Name: Due: September 21 st Physics 7230 Laboratory 3: High Resolution SEM Imaging
|
|
|
- Noreen Murphy
- 9 years ago
- Views:
Transcription
1 Name: Due: September 21 st 2012 Physics 7230 Laboratory 3: High Resolution SEM Imaging 1. What is meant by the term resolution? How does this differ from other image variables, such as signal to noise ratios and contrast/brightness? 2. Compare the images at a WD of 20mm and 5mm with the upper and lower detectors. Explain why you think the difference exists and whether resolution is affected. 3. What do you observe in the images taken at different accelerating voltages (Vacc)? Discuss your observations in terms of resolution. 4. What did you observe in terms of resolution in the images taken at working distances of 5 and 20 mm? Discuss any differences you see and why they may have occurred. 5. What did you observe in the images taken at the two Cond Lens 1 settings? Define spot size and relate this idea to any differences you might have observed.
2 Physics 7230 Laboratory 3: High Resolution SEM Imaging INTRODUCTION: This lab explores the necessary conditions to obtain high quality, high resolution images. These include working distance, spot size and accelerating voltage. SAMPLE: Evaporated Gold on Carbon Resolution Standard OBJECTIVES: High magnification imaging with a cold cathode FESEM.
3 INSTRUMENT PARAMETERS: 5-30 kev, 10 µamps, Ultra High Resolution setting A. SET-UP and OBTAIN IMAGES: Working Distance and Upper/Lower Detectors 1. Log in. 2. Check VACUUM conditions. See CHAMBER VACUUM panel. Both SC and SEC lights should be green. 3. The sample is already in the specimen chamber. See TV monitor 4. Set the Vacc (kv) to 10 and the le (emission current) to 10. Turn HV on. 5. Click on the H/L button. This puts you in the low magnification [LM] mode. 6. Set the mode to NORMAL (SETUP-COLUMN). 7. Set the CON LENS 1 to An image of the holder should be present at ~25-30X with the hole in the middle. If not, check CONTRAST and BRIGHTNESS. 9. Set your scan speed to FAST1. This is the slower of the two fast scan modes and is less noisy. 10. Switch back to HIGH MAG mode. 11. Make sure the UPPER SE detector is on (SETUP-COLUMN). 12. Set sample working distance to 5mm. (Change WD to 5mm with Coarse Focus, then use Z-axis to focus sample on image screen. 13. Change mag to 200Kx or higher and focus and check stigmation. 14. Change mag to 100kx, set scan speed to SLOW1, adjust Contrast and Brightness, and take an image. Do not refocus after zooming out! Label image 10SEU_WD Switch to Lower Detector (SETUP-COLUMN). Take a second image after adjusting contrast and brightness. Label image 10SEL_WD Decrease mag to around 300x. 17. Set the WD to 20mm with the coarse focus. With the STAGE Z control, lower the sample into focus. 18. Increase mag to 200kX or higher, focus and stigmate image.
4 19. Decrease mag to 100kX and take image as before. Label it 10SEL_WD Switch to Upper Detector and take a second image, as before. Label it 10SEU_WD20. B. Obtain Images: Spot Size 1. Keep Vacc at 10KV, WD to 10, Cond 1 to 3 (Setup-Column). Set Detector to Mixed. 2. Set mag to 200KX. 3. Adjust Contrast/Brightness 4. Focus and align/stigmate the image. 5. Decrease mag down to 100KX. Do not refocus. 6. Set scan speed to Slow1, do final adjustment of Contrast/Brightness. 7. Obtain image and label it 10SEM_SS3_WD Change Cond 1 to Increase mag to 200KX, Focus, align the aperture and correct for astigmatism if needed. 10. Decrease mag to 100KX, go to Slow1 and obtain an image. Label it 10SEU_SS8_WD10. C. Obtain Images: Accelerating Voltage 1. Set Cond 1 to 8 and Vacc to 5KV. Obtain an image on the screen and go to a working distance of 10 mm. (Adjust the coarse focus until WD=10, then bring specimen into focus with Z-axis control. Watch the image on the TV screen so nothing touches the final lens polepiece!). Set Detector to Mixed. 2. Set mag to 200KX. Align, focus and stigmate the image.
5 3. Reduce mag to 100KX. Do not refocus. 4. Set scan speed to Slow1, correct contrast/brightness and obtain an image. Label it 5SEM-WD Set Vacc to 30KV. 6. Increase mag to 200KX and repeat aperture centering and astigmatism correction. 7. Reduce mag to 100KX. Do not refocus. 8. Slow scan speed to Slow1, adjust brightness/contrast. Record an image and label it 30SEM_WD10. When finished, please leave sample in microscope.
The Basics of Scanning Electron Microscopy
The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single
View of ΣIGMA TM (Ref. 1)
Overview of the FESEM system 1. Electron optical column 2. Specimen chamber 3. EDS detector [Electron Dispersive Spectroscopy] 4. Monitors 5. BSD (Back scatter detector) 6. Personal Computer 7. ON/STANDBY/OFF
Start Up. On the right screen, click on the following icon in order to launch the program for EDS analysis.
Start Up On the right screen, click on the following icon in order to launch the program for EDS analysis. You must start up SEM application before starting EDS application. Don t forget to switch off
INTRODUCTION TO THE XL30-FEG SEM. 1.1 The mouse. 1.2. The monitor. Figure 1.1 1. THE USER INTERFACE
INTRODUCTION TO THE XL30-FEG SEM All software used to control the microscope runs in the MS-Windows environment. This environment is loaded on the Windows2000 operating system. However, it is not really
SEMTech Solutions. Leaders in Refurbished SEMs. SEMTech Solutions Windows 7 SOFTWARE CONTROL SYSTEM
SEMTech Solutions Leaders in Refurbished SEMs SEMTech Solutions Windows 7 SOFTWARE CONTROL SYSTEM Recertification Process Our Goal: Value Added Technologies Demo Outgoing Inspection Can Include: New PC
Electron Microscopy 3. SEM. Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts
Electron Microscopy 3. SEM Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts 3-1 SEM is easy! Just focus and shoot "Photo"!!! Please comment this picture... Any
Detailed Alignment Procedure for the JEOL 2010F Transmission Electron Microscope
Detailed Alignment Procedure for the JEOL 2010F Transmission Electron Microscope by Wendy Sarney ARL-MR-603 December 2004 Approved for public release; distribution unlimited. NOTICES Disclaimers The findings
Ion Beam Sputtering: Practical Applications to Electron Microscopy
Ion Beam Sputtering: Practical Applications to Electron Microscopy Applications Laboratory Report Introduction Electron microscope specimens, both scanning (SEM) and transmission (TEM), often require a
Scanning Electron Microscopy Primer
Scanning Electron Microscopy Primer Bob Hafner This primer is intended as background for the Introductory Scanning Electron Microscopy training offered by the University of Minnesota s Characterization
7. advanced SEM. Latest generation of SEM SEM
7. advanced SEM SEM Low voltage SE imaging Condition of the surface, coatings, plasma cleaning Low voltage BSE imaging Polishing for BSE, EDX and EBSD, effect of ion beam etching/polishing 1 Latest generation
Balzers Sputter Coater SCD 050
Balzers Sputter Coater SCD 050 The SCD 050 is a bench top, sputter deposition system designed for thin films on substrates up to 6 inches. Morphology and thickness is user controlled using power, pressure,
Cathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
SALES SPECIFICATION. SC7640 Auto/Manual High Resolution Sputter Coater
SALES SPECIFICATION SC7640 Auto/Manual High Resolution Sputter Coater Document Number SS-SC7640 Issue 1 (01/02) Disclaimer The components and packages described in this document are mutually compatible
Nanometer-scale imaging and metrology, nano-fabrication with the Orion Helium Ion Microscope
[email protected] Nanometer-scale imaging and metrology, nano-fabrication with the Orion Helium Ion Microscope Bin Ming, András E. Vladár and Michael T. Postek National Institute of Standards and Technology
THERMO NORAN SYSTEM SIX ENERGY DISPERSIVE X- RAY SPECTROMETER. Insert Nickname Here. Operating Instructions
THERMO NORAN SYSTEM SIX ENERGY DISPERSIVE X- RAY SPECTROMETER Insert Nickname Here Operating Instructions Table of Contents 1 INTRODUCTION Safety 1 Samples 1 2 BACKGROUND Background Information 3 References
VCR Ion Beam Sputter Coater
VCR Ion Beam Sputter Coater Sputtering Process and Rates 2 Vacuum System 3 Loading the Sputter Chamber 4 Sputter Coating 5 Removing Samples from Chamber 6 Appendix A: VCR High Vacuum Gauge Conditioning
Mass production, R&D Failure analysis. Fault site pin-pointing (EM, OBIRCH, FIB, etc. ) Bottleneck Physical science analysis (SEM, TEM, Auger, etc.
Failure Analysis System for Submicron Semiconductor Devices 68 Failure Analysis System for Submicron Semiconductor Devices Munetoshi Fukui Yasuhiro Mitsui, Ph. D. Yasuhiko Nara Fumiko Yano, Ph. D. Takashi
Zeiss 780 Training Notes
Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn
4.1 DESCRIPTION OF CONTROLS cont.
4.1 DESCRIPTION OF CONTROLS cont. EGPS-1022 Power Supply Remote Control Computer Panel (ADDITIONAL CONTROLS AND INDICATORS BELOW MAIN PANEL) Fig. 4.1-4 Screen captures of LabVIEW TM program for EGPS-1022
Pictorial User s Guide
S-T IMAGING Pictorial User s Guide Copyright 2008 ST Imaging, Inc. a division of Digital Check Corp. All Rights Reserved. Table of Contents Getting Started... 1 Adjust for Viewing...1 Loading Microfilm...2
Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES
BIODIVERSITY I BIOL1051 Microscopy Professor Marc C. Lavoie [email protected] MAJOR FUNCTIONS OF MICROSCOPES MAGNIFY RESOLVE: => INCREASE CONTRAST Microscopy 1. Eyepieces 2. Diopter adjustment
Firearms & Tool Marks Comparison Microscope. Discovery. Leeds Forensic Systems. Your Forensic Imaging Source
Firearms & Tool Marks Comparison Microscope Leeds Forensic Systems Your Forensic Imaging Source Leeds Discovery The Leeds Discovery Firearms & Tool Marks Comparison Microscope is an innovative microscope
CSCI 4974 / 6974 Hardware Reverse Engineering. Lecture 8: Microscopy and Imaging
CSCI 4974 / 6974 Hardware Reverse Engineering Lecture 8: Microscopy and Imaging Data Acquisition for RE Microscopy Imaging Registration and stitching Microscopy Optical Electron Scanning Transmission Scanning
Products - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available!
Products - Microarray Scanners - SpotLight Two-Color Microarray Fluorescence Scanners New! SpotLight 2 now available! Arrayit SpotLight Two-Color Microarray Scanners provide the market s most affordable
Infrared Viewers. Manual
Infrared Viewers Manual Contents Introduction 3 How it works 3 IR viewer in comparison with a CCD camera 4 Visualization of infrared laser beam in mid-air 4 Power Density 5 Spectral sensitivity 6 Operation
Electron Microscopy 3. SEM. Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts
Electron Microscopy 3. SEM Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts SEM is easy! Just focus and shoot "Photo"!!! Please comment this picture... Any idea
Care and Use of the Compound Microscope
Revised Fall 2011 Care and Use of the Compound Microscope Objectives After completing this lab students should be able to 1. properly clean and carry a compound and dissecting microscope. 2. focus a specimen
PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES
British Made Portable Microscopes with Image Erection PCB Through - Hole Scope Depthscope Brinell Hardness Impression Microscope Cathode Ray Tube Inspector Portable Video Scope Metallurgical Microscope
Introduction to the Scanning Electron Microscope
Introduction to the Scanning Electron Microscope Theory, Practice, & Procedures Prepared by Michael Dunlap & Dr. J. E. Adaskaveg Presented by the FACILITY FOR ADVANCED INSTRUMENTATION, U. C. Davis 1997
Tuning & Mass Calibration
Tuning & Mass Calibration 1 1 The Sample List Sample List Name Project Name 2 The sample list is the top level screen in the TurboMass Gold Software. Data storage is set up in PROJECT files and within
1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002
05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical
Fundamentals of Scanning Electron Microscopy
1 Fundamentals of Scanning Electron Microscopy Weilie Zhou, Robert P. Apkarian, Zhong Lin Wang, and David Joy 1. Introduction The scanning electron microscope (SEM) is one of the most versatile instruments
INCA Energy. Operator Manual. INCA Energy Operator Manual. fëëìé=okn= g~åì~êó=omms
INCA Energy Operator Manual INCA Energy Operator Manual fëëìé=okn= g~åì~êó=omms Oxford Instruments Analytical Halifax Road High Wycombe HP12 3SE, UK Section 1 - INCA Energy System Overview 1 Section 2
The service note describes the basic steps to install a ip camera for the DVR670
Tech note Description Adding IP camera to DVR670 General The service note describes the basic steps to install a ip camera for the DVR670 Steps involved: 1) Configuration Manager application 2) Camera
Forensic Science: The Basics. Microscopy
Forensic Science: The Basics Microscopy Chapter 6 Jay A. Siegel,Ph.D. Power point presentation by Greg Galardi, Peru State College, Peru Nebraska Presentation by Greg Galardi, Peru State College CRC Press,
Usage of AFM, SEM and TEM for the research of carbon nanotubes
Usage of AFM, SEM and TEM for the research of carbon nanotubes K.Safarova *1, A.Dvorak 2, R. Kubinek 1, M.Vujtek 1, A. Rek 3 1 Department of Experimental Physics, Faculty of Science, Palacky University,
Basics of Image and data analysis in 3D
Basics of Image and data analysis in 3D outline Why image processing, and how? Image processing in 2D What is an ideal image? Histogram tells stories! Before taking the image: the right imaging conditions!
FEATURE GUIDE AeroSphère / Small / Large / Base
FEATURE GUIDE AeroSphère / Small / Large / Base Content: Wi-Fi Setup - advanced topics WPS - advanced topics Wi-Fi Setup with the AeroSphère Network Setup webpage Troubleshooting & FAQ Version v0.52 WI-FI
COLOR TFT LCD MONITOR. User Manual
COLOR TFT LCD MONITOR User Manual GENERAL INFORMATION Thank you for choosing our TFT LCD(liquid crystal display) monitor. This product employs integrate circuits, low power consumption, and no radiation
Outlook Profile Setup Guide Exchange 2010 Quick Start and Detailed Instructions
HOSTING Administrator Control Panel / Quick Reference Guide Page 1 of 9 Outlook Profile Setup Guide Exchange 2010 Quick Start and Detailed Instructions Exchange 2010 Outlook Profile Setup Page 2 of 9 Exchange
User Guide LUXXOR VIDEO MICROSCOPE. 2 Luxxor Video Microscope Set Up
2 Luxxor Video Microscope Set Up User Guide LUXXOR VIDEO MICROSCOPE Column When removed from its packaging, the Luxxor Video Microscope will be configured as shown, with the Vertical Slide and Slide Stop
CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL
CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL Robert Bagnell 2006 This tutorial covers the following CLSM topics: 1) What is the optical principal behind CLSM? 2) What is the spatial resolution in X, Y,
X-Ray Fluorescence - Data Collection and Analysis
XRF User Manual 1 X-Ray Fluorescence - Data Collection and Analysis Matthias Zeller, Jerry Hunter, Sept. 19 th 2008 A) Data Collection Logging In: The login and password to log into the program as a normal
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
Scanning Electron Microscope Operation Zeiss Supra-40
Scanning Electron Microscope Operation Zeiss Supra-40 Roger Robbins 9/10/2010 Update: 9/29/2015 Erik Jonsson School of Engineering Title: Scanning Electron Microscope Operation Page 1 of 85 Scanning Electron
Name Class Date Laboratory Investigation 4B Chapter 4: Cell Structure
Name Class Date Laboratory Investigation 4B Chapter 4: Cell Structure The Microscope: A Tool of the Scientist You may refer to pages 66-67, 72-73 in your textbook for a general discussion of microscopes.
5. Tutorial. Starting FlashCut CNC
FlashCut CNC Section 5 Tutorial 259 5. Tutorial Starting FlashCut CNC To start FlashCut CNC, click on the Start button, select Programs, select FlashCut CNC 4, then select the FlashCut CNC 4 icon. A dialog
EXPERIMENT #1: MICROSCOPY
EXPERIMENT #1: MICROSCOPY Brightfield Compound Light Microscope The light microscope is an important tool in the study of microorganisms. The compound light microscope uses visible light to directly illuminate
Operating Manual for UVEX-p (JAN Scientific, Inc. Aug 2014)
Operating Manual for UVEX-p (JAN Scientific, Inc. Aug 2014) [The beige shaded areas introduce the user interface in detail. For turn on and run operation, you may skip these descriptions. The blue shaded
Chapter 4. Microscopy, Staining, and Classification. Lecture prepared by Mindy Miller-Kittrell North Carolina State University
Chapter 4 Microscopy, Staining, and Classification 2012 Pearson Education Inc. Lecture prepared by Mindy Miller-Kittrell North Carolina State University Microscopy and Staining 2012 Pearson Education Inc.
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
Institut für Experimentelle Immunologie und Bildgebung
Page 1 von 6 Starting Up the System Start up the computer, start Windows and log in as IMCES_user start BD FACSDiva software with your Groupname Turn on the cytometer main power Wait a moment and Check
How to Configure Outlook 2010 E-mail Client for Exchange
LAUSD IT Help Desk How to Configure Outlook 2010 E-mail Client for Exchange (03/04/2012) LAUSD IT Help Desk 333 S. Beaudry Ave. 9 th Floor Phone 213.241.5200 Table of Contents Configure Outlook 2010 for
Lacerta MGEN Quick Start Guide
Lacerta MGEN Quick Start Guide While The Lacerta MGEN Superguider provides many functions for the advanced, it is actually quite simple to use even for a beginner. Here we describe the basic steps to start
Temperature Datalogger Users Manual
Temperature Datalogger Users Manual Page 1 TABLE OF CONTENTS FEATURES & SPECIFICATIONS... 3 APPLICATIONS... 4 START UP AND GENERAL OPERATION... 5 LCD Display... 5 Turn On Temperature Data Logger... 6 Switch
Preface Light Microscopy X-ray Diffraction Methods
Preface xi 1 Light Microscopy 1 1.1 Optical Principles 1 1.1.1 Image Formation 1 1.1.2 Resolution 3 1.1.3 Depth of Field 5 1.1.4 Aberrations 6 1.2 Instrumentation 8 1.2.1 Illumination System 9 1.2.2 Objective
Optical Lab Room 1122 Nexus 670 and Magna 560 Instructions Feb. 26, 2013
Optical Lab Room 1122 Nexus 670 and Magna 560 Instructions Feb. 26, 2013 To get set up to use the FTIR systems, first contact Scott at extension 51806 e- mail [email protected] to make appointment for training/orientation
Operating instructions Diffuse reflection sensor. OJ50xx 701396 / 01 07 / 2004
Operating instructions Diffuse reflection sensor OJ50xx 7096 / 0 07 / 004 Contents Preliminary note. Symbols used Function and features Installation. Installation of the supplied mounting fixture 4 4 Electrical
for Low power Energy Harvesting Sun to fiber' Solar Devices
Nanostructured Energy Conversion for Low power Energy Harvesting Devices and Beyond for High power Sun to fiber' Solar Devices Michael Oye and Nobuhiko Nobby Kobayashi Advanced Studies Laboratories and
Starting a Videoconference with a portable unit:
Starting a Videoconference with a portable unit: 1. Plug in the power bar with the TV & Video conference unit and the Network Line. 2. If the LED above the power button on the video conference is off,
Learning Objectives:
Proteomics Methodology for LC-MS/MS Data Analysis Methodology for LC-MS/MS Data Analysis Peptide mass spectrum data of individual protein obtained from LC-MS/MS has to be analyzed for identification of
Instructions for the Epson Perfection 1670 Scanner
Instructions for the Epson Perfection 1670 Scanner Mathew Dando 7-11-05 Basic Scanning One file at a time Double click on the Epson Scan icon on the desktop Or Start -> Programs -> Epson Scan Put the object
EF-S18-55mm f/3.5-5.6 IS II COPY ENG. Instruction
EF-S18-55mm f/3.5-5.6 IS II ENG Instruction Thank you for purchasing a Canon product. The Canon EF-S18-55mm f/3.5-5.6 IS II lens is a high-performance standard zoom lens developed for digital SLRs compatible
TwinCAT NC Configuration
TwinCAT NC Configuration NC Tasks The NC-System (Numeric Control) has 2 tasks 1 is the SVB task and the SAF task. The SVB task is the setpoint generator and generates the velocity and position control
INSTRUCTION MANUAL FOR MODEL S-4800 FIELD EMISSION SCANNING ELECTRON MICROSCOPE
INSTRUCTION MANUAL FOR MODEL S-4800 FIELD EMISSION SCANNING ELECTRON MICROSCOPE Please read through this manual carefully before using the instrument. Before using the instrument, read the safety instructions
Electron Microscopy SEM and TEM
Electron Microscopy SEM and TEM Content 1. Introduction: Motivation for electron microscopy 2. Interaction with matter 3. SEM: Scanning Electron Microscopy 3.1 Functional Principle 3.2 Examples 3.3 EDX
GETTING STARTED. 6. Click the New Specimen icon.
BD FACSDiva Software Basic Experiment Guide Please note, this guide is not intended to take the place of in-person training but is to be used as a reference as needed after completing training with Unified
User Manual CROMLAWATCH. Data Logging Software. Version 1.6. For use with Color Sensors CROMLAVIEW CR100 CROMLAVIEW CR200 CROMLAVIEW CR210
User Manual CROMLAWATCH Data Logging Software Version 1.6 For use with Color Sensors CROMLAVIEW CR100 CROMLAVIEW CR200 CROMLAVIEW CR210 CROMLAWATCH User manual Contents Notes The information contained
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM
picture real beauty Capture real beauty with the one you love
picture real beauty Capture real beauty with the one you love Lina Dove invites you to experience photography for yourself and capture images of beauty that live around you. To help, we ve taken the principles
INSTRUCTIONS Please read all instructions carefully before using
Car Multimedia System XM-2DTSB79 AUX IR MIC RESET USB INSTRUCTIONS Please read all instructions carefully before using Touch calibration If touch calibration error occurs, you can adjust the touch coordinate
Micro Cam Software. User Manual V1.3
Micro Cam Software User Manual V1.3 CONTENT CHAPTER 1: MICRO CAM SOFTWARE INSTALLATION AND CONNECTION... - 1-1.1 SOFTWARE MICRO CAM INSTALLATION... - 1-1.2 WIRED DEVICE CONNECTION... - 4-1.3 SOFTWARE OPERATION
Exercise 2. The Compound Light Microscope
6 Exercise 2 The Compound Light Microscope INTRODUCTION: Student Learning Objectives: After completing this exercise students will: a. Demonstrate proficient use of the microscope using low, high dry,
The Focused Ion Beam Scanning Electron Microscope: A tool for sample preparation, two and three dimensional imaging. Jacob R.
The Focused Ion Beam Scanning Electron Microscope: A tool for sample preparation, two and three dimensional imaging Jacob R. Bowen Contents Components of a FIB-SEM Ion interactions Deposition & patterns
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)
SE05: Getting Started with Cognex DataMan Bar Code Readers - Hands On Lab Werner Solution Expo April 8 & 9
SE05: Getting Started with Cognex DataMan Bar Code Readers - Hands On Lab Werner Solution Expo April 8 & 9 Learning Goals: At the end of this lab, the student should have basic familiarity with the DataMan
XMEye Android Mobile Monitoring Client English Manual
XMEye Android Mobile Monitoring Client English Manual Written by: The test department Prepared Date: 2014-03-25 Version: V1.0 目 录 1. Introduction... 3 1.1. Brief Introduction... 3 1.2. Main Functions...
PROJECT WORKPLACE DEVELOPMENT
CISCO PROJECT WORKPLACE DEVELOPMENT New York Room Basic Meeting Room Integration Price $15,000.00 Annual Service Contract $2,250.00 Zdi is the valued Cisco partner that integrates this room s technology
How to use the Microfilm Reader/Scanner
How to use the Microfilm Reader/Scanner You can print each page directly to the printer on the left of the machine. This printer can only print on one side of the paper. OR, you can scan the entire article
LCD MONITOR TOUCH PANEL DRIVER 2 OPERATION MANUAL. for Mac. Version 1.0 PN-L703A/PN-70TA3/PN-L703B/PN-70TB3/PN-L603A/PN-60TA3/PN-L603B/PN-60TB3
LCD MONITOR TOUCH PANEL DRIVER 2 OPERATION MANUAL for Mac Version 1.0 Applicable models PN-L703A/PN-70TA3/PN-L703B/PN-70TB3/PN-L603A/PN-60TA3/PN-L603B/PN-60TB3 Contents Introduction...3 System Requirements...3
GENESIS SPECTRUM USER S MANUAL. COPYRIGHT EDAX INC. ALL RIGHTS RESERVED EDAX INC. 91 McKEE DRIVE MAHWAH, NJ 07430 USA
GENESIS SPECTRUM USER S MANUAL COPYRIGHT EDAX INC. ALL RIGHTS RESERVED EDAX INC. 91 McKEE DRIVE MAHWAH, NJ 07430 USA 9499.203.50000 January 2003 Revision 3.5 TABLE OF CONTENTS 1 INTRODUCTION TO GENESIS
Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F)
Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F) Please contact Dr. Amanda Young for training requests and assistance: 979-862-6845, [email protected] Hardware Our spectrophotometer is made up of
Win 7 - STIMA CLS / CMP Driver Installation Quick Guide
Win 7 - STIMA CLS / CMP Driver Installation Quick Guide This guide is intended to demonstrate a quick installation and setup of ticket alignment under Windows 7 for the Stima CLS and CMP printers using
Tecnai on-line help Working with a FEG 1 FEG_A4.doc Software version 2
Tecnai on-line help Working with a FEG 1 Tecnai on-line help manual -- Working with a FEG Table of Contents 1 Introduction...2 2 FEG Safety...2 2.1 The column valves...2 3 FEG States...3 4 Starting the
Looking through the fish-eye the Electron Ronchigram. Duncan T.L. Alexander CIME seminar May 24, 2012
Looking through the fish-eye the Electron Ronchigram Duncan T.L. Alexander CIME seminar May 24, 2012 Introduction Aim of the seminar: open a discussion on the Electron Ronchigram How is it formed? What
Carl Zeiss NTS - Nano Technology System Division. ΣIGMA Field Emission Scanning Electron Microscope. Instruction Manual. Enabling the Nano-Age World
Carl Zeiss NTS - Nano Technology System Division ΣIGMA Field Emission Scanning Electron Microscope Instruction Manual Enabling the Nano-Age World Operator s User Guide ΣIGMA FESEM Original instructions
CONCEPT OF DETERMINISTIC ION IMPLANTATION AT THE NANOSCALE
CONCEPT OF DETERMINISTIC ION IMPLANTATION AT THE NANOSCALE Daniel Spemann Jan Meijer 1, Jürgen W. Gerlach, Paul Räcke 1, Susann Liedtke, Stephan Rauschenbach 2, Bernd Rauschenbach 1 University of Leipzig,
Contact Center CC-165 INSTALLATION & USER'S GUIDE
Contact Center CC-165 INSTALLATION & USER'S GUIDE Table of Contents Introduction... 2 Specifications... 2 J4 Host Communications... 2 J5 and J12 (upper board)... 3 J6, J13, and 14 Input Connections...
Chapter I Model801, Model802 Functions and Features
Chapter I Model801, Model802 Functions and Features 1. Completely Compatible with the Seventh Generation Control System The eighth generation is developed based on the seventh. Compared with the seventh,
Measuring the Point Spread Function of a Fluorescence Microscope
Frederick National Laboratory Measuring the Point Spread Function of a Fluorescence Microscope Stephen J Lockett, PhD Principal Scientist, Optical Microscopy and Analysis Laboratory Frederick National
BB-18 Black Body High Vacuum System Technical Description
BB-18 Black Body High Vacuum System Technical Description The BB-18 Black Body is versatile and is programmed for use as a fixed cold target at 80 K or variable target, at 80 K- 350 K no extra cost. The
D/N VANDAL RES. VARI-FOCAL IR DOME CAMERA
D/N VANDAL RES. VARI-FOCAL IR DOME CAMERA PIH-0746XW/0746XS INSTRUCTION MANUAL IMPORTANT SAFEGUARDS CAUTION RISK OF ELECTRIC SHOCK DO NOT OPEN CAUTION TO REDUCE THE RISK OF ELECTRIC SHOCK, DO NOT REMOVE
1. Introduction... 3. 2.Fixture exterior view... 3. 3. Connecting the Robe Universal Interface... 4. 3.1 Software update of the fixture...
1 Table of contests 1. Introduction... 3 2.Fixture exterior view... 3 3. Connecting the Robe Universal Interface... 4 3.1 Software update of the fixture... 4 3.2 Connecting a DMX console to the Media Fusion...
MAINTENANCE & TROUBLESHOOTING
MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens clean the fan intake filter replace the projection lamp replace the batteries in the remote control use the Kensington lock feature
Index. ROPIX P18 Hardware... 3. ROPIX P18 Software... 12 ROPIX P18
Index ROPIX P18 Hardware... 3 I. ROPIX P18 Profile Overview... 3 II. Parameter... 5 Sending Card... 6 Receiving Card... 8 Maintenance Tools... 8 III. ROPIX P18 Installation... 9 Cabinet Assembly... 9 Controlling
DIGA Player Instruction manual for ios device
DIGA Player Instruction manual for ios device Index 1.Activate Remote Recording function on DIGA 2.Let s use DIGA Player 2.1. How to use Remote Recording Service 2.1.1. Make your account (User registration)
Introduction to Autodesk Inventor for F1 in Schools
Introduction to Autodesk Inventor for F1 in Schools F1 in Schools Race Car In this course you will be introduced to Autodesk Inventor, which is the centerpiece of Autodesk s digital prototyping strategy
MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope.
MICROSCOPY I. OBJECTIVES To demonstrate skill in the proper utilization of a light microscope. To demonstrate skill in the use of ocular and stage micrometers for measurements of cell size. To recognize