MRI SCANNERS: A BUYER S GUIDE

Size: px
Start display at page:

Download "MRI SCANNERS: A BUYER S GUIDE"

Transcription

1 09 MRI SCANNERS: A BUYER S GUIDE MRI SCANNERS: A BUYER S GUIDE D. Price, I. Delakis, C. Renaud and R. Dickinson Correspondence Centre for Evidence-based Purchasing 152C Skipton House 80 London Road London SE1 6LH UK Website: The UK National Health Service (NHS) Purchasing and Supply Agency (PASA) Centre for Evidence-based Purchasing (CEP): has produced a series of comparative reports on the specifications and technical performance of both 1.5 and 3 Tesla (T) closed system magnetic resonance imaging (MRI) equipment, to help inform buyers purchasing decisions. Seven 1.5 T models, produced by GE, Phillips, Siemens and Toshiba, and four 3 T models, produced by GE, Phillips and Siemens, were evaluated (figure 1; table 1). The latest reports, which were first published in May 2007, form the basis of this article [1, 2]. Developments in MRI Effective diagnostic imaging has a key role in ensuring that patients have speedy access to appropriate treatment. The UK Department of Health has prioritised imaging as part of its progress towards meeting its 18-week referral-to-treatment target, due to be implemented by the end of Most hospitals buy 1.5 T systems. These have an important role in cancer diagnostics, including the identification of metastases from primary tumours of the respiratory system. Until recently, the more expensive 3 T systems have been used primarily for research, but they are becoming increasingly 138 THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS

2 MRI SCANNERS: A BUYER S GUIDE 09 Table T and 3 T models evaluated. GE Philips Siemens Toshiba 1.5 T Signa HDx Signa HDe Intera Achieva MAGNETOM MAGNETOM Excelart model Symphony Avanto I Vantage class & T class 3 T Signa HDx Achieva Allegra Trio, A Tim model X series System I class & T class popular for clinical work. They offer the possibility of higher image quality and shorter scanning times. Hitherto, MRI has not been the modality of choice for imaging the respiratory system. The air content in the lungs usually makes it difficult to produce sufficient image contrast at an appropriate spatial resolution. Blurring can be present in the images because of patient motion and it can also occur because the magnetic resonance signal from the lungs decays very rapidly. This article discusses the technological features available on the latest MRI systems that improve image quality and imaging speed and have enabled studies of lung structure and function using MRI [3]. These have included examinations of blood flow in the pulmonary vessels and blood perfusion in lung tissue [4, 5]. Normally these studies are performed with the use of injected contrast agents. Respiratory gating may be needed to coincide the imaging with the patient s breathing. Such gating can reduce motion blurring although ideally, if the scan time can be reduced enough, images can be acquired in a single breath-hold. Further developments are opening up the potential for MRI to be used much more widely in the diagnosis and management of respiratory disease. These include inhaled gas contrast agents (such as hyperpolarised helium), which enable MRI to track how air is being used in the lungs [5, 6]. How does MRI work? MRI provides excellent structural and functional images of the soft tissues, organs and vasculature in any desired spatial plane. This is achieved when the patient is placed inside a powerful magnetic field. Most modern scanners are based on superconducting magnets, which are cooled with liquid helium. The billions of protons that constitute the nuclei of hydrogen atoms in the body have magnetic properties. These continually spin in different directions, but line up with the pull of the main magnetic field. Once the protons are spinning in unison, they can all be tipped out of their alignment with the main magnetic field by applying a pulse of energy in the radiofrequency (RF) range. This is achieved with an RF transmit coil that is normally built into the scanner. After the RF pulse has been applied, the protons return to their original positions, releasing energy in the form of a signal, which carries information about specific tissues. The signal is collected by an RF receive coil, a range of which are provided for different areas of the body. The pulse sequences, which describe a pattern of pulses, are designed to provide differing image contrast for different body tissues, such as the brain or the heart. Three further coils built into the scanner, known as the gradient coils, are switched rapidly on and off during the scan. The magnetic fields applied by the gradient coils vary with position, encoding the signal with spatial information so that an image can be constructed. The timing of the RF and gradient pulses, and conversion of the signals into images, is controlled by computer. Design features The latest ranges of cylindrical models have been designed with patient comfort in mind, offering wider flared openings to the cylinder, shorter scan lengths, and less acoustic noise. Higher specifications usually mean better-quality images and/or reduced scan times, plus the ability to perform the most advanced applications. But it might be that THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS 139

3 09 MRI SCANNERS: A BUYER S GUIDE not all the various pulse sequence packages that come with these models are required. Inevitably, higher-specification machines also tend to be more expensive. However, these options could be worth considering, as it may be easier to add on further applications as the technology and clinical need develop. All models come with certain standard features. These include: A set of core pulse sequence packages to cover the entire anatomy. The higher-specification models have a range of optional additions for specific applications, such as spectroscopy (see below). A suite of RF coils to cover the entire anatomy: head, body, shoulder, spine coils, etc. Key purchasing factors to consider 1. Technical features The strength of the main magnetic field will affect image quality and the number of applications for which the scanner can be used. For instance, advanced neurological imaging techniques, such as diffusion MRI, are normally performed at 1.5 T and above. Low field systems can produce images of excellent diagnostic quality for many applications. However in general, the higher the strength, the better will be the quality of the images produced, because a higher magnetic field provides a higher signal-to-noise ratio. It should be noted, however that the magnetic resonance signal from lung may decay even quicker at higher fields, so any benefit may not be realised unless the imaging time can be reduced further through the use of more powerful gradient coil systems. The signal-to-noise ratio (SNR) is the most important parameter defining image quality in MRI. A low SNR means that the contrast between different tissues can be obscured by background noise. On the other hand, a high SNR provides more options. For example, it can be partially traded off to obtain improved spatial resolution (smaller pixels and finer detail) or faster images through parallel imaging. Faster scanning reduces the amount of blurring caused by movement during breathing. Field uniformity and stability are important for a good-quality image. Good magnetic field homogeneity over a large volume, which is reflected in a large maximum field of view, means that image quality is maintained even on the peripheral areas of the body. Sometimes the high signal from fat tissue will need to be suppressed, because it can obscure important detail in the image. This can be achieved using special pulse sequences but it may not be successfully suppressed if magnetic field homogeneity is poor. Homogeneity is normally defined as the variation in magnetic field in parts per million over a defined spherical volume. Shim coils, which produce a relatively small current, readjust the disturbance to the magnetic field caused by the patient inside the scanner, a process called active shimming. The gradient coils are used for active shimming, but optional dedicated shim coils may also be available. Field uniformity and stability are even more important for spectroscopy, which analyses chemical information at the molecular level and is primarily used in oncology. The radiofrequency system refers to the number of independent channels that can receive signals from the RF coils. Multichannel RF coils feature several receiving elements, each of which can feed into an independent channel in the system. A higher number of channels boosts the SNR and offers the option of parallel imaging. A higher SNR could be useful for lung imaging, where there is more air than tissue. Multichannel body coils are likely to be the most important component for this application. Gradient coil systems provide the images in any desired plane (figure 1). The magnetic field gradient applied by the coil is measured in mt m -1 while the rate at which the gradient can be changed, the slew rate, is measured in mt m -1 ms-1. Higheramplitude gradients, which can be switched on and off quickly (high slew rate) offer a stronger gradient pulse within a shorter time frame. This provides the same or even better spatial resolution for a shorter scan time. Various imaging parameters can be selected to determine the timings of the RF and gradient pulses. For instance, the repetition time, defined as the time between successive RF pulses, affects image contrast and scan time. Since the MR signal from lung tissue decays rapidly short time-imaging parameters are often essential. High-specification gradients are needed for this. Parallel imaging techniques use information from the RF coils to help spatially encode the magnetic resonance signal, thus reducing the number of times that the gradient coils have to be switched on and off, and speeding up the imaging process. Various commercial packages are available. The factor by which the scan time is reduced is known as the parallel imaging factor. A factor of 2 will suit most applications. This technique is widely used in breath-hold imaging. But higher parallel imaging factors reduce SNR, so higher-order factors have not been used widely in clinical practice to date. Multichannel coils are required for parallel imaging. 2. Ease of use and safety Patient comfort: about one in 10 patients feel claustrophobic inside the closed cylinder and models with wider apertures can help minimise this. Wider, longer tables will accommodate heavier patients, which may become more important if the prevalence of obese patients continues to rise. Large RF coils are quite heavy and some models incorporate spine coils built into the 140 THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS

4 MRI SCANNERS: A BUYER S GUIDE 09 patient couch. Some also allow all the coils to be plugged in at once, which speeds up the process for both patient and staff. Safety: MRI equipment has a good safety profile. Most adverse incidents are the result of incorrect procedures or are related to the design of the MRI site. Active shielding curbs the extent of the fringe magnetic field around the scanner, while built-in limits in the scanner protect the patient from excessive RF energy. Z Y X Ferromagnetic objects, such as oxygen cylinders, can be sucked into the main magnetic field and so present a projectile hazard. It s also important to check whether patients and staff have implants, metal pins, or pacemakers. Noise levels can exceed 100 db, although some models now feature noise-control technology that reduces this to <90 db for most scans. Hearing protection is essential for patients and for staff remaining in the scan room during imaging. Dimensions: MRI scanners take up a lot of space because of the need to allow for the extent of the magnetic field, and the heavier models may require the floor to be reinforced. Manufacturers specify the minimum installation area, but if space is constrained, large amounts of magnetic field shielding may be required to prevent the fringe field from extending into public areas. Compatibility with other existing equipment needs to be checked to ensure it is safe to use near the scanner and that it will not interfere with the images. Normally there are safe operating conditions that must be followed: for instance, certain equipment may be safe only if kept a certain distance from the magnet. 3. Costs Capital outlay: you can expect to pay a minimum of 700,000 Figure 1. MRI gradient directions X, Y and Z. ( 1 million) for a 1.5 T model and from 1 million ( 1.5 million) for a 3 T model. Some level of service and on-site training should be included in this price. Manufacturers will sometimes include accessories in the overall bundle of costs, but these can vary, so check which accessories are included. Manufacturers can offer various financial packages. Accessories include an independent workstation, which allows you to view images and perform image processing away from the main console. Other accessories you may require include: contrast injectors, cardiac, respiratory and oxygen monitors, additional beds and detachable trolleys. If these are not included in the package, you need to consider where you are going to source them. Similarly, check what sequence packages are included, and whether these suit your clinical needs. What RF coils are included and what come as optional extras? Do you need to buy extra RF channels to use some of these coils? Maintenance: high-specification machines are expensive to buy and run. The liquid helium in the magnet slowly boils off and it will need to be topped up every few years. What type of service contract will you get? A 6-month service is common and frequently included in the price, but check. Are there options to extend it? How much will these cost? Before you buy: as well as reading the brochures and talking to the manufacturers, it is always worth visiting the factory or a site where your selected equipment is already in use before you buy. 4. Customer support Training: when the equipment is installed by the manufacturer, training on how to use it will automatically be provided. Check how much training will be provided and for whom, and whether this suits your needs. For instance, you may need extra training if you are upgrading from a 1.5 T to a 3 T model. Will all the training be on site? Operator and reference manuals: will these be provided in electronic or print format, or both? THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS 141

5 09 MRI SCANNERS: A BUYER S GUIDE Table 2. Technical specifications for 1.5 T equipment. GE Signa GE Signa Philips Philips Siemens Siemens Toshiba Excelart HDx HDe Intera Achieva Symphony Avanto Vantage Main magnetic <0.27 < <1.0 field homogeneity (40 cm DSV ppm) Max field of view mm (isotropic) (isotropic) (isotropic) (isotropic) (isotropic) (isotropic) (isotropic) For Atlas ZGV: 550 for x & y directions and 500 for z direction Min field of view mm Helium refill >3 yrs >3 yrs 3 yrs 3 yrs 2 yrs 10 yrs 2 3 yrs Min installed <33 < area m 2 RF system EXCITE HDx EXCITE HDe Synergy Freewave Advanced/ TM AGV/XGV/ Whole body 32x8/76x18/ ZGV/Atlas Array 76x32 No of 8/16/32 4/8 4/6 8/16/32 Advanced=4 8/18/32 AGV=4 independent Whole body=8 XGV=4,8 RF receiver ZGV=8 channels Atlas=16 Gradient HDx Twin HDe Pulsar Pulsar Nova Nova Quantum Q SQ AGV XGV ZGV system HP HP Dual engine engine HP Max amplitude , mt m -1 X, Y, Z (45 Z) Slew rate , mt m-1 ms-1 X, Y, Z Min TR ms 3D gradient echo DSV: defined spherical volume; RF: radiofrequency; TR: repetition time. Table 3. Technical specifications for 3 T equipment. GE Signa HDx Philips Achieva Siemens Allegra Siemens Trio X series Main magnetic field homogeneity (40 cm DSV ppm) Helium refill 4 yrs NA 7 months 9 months Minimum installed area m <33 RF system HDx Freewave ipat/ipat plus 102x8/102x18/102x32 No of independent RF 8 /16/32 8 /16/32 ipat=4 102x8=81/102x18=18/ receive channels ipat plus=8 102x32=32 Max amplitude mt m -1 X, Y, Z 50 Standard mode = (45 for z direction) Enhanced mode = 80 Slew rate mt m -1 ms-1 X, Y, Z 150 Standard mode = Enhanced mode = 200 Min TR ms 3D gradient echo Max field of view mm 450 (x & y directions) 530 (isotropic) 220 (isotropic) 500 (isotropic) 480 (z directions) Min field of view mm DSV: defined spherical volume; RF: radiofrequency; TR: repetition time. 142 THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS

6 MRI SCANNERS: A BUYER S GUIDE 09 Table 4. Patient comfort: 1.5 T models. Patient aperture at narrowest cm Width x height (couch to pole) GE Signa HDx GE Signa HDe Philips Intera Philips Achieva Siemens Symphony Siemens Avanto Toshiba Excelart Vantage 60x x x42 60x42 60x x x48.3 Total length cm Couch 97/69 97/69 89/52 89/52 100/45 89/ /42 max/min height cm Body mass limit kg DSV: defined spherical volume; RF: radiofrequency; TR: repetition time. Table 5. Patient comfort: 3T models. GE Signa HDx Philips Achieva X series Siemens Allegra Siemens Trio Patient aperture at narrowest cm Width x height (couch to pole) 60x105 60x42 35x35 60x45.5 Total length cm Couch max/min height cm 69/97 52/89 46/80 57/100 Body mass limit kg Service contract: find out if there are enough engineers to fix any potential problems and how soon the manufacturer guarantees to rectify them. How widely available are parts? How quickly can they be shipped in? Does the manufacturer guarantee a certain number of working days in the year (guaranteed uptime )? Remote diagnostics: some manufacturers operate remote diagnostics, whereby the system is automatically linked to the factory. This can speed up detection and resolution of problems. Check whether this is available. All CEP reports since 2002 are available to download from the organisation's website. An alert service is also available, by contacting cep@pasa.nhs.uk REFERENCES 1. NHS Purchasing and Supply Agency. Report T MRI Systems. Issue 4. May NHS Purchasing and Supply Agency. Report T MRI Systems. Issue 7. May Heidemann RM, Griswold MA, Kiefer B, et al. Resolution enhancement in lung 1 H imaging using parallel imaging methods. Magn Reson Med 2003; 49: Kluge A, Gerriets T, Lange U, Bachman G. MRI for short-term follow-up of acute pulmonary embolism. Assessment of thrombus appearance and pulmonary perfusion: a feasibility study. Eur Radiol 2005; 15: Mills GH, Wild JM, Eberle B, Van Beek EJR. Functional magnetic resonance imaging of the lung. Br J Anaesth 2003; 91: Fain SB, Korosec FR, Holmes JH, et al. Functional lung imaging using hyperpolarized gas MRI. J Magn Reson Imaging 2007; 25: MANUFACTURERS WEBSITES Philips Medical Systems Siemens Medical Solutions GE Healthcare Toshiba Medical Systems (Europe) THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS 143

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Excelart Vantage Magnet Configuration: Ultra-short-bore Strength (or W x H): 1.5 T Homogeneity, ppm V-RMS: Dimensions of maximum

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

Basic Principles of Magnetic Resonance

Basic Principles of Magnetic Resonance Basic Principles of Magnetic Resonance Contents: Jorge Jovicich jovicich@mit.edu I) Historical Background II) An MR experiment - Overview - Can we scan the subject? - The subject goes into the magnet -

More information

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications Safety Issue 3.0 T MRI Pro 3.0 T MRI Con Immediate fringe field surrounding magnet A ferromagnetic object inadvertently brought into the scan room will experience a sharp increase in attraction toward

More information

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: kturek@mri-tech.pl

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: kturek@mri-tech.pl Cirrus 0.2T MRI for Everyone North America, Asia, Europe contact: kturek@mri-tech.pl MRI-TECH inc. Cirrus MRI system for all your needs: Low costs Low maintenance High quality Open geometry Imaging of

More information

5 Factors Affecting the Signal-to-Noise Ratio

5 Factors Affecting the Signal-to-Noise Ratio 5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed

More information

Purchasing a cardiac CT scanner: What the radiologist needs to know

Purchasing a cardiac CT scanner: What the radiologist needs to know Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage

More information

University Children s Hospital Basel

University Children s Hospital Basel Cooperation Siemens International Reference Center Pediatric MRI www.siemens.com/skyra University Children s Hospital Basel Siemens International Reference Center Pediatric MRI Answers for life. Partnering

More information

It takes two to connect the dots in MRI.

It takes two to connect the dots in MRI. It takes two to connect the dots in MRI. www.siemens.com/healthcare Tim+ Tim (Total imaging matrix) technology has made fast, flexible and accurate scans the rule, rather than the exception. But we know

More information

Who can have an MRI scan?

Who can have an MRI scan? i MRI Scan Contents What is an MRI scan? 1 Why am I having an MRI scan? 1 Who can have an MRI scan? 1 What preparation will I need? 2 What happens when I come for the scan? 2 Who will I see when I have

More information

Musculoskeletal MRI Technical Considerations

Musculoskeletal MRI Technical Considerations Musculoskeletal MRI Technical Considerations Garry E. Gold, M.D. Professor of Radiology, Bioengineering and Orthopaedic Surgery Stanford University Outline Joint Structure Image Contrast Protocols: 3.0T

More information

MRI SEQUENCES. 1 Gradient Echo Sequence

MRI SEQUENCES. 1 Gradient Echo Sequence 5 An MRI sequence is an ordered combination of RF and gradient pulses designed to acquire the data to form the image. In this chapter I will describe the basic gradient echo, spin echo and inversion recovery

More information

Breast MRI Quality Control

Breast MRI Quality Control Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Discuss the importance of breast MRI quality control (QC). Provide an overview of the new ACR Breast MRI Accreditation

More information

Shorter Scanning Times Through Efficient Software. Annette Tuffs. Article from the customer magazine Medical Solutions, December 2010

Shorter Scanning Times Through Efficient Software. Annette Tuffs. Article from the customer magazine Medical Solutions, December 2010 Shorter Scanning Times Through Efficient Software Annette Tuffs Article from the customer magazine Medical Solutions, December 2010 www.siemens.com/healthcare-magazine Highlight Topic Efficiency: MRI A

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals

More information

MRI SAFETY JEOPARDY (Technologist Edition) Questions

MRI SAFETY JEOPARDY (Technologist Edition) Questions MRI SAFETY JEOPARDY (Technologist Edition) Questions For MRI Safety Week, 2009 Provided, Inc. www.mednovus.com PowerPoint MRI Safety Quiz available for free download from http://mrimetaldetector.com/media/proving_grounds/mri_safety_week.html

More information

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI Douglas C. Noll* and Walter Schneider Departments of *Radiology, *Electrical Engineering, and Psychology University

More information

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi Diffusione e perfusione in risonanza magnetica E. Pagani, M. Filippi DW-MRI DIFFUSION-WEIGHTED MRI Principles Diffusion results from a microspic random motion known as Brownian motion THE RANDOM WALK How

More information

Introduction to Nuclear Magnetic Resonance Spectroscopy

Introduction to Nuclear Magnetic Resonance Spectroscopy Introduction to Nuclear Magnetic Resonance Spectroscopy Dr. Dean L. Olson, NMR Lab Director School of Chemical Sciences University of Illinois Called figures, equations, and tables are from Principles

More information

GE Medical Systems Training in Partnership. Module 12: Spin Echo

GE Medical Systems Training in Partnership. Module 12: Spin Echo Module : Spin Echo Spin Echo Objectives Review the SE PSD. Review the concepts of T, T, and T*. Spin Echo PSD RF Gz Gy 90 80 Gx Spin Echo - SE Spin echo is a standard pulse sequence on Signa MRi/LX and

More information

SITE IMAGING MANUAL ACRIN 6698

SITE IMAGING MANUAL ACRIN 6698 SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012

More information

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System 1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro Corp. 1800 Bridge Parkway

More information

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a BASIC PRINCIPLES INTRODUCTION TO NUCLEAR MAGNETIC RESONANCE (NMR) 1. The nuclei of certain atoms with odd atomic number, and/or odd mass behave as spinning charges. The nucleus is the center of positive

More information

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R GE 3.0T 3.0T WRIST Invivo 8CH Wrist Coil Sequence Ax T2 Cor PD Cor PDFS Cor T1 Cor PD (Small FOV) FOV (mm) 80 80 80 80 40 Matrix 384x224 384x256 320x256 384x320 320x192 Phase Direction RL RL RL RL RL #

More information

MRI Department Goals. Effective Ways to Improve Patient Cooperation and Safety in MRI Exam. Lecture Topics

MRI Department Goals. Effective Ways to Improve Patient Cooperation and Safety in MRI Exam. Lecture Topics Effective Ways to Improve Patient Cooperation and Safety in MRI Exam lowens2@clarian.org Lecture Topics MRI patient preparation Department MRI Safety Policies and Procedures MRI patient safety screening

More information

FDA Guidelines for Magnetic Resonance Equipment Safety

FDA Guidelines for Magnetic Resonance Equipment Safety FDA Guidelines for Magnetic Resonance Equipment Safety Loren A. Zaremba, Ph.D. Center for Devices and Radiological Health Food and Drug Administration Outline I. Introduction II. Static Magnetic Field

More information

MRI QA Technologist s Tests

MRI QA Technologist s Tests AAPM 55th Annual Meeting MRI QA Technologist s Tests Guang Jia, PhD Associate Professor Department of Physics Louisiana State University Background Outline QC importance Technologist s role QC test tips

More information

Nuclear Magnetic Resonance

Nuclear Magnetic Resonance Nuclear Magnetic Resonance NMR is probably the most useful and powerful technique for identifying and characterizing organic compounds. Felix Bloch and Edward Mills Purcell were awarded the 1952 Nobel

More information

Nuclear Magnetic Resonance Spectroscopy

Nuclear Magnetic Resonance Spectroscopy Nuclear Magnetic Resonance Spectroscopy Introduction NMR is the most powerful tool available for organic structure determination. It is used to study a wide variety of nuclei: 1 H 13 C 15 N 19 F 31 P 2

More information

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager Technique and Safety of performing EEG/fMRI measurements by Pierluigi Castellone, Electronics Engineer Brain Products General Manager Contents of the presentation Why recording simultaneous EEG and fmri?

More information

2012 Operating Room Design Manual

2012 Operating Room Design Manual CHAPTER 16 REMOTE AND HAZARDOUS LOCATIONS Lead Author: John Rogoski, DO, Assistant Professor of Clinical Anesthesiology, Ohio State University College of Medicine Checklist: Is the room size large enough

More information

32-Channel Head Coil Imaging at 3T

32-Channel Head Coil Imaging at 3T 32-Channel Head Coil Imaging at 3T Thomas Benner Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA

More information

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT GUIDE TO SETTING UP AN MRI RESEARCH PROJECT Formal requirements and procedures OVERVIEW This document is intended to help a principle investigator set up a research project using magnetic resonance imaging

More information

Preserving Hyperpolarization of 129 Xe for Transport Submitted by: Zach Zytner Project Supervisor: Dr. Giles Santyr Medical Biophysics 3970 Date:

Preserving Hyperpolarization of 129 Xe for Transport Submitted by: Zach Zytner Project Supervisor: Dr. Giles Santyr Medical Biophysics 3970 Date: Preserving Hyperpolarization of 129 Xe for Transport Submitted by: Zach Zytner Project Supervisor: Dr. Giles Santyr Medical Biophysics 3970 Date: April 1, 2011 Introduction and Theory The effective use

More information

13C NMR Spectroscopy

13C NMR Spectroscopy 13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number

More information

Adding Heart to Your Technology

Adding Heart to Your Technology RMCM-01 Heart Rate Receiver Component Product code #: 39025074 KEY FEATURES High Filtering Unit Designed to work well on constant noise fields SMD component: To be installed as a standard component to

More information

Functional neuroimaging. Imaging brain function in real time (not just the structure of the brain).

Functional neuroimaging. Imaging brain function in real time (not just the structure of the brain). Functional neuroimaging Imaging brain function in real time (not just the structure of the brain). The brain is bloody & electric Blood increase in neuronal activity increase in metabolic demand for glucose

More information

High Field MRI: Technology, Applications, Safety, and Limitations. Brief overview of high-field MRI. The promise of high-field MRI

High Field MRI: Technology, Applications, Safety, and Limitations. Brief overview of high-field MRI. The promise of high-field MRI High Field MRI: Technology, Applications, Safety, and Limitations R. Jason Stafford, Ph.D. Department of Imaging Physics The University of Texas M. D. Anderson Cancer Center Houston, TX Brief overview

More information

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading:

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading: Applied Spectroscopy Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading: Banwell and McCash Chapter 7 Skoog, Holler Nieman Chapter 19 Atkins, Chapter 18 Relaxation processes We need

More information

Ventilation Perfusion Relationships

Ventilation Perfusion Relationships Ventilation Perfusion Relationships VENTILATION PERFUSION RATIO Ideally, each alveolus in the lungs would receive the same amount of ventilation and pulmonary capillary blood flow (perfusion). In reality,

More information

MRI SAFETY JEOPARDY (NON Technologist Edition) Questions

MRI SAFETY JEOPARDY (NON Technologist Edition) Questions MRI SAFETY JEOPARDY (NON Technologist Edition) Questions For MRI Safety Week, 2009 Provided, Inc. www.mednovus.com PowerPoint MRI Safety Quiz available for free download from http://mrimetaldetector.com/media/proving_grounds/mri_safety_week.html

More information

REVEAL LINQ LNQ11. Insertable Cardiac Monitor MRI procedural information. MRI Technical Manual

REVEAL LINQ LNQ11. Insertable Cardiac Monitor MRI procedural information. MRI Technical Manual REVEAL LINQ LNQ11 Insertable Cardiac Monitor MRI procedural information MRI Technical Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician. The following list

More information

Magnetic Resonance Imaging (MRI) Scans

Magnetic Resonance Imaging (MRI) Scans Magnetic Resonance Imaging (MRI) Scans Patient Information Radiology Department Author ID: ST Leaflet Number: Rad 004 Version: 3 Name of Leaflet: MRI Scan Date Produced: March 2015 Review Date: March 2017

More information

Projects at the Danish Research Centre for Magnetic Resonance

Projects at the Danish Research Centre for Magnetic Resonance Projects at the Danish Research Centre for Magnetic Resonance Five projects involving Magnetic Resonance Imaging (MRI) Hvidovre Hospital Example: [1-13C]Pyruvate signal enhanced by ~50,000x! Magnetic Resonance

More information

The smart approach to digital MRI

The smart approach to digital MRI Philips Healthcare is part of Royal Philips Electronics How to reach us: www.philips.com/healthcare healthcare@philips.com Product information: www.philips.com/smartpathtodstream The smart approach to

More information

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter

More information

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research. ParaVision 6 The Next Generation of MR Acquisition and Processing for Preclinical and Material Research Innovation with Integrity Preclinical MRI A new standard in Preclinical Imaging ParaVision sets a

More information

MRI for Paediatric Surgeons

MRI for Paediatric Surgeons MRI for Paediatric Surgeons Starship David Perry Paediatric Radiologist Starship Children s Hospital CHILDREN S HEALTH What determines the brightness of a pixel in MRI? i.e. What determines the strength

More information

510(k) Summary. This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.

510(k) Summary. This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807. MAR, 2009 / \MLECH MRI-TECH Sp. z o.o., ul. Zielihska 3, 31-227 Krak6w, Poland MRI-TECH Canada, Inc. 206-3820 Cambie Street Vancouver BC V5Z 2X7 510(k) Summary This summary of 510(k) safety and effectiveness

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging What are the uses of MRI? To begin, not only are there a variety of scanning methodologies available, but there are also a variety of MRI methodologies available which provide

More information

BIOMEDICAL ULTRASOUND

BIOMEDICAL ULTRASOUND BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic

More information

MRC High Resolution. MR-compatible digital HD video camera. User manual

MRC High Resolution. MR-compatible digital HD video camera. User manual MRC High Resolution MR-compatible digital HD video camera User manual page 1 of 12 Contents 1. Intended use...2 2. System components...3 3. Video camera and lens...4 4. Interface...4 5. Installation...5

More information

Investigation of Magnetic Resonance Imaging Effects when Using Bone Conduction Implants

Investigation of Magnetic Resonance Imaging Effects when Using Bone Conduction Implants Investigation of Magnetic Resonance Imaging Effects when Using Bone Conduction Implants Master of Science Thesis in Biomedical Engineering, MPBME Karl-Johan Fredén Jansson Department of Signals and Systems

More information

RE: Invitation for Bid (IFB) No. Q14-32 Type of Service: Magnetic Resonance Imaging System, Maintenance Agreement and Mobile MRI Unit. Addendum No.

RE: Invitation for Bid (IFB) No. Q14-32 Type of Service: Magnetic Resonance Imaging System, Maintenance Agreement and Mobile MRI Unit. Addendum No. VIA EMAIL DATE: June 16, 2014 TO: FROM: Prospective Proposers Maureen Crystal Director of Contracts RE: Invitation for Bid (IFB) No. Q14-32 Type of Service: Magnetic Resonance Imaging System, Maintenance

More information

8.1 Radio Emission from Solar System objects

8.1 Radio Emission from Solar System objects 8.1 Radio Emission from Solar System objects 8.1.1 Moon and Terrestrial planets At visible wavelengths all the emission seen from these objects is due to light reflected from the sun. However at radio

More information

C1 Medical Imaging Modalities & Characteristics. 4005-759 Linwei Wang

C1 Medical Imaging Modalities & Characteristics. 4005-759 Linwei Wang C1 Medical Imaging Modalities & Characteristics 4005-759 Linwei Wang Major Types of Medical Imaging Modalities X-ray Imaging Computed Tomography (CT) Magnetic Resonance Imaging (MRI) Nuclear Imaging Positron

More information

MRI Physics for Radiologists

MRI Physics for Radiologists Alfred L. Horowitz MRI Physics for Radiologists A Visual Approach Second Edition With 94 Illustrations Springer-Verlag New York Berlin Heidelberg London Paris Tokyo Hong Kong Barcelona Budapest Alfred

More information

Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis

Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis Main points of the chapter 1. Hydrogen Nuclear Magnetic Resonance a. Splitting or coupling (what s next to what) b. Chemical shifts (what type is it) c. Integration

More information

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine G16.4426/EL5823/BE6203 Medical Imaging MRI Instrumentation, Data Acquisition, Image Reconstruction Riccardo Lattanzi, Ph.D. Assistant Professor Department of Radiology, NYU School of Medicine Department

More information

PET/CT-MRI First clinical experience

PET/CT-MRI First clinical experience 20 th April 2013, Barcelona, Sp PET/CT-MRI First clinical experience Philippe Appenzeller, MD Staff Radiologist and Nuclear Medicine Physician Department Medical Imaging, University Hospital Zurich PET/CT-MR

More information

NMR and MRI. Seppo Vahasalo Philips Medical Systems MR Finland 2008-06-17

NMR and MRI. Seppo Vahasalo Philips Medical Systems MR Finland 2008-06-17 NMR and MRI Seppo Vahasalo Philips Medical Systems MR Finland 2008-06-17 Contents Some local MRI history NMR and MRI Physics Magnet technology and electronics Recent trends in MRI and magnet technology

More information

THE CAPITAL COST AND PRODUCTIVITY OF MRI IN A BELGIAN SETTING*

THE CAPITAL COST AND PRODUCTIVITY OF MRI IN A BELGIAN SETTING* JBR BTR, 2010, 93: 92-96. SPECIAL ARTICLE THE CAPITAL COST AND PRODUCTIVITY OF MRI IN A BELGIAN SETTING* C. Obyn 1, I. Cleemput 2 Against the Belgian background of national planning of MRI units and a

More information

Background A nucleus with an odd atomic number or an odd mass number has a nuclear spin that can be observed by NMR spectrometers.

Background A nucleus with an odd atomic number or an odd mass number has a nuclear spin that can be observed by NMR spectrometers. NMR Spectroscopy I Reading: Wade chapter, sections -- -7 Study Problems: -, -7 Key oncepts and Skills: Given an structure, determine which protons are equivalent and which are nonequivalent, predict the

More information

RF Coils... They ve Come a Long, Long Way

RF Coils... They ve Come a Long, Long Way GE Healthcare MR Field Notes The sensitivity area relates to the diameter of the coil. By Increasing coil size, the area of sensitivity also increases. But as with most MR princicloser a coil s position

More information

Nuclear Magnetic Resonance Spectroscopy

Nuclear Magnetic Resonance Spectroscopy Nuclear Magnetic Resonance Spectroscopy Nuclear magnetic resonance spectroscopy is a powerful analytical technique used to characterize organic molecules by identifying carbonhydrogen frameworks within

More information

INTEGRATED CIRCUITS DATA SHEET. TDA7000 FM radio circuit. Product specification File under Integrated Circuits, IC01

INTEGRATED CIRCUITS DATA SHEET. TDA7000 FM radio circuit. Product specification File under Integrated Circuits, IC01 INTEGRATED CIRCUITS DATA SHEET File under Integrated Circuits, IC01 May 1992 GENERAL DESCRIPTION The is a monolithic integrated circuit for mono FM portable radios, where a minimum on peripheral components

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Material (ESI) for Physical Chemistry Chemical Physics. This journal is the Owner Societies 2016 Electronic Supplementary Information Achieving High Resolution and Controlling

More information

Chemistry Department

Chemistry Department Chemistry Department NMR/Instrumentation Facility Users Guide - Rules, safety and system information Prepared by Leila Maurmann The NMR/Instrumentation facility at the Chemistry Department at Kansas State

More information

I i nstrumentat on for ESR M icroscopy Microscopy CURT R DUNNAM 1

I i nstrumentat on for ESR M icroscopy Microscopy CURT R DUNNAM 1 Instrumentation for ESR Microscopy CURT R DUNNAM 1 ACERT ESRM Instrumentation Development elopment to Date: CW (9 GHz) Modified Varian E-series X-band Spectrometer (add-on system) stem) Capable of Projection

More information

RADIOLOGY SERVICES. By Dr Lim Eng Kok 1

RADIOLOGY SERVICES. By Dr Lim Eng Kok 1 INTRODUCTION RADIOLOGY SERVICES By Dr Lim Eng Kok 1 Radiology is the branch of medicine that deals with the use of ionising (e.g. x- rays and radio-isotopes) and non-ionising radiation (e.g. ultrasound

More information

mri : Physics For anyone who does not have a degree in physics Evert J Blink Application Specialist MRI

mri : Physics For anyone who does not have a degree in physics Evert J Blink Application Specialist MRI B A S I C mri : Physics For anyone who does not have a degree in physics Evert J Blink Application Specialist MRI 0 Preface Over the years Magnetic Resonance Imaging, hereafter referred to as MRI, has

More information

The MRI Study Guide for Technologists

The MRI Study Guide for Technologists The MRI Study Guide for Technologists Kenneth S. Meacham The MRI Study Guide for Technologists With 51 Illustrations Springer-Verlag New York Berlin Heidelberg London Paris Tokyo Hong Kong Barcelona Budapest

More information

NMR SPECTROSCOPY A N I N T R O D U C T I O N T O... Self-study booklet NUCLEAR MAGNETIC RESONANCE. 4 3 2 1 0 δ PUBLISHING

NMR SPECTROSCOPY A N I N T R O D U C T I O N T O... Self-study booklet NUCLEAR MAGNETIC RESONANCE. 4 3 2 1 0 δ PUBLISHING A N I N T R O D U T I O N T O... NMR SPETROSOPY NULEAR MAGNETI RESONANE 4 3 1 0 δ Self-study booklet PUBLISING NMR Spectroscopy NULEAR MAGNETI RESONANE SPETROSOPY Origin of Spectra Theory All nuclei possess

More information

Report. Magnetic Resonance Equipment (MRI) - Study on the potential for environmental improvement by the aspect of energy efficiency

Report. Magnetic Resonance Equipment (MRI) - Study on the potential for environmental improvement by the aspect of energy efficiency Report Magnetic Resonance Equipment (MRI) - Study on the potential for environmental improvement by the aspect of energy efficiency Title of the Study: Magnetic resonance Equipment (MRI) - Study on the

More information

Ultrasonic Gas Leak Detection

Ultrasonic Gas Leak Detection Ultrasonic Gas Leak Detection What is it and How Does it Work? Because every life has a purpose... Ultrasonic Gas Leak Detection Introduction Ultrasonic gas leak detection (UGLD) is a comparatively recent

More information

Lecture #7 (2D NMR) Utility of Resonance Assignments

Lecture #7 (2D NMR) Utility of Resonance Assignments Lecture #7 (2D NMR) Basics of multidimensional NMR (2D NMR) 2D NOESY, COSY and TOCSY 2/23/15 Utility of Resonance Assignments Resonance Assignments: Assignment of frequency positions of resonances (peaks)

More information

Power Injectors for Diagnostic Imaging

Power Injectors for Diagnostic Imaging Power Injectors for Diagnostic Imaging Contrast under Control A resource for all your contrast delivery needs Contrast under Control 2 1 3 Mallinckrodt provides contrast delivery systems for Computed

More information

3/30/2013. Disclosure. Advanced Neuro MRI: Imaging Techniques and Protocol Optimization. MRI, 35 year ago. MRI Today. Outlines

3/30/2013. Disclosure. Advanced Neuro MRI: Imaging Techniques and Protocol Optimization. MRI, 35 year ago. MRI Today. Outlines http://www.magnet.fsu.edu Disclosure Advanced Neuro MRI: Imaging Techniques and Protocol Optimization Research funding provided by Siemens Healthcare. Chen Lin, PhD DABR Indiana University School of Medicine

More information

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der

More information

Advances in scmos Camera Technology Benefit Bio Research

Advances in scmos Camera Technology Benefit Bio Research Advances in scmos Camera Technology Benefit Bio Research scmos camera technology is gaining in popularity - Why? In recent years, cell biology has emphasized live cell dynamics, mechanisms and electrochemical

More information

Standardized MRI Protocol for Brain Tumor Clinical Trials. Benjamin M. Ellingson, Ph.D. Assistant Professor of Radiology at UCLA

Standardized MRI Protocol for Brain Tumor Clinical Trials. Benjamin M. Ellingson, Ph.D. Assistant Professor of Radiology at UCLA Standardized MRI Protocol for Brain Tumor Clinical Trials Benjamin M. Ellingson, Ph.D. Assistant Professor of Radiology at UCLA Standardized MRI Protocol for Therapeutic Studies FDA Meeting in January

More information

Case Report: Whole-body Oncologic Imaging with syngo TimCT

Case Report: Whole-body Oncologic Imaging with syngo TimCT Case Report: Whole-body Oncologic Imaging with syngo TimCT Eric Hatfield, M.D. 1 ; Agus Priatna, Ph.D. 2 ; John Kotyk, Ph.D. 1 ; Benjamin Tan, M.D. 1 ; Alto Stemmer 3 ; Stephan Kannengiesser, Ph.D. 3 ;

More information

University College Hospital. Having a CT scan. Imaging Department

University College Hospital. Having a CT scan. Imaging Department University College Hospital Having a CT scan Imaging Department If you need a large print, audio, or a translated copy of this document, please contact us on: 020 345 67002 We will do our best to meet

More information

Subject: Magnetic Resonance Imaging (MRI) Information on Sorin Group Heart Valve Prostheses and Annuloplasty Devices

Subject: Magnetic Resonance Imaging (MRI) Information on Sorin Group Heart Valve Prostheses and Annuloplasty Devices Saluggia, 18 December 2013 Subject: Magnetic Resonance Imaging (MRI) Information on Sorin Group Heart Prostheses and Annuloplasty Devices To whom it may concern This letter summarizes the required MRI

More information

Replacement Heart Valve Product Description (Stented Tissue) Models Reference

Replacement Heart Valve Product Description (Stented Tissue) Models Reference Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences LLC

More information

Kap 8 Image quality, signal, contrast and noise

Kap 8 Image quality, signal, contrast and noise 4/5/ FYS-KJM 474 contrast SNR MR-teori og medisinsk diagnostikk Kap 8 Image qualit, signal, contrast and noise resolution vailable MRparameters speed Main source of noise in MRI: Noise generated within

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Data Sheet FUJITSU PRIMECENTER M2 19-inch Racks

Data Sheet FUJITSU PRIMECENTER M2 19-inch Racks Data Sheet FUJITSU PRIMECENTER M2 19-inch Racks Data Sheet FUJITSU PRIMECENTER M2 19-inch Racks 19-inch Industry Standard Rack Systems The trend to consolidation as a result of virtualization and optimization

More information

Tetramethylsilane (TMS) Trimethylsilyl d 4. -propionic acid (TMSP) Dioxane. O - Na + Dimethylfura n. Potassium Hydrogen Phthalate. Sodium Maleate CH 3

Tetramethylsilane (TMS) Trimethylsilyl d 4. -propionic acid (TMSP) Dioxane. O - Na + Dimethylfura n. Potassium Hydrogen Phthalate. Sodium Maleate CH 3 Practical Aspects of Quantitative NMR Experiments This discussion presumes that you already have an understanding of the basic theory of NMR. There are a number of issues that should be considered when

More information

Used to determine relative location of atoms within a molecule Most helpful spectroscopic technique in organic chemistry Related to MRI in medicine

Used to determine relative location of atoms within a molecule Most helpful spectroscopic technique in organic chemistry Related to MRI in medicine Structure Determination: Nuclear Magnetic Resonance CHEM 241 UNIT 5C 1 The Use of NMR Spectroscopy Used to determine relative location of atoms within a molecule Most helpful spectroscopic technique in

More information

Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.

Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source

More information

Current Probes. User Manual

Current Probes. User Manual Current Probes User Manual ETS-Lindgren L.P. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall

More information

Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint

Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint Tomohiro Sahara, Hiroshi Tsutsui, Shigehiro Hashimoto, Shuichi Mochizuki, Kenichi Yamasaki, Hideo Kondo, Teruo Miyazaki Department

More information

SZI 1029. IR Audio Transmission Technology Modulators/Radiators

SZI 1029. IR Audio Transmission Technology Modulators/Radiators General Description The is a 5-W high-power radiator for use with all wideband and narrowband modulators. The radiator is automatically switched on by the RF carrier and can cover areas of up to 800 m

More information

Siemens Remote Service

Siemens Remote Service Siemens Remote Service Capitalize on a new dimension in system support Introduction / Contents Siemens Remote Services (SRS) Mastery of growth, dynamics and big diversity of systems in a complex environment

More information

YOUR MRI EXAM: QUESTIONS AND ANSWERS

YOUR MRI EXAM: QUESTIONS AND ANSWERS YOUR MRI EXAM: QUESTIONS AND ANSWERS What is MRI? MRI is short for Magnetic Resonance Imaging. MRI is an advanced technology that lets your doctor see internal organs, blood vessels, muscles, joints, tumors,

More information

Handling Gas: Cylinders vs Gas Generators

Handling Gas: Cylinders vs Gas Generators Handling Gas: Cylinders vs Gas Generators Lab Manager Tech Trends Webinar April 17, 2014 Presented by John Speranza of Proton On Site What you will learn more about today The various advantages and disadvantages

More information

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers Data Sheet Specifications Specifications are only valid for the stated operating frequency, and apply over 0 C to +55 C unless otherwise

More information

Proton Nuclear Magnetic Resonance Spectroscopy

Proton Nuclear Magnetic Resonance Spectroscopy CHEM 334L Organic Chemistry Laboratory Revision 2.0 Proton Nuclear Magnetic Resonance Spectroscopy In this laboratory exercise we will learn how to use the Chemistry Department's Nuclear Magnetic Resonance

More information