert Min, MD Objective. The purpose of this study was to determine the characteristics and
AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz
AN1200.04 Application Note: FCC Regulations for ISM Band Devices: Copyright Semtech 2006 1 of 15 www.semtech.com 1 Table of Contents 1 Table of Contents...2 1.1 Index of Figures...2 1.2 Index of Tables...2
PeakVue Analysis for Antifriction Bearing Fault Detection
August 2011 PeakVue Analysis for Antifriction Bearing Fault Detection Peak values (PeakVue) are observed over sequential discrete time intervals, captured, and analyzed. The analyses are the (a) peak values
Light and its effects
Light and its effects Light and the speed of light Shadows Shadow films Pinhole camera (1) Pinhole camera (2) Reflection of light Image in a plane mirror An image in a plane mirror is: (i) the same size
PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany
ABSTRACT: PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany Today, in-line inspection tools are used routinely
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide
Experiment 5. Lasers and laser mode structure
Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,
Transmission Line and Back Loaded Horn Physics
Introduction By Martin J. King, 3/29/3 Copyright 23 by Martin J. King. All Rights Reserved. In order to differentiate between a transmission line and a back loaded horn, it is really important to understand
Laboratory Guide. Anatomy and Physiology
Laboratory Guide Anatomy and Physiology TBME04, Fall 2010 Name: Passed: Last updated 2010-08-13 Department of Biomedical Engineering Linköpings Universitet Introduction This laboratory session is intended
A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)
chapter 2 Pulsed Ultrasound In diagnostic ultrasound imaging, short bursts, or pulses, of acoustic energy are used to create anatomic images. Continuous wave sound cannot create anatomic images. Analogy
The Calculation of G rms
The Calculation of G rms QualMark Corp. Neill Doertenbach The metric of G rms is typically used to specify and compare the energy in repetitive shock vibration systems. However, the method of arriving
Impedance Matching of Filters with the MSA Sam Wetterlin 2/11/11
Impedance Matching of Filters with the MSA Sam Wetterlin 2/11/11 Introduction The purpose of this document is to illustrate the process for impedance matching of filters using the MSA software. For example,
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture
ε: Voltage output of Signal Generator (also called the Source voltage or Applied
Experiment #10: LR & RC Circuits Frequency Response EQUIPMENT NEEDED Science Workshop Interface Power Amplifier (2) Voltage Sensor graph paper (optional) (3) Patch Cords Decade resistor, capacitor, and
Atomic Force Microscope
Atomic Force Microscope (Veeco Nanoman) User Manual Basic Operation 4 th Edition Aug 2012 NR System Startup If the system is currently ON To start the NanoScope software, double-click the NanoScope startup
Little LFO. Little LFO. User Manual. by Little IO Co.
1 Little LFO User Manual Little LFO by Little IO Co. 2 Contents Overview Oscillator Status Switch Status Light Oscillator Label Volume and Envelope Volume Envelope Attack (ATT) Decay (DEC) Sustain (SUS)
The promise of ultrasonic phased arrays and the role of modeling in specifying systems
1 modeling in specifying systems ABSTRACT Guillaume Neau and Deborah Hopkins This article illustrates the advantages of phased-array systems and the value of modeling through several examples taken from
Twelve. Figure 12.1: 3D Curved MPR Viewer Window
Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from
Lecture 14. Point Spread Function (PSF)
Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect
GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time
Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs
Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications
Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Janet E. Semmens Sonoscan, Inc. 2149 E. Pratt Boulevard Elk Grove Village, IL 60007 USA Phone: (847)
Philips CX50 CompactXtreme System. A New Class of Compact Echocardiography
Philips CX50 CompactXtreme System A New Class of Compact Echocardiography Insights from compact ultrasound users Diagnostic confidence is paramount Ultrasound labs are busy and workflow efficiency is crucial
Acousto-optic modulator
1 of 3 Acousto-optic modulator F An acousto-optic modulator (AOM), also called a Bragg cell, uses the acousto-optic effect to diffract and shift the frequency of light using sound waves (usually at radio-frequency).
SAMSUNG ULTRASOUND RS80A
Samsung Medison is a global leading medical device company. Founded in 1985, the company sells cutting-edge diagnostic ultrasound devices around the world in various medical fields. The company has attracted
Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide)
Waves: Recording Sound Waves and Sound Wave Interference (Teacher s Guide) OVERVIEW Students will measure a sound wave by placing the Ward s DataHub microphone near one tuning fork A440 (f=440hz). Then
Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated testing of square and round bars
11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated
STATISTICAL ANALYSIS OF ULTRASOUND ECHO FOR SKIN LESIONS CLASSIFICATION HANNA PIOTRZKOWSKA, JERZY LITNIEWSKI, ELŻBIETA SZYMAŃSKA *, ANDRZEJ NOWICKI
STATISTICAL ANALYSIS OF ULTRASOUND ECHO FOR SKIN LESIONS CLASSIFICATION HANNA PIOTRZKOWSKA, JERZY LITNIEWSKI, ELŻBIETA SZYMAŃSKA *, ANDRZEJ NOWICKI Institute of Fundamental Technological Research, Department
1) The time for one cycle of a periodic process is called the A) wavelength. B) period. C) frequency. D) amplitude.
practice wave test.. Name Use the text to make use of any equations you might need (e.g., to determine the velocity of waves in a given material) MULTIPLE CHOICE. Choose the one alternative that best completes
Introduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
A comparison of radio direction-finding technologies. Paul Denisowski, Applications Engineer Rohde & Schwarz
A comparison of radio direction-finding technologies Paul Denisowski, Applications Engineer Rohde & Schwarz Topics General introduction to radiolocation Manual DF techniques Doppler DF Time difference
Describing Sound Waves. Period. Frequency. Parameters used to completely characterize a sound wave. Chapter 3. Period Frequency Amplitude Power
Parameters used to completely characterize a sound wave Describing Sound Waves Chapter 3 Period Frequency Amplitude Power Intensity Speed Wave Length Period Defined as the time it take one wave vibrate
FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER
2014 Amplifier - 1 FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER The objectives of this experiment are: To understand the concept of HI-FI audio equipment To generate a frequency response curve for an audio
CBS RECORDS PROFESSIONAL SERIES CBS RECORDS CD-1 STANDARD TEST DISC
CBS RECORDS PROFESSIONAL SERIES CBS RECORDS CD-1 STANDARD TEST DISC 1. INTRODUCTION The CBS Records CD-1 Test Disc is a highly accurate signal source specifically designed for those interested in making
Ultrasound Technologies. PD1v. vascular pocket doppler. Operating Instructions
Ultrasound Technologies PD1v vascular pocket doppler Operating Instructions Before using your Pocket Doppler for the first time, please read these operating instructions carefully. 1 INTRODUCTION The UltraTec
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy Application Note RF & Microwave Spectrum Analyzers Table of Contents 3 3 4 4 5 7 8 8 13 13 14 16 16 Introduction Absolute versus relative
High-quality/High-resolution Digital Ultrasound Diagnostic Scanner
Hitachi Review Vol. 52 (2003), No. 4 207 High-quality/High-resolution Digital Ultrasound Diagnostic Scanner Tsuyoshi Otake Toshihiko Kawano Takashi Sugiyama Tsuyoshi Mitake Shinichiro Umemura, Ph. D. OVERVIEW:
application note Directional Microphone Applications Introduction Directional Hearing Aids
APPLICATION NOTE AN-4 Directional Microphone Applications Introduction The inability to understand speech in noisy environments is a significant problem for hearing impaired individuals. An omnidirectional
Signal to Noise Instrumental Excel Assignment
Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a
Introduction to acoustic imaging
Introduction to acoustic imaging Contents 1 Propagation of acoustic waves 3 1.1 Wave types.......................................... 3 1.2 Mathematical formulation.................................. 4 1.3
Uses of Derivative Spectroscopy
Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative
Vascular Laboratory Fundamentals. Robert Mitchell MD, RPVI Duke University Medical Center 5/2/08
Vascular Laboratory Fundamentals Robert Mitchell MD, RPVI Duke University Medical Center 5/2/08 Goals of Lecture Understand basic ultrasound principles Understand normal and abnormal arterial hemodynamics
INTERFERENCE OF SOUND WAVES
2011 Interference - 1 INTERFERENCE OF SOUND WAVES The objectives of this experiment are: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves. To observe interference phenomena