1 Chapter 10 Environmental Monitoring By the end of this chapter, you will be able to: l List the types of environmental testing l Name three types of checks that are carried out to ensure the isolator is working correctly Introduction This chapter examines the ways in which we monitor the cleanroom environments in which we work and also the operators performing aseptic manipulations. Why do we need to monitor the environment? Regular monitoring of the environment, process and finished product must occur according to a written procedure and in line with the published written standards and guidelines. All areas associated with the aseptic preparation process should be assessed for compliance with the standards on commissioning, after routine maintenance and routinely at a defined frequency. Written reports should be completed and assessed so that appropriate action can be taken if abnormal results are obtained. Techniques used for monitoring should be easy to perform, produce meaningful results and must not contribute to contamination. Chapter 10: Environmental Monitoring 1
2 Aseptic Processing What is environmental monitoring? Environmental monitoring is testing of the equipment, cleanrooms and people to: l check that the cleanrooms, laminar flow cabinets and isolators are performing as expected; l ensure that the facilities and equipment are controlling the environment within the correct limits; and l ensure that the SOPs are being followed to maintain the cleanroom environment. Types of environmental testing: l Physical Tests l Microbiological Tests l Chemical Tests l Integrity of the filters l Particle count l Room air change rates l LAFC/Isolator velocities These tests are usually carried out by quality control (QC) Microbiological tests include: Physical tests include: l Gauge checking for pressure and filter status of rooms l Pressure and filter status of Laminar Air Flow Cabinet (LAFC) l Glove leak test in isolator l Isolator leak test These tests are usually carried out by operators l Settle Plates l Contact Plates l Swabs l Finger Dabs (touch plates) l Active Air samples Chemical tests include: l Surface swabs for chemical residues e.g. cytotoxics l Surface wipes for chemical residues 2 Chapter 10: Environmental Monitoring
3 Chapter 10 Environmental Monitoring Test Yourself Complete the following sentence: Environmental monitoring is testing of the Write down three reasons why we carry out environmental testing Chapter 10: Environmental Monitoring 3
4 Aseptic Processing What methods and equipment do we use to test the environment? Passive Air Sampling Microbiological Testing Active Air Sampling Biotest RSC Plus Air Samplers Active air sampling machines are used to test the air. There are several types available. Bacterial counts are monitored when a sample of the air is drawn over a media on which microbes will grow. Tests are carried out in the manned (operational) state and results compared to standards set for the appropriate grade of environment. Settle Plates Settle plates are used to detect any bacteria that may be present in the environment that fall onto the plates whilst they are exposed, usually for 4 hours. The plates contain a medium that is ideal for most bacteria to grow on. Plates are exposed during each working session in the work zone. Plates are usually also exposed at least weekly in the clean room and hatches. Plates are incubated at a specific temperature and then checked for growth after a set period of incubation. The number of colonies grown is recorded and the levels compared against standards in the Orange Guide. 4 Chapter 10: Environmental Monitoring
5 Chapter 10 Environmental Monitoring Surface Sampling Operator Sampling Surface Swabs Swabs are taken from areas where dirt and bacteria are most likely to be present to check that the cleaning is being carried out effectively. Finger Dabs Contact Plates Image supplied by BiotestUK Contact plates have a raised agar surface. They are placed lightly onto the surface for 15 seconds and then incubated. They also check that cleaning has been carried out properly. Finger dabs are carried out by all operators after a working session. Sometimes an un-gloved before dab is used to assess hand washing techniques to ensure all dirt and bacteria are removed when scrubbing up. Finger dabs assess operator technique and the efficiency of item transfer techniques. If bacteria are present, this could be due to poor operator technique or poor spraying and wiping technique when bacteria are still present on the outer surface of bottles and containers or components which may have been passed into the LFC or Isolator. Chapter 10: Environmental Monitoring 5
6 Aseptic Processing Physical Testing Particle Testing Room Pressure Testing Gauges (Manometers) The differential pressure between the cleanrooms and the outer environment should be measured regularly. A schedule is provided in the Yellow Guide. Biotest APC handheld Airborne Particle Counter. At a suitable frequency, the Quality Control Department will visit to carry out non-viable airborne particulate counting. This assesses the number of particles which are present in the air to check on the level of contamination present at that time, both in the manned and unmanned states. Results are compared to standards. If the results are higher than the limit, investigations must be carried out to determine the cause of the problem. This may be due to problems with the air filtration systems. DOP (Dispersed Oil Particulates) Test This test is carried out to determine the filter integrity to ensure the filter has not been damaged. It also checks the filter housing to ensure there is no air by-pass due to poor seals. DOP is introduced as a 0.3 micron sized aerosol into the duct, upstream of the filter and a sample of the downstream air is passed across a light source and photometer. A photometer analyses the penetration and gives a pass/fail result. All operators should check the room pressure before using a cleanroom. Temperature and Humidity Thermometer The temperature of cleanrooms should be 20 o C ± 2 o C This should be checked regularly together with the relative humidity. 6 Chapter 10: Environmental Monitoring
7 Chapter 10 Environmental Monitoring Air Flow and Room Air Change Rate Anenometer Image supplied by Envair Laminar flow cabinets should provide airflow at 0.45 meters per second on average over the whole filter face. Airflow into a cleanroom is also measured so that the air change rate can be calculated. Other Testing In addition, lighting and noise levels should also be tested periodically. Isolator Leak Testing and Glove Leak Testing Isolators are leak tested on a weekly basis. Pressure decay (Isolator leak) tests are also carried out to ensure the integrity of the internal environment is maintained. Glove leak tests are carried out to check for holes that may not be visible and that the glove has been fitted correctly. Glove leak tests also check the sleeves for holes. Chapter 10: Environmental Monitoring 7
8 Aseptic Processing Environmental Monitoring In every manufacturing facility there is a schedule for testing which contributes to validation of cleaning and disinfection. All results of testing are recorded and results need to be within acceptable levels. Below is an example of a schedule of monitoring Type Frequency Where Manometer readings (gauges) Dailly Room Sessional LAFC/Isolator Settle Plates Weekly Room Surface Samples (contact plates, swabs) Sessional Weekly LAFC/Isolator Room Finger Dabs Sessional Workzone Below are the acceptable levels for all the different types of environmental monitoring. Grade of room Finger dabs /cfu/hand Settle plates 90mm cfu/4 hrs Contact plates 55mm cfu/m 3 Active air sample cfu/m 3 A <1 <1 <1 <1 B C N/A D N/A Reference: Orange Guide 2007 Any results which are out of limits i.e. outside the acceptable levels, must be reported immediately to the clean room supervisor or person in charge. It must also be documented as laid down in your SOP. Further actions or remedies must also be recorded. This is very important to help maintain the quality of the environment. Validation Tests Staff are expected to undertake evaluation tests both initially and at a designated frequency to check their aseptic technique to ensure their working practices do not contaminate the products. Further Reading: See Chapter 8 Quality Systems for Operators for detailed information on staff validation tests. 8 Chapter 10: Environmental Monitoring
9 Chapter 10 Environmental Monitoring Control of Air Flow: Cleanroom Monitoring Positive Air Flow In order to prevent dirty air entering the facility when doors or transfer devices are opened, the facility is maintained at positive pressure. This means that air will always flow out of the facility and dirty air cannot enter and contaminate the environment. Pressure gauges (manometers) between rooms should be checked in accordance with the SOP to ensure that the pressure is correct and that air is flowing in the right direction. Gauges are present that display the pressure differences between the areas in the facility. The gauges record values, which are between limits marked on the gauges or laid down in standard operating procedures (SOPs). Limits are 10Pa difference between cleanto-clean areas and 15Pa difference between clean-to-dirty areas. If the limits are not achieved, then no work is to take place until the correct pressure is established. This decision will be made by the Unit Manager. If the agreed values are not met/exceeded, then there may be a problem with the filters and so work should be suspended until this can be investigated. Calibration checks are carried out on manometers by Quality Control to assess the accuracy of the gauges. Pa = Pascals: This is the international unit of pressure. Control of Airflow: Workstation Monitoring HEPA Filter Gauges The HEPA gauge on workstation cabinets should be checked to monitor the blocking of the HEPA filter. The Laminar Air Flow Cabinet (LAFC) gauge should be checked to ensure: l The cabinet is switched on and functioning within acceptable limits l The HEPA filter is functioning within limits Isolator gauges should be checked to ensure: l The isolator is switched on and operating to protect the operator and product l The HEPA filter is functioning within limits l The total air change rate is operating within acceptable limits. The Laminar Air Flow Cabinets and isolators are specially designed workstations with super clean air in which the products should be prepared. Laminar Air Flow Cabinets should be left running at all times. Clean up times after interruptions, i.e. power failures, should be validated. Chapter 10: Environmental Monitoring 9
10 Aseptic Processing Environmental Monitoring Systems In this example we will look at a computer system which monitors the particles and pressures using monitors placed throughout an aseptic unit. Overview screen This screen shows a graph of the differential pressures of the unit for one day. Peaks and troughs show when the unit is occupied and operators move from one area to another. A Particle Monitoring System This screen display shows the full plan of an aseptic unit. Green symbols indicate that these zones are within limits. In an alarm state the symbols change to red, indicating increased levels of particles or pressures. Differential Pressure displays The software allows a number of ways of viewing the displays. Here is a close up of one of the particle monitors in the unit. This one is situated on a wall in the assembly room. They are also fitted into the ceilings. This unit has 24 ports situated throughout the entire unit. This screen shows the pressure differentials between rooms. Positive pressure must be maintained throughout. The green squares will change to red if the pressure falls below 10 Pascals (minimum standard between cleanrooms). 10 Chapter 10: Environmental Monitoring
11 Chapter 10 Environmental Monitoring This image shows a particle monitor within a LAFC. It is connected via tubing to an outlet in the wall. The pump in the plant room draws the air sample into the machine which then counts the particles in the area. (The other device is a CCTV camera used for checking see Chapter 8: Aseptic Services CCTV Checking Systems for further information). A screen displays the values received by the particle monitors around the unit. The system can count particles ranging is size from 0.3 micron to 25 micron. The air samples are pulled into the main controls which are situated in the pharmacy plant room by a system of ultra clean Bevaline tubing. Chapter 10: Environmental Monitoring 11
12 Aseptic Processing Questions Q1 Complete the table to show examples of each type of environmental test. Physical Microbiological Chemical Q2 Air: What methods do we use to test the quality of the following? Surfaces: People: 12 Chapter 10: Environmental Monitoring
13 Chapter 10 Environmental Monitoring 1. Q3 Give 2 examples of tests that we use to monitor HEPA filters. 2. Q4 How often are isolators leak tested? Name the types of leak tests for Isolators? (4) Chapter 10: Environmental Monitoring 13
14 Aseptic Processing 14 Chapter 10: Environmental Monitoring
Environmental Monitoring of Clean Rooms
Environmental Monitoring of Raul Duarte President DDK Scientific, Corp. Copyright DDK Scientific, Corp. 2008, 2009, DDK Scientific, Corp. Proprietary A manufacturing facility for pharmaceutical products
More informationEnvironmental Monitoring
Environmental Monitoring Purpose of Environmental Monitoring Critical process within the pharmaceutical and biotechnology industries. Determines the microbial and particulate content of cleanroom air and
More informationENVIRONMENTAL MONITORING
ENVIRONMENTAL MONITORING Assessment and verification of the adequacy of the aseptic compounding environment is essential. Environmental monitoring programs are designed to promptly identify potential sources
More informationCLEAN ROOM MONITORING REGULATORY STANDARDS
CLEAN ROOM MONITORING REGULATORY STANDARDS Air Classification as per Schedule M Maximum permitted number of particles / m3 equal or above at rest in operation 0.5µm 5.0µm 0.5µm 5.0µm A 3,520 29 3,500 29
More informationGUIDELINES ON TEST METHODS FOR ENVIRONMENTAL MONITORING FOR ASEPTIC DISPENSING FACILITIES
GUIDELINES ON TEST METHODS FOR ENVIRONMENTAL MONITORING FOR ASEPTIC DISPENSING FACILITIES PRODUCED BY A WORKING GROUP OF THE SCOTTISH QUALITY ASSURANCE SPECIALIST INTEREST GROUP SECOND EDITION FEBRUARY
More informationCleaning. By the end of this chapter, you will be able to: Introduction. Definitions. Chapter 9
Chapter 9 By the end of this chapter, you will be able to: l Define the terms cleaning, decontamination, disinfection l List the factors that affect disinfection l Describe the aims of a cleaning schedule
More informationCleanroom. For. Sterile Manufacturing Facilities
Cleanroom For Sterile Manufacturing Facilities Praphon Angtrakool Food and Drug Administration 1 WHO TRS No. 823 Annex 1, 1992 (1) General 17.1 The production of sterile preparations should be carried
More informationGuide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested
DOP Solutions Technical Document: Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested Issue 1: Prepared by Neil Stephenson and Tim Triggs and edited by
More informationParticle Monitoring Requirements in Pharmaceutical Cleanrooms
Particle Monitoring Requirements in Pharmaceutical Cleanrooms All drugs must be manufactured in accordance with the current Good Manufacturing Practice (cgmp) regulations. Pharmaceutical manufacturers
More informationMOHD NASRUL MOHAMAD NOOR
QUALITY ASSURANCE IN ASEPTIC WHO Collaborating Centre For Regulatory Control of Pharmaceuticals DISPENSING : MAINTENANCE OF CLEAN ROOM & EQUIPMENT Member of Pharmaceutical Inspection Cooperation Scheme
More informationEverything you want to know about USP 797
Everything you want to know about USP 797 Dr. Linda D. Lee Director of Operations Waste Management Healthcare Solutions September. 22, 2009 Goals High level understanding of USP 797 Who does it apply too
More informationEU GMP Annex 1 Update 2008: Airborne Particle Counting
EU GMP Annex 1 Update 2008: Airborne Particle Counting Airborne Particle Counting for Pharmaceutical Facilities: Update 2008, EU GMP Annex 1 Morgan Polen - VP of Applications Technology, Lighthouse Worldwide
More informationLaminar Flow Cabinets and Workstations
VERTICAL AND HORIZONTAL HEPA FILTERED SYSTEMS FOR PROCESS OR OPERATOR PROTECTION Laminar Flow Cabinets and Workstations For procedures requiring protection from particulate contamination and for aseptic
More informationInternational Standard ISO 14644
International Standard ISO 14644 Cleanrooms and associated controlled environments Praphon Angtrakool Food and Drug Administration 1 International Standard : ISO 14644 ISO 14644 consists of the following
More informationGMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE OF STERILE MEDICINAL PRODUCTS
PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PI 032-2 8 January 2010 RECOMMENDATION GMP ANNEX 1 REVISION 2008, INTERPRETATION OF MOST IMPORTANT CHANGES FOR THE MANUFACTURE
More informationManufacturing. Environmental Monitoring Particle Counts Are Easy B IOP ROCESSTECHNICAL. Scott E. Mackler
B IOP ROCESSTECHNICAL Environmental Monitoring Particle Counts Are Easy Scott E. Mackler Manufacturing pharmaceutical and biotechnology products requires that the appropriate level of quality be designed
More informationValidating and Monitoring the Cleanroom
6 Validating and Monitoring the Cleanroom Kevina O Donoghue Overview... 6-1 Cleanroom Design... 6-1 Principles of Cleanroom Validation and Testing... 6-2 Cleanroom Validation... 6-2 So How Does One Determine
More informationQualification of Cleanrooms and Environmental Monitoring - Practical Approaches in Europe
Qualification of Cleanrooms and Environmental Monitoring - Practical Approaches in Europe 1 Dr. Doris Herrmann Zum Eselsbachtal 20 67657 Kaiserslautern Doris.Herrmann@GMP-unlimited.de www.gmp-unlimited.com
More informationWorld-Class Oil Sampling It is Possible
www.lubecontrol.com.au World-Class Oil Sampling It is Possible Lube Control Pty Ltd Providing lubrication solutions! Unit 5/53 Norfolk Rd, Marion, SA 5043 t: (08) 8298 5563 f: (08) 8298 6253 e: info@lubecontrol.com.au
More informationQualification of an Environmental Monitoring Program
[ Scott Sutton Qualification of an Environmental Monitoring Program Scott Sutton Microbiology Topics discusses various topics in microbiology of practical use in validation and compliance. We intend this
More informationLab Exercise 3: Media, incubation, and aseptic technique
Lab Exercise 3: Media, incubation, and aseptic technique Objectives 1. Compare the different types of media. 2. Describe the different formats of media, plate, tube etc. 3. Explain how to sterilize it,
More informationUSP <797> Cleanroom Design and Environmental Monitoring. Andrew King, USP <797> Specialist CETA Member RCCP-SC
USP Cleanroom Design and Environmental Monitoring Andrew King, USP Specialist CETA Member RCCP-SC Objectives The objectives of this presentation: To summarize basic cleanroom concepts and how
More informationCLASSIFICATION OF CLEANROOMS
CLASSIFICATION OF CLEANROOMS Cleanrooms are classified by the cleanliness of their air. The method most easily understood and most universally applied is the one suggested in the earlier versions (A to
More informationParticle Counting and Environmental Monitoring In Pharmaceutical / Life Sciences
Particle Counting and Environmental Monitoring In Pharmaceutical / Life Sciences LIGHTHOUSE WORLDWIDE SOLUTIONS WWW.GOLIGHTHOUSE.COM Introduction The pharmaceutical and medical device industries are developing
More informationStainless Steel Worksurface 3 Feet Hood A10-WS-PR3 A10-WS-SR3 4 Feet Hood A10-WS-PR4 A10-WS-SR4 6 Feet Hood A10-WS-PR6 A10-WS-SR6.
Product Sampler #1 Laminar Flow Hoods / Laminar Flow Cabinets: 1- Benchtop Vertical Laminar Flow Hoods- Air Recirculation Available s: Clean Bench Class 100 Bench-top Vertical Laminar Flow Hoods with Air-recirculation
More informationAir and Water Balancing - Sound and Vibration Testing - Cleanroom Services - Building Commissioning Building Envelope Testing Thermal Imaging
NEUDORFER ENGINEERS, INC. provides our customer with unbiased, professional, performance testing of Cleanrooms. We are committed to providing you, our customer, with value-added expertise in an effort
More informationLiabilities of Vented Crawl Spaces And Their Impacts on Indoor Air Quality in Southeastern U.S. Homes
Liabilities of Vented Crawl Spaces And Their Impacts on Indoor Air Quality in Southeastern U.S. Homes Jonathan Coulter, Bruce Davis, Cyrus Dastur, Melissa Malkin-Weber and Tracy Dixon Advanced Energy,
More informationThe use of risk assessment tools for microbiological assessment of cleanroom environments. by Tim Sandle
The use of risk assessment tools for microbiological assessment of cleanroom environments by Tim Sandle Email: tim.sandle@bpl.co.uk / timsandle@btinternet.com Web: www.pharmig.blogspot.com Environmental
More informationLaminar Flow Cabinets
Laminar Flow Cabinets The World s Most Practical Selection of Benchtop Laminar Flow Cabinets. 24 36 48 Simple, Effective Protection for Samples and Work Processes, Uncompromised Performance, Economical
More informationG U I D E L CERTIFICATION EDUCATION TRAINING. Procurement Guideline. National Air Filtration Association Recommended Practice
G U National Air Filtration Association Recommended Practice I D E L I N E Procurement Guideline TRAINING EDUCATION CERTIFICATION HVAC Air Filter Purchasing Specification Guideline 1.0 Purpose To provide
More informationRisk Assessment: Biosafety Cabinet
Risk Assessment: Biosafety Cabinet Biosafety cabinets (BSCs) serve as primary containment devices for operations involving potential splashes, spills or aerosolization of hazardous biological materials.
More informationCleaning and Disinfection Agenda. 1. Cleanroom Classification. Cleanroom Classification. What does contamination mean? Types of contamination
Cleaning and Disinfection Agenda CLEANING AND DISINFECTION TECHNIQUES Barbara Wessels Key Account Manager Ecolab Contamination Control October 2014 1. Cleanroom classification and definition 2. Sources
More informationHow To Control A Medical Device With A Pc System
Clean room monitoring and disinfection FMS 5.0, the Leading Facility Monitoring Software TSI FMS system is an advanced, reliable, and user-friendly monitoring software suite that features a truly open
More informationHEPA-filtered Negative Air Machines and their Role in Creating Airborne Infection Isolation Rooms with the ECU AnteRoom. Frequently Asked Questions
HEPA-filtered Negative Air Machines and their Role in Creating Airborne Infection Isolation Rooms with the ECU AnteRoom Frequently Asked Questions What is a negative air machine and why do I need one?
More informationHVAC Design Requirements
Maintenance of HVAC HVAC Design Requirements Air Flow Diagram แบบ Recirculation Air Flow Diagram แบบ Once Through Definition Of Conditions As built At rest In operation air air air Maintenance GMP Requirements
More informationPharmagraph. GMP/GLP Monitoring System
Pharmagraph GMP/GLP Monitoring System Pharmagraph GMP/GLP Monitoring System Demonstrates GMP/GLP compliance for cleanrooms and laboratories Monitors, Alarms, Logs and Reports Airborne particle counts,
More informationEACO DOP / PAO TESTING GUIDELINE 2013
2013 Foreword The guideline promotes industry standards for training, equipment, and methodologies for HEPA filter integrity testing of negative air units, vacuums and any other HEPA filtered equipment
More informationCleanroom Solutions Ltd T: 01273 760 060 F: 01273 760 061 E: mail@cleanroom-solutions.co.uk W: www.cleanroom-solutions.co.uk
Cleanroom Solutions Ltd T: 01273 760 060 F: 01273 760 061 E: mail@cleanroom-solutions.co.uk W: Introduction Cleanroom Solutions Ltd has a wealth of experience together with a solid track record in the
More informationTSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS
TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS APPLICATION NOTE CC-105 (US) Introduction This application note will demonstrate how real-time viable particle count data will
More informationRecommendations for the Safe Use of Handling of Cytotoxic Drugs
Recommendations for the Safe Use of Handling of Cytotoxic Drugs Introduction Cytotoxic drugs are toxic compounds and are known to have carcinogenic, mutagenic and/or teratogenic potential. With direct
More informationDesign, Operation and Management of GTP/GMP Cell Engineering Facilities
Design, Operation and Management of GTP/GMP Cell Engineering Facilities Scott R. Burger, MD Advanced Cell & Gene Therapy BFDA 2007 International Symposium on Regulation of Human Cell and Tissue- Based
More informationSPECIAL APPLICATIONS
Page 1 SPECIAL APPLICATIONS The protection of microelectronics clean rooms using aspirating smoke detection systems requires special consideration. The physical configuration of the room(s) and the direction
More informationBoth health-care facility classification and area classification, as per CSA, determine the design requirements.
HEATING, VENTILATION AND AIR-CONDITIONING SYSTEM (HVAC) Class 1 Facility: General Anesthesia Heating, ventilation and air-conditioning system standards are a primary infection prevention and control strategy
More informationIndoor air quality monitoring made easy
Indoor air quality monitoring made easy with the Fluke 985 Particle Counter Application Note Introduction Energy savings is top of facilities owners agenda. HVAC system tends to be the largest contributor
More informationDIF O.T. Ceiling filtration system for operating theatres
DIF O.T. Ceiling filtration system for operating theatres Air filtration and distribution for operating theatres The operating rooms are the hospital areas where it is required the most accurate control
More informationCLEANROOMS IN PHARMACEUTICAL PRODUCTION
Mikhail Kitain CLEANROOMS IN PHARMACEUTICAL PRODUCTION Bachelor s thesis Building Services March 2010 DESCRIPTION Date of the bachelor's thesis 14.03.2010 Author(s) Mikhail Kitain Degree programme and
More informationDesign, Construction, Commission, and Qualification of Critical Utility Systems: Part III
Design, Construction, Commission, and Qualification of Critical Utility Systems: Part III BY DAVID W. VINCENT AND HERBERT MATHESON HEATING, VENTILATION, AND AIR CONDITIONING (HVAC) SYSTEMS INTRODUCTION
More informationLaboratory Manual for Air Quality Sampling and Analysis
Laboratory Manual for Air Quality Sampling and Analysis Department of Civil Engineering Indian Institute of Technology Delhi New Delhi-110016 MONITORING OF RESPIRABLE PARTICULATE MATTER 1. Objective Sampling
More informationBIOLOGICAL SAFETY CABINET PROGRAM
BIOLOGICAL SAFETY CABINET PROGRAM TABLE OF CONTENTS: 1. INTRODUCTION 2. PURPOSE AND SCOPE 3. SELECTION AND INSTALLATION 4. USE OF BIOLOGICAL SAFETY CABINETS 5. CERTIFICATION OF BIOLOGICAL SAFETY CABINETS
More informationBoth health-care facility classification and area classification, as per CSA, determine the design requirements.
HEATING, VENTILATION AND AIR-CONDITIONING SYSTEM (HVAC) Class 3 Facility: Local Anesthesia Heating, ventilation and air-conditioning system standards are a primary infection prevention and control strategy
More informationFacility Cleaning & Monitoring. Cleaning
Facility Cleaning & Monitoring Cleaning GTP Regulations: 21CFR Part 1271.190 Maintain facility in clean, sanitary & orderly manner Establish & maintain procedures for facility cleaning Assign responsibility
More informationCETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG-007-2010 March 24, 2010
CETA Application Guide for the Exhaust System Requirements of Class II, Type B Biosafety Cabinets CAG-007-2010 March 24, 2010 1. Background Class II Type B Biosafety Cabinets (BSCs) are unique laboratory
More informationRegulations Concerning
Regulations Concerning Airborne Particle Counting Two types of activities performed with an airborne particle counter: Classification focuses primarily on the environment Particle Counting Applications
More informationSample Pharmacy CLEANING AND SANITIZING PROCEDURE P-304.1
P-304.1 1.0 Definition and Purpose: To standardize and define the process by which the controlled Cleanroom environments (ISO Class 5, 7, and 8) and the general pharmacy prep area is cleaned, disinfected,
More informationProcedures for the Effective Use of Biological Safety Cabinets
Procedures for the Effective Use of Biological Safety Cabinets APPROVAL DATE BY BIOHAZARDS SUBCOMMITTEE: August 10, 2012 SUPERSEDES: March 1, 2012 This corporate guideline/procedure is intended as a minimum
More informationLiabilities of Vented Crawl Spaces, Their Impacts on Indoor Air Quality in Southeastern U.S. Homes and One Intervention Strategy
Liabilities of Vented Crawl Spaces, Their Impacts on Indoor Air Quality in Southeastern U.S. Homes and One Intervention Strategy Jonathan Coulter 1 ABSTRACT This paper documents the influences vented crawl
More informationPI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3)
PI s Name Date Bldg./Rm# CDC Biosafety Level 3 (BSL-3) Yes No 1. Is access to the laboratory limited or restricted at the discretion of the laboratory director when experiments are in progress? Yes No
More informationFumigation of Safety Cabinets
Fumigation of Safety Cabinets General Microbiological safety cabinets should always be fumigated if a large spillage of infectious material occurs within them, before filters are changed or before any
More informationHEPA FILTER DEFECTS ARE COMMON
HEPA FILTER DEFECTS ARE COMMON Kenneth F. Cherry 1 and Kenneth A. Smigielski 2 1 NCEC Inc. 2 First Solar LLC. ABSTRACT HEPA filters (High Efficiency Particulate Air) filters are claimed by vendors to have
More informationRisk-Based Environmental Monitoring. Marsha Stabler Hardiman Senior Consultant Concordia ValSource Wednesday September 17, 2014 FDA/PQRI
Risk-Based Environmental Monitoring Marsha Stabler Hardiman Senior Consultant Concordia ValSource Wednesday September 17, 2014 FDA/PQRI Presenter Marsha Stabler Hardiman Over 20 years experience in the
More informationApproval Block. Prepared by: Signature Date. Evan Parnell 08 NOV 2013. Reviewed by: Signature Date. Approved by: Signature Date
ATS-ATP-3460 Page: 1 of 5 Approval Block Prepared by: Signature Date Evan Parnell 08 NOV 2013 Reviewed by: Signature Date Brian Flynn 08 NOV 2013 Approved by: Signature Date Kristal Jewell 08 NOV 2013
More informationSubpart 1. Installation. All plumbing systems must be. installed and tested according to this chapter and chapter 4715,
4658.4500 PLUMBING SYSTEMS; NEW CONSTRUCTION. Subpart 1. Installation. All plumbing systems must be installed and tested according to this chapter and chapter 4715, the Minnesota Plumbing Code. Subp. 2.
More informationFlowFAST V. Vertical Laminar Airflow Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.
Vertical Laminar Airflow Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE. Vertical Laminar Flow Cabinets BEYOND MINIMUM SAFETY REQUIREMENT Cabinets are ISO Class 3 vertical laminar flow cabinets which
More informationCORESTA RECOMMENDED METHOD Nº 81
CORESTA RECOMMENDED METHOD Nº 81 ROUTINE ANALYTICAL MACHINE FOR E-CIGARETTE AEROSOL GENERATION AND COLLECTION DEFINITIONS AND STANDARD CONDITIONS (June 2015) 0. INTRODUCTION This Method includes the requirements
More informationEUROPEAN COMMISSION ENTERPRISE AND INDUSTRY DIRECTORATE-GENERAL. EudraLex The Rules Governing Medicinal Products in the European Union
EUROPEAN COMMISSION ENTERPRISE AND INDUSTRY DIRECTORATE-GENERAL Consumer goods Pharmaceuticals Brussels, 25 November 2008 (rev.) EudraLex The Rules Governing Medicinal Products in the European Union Volume
More informationASBESTOS SCIENCE TECHNOLOGIES, INC.
ASBESTOS SCIENCE TECHNOLOGIES, INC. P.O. Box 505 Bangor, CA 95914 Phone (530) 518-0934 Email - astinc17@yahoo.com Report for Lincoln Elementary School 1582 Lincoln Road Yuba City, Ca. 95993 Mold Report
More informationATS-SOI-5910 Page: 1 of 7. Approval Block. Prepared by: Signature Date. Evan Parnell 04 AUG 2014. Reviewed by: Signature Date. Brian Flynn 04 AUG 2014
ATS-SOI-5910 Page: 1 of 7 Approval Block Prepared by: Signature Date Evan Parnell 04 AUG 2014 Reviewed by: Signature Date Brian Flynn 04 AUG 2014 Approved by: Signature Date Kristal Jewell 04 AUG 2014
More informationLaboratory Biosafety Level 3 Criteria
Laboratory Biosafety Level 3 Criteria Biosafety in Microbiological and Biomedical Laboratories (BMBL) 5th Edition Section IV Biosafety Level 3 (BSL-3): Biosafety Level 3 is applicable to clinical, diagnostic,
More informationEnvironmental Monitoring Risk Assessment
Environmental Monitoring Risk Assessment By Tim Sandle INTRODUCTION Environmental Monitoring describes the microbiological testing undertaken in order to detect changing trends of microbial counts and
More informationENVIRONMENTAL MONITORING / TESTING TO CONFORM WITH ISO14644 & TGA GMP REQUIREMENTS
ENVIRONMENTAL MONITORING / TESTING TO CONFORM WITH ISO14644 & TGA GMP REQUIREMENTS PRESENTATION TO RACI 20 TH APRIL 2016 PAUL MORGAN DIRECTOR CRITICAL CERTIFICATION PTY LTD CHANGES TO ISO14644 PART 1 CHANGES
More informationAsbestos in the Workplace: A Guide to Removal of Friable Asbestos Containing Material
Asbestos in the Workplace: A Guide to Removal of Friable Asbestos Containing Material Revised November 21, 2013 Application Code of Practice Where asbestos is present or believed to be present in a workplace
More informationDIESEL FUEL CONDITIONING
COD FILTRATION DIESEL FUEL CONDITIONING REMOVE WATER & PARTICLES TO EXTEND FUEL INJECTOR LIFE & INCREASE COMBUSTION ENGINE FUEL EFFICIENCY Remove free water to saturation point with high single pass efficiency
More informationStandard Hydraulic Piping Flushing Procedure
Page 1 of 6 1.0 Scope This procedure covers the minimum technical requirements for cleaning and flushing of Hydraulic and Piping, and related accessories. 2.0 Introduction In hydraulic fluid power systems,
More informationPERFORMANCE QUALIFICATION OF A CLEAN ROOM FROM A QUALITY PERSPECTIVE
PERFORMANCE QUALIFICATION OF A CLEAN ROOM FROM A QUALITY PERSPECTIVE Every facility design is unique for its specific needs, and validation programs must be customized to meet these challenges. While there
More informationCal Water 1961 Petra Lane, Placentia, CA 92870 (800) CAL-WATER (800) 225-9283 FAX: (714) 792-0794 http://www.cal-water.com
Industrial Water Purification MAINTAINING A STERILE HIGH PURITY WATER SYSTEM Introduction Maintaining a microbe free deionized water system is like trying to maintain a vacuum, and nature abhors a vacuum.
More informationGENANO 4500 MEDICAL AIR PURIFICATION DEVICE
GENANO AIR PURIFICATION TECHNOLOGY GENANO 4500 MEDICAL AIR PURIFICATION DEVICE OPERATING INSTRUCTION 1 GENANO LTD Kimmeltie 3 FI- 02110 Espoo, Finland Tel + 358 9 7743 870 Fax + 358 9 7533 130 www.genano.fi
More informationPresentation at the 3 rd SAFOODNET seminar
HYGIENE SURVEY IN ROMANIAN BAKERIES ALINA DOBRE, IBA, ROMANIA & SATU SALO, VTT, FINLAND INTRODUCTION Hygiene survey was performed in six Romanian bakeries. The work was carried out at VTT Technical Research
More informationDRAFT GUIDANCE. This guidance document is being distributed for comment purposes only.
Guidance for Industry Current Good Manufacturing Practice Interim Guidance for Human Drug Compounding Outsourcing Facilities Under Section 503B of the FD&C Act DRAFT GUIDANCE This guidance document is
More informationEASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration
More informationQuestions & Answers (Q & A) Regarding Asbestos Treatment with the Fire Suppression and Renovation Project
Questions & Answers (Q & A) Regarding Asbestos Treatment with the Fire Suppression and Renovation Project The Watergate Community is currently in the process with the fire suppression and interior renovations
More informationVALIDATION OF CLEAN ROOMS FOR ASEPTIC MANUFACTURING. F.R.Smal Pharmaceutical Director ICCE ICCE a member of SNC-Lavalin Group
VALIDATION OF CLEAN ROOMS FOR ASEPTIC MANUFACTURING F.R.Smal Pharmaceutical Director ICCE ICCE a member of SNC-Lavalin Group tel + 3226431600 E-Mail: francis.smal@be.snclavalin.com VALIDATION OF CLEAN
More informationUSER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 THANK YOU FOR CHOOSING YOUR NEW DEHUMIDIFIER. BEFORE USING THE UNIT READ THESE INSTRUCTIONS FULLY AND RETAIN THEM FOR FUTURE REFERENCE
More informationContamination Control and Environmental Monitoring Systems
Life Science Contamination Control and Environmental Monitoring Systems Hospital Electronics Brookhaven Instruments deliver, install and validate particle counters, Microbial Air Samplers and Monitoring
More informationTransferring a Broth Culture to Fresh Broth
Sterile Technique It is very important in microbiology to work with pure cultures. Unfortunately this is difficult. The world around us is covered with microorganisms. Microorganisms are even carried on
More informationNational Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook
National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook Trainee:... Training Provider:... Year:... Trainee Contact no:... Competency Based Skill Training Logbook Introduction
More informationA Basic Introduction to Clean Rooms
A Basic Introduction to Clean Rooms A cleanroom is a controlled environment where products are manufactured. It is a room in which the concentration of airborne particles is controlled to specified limits.
More informationWhite Paper. Life Sciences Industry Solution Blueprint for Environmental Monitoring Systems (EMS) ISPE Guidance and Best Practices.
White Paper Life Sciences Industry Solution Blueprint for Environmental Monitoring Systems (EMS) Author: Roberto Zerbi, Global Industries Solutions Life Sciences, Invensys What s Inside: 1. Introduction
More informationBasic Requirements For Aseptic Manufacturing Of Sterile Medicinal Products A Comparison Between Europe And USA
Basic Requirements For Aseptic Manufacturing Of Sterile Medicinal Products A Comparison Between Europe And USA Wissenschaftliche Prüfungsarbeit zur Erlangung des Titels Master of Drug Regulatory Affairs
More informationInfection Control Risk Assessment Matrix of Precautions for Construction & Renovation
Infection Control Risk Assessment Matrix of Precautions for Construction & Renovation Step One: Using the following table, identify the Type of Construction Project Activity (Type A-D) TYPE A TYPE B TYPE
More informationCHILDCARE SETTINGS. Childcare Settings
Childcare Settings This document provides guidelines and recommendations for cleaning childcare settings such as nurseries, crèches and schools in the event of a disease outbreak in the vicinity. Cleaning
More informationParticle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations
Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations B e c k m a n C o u l t e r L i f e S c i e n c e s 2 5 0 S. K r a e m e r B o u l e
More informationHEPA Filters for Pharma Applications Cleanrooms & Safety Applications. Gordon Farquharson June 2012
HEPA Filters for Pharma Applications Cleanrooms & Safety Applications Gordon Farquharson June 2012 1 Air Filters at work Images courtesy of CamfilFarr 2 How filters work Sieving Effect Straining 3 Filter
More informationClass II Type B2 Laminar Flow Biological Safety Cabinets at Ordering Criteria for the National Institutes of Health
Background Class II Type B2 Laminar Flow Biological Safety Cabinets at Ordering Criteria for the National Institutes of Health This document provides guidelines for the design and installation of Class
More informationRaleigh, NC May 12 th & 13 th 2016. Meeting The New ISO Cleanroom Standards And Changing EU Guidance How Will You Comply?
Raleigh, NC May 12 th & 13 th 2016 Meeting The New ISO Cleanroom Standards And Changing EU Guidance How Will You Comply? The International Cleanroom Standards (ISO 14644-1 & 2) have been updated, these
More informationAseptic Technique. A GMP/GTP Training Module
Aseptic Technique A GMP/GTP Training Module Aseptic Technique The GMP Facility manufactures products for clinical use These products must meet a number of requirements, one of which is that they are sterile
More informationCARE HOMES AND NURSING HOMES
Care Homes and Nursing Homes: This document provides guidelines and recommendations for cleaning care homes and nursing homes in the event of a disease outbreak in the vicinity. Cleaning and disinfection
More informationFume Extraction: Guide to Safely Managing Solder Fumes in the Workplace
Fume Extraction: Guide to Safely Managing Solder Fumes in the Workplace Our thanks to OK International for allowing us to reprint the following. Why do we need fume extraction? Today s Electronics workplace
More informationWelcome to chapter 10. The following chapter is called Equipment maintenance and monitoring. The author is Ronny Janssens.
Welcome to chapter 10. The following chapter is called Equipment maintenance and monitoring. The author is Ronny Janssens. 1 After finishing this chapter, the student should be familiar with - the requirements
More informationAir Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection
Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection A wide range of products to meet the needs of transferring animals, operators,
More information#9. EFFECTIVE CLEANING, MAINTENANCE, AND PEST CONTROL PROGRAMS FOR YOUR HACCP PLAN
#9. EFFECTIVE CLEANING, MAINTENANCE, AND PEST CONTROL PROGRAMS FOR YOUR HACCP PLAN by O. Peter Snyder, Jr., Ph.D. Hospitality Institute of Technology and Management In past articles, I have discussed supplier
More information