User Guide. RDO Optical Dissolved Oxygen Sensor
|
|
|
- Erin McKenzie
- 10 years ago
- Views:
Transcription
1 User Guide RDO Optical Dissolved Oxygen Sensor
2 ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion, phisa, phuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, phuture MMS, phuture Pentrode, phuture Quatrode, phuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable ph, Universal Access are trademarks of Thermo Fisher. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher. PerpHecT meters are protected by U.S. patent 6,168,707. PerpHecT ROSS are protected by U.S. patent 6,168,707. ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results solutions are protected by US Patent 5,830,338. ROSS Ultra electrodes are protected by US patents 6,793,787. Orion ORP Standard is protected by US Patent 6,350,367. Orion NoCal electrodes are protected by US Patent 7,276, Thermo Fisher Scientific Inc. All rights reserved. RDO is a registered trademark of In-Situ Inc., Fort Collins, CO USA. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice. This publication supersedes all previous publications on this subject.
3 Table of Contents Introduction Required Equipment Important Notes Read Before Assembling or Operating the Equipment RDO Sensor Preparation RDO Sensor Calibration Calibration Sleeve Preparation for Water-Saturated Air Calibration Zero Oxygen Solution for Zero Point Calibration Stainless Steel Protective Guard RDO Sensor Measurements RDO Sensor Storage RDO Sensor Maintenance Cleaning the RDO Cap Cleaning the RDO Sensor Body Cleaning the Optical Window Yearly Replacement of the RDO Cap Troubleshooting Troubleshooting Chart Interfering Substances Substances That Damage the RDO Sensor Assistance Warranty WEEE Compliance Ordering Information Specifications RDO Optical Dissolved Oxygen Sensor User Guide
4 II RDO Optical Dissolved Oxygen Sensor User Guide
5 Introduction This user guide contains information on the preparation, operation and maintenance of the Thermo Scientific Orion RDO optical dissolved oxygen sensors. These dissolved oxygen sensors are compatible with a wide range of applications including rugged field applications, water and wastewater analysis, beer and wine testing and BOD (biochemical oxygen demand) analysis. All of the RDO sensors feature automatic temperature compensation and convenient press-on RDO caps. Orion Star RDO Optical Dissolved Oxygen Sensors The Orion Star RDO optical dissolved oxygen sensors have a 9 pin minidin connector that ensures a waterproof and secure connection when used with the Orion Star Plus RDO optical dissolved oxygen meters. Figure 1 RDO Optical Dissolved Oxygen Sensor Temperature Sensor O-rings Spring Contacts Thread Ring Optical Window RDO Cap RDO Optical Dissolved Oxygen Sensor User Guide 1
6 Required Equipment 1. Thermo Scientific Orion Star Plus RDO optical dissolved oxygen meter. 2. Thermo Scientific Orion RDO optical dissolved oxygen sensor. 3. RDO cap, Cat. No Calibration sleeve, Cat. No Stainless steel protective sensor guard (optional), Cat. No Magnetic stirrer or stirrer probe, Cat. No The stirrer probe can be used with 3-Star, 4-Star and 5-Star Plus benchtop RDO meters. 7. Distilled or deionized water. Important Notes Read Before Assembling or Operating the Equipment Note: Do not touch the bottom of the RDO cap, optical window or spring contacts on the RDO sensor. Note: The RDO sensor has an internal clock that counts down the 365 day lifespan of a new RDO cap. The countdown begins when the RDO cap is installed on the RDO sensor, the sensor is connected to the meter and the first measurement is taken. The 365 day countdown cannot be reset or changed once the first measurement is taken. 2 RDO Optical Dissolved Oxygen Sensor User Guide
7 RDO Sensor Preparation 1. Remove the protective shipping cap from the end of the RDO sensor 2. Lubricate the two O-rings on the RDO sensor with a thin film of the silicone grease that is supplied with the sensor. 3. Align the arrow on the RDO cap with the arrow on the RDO sensor. See Figure 2. Figure 2 4. Slide and press the RDO cap onto the RDO sensor until the cap is flush with the sensor body. See Figure 3. Note: Do not twist the RDO cap. Once the cap is installed, the cap should not be removed from the RDO sensor until a new cap needs to be installed or troubleshooting needs to be performed. Figure 3 5. Connect the RDO sensor to the 9 pin minidin input on the meter. See Figure 4. Figure 4 Note: An Err 880 message on the meter indicates that the RDO sensor is not fully connected to the meter. If this error appears, reconnect the sensor by firmly pushing the connector into the 9 pin minidin meter input. Note: An Err 882 message on the meter indicates that the RDO cap is missing or not properly installed on the sensor. RDO Optical Dissolved Oxygen Sensor User Guide 3 Temperature Sensor Thread Cap
8 RDO Sensor Calibration Calibration of an RDO optical dissolved oxygen sensor can be done quickly and conveniently by performing a water-saturated air calibration. Under equilibrium conditions, the partial pressure of oxygen in air-saturated water is equal to the partial pressure of oxygen in water-saturated air. This means that a sensor calibrated in water-saturated air will correctly read the partial pressure of oxygen in water samples. When measuring low concentration samples (less than 1 ppm), a second calibration with a zero oxygen standard is recommended. Calibration Sleeve Preparation for Water-Saturated Air Calibration 1. Remove the cap from the calibration sleeve and remove the sponge from the cap. 2. Saturate the sponge with distilled water and squeeze the excess water out of the sponge. 3. Reassemble the calibration sleeve. 4. Unscrew and remove the thread ring from the RDO sensor. See Figure 5. Slide the calibration sleeve over the front of the RDO sensor and screw the sleeve onto the sensor. 5. Wait at least five minutes for the RDO sensor to equilibrate. 6. Follow the instructions in the meter user guide for performing a water-saturated air calibration. 4 RDO Optical Dissolved Oxygen Sensor User Guide
9 Zero Oxygen Solution for Zero Point Calibration 1. Perform a water-saturated air calibration. 2. Prepare a sodium sulfite solution by dissolving about 15.0 g of Na 2 SO 3 in about 250 ml of distilled water. Transfer the solution to a BOD bottle or flask. Note: A small amount of cobalt salt can be added to the sodium sulfite solution. The cobalt salt will act as an indicator and change color when the sodium sulfite solution no longer has a zero oxygen content. 3. Add a stir bar to the bottle or flask. Insert the RDO sensor into the bottle or flask and use parafilm to seal the open area between the bottle or flask and the sensor. 4. Place the bottle or flask on a magnetic stir plate and gently stir the solution. 5. Wait at least five minutes for the RDO sensor to equilibrate. 6. Follow the instructions in the meter user guide for performing a zero point calibration. 7. Thoroughly rinse the RDO sensor under running water and blot it dry with a lint-free tissue. Note: If the RDO sensor is sluggish or inaccurate after a zero point calibration, not all of the sodium sulfite was removed from the sensor. A very thorough soaking and rinsing of the RDO sensor in distilled water is required to remove all of the sodium sulfite solution and restore the sensor performance. Soak the RDO sensor in distilled water for 30 minutes, blot it dry with a lint-free tissue, use a wash bottle to rinse the sensor with distilled water, soak the sensor for another 30 minutes in fresh distilled water and blot it dry with a lint-free tissue. RDO Optical Dissolved Oxygen Sensor User Guide
10 Stainless Steel Protective Guard The RDO optical dissolved oxygen sensor can be used with the stainless steel protective guard, Cat. No The guard protects the RDO sensor from damage and adds weight to the sensor for low depth measuring applications. Installation of the Protective Guard 1. Unscrew and remove the thread ring from the RDO sensor. See Figure 5. Figure 5 Thread Ring 2. Slide the protective guard over the front of the RDO sensor and screw the guard onto the sensor. See Figure 6. Figure 6 RDO Sensor with Protective Guard Thread Cap Temperature Sensor O-rings Spring Contacts 6 RDO Optical Dissolved Window Oxygen Sensor User Guide RDO Cap
11 RDO Sensor Measurements For accurate measurements, the RDO sensor must be immersed in solution until the temperature sensor is covered by at least a ½ inch of the solution. If any salt or mineral deposits are observed on the RDO sensor, immediately rinse them off with distilled or deionized water. 1. Rinse the RDO sensor with distilled water, blot it dry with a lint-free tissue and insert it into the sample. 2. Follow the instructions in the meter user guide for performing a measurement. 3. Remove the RDO sensor from the sample, rinse it with distilled water, blot it dry with a lint-free tissue, insert the sensor into the next sample and repeat step 2 or refer to the RDO Sensor Storage section for the recommended storage procedures. RDO Sensor Storage Short Term Storage Between measurements and overnight, leave the RDO cap on the RDO sensor and keep the sensor in the calibration sleeve, making sure to wet the sponge with distilled water, or in a beaker with distilled water. Keep the RDO sensor away from direct sunlight during storage. Long Term Storage Leave the RDO cap on the RDO sensor and keep the sensor in the calibration sleeve, making sure to wet the sponge with distilled water, or in a beaker with distilled water. Keep the RDO sensor away from direct sunlight during storage. RDO Optical Dissolved Oxygen Sensor User Guide 7
12 RDO Sensor Maintenance Figure 7 RDO Sensor and RDO Cap Temperature Sensor O-rings Spring Contacts Optical Window RDO Cap Cleaning the RDO Cap 1. Make sure that the RDO cap is properly installed on the RDO sensor and rinse the cap and temperature sections of the sensor with distilled water from a squirt bottle or spray bottle and blot dry with a lint-free soft cloth. 2. Inspect the RDO cap for scratches or discoloration and replace the cap if any flaws are observed. If algal growth is present on the RDO cap, gently clean it with a soft brush. Note: Do not use solvents or soaps to clean the RDO cap and do not rub the cap with abrasive material. Use of these materials will damage the RDO cap and void the warranty. Do not let salt or mineral deposits form on the RDO cap. To remove salt or mineral deposits from the cap, refer to step 2 in the Cleaning the RDO Sensor Body section. Cleaning the RDO Sensor Body 1. Make sure that the RDO cap is properly installed on the sensor and gently scrub the sensor body with a soft brush. 2. Calcareous fouling can normally be dissolved with household vinegar. Disconnect the RDO sensor from the meter and soak the assembled sensor, with the RDO cap installed on the sensor, in vinegar overnight. If the marine growth remains, use clean cotton swabs to gently wipe the growth off after it has been softened by soaking in vinegar. 3. After cleaning the RDO sensor, thoroughly rinse the sensor in distilled water and blot it dry with a lint-free tissue before using or storing it. After cleaning the RDO sensor, a watersaturated air calibration should be performed. 8 RDO Optical Dissolved Oxygen Sensor User Guide
13 Cleaning the Optical Window 1. Remove the RDO cap and gently wipe the optical window with a clean lens cloth. Note: Do not wet the optical window area with water or any other solution. Yearly Replacement of the RDO Cap The RDO cap must be replaced every 365 days. The meter will display an Err 881 message when the RDO cap needs to be replaced. To view the days remaining before the RDO cap must be replaced, refer to the meter user guide. 1. Turn the power to the meter off. 2. Remove the old RDO cap from the RDO sensor. Pull the cap straight off of the sensor and do not twist the cap. 3. Remove the existing two O-rings from the RDO sensor. 4. Remove any moisture from the RDO sensor body with a lint-free tissue. Do not touch the optical window or spring contacts with anything other than a lint-free tissue. Make sure that there is no moisture in the O-ring grooves. 5. Apply a thin film of silicone grease around the O-ring grooves and install the two new O-rings on the sensor. Only use the silicone grease and O-rings that are supplied with the RDO cap. 6. Lubricate the two O-rings on the RDO sensor with a thin film of silicone grease. Do not allow any silicone grease to touch the optical window or spring contacts. 7. Examine the optical window for scratches or dirt and clean it if necessary. To clean the optical window, gently wipe it with a clean lens cloth. 8. Install a new RDO cap onto the RDO sensor by aligning the arrow on the cap with the arrow on the sensor and pressing the cap onto the sensor until the cap is flush with the sensor body. Do not twist the RDO cap. Make sure that the O-rings are not pinched or rolled between the cap and the sensor. 9. Turn the power to the meter on. 10. After replacing the RDO cap, a water-saturated air calibration should be performed. RDO Optical Dissolved Oxygen Sensor User Guide
14 Troubleshooting The most important principle in troubleshooting is to isolate the components of the system and check each in turn. The components of the system include the meter, RDO sensor, sample and technique. Meter The meter is the easiest component to eliminate as a possible cause of error. Thermo Scientific Orion meters include an instrument checkout procedure in the meter user guide. Consult the meter user guide for complete instructions and verify that the meter operates correctly and is stable in all steps. RDO Sensor 1. Rinse the RDO sensor thoroughly with distilled water, blot it dry with a lint-free tissue and inspect the RDO cap for scratches or discoloration. 2. Remove the RDO cap from the RDO sensor and make sure that there is no water inside the cap, the optical window is clean and clear, the O-rings are intact and have a thin coating of silicone grease and the spring contacts are clean and undamaged. 3. If readings continue to be erratic and unstable, the RDO cap may need to be replaced. Refer to the RDO Sensor Maintenance section. Sample If the RDO sensor works properly in standards but not in the sample, look for possible interferences or substances in the sample that could alter the sensor response or physically damage the sensor. If possible, determine the composition of the sample and check for issues. Technique Check if the method of analysis is compatible with your sample. 10 RDO Optical Dissolved Oxygen Sensor User Guide
15 Troubleshooting Chart Issue Unable to calibrate the RDO sensor Dissolved oxygen measurements are unstable Dissolved oxygen measurement is too low Wrong temperature displayed Bad zero point calibration Bad reading after zero point calibration Err 880 error message Err 881 error message Err 882 error message Recommended Action Verify the calibration setup and procedure. Make sure that no water droplets are on the surface of the RDO cap. Perform RDO sensor maintenance. Refer to the Troubleshooting and RDO Sensor Maintenance sections. Measurements may take longer if the solution temperature is unstable. Perform RDO sensor maintenance. Refer to the Troubleshooting and RDO Sensor Maintenance sections. Interfering substance may be present. Salt may be present in the sample. Set the salinity factor in the meter. Perform RDO sensor maintenance. Refer to the Troubleshooting and RDO Sensor Maintenance sections. Verify that the temperature sensor on the RDO sensor is immersed in the solution. Contact Technical Support. The RDO sensor must be in an oxygenfree solution for at least five minutes. Make sure that the open area between the sensor and bottle is covered with parafilm. Soak the RDO sensor in distilled water for 30 minutes, use a wash bottle to thoroughly rinse the sensor with distilled water and then soak the sensor for another 30 minutes in fresh distilled water. The RDO sensor is not fully connected to the meter. Reconnect the RDO sensor. The RDO cap has expired. Install a new RDO cap. Verify that the RDO cap is fully seated on the sensor body. Install a new RDO cap. RDO Optical Dissolved Oxygen Sensor User Guide 11
16 Interfering Substances Alcohols greater than 5%, hydrogen peroxide (H 2 O 2 ) greater than 3%, sodium hypochlorite (NaClO, commercial bleach) greater than 3%, gaseous sulfur dioxide (SO 2 ) and gaseous chlorine (Cl 2 ) may interfere with the dissolved oxygen measurements. If the measurements are unstable and other troubleshooting procedures do not correct the problem, there may be interfering substances present in the solution. Carbon dioxide (CO 2 ), ammonia (NH 3 ), ph, any ionic species like sulfide (S 2- ), sulfate (SO 4-2 ), chloride (Cl - ) and hydrogen sulfide (HS - ) do not interfere with the dissolved oxygen measurements. Substances That Damage the RDO Sensor Do not use the RDO sensor in solutions that contain organic solvents, such as acetone, chloroform or methylene chloride. Assistance After troubleshooting all components of your measurement system, contact Technical Support. Within the United States call and outside the United States call or fax In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit For the latest application and technical resources for Thermo Scientific Orion products, visit Warranty For the most current warranty information, visit 12 RDO Optical Dissolved Oxygen Sensor User Guide
17 WEEE Compliance This product is required to comply with the European Union s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the symbol on the left. We have contracted with one or more recycling/disposal companies in each EU Member State and this product should be disposed of or recycled through them. Further information on compliance with these directives, the recyclers in your country and information on Thermo Scientific Orion products that may assist the detection of substances subject to the RoHS directive are available at Ordering Information Cat. No. Description MD RDO optical dissolved oxygen sensor with 3 meter cable, RDO cap, calibration sleeve and stainless steel protective guard MD RDO optical dissolved oxygen sensor with 6 meter cable, RDO cap, calibration sleeve and stainless steel protective guard MD RDO optical dissolved oxygen sensor with 10 meter cable, RDO cap, calibration sleeve and stainless steel protective guard MD RDO optical dissolved oxygen sensor with 15 meter cable, RDO cap, calibration sleeve and stainless steel protective guard MD RDO optical dissolved oxygen sensor with 30 meter cable, RDO cap, calibration sleeve and stainless steel protective guard Replacement RDO cap Stainless steel protective guard Calibration sleeve RDO Optical Dissolved Oxygen Sensor User Guide 13
18 Specifications Dissolved Oxygen Range Dissolved Oxygen Accuracy Dissolved Oxygen Resolution 0 to 20 mg/l 0 to 200 % ± 0.1 mg/l up to 8 mg/l ± 0.2 mg/l from 8 to 20 mg/l 0.01 mg/l Dissolved Oxygen Response Time 90% of final value in 30 seconds, 95% of final value in 90 seconds Temperature Range 5 to 45 C Temperature Accuracy ± 0.3 C Minimum Immersion Depth Sensor Length (Tip) Sensor Diameter (Tip) Total Sensor Length Total Sensor Diameter Cable Length RDO Cap Expected Lifespan 40 mm 70 mm 20 mm 190 mm 35 mm MD 3 meters MD 6 meters MD 10 meters MD 15 meters MD 30 meters 1 year Specifications are subject to change without notice 14 RDO Optical Dissolved Oxygen Sensor User Guide
19 Thermo Fisher Scientific Environmental Instruments Water Analysis Instruments North America 166 Cummings Center Beverly, MA USA Toll Free: Tel: Dom. Fax: Int l Fax: Europe P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 BN Nijkerk, The Netherlands Tel: (31) Fax: (31) Rev. A Asia Pacific Blk 55, Ayer Rajah Crescent #04-16/24, Singapore Tel: Fax:
User Guide. 9708 Dissolved Oxygen Probes
User Guide 9708 Dissolved Oxygen Probes Introduction The Thermo Scientific Orion 9708 dissolved oxygen probe simplifies measurements of dissolved oxygen, particularly Biochemical Oxygen Demand (BOD).
DISSOLVED OXYGEN PROBE
51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor
Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
HOBO Dissolved Oxygen Logger (U26-001) Manual
HOBO Dissolved Oxygen Logger (U26-001) Manual The HOBO Dissolved Oxygen logger is a standalone logger that uses RDO Basic Technology to measure dissolved oxygen (DO). The logger has an optical sensor that
Orion Star Series Meter
Analyze Detect Measure Control Orion Star Series Meter User s Guide ROSS and the COIL trade dress are trademarks of. AQUAfast, Cahn, EZ Flash, Ionalyzer, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT,
STANDARD OPERATING PROCEDURE Dissolved Oxygen Measurement, Meter Calibration and Maintenance
Page 1 of 9 KEY WORDS- Dissolved Oxygen, calibration, probe preparation, meter preparation APPROVALS APPROVED BY: Kean S. Go, DATE: +/?? APPROVED BY: A Pement DATE: a// A/? f Lisa Ross, EHAP Senior Scientist
Dissolved Oxygen Sensors
Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit
Handbook. Thermo Scientific. ph Electrode Selection Handbook
Handbook Thermo Scientific ph Electrode Selection Handbook Table of Contents ph Theory... 4 ph Electrode Theory... 4 ph versus Temperature Theory... 5 ph Electrode Selection Guide... 6 Thermo Scientific
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table
Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination
Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL 970 350/REV A/10-03 MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1 Installation 2 Displays 2-3 Controls 4 Preparation
6 H2O + 6 CO 2 (g) + energy
AEROBIC RESPIRATION LAB DO 2.CALC From Biology with Calculators, Vernier Software & Technology, 2000. INTRODUCTION Aerobic cellular respiration is the process of converting the chemical energy of organic
Professional Plus. Calibration Tips
Professional Plus Calibration Tips Table of Contents Introduction...1 Calibration Worksheet...1 Temperature...3 Calibration Tips...3 Troubleshooting Tips...3 Conductivity...4 Calibration Tips...4 Troubleshooting
SINGLE RANGE DS METERS Operation Instructions
SINGLE RANGE DS METERS Operation Instructions User Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10 6115 Corte del Cedro Carlsbad, CA 92009-1516 USA Tel 760-438-2021 Fax
User Guide. Thermo Scientific Orion ROSS Ultra, ROSS Ultra Triode, ROSS and PerpHecT ROSS Electrodes
User Guide Thermo Scientific Orion ROSS Ultra, ROSS Ultra Triode, ROSS and PerpHecT ROSS Electrodes ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. U.S. patent 6,793,787.
USEPA Electrode Method Method 8156
ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method
Instruction Manual HI 9146. Portable Waterproof Microprocessor Dissolved Oxygen Meter
Instruction Manual HI 9146 Portable Waterproof Microprocessor Dissolved Oxygen Meter w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction
AIRMoN, ph and Conductivity Field Measurement
AIRMoN, ph and Conductivity Field Measurement Items needed: Sample to be measured (sample volume 50 ml) Field Observer Form (FOF) for the sample to be measured conductance cell ph vials, 4 conductivity
Trace Dissolved Oxygen Sensor
Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION
Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes
Tech Note Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes Probe Care and Maintenance Periodic cleaning and membrane changes are required to keep your sensors operating properly. Follow
Measuring the Conductivity of Pure Water Using USP<645>
Measuring the Conductivity of Pure Water Using USP Water Analysis Instruments, Thermo Fisher Scientific Application Note 002 Key Words Water conductivity, ultrapure water (UPW), USP, water quality,
ph Measurements of Common Substances
Chem 100 Section Experiment 10 Name Partner s Name Introduction ph Measurements of Common Substances The concentration of an acid or base is frequently expressed as ph. Historically, ph stands for the
PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:
PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA
User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
User Manual. Thermo Scientific Orion VERSA STAR
User Manual Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 68X006501 Revision A April 2014 Contents Chapter 1 Meter Introduction... 6 Meter Overview... 6 Packing List... 7
Appendix C. Vernier Tutorial
C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features 2 Quick guide 2 To operate 3 Cleaning 4 Battery replacement 4 Calibration 5 Error messages 5 Troubleshooting guide 6 Technical specifications 6 Information about
Thermo Scientific Rotor Care Guide. Protect your rotor investment
Thermo Scientific Rotor Care Guide Protect your rotor investment Thermo Scientific Rotor Care Guide Thermo Scientific rotors meet the high standards we set for all of our laboratory products. The precision-engineered
Thermo Scientific Orion 2111XP Sodium Analyzer. User Guide
Thermo Scientific Orion 2111XP Sodium Analyzer User Guide ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. U.S. patent 6,793,787. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION,
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide
Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3
Measuring ph in Water or CaCl2 Using a ph Meter Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 PREPARING THE ph METER FOR USE (Calibration and Buffer Solutions) 4 When is Calibration
Dissolved Oxygen Probe (Order Code DO-BTA or DO-DIN)
Dissolved Oxygen Probe (Order Code DO-BTA or DO-DIN) The Dissolved Oxygen Probe can be used to measure the concentration of dissolved oxygen in water samples tested in the field or in the laboratory. You
User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600
User's Guide ExStik DO600 Dissolved Oxygen Meter II RECALL Dissolved Oxygen DO600 Introduction Congratulations on your purchase of the ExStik DO600 dissolved oxygen / temperature meter which simultaneously
Orion High-Level Chloride Monitor
Orion 1817HL Orion High-Level Chloride Monitor INSTRUCTION MANUAL rifos noglib rifos dome glid off a b c d Pure Water MONITORS 1800 series release latches to access Analyze Detect Measure Control AQUAfast,
YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen
YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen 1. Check the log-book to see if somebody else has already checked the meter for accuracy within the timeframes below. If the meter hasn
Model 1811EL Low-Level Sodium Monitor
Model 1811EL Low-Level Sodium Monitor INSTRUCTION MANUAL test values C mv slope E 0 sample values ppm ppb error stable enter/ done sample test program cal. begin cal two pt. cal. complete error add std.
Care and use guide. Contents. page
Care and use guide Contents page Features Bluelab Combo Meter Introduction Preparing for use Calibration Changing nutrient and temperature display units 6 Measuring hydroponic elements 7 Battery replacement
ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.
Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE
Instruction Manual. Image of SP-3015 & SP-3815. Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function
Important Safeguards READ ALL INSTRUCTIONS BEFORE USE. Instruction Manual Automatic Dispensing Hot Water Pot with Reboil Function Image of SP-3015 & SP-3815 SP-3015: 3.0L SP-3815: 3.8L SP-3017: 3.0L (Stainless
User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
Quick Start Guide See Inside for Use and Safety Information
3 rd Generation Personal 3D Printer Quick Start Guide See Inside for Use and Safety Information The USB Mass Storage Device Contains the User Guide and Quick Start Guide in other Languages Congratulations
Winemaker s Guide to Chemistry Analysis
Thermo Scientific Orion Product and Application Guide Winemaker s Guide to Chemistry Analysis Recommended electrochemistry instrumentation and detailed application notes for winemaking Chemistry Analysis
Neutralizing an Acid and a Base
Balancing Act Teacher Information Objectives In this activity, students neutralize a base with an acid. Students determine the point of neutralization of an acid mixed with a base while they: Recognize
Portable Waterproof ph/dissolved Oxygen Meter
Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35632-Series Portable Waterproof /Dissolved Oxygen Meter WTERPROOF REDY MES % T /Dissolved Oxygen/ /
For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588).
User Bulletin ABI PRISM 310 Genetic Analyzer 01/2007 SUBJECT: Syringe Care and Maintenance Overview Safety Information Warranty Information In this User Bulletin This user bulletin describes the care,
OxyGuard. Dissolved Oxygen Probe
OxyGuard Dissolved Oxygen Probe Standard type USER MANUAL D02 Standard Probe manual gb 0903 CONTENTS 1. DESCRIPTION 1.1 The OxyGuard Probe... 1 1.3 Which kind of transmitter can be used with the OxyGuard
Why Expanded Beam types are essential in military and other harsh environment applications
White paper Fiber Optic Connectors Bliek 18 4941 SG RAAMSDONKSVEER NEDERLAND Telephone: +31 162 518 208 Fax: +31 162 518 216 E-mail: [email protected] Website: www.mcap.nl ABN-AMRO: 52.14.96.500 IBAN: NL28ABNA0521496500
PATIENT GUIDE. Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Clinician s Signature: Every drainage Weekly
PATIENT GUIDE Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Every drainage Weekly Clinician s Signature: ACCESS SYSTEMS Pleural Space Insertion Site Cuff Exit Site Catheter
Reasons for reissue are provided in Section 6, REVISION SUMMARY.
Reasons for reissue are provided in Section 6, REVISION SUMMARY. The FRONT OF TOOL has the AMP marking on the link. The BACK OF TOOL (Wire Side), into which the wire is inserted, has the wire size marked
DISSOLVED OXYGEN METER Model : DO-5509
DISSOLVED OXYGEN METER Model : DO-5509 Your purchase of this DISSOLVED OXYGEN METER marks a step forward for you into the field of precision measurement. Although this DISSOLVED OXYGEN METER is a complex
KODAK Developer System Cleaner and Neutralizer
KODAK Developer System Cleaner and Neutralizer TECHNICAL DATA / CHEMICAL March 2010 TI-2000 GENERAL INFORMATION KODAK Developer System Cleaner and Neutralizer is designed to remove the buildup of silver
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL 430 350/REV A/10-03 MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL CONTENTS Introduction 1 Specification 2 Installation 3 Displays 3-4
PRETREATMENT TECHNICAL DATA SHEET WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION
INDUSTRIAL COATINGS RA260 WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION is water emulsifiable rust preventative oil, which provides outstanding corrosion protection for all phosphate metals.
UltraBasic Benchtop ph Meter Operation Manual
UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.
To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.
Titration Curves PURPOSE To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. GOALS 1 To gain experience performing acid-base titrations with a ph meter. 2
Operations Manual. YSI 550A Dissolved Oxygen Instrument
YSI 550A Dissolved Oxygen Instrument Operations Manual CONTENTS CONTENTS...iii GENERAL DESCRIPTION...1 WARRANTY...2 FEATURES OF THE YSI 550A...3 PRINCIPLES OF OPERATION...5 PREPARING THE PROBE...5 DISSOLVED
Global Water Instrumentation, Inc.
Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 Temperature Sensor: WQ101 ph Sensor: WQ201 Conductivity Sensor: WQ301
Household Acids and Bases
Household Acids and Bases GRADE LEVEL INDICATORS Experiment Demonstrate that the ph scale (0-14) is used to measure acidity and classify substances or solutions as acidic, basic, or neutral. 21 Develop
Instruction Manual Temp 4/5/6 Temperature Meter Series
Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.
ExStik TM ph Waterproof Meters
User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL
PART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
How to prepare the Green Cleaning Recipes for a Healthy Home pages:
How to prepare the Recipes for a Healthy Home pages: The following pages, Recipes for a Healthy Home, should be printed back to back on heavy weight paper. The odd number recipes will be on top and the
& CLEANING GUIDE MAINTAIN BEAUTY & PERFORMANCE
& CLEANING GUIDE MAINTAIN BEAUTY & PERFORMANCE BATHS, SHOWERS & WHIRLPOOLS MADE OF SOLID VIKRELL Everything Sterling puts its name on is painstakingly designed and precision-engineered to ensure excellence.
Installation Instructions 4662.003
Installation Instructions.00 SINGLE CONTROL KITCHEN FAUCET Thank you for selecting American-Standard...the benchmark of fine quality for over 00 years. To ensure that your installation proceeds smoothly--please
Char-Lynn Hydraulic Motor. Repair Information. 10 000 Series. October, 1997
Char-Lynn Hydraulic Motor October, 1997 Repair Information Geroler Motor Two Speed 001 27 Retainer inside bore of valve plate bearingless motors only 4 15 16 3 6 35 Parts Drawing 25 2 2 1 19 17 36 40 47
Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information
Chlorine, Total DOC316.53.01028 DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials Scope and application: For testing higher levels of total (free plus combined) chlorine in drinking water,
EZ-PAGE Electrophoresis System USER MANUAL
EZ-PAGE Electrophoresis System USER MANUAL Table of Contents Safety Information.. 2 Product Description... 2 Product Contents..... 3 Specifications & Storage Conditions.. 3 Product Use..... 3 Getting Started
Thermo Scientific Rotor Care Guide. Protect your rotor investment
Thermo Scientific Rotor Care Guide Protect your rotor investment pg 2 pg 4 pg 6 pg 8 Routine Evaluation and Care of Your Rotor Rotor Maintenance Rotor Inspection Services Rotor Life Management Thermo Scientific
2006 JUDY SERVICE GUIDE
2006 JUDY SERVICE GUIDE For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com. Information contained in this publication is subject
CONDUCTIVITY SENSOR. Description D0382. Figure 1. The Conductivity Sensor
CONDUCTIVITY SENSOR Description D0382 Figure 1. The Conductivity Sensor Short Description The Conductivity Sensor can be used to measure either solution conductivity or total ion concentration of aqueous
Chloride and Salinity
INTRODUCTION Chloride Chloride and Chloride, in the form of the Cl ion, is one of the major inorganic anions, or negative ions, in saltwater and freshwater. It originates from the dissociation of salts,
HP Scanjet G4000 series. User Guide
HP Scanjet G4000 series User Guide Contents 1 How to use the scanner...2 Where to get additional information...2 Accessibility...2 How to use the HP Photosmart software...3 Front panel and accessories
STANDARD OPERATING PROCEDURE
STANDARD OPERATING PROCEDURE DATA ACQUISITION USING THE C1 MEMOCONE SYSTEM REVISION DATE: OKT 21, 2011 Prepared by Per Ljunggren TABLE OF CONTENTS STANDARD OPERATING PROCEDURE...3 DATA ACQUISITION USING
Matte Finish Paint Owner s Guide. Matte Finish Paint Owner s Guide. Matte Finish Paint Owner s Guide 3. NP150-V2013 (Rev 06/11/12) Printing xx/xx/12
Matte Finish Paint Owner s Guide Matte Finish Paint Owner s Guide NP150-V2013 (Rev 06/11/12) Printing xx/xx/12 Matte Finish Paint Owner s Guide 3 TABLE OF CONTENTS Special Care Needed Introduction and
SenTix 950 SenTix 980
SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the
User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
Sip/Puff Switch User Guide
Sip/Puff Switch User Guide www.orin.com Table of Contents Table of Contents... 2 Legal Notices... 3 Declaration of Conformity with European Directives... 3 Application Disclaimer... 4 Introduction... 5
Bluelab Soil ph Meter Instruction Manual
Instruction Manual www.getbluelab.com Table of Contents 1.0 Information about measuring the ph of soils/media. 2 2.0 Introduction to Bluelab Soil ph Meter 4 3.0 Preparing the Bluelab Soil ph Meter for
MZ Series Overview. Introduction. (MZ 220 Illustrated)
Thank you for choosing one of our Zebra MZ Series Mobile s. These rugged printers will become productive and efficient additions to your workplace thanks to their innovative design. Because they are made
Ignition Coil Sealing Instructions Revision 1.0
Ignition Coil Sealing Instructions Revision 1.0 Parts Required (available at your local auto parts store): - Dielectric Grease - High Temp RTV Silicone - Waterproof crimp on terminals. 0.250" female, sealed,
Toroidal Conductivity Sensor
Instruction Sheet PN 51A-/rev.C June 2012 Toroidal Conductivity Sensor For additional information, please visit our website at www.emersonprocess.com/rosemountanalytical.com SPECIFICATIONS Wetted Materials:
Refer to page 6 to page 19 for detailed instructions.
CODE:GZ0000297065 Refer to page 6 to page 19 for detailed instructions. Preface This manual describes the operation of the COMPACT CONDUCTIVITY METER, B-771. Be sure to read this manual before using the
BASYS INSTALLATION INSTRUCTIONS FOR ELECTRONIC CAPACITANCE SENSING LAVATORY FAUCETS EFX-3XX Pedestal, Sensor Activated Lavatory Faucets
BASYS INSTALLATION INSTRUCTIONS FOR ELECTRONIC CAPACITANCE SENSING LAVATORY FAUCETS EFX-3XX Pedestal, Sensor Activated Lavatory Faucets Any Application. Any Environment. Code No: 0816407 Rev. 4 (07/12)
Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements
Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
Operator Quick Guide EC SENSOR
Operator Quick Guide EC SENSOR Revision G - 24/04/2008 General Information About This Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra
For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com.
For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com. 2 0 0 5 D U K E A I R X C / S L / R A C E S E R V I C E G U I D E Information
Eutech 2700 Series. Bench Meters. ph ORP ION Conductivity TDS Salinity Resistivity Dissolved Oxygen Temperature
Eutech 2700 Series Bench Meters ph ORP ION Conductivity TDS Salinity Resistivity Dissolved Oxygen Temperature Features Comprehensive Backlit Display All Measurements Include Temperature Display Up To 500
AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU680 Weekly (Analyzer and ISE) Maintenance
AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance Step 1 Action Perform the Manually Clean the ISE Mix Bar, Liquid Level Sensors, Sample Pot and Sample Pot Tubing procedure AU680 Weekly (Analyzer
Owner s Guide Brugervejledning Bedienungsanleitung Guía de usuario Omistajan opas Notice d utilisation Felhasználói útmutató Manuale di istruzioni
Owner s Guide Brugervejledning Bedienungsanleitung Guía de usuario Omistajan opas Notice d utilisation Felhasználói útmutató Manuale di istruzioni Gebruiksaanwijzing Podręcznik użytkownika Bruksanvisningen
The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.
i800 AC Current Clamp Instruction Sheet Introduction The i800 AC Current Clamp, the Clamp, has been designed for use with multimeters, recorders, power analyzers, safety testers, etc., for accurate non-intrusive
Coating and Extraction of Honeycomb Denuders
QSM Approval: Coating and Extraction of Honeycomb Denuders 1 INTRODUCTION 1.1 The following procedures are used for the coating of honeycomb denuders with citric acid and sodium carbonate solutions, and
ph Measurement and Control
ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components
KUSTER ELECTRONIC GAUGES
Kuster Company 2900 E. 29th Street Long Beach, CA USA 90806 Telephone 562-595-0661 Fax 562-426-7897 E-Mail [email protected] KUSTER ELECTRONIC GAUGES K8 EMR 25, K8 EMR 39, K4 EMR ¾, K4 EMR ¾ QUARTZ SERVICE
3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE
4 3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE A. Changing the Control Board B. Replacing a Heating Element C. Thermistor Replacement D. High Limit Switch Replacement E. Level Detector Replacement
ClikSTAR - Important facts about your new insulin delivery device.
ClikSTAR - Important facts about your new insulin delivery device. Instruction for Use ClikSTAR Insulin delivery device Before you start: Read these instructions and follow them completely each time you