Tradeoffs in CT Image Quality and Dose

Similar documents
How To Improve Your Ct Image Quality

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

Assessing Radiation Dose: How to Do It Right

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

American College of Radiology CT Accreditation Program. Testing Instructions

Purchasing a cardiac CT scanner: What the radiologist needs to know

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Fundamentals of Cone-Beam CT Imaging

Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.

Physics testing of image detectors

Patient-centered CT imaging: New methods for patient-specific optimization 1 of image quality and radiation dose

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver /9/07) by M. Jensen and J. E.

WHERE IN THE WORLD JILL LIPOTI?

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Digital Mammography Update: Design and Characteristics of Current Systems

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Table of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

The disclaimer on page 1 is an integral part of this document. 1. Copyright December 14, 2015 by AAPM. All rights reserved.

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Quality Control and Maintenance Programs

The AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement.

Experimental study of beam hardening artefacts in photon counting breast computed tomography

Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives

3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?

Multi-slice Helical CT Scanning of the Chest

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

The Challenge of CT Dose Records

Concepts for High-Resolution Low-Dose CT of the Breast

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

Performance testing for Precision 500D Classical R/F System

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

Diagnostic Exposure Tracking in the Medical Record

Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad.

Quality Control of Full Field Digital Mammography Units

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

Checklist for the Qualification of Digital Detector Array Systems

Techniques and Applications of Automatic Tube Current Modulation for CT 1

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

Prepublication Requirements

MRI QA Technologist s Tests

Practical exercise: Effective dose estimate in CT

DICOM Correction Item

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

The effect of pitch in multislice spiral/helical CT

Digital Breast Tomosynthesis QC Requirements

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

Bergen Community College Division of Health Professions/ Radiography Program. Course Syllabus and Course of Study for Lecture

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Copyright March 1, 2016 by AAPM. All rights reserved.

Interim policy effective January 1, 2016, through December 31, 2017

INTRODUCTION. A. Purpose

Computed Tomography Radiation Safety Issues in Ontario

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

First floor, Main Hospital North Services provided 24/7 365 days per year

Recommendations for a technical quality control program for diagnostic X-ray equipment

Data. microcat +SPECT

Development of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

MEDICAL IMAGING 2nd Part Computed Tomography

State-of-the-Art Technology in Cardiac CT

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

Automated EMR Dose History Extraction and Monitoring

Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction

Use of 3D Printers in Proton Therapy

CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

Lecture 14. Point Spread Function (PSF)

Breast MRI Quality Control

Patient Prep Information

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays

DENTAL Cone beam 3D X-RAY SYSTEM with

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Assessment of calcium scoring performance in cardiac computed tomography

An iterative approach to the beam hardening correction in cone beam CT

Transcription:

Tradeoffs in CT Image Quality and Dose Michael F. McNitt-Gray, PhD, DABR Associate Professor Depart of Radiology David Geffen School of Medicine at UCLA In CT scanning, image quality has many components and is influenced by many technical parameters. While image quality has always been a concern for the physics community, clinically acceptable image quality has become even more of an issue as strategies to reduce radiation dose especially to pediatric patients become a larger focus. The purpose of this presentation will be to first describe several of the components of CT image quality noise, slice thickness (Z-axis resolution), low contrast resolution, and high contrast resolution as well as radiation dose and to describe how each of these may be affected by technical parameter selection. This presentation will pay particular attention to the tradeoffs that exist between different aspects of image quality, especially when the reduction of radiation dose is one of the objectives. The presentation will then explore several mechanisms that can be used to reduce radiation dose in CT exams and the implications for the diagnostic image quality of the exam. Specifically, the implications of varying the tube current*time product (mas), pitch or tablespeed (or for axial imaging, the table increment), slice thickness, beam energy (kvp), patient (or phantom) size and dose reduction options (such as varying tube current during exposure) will be described for both radiation dose and diagnostic image quality. These strategies will be described for both single and various multidetector scanners as, unfortunately, the tradeoffs are not uniform across scanners. In addition to these direct effects, this presentation will describe possible indirect effects that both collimation and reconstruction algorithm may have on radiation dose and image quality when compensatory actions are taken to overcome increased noise levels. Finally, this presentation will emphasize that the tradeoffs between radiation dose and image quality are clinical-task dependent; this will be illustrated with examples from selected diagnostic imaging exams.

CT Image Quality- A Summary Image Noise Image noise, in its most simple definition, is measured as the standard deviation of voxel values in a homogenous (typically water) phantom. There are, of course, much more complete definitions of noise that take into account the contrast scale of the scanner (for example, the definition described in AAPM report #39[1]). In this context, we will consider this simple definition and observe that image noise is influenced by a large number of parameters, including: kvp ma Exposure time Collimation/Reconstructed Slice Thickness Reconstruction algorithm or filter Helical Pitch/Table speed Helical Interpolation Algorithm Others (Focal spot to isocenter distance, detector efficiency, etc.) For example, reducing the mas is expected to increase the noise (measured standard deviation) by 1. Therefore, if the mas is reduced by ½, then noise should increase by 2 =1.414 à(40% mas increase). This is illustrated in the water phantom images below where the image on the left was scanned with 100 ma and 0.8 sec scan (80 mas), while the image of the right was scanned with 50 ma and 0.8 sec scan (40 mas). The measured standard deviation is nearly 40% higher for the lower mas scan (17.91 vs. 13.03). (a) (b) Figure 1: (a) water phantom scanned at 100 ma, 0.8 sec scan time (80 mas), and same phantom scanned at 50 ma, 0.8 sec scan time (40 mas) and all other parameters identical.

Slice Thickness (Z-axis resolution) The reconstructed slice thickness of an image has become much more interesting and complex when going from axial scanning to helical scanning to multidetector helical scanning. For this discussion, we will focus only on the reconstructed slice width in helical scanning and the factors that may influence it, which may include: X-ray Beam Collimation (especially in single slice CT scanners) Detector Width (especially in multidetector CT scanners) Helical Pitch/Table speed Helical Interpolation Algorithm Note: For some manufacturers multidetector scanners, the reconstructed slice thickness is independent of table speed; this is because of the interpolation algorithm used. Hence, these last two items are tightly linked. For single detector helical scanners using either the 180 LI or 360 LI interpolation algorithm, the results were quite clear that higher table speeds produced larger effective slice thicknesses [2, 3]. For multidetector helical scanners, these trends are not quite so clear because of the ability to interpolate data collected from multiple detectors and because of the different interpolation algorithms available [4]. Figure 2 illustrates that in one multidetector scanner, the table speed can affect the slice sensitivity profile and effective slice thickness. The top curve was obtained using 4 x 5mm collimation mode, pitch of 0.75 (table speed of 15 mm/rot) and a nominal reconstructed slice thickness of 5mm; the calculated FWHM was 5.31 mm. The bottom curve was obtained using the same parameters except that the pitch was increased to 1.5 (30 mm/rot) and the resulting calculated FWHM was 6.24 mm; a 17% increase. Slice Sensitivity Profiles Intensity in HU 200 180 160 140 120 100 80 60 40 20 0 0 5 10 15 20 distance in mm 5 mm HQ 15mm/rot 5 mm HS 30 mm/rot Figure 2. Differences in Slice Sensitivity Profile due to differences in table speed with all other factors held constant in a Multidetector helical CT scanner (GE LightSpeed Qx/I, GE Medical Systems, Milwaukee, WI).

High Contrast (Spatial) Resolution The high contrast or spatial resolution of the system is often determined using objects having a large signal to noise ratio. This test measures the system s ability to resolve high contrast objects of increasingly smaller sizes (increasing spatial frequencies). Several quantitative methods (scanning a wire to calculate the modulation transfer function - MTF - or scanning a bar pattern phantom to calculate MTF using the Droege-Morin approach [5]) have been described. High contrast spatial resolution is influenced by factors including: System geometric resolution limits focal spot size, detector width and ray sampling, Pixel size Properties of the convolution kernel/mathematical reconstruction filter Typically, those reconstruction filters that enhance or preserve the higher spatial frequencies (smaller objects) do so at the cost of increased noise in the image. For high signal to noise objects, this tradeoff is usually acceptable. These concepts are illustrated below. Figure 3 shows the MTF of two different reconstruction filters (using the Droege-Morin method) measured on a multidetector scanner (GE LightSpeed Qx/I). This figure shows that the Bone algorithm maintains a higher amplitude modulation than the Standard algorithm. Figure 4 illustrates the impact of these two reconstruction algorithms on noise for two water phantom images. The measured standard deviation from Bone algorithm is 45.36, and 11.88 from Standard algorithm. MTF GE LightSpeed 1 Amplitude 0.75 0.5 0.25 Standard Bone Figure 3. Modulation Transfer Function (MTF) of two reconstruction algorithms from the same multidetector CT scanner (GE LightSpeed Qx/I). 0 0 2 4 6 8 10 12 spatial frequency lp/cm (a) (b) Figure 4: Water phantom reconstructed with (a) Standard algorithm and (b) Bone algorithm

Low Contrast Resolution The low contrast resolution of the system is often determined using objects having a very small difference from background (typically from 4-10 HU difference). In this case, because the signal (the difference between object and background) is so small, noise is a significant factor in this test. This test measures the system s ability to resolve low contrast objects of increasingly smaller sizes (increasing spatial frequencies). While several purely quantitative methods have been proposed, the most widely accepted methods are still those that require an observer to subjectively detect objects as distinct. An example of a low contrast resolution phantom is that in use by the American College of Radiology (ACR) CT Accreditation program and is shown in Figure 5. This phantom consists of 1 25mm rod for reference and measurements, and then sets of 4 rods where each set is decreasing in diameter, starting with 6mm, 5mm, 4mm on down to 2mm (typically not visible unless a very, very high technique is used). Low contrast resolution is influenced by a large number of factors that include a similar list to that described above for image noise. The actual test results can also be influenced by factors such as window and level settings, the monitor or film calibration curve and the observer themselves. Figure 5 illustrates just one aspect of the low contrast resolution test and how noise as created by different mas levels can influence the results of this test (Note that the reader s printer or computer monitor may also significantly affect the appearance of these images). Under good viewing conditions, the 4 mm diameter rods can be seen on the image on the left; while even the 6 mm diameter rods are questionable on the right hand image. (a) (b) Figure 5. ACR CT accreditation phantom low contrast module imaged under two different noise conditions on the same scanner with: (a) 240 mas and (b) 80 mas.

Radiation Dose The radiation dose measured in phantoms by a particular imaging protocol as described by CTDI vol and described in [6] - is influenced my many technical factors, including: kvp ma Exposure time X-ray Beam collimation Pitch (Table Speed) Dose Reduction Options (such as tube current modulation techniques) Scanner make and model specifics (which include beam spectra, bowtie filter and geometry such as distance from focal spot to isocenter) Indirect Effects of Algorithm and Collimation Several of these parameters have been described as possible mechanisms to reduced radiation dose to patients and to pediatric and small adult patients in particular [7]. These included recommendations to: Reduce tube current (ma) Increase table increment (axial) or pitch Develop ma settings based on patient weight (or diameter) and body region Reduce number of multiple scans w/contrast Eliminate inappropriate referrals for CT While the last two items have to do with referral patterns and appropriate usage of CT, the first three have implications for image quality.

Tradeoffs between Radiation Dose and Image Quality As parameters are varied to reduce radiation dose, the impacts on image quality must be considered as there are often direct tradeoffs between the two. Where the level of acceptable tradeoff in image quality lies may be dependent on the clinical task at hand. Some examples of tradeoffs include: Reducing mas: o Reduces radiation dose in proportion to reduction in mas, o But increases image noise in proportion to ( mas ) /( mas ). Therefore, if the mas is reduced to ½ of the original, then the noise is expected to increase by 1.41 (41% increase). As illustrated above, this should degrade the low contrast resolution performance Increase table speed or pitch: o Reduces radiation dose in proportion to the increase in pitch (UNLESS scanner automatically increases ma with increase in pitch as Siemens and Philips multidetector CT scanners do). o But, may increase slice sensitivity profile, creating larger effective slice thickness and reducing z-axis resolution. Reducing kvp: o Reduces radiation dose (though not linearly with kvp), o May increase signal contrast for some tissues and iodine (High Z) due to increased photoelectric o May significantly increase beam hardening artifact if beam energy gets too low (e.g. 80 kvp) Indirect Effects: o When thinner slices or high spatial frequency filters are used, with all other factors held constant, noise increases. o To compensate for increase in noise, user may be inclined to increase the mas, which in turn will increase radiation dose. original reduced Influence of the Clinical Task on Determining Acceptable Image Quality Different clinical tasks have different requirements for image quality. Some examples are: High Signal to Noise Tasks o Solid nodule detection in the lung (Exception is the ground glass nodule) o Coronary artery calcium detection. o Identification of Emphysema in lung Low Signal to Noise Tasks o Abdominal Scans (liver or kidney lesion detection) o Diffuse Lung Disease Medium Signal to Noise Tasks o Brain o Peds Abdomen/Chest (Rule out Bleeds) Lung Disease

Summary: There are many tradeoffs in image quality and many times the different image quality characteristics are inter-related. Noise versus spatial resolution radiation dose is one example. Methods to reduce radiation dose may involve tradeoffs in image quality (increase in noise, decrease in low contrast resolution, etc.). Different clinical imaging tasks have different requirements in terms of acceptable image quality. The more clearly defined the objectives of a clinically indicated study, then the more clearly the image quality requirements can be determined. In an effort to reduce radiation dose, the effect on physical image quality and the ability to carry out the clinical imaging task must be considered. References [1] AAPM Report # 39 Specification and Acceptance Testing of Computed Tomography Scanners [2]. Polacin A; Kalender WA; Marchal G. Evaluation of section sensitivity profiles and image noise in spiral CT. Radiology, 1992 Oct, 185(1):29-35. [3]. Polacin A; Kalender WA; Brink J; Vannier MA. Measurement of slice sensitivity profiles in spiral CT. Medical Physics, 1994 Jan, 21(1):133-40. [4] McCollough CH, Zink FE. Performance evaluation of a multi-slice CT system. Med Phys 1999; 26:2223 2230. [5] Droege RT, Morin RL A practical method to measure the MTF of CT scanners. Med Phys 1982 Sep-Oct; 9(5):758-760. [6] McNitt-Gray, Radiation dose in CT, Radiographics 22:1541-1553 (2002) [7] FDA Notice dated 11-2-01 http://www.fda.gov/cdrh/safety/110201-ct.html