Undulators at PETRA: Experience and Perspectives
|
|
|
- Merry Stewart
- 10 years ago
- Views:
Transcription
1 Undulators at PETRA: Experience and Perspectives Markus Tischer for the Insertion Devices Group at DESY 3-Way Meeting, Argonne, Aug 1 st - 2 nd 2013
2 Outline > PETRA III IDs > Commissioning experience > Radiation Damage Issues > PETRA III Extension IDs Coworkers: A. Schöps, G.Naulin P. Vagin, S. Tripathi, O. Zeneli P. Neumann, T. Vielitz, T. Cheung M. Tischer 3-Way Meeting, Page 2
3 Parameters and Spectral Performance of PETRA III Undulators U29_5m U29 U32 U23 UE65 * U19 U32_10m Minimum magnetic gap [mm] Period length λ U [mm] Length L [m] Periods Peak field B 0 [T] Deflection parameter K max st Harmonic E 1 [kev] Total power P tot [kw] On-axis power dens. [kw/mrad 2 ] Power in 1x1mm 2 at 40m [W] High-β source (10keV) size : 140 x 5.6 µm 2 divergence : 7.9 x 4.1 µrad 2 Low-β source (10keV) size : 36 x 6.1 µm 2 divergence : 28 x 4.0 µrad 2 * in helical mode M. Tischer 3-Way Meeting, Page 3
4 Operation Experience at PETRA III > Only small closed orbit distortions Corrected by air coils (CODs ±20 / ±5µm (horiz./vertical) No problems with injection at closed gap > Gap dependent pointing direction Some very sensitive experiments required sub-µrad corrections to assure pointing accuracy > Tune shift of UE65 (APPLE2) 3kHz, 0.5kHz (horiz., vertical), shift dependent 3x reduced by dynamic multipole shims Further reduction possible by feed-forward (not needed at present) > Radiation Damage at several undulators Known since quite a while M. Tischer 3-Way Meeting, Page 4
5 Radiation Damage Issues > Summary of previous observations High dose rate in the tunnel (~kgy per 5 weeks on TLDs) Damaged components (several motor and linear encoders at IDs, various other electronics) Dosis within 1 run [Sv] PETRA III ID Dosis "New Oktant" assorted selection of different run periods (~4-5 weeks) 2012_6 2013_3 2013_4 SE PU01SA PU01SE PU02SE PU03SA PU03SE PU04SA PU04SE PU05SE PU06SA PU06SE PU07SA PU07SE PU08SE PU09SA PU09SE PU10SA PU10SE PU11SE PU12SA PU12SE PU13SE PU14SA PU14 PU01 PU07 PU PETRA III Dose Rate "Undulators in New Octant" (starting at installation date of particular ID) Grand Total [ksv] M. Tischer 3-Way Meeting, Page 5 PU01 PU07 PU14 SE PU01SA PU01SE PU02SE PU03SA PU03SE PU04SA PU04SE PU05SE PU06SA PU06SE PU07SA PU07SE PU08SE PU09SA PU09SE PU10SA PU10SE PU11SE PU12SA PU12SE PU13SE PU14SA PU14
6 Radiation Damage Issues > Summary of previous observations Strong dependence of dose rate as function of vertical angle through the upstream dipole (SR) Corrosion of magnet structure due to radiochemistry (humidity problem solved) BeamLossMonitors (pin diodes) were installed at all IDs as online-diagnostics (but not usable by normal shift operators) BLMs Dipole Absorber B (horizontal) Taper T (vertical) 9mm Undulators 7mm aperture 6m M. Tischer 3-Way Meeting, Page 6
7 Radiation Damage Issues Gap [mm] Undulator Gap of PU01 vs Time 14.3 tuning at E=14.4keV PU01a 14 PU01b Date Gap [mm] Undulator Gap of PU02 vs Time tuning at E=60keV Date Minimum Gap = 9.95 mm (Reached in December 2013) > Gap decrease for a desired energy Systematic decrease of gap for a particular energy over time Gap optimized for max. throughput for otherwise same beamline settings Observed at some beamlines which do not change the setup frequently M. Tischer 3-Way Meeting, Page 7
8 Radiation Damage Issues > Spectral Degradation Broadened undulator harmonics, even with satellites Intensity loss Spoiled line shape can only be partly cured by applying a taper At PU08: Upstream end suffered magnetization loss (corresponding to a negative taper for correction) Total intensity drop by at least a factor of 2 since installation 5 th harm. line shape for different taper at undulator PU08 M. Tischer 3-Way Meeting, Page 8
9 Radiation Damage Issues > Magnetic Measurements Performed in the tunnel at the installed ID Hall probe moved manually over the magnet structure in a carriage Comparison of peak field values Coarse measurement, but error bar below <0.2% M. Tischer 3-Way Meeting, Page 9
10 Radiation Damage Issues > Magnetic Measurements (cont d) Both 5m IDs at the beginning of the new octant (2012): M. Tischer 3-Way Meeting, Page 10
11 Radiation Damage Issues > BLMs using Cherenkov Radiation in optical fibers Present test-setup at 2 straight sections in PETRA (under construction) Well established system at FLASH IDs and other locations (e.g. DIPAC (2009) 411; EPAC (2008) 1032) Principle: Radiation shower is converted into Cerenkov light Detection by photomultiplier + counter and/or fast ADC ( dose rate) Position-sensitive Extremely high temporal resolution (depending on fiber length) Will allow to distinguish different operation states INTEGRATOR TIME MEASUREMENT PHOTODETECTOR IRRADIATION INTENSITY POSITION FIBER OPTICAL CABLE ACCELERATOR M. Körfer, 2001 > No consistent picture up to now Only particular IDs are affected No correlation to visible corrosion patterns TLD results do not match well to the damage locations BLM and TLD measurements agree only qualitatively mechanism not yet understood, but both electrons and SR are involved at different locations F.Wulf, M.Körfer, DIPAC (2009) 411 M. Tischer 3-Way Meeting, Page 11
12 Undulators for PETRA III Extension BL length period energy technique P61 10 x 4 m 200 mm kev mater. sci., PDF P62 N = 12 Tbd 6-35 kev SAXS P63 N = mm 3-44 kev micro XRF P64 2 m 32 mm 3-44 kev (Q)EXAFS P65 N = mm 4-44 kev XAFS BL length period energy technique P21a 4 m IVU kev mater. sci. P21b 2 m wiggler kev imaging P21c 2 m 29mm ~100 kev mater. sci. P22 2 m tbd 3-50 kev nano XRF P23 2 m tbd 5-50 kev nano XRD P24 N = 12 tbd 8-40 kev (Q)EXAFS P25 N = 12 tbd tbd education ID access Vehicle for undulator transport in the tunnel (several 100 meters) due to lack of crane access. M. Tischer 3-Way Meeting, Page 12
13 Undulators for PETRA III Extension > Boundaries for ID design, transport, operation Various obstacles for transport in the present tunnel (existing installations, cable trays, doors, steps) Different beam height in the new arcs (1400mm) and in the old tunnel (~1.2m) need for 2 different support structures Temperature control: ~slim concept with potential for upgrade > Undulator design Same concept for FLASH II and PETRA Extension IDs Various mechanical improvements (stiffness, adjustment, floor mount) Revised drive system: 2 motors and 2 pairs of left/right-handed spindles Additional chamber touch sensors New linear encoder gap measurement system M. Tischer 3-Way Meeting, Page 13
14 Linear Encoder Systems (FLASH II) Rotary motor encoders - systematic deviation reflects magnetic forces, compensated by feed-forward correction - resulting accuracy below 10µm, i.e. very precise drive mechanics Linear encoders - different versions built & tested - open Renishaw system - hysteresis reduced to less than 1µm - reproducibility is much better (sub-µm) 1 µm M. Tischer 3-Way Meeting, Page 14
15 Undulators for PETRA III Extension > Developments for magnetic measurements and tuning Touch probe measurement of pole contour 3D metrology with the measurement bench for mechan. characterization and transfer measurements Investigation of Hall probe calibration errors, wire probe etc. Use additional pole movements for correction of higher order magnetic errors > Improved tuning results for FLASH II IDs Reduced phase error Enhanced trajectory straightness Smaller gap dependence of field integrals Local correction of quadrupole errors Beneficial improvements in particular for P3 extension IDs M. Tischer 3-Way Meeting, Page 15
16 Undulator Source for EXAFS > EXAFS requires frequent scanning of the energy, i.e. gap! Tapered Undulator? > Tapered Undulator: Spatial distribution changes with energy! M. Tischer 3-Way Meeting, Page 16
17 Undulator Source for EXAFS > Simulated EXAFS with Tapered Undulator Inhomogeneous test sample: - Zn-K edge = 9669 ev - Area: 700 µm x 700 µm - Gap: 350 µm µd I = ln I 0 = ln I 0 Courtesy of O.Müller, Univ. Wuppertal ( ) ( ) ( ) ( ) I0 E, x, y da µ E, x, y d x, y E, x, y e da Non-uniform illumination of an inhomogeneous sample large artefacts in the spectrum (show stopper) M. Tischer 3-Way Meeting, Page 17
18 Undulator Source for EXAFS > Real EXAFS Measurement with Tapered Undulator Courtesy of O.Müller, Univ. Wuppertal Undulator P06: - E Undulator = ev - Taper = 1 mm - λ U = 31.4 mm; L=2m - Source Distance = 90 m > Tapered spectrum shows sub-structure > XAFS spectra strongly depend on the orientation of the test sample > Tapered undulator is an inadequate source for (most) XAFS applications On-the-fly measurements with synchronized monochromator and undulator > Wiggler not possible due to power restrictions, short wiggler M. Tischer as bad 3-Way Meeting, as tapered undulator Page 18
19 Energy Scan Monochromator vs. Undulator > Energy Scan 20 ev / sec (und. velocity control OK) > Preliminary tests very promising, next: Include the undulator velocity control in a TANGO device Treat undulator gap as slave of the Bragg axis Feeding the Bragg-encoder signals directly into the undulator NC (still the previous status) A. Schöps G. Wellenreuther M. Tischer 3-Way Meeting, Page 19
20 Damping Wigglers for PETRA III Extension PU61 92m Damping section North 6 m long final absorber M. Tischer 3-Way Meeting, Page 20
21 Damping Wigglers: Parameters Magnet structure Peak field B T Magnetic gap 24 mm Period length 0.2 m Number of poles ½ Magnet volume / period 2200 cm 3 Field quality at x 0 = 10 mm < 10 3 Damping integral / segment 3.9 T 2 m SR characteristics SR critical energy 35.8 kev K -Parameter 28.4 Wiggler SR power* 21 kw Vertical SR spread 170 µrad Horizontal SR spread 4.84 mrad * I=100mA Overall length / segment 3.97 m Number of wiggler segments (West + North) NdFeB magnets built in collaboration with BINP Proceedings EPAC08, Genoa (2008) 2317 B r = 1.3T H cj = 1430kA/m (17.9kOe) M. Tischer 3-Way Meeting, Page 21
22 Damping Wigglers: Magnetic Design End plate Pole Long. magnet Non-magnetic spacer Side magnet Iron yoke Corrector bolt Iron yoke Wedged poles 290 Modified hybrid structure Wedge-shaped iron poles also powered by side magnets Iron enclosure Magnetic yoke Mechanical support Iron spacer 262 Aluminium spacer Zero potential surface Axial magnets split into 2 halves by soft iron V-notches sitting on the yoke almost no cross talk between poles M. Tischer 3-Way Meeting, Page 22
23 Conclusion > PETRA III Reliable operation of all IDs Several magnet structures suffer from radiation damage; investigation will continue IVU for PU07 in 2014 > PETRA Extension IDs ID parameters ~specified, construction partly started, even more safety margin needs to be included in the magnet design IVU for PU21 EXAFS-IDs shall be undulators > Measurement Lab Assembly and tuning of 12 SASE IDs for FLASH II presently ongoing In this context, improvements and extensions of calibration and measurement techniques and also tuning procedures M. Tischer 3-Way Meeting, Page 23
Damping Wigglers in PETRA III
Damping Wigglers in PETRA III WIGGLE2005, Frascati 21-22.2.2005 Winni Decking, DESY-MPY Introduction Damping Wiggler Parameters Nonlinear Dynamics with DW Operational Aspects Summary DESY and its Accelerators
Undulators and wigglers for the new generation of synchrotron sources
Undulators and wigglers for the new generation of synchrotron sources P. Elleaume To cite this version: P. Elleaume. Undulators and wigglers for the new generation of synchrotron sources. Journal de Physique
A Quick primer on synchrotron radiation: How would an MBA source change my x-ray beam. Jonathan Lang Advanced Photon Source
A Quick primer on synchrotron radiation: How would an MBA source change my x-ray beam Jonathan Lang Advanced Photon Source APS Upgrade - MBA Lattice ε ο = 3100 pm ε ο = 80 pm What is emi7ance? I don t
ESRF Upgrade Phase II: le nuove opportunitá per le linee da magnete curvante
LUCI DI SINCROTRONE CNR, ROMA 22 APRILE 2014 ESRF Upgrade Phase II: le nuove opportunitá per le linee da magnete curvante Sakura Pascarelli [email protected] Page 2 INCREASE IN BRILLIANCE H emittance V emittance
Information about the T9 beam line and experimental facilities
Information about the T9 beam line and experimental facilities The incoming proton beam from the PS accelerator impinges on the North target and thus produces the particles for the T9 beam line. The collisions
Lecture VI Magnetic Design
Lecture VI Magnetic Design Field Quality Adjustment After Construction (or After the Start of Production Run) Ramesh Gupta Superconducting Brookhaven National Laboratory US Particle Accelerator School
Measurement of RF Emissions from a Final Coat Electronics Corrosion Module
Engineering Test Report No. 37802-02 Rev. A Measurement of RF Emissions from a Final Coat Electronics Corrosion Module For : Canadian Auto Preservation, Inc. 390 Bradwick Drive Concord, Ontario CANADA
Insertion Devices Lecture 4 Permanent Magnet Undulators. Jim Clarke ASTeC Daresbury Laboratory
Insertion Devices Lecture 4 Permanent Magnet Undulators Jim Clarke ASTeC Daresbury Laboratory Introduction to Lecture 4 So far we have discussed at length what the properties of SR are, when it is generated,
Robot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
Catherine LeCocq, Robert Ruland. February, 2008
Status Report on Alignment Activities at SLAC Catherine LeCocq, Robert Ruland Budget Situation SSRL / SPEAR PEPII / Babar LCLS Instrumentation / Calibration 1 Alignment Activities for SPEAR Ring mapping
Practical Alignment Tools and Techniques
Practical Alignment Tools and Techniques Bruce Lehmann Sr. Engineer, Thin Kerf Technologies Inc Accurate sawing requires attention to proper alignment and maintenance. Alignment is critical to modem sawmills
E190Q Lecture 5 Autonomous Robot Navigation
E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator
The half cell of the storage ring SESAME looks like: Magnetic length =
6. Magnets 6.1 Introduction SESAME will be perhaps erected in steps. The number of steps depends upon the available money. The cheapest way is to use the quadrupole and sextupoles from BESSY I. In this
Cross-beam scanning system to detect slim objects. 100 mm 3.937 in
891 Object Area Sensor General terms and conditions... F-17 Related Information Glossary of terms... P.1359~ Sensor selection guide...p.831~ General precautions... P.1405 PHOTO PHOTO Conforming to EMC
Digital vs. Analogue Control Systems
Digital vs. Analogue Control Systems Presented at the 2011 Annual Meeting of the American College of Medical Physics, Chattanooga, TN, May 1, 2011 Ivan A. Brezovich, PhD, Dept. of Rad Onc, Univ of Alabama
Brookhaven Magnet Division - Nuclear Physics Program Support Activities
Brookhaven - Nuclear Physics Program Support Activities Spin Program support E-Cooling R&D RIA RHIC Operations Support Funding AGS Warm Siberian Snake AGS Warm Snake Magnetic element designed and built
Zero Degree Extraction using an Electrostatic Separator
Zero Degree Extraction using an Electrostatic Separator L. Keller Aug. 2005 Take another look at using an electrostatic separator and a weak dipole to allow a zero degree crossing angle a la the TESLA
SpectraTec II. Polarized Multi-Laser Source BLUE SKY RESEARCH WAVELENGTHS. The SpectraTec II
BLUE SKY RESEARCH The SpectraTec II, two wavelength laser module is a highly integrated system comprised of two lasers, individual driving and temperature control electronics, wavelength combining, and
Measurement of Charge-to-Mass (e/m) Ratio for the Electron
Measurement of Charge-to-Mass (e/m) Ratio for the Electron Experiment objectives: measure the ratio of the electron charge-to-mass ratio e/m by studying the electron trajectories in a uniform magnetic
Development of Radiation Resistant Quadrupoles Based on High Temperature Superconductors for the Fragment Separator
Development of Radiation Resistant Quadrupoles Based on High Temperature Superconductors for the Fragment Separator R. Gupta and M.Harrison, Brookhaven National Laboratory A. Zeller, Michigan State University
Shielding and Radiation Measurements at ESRF
Shielding and Radiation Measurements at ESRF Radiation Safety Meeting SOLEIL Synchrotron, 18 October 26 P. Berkvens European Synchrotron Radiation Facility Radiation protection policy at the ESRF Storage
We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement
MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters
ENGINEERING METROLOGY
ENGINEERING METROLOGY ACADEMIC YEAR 92-93, SEMESTER ONE COORDINATE MEASURING MACHINES OPTICAL MEASUREMENT SYSTEMS; DEPARTMENT OF MECHANICAL ENGINEERING ISFAHAN UNIVERSITY OF TECHNOLOGY Coordinate Measuring
International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters
www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in
CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS
CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS E. Batzies 1, M. Kreutzer 1, D. Leucht 2, V. Welker 2, O. Zirn 1 1 Mechatronics Research
L-LAS-TB-CL serie. laser light curtains for inline measuring tasks
L-LAS-TB-CL serie laser light curtains for inline measuring tasks Maximum distance 2.000 mm Measurement range 6 up to 98 mm Resolution up to 4 µm up to 1,5 khz scan rate 0-10 V or 4-20 ma analogue output
5-Axis Test-Piece Influence of Machining Position
5-Axis Test-Piece Influence of Machining Position Michael Gebhardt, Wolfgang Knapp, Konrad Wegener Institute of Machine Tools and Manufacturing (IWF), Swiss Federal Institute of Technology (ETH), Zurich,
BB-18 Black Body High Vacuum System Technical Description
BB-18 Black Body High Vacuum System Technical Description The BB-18 Black Body is versatile and is programmed for use as a fixed cold target at 80 K or variable target, at 80 K- 350 K no extra cost. The
Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection
w Technical Product Information Precision Miniature Load Cell with Overload Protection 1. Introduction The load cells in the model 8431 and 8432 series are primarily designed for the measurement of force
National Performance Evaluation Facility for LADARs
National Performance Evaluation Facility for LADARs Kamel S. Saidi (presenter) Geraldine S. Cheok William C. Stone The National Institute of Standards and Technology Construction Metrology and Automation
FORCE ON A CURRENT IN A MAGNETIC FIELD
7/16 Force current 1/8 FORCE ON A CURRENT IN A MAGNETIC FIELD PURPOSE: To study the force exerted on an electric current by a magnetic field. BACKGROUND: When an electric charge moves with a velocity v
Cathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
PCB Component Placement Inspection
Executive Summary PCB Component Placement Inspection Optimet s ConoProbe Mark10 HD with a 50 mm focal length lens was used to inspect PCB component placement. The PCB board inspected contained both individual
Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows
LS-296 Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows T. L. Smith ASD / RF Group Advanced Photon Source Argonne National Laboratory June 26, 2002 Table of Contents 1) Introduction
Making Better Medical Devices with Multisensor Metrology
Making Better Medical Devices with Multisensor Metrology by Nate J. Rose, Chief Applications Engineer, Optical Gaging Products (OGP) Multisensor metrology is becoming a preferred quality control technology
A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS
A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS Joseph J. Stupak Jr, Oersted Technology Tualatin, Oregon (reprinted from IMCSD 24th Annual Proceedings 1995) ABSTRACT The
HIGH PRODUCTION DUAL DRIVE HORIZONTAL MACHINING CENTER
HIGH PRODUCTION DUAL DRIVE HORIZONTAL MACHINING CENTER 14001 認 可 登 錄 14001 認 可 登 錄 1 Introducing all new H630B high speed horizontal machining center. Incorporates state of the art Dual Drive technology
The accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
Prelab Exercises: Hooke's Law and the Behavior of Springs
59 Prelab Exercises: Hooke's Law and the Behavior of Springs Study the description of the experiment that follows and answer the following questions.. (3 marks) Explain why a mass suspended vertically
Wir schaffen Wissen heute für morgen
Diffractive optics for photon beam diagnostics at hard XFELs Wir schaffen Wissen heute für morgen PSI: SLAC: ESRF: SOLEIL: APS: SACLA: EuroXFEL C. David, S. Rutishauser, P. Karvinen, Y. Kayser, U. Flechsig,
Bar Code Label Detection. Size and Edge Detection APPLICATIONS
Bar Code Label Detection The EE-SY169(-A) and EE-SX199 sensors are used for bar code label detection. The EE-SX199 detects the absence or presence of labels (see Bar Code Printer illustration at right).
WOOD WEAR TESTING USING TRIBOMETER
WOOD WEAR TESTING USING TRIBOMETER Prepared by Duanjie Li, PhD 6 Morgan, Ste156, Irvine CA 92618 P: 949.461.9292 F: 949.461.9232 nanovea.com Today's standard for tomorrow's materials. 2015 NANOVEA INTRO
Products. Air Bearing- Linear systems 06-2009 EITZENBERGER. Luftlagertechnik GmbH. Feichtmayrstraße 17 D-82405 Wessobrunn
Products Air Bearing- Linear systems 06-2009 ITZNBRGR D- mail: URL. Index - Air Bearing linear Systems Z 0526 Air Bearing Linear Stage with high resolution encoder and iron free linear motor Z 0512 XY-Air
COMBIMASS. Technical Data COMBIMASS eco-bio +
COMBIMASS Technical Data THE SYSTEM COMBIMASS The field transmitters of the COMBIMASS eco series are suitable for gas flow measurement and cover a wide range of different applications. The instruments
Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
High Performance Visible Laser Diode Scanners
A B ALLEN-BRADLEY High Performance Visible Laser Diode Scanners (Catalog No. 27-LD4z1, -LD4z4 27-LD8z1, -LD8z4) Product Data These high performance, fixed-mount scanners use a Visible Laser Diode (VLD)
Review Questions PHYS 2426 Exam 2
Review Questions PHYS 2426 Exam 2 1. If 4.7 x 10 16 electrons pass a particular point in a wire every second, what is the current in the wire? A) 4.7 ma B) 7.5 A C) 2.9 A D) 7.5 ma E) 0.29 A Ans: D 2.
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.
Magnetic Field and Magnetic Forces
Chapter 27 Magnetic Field and Magnetic Forces PowerPoint Lectures for University Physics, Thirteenth Edition Hugh D. Young and Roger A. Freedman Lectures by Wayne Anderson Goals for Chapter 27 Magnets
Status of the FERMI@Elettra Free Electron Laser
Status of the FERMI@Elettra Free Electron Laser E. Allaria on behalf of the FERMI team Work partially supported by the Italian Ministry of University and Research under grants FIRB-RBAP045JF2 and FIRB-RBAP06AWK3
Series: IDAM Servo Drive E Digital Motor Drive - DMD-078.12
Series: IDAM Servo Drive E Digital Motor Drive - DMD-078.12 inside Integrated amplifiers for 8 single-phase motors, 4 two-phases motors or 4 three-phases motors or combinations of them in one device Position
[ means: Save time, money and space! MAXXMILL 500. Vertical milling center for 5-side machining
[ E[M]CONOMY] means: Save time, money and space! MAXXMILL 500 Vertical milling center for 5-side machining MAXXMILL 500 MAXXMILL 500 is the ideal vertical milling center for the for the 5-axis operation
Collision Prevention and Area Monitoring with the LMS Laser Measurement System
Collision Prevention and Area Monitoring with the LMS Laser Measurement System PDF processed with CutePDF evaluation edition www.cutepdf.com A v o i d...... collisions SICK Laser Measurement Systems are
Specifications for DRIVER S REAR VIEW CAMERA
Specifications for DRIVER S REAR VIEW CAMERA August 2001 Sekai Electronics Inc. 13570 Larwin Circle Santa Fe Springs CA-90670 U.S.A. General Description Introduction A camera system designed to assist
Types of 3D Scanners and 3D Scanning Technologies.
Types of 3D Scanners and 3D Scanning Technologies. There are many types of 3D scanners and 3D scanning technologies. Some are ideal for short range scanning while others are better suited for mid or long
Precise heavy-duty cutting
PR 130 / PR 150 / PR 160 / PR 180 / PR 200 / PR 260 Precise heavy-duty cutting P-SERIES K-SERIES T-SERIES MILLFORCE 02 / Fields of application 03 / Machine concept 04 / Machine technology 05 / Options
Three Channel Optical Incremental Encoder Modules Technical Data
Three Channel Optical Incremental Encoder Modules Technical Data HEDS-9040 HEDS-9140 Features Two Channel Quadrature Output with Index Pulse Resolution Up to 2000 CPR Counts Per Revolution Low Cost Easy
Fraunhofer Diffraction
Physics 334 Spring 1 Purpose Fraunhofer Diffraction The experiment will test the theory of Fraunhofer diffraction at a single slit by comparing a careful measurement of the angular dependence of intensity
Atomic Force Microscope
Atomic Force Microscope (Veeco Nanoman) User Manual Basic Operation 4 th Edition Aug 2012 NR System Startup If the system is currently ON To start the NanoScope software, double-click the NanoScope startup
Atomic Force Microscope and Magnetic Force Microscope Background Information
Atomic Force Microscope and Magnetic Force Microscope Background Information Lego Building Instructions There are several places to find the building instructions for building the Lego models of atomic
Traffic Monitoring Systems. Technology and sensors
Traffic Monitoring Systems Technology and sensors Technology Inductive loops Cameras Lidar/Ladar and laser Radar GPS etc Inductive loops Inductive loops signals Inductive loop sensor The inductance signal
Module 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
Status of the SOLEIL project Commissioning from Linac to beamlines
Status of the SOLEIL project Commissioning from Linac to beamlines On behalf of the commissioning team 2 D01-1-CX1/DT/DTC/absorption Y Axis Title 0-200 0 200 400 600 800 1000 1200 1400 X Axis Title 1 Site
DUGARD. Machine Tools Since 1939. Dugard 400 Slant Bed High Precision CNC Lathe. www.dugard.com
DUGARD Machine Tools Since 1939 Dugard 400 Slant Bed High Precision CNC Lathe www.dugard.com Superb Performance, Maximum Stability, Maximum Reliability Precision, Power and Capacity Make the Dugard 400
Development of on line monitor detectors used for clinical routine in proton and ion therapy
Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
Electromagnetic Compatibility Test Report Test results of Floww equipment
Electromagnetic Compatibility Test Report Test results of Floww equipment Reference number Status test report Brand Model number : 10C00357RPT01.doc : Final : Floww : mobilefloww screenfloww pocketfloww
Short overview of TEUFEL-project
Short overview of TEUFEL-project ELAN-meeting may 2004 Frascati (I) Contents Overview of TEUFEL project at Twente Photo cathode research Recent experience Outlook Overview FEL Drive laser Photo cathode
Alternative Linear Motion Systems. Iron Core Linear Motors
Alternative Linear Motion Systems ME EN 7960 Precision Machine Design Topic 5 ME EN 7960 Precision Machine Design Alternative Linear Motion Systems 5-1 Iron Core Linear Motors Provide actuation forces
Testing thermo-acoustic sound generation in water with proton and laser beams
International ARENA Workshop DESY, Zeuthen 17th 19th of May 25 Testing thermo-acoustic sound generation in water with proton and laser beams Kay Graf Universität Erlangen-Nürnberg Physikalisches Institut
How To Use An Ionsonic Microscope
Sample Analysis Design Element2 - Basic Software Concepts Scan Modes Magnetic Scan (BScan): the electric field is kept constant and the magnetic field is varied as a function of time the BScan is suitable
IP-S2 Compact+ 3D Mobile Mapping System
IP-S2 Compact+ 3D Mobile Mapping System 3D scanning of road and roadside features Delivers high density point clouds and 360 spherical imagery High accuracy IMU options without export control Simple Map,
3D laser removal and texturing in a new dimension
www.dmgmori.com LASERTEC 45 Shape 3D material removal by laser Laser texturing LASERTEC 45 Shape 3D laser removal and texturing in a new dimension LASERTEC LASERTEC 45 Shape: High-precision 3D laser removal
Solar Tracking Application
Solar Tracking Application A Rockwell Automation White Paper Solar trackers are devices used to orient photovoltaic panels, reflectors, lenses or other optical devices toward the sun. Since the sun s position
SUN SIMULATION AT HIGH FLUX LEVEL
SUN SIMULATION AT HIGH FLUX LEVEL Paul-Eric DUPUIS, Thibault RAVILY, Emmanuel SARAIVA, David LAFFONT, Nicolas GROTTO, Fabien CABARET (1) (1) INTESPACE, 2 rond point Pierre Guillaumat, BP 64356,31029 Toulouse
MEASURING AND LEVELING TOOLS
MEASURING AND LEVELING TOOLS LASER LEVELING FAST AND ACCURATE LEVELING Earth moving, laying of floors, interior design and set up; all applications for our laser and optical levels to prove their accuracy,
Index-Velocity Rating Development for Rapidly Changing Flows in an Irrigation Canal Using Broadband StreamPro ADCP and ChannelMaster H-ADCP
Index-Velocity Rating Development for Rapidly Changing Flows in an Irrigation Canal Using Broadband StreamPro ADCP and ChannelMaster H-ADCP HENING HUANG, RD Instruments, 9855 Businesspark Avenue, San Diego,
Accurate Air Flow Measurement in Electronics Cooling
Accurate Air Flow Measurement in Electronics Cooling Joachim Preiss, Raouf Ismail Cambridge AccuSense, Inc. E-mail: [email protected] Air is the most commonly used medium to remove heat from electronics
Soil degradation monitoring by active and passive remote-sensing means: examples with two degradation processes
Soil degradation monitoring by active and passive remote-sensing means: examples with two degradation processes Naftaly Goldshleger, *Eyal Ben-Dor,* *Ido Livne,* U. Basson***, and R.Ben-Binyamin*Vladimir
Active Vibration Isolation of an Unbalanced Machine Spindle
UCRL-CONF-206108 Active Vibration Isolation of an Unbalanced Machine Spindle D. J. Hopkins, P. Geraghty August 18, 2004 American Society of Precision Engineering Annual Conference Orlando, FL, United States
DUGARD. Machine Tools Since 1939. Dugard ECO Series Driven Tool, Slant Bed CNC Lathes. www.dugard.com
DUGARD Machine Tools Since 1939 ECO Series Driven Tool, Slant Bed CNC Lathes www.dugard.com ECO Series - CNC Lathes Amazing speed, power, precision and durability all at an affordable price Rigid and Trustworthy
Mark Jaski. Multi-Bend-Achromat (MBA) Magnets
Multi-Bend-Achromat (MBA) Magnets Outline Multi-Bend-Achromat (MBA) Functional Requirements Document Sector Layout Types of Magnets Magnet requirements Magnet Designs Modular design Coil Designs Alternative
Newland Machine Tool Group Inc.
Newland Machine Tool Group Inc. HEAVY MACHINE TOOLS OVERVIEW Horizontal Boring Mills KB Series T-type HBM with 130 or 165 mm extendable spindle (W-axis), column movable in the spindle direction (Z-axis),
DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements
DMS 680 Universal length measuring system gauge inspection for»iso 9000«requirements Universal length measuring system DMS 680 Wide range of applications For periodic inspection of gauges, reference gauges,
CNC HARDWARE & TOOLING BASICS
Computer Aided Manufacturing (CAM) CNC HARDWARE & TOOLING BASICS Assoc. Prof. Dr. Tamer S. Mahmoud 1. Parts of CNC Machine Tools Any CNC machine tool essentially consists of the following parts: Part Program,
HIGH SPEED PERMANENT MAGNET SYNCHRONOUS MOTOR / GENERATOR DESIGN FOR FLYWHEEL APPLICATIONS
HIGH SPEED PERMANENT MAGNET SYNCHRONOUS MOTOR / GENERATOR DESIGN FOR FLYWHEEL APPLICATIONS Aleksandr Nagorny, Ph.D. National Research Council Outline Introduction Selection of the Rated Point The major
DRS-6/1 Doppler Radar Sensor
DRS-6/1 Doppler Radar Sensor B+S Software und Messtechnik GmbH Lochhamer Schlag 17 D-82166 Graefelfing Germany October 1999 User's Manual DRS-6 SENSOR for speed measurements 1. DRS-6 Functional Description
TA-A BED TYPE MILLING CENTRE
TA-A BED TYPE MILLING CENTRE TA-A BED TYPE MILLING CENTRE MILLING CENTRE HIGH CAPACITY IN A COMPACT MACHINE High stock removal capacity TA-A BED TYPE MILLING CENTRE 2 TA-A The TA-A milling centre is a
PoS(PhotoDet 2012)068
Characterization of the Hamamatsu R11265 multi-anode photomultiplier tube with single photon signals Luca CADAMURO, Marta CALVI, Andrea GIACHERO,, Matteo MAINO, Clara MATTEUZZI, Gianluigi PESSINA Dipartimento
DREXELBROOK. USonic-R Series. Ultrasonic Level Measurement System. User Friendly
DREXELBROOK A Leader In Level Measurement Solutions USonic-R Series Ultrasonic Level Measurement System User Friendly Set the measurement range directly in inches, feet, millimeters, centimeters, or meters
MoveInspect HF HR. 3D measurement of dynamic processes MEASURE THE ADVANTAGE. MoveInspect TECHNOLOGY
MoveInspect HF HR 3D measurement of dynamic processes MEASURE THE ADVANTAGE MoveInspect TECHNOLOGY MoveInspect HF HR 3D measurement of dynamic processes Areas of application In order to sustain its own
Givare NC-SRS 280 och 880
Givare NC-SRS 280 och 880 Aratron AB/Nordic Control, 02 03 SRS280SEALED ROTARY SENSOR PERFORMANCE ±2 Resistance ±20% Hysteresis (repeatability) Accuracy Applied voltage maximum Resolution Output smoothness
PASSIVE INFRARED INTRUSION DETECTOR PASSIVE INFRAROOD DETECTOR DETECTEUR D INTRUSION PASSIF INFRAROUGE
MODEL HAA51 PASSIVE INFRARED INTRUSION DETECTOR PASSIVE INFRAROOD DETECTOR DETECTEUR D INTRUSION PASSIF INFRAROUGE OPERATING MANUAL HAA51 PASSIVE INFRARED INTRUDER DETECTOR INTRODUCTION HAA51 Passive Infrared
A new extraction system for the upgraded AIC-144 cyclotron
NUKLEONIKA 2001;46(2):51 57 ORIGINAL PAPER A new extraction system for the upgraded AIC-144 cyclotron Edmund Bakewicz, Krzysztof Daniel, Henryk Doruch, Jacek Sulikowski, Ryszard Taraszkiewicz, Nikolaj
Modern Physics Laboratory e/m with Teltron Deflection Tube
Modern Physics Laboratory e/m with Teltron Deflection Tube Josh Diamond & John Cummings Fall 2010 Abstract The deflection of an electron beam by electric and magnetic fields is observed, and the charge
TENSIONMETER ** code 183_LNRSTW **
Electronic tension meter for applications in textile and wire industry. 9 Tension ranges from 0.1-200 cn to 5-50 dan. Versatile instruments usable for applications in textile industry: e. g. 50 up to max.
DRTS 33. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters
The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters Testing all relay technologies: electromechanical, solid state, numerical and IEC61850 Manual