Specification of Requirements for Ventilated Cage Rack System

Size: px
Start display at page:

Download "Specification of Requirements for Ventilated Cage Rack System"

Transcription

1 Specification of Requirements for Ventilated Cage Rack System 1.0 General 1.1 This specification comprises minimum requirements for purchase and installation of ten (10) each Ventilated Cage Racks with full complement of 126 complete cage units each, and five (5) each blower units complete with all components, parts, installation, connectors, trolleys, assembly, labor and accessories, as required to provide a complete and functioning system including animal cages to house mice. In addition to the Ventilated Cage Racks and Blower Units, the vendor shall provide and deliver additional complete cage units equal to 50% of the base cage count (630 cages in additional to the 1,260 cages routinely provided as part of the 10 Ventilated Cage Racks for a total cage count of 1,890) and water bottles equal to 150% of the base cage count (for a total count of 1,890 water bottles). 1.2 Any deviations from the requirements of the specifications shall be clearly shown in the RFP with reference to the specific clauses in the specification. Any additional information, including but not limited to, manufacturer s preprinted product literature, material qualities, proprietary details, test reports, etc., not described shall be submitted as an attachment to the RFP. 1.3 When RFP s are evaluated, an assessment will be made of the general design of the units, including but not limited to choice of material proposed for manufacture of the ventilated cage rack system, operational reliability, design details, in addition to the requirements of the specification. 1.4 As part of the assessment of the RFP, emphasis will be placed on compliance with the specification and on the total number of animals that can be kept in the ventilated cage racks proposed, so that the animal rooms may be utilized both practically and optimally. The number of animals that can be safely and efficiently housed depends on both the total number of cages and the size of the animals housed. All cages shall comply with or exceed the minimum standards contained in the Guide for the Care and Use of Laboratory Animals (National Research Council, 1996, National Academy Press, Washington, DC) for the number of animals recommended per cage size. 1.5 All electrical equipment shall carry the UL or CE mark and shall comply with the National Electrical Code (NEC) as applicable. 1.6 ISO 9002 must be in-force at the manufacturing facility at all times during manufacture of all components of the ventilated cage rack system. The vendor shall provide evidence of ISO 9002 certification as an attachment the RFP response. 1.7 Use of Audible alarms shall not be permitted in Animal Holding rooms. Page 1 of 8

2 2.0 Design 2.1 Cage Racks Approved Manufacturers: a. Techniplast USA, Inc 180 Gordon Drive, Suite 109 Exton, PA Fax: Contact: Bill Stacy, Fax b. Allentown Caging Equipment Co., Inc. PO Box 698, Rt. 526 Allentown, NJ Fax: Contact: William Bean, Fax c. Lab Products, Inc. 742 Sussex Avenue, P.O. Box 639 Seaford, DE Fax: Contact: Robert Graves, Fax: The ventilated cage rack shall be designed to hold cages designed for mice The ventilated cage rack system shall be designed to be used either single-sided or double-sided, i.e. cages placed on one side or both sides All components of the ventilated cage rack system shall be designed to be cleaned, disinfected and sterilized in a mechanical cage washer and/or autoclave All electrical components shall be mounted in water-resistant NEMA rated sealed boxes to withstand disinfecting procedures All cage racks shall be provided with fully swiveling, lockable casters manufactured from a tread material that does not damage, discolor, or otherwise mark or deface floor coverings or walls. The dimension of the wheels (at least 4 inches) shall allow one person to easily maneuver the cage rack without assistance The ventilated cage rack system shall coordinate with, and connect to the automatic watering system, as manufactured and installed by Edstrom Industries, Inc. in the BSRB Facility. The cage rack system shall incorporate integral animal drinking manifolds mounted directly to the ventilated cage rack. The animal drinking manifolds shall incorporate on line rack flushing with both supply inlet and discharge outlet on the same side of the rack. Digital photographs of the Page 2 of 8

3 BSRB animal watering system are available upon request. Arrangement of animal watering system components, including but not limited to valves, shall not obstruct view of cage interiors and shall allow for tight stacking of cages during cleaning. 2.2 Animal Cages Animal cages shall be provided in sizes suitable for mice. When in use, the cage base and filter top will positively seal by means of an integrally designed gasket manufactured of nylon or other suitable material The filter top shall be of low profile type easily mounted on the cage without use of specialized tools or parts. The filter top shall include an integrally designed depression for a water bottle that can be removed from the cage without requiring removal of the lid from the cage, or the cage from the rack. When a bottle is in use, it shall seal against the inlet grommet. The water bottle inlet seal shall be manufactured from silicone or some other inert plastic material that remains flexible. When cages or bottles are removed from the ventilated cage rack, all grommet openings shall seal by means of a self-closing device, which will be positive locking and autoclavable The top shall not be permanently fixed to the cage rack in any way. When an animal cage is removed from the ventilated cage rack system, the filter top shall remain in place on the animal cage by means of a secure clip mechanism to preserve the barrier protection Filter tops shall be provided with 0.2 micron efficient filter paper The animal cages shall be easy to insert and remove from the ventilated cage rack system. When the animal cage is placed in the correct position in the ventilated cage rack system, the cage shall be ventilated automatically without additional manual adjustments being necessary. 3.0 Dimensions 3.1 The maximum external dimensions of the ventilated cage rack system shall not exceed 78 inches x 35 inches including any protrusions such as rails or bumpers. 3.2 The total height of the ventilated cage rack system shall not exceed 82 inches (to permit movement through doors and elevators). 3.3 Mouse cages shall be provided with a minimum floor area of 82 in. 2, with a usable minimum internal height of 5.5 inches. Rat cages shall be provided with a minimum floor area of 215 in. 2, with a minimum usable internal height of 8.3 inches. 4.0 Material and Surface Treatment 4.1 All metal used in the ventilated cage rack system shall be 304 Stainless Steel. 4.2 Casters shall be sealed (free of grease), shall be maintenance free, and shall be designed to withstand autoclaving. 4.3 Animal cages and filter tops shall be molded of Udel (Polysulfone) plastic and shall be designed to withstand routine cleaning and autoclaving. Cage unit components shall be completely autoclavable without deformation so the barrier protection is maintained at all times. Page 3 of 8

4 4.4 Ventilation tubes (air manifolds) shall be round or oval for optimal air circulation and static pressure, and shall be easily removable from the rack. Manifolds constructed of square tubing are not permitted. Ventilation tubes, manifolds and other components shall be designed for washing and disinfection in a mechanical rack washer with a maximum working temperature of F (135 0 C). Each tube section, both vertical and horizontal, shall include access openings for ease of tube cleaning. All access openings shall be sealed air-tight when in use. 5.0 Protective Function 5.1 The purpose of the ventilated cage rack system is to form a double-sided protective barrier around the animals against the surrounding conditions, partly to protect the animal against airborne pollution and infection, partly to minimize the spread of allergens among staff when housing animals and handling animal cages. 5.2 Each cage shall maintain Positive or Negative pressure at all times when placed in a positive or negative (exhaust) mode. Positive and negative air flows shall be fully adjustable by the user while the blower units are in operation without requiring use of special or proprietary tools. 5.3 When placed in the Negative mode, air exhausted from the cages shall be taken directly from within the cage environment and shall exhaust directly into the building general exhaust system. 6.0 Noise and Vibration Isolation 6.1 Noise generating from ventilated cage rack components shall not exceed 50dBA when measured at all points inside animal cages once placed inside the cage rack system. Within the octave band 32kHz, the noise level shall not exceed 45dBA when measured at all points inside animal cages once placed inside the cage rack system. 6.2 Total noise generated from the ventilated cage rack system shall not exceed 45 dba when measured inside the room. Measurements will be made with all cage racks operating. 6.3 The air supply fans or associated equipment shall not cause vibration that can spread to the animal cage. Fan wheels shall be statically and dynamically balanced in the factory to meet or exceed grade G6.3 as defined in ANSI S Ventilation 7.1 Air Conduction The ventilated cage rack system shall be of recirculating type; i.e. room air shall be filtered and delivered to the animal cages via the top of the cage. Exhaust shall also take place via the cage top, and all exhaust air from the cage rack shall be filtered and exhausted directly from the room through flexible connection to a 6 inch round exhaust duct provided at the ceiling of the room. The ventilated cage rack system shall be designed to deliver a wide range of air changes per hour which can be adjusted by the user while the unit is in operation. The ventilated cage rack system shall be provided complete with supply and exhaust filter/fan modules that are connected to the cage rack with flexible hoses. These modules shall be mounted on a mobile equipment trolley that can be placed in various configurations around the rack. It Page 4 of 8

5 shall also be possible to interconnect several racks to this trolley. The exhaust output shall be connected to the building exhaust by means of a flexible thimble connection. 7.2 Filtering Both supply and exhaust air shall be filtered with a HEPA filter with an efficiency of >99.99% according to the Warm DOP Method for 0.2 micron particles. A prefilter shall be placed in front of the HEPA filter to prolong its life. The prefilter shall be easily accessible for filter replacement. 7.3 Air flow for different animal cage placement configurations in the cage rack system shall be maintained within 15%+/- of the air flow the unit is set to deliver irrespective of the positions of the individual cages in the rack. 7.4 The relation between supply and exhaust air shall be easy adjustable so the pressure differential between animal cages and room can be positive or negative. Supply and exhaust air flows shall be adjustable by the user while the unit is in operation. Controls shall be located at the blower units. 8.0 Cage Ventilation and Airflow 8.1 As part of the acceptance of the cage rack systems, confirmation of air movement and velocity will be provided by an independent testing agency, approved by the Owner. Testing shall include such techniques as Computation Fluid Dynamic modeling and/or smoke tests as a minimum. Ventilation airflow and velocity information shall be provided to the Owner as part of the project close-out documents in the form of a printed, certified report with a copy provided in CD format. 8.2 The Owner reserves the right to confirm airflow and pressure relationships using an independent testing agent as a condition of acceptance of the cage rack systems. The cost of this testing will be borne by the Owner. Any discrepancies found or adjustments required will be corrected by the vendor at no additional cost to the Owner. 9.0 Control and Safety Equipment 9.1 The ventilated cage rack system shall include an alarm or test system that alerts faults in the ventilation system. At a minimum, the system shall: 1) HEPA filter test circuit to continuously monitor and assess condition of the HEPA filter including an hour-meter to track time since last filter maintenance, 2) A pre-filter timer to visually indicate completion of a preset cleaning interval, 3) Digital readout of both Supply and Exhaust fan units that converts to air changes per hour, 4) Continuous monitoring and digital readout of pressure differential between room pressure and pressure inside the cage. 9.2 The Ventilated Cage Rack System shall provide clean contacts or relays for power failure alert via the Building Automation and Management System provided in the BSRB Facility. As a condition of acceptance of the cage rack systems, the vendor shall demonstrate successful communication with the Building Automation System (BAS) through activation of a low pressure alarm at the BAS panel. 9.3 As an attachment to the bid proposal, the vendor shall provide a proposal to provide and install manufacturer s optional local alarms at all ventilated cage racks. The local alarm panel shall be Page 5 of 8

6 mounted directly on the Ventilated Cage Rack in a highly visible location. Use of audible alarms is not permitted at the Ventilated Cage Rack System. 9.4 Exhaust air shall be captured via an indirect thimble connection and vented into the building general exhaust system. Thimble connections shall be provided and installed by the vendor. Thimble connections require a 4 flexible hose on the discharge side of the rack and shall be mounted to a 6 inch diameter round exhaust duct in the ceiling Ergonomic Considerations 10.1 The following items are minimum requirements for the Ventilated Cage Rack System. 1) Weight Racks and trolleys shall be lightweight and shall roll freely and easily. Use of shelves in the ventilated cage rack system is not permitted. 2) Visibility Rack design shall allow clear, unobstructed view of cage interiors to observe animals and to check water levels in animal watering bottles. Rack components shall not obstruct visibility of cage interiors. 3) Access to Equipment Supply and Exhaust units, filters and controls shall be accessible without requiring removal of racks. Associated equipment shall be mounted on mobile trolleys that can be easily moved and serviced. At no time shall equipment require lifting or removal for service or cleaning. 4) Cage Docking Racks shall NOT utilize individual cage locking devices that could potentially lead to stress related disorders, i.e., carpel tunnel system, etc. Cages shall seat firmly in place by inserting the cage into the rack with one clean motion Documentation & Submittals 11.1 One copy each of concise preprinted user instructions shall accompany each unit. User instructions shall be provided in both English and Spanish. The instructions shall contain all necessary information to achieve a satisfactory, safe protective function at all times when the cage rack system is in use. Use of symbols, flow charts, and simple schematic drawings is recommended Operation and Maintenance Manuals including but not limited to the following documentation shall be submitted to the Owner: 1) Operating instructions, 2) Description of design and function, 3) Regulations for safety and handling, 4) Description of control and alarm systems, 5) Maintenance instructions and schedules for all materials and components, 6) Replacement parts pictures and part numbers, 11.3 Vendor shall provide a minimum of three (3) copies of submittals, product data, and/or shop drawings as required to fully describe the proposed cage rack system, for review and approval, prior to beginning of fabrication. Submittals shall clearly detail and confirm the interface between the animal watering component provided as part of the cage rack system and the animal watering system provided in the building. Submittals shall also show the relationship between animal cages and the ventilated cage rack components. Page 6 of 8

7 11.4 Manufacturer shall provide a list of recommended spare parts and perishable supplies to be stocked by Owner. The recommended spare parts list shall be provided in the following format. M.D. Anderson Spare Parts List Spare Part Description Model/ Part # Replacement frequency Unit Price Count to keep in stock Total Cost subtotal shipping grand total 11.5 Engraved, waterproof name plates shall be permanently affixed to all equipment marked with, at a minimum, manufacturer name, manufacturer address and telephone number, model number, type designation and serial number Installation 12.1 Ventilated cage rack systems, including all parts, components and associated paperwork, shall be delivered directly to the final equipment locations in the Vivarium of the BSRB Facility located at 6767 Bertner Road, Houston, Texas All equipment and components shall be received, uncrated, inspected and accepted and moved into final place by the vendor. The delivery route through the building shall include delivery and unloading at the BSRB loading dock, use of the freight elevator located adjacent to the dock area, to move materials from the dock to the Basement and delivery to the final locations on the basement. An electronic copy of the floor plans of the facility is available upon request. The actual date and time of delivery shall be coordinated with the Owner a minimum of two (2) weeks before delivery and installation Ventilated cage rack systems will be installed in the following locations: Room No. XB.8452 XB.8455 XB.8293 Room Name: XRO Isolator Room XRO Breeding Room GEM Housing Room Description of Equipment: Two (2) Ventilated Cage Rack Systems + one (1) blower unit Six (6) Ventilated Cage Rack Systems + three (3) blower units Two (2) Ventilated Cage Rack Systems + one (1) blower unit 12.3 All equipment shall be installed, cleaned, disinfected, started, commissioned, tested and certified by the vendor. All connections, included but not limited to, electrical connections, HVAC connections including flexible thimble connections shall be the responsibility of the vendor. Vendor shall demonstrate operation of all equipment to the Owner as a condition of acceptance. Page 7 of 8

8 12.4 Removal and disposal of all unused materials, trash, debris, crating materials, etc. shall be the responsibility of the vendor. Spaces shall be left clean and in the same condition as before installation of the ventilated cage rack systems Inspection 13.1 All control and alarm equipment shall be tested with a built-in test program in the presence of the Owner All control and alarm equipment will be calibrated by the supplier All racks and equipment trolleys shall be thoroughly tested and documented at the factory and test data shall be recorded in the operation manuals included with each ventilated cage rack system Warranties, Maintenance Agreements and Preventative Maintenance 14.1 Vendor shall provide the manufacturer s standard one year warranty against defects in manufacture, workmanship and design. Warranty coverage shall begin upon date of Owner s final acceptance of equipment Vendor shall provide a proposal for an additional one year extension to the standard manufacturer s one year maintenance agreement providing all materials, parts, and labor required to maintain Ventilated Cage Rack systems in place without interruption of service Vendor shall provide a schedule showing all maintenance activities, including preventative maintenance activities, along with scheduled dates for each activity. End of Specifications Page 8 of 8

DRYER VENTING SPECIFICATIONS DRYER SAFETY

DRYER VENTING SPECIFICATIONS DRYER SAFETY DRYER VENTING SPECIFICATIONS Table of Contents DRYER SAFETY...1 INSTALLATION REQUIREMENTS...4 Venting Requirements...5 DRYER INSPECTION AND CLEANING...7 Frequency of Exhaust System Cleaning...7 Inspecting

More information

All work must conform to the National Electric Code, latest edition, and all other applicable codes and regulations.

All work must conform to the National Electric Code, latest edition, and all other applicable codes and regulations. DIVISION 16 ELECTRICAL SECTION 16100 - ELECTRICAL PART 1 - GENERAL SCOPE All electrical work as shown on the drawings and as necessary to provide a complete electrical system. Include primary service,

More information

Horizontal Steam Sterilizers. Healthcare Line. Since

Horizontal Steam Sterilizers. Healthcare Line. Since Horizontal Steam Sterilizers Healthcare Line Since 1925 Horizontal Steam Sterilizers for Healthcare Applications The Tuttnauer line of healthcare sterilizers are designed for hospital operating rooms,

More information

Laminar Flow Cabinets

Laminar Flow Cabinets Laminar Flow Cabinets The World s Most Practical Selection of Benchtop Laminar Flow Cabinets. 24 36 48 Simple, Effective Protection for Samples and Work Processes, Uncompromised Performance, Economical

More information

SECTION 23 00 01 HEATING, VENTILATION AND AIR CONDITIONING EQUIPMENT

SECTION 23 00 01 HEATING, VENTILATION AND AIR CONDITIONING EQUIPMENT SECTION 23 00 01 HEATING, VENTILATION AND AIR CONDITIONING EQUIPMENT PART 1 - GENERAL 1.1 DESCRIPTION A. The Work specified in this Section consists of designing, furnishing and installing factory-assembled

More information

Element D Services Heating, Ventilating, and Air Conditioning

Element D Services Heating, Ventilating, and Air Conditioning PART 1 - GENERAL 1.01 OVERVIEW A. This section supplements Design Guideline Element D3041 on air handling distribution with specific criteria for projects involving design of a Data Center spaces B. Refer

More information

Heating oven with mechanical convection

Heating oven with mechanical convection FD series 53 Drying ovens / Heating ovens Heating oven with mechanical convection A BINDER FD series heating oven is always used when fast drying and sterilization is required. Thanks to its fully homogeneous

More information

Subpart 1. Installation. All plumbing systems must be. installed and tested according to this chapter and chapter 4715,

Subpart 1. Installation. All plumbing systems must be. installed and tested according to this chapter and chapter 4715, 4658.4500 PLUMBING SYSTEMS; NEW CONSTRUCTION. Subpart 1. Installation. All plumbing systems must be installed and tested according to this chapter and chapter 4715, the Minnesota Plumbing Code. Subp. 2.

More information

Class II Type B2 Laminar Flow Biological Safety Cabinets at Ordering Criteria for the National Institutes of Health

Class II Type B2 Laminar Flow Biological Safety Cabinets at Ordering Criteria for the National Institutes of Health Background Class II Type B2 Laminar Flow Biological Safety Cabinets at Ordering Criteria for the National Institutes of Health This document provides guidelines for the design and installation of Class

More information

LAMINAR FLOW CABINETS

LAMINAR FLOW CABINETS LAMINAR FLOW CABINETS Vertical Laminar Flow Cabinets VLF 550 VLF650 VLF1000 VLF1200 VLF 1500 VLF1800 Monmouth Scientific's brand new range of Vertical laminar Flow (VLF) cabinets is available in six different

More information

Cooktop Low-Profile Ventilation Hoods

Cooktop Low-Profile Ventilation Hoods INSTALLATION GUIDE Cooktop Low-Profile Ventilation Hoods Contents Wolf Cooktop Low-Profile Ventilation Hoods........ 3 Cooktop Low-Profile Hood Specifications.......... 4 Cooktop Low-Profile Hood Installation............

More information

Heating oven with mechanical control

Heating oven with mechanical control ED series 53 Drying ovens / Heating ovens Heating oven with mechanical control The strength of the BINDER ED series heating oven: routine drying and sterilization applications up to 300 C (572 F). Thanks

More information

FIRE DAMPER APPLICATION GUIDE

FIRE DAMPER APPLICATION GUIDE FIRE AND CEILING RADIATION DAMPERS FIRE DAMPER APPLICATION GUIDE Use this application guide to help determine what type of fire damper best suits the application. Below are listed the ten Metal-Fab fire

More information

Material test chamber with mechanical convection

Material test chamber with mechanical convection FP series 400 Material test chambers Material test chamber with mechanical convection A BINDER material test chamber with mechanical convection of the FP series provides reliably short drying times and

More information

SECTION 16853 ZONED (DC LOOP) FIREALARM SYSTEM

SECTION 16853 ZONED (DC LOOP) FIREALARM SYSTEM SECTION 16853 ZONED (DC LOOP) FIRE-ALARM SYSTEM PART 1 - GENERAL 1.1 RELATED DOCUMENTS A. Drawings and general provisions of the Contract, including General and Supplementary Conditions and Division 1

More information

The Animal Welfare Act Regulations for the Transportation of Wildlife

The Animal Welfare Act Regulations for the Transportation of Wildlife The Animal Welfare Act Regulations for the Transportation of Wildlife The following sections of the Code of Federal Regulations Title 9 are the most relevant to the transportation of animals by UIUC researchers,

More information

Installation and Operating Instructions Installation Instructions for SS EPE-316L Series

Installation and Operating Instructions Installation Instructions for SS EPE-316L Series INSTR3010 0406 Installation and Operating Instructions Installation Instructions for SS EPE-316L Series Congratulations on your purchase of this Aqua-Pure high flow, single housing filtration system. This

More information

Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection

Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection A wide range of products to meet the needs of transferring animals, operators,

More information

Convection Ovens. BX Classic. Models BX4 / FG 189C, BX4-6040 / FG 158C, BX10 / FG 180C. Operator's Manual

Convection Ovens. BX Classic. Models BX4 / FG 189C, BX4-6040 / FG 158C, BX10 / FG 180C. Operator's Manual Convection Ovens BX Classic Models BX4 / FG 189C, BX4-6040 / FG 158C, BX10 / FG 180C Operator's Manual Belshaw Bros., Inc. 814 44 th Street NW, Suite 103 Auburn, WA 98001 USA Phone: 206-322-5474 Fax: 206-322-5425

More information

ELECTRICAL CONTRACTORS THE UNIVERSITY OF TENNESSEE KNOXVILLE, TENNESSEE SECTION 16050 PAGE 1

ELECTRICAL CONTRACTORS THE UNIVERSITY OF TENNESSEE KNOXVILLE, TENNESSEE SECTION 16050 PAGE 1 KNOXVILLE, TENNESSEE SECTION 16050 PAGE 1 PART 1 - GENERAL 1.01 RELATED DOCUMENTS: SECTION 16050 A. Drawings and general provisions of Contract, including General and Supplementary Conditions and Division-1

More information

Part 3 Ancillary technologies. Chapter 3.2. Mechanical ventilation with heat recovery

Part 3 Ancillary technologies. Chapter 3.2. Mechanical ventilation with heat recovery Part 3 Ancillary technologies Chapter CONTENTS SCOPE Clause Page DESIGN Design standards D1 1 Statutory requirements D2 1 System design D3 1 Building integration D4 2 Noise D5 2 Access and controls D6

More information

FIRE ALARM SYSTEM. A. Shop drawings shall be submitted as follows: 1. Manufacturer's published literature. for aspproval.

FIRE ALARM SYSTEM. A. Shop drawings shall be submitted as follows: 1. Manufacturer's published literature. for aspproval. Polk School District Specifications for the New Fire Alarm in the Existing Building Eastside Elementary School FIRE ALARM SYSTEM 1.01 SUBMITTALS A. Shop drawings shall be submitted as follows: 1. Manufacturer's

More information

LOADING DOCK EQUIPMENT. A. Section 03100 - Concrete Forms and Accessories: Placement of anchors into concrete.

LOADING DOCK EQUIPMENT. A. Section 03100 - Concrete Forms and Accessories: Placement of anchors into concrete. LOADING DOCK EQUIPMENT PART 1 GENERAL 1.1 SECTION INCLUDES A. Loading dock equipment of the following types: 1. Dock lifts. 2. Control stations. 1.2 RELATED SECTIONS A. Section 03100 - Concrete Forms and

More information

OPERATING MANUAL CONTRACTOR STYLE AXIAL FANS

OPERATING MANUAL CONTRACTOR STYLE AXIAL FANS Manual No. BLWR043 Rev. 4 September 2012 OPERATING MANUAL CONTRACTOR STYLE AXIAL FANS CVF-8AC CVF-8AC50 CVF-6ACAN CVF-15ACAN CVF-25ACAN CVF-8A15KIT CVF-8A25KIT MODELS: CVF-8DC CVF-6DCAN CVF-15DCAN CVF-25DCAN

More information

PRE-ASSEMBLED INSTALLATION INSTRUCTIONS FOR PACB SERIES COAL FIRED OVENS

PRE-ASSEMBLED INSTALLATION INSTRUCTIONS FOR PACB SERIES COAL FIRED OVENS PRE-ASSEMBLED INSTALLATION INSTRUCTIONS FOR PACB SERIES COAL FIRED OVENS OVENS MANUFACTURED BY: EARTHSTONE OVENS-6717 SAN FERNANDO RD GLENDALE CA 91201 TELEPHONE: 800-840-4915 818-553-1134 FAX: 818-553-1133

More information

SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS

SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS ADVANTAGE PLUS SITE REQUIREMENTS SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS 50096-927 Rev A For additional information contact your Medivators Pg 1 of 18 SITE REQUIREMENTS The purpose of this section

More information

CLEAR-DIVISIONS CENTERFOLD

CLEAR-DIVISIONS CENTERFOLD CLEAR-DIVISIONS CENTERFOLD Introduction: The following three (3) part specification offers the Standard and Optional features for the CLEAR-DIVISIONS CENTERFOLD moveable The yellow highlighted areas in

More information

A. Air-Conditioning and Refrigeration Institute: 1. ARI 850 - Commercial and Industrial Air Filter Equipment.

A. Air-Conditioning and Refrigeration Institute: 1. ARI 850 - Commercial and Industrial Air Filter Equipment. SECTION 23 40 00 - HVAC AIR CLEANING DEVICES PART 1 - GENERAL 1.1 SUMMARY A. Section Includes: 1. Activated carbon filters. 2. Extended surface rigid air filter with synthetic media. 3. Disposable panel

More information

SPECIAL APPLICATIONS

SPECIAL APPLICATIONS Page 1 SPECIAL APPLICATIONS The protection of microelectronics clean rooms using aspirating smoke detection systems requires special consideration. The physical configuration of the room(s) and the direction

More information

Heating, Ventilation, Air Conditioning and Refrigeration (HVACR)

Heating, Ventilation, Air Conditioning and Refrigeration (HVACR) Heating, Ventilation, Air Conditioning and Refrigeration (HVACR) I. Demonstrate safety skills in typical HVACR work situations to NATE Core Installer Knowledge Areas for Technician Excellence for Safety

More information

Element D Services Plumbing

Element D Services Plumbing Medical Vacuum and Gas PART 1 - GENERAL 1.01 OVERVIEW A. This section addresses medical vacuum, waste anesthetic gas disposal, compressed air, oxygen, nitrous oxide, nitrogen and carbon dioxide systems.

More information

E-PDD Duct Smoke Detector Installation Sheet

E-PDD Duct Smoke Detector Installation Sheet E-PDD Duct Smoke Installation Sheet Operation The duct smoke detector's primary purpose is to provide early warning of an impending fire and shut down the HVAC unit in order to prevent smoke from circulating

More information

Submit shop drawings for equipment provided under this section Shop drawings shall indicate:

Submit shop drawings for equipment provided under this section Shop drawings shall indicate: Section 16435 - SWITCHBOARDS Introduction Part 1 - General Reference The work under this section is subject to requirements of the Contract Documents including the General Conditions, Supplementary Conditions,

More information

Disposable Rodent Caging System

Disposable Rodent Caging System Since introducing the Innovive Disposable Rodent Caging System in 2005, Innovive continues to bring innovation to the animal husbandry industry. Our suite of research-ready caging products now includes

More information

General Purpose Laminar Flow Cabinets, Horizontal and Vertical

General Purpose Laminar Flow Cabinets, Horizontal and Vertical 24 36 48 60 72 96 The World s Most Practical Selection of Benchtop Laminar Flow Cabinets. VLF-48 Provides reliable protection for samples and work processes for a multitude of applications. p:2 Product

More information

Ceiling Mount Air Handler Manual

Ceiling Mount Air Handler Manual www.surna.com 303.993.5271 Ceiling Mount Air Handler Manual Models: CMAH12, CMAH18, CMAH24, CMAH30, CMAH36, CMAH48, CMAH60 Revised: September 2014 Table of Contents Warranty Information 4 Limited Warranty

More information

SECTION 27 05 36 CABLE TRAYS FOR COMMUNICATIONS SYSTEMS

SECTION 27 05 36 CABLE TRAYS FOR COMMUNICATIONS SYSTEMS SECTION 27 05 36 CABLE TRAYS FOR COMMUNICATIONS SYSTEMS PART 1 GENERAL 1.01 DESCRIPTION A. The work covered by this section of the Specifications includes all labor necessary to perform and complete such

More information

AABC Specifications for Testing and Balancing HVAC Systems

AABC Specifications for Testing and Balancing HVAC Systems AABC Specifications for Testing and Balancing HVAC Systems The Associated Air Balance Council (AABC) Total System Balancing Specifications for HVAC Testing, Balancing and Performance Verification TOTAL

More information

Utility Distribution Systems

Utility Distribution Systems Utility Distribution Systems 6/2012 A0011037 1 WARRANTY This equipment is warranted to be free from defects in materials and workmanship, under normal use and service, for a period of 12 months from date

More information

Portable Air Conditioner

Portable Air Conditioner Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE

More information

Spetec. Laminar Flow Box. Portable Clean Room Technology

Spetec. Laminar Flow Box. Portable Clean Room Technology Spetec Laminar Flow Box Portable Clean Room Technology 1 Fundamentals Portable Clean Room Technology Laminar Flow The basis for the development of the clean room technology was first established in the

More information

ASX7-4-IOM-7 J30-08579

ASX7-4-IOM-7 J30-08579 INSTALLATION INSTRUCTIONS 4" COMBUSTION AIR INLET KIT TUBULAR GAS FIRED DIRECT SPARK PROPELLER UNIT HEATERS UNIT CONVERSION & CATEGORY III VENTING FOR SEPARATED COMBUSTION For 30,000 to 75,000 BTU/HR SUPPLEMENT

More information

FOXAIR 60/50 ALL MODELS

FOXAIR 60/50 ALL MODELS FOXAIR 60/50 ALL MODELS OPERATION AND SERVICE MANUAL UNIT WEIGHT - ALL MODELS APPROX. 300 POUNDS FOXTRONICS, INC. 38 L x 28 W x 24 H Love Field-Dallas TX MADE IN THE USA 3448 West Mockingbird Lane Dallas,

More information

Stainless Steel Worksurface 3 Feet Hood A10-WS-PR3 A10-WS-SR3 4 Feet Hood A10-WS-PR4 A10-WS-SR4 6 Feet Hood A10-WS-PR6 A10-WS-SR6.

Stainless Steel Worksurface 3 Feet Hood A10-WS-PR3 A10-WS-SR3 4 Feet Hood A10-WS-PR4 A10-WS-SR4 6 Feet Hood A10-WS-PR6 A10-WS-SR6. Product Sampler #1 Laminar Flow Hoods / Laminar Flow Cabinets: 1- Benchtop Vertical Laminar Flow Hoods- Air Recirculation Available s: Clean Bench Class 100 Bench-top Vertical Laminar Flow Hoods with Air-recirculation

More information

PI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3)

PI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3) PI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3) Yes No 1. Is access to the laboratory limited or restricted at the discretion of the laboratory director when experiments are in progress? Yes No

More information

CORNER FRIDGE REFRIGERATION UNIT INSTALLATION AND OPERATION INSTRUCTIONS

CORNER FRIDGE REFRIGERATION UNIT INSTALLATION AND OPERATION INSTRUCTIONS CORNER FRIDGE REFRIGERATION UNIT INSTALLATION AND OPERATION INSTRUCTIONS D E F Y I N G C O N V E N T I O N Congratulations on your new Corner Fridge Your new corner fridge may have different functions

More information

SECTION 11014 WINDOW WASHING SYSTEMS

SECTION 11014 WINDOW WASHING SYSTEMS SECTION 11014 WINDOW WASHING SYSTEMS PART 1 - GENERAL 1.1 SUMMARY A. Section includes, but is not limited to, design and furnishing portable davits, davit sleeves, dvit bases, four (4) removable outriggers,

More information

HEPA-filtered Negative Air Machines and their Role in Creating Airborne Infection Isolation Rooms with the ECU AnteRoom. Frequently Asked Questions

HEPA-filtered Negative Air Machines and their Role in Creating Airborne Infection Isolation Rooms with the ECU AnteRoom. Frequently Asked Questions HEPA-filtered Negative Air Machines and their Role in Creating Airborne Infection Isolation Rooms with the ECU AnteRoom Frequently Asked Questions What is a negative air machine and why do I need one?

More information

STABITHERM INSTALLATION & SERVICE BOOKLET

STABITHERM INSTALLATION & SERVICE BOOKLET STABITHERM INSTALLATION & SERVICE BOOKLET TABLE OF CONTENTS Introduction...3 Warranty...3 Product Range and Physical Description...4 Method of Operation...5 Recommended Pressures and Temperatures...6 Flow

More information

Environmental Monitoring of Clean Rooms

Environmental Monitoring of Clean Rooms Environmental Monitoring of Raul Duarte President DDK Scientific, Corp. Copyright DDK Scientific, Corp. 2008, 2009, DDK Scientific, Corp. Proprietary A manufacturing facility for pharmaceutical products

More information

FIRE ALARM AND DETECTION SYSTEMS SECTION 16721

FIRE ALARM AND DETECTION SYSTEMS SECTION 16721 PART 1 - GENERAL 1.01 WORK INCLUDED FIRE ALARM AND DETECTION SYSTEMS SECTION 16721 A. Provide a complete fully addressable, power limited, fire detection and evacuation system. The system shall be connected

More information

Select Radiators Installation Guide

Select Radiators Installation Guide Select Radiators Installation Guide Table of Contents Informational Symbols...3 Before You Begin...4 Select Rough-In... 5 Connection Installation...6 Optional Piping Arrangements...7 Conventional Wall

More information

SECTION 11 78 13 MORTUARY REFRIGERATORS

SECTION 11 78 13 MORTUARY REFRIGERATORS PART 1 - GENERAL 1.1 DESCRIPTION SECTION 11 78 13 MORTUARY REFRIGERATORS SPEC WRITER NOTE: Delete between //---// if not applicable to project. Also delete any other item or paragraph not applicable in

More information

How To Install A Perfect 16 Id Series Filter System

How To Install A Perfect 16 Id Series Filter System IQAir Perfect 16 ID-Series MERV16 Micro-Particle Filtration System Perfect 16 Installation Manual Product Data High-Performance Medical-Grade Residential/Commercial Air Filtration System Application The

More information

DATA CENTER RACK SYSTEMS: KEY CONSIDERATIONS IN TODAY S HIGH-DENSITY ENVIRONMENTS WHITEPAPER

DATA CENTER RACK SYSTEMS: KEY CONSIDERATIONS IN TODAY S HIGH-DENSITY ENVIRONMENTS WHITEPAPER DATA CENTER RACK SYSTEMS: KEY CONSIDERATIONS IN TODAY S HIGH-DENSITY ENVIRONMENTS WHITEPAPER EXECUTIVE SUMMARY Data center racks were once viewed as simple platforms in which to neatly stack equipment.

More information

COMMERCIAL COOKING HOODS, VENTILATION & FIRE SUPPRESSION SYSTEM GUIDELINES AND PROCEDURES

COMMERCIAL COOKING HOODS, VENTILATION & FIRE SUPPRESSION SYSTEM GUIDELINES AND PROCEDURES COMMERCIAL COOKING HOODS, VENTILATION & FIRE SUPPRESSION SYSTEM GUIDELINES AND PROCEDURES These guidelines are to be used for ALL commercial cooking hoods, ventilation systems and related fire suppression

More information

SWIMMING POOL, SPA, & HOT TUB GUIDELINES

SWIMMING POOL, SPA, & HOT TUB GUIDELINES SWIMMING POOL, SPA, & HOT TUB GUIDELINES Pool permit cost is based on the total value of the project. See Permit Fee Schedule Table 1A. A. Adopted construction codes and installation requirements 1. 2015

More information

LTR Series Uninterruptible Power Systems 700 VA - 2.1 KVA. General Specification

LTR Series Uninterruptible Power Systems 700 VA - 2.1 KVA. General Specification 08/17/12 Rev2 LTR Series General Specification 700 VA to 2.1 KVA 1.0 General LTR Series Uninterruptible Power Systems 700 VA - 2.1 KVA General Specification This specification describes the features and

More information

DIAMOND Retractable Rodding Robot Model SPRAYROD-R

DIAMOND Retractable Rodding Robot Model SPRAYROD-R 2004-12-21 2 1 (23) DIAMOND Retractable Rodding Robot Model SPRAYROD-R 2004-12-21 2 2 (23) Table of contents 1 TECHNICAL DESCRIPTION...4 1.1 MAIN DETAILS...5 1.2 COMPONENTS DESCRIPTION...5 1.2.1 Pneumatic

More information

SAM HOUSTON STATE UNIVERSITY

SAM HOUSTON STATE UNIVERSITY PART 1: GENERAL 1.01 General Requirements A. This standard is intended to provide useful information to the Professional Service Provider (PSP) to establish a basis of design. The responsibility of the

More information

HVAC Checklist - Long Form

HVAC Checklist - Long Form HVAC Checklist - Long Form Page 1 of 14 Appendix B discusses HVAC system components in relation to indoor air quality. utside Air Intake Location pen during occupied hours? Unobstructed? Standing water,

More information

26 3213.13 Diesel Engine Driven Generators Page 1 of 6

26 3213.13 Diesel Engine Driven Generators Page 1 of 6 Last Update: December 8, 2014 A. Description of System Consultant s Handbook Page 1 of 6 1. Provide a diesel engine driven electric generating unit, factory assembled, tested and certified to operate at

More information

CEILING CASSETTE FAN COIL UNIT

CEILING CASSETTE FAN COIL UNIT CEILING CASSETTE FAN COIL UNIT Operation and Installation Manual MHCFC4W-04, 08, 12, 16 Four way ceiling cassette fan coil unit Please read this manual before using the fan coil unit. Installation and

More information

Industrial Ventilation Design Guidebook Chapter 10 LOCAL VENTILATION

Industrial Ventilation Design Guidebook Chapter 10 LOCAL VENTILATION Industrial Ventilation Design Guidebook Chapter 10 LOCAL VENTILATION LARS OLANDER, LORRAINE CONROY, ILPO KULMALA, RICHARD P. GARRISON, MICHAEL ELLENBECKER, BERNHARD BIEGERT, BERNARD FLETCHER, HOWARD D.

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MN12CES / MN10CESWW Voltage rating: 120V~60Hz Power rating : 1100W (MN12CES) Power rating : 900W (MN10CESWW) Customer

More information

Standard Hydraulic Piping Flushing Procedure

Standard Hydraulic Piping Flushing Procedure Page 1 of 6 1.0 Scope This procedure covers the minimum technical requirements for cleaning and flushing of Hydraulic and Piping, and related accessories. 2.0 Introduction In hydraulic fluid power systems,

More information

PANDA SPRAY BOOTH FILTER INFORMATION

PANDA SPRAY BOOTH FILTER INFORMATION PANDA SPRAY BOOTH FILTER INFORMATION Our Spray Booth has been designed to reduce the risk of potential hazardous situations occurring. You must adhere to the specifications chart listed below when replacing

More information

Technical Concept. Filter Fan Unit COMPACT. Type COMPACT AC Type COMPACT EC

Technical Concept. Filter Fan Unit COMPACT. Type COMPACT AC Type COMPACT EC Technical Concept Filter Fan Unit COMPACT Type COMPACT AC Type COMPACT EC Ultra Filter Fan Unit COMPACT Technical Concept Product Description Filter Fan Units (FFUs) from M+W Products are designed to provide

More information

Laminar Flow Cabinets and Workstations

Laminar Flow Cabinets and Workstations VERTICAL AND HORIZONTAL HEPA FILTERED SYSTEMS FOR PROCESS OR OPERATOR PROTECTION Laminar Flow Cabinets and Workstations For procedures requiring protection from particulate contamination and for aseptic

More information

D. Infrared scanning reports with pictures printed in a final report with any deficiencies and actions taken to rectify.

D. Infrared scanning reports with pictures printed in a final report with any deficiencies and actions taken to rectify. SECTION 263354 STATIC UPS TESTING AND COMMISSIONING PART 1 - GENERAL 1.1 SUMMARY A. Section Includes 1. System start-up services. 2. Battery charging. 3. Load bank testing. 4. Battery load testing. 5.

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CCS Customer Support : 1-800-474-2147 Voltage rating: 115V~60Hz Power rating : 1400W For product inquiries or support

More information

Fire and Smoke Dampers: Best Practice Design Tips. An AMCA International White Paper. July 2015

Fire and Smoke Dampers: Best Practice Design Tips. An AMCA International White Paper. July 2015 July 2015 Fire and Smoke Dampers: Best Practice Design Tips An AMCA International White Paper AMCA International 30 West University Dr. Arlington Heights, IL 60004 USA www.amca.org June 2015 This white

More information

Generic - Hearing Loop - (AFILS) U.S. System Specification

Generic - Hearing Loop - (AFILS) U.S. System Specification This document is a generic specification for any Hearing Loop (Audio Frequency Induction Loop System). For the remainder of the document, we will refer to using the term Hearing Loop rather than Audio

More information

Laboratory & Scientific Line. Horizontal Steam Sterilizers

Laboratory & Scientific Line. Horizontal Steam Sterilizers Laboratory & Scientific Line Horizontal Steam Sterilizers Horizontal Steam Sterilizers for Laboratory & Scientific Applications Sterilization in a laboratory environment has its unique requirements. The

More information

Revised April (May) 2015

Revised April (May) 2015 Section 19.0 Cross Connection Control Policy 19.1 Responsibility 19.1.1 The Elkhart Water Utility has the responsibility to supply safe, potable water from the source to the point of delivery defined as

More information

HOSPITAL ISOLATION ROOM HVAC SYSTEM DESIGN

HOSPITAL ISOLATION ROOM HVAC SYSTEM DESIGN HOSPITAL ISOLATION ROOM HVAC SYSTEM DESIGN Presented to: ASHRAE YEA Illinois Chapter Specialty Environment Design Conference March 24, 2014 Jean O. Gibbons, PE, LEED AP Senior Vice President Grumman/Butkus

More information

Procedures for the Effective Use of Biological Safety Cabinets

Procedures for the Effective Use of Biological Safety Cabinets Procedures for the Effective Use of Biological Safety Cabinets APPROVAL DATE BY BIOHAZARDS SUBCOMMITTEE: August 10, 2012 SUPERSEDES: March 1, 2012 This corporate guideline/procedure is intended as a minimum

More information

Customized Server Cabinets

Customized Server Cabinets GOING GREEN WITH CUSTOMIZABLE CABINETS AND RACKS Designed to provide industry leading airflow, CyrusOne s cabinets and rack solutions promote optimal cooling and contribute to reduced power consumption.

More information

Fire. The Fire Installers Mate. A guide to fire alarm systems design

Fire. The Fire Installers Mate. A guide to fire alarm systems design Fire The Fire Installers Mate A guide to fire alarm systems design BS 5839 Part 1:2002 A guide to BS 5839 Part 1:2002 Disclaimer This booklet is not intended to be a comprehensive guide to all aspects

More information

Incubator with mechanical convection

Incubator with mechanical convection BF series 400 Incubators and cooling incubators Incubator with mechanical convection The BINDER incubator of the BF series with mechanical convection is suitable for all gentle incubation applications,

More information

Area of Rescue Assistance Communication System ADA Code Applicable excerpts:

Area of Rescue Assistance Communication System ADA Code Applicable excerpts: Area of Rescue Assistance Communication System ADA Code Applicable excerpts: CONTENTS (Guidelines and Definitions) Page 1 Area of Rescue Assistance Requirements Page 2 Emergency Communication Equipment

More information

GETINGE GEW 2100LS CAGE AND RACK WASHER PREVENTING CONTAMINATION IN ANIMAL CARE

GETINGE GEW 2100LS CAGE AND RACK WASHER PREVENTING CONTAMINATION IN ANIMAL CARE GETINGE GEW 2100LS CAGE AND RACK WASHER PREVENTING CONTAMINATION IN ANIMAL CARE GETINGE GEW 2100LS Cage and Rack Washer - Preventing contamination in animal care 3 CLEAN CAGES, RACKS AND BOTTLES WITH

More information

Building Commissioning 17900-1

Building Commissioning 17900-1 SECTION 17900 BUILDING COMMISSIONING PART 1 - GENERAL 1.01 WORK DESCRIPTION: A. General: 1. The Commissioning process is a joint team effort to ensure that all mechanical equipment, controls, and systems

More information

E-Version SECTION 15318 - CHLORINATION SYSTEM PART 1 - GENERAL PART 2 - PRODUCT

E-Version SECTION 15318 - CHLORINATION SYSTEM PART 1 - GENERAL PART 2 - PRODUCT SECTION 15318 - CHLORINATION SYSTEM PART 1 - GENERAL 1.1. SCOPE OF WORK A. Under this item the Contractor shall furnish and install a sodium hypochlorite solution metering system complete with all accessories

More information

Environmental Monitoring

Environmental Monitoring Chapter 10 Environmental Monitoring By the end of this chapter, you will be able to: l List the types of environmental testing l Name three types of checks that are carried out to ensure the isolator is

More information

D Youville College July 31, 2008 New Academic Building Cannon Design Project No. 002935.00

D Youville College July 31, 2008 New Academic Building Cannon Design Project No. 002935.00 SECTION 22 6119 - COMPRESSED-AIR EQUIPMENT FOR LABORATORY AND HEALTHCARE PART 1 - GENERAL 1.1 SUMMARY A. Section Includes: 1. Packaged, oilless scroll air compressors. 1.2 DEFINITIONS A. Actual Air: Air

More information

SECTION 10680 MediaStation Model 125 Electric Lateral Files (ELF)

SECTION 10680 MediaStation Model 125 Electric Lateral Files (ELF) SECTION 10680 MediaStation Model 125 Electric Lateral Files (ELF) PART 1-GENERAL 1.01 RELATED WORK SPECIFIED IN OTHER SECTIONS A. Electrical connections see electrical specifications. 1.02 SUBMITTALS A.

More information

SOHOline and SOHO Slimline The SOHOline and SOHO Slimline offer a modular, sturdy and compact design

SOHOline and SOHO Slimline The SOHOline and SOHO Slimline offer a modular, sturdy and compact design SOHOline and SOHO Slimline The SOHOline and SOHO Slimline offer a modular, sturdy and compact design Ideal installation for; Offices Branch Offices Schools Computer Aided Classes Government Offices Hotels

More information

SECTION 23 05 93 TESTING, ADJUSTING, AND BALANCING FOR HVAC INTRODUCTORY INFORMATION

SECTION 23 05 93 TESTING, ADJUSTING, AND BALANCING FOR HVAC INTRODUCTORY INFORMATION SECTION 23 05 93 TESTING, ADJUSTING, AND BALANCING FOR HVAC INTRODUCTORY INFORMATION The purpose of this guide specification is to assist the specifier in correctly specifying Mechanical System Testing

More information

Medical Air Systems: Scroll

Medical Air Systems: Scroll GENERAL The Chemetron skid mount scroll medical air system is designed to provide medical breathing air for hospital and medical institutions. This system meets NFPA 99 requirements for Level 1 breathing

More information

RPV - RP - RA air handling units

RPV - RP - RA air handling units APPLICATIONS Heating, ventilation, cooling >> FACTORIES >> WAREHOUSES >> HALLS >> DISTRIBUTION CENTRES >> COMMERCIAL PREMISES >> PLACES OF WORSHIP RPV - RP - RA air handling units www.reznor.eu RPV - RP

More information

Solar air collectors for industry and larger halls S Ø. Efficient dehumidification and air heating for free...

Solar air collectors for industry and larger halls S Ø. Efficient dehumidification and air heating for free... Solar air collectors for industry and larger halls N S Ø Efficient dehumidification and air heating for free... This booklet contains information about SolarVenti Industrial air solar system. The system

More information

All work shall comply with applicable codes and ordinances. Work shall comply with the BOCA National Plumbing Code, latest edition.

All work shall comply with applicable codes and ordinances. Work shall comply with the BOCA National Plumbing Code, latest edition. DIVISION 15 MECHANICAL SECTION 15100 - PLUMBING PART 1 - GENERAL SCOPE Plumbing work includes all domestic water, gas, and sanitary sewer systems as shown on plans. Also all fixtures and equipment which

More information

SECTION 1 GENERAL REQUIREMENTS

SECTION 1 GENERAL REQUIREMENTS Page 1 of 6 SECTION 1 GENERAL REQUIREMENTS 1. SCOPE OF WORK: The work to be performed under the provisions of these documents and the contract based thereon includes furnishing all labor, equipment, materials,

More information

PIEZOMETER AIRFLOW MEASURING RING

PIEZOMETER AIRFLOW MEASURING RING PIEZOMETER AIRFLOW MEASURING RING M.K. Plastics is pleased to introduce The Piezometer Airflow Measuring Ring, now available as an accessory for our steel Axijet fans. The system consists of a Piezometer

More information

UB1 AIR CONDITIONING UNIT INSTALLATION INSTRUCTIONS

UB1 AIR CONDITIONING UNIT INSTALLATION INSTRUCTIONS UB1 AIR CONDITIONING UNIT INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS: Carefully read these instructions before installing your new air-conditioner. AUSTRALIAN AUTOMOTIVE AIR AL00500054E 1 Table

More information

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W

Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W MODE ALARM Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MF08CESWW Voltage rating: 115V~60Hz Power rating : 800W Customer Support : 1-800-474-2147 For product inquiries

More information

SERIES ASM NEOPRENE/EPMD FLANGED SINGLE SPHERE CONNECTOR CONNECTORS. Pressures to 225 PSIG (15.51 barg) Temperatures to 230ºF (110ºC)

SERIES ASM NEOPRENE/EPMD FLANGED SINGLE SPHERE CONNECTOR CONNECTORS. Pressures to 225 PSIG (15.51 barg) Temperatures to 230ºF (110ºC) APPLICATIONS Process Industry Weak Acids Alkalies Compressed Air Pulp & Paper MODELS ASM - Flanged Connection OPTIONS Control Rods Oil & Gas Water & Waste Pump suction & discharge Sea water Chemical lines

More information

LAMINAR FLOW FOR ALL YOUR NEEDS

LAMINAR FLOW FOR ALL YOUR NEEDS LAMINAR FLOW FOR ALL YOUR NEEDS 2 LAMINAR FLOW FOR ALL YOUR NEEDS EXPLORE TECNIPLAST COMPLETE RANGE OF LAMINAR FLOW SOLUTIONS: OUR CUTTING-EDGE TECNOLOGY IS AT YOUR SERVICE, TO ASSURE TOP ENVIRONMENTAL,

More information

FVA/A1 saturated-steam sterilizer

FVA/A1 saturated-steam sterilizer FVA/A1 where technology innovation and design come together FVA/A1 saturated-steam sterilizer www.fedegariautoklaven.ch Fedegari's new FVA/A1 series of automatic vertical sterilizers has been expressly

More information