Detailed Alignment Procedure for the JEOL 2010F Transmission Electron Microscope
|
|
|
- Thomas Hunter
- 10 years ago
- Views:
Transcription
1 Detailed Alignment Procedure for the JEOL 2010F Transmission Electron Microscope by Wendy Sarney ARL-MR-603 December 2004 Approved for public release; distribution unlimited.
2 NOTICES Disclaimers The findings in this report are not to be construed as an official Department of the Army position, unless so designated by other authorized documents. Citation of manufacturers or trade names does not constitute an official endorsement or approval of the use thereof. DESTRUCTION NOTICE Destroy this report when it is no longer needed. Do not return it to the originator.
3 Army Research Laboratory Adelphi, MD ARL-MR-603 December 2004 Detailed Alignment Procedure for the JEOL 2010F Transmission Electron Microscope Wendy Sarney Sensors and Electron Devices Directorate, ARL Approved for public release; distribution unlimited.
4 REPORT DOCUMENTATION PAGE Form Approved OMB No Public reporting burden for this collection of information is estimated to average 1 hour per response, including the time for reviewing instructions, searching existing data sources, gathering and maintaining the data needed, and completing and reviewing the collection information. Send comments regarding this burden estimate or any other aspect of this collection of information, including suggestions for reducing the burden, to Department of Defense, Washington Headquarters Services, Directorate for Information Operations and Reports ( ), 1215 Jefferson Davis Highway, Suite 1204, Arlington, VA Respondents should be aware that notwithstanding any other provision of law, no person shall be subject to any penalty for failing to comply with a collection of information if it does not display a currently valid OMB control number. PLEASE DO NOT RETURN YOUR FORM TO THE ABOVE ADDRESS. 1. REPORT DATE (DD-MM-YYYY) December TITLE AND SUBTITLE 2. REPORT TYPE Summary 3. DATES COVERED (From - To) 01/ a. CONTRACT NUMBER Detailed Alignment Procedure for the JEOL 2010F Transmission Electron Microscope 6. AUTHOR(S) Wendy Sarney 5b. GRANT NUMBER 5c. PROGRAM ELEMENT NUMBER 5d. PROJECT NUMBER 5e. TASK NUMBER 5f. WORK UNIT NUMBER 7. PERFORMING ORGANIZATION NAME(S) AND ADDRESS(ES) U.S. Army Research Laboratory Sensors & Electron Devices Directorate (ATTN: AMSRD-ARL-SE-EI) [email protected] Adelphi, MD SPONSORING/MONITORING AGENCY NAME(S) AND ADDRESS(ES) ARL 2800 Powder Mill Road Adelphi, MD DISTRIBUTION/AVAILABILITY STATEMENT Approved for public release; distribution unlimited. 8. PERFORMING ORGANIZATION REPORT NUMBER ARL-MR SPONSOR/MONITOR'S ACRONYM(S) 11. SPONSOR/MONITOR'S REPORT NUMBER(S) 13. SUPPLEMENTARY NOTES 14. ABSTRACT The transmission electron microscope (TEM) allows lattice-resolution imaging of specimens. At ARL, we rely on TEM imaging for detailed structural characterization. Typical experiments involve examining the crystalline structure, interface quality, and defect morphology of semiconductor materials. The TEM consists primarily of an electron gun, electromagnetic lenses, apertures, stigmators, deflectors, and a viewing system. The alignment affects the propagation of the beam down the column and its interaction with the specimen and lenses. This in turn affects the quality of the image and the ease of obtaining it. This report discusses the alignment procedure for the JEOL 2010F TEM. 15. SUBJECT TERMS Transmission Electron Microscopy (TEM) 16. SECURITY CLASSIFICATION OF: a. REPORT UNCLASSIFIED b. ABSTRACT UNCLASSIFIED c. THIS PAGE UNCLASSIFIED 17. LIMITATION OF ABSTRACT UNCLASSIFIED 18. NUMBER OF PAGES 15 19a. NAME OF RESPONSIBLE PERSON Wendy Sarney 19b. TELEPHONE NUMBER (Include area code) (301) Standard Form 298 (Rev. 8/98) ii
5 Contents 1. Introduction 1 2. Definitions and Related Figures F Column Schematic Index of Knobs, Dials, and Levers F TEM Alignment Procedure Gun and Condenser Lens Alignment Objective Lens Alignment Intermediate Lens Alignment Conclusion 7 5. Reference 8 Distribution list 9 Figure Figure 1. Schematic of lenses, apertures, stigmators, and deflectors pertinent to the TEM alignment....2 iii
6 This page is intentionally left blank. iv
7 1. Introduction The Sensors and Electron Devices Directorate (SEDD) operates two transmission electron microscopes (TEMs). One is a JEOL 2010 with a thermionic electron gun using a LaB6 filament and the other is the JEOL 2010F with a field emission gun. Both TEMs normally operate at 200 kev. The 2010F also contains a scanning image observation device (ASID) and an energy dispersive x-ray detector (EDAX). The TEMs support numerous projects throughout the directorate, but primarily in the Electro-Optics & Photonics and the RF Electronics divisions. Typical experiments involve examining the crystalline structure, interface quality, and defect morphology of semiconductor materials. Characterization at the nano-scale requires high-resolution imaging. Some alignments, particularly the objective lens astigmatism and the high voltage centering, are critical to the TEM s performance and are touched up throughout a single TEM session. Other alignments, such as shift and tilt purity adjustments, are done less frequently. We typically check the less-critical alignments when it becomes difficult, but not impossible, to obtain a high-resolution image. Since the TEM sample tends to degrade with prolonged exposure to the electron beam, it is important to keep the microscope aligned so that we can minimize the amount of time needed to obtain a quality image. For that same reason, when performing the alignments that require the sample to be in the path of the beam (such as the objective lens astigmatism and the high voltage centering), we select an area of the sample that is near, but not the exact region to be examined during the experiment. JEOL service engineers check all customer and engineer alignments at least twice per year during a routine maintenance visit as part of the service contract. 1
8 2. Definitions and Related Figures F Column Schematic Figure 1 shows a schematic of the lenses, deflectors, and apertures pertinent to the alignment of the 2010F. Refer to the JEOL manual for more detailed schematics. Electron gun chamber Emitter Electrostatic lens Gun beam deflectors 1 st and 2 nd condenser lenses Condenser lens aperture Condenser lens stigmators Condenser lens deflectors Condenser mini-lens Microscope Column Objective lens aperture Objective lens Objective lens stigmators Objective mini-lens Image shifts Selected area aperture Intermediate lens stigmators 1 st, 2 nd, and 3 rd intermediate lenses Projector lens deflectors Projector lens Fluorescent screen Figure 1. Schematic of lenses, apertures, stigmators, and deflectors pertinent to the TEM alignment. 2
9 2.2 Index of Knobs, Dials, and Levers In this section, we define the user controls necessary for alignment. α SELECTOR knob Controls the current through the condenser mini-lens (CM), which in turn affects the angle of incidence of electrons on the specimen. Beam SHIFT X and SHIFT Y knobs BRIGHTNESS knob BRIT TILT Condenser lens apertures Deflector coils Shifts the electron beam by controlling the current in the condenser lens beam deflectors. Allows the variation of the second condenser lens (CL3) current. This modifies the beam intensity. Used with the DEF X,Y knobs on the panel to tilt the beam. Restrict the number of electrons traveling down the column and striking the specimen. Reduces the maximum allowed angle between the traveling electrons and the optical axis. Restricting the quantity and the incidence angle of electrons in the beam reduces the number of x-rays generated by electrons striking parts of the column. There are 5 condenser lens aperture sizes having diameters of 10, 20, 50, 70, and 150 µm. Allow the beam to be shifted or tilted, to correct for mechanical misalignments in the column or to achieve special conditions, such as dark field imaging. Gun DEF X and Y knobs Tilts the beam by varying the current in the electron gun beam deflector coil. Select the Gun DEF knobs by depressing the GUN deflector button, located inside the right hand drawer. Gun SHIFT X and SHIFT Y knobs DIFF FOCUS knob FOCUS knob WOBBLER-HT SPOT SIZE switch Translates the beam by varying the current in the electron gun beam deflector coil. The shift knobs are selected by depressing the GUN deflector button. These controls are located inside the right hand drawer. Varies the current in the 1 st intermediate lens. Controls the focus of the image by modifying the current in the objective lens. Varies the high voltage, used for high voltage centering. Changes the first condenser lens (CL1) current. The spot size refers to the size of the converged electron beam obtained for maximum BRIGHTNESS. Larger numerical size numbers refer to smaller spot sizes (Spot size 1 is the largest, spot size 5 is the smallest). 3
10 Stigmators Z CONT switch Allow for the correction of lens imperfections. We can adjust the condenser, objective, and intermediate lens astigmatism. Moves the specimen upward and downward and provides rough focusing F TEM Alignment Procedure For full details, please see the JEOL manual. Some alignments are done infrequently, and some are usually only done by the JEOL engineers when they service the microscope. 1. Press the VALVE button to allow the beam to come down the column. 2. Check the DV reading (on the screen). Adjust the FOCUS knob until DV=0. The microscope is calibrated at DV=0. It is very difficult to obtain a high quality image if you deviate far from DV=0. 3. Focus the image with the Z control. To avoid deviating far from DV=0, perform all rough focusing with the Z control rather than the FOCUS knob. 4. Using the SPOT SIZE lever and the α-selector knob, set the spot size to 1 and the convergent angle to 3. The first page on the TV screen will display TEM Gun and Condenser Lens Alignment The 2010F has a three-stage condenser lens system consisting of the 1 st (CL1), 2 nd (CL3), and mini condenser lens (CM). It is important to note that this is different from the 2010, which has a four-stage condenser lens system consisting of a 1 st, 2 nd, 3 rd, and mini condenser lens. The 2010F needs one less condenser lens because the size of the original beam from the field emission gun is smaller than can be generated from the LaB 6 filament in the This causes discrepancies and confusion in the JEOL 2010F manual. They inconsistently refer to the second condenser lens as CL3, the third condenser lens, or CL2. This happened because the 2010F manual is an expanded and modified 2010 manual, resulting in several inconsistencies. In this document, we will always refer to it as CL3, since that is the convention used by the JEOL service engineers. We control the current through CL3 with the BRIGHTNESS knob. We perform most of the condenser lens alignments at a magnification of 40000X. 1. Align the electron gun axis. a. Select the BRIT TILT button on the panel. b. Press the ANODE WOBBLER button. c. Use the TILT knobs on the panel to make the beam expand and contract concentrically. 2. Alignment of the gun to the condenser lens this step aligns the illumination along the optical axis. 4
11 a. Insert a condenser lens aperture. b. Decrease the spot size to 5 (TEM5-3) using the SPOT SIZE switch. c. Adjust CL3 to obtain crossover and center the spot with the SHIFT knobs on the right and left side console panels. d. Increase the spot size to 1 (TEM1-3) and adjust CL3 to obtain crossover. e. Center the spot with the GUN SHIFT knobs located in the drawer f. Repeat steps b-e until the spot stays centered whether CL1 is minimized or maximized. If the beam does not remain centered or look like it is improving after a few iterations, repeat the gun alignment described in section 3.1 and then retry steps a-e. 3. High Voltage Centering a. Select an easily recognizable feature in the sample. Increase the magnification to 100X. Make sure that the Z-height is correct. b. Press the WOBBLER-HT button. c. Raise the view screen. d. Select the BRIGHT TILT. e. Minimize the motion of the specimen by adjusting the panel beam deflectors (DEF x and y). 4. Condenser aperture alignment a. Insert the condenser lens aperture that you plan to work with. b. With the BRIGHTNESS knob, obtain the smallest possible beam (crossover). Center the beam with the SHIFT knobs on the panel. c. Spread the beam until it is about 10 cm wide. Center the beam with the x and y condenser aperture controls (two knobs located next to the condenser lens aperture). d. Repeat b-c until beam remains centered as CL3 is maximized and minimized. 5. Condenser lens astigmatism correction (correct at each spot size needed during the session). a. Press the COND STIG button on the panel. Slightly turn the BRIGHTNESS knob back and forth from crossover. Correct the ellipticity of the beam with the DEF knobs on the panel. b. Before continuing with other alignments turn off the COND STIG by pressing the BRIGHT FIELD key. 6. Beam tilt purity alignment (adjusts the current balance in the condenser deflectors, done only as needed). This alignment ensures that the beam does not shift and tilt simultaneously. a. Set the magnification at 100,000X. b. Set the SPOT SIZE to TEM1-3. Obtain crossover with the CL3 knob. Center the beam with the COND SHIFTS knobs in the drawer. c. Depress the COND DEF ADJ TILT button and turn the TILT X/Y switch to the X side. These knobs and switches are in the right hand drawer. d. Make the separated spots overlap using the SHIFT X and DEF X knobs in the drawer. e. Turn the TILT X/Y switch to the Y side and adjust the SHIFT Y and DEF Y knobs until the separated spots overlap. 5
12 f. Do a rough voltage center alignment spread the beam, center it, press WOBBLER HT, and minimize the beam motion with the gun deflectors located on the panel. 7. Shift interlock adjustment (adjusts the current balance in the condenser deflectors, done only as needed). This alignment ensures that the beam does not shift and tilt simultaneously. a. Press the DIFF button and set the camera length to 200 cm. b. Press the COND DEF ADJ SHIFT knob and switch the SHIFT X/Y switch to X. c. Adjust the SHIFT X and DEF X knobs (drawer) until the separated caustic spots overlap. d. Switch the SHIFT X/Y switch to Y. e. Adjust the SHIFT Y and DEF Y knobs (drawer) until the separated caustic spots overlap. 3.2 Objective Lens Alignment 1. The JEOL manual describes a procedure for obtaining the proper objective lens current. This current was determined at the installation, and we should not need to make adjustments. 2. Objective lens astigmatism correction a. The best magnifications for correcting objective lens astigmatism falls in the range of kX. b. Find and center an area of the specimen with an amorphous edge c. Raise the screen and view the image on the TV monitor. d. Slightly under-focus the image and examine the bright Fresnel fringes at the edge. e. Select the OBJ STIG button and adjust X and Y DEF until the contrast at the edges becomes sharp. f. Adjust the FOCUS knob closer to optimal focus and repeat step e. g. The objective lens astigmatism is optimized when very little contrast is seen at the sample edge. h. After completing the entire alignment procedure, touch up the astigmatism at the magnification needed for the current TEM session by repeating steps b-g. 3.3 Intermediate Lens Alignment 1. Diffraction focus alignment (This alignment is needed when an unusual number of diffraction spots seem to fit through the objective lens aperture) a. Press the DIFF button and set the camera length to 200 cm. b. Maximize CL3. c. Minimize the diameter of the caustic spot by adjusting the DIFF FOCUS knob. 2. Intermediate lens astigmatism correction (Needed when the diffraction spots are not perfectly round). a. Select the STIG INT button in the drawer and adjust the deflectors until the caustic spot is round. 6
13 b. Slightly under and over focus the caustic spot with the DIFF FOCUS knob to make sure it stays round. If the spot does not stay round as it passes through focus, repeat step a. c. Minimize the diameter of the caustic spot by adjusting the DIFF FOCUS knob. 4. Conclusion The TEM is ready for high-resolution imaging after completing the alignment procedure. The high voltage centering and the astigmatism may have to be touched up throughout the session. If the alignment ever seems to become very bad, it is usually best to push the N button and then the DEFLECTOR, STIGMATOR, and/or COND DEF ADJ buttons to return to the alignment conditions set by the service engineer. These computer-stored alignments serve as a good starting point for small alignment modifications. If the alignment is still unrecoverable, there is likely a larger problem. In this situation, call the service engineers. 7
14 5. Reference JEOL 2010F Instruction Manual 8
15 Distribution List Admnstr Defns Techl Info Ctr ATTN DTIC-OCP (Electronic copy) 8725 John J Kingman Rd Ste 0944 FT Belvoir VA US Army Rsrch Lab ATTN AMSRD ARL-SE-EI J Little ATTN AMSRD-ARL-CI-IS Mail & Records Mgmt ATTN AMSRD-ARL-CI-OK-T Techl Pub (2 copies) ATTN AMSRD-ARL-CI-OK-TL Techl Lib (2 copies) ATTN AMSRD-ARL-SE-E G Simonis ATTN AMSRD-ARL-SE-EI N Dhar ATTN AMSRD-ARL-SE-EI W Sarney (20 copies) ATTN AMSRD-ARL-SE-EM D Beekman ATTN AMSRD-ARL-SE-RL S Svensson 9
Simulation of Air Flow Through a Test Chamber
Simulation of Air Flow Through a Test Chamber by Gregory K. Ovrebo ARL-MR- 0680 December 2007 Approved for public release; distribution unlimited. NOTICES Disclaimers The findings in this report are not
Lenses and Apertures of A TEM
Instructor: Dr. C.Wang EMA 6518 Course Presentation Lenses and Apertures of A TEM Group Member: Anup Kr. Keshri Srikanth Korla Sushma Amruthaluri Venkata Pasumarthi Xudong Chen Outline Electron Optics
View of ΣIGMA TM (Ref. 1)
Overview of the FESEM system 1. Electron optical column 2. Specimen chamber 3. EDS detector [Electron Dispersive Spectroscopy] 4. Monitors 5. BSD (Back scatter detector) 6. Personal Computer 7. ON/STANDBY/OFF
Name: Due: September 21 st 2012. Physics 7230 Laboratory 3: High Resolution SEM Imaging
Name: Due: September 21 st 2012 Physics 7230 Laboratory 3: High Resolution SEM Imaging 1. What is meant by the term resolution? How does this differ from other image variables, such as signal to noise
The Basics of Scanning Electron Microscopy
The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single
Report Documentation Page
(c)2002 American Institute Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the collection of information is estimated to average 1 hour per response, including the
Overview Presented by: Boyd L. Summers
Overview Presented by: Boyd L. Summers Systems & Software Technology Conference SSTC May 19 th, 2011 1 Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the collection
OPTICAL IMAGES DUE TO LENSES AND MIRRORS *
1 OPTICAL IMAGES DUE TO LENSES AND MIRRORS * Carl E. Mungan U.S. Naval Academy, Annapolis, MD ABSTRACT The properties of real and virtual images formed by lenses and mirrors are reviewed. Key ideas are
DEFENSE CONTRACT AUDIT AGENCY
DEFENSE CONTRACT AUDIT AGENCY Fundamental Building Blocks for an Acceptable Accounting System Presented by Sue Reynaga DCAA Branch Manager San Diego Branch Office August 24, 2011 Report Documentation Page
Guide to Using DoD PKI Certificates in Outlook 2000
Report Number: C4-017R-01 Guide to Using DoD PKI Certificates in Outlook 2000 Security Evaluation Group Author: Margaret Salter Updated: April 6, 2001 Version 1.0 National Security Agency 9800 Savage Rd.
Physics 441/2: Transmission Electron Microscope
Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This
Human Body Mesh With Maya
Animating a Human Body Mesh with Maya for Doppler Signature Computer Modeling by Getachew Kirose ARL-TN-0351 June 2009 Approved for public release; distribution unlimited. NOTICES Disclaimers The findings
Chapter 4. Microscopy, Staining, and Classification. Lecture prepared by Mindy Miller-Kittrell North Carolina State University
Chapter 4 Microscopy, Staining, and Classification 2012 Pearson Education Inc. Lecture prepared by Mindy Miller-Kittrell North Carolina State University Microscopy and Staining 2012 Pearson Education Inc.
Tecnai on-line help Working with a FEG 1 FEG_A4.doc Software version 2
Tecnai on-line help Working with a FEG 1 Tecnai on-line help manual -- Working with a FEG Table of Contents 1 Introduction...2 2 FEG Safety...2 2.1 The column valves...2 3 FEG States...3 4 Starting the
Asset Management- Acquisitions
Indefinite Delivery, Indefinite Quantity (IDIQ) contracts used for these services: al, Activity & Master Plans; Land Use Analysis; Anti- Terrorism, Circulation & Space Management Studies; Encroachment
Using the Advancement Degree of Difficulty (AD 2 ) as an input to Risk Management
Using the Advancement Degree of Difficulty (AD 2 ) as an input to Risk Management James W. Bilbro JB Consulting International Huntsville, AL Multi-dimensional Assessment of Technology Maturity Technology
EAD Expected Annual Flood Damage Computation
US Army Corps of Engineers Hydrologic Engineering Center Generalized Computer Program EAD Expected Annual Flood Damage Computation User's Manual March 1989 Original: June 1977 Revised: August 1979, February
Cathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
INTRODUCTION TO THE XL30-FEG SEM. 1.1 The mouse. 1.2. The monitor. Figure 1.1 1. THE USER INTERFACE
INTRODUCTION TO THE XL30-FEG SEM All software used to control the microscope runs in the MS-Windows environment. This environment is loaded on the Windows2000 operating system. However, it is not really
DCAA and the Small Business Innovative Research (SBIR) Program
Defense Contract Audit Agency (DCAA) DCAA and the Small Business Innovative Research (SBIR) Program Judice Smith and Chang Ford DCAA/Financial Liaison Advisors NAVAIR 2010 Small Business Aviation Technology
Optical Blade Position Tracking System Test
National Renewable Energy Laboratory Innovation for Our Energy Future A national laboratory of the U.S. Department of Energy Office of Energy Efficiency & Renewable Energy Optical Blade Position Tracking
RT 24 - Architecture, Modeling & Simulation, and Software Design
RT 24 - Architecture, Modeling & Simulation, and Software Design Dennis Barnabe, Department of Defense Michael zur Muehlen & Anne Carrigy, Stevens Institute of Technology Drew Hamilton, Auburn University
Issue Paper. Wargaming Homeland Security and Army Reserve Component Issues. By Professor Michael Pasquarett
Issue Paper Center for Strategic Leadership, U.S. Army War College May 2003 Volume 04-03 Wargaming Homeland Security and Army Reserve Component Issues By Professor Michael Pasquarett Background The President
Leveraging Open Source Software to Create Technical Animations of Scientific Data
Leveraging Open Source Software to Create Technical Animations of Scientific Data by John M. Vines ARL-TR-3918 September 2006 Approved for public release; distribution is unlimited. NOTICES Disclaimers
Integrated Force Method Solution to Indeterminate Structural Mechanics Problems
NASA/TP 2004-207430 Integrated Force Method Solution to Indeterminate Structural Mechanics Problems Surya N. Patnaik Ohio Aerospace Institute, Brook Park, Ohio Dale A. Hopkins and Gary R. Halford Glenn
Looking through the fish-eye the Electron Ronchigram. Duncan T.L. Alexander CIME seminar May 24, 2012
Looking through the fish-eye the Electron Ronchigram Duncan T.L. Alexander CIME seminar May 24, 2012 Introduction Aim of the seminar: open a discussion on the Electron Ronchigram How is it formed? What
Advanced Micro Ring Resonator Filter Technology
Advanced Micro Ring Resonator Filter Technology G. Lenz and C. K. Madsen Lucent Technologies, Bell Labs Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the collection
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control
John Mathieson US Air Force (WR ALC) Systems & Software Technology Conference Salt Lake City, Utah 19 May 2011
John Mathieson US Air Force (WR ALC) Systems & Software Technology Conference Salt Lake City, Utah 19 May 2011 Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the
Headquarters U.S. Air Force
Headquarters U.S. Air Force I n t e g r i t y - S e r v i c e - E x c e l l e n c e Air Force Technology Readiness Assessment (TRA) Process for Major Defense Acquisition Programs LtCol Ed Masterson Mr
Scanning Electron Microscopy Primer
Scanning Electron Microscopy Primer Bob Hafner This primer is intended as background for the Introductory Scanning Electron Microscopy training offered by the University of Minnesota s Characterization
Preface Light Microscopy X-ray Diffraction Methods
Preface xi 1 Light Microscopy 1 1.1 Optical Principles 1 1.1.1 Image Formation 1 1.1.2 Resolution 3 1.1.3 Depth of Field 5 1.1.4 Aberrations 6 1.2 Instrumentation 8 1.2.1 Illumination System 9 1.2.2 Objective
MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope.
MICROSCOPY I. OBJECTIVES To demonstrate skill in the proper utilization of a light microscope. To demonstrate skill in the use of ocular and stage micrometers for measurements of cell size. To recognize
How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
Pima Community College Planning Grant For Autonomous Intelligent Network of Systems (AINS) Science, Mathematics & Engineering Education Center
Pima Community College Planning Grant For Autonomous Intelligent Network of Systems (AINS) Science, Mathematics & Engineering Education Center Technical Report - Final Award Number N00014-03-1-0844 Mod.
Mr. Steve Mayer, PMP, P.E. McClellan Remediation Program Manger Air Force Real Property Agency. May 11, 2011
Mr. Steve Mayer, PMP, P.E. McClellan Remediation Program Manger Air Force Real Property Agency May 11, 2011 Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the collection
Evaluation of a 10 kv, 400 ka Si SGTO at High di/dt
Evaluation of a kv, ka Si SGTO at High di/dt Heather O Brien, William Shaheen, Timothy Crowley Berkeley Research Associates, U.S. Army Research Laboratory, Attn: Berkeley Contractor 8 Powder Mill Road,
Electron Microscopy SEM and TEM
Electron Microscopy SEM and TEM Content 1. Introduction: Motivation for electron microscopy 2. Interaction with matter 3. SEM: Scanning Electron Microscopy 3.1 Functional Principle 3.2 Examples 3.3 EDX
A GPS Digital Phased Array Antenna and Receiver
A GPS Digital Phased Array Antenna and Receiver Dr. Alison Brown, Randy Silva; NAVSYS Corporation ABSTRACT NAVSYS High Gain Advanced GPS Receiver (HAGR) uses a digital beam-steering antenna array to enable
Diffraction of a Circular Aperture
Diffraction of a Circular Aperture Diffraction can be understood by considering the wave nature of light. Huygen's principle, illustrated in the image below, states that each point on a propagating wavefront
An Oil-Free Thrust Foil Bearing Facility Design, Calibration, and Operation
NASA/TM 2005-213568 An Oil-Free Thrust Foil Bearing Facility Design, Calibration, and Operation Steve Bauman Glenn Research Center, Cleveland, Ohio March 2005 The NASA STI Program Office... in Profile
Z-scan Measurement of the Upconversion Coefficient in Er:YAG
Z-scan Measurement of the Upconversion Coefficient in Er:YAG by Jeffrey O. White, Robert Dibiano, Alexander B. Fick, and John E. McElhenny ARL-TR-4994 September 2009 Approved for public release; distribution
Microstructural Evaluation of KM4 and SR3 Samples Subjected to Various Heat Treatments
NASA/TM 2004-213140 Microstructural Evaluation of KM4 and SR3 Samples Subjected to Various Heat Treatments David Ellis and Timothy Gabb Glenn Research Center, Cleveland, Ohio Anita Garg University of Toledo,
DEFENSE BUSINESS PRACTICE IMPLEMENTATION BOARD
Defense Business Practice Implementation Board DEFENSE BUSINESS PRACTICE IMPLEMENTATION BOARD Report to the Senior Executive Council, Department of Defense MANAGEMENT INFORMATION TASK GROUP Report FY02-3
Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012
Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012 Figure 1.1 illustrates the parts of an upright compound microscope and indicates the terminology that I use in these notes. Figure 1.1. Parts of a Compound
2010 2011 Military Health System Conference
2010 2011 Military Health System Conference Population Health Management The Missing Element of PCMH Sharing The Quadruple Knowledge: Aim: Working Achieving Together, Breakthrough Achieving Performance
Measuring the Point Spread Function of a Fluorescence Microscope
Frederick National Laboratory Measuring the Point Spread Function of a Fluorescence Microscope Stephen J Lockett, PhD Principal Scientist, Optical Microscopy and Analysis Laboratory Frederick National
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND THE THREE-DIMENSIONAL DISTRIBUTION OF THE RADIANT FLUX DENSITY AT THE FOCUS OF A CONVERGENCE BEAM
Synchronization of Unique Identifiers Across Security Domains
Synchronization of Unique Identifiers Across Security Domains by Mark R. Mittrick and Gary S. Moss ARL-MR-712 February 2009 Approved for public release; distribution is unlimited. NOTICES Disclaimers The
Revision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away.
Revision problem Chapter 18 problem 37 page 612 Suppose you point a pinhole camera at a 15m tall tree that is 75m away. 1 Optical Instruments Thin lens equation Refractive power Cameras The human eye Combining
An Implicit Three-Dimensional Meteorological Message for Artillery Trajectory Calculation
An Implicit Three-Dimensional Meteorological Message for Artillery Trajectory Calculation by Patrick A. Haines, David J. Epler II, James L. Cogan, Sean G. O Brien, Benjamin T. MacCall, and Brian P. Reen
SafetyScan Lens. User Manual
SafetyScan Lens User Manual Contents Introduction 2 Lens selection 3 Beam quality parameter Using the tables Use with lasers whose M 2 is 1.2 4 Use with lasers whose M 2 is 3.0 5 Use with lasers whose
Near-field scanning optical microscopy (SNOM)
Adviser: dr. Maja Remškar Institut Jožef Stefan January 2010 1 2 3 4 5 6 Fluorescence Raman and surface enhanced Raman 7 Conventional optical microscopy-limited resolution Two broad classes of techniques
Be careful not to scratch or hit front edge of the side viewing micro prisms onto hard objects!
Instructions Manual Flexia BGA Inspection Systems This manual describes how to use Flexia BGA Inspection System Optilia Instruments 1 AB Contents 1. Safety and maintenance Instructions 3 2. About Flexia
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,
Dynamic IR Scene Projector Based Upon the Digital Micromirror Device
Dynamic IR Scene Projector Based Upon the Digital Micromirror Device D. Brett Beasley, Matt Bender, Jay Crosby, Tim Messer, and Daniel A. Saylor Optical Sciences Corporation www.opticalsciences.com P.O.
CSCI 4974 / 6974 Hardware Reverse Engineering. Lecture 8: Microscopy and Imaging
CSCI 4974 / 6974 Hardware Reverse Engineering Lecture 8: Microscopy and Imaging Data Acquisition for RE Microscopy Imaging Registration and stitching Microscopy Optical Electron Scanning Transmission Scanning
E X A D A T A NEW EXPLOSIVES ACCIDENT DATABASE
E X A D A T A NEW EXPLOSIVES ACCIDENT DATABASE Peter O. Kummer dipl. Ing. HTL/SIA, EUR ING BIENZ, KUMMER & PARTNER LTD, Langägertenstrasse 6 CH-8125 ZOLLIKERBERG, SWITZERLAND Tel: int 41 1 391 27 37 FAX:
- the. or may. scales on. Butterfly wing. magnified about 75 times.
Lecture Notes (Applications of Diffraction) Intro: - the iridescent colors seen in many beetles is due to diffraction of light rays hitting the small groovess of its exoskeleton - these ridges are only
Use the BET (after Brunauer, Emmett and Teller) equation is used to give specific surface area from the adsorption
Number of moles of N 2 in 0.129dm 3 = 0.129/22.4 = 5.76 X 10-3 moles of N 2 gas Module 8 : Surface Chemistry Objectives Lecture 37 : Surface Characterization Techniques After studying this lecture, you
CAPTURE-THE-FLAG: LEARNING COMPUTER SECURITY UNDER FIRE
CAPTURE-THE-FLAG: LEARNING COMPUTER SECURITY UNDER FIRE LCDR Chris Eagle, and John L. Clark Naval Postgraduate School Abstract: Key words: In this paper, we describe the Capture-the-Flag (CTF) activity
Determination of Volatile Organic Compounds Content Using Novel Ash and Total Solids Analysis for Chemical Agent Resistant Coating Systems
Determination of Volatile Organic Compounds Content Using Novel Ash and Total Solids Analysis for Chemical Agent Resistant Coating Systems by Pauline Smith, William Lum, and Daniel Pope ARL-TR-4625 October
Understanding astigmatism Spring 2003
MAS450/854 Understanding astigmatism Spring 2003 March 9th 2003 Introduction Spherical lens with no astigmatism Crossed cylindrical lenses with astigmatism Horizontal focus Vertical focus Plane of sharpest
Scanning Electron Microscope Operation Zeiss Supra-40
Scanning Electron Microscope Operation Zeiss Supra-40 Roger Robbins 9/10/2010 Update: 9/29/2015 Erik Jonsson School of Engineering Title: Scanning Electron Microscope Operation Page 1 of 85 Scanning Electron
EXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab
EXPERIMENT O-6 Michelson Interferometer Abstract A Michelson interferometer, constructed by the student, is used to measure the wavelength of He-Ne laser light and the index of refraction of a flat transparent
Fundamentals of Scanning Electron Microscopy
1 Fundamentals of Scanning Electron Microscopy Weilie Zhou, Robert P. Apkarian, Zhong Lin Wang, and David Joy 1. Introduction The scanning electron microscope (SEM) is one of the most versatile instruments
Usage of AFM, SEM and TEM for the research of carbon nanotubes
Usage of AFM, SEM and TEM for the research of carbon nanotubes K.Safarova *1, A.Dvorak 2, R. Kubinek 1, M.Vujtek 1, A. Rek 3 1 Department of Experimental Physics, Faculty of Science, Palacky University,
THE MIMOSA OPEN SOLUTION COLLABORATIVE ENGINEERING AND IT ENVIRONMENTS WORKSHOP
THE MIMOSA OPEN SOLUTION COLLABORATIVE ENGINEERING AND IT ENVIRONMENTS WORKSHOP By Dr. Carl M. Powe, Jr. 2-3 March 2005 Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden
ELECTRONIC HEALTH RECORDS. Fiscal Year 2013 Expenditure Plan Lacks Key Information Needed to Inform Future Funding Decisions
United States Government Accountability Office Report to Congressional July 2014 ELECTRONIC HEALTH RECORDS Fiscal Year 2013 Expenditure Plan Lacks Key Information Needed to Inform Future Funding Decisions
Award Number: W81XWH-10-1-0739
AD Award Number: W81XWH-10-1-0739 TITLE: Voxel-Wise Time-Series Analysis of Quantitative MRI in Relapsing-Remitting MS: Dynamic Imaging Metrics of Disease Activity Including Prelesional Changes PRINCIPAL
Study Guide for Exam on Light
Name: Class: Date: Study Guide for Exam on Light Multiple Choice Identify the choice that best completes the statement or answers the question. 1. Which portion of the electromagnetic spectrum is used
AFRL-RX-WP-TP-2008-4023
AFRL-RX-WP-TP-2008-4023 HOW KILLDEER MOUNTAIN MANUFACTURING IS OPTIMIZING AEROSPACE SUPPLY CHAIN VISIBILITY USING RFID (Postprint) Jeanne Duckett Killdeer Mountain Manufacturing, Inc. FEBRUARY 2008 Final
MINDS: A NEW APPROACH TO THE INFORMATION SECURITY PROCESS
MINDS: A NEW APPROACH TO THE INFORMATION SECURITY PROCESS E. E. Eilertson*, L. Ertoz, and V. Kumar Army High Performance Computing Research Center Minneapolis, MN 55414 K. S. Long U.S. Army Research Laboratory
Nanoelectronics 09. Atsufumi Hirohata Department of Electronics. Quick Review over the Last Lecture
Nanoelectronics 09 Atsufumi Hirohata Department of Electronics 12:00 Wednesday, 4/February/2015 (P/L 006) Quick Review over the Last Lecture ( Field effect transistor (FET) ): ( Drain ) current increases
Optical Communications
Optical Communications Telecommunication Engineering School of Engineering University of Rome La Sapienza Rome, Italy 2005-2006 Lecture #2, May 2 2006 The Optical Communication System BLOCK DIAGRAM OF
VCR Ion Beam Sputter Coater
VCR Ion Beam Sputter Coater Sputtering Process and Rates 2 Vacuum System 3 Loading the Sputter Chamber 4 Sputter Coating 5 Removing Samples from Chamber 6 Appendix A: VCR High Vacuum Gauge Conditioning
Small PV Systems Performance Evaluation at NREL's Outdoor Test Facility Using the PVUSA Power Rating Method
National Renewable Energy Laboratory Innovation for Our Energy Future A national laboratory of the U.S. Department of Energy Office of Energy Efficiency & Renewable Energy Small PV Systems Performance
PREPARED FOR: U.S. Army Medical Research and Materiel CommandFort Detrick, Maryland 21702 5012
AD Award Number: W81XWH 07 1 0542 TITLE: Is Nuclear Structure Altered in Breast Cancer Cells? PRINCIPAL INVESTIGATOR: Han Htun, Ph.D. CONTRACTING ORGANIZATION: University of California Los Angeles, CA
Laue lens for Nuclear Medicine
Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of
Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012
Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012 You are sitting at your microscope working at high magnification trying to sort out the three-dimensional compartmentalization
Addressing the Real-World Challenges in the Development of Propulsion IVHM Technology Experiment (PITEX)
NASA/CR 2005-213422 AIAA 2004 6361 Addressing the Real-World Challenges in the Development of Propulsion IVHM Technology Experiment (PITEX) William A. Maul, Amy Chicatelli, and Christopher E. Fulton Analex
TS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 Instructions
TS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 ENG Instructions Thank you for purchasing a Canon product. Canon s TS-E lenses are tilt-shift lenses designed for EOS cameras. The tilt-shift mechanism enables
Firearms & Tool Marks Comparison Microscope. Discovery. Leeds Forensic Systems. Your Forensic Imaging Source
Firearms & Tool Marks Comparison Microscope Leeds Forensic Systems Your Forensic Imaging Source Leeds Discovery The Leeds Discovery Firearms & Tool Marks Comparison Microscope is an innovative microscope
Care and Use of the Compound Microscope
Revised Fall 2011 Care and Use of the Compound Microscope Objectives After completing this lab students should be able to 1. properly clean and carry a compound and dissecting microscope. 2. focus a specimen
Elastomer Compatibility Testing of Renewable Diesel Fuels
National Renewable Energy Laboratory Innovation for Our Energy Future A national laboratory of the U.S. Department of Energy Office of Energy Efficiency & Renewable Energy Elastomer Compatibility Testing
73rd MORSS CD Cover Page UNCLASSIFIED DISCLOSURE FORM CD Presentation
73rd MORSS CD Cover Page UNCLASSIFIED DISCLOSURE FORM CD Presentation 21-23 June 2005, at US Military Academy, West Point, NY 712CD For office use only 41205 Please complete this form 712CD as your cover
1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002
05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical
IISUP-. NAVAL SUPPLY SVSTE:MS COMMAND. Ready. Resourceful. Responsive!
~ IISUP-. NAVAL SUPPLY SVSTE:MS COMMAND Ready. Resourceful. Responsive! Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the collection of information is estimated
Metamaterials and Transformation Optics
AFRL-AFOSR-UK-TR-2011-0021 Metamaterials and Transformation Optics John Pendry Imperial College London Physics Department South Kensington Campus Exhibition Road London, United Kingdom SW7 2AZ EOARD GRANT
Application Report: Running µshape TM on a VF-20 Interferometer
: Running µshape TM on a VF-20 Interferometer General This report describes how a fiber interferometer from Arden Photonics Ltd was used together with the µshape TM Generic software package. The VF-20
CERT Virtual Flow Collection and Analysis
CERT Virtual Flow Collection and Analysis For Training and Simulation George Warnagiris 2011 Carnegie Mellon University Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden