1 Routine Ultrasound Equipment Tests for Quality Assurance James A. Zagzebski, Ph.D. Nick Rupert, M.S. Ryan DeWall, M.S. Tomy Varghese, Ph.D. Dept. of Medical Physics University of Wisconsin, Madison, WI, USA Contents Elements of a Routine Test Program What system components are considered most important in routine testing? Concentrate on transducers! Steps to carry out tests on a regular basis Activity of Ultrasound Lab Accreditation Groups Basic QA methods Phantoms and tests 1
2 Routine QA Program Guidelines available from US facility accreditation bodies (ACR; AIUM) Emphasis is given to safety, cleanliness, and very basic performance tests. Issues related to image performance, such as spatial resolution, are not included (except ACR breast). Traditional parameters such as distance measurements are believed by many to be unnecessary. Includedi n some protocols, but only briefly. Routine QA (ACR General US Program) System sensitivity and/or penetration capability 1999: Do for 2 probes (?) Image Uniformity Look for element dropout Photography and other hard copy recording 2010: workstation monitor display Low contrast object detectability Assurance of electrical and mechanical safety and sterility Verify measurement accuracy during program initiation 2
3 American College of Radiology Requirements: General Sonography ACR physics committee developing this standard based it on measurements using a phantom. A phantom was selected (fall,1998) but turned out to be much more costly than originally promised. Physics Committee chose to leave the choice of phantoms to the user ACR staff published the following: Using a phantom will be helpful in responding to questions about low contrast detectability. However, the use of a phantom is optional at this time. Questions relating to characteristics associated with system sensitivity and image uniformity may be answered without the use of a phantom or test object. American Institute of Ultrasound in Medicine: General US QC 2008 Originally called QA in the Clinic Sonographers and clinicians helped draft Outlines what to do Sonographers Physicists/engineers Limited information on methodology Requires a phantom Phantom left to users See 3
4 American Institute of Ultrasound in Medicine: General US QC American Institute of Ultrasound in Medicine: General Sonography (physicist, engineer) Transducer Uniformity Phantom Electronic probe tester Sensitivity, Maximum depth of visualization into a phantom Distance measurement accuracy Target detection and imaging Focal targets such as simulated cysts or low contrast objects Choice of phantoms left to users Image display fidelity Cables OK, air filters clean, mechanical and electrical inspection Minimum frequency Annual for physicists/engineer tests Program includes daily, weekly checks by sonographers 4
5 Water-based gels Currently used materials: Advantages: Speed of sound = 1540 m/s Attenuation proportional to frequency Backscatter Disadvantages: Subject to desiccation Must be kept in containers Requires scanning window Currently used TM materials: Solid, non-water-based materials (urethane) Advantages: Not subject to desiccation No need for scanning window Produce tissue-like backscatter Disadvantages: C= m/s Attenuation not proportional to frequency (~f 1.6 ) Surface easily damaged if not cleaned regularly to remove gels 5
6 Maximum Depth of Visualization Considered by many as a good overall check of the integrity of the system FOV at 18 cm (or set to match the phantom) Output power (MI) at max Transmit focus at deepest settings Gains, TGC for visualization to the maximum distance possible Maximum Depth of Visualization How far can you see the speckle pattern in the material? 6
7 Objective Maximum Depth of Visualization Gibson, Dudley, Griffith, A computerized Quality Control System for B-mode Ultrasound, Ultrasound Med Biol 27:1697 : , 1711, Shi, Al-Sadah, Mackie, Zagzebski, Signal to Noise Ratio Estimates on Ultrasound Depth of Penetration (abstract only), Medical Physics 30: : 11367, Gorny, Tradup, Bernatz, Stekel, and Hangiandreou, Evaluation of an Automated Depth of Penetration Measurement ofr the Purpose of Ultrasonic Scanner Comparison, (abstract only), J. Ultrasound Med 23: S76, MPV Compute mean pixel value vs depth for phantom (signal) and for noise only (noise) Depth where signal/noise = 1.5 =DOP SNR -DOP tracks well Observer-DOP 16 Observers-DOP & SNR-DOP comparision (ATL C5-2) Depth (cm) (Al-Sadah, UW-Madison) Observer's DOP DOP using SNR'= Mechanical Index MI 7
8 Routine QA: Image Uniformity Considered to be the most impor- tant and useful test! Ideally: No evidence of element dropout No vertical shadows No loss of sensitivity near edges Uniformity Turn off spatial compounding, speckle reduction Use 1 tx focus Search for shadows emanating from transducer Common in new and old probes! 8
9 Coupling to a flat surface phantom scanning window with curvilinear transducers Coupling to a flat surface phantom scanning window with curvilinear transducers Solution 1: rock transducer from side to side 9
10 Coupling to a flat surface phantom scanning window with curvilinear transducers Solutside to side Coupling to a flat surface phantom scanning window with curvilinear transducers 10
11 Coupling to a flat surface phantom scanning window with curvilinear transducers Solution 2: Use a liquid TM material Coupling to a flat surface with curvilinear transducers 1 Sensitivity 0.8 Volts p-p Elements Electronic Probe test Solution 2: Use a liquid TM material (pseudo liquid TM slurry; Madsen et al.) 11
12 Coupling to a flat surface phantom scanning window with curvilinear transducers Solution 3: Use a phantom having concave windows (Goodsitt et al) Visualizing 1-2 element dropouts Use persistence; translate transducer. 12
13 Visualizing 1-2 element dropouts Use persistence; translate transducer. Visualizing 1-2 element dropouts Record Image loops while translating the transducer; Compress a 100 frame loop into 1 image (averages speckle) 1 image Ultrasonix SonixTouch Enables masking off individual elements Elements 95 and 96 masked 13
14 Visualizing 1-2 element dropouts Record Image loops while translating the transducer; Compress a 100 frame loop into 1 image (averages speckle) Average image Ultrasonix SonixTouch Enables masking off individual elements Elements 95 and 96 masked Visualizing 1-2 element dropouts Use persistence; translate transducer Use Record Image loops while translating the transducer; compress a 100 frame loop into 1 image (averages speckle) Phased array transducer performance Need electronic probe tester Transducer test modes on the ultrasound instrument are possible; manufacturers need to be encouraged to provide these Future 2-D probes, probes that do part of the beam forming within the probe housing 14
15 Gray Scale Photography, Workstation Fidelity Important for monitor on machine to be set up properly to view all echo levels available and entire gray bar pattern. Set up during acceptance testing Take steps to avoid the meat hook effect (mark or inscribe contrast and brightness controls) Workstation monitors and/or hardcopy devices should be matched to the machine monitor. AAPM, July, 2010 Check hardcopy, workstation displays. Are all gray bar transitions visible? 15
16 SMPTE Test Pattern Available on most scanners 0% to 100% gray bar pattern Squares for detecting geometric distortion Are all gray transitions visible? Distance Measurement Accuracy: Vertical Actual 2.5 Measure 2.46 error 1.5% Acceptable 16
17 Distance Measurement Accuracy: Horizontal Actual 1.0 Measure 1.01 error <1% Acceptable Routine QA (ACR General US Program) Distance Measurement Accuracy tests as needed Necessary? ( Scanner is a transducer tied to a computer. ) May be important for specific uses Images reregistered from 3-D data sets Workstation measurements Radiation seed implants 17
18 ACR Recommended QC for Breast US Maximum Depth of visualization Vertical and horiz measurement accuracy Semiannually Semiannually Service eng./physicist Service eng./physicist Hardcopy Recording Semiannually Service eng./physicist Uniformity Semiannually Service eng./physicist Elec-mechanical condition Semiannually Service eng./physicist Anechoic void perception Semiannually Service eng./physicist Ring down Semiannually Service eng./physicist Lateral Resolution Semiannually Service eng./physicist QC Checklist Semiannually Service eng./physicist Adherence to infection control procedures After each biopsy Sonographer Clean transducers After each biopsy Sonographer Vertical and horiz measurement accuracy Quarterly Sonographer Grey-scale photography Quarterly Sonographer Dead Zone? No suitable targets for dead zone testing, even in phantoms that advertise deadzone tragets! 18
19 Anechoic Sphere Imaging, a better choice Anechoic Sphere Imaging 19
20 Image of a phantom is useful for qualitative comparisons! Conventional Spatial Compounding Images obtained during routine Breast QC testing, 3/2010 Image of a phantom is useful! Conventional Spatial Compounding Images obtained 1 month later, after a software change; 20
21 Routine QA: Cleanliness, safety Transducer Inspection Delaminations Frayed cables Proper cleaning Console Air filters Viewing monitor, keyboard clean Lights, indicators Wheels, wheel locks Proper cleaning Summary Setting up, maintaining an equipment QA program is straight forward The ACR listed procedures are a useful, basic QA program Directed by physicist or lab personnel Integrated effort including lab and technical staff Requires a Phantom Closely correlates with AIUM list of factors to test Transducer uniformity is a frequent fault in today s scanning machines Computational methods can be developed for objective tests 21
22 References Goodsitt et al, Real-time B-mode ultrasound quality control test procedures, Medical Physics 1998; 25: J Zagzebski, US quality assurance with phantoms. In Categorical Course in Diagnostic Radiology Physics: CT and US Cross-Sectional Imaging, Edited by L. Goldman and B. Fowlkes, 2000, Oak Brook, IL: Radiological Society of North America, pp J Zagzebski and J Kofler, Ultrasound Equipment Quality Assurance, in Quality Management in the Imaging Sciences, ed. By J Papp, 2002, St. Louis, Mosby, pp QA Manual for Gray Scale Ultrasound Scanners, 1995, American Institute of Ultrasound in Medicine, Laurel, MA. D Groth et al, Cathode ray tube quality control and acceptance program: initial results for clinical PACS displays, Radiographics 2001, 21:
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
More informationDiagnostic Ultrasound Performance Testing & Accreditation
Diagnostic Ultrasound Performance Testing & Accreditation Nicholas J Hangiandreou, PhD Associate Professor of Medical Physics Department of Radiology Mayo Clinic and Foundation Rochester, Minnesota hangiandreou@mayo.edu
More informationUltrasound Phantoms: B-Mode, Doppler and Others. Heather M. Pierce, B.S. Norfolk, VA
Ultrasound Phantoms: B-Mode, Doppler and Others Heather M. Pierce, B.S. Norfolk, VA Overview Design of ultrasound phantoms Types of phantoms available Use of phantoms in a QA/QC program Advantages and
More informationINTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
More informationHealth. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors
Health NYC Quality Assurance Guidelines for Primary Diagnostic Monitors June 2015 Content I. Overview... 1 II. Definitions... 1 III. Acceptance Testing Requirements... 4 IV. Bi-weekly PDM Testing... 4
More informationQuality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
More informationPrepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
More informationScanning Acoustic Microscopy Training
Scanning Acoustic Microscopy Training This presentation and images are copyrighted by Sonix, Inc. They may not be copied, reproduced, modified, published, uploaded, posted, transmitted, or distributed
More informationBreast MRI Quality Control
Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Discuss the importance of breast MRI quality control (QC). Provide an overview of the new ACR Breast MRI Accreditation
More informationPhysics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
More informationJoon-Il Choi, MD, Seung Eun Jung, MD, Pyo Nyun Kim, MD, Sang Hoon Cha, MD, Jae Kwan Jun, MD, Hoo-Yeon Lee, MD, Eun-Cheol Park, MD
ORIGINAL RESEARCH Quality Assurance in Ultrasound Screening for Hepatocellular Carcinoma Using a Standardized Phantom and Standard Clinical Images A 3-Year National Investigation in Korea Joon-Il Choi,
More informationAn abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen.
Scan for mobile link. Ultrasound - Abdomen Ultrasound imaging of the abdomen uses sound waves to produce pictures of the structures within the upper abdomen. It is used to help diagnose pain or distention
More informationREGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
More informationQuality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
More informationUltrasonic Wave Propagation Review
Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic
More informationApplicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2
Version 3.1 Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 MQSA Quality Control Manual for Monochrome Monitors for Mammography Foreword Purpose of this Document This document is for
More informationHIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950
HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple
More informationSUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
More informationChecklist for the Qualification of Digital Detector Array Systems
Paper No. 010-12 Checklist for the Qualification of Digital Detector Array Systems 1 February 2012 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction
More informationRobot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
More informationWHERE IN THE WORLD JILL LIPOTI?
WHERE IN THE WORLD IS JILL LIPOTI? HELLO FROM NEW JERSEY CRCPD - National Symposium on Fusion Imaging and Multimodalities February 18-20, 2004 Kansas City, Missouri New Jersey s Requirements As They Pertain
More informationPipeline External Corrosion Analysis Using a 3D Laser Scanner
Pipeline Technology Conference 2013 Pipeline External Corrosion Analysis Using a 3D Laser Scanner Pierre-Hugues ALLARD, Charles MONY Creaform, www.creaform3d.com 5825 rue St-Georges, Lévis (QC), Canada,
More informationAccreditation Is Coming
Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt
More informationThe promise of ultrasonic phased arrays and the role of modeling in specifying systems
1 modeling in specifying systems ABSTRACT Guillaume Neau and Deborah Hopkins This article illustrates the advantages of phased-array systems and the value of modeling through several examples taken from
More informationPerformance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
More informationMQSA Quality Control Manual for Monochrome Displays for Mammography
MQSA Quality Control Manual for Monochrome Displays for Mammography Ver1.0 Applicable Models LMD-DM50, LMD-DM30 Sony Corporation - 1 - MQSA Quality Control Manual for Monochrome Displays for Mammography
More informationThe American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions
The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions Table of Contents APPLICATION - GENERAL... 1 MOVED FACILITIES AND UNITS... 3 EQUIPMENT... 4 PERSONNEL...
More informationMedical imaging monitors specification guidelines
Medical imaging monitors specification guidelines Document details Contact for enquiries and proposed changes If you have any questions regarding this document or if you have a suggestion for improvements,
More informationASSESSMENT REPORT. Program: AAS Diagnostic Medical Sonography Cardiac Vascular Tract
ASSESSMENT REPORT Date Submitted: February 9, 2015 School: Engelstad School of Health Sciences Program: AAS Diagnostic Medical Sonography Cardiac Vascular Tract Submitted By: Tracy Lopez, Program Director
More informationRole of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
More informationSurveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,
More informationCompliance Guidance for FLUOROSCOPIC QUALITY CONTROL
Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp
More informationAutomated Optical Inspection is one of many manufacturing test methods common in the assembly of printed circuit boards. This list includes:
What is AOI? Automated Optical Inspection is one of many manufacturing test methods common in the assembly of printed circuit boards. This list includes: Test methods for electronic assemblies: - FT (Functional
More informationPET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
More informationQUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.
More informationQuality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
More informationStudy the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
More informationACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
More informationSynthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
More informationA pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)
chapter 2 Pulsed Ultrasound In diagnostic ultrasound imaging, short bursts, or pulses, of acoustic energy are used to create anatomic images. Continuous wave sound cannot create anatomic images. Analogy
More informationHunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality
Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL
More informationUpdate on ACR Digital Mammography QC Manual
Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual
More informationAxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy
Microscopy from Carl Zeiss AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy New Dimensions in Performance AxioCam MR from Carl Zeiss Both
More informationModule 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
More informationkv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
More informationEncoded Phased Array Bridge Pin Inspection
Encoded Phased Array Bridge Pin Inspection James S. Doyle Baker Testing Services, Inc. 22 Reservoir Park Dr. Rockland, MA 02370 (781) 871-4458; fax (781) 871-0123; e-mail jdoyle@bakertesting.com Product
More informationScan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
More informationPIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany
ABSTRACT: PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany Today, in-line inspection tools are used routinely
More informationBIOMEDICAL ULTRASOUND
BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic
More informationAn Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
More informationA PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA
A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA N. Zarrinpanjeh a, F. Dadrassjavan b, H. Fattahi c * a Islamic Azad University of Qazvin - nzarrin@qiau.ac.ir
More informationR/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
More informationQuality control of CT systems by automated monitoring of key performance indicators: a two-year study
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Quality control of CT systems by automated monitoring of key performance indicators: a two-year study Patrik Nowik, a Robert Bujila,
More informationStaff Doses & Practical Radiation Protection in DEXA
Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /
More informationUnderstanding Digital Modalities: System Integration and Use
Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information
More informationIAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.
IAC Ch 41, p.1 641 41.7 (136C) X-ray machines used for stereotactically guided breast biopsy. 41.7(1) Definitions. In addition to the definitions provided in rules 641 38.2(136C), 641 40.2(136C), and 641
More informationAIUM Ultrasound Practice Accreditation. Paula S. Woletz, MPH, RDMS, RDCS American Institute of Ultrasound in Medicine Laurel, MD
AIUM Ultrasound Practice Accreditation Paula S. Woletz, MPH, RDMS, RDCS American Institute of Ultrasound in Medicine Laurel, MD Accreditation: A voluntary peer review process that Assesses the strengths
More informationPersonal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture
More informationDigital Breast Tomosynthesis QC Requirements
Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist
More informationScanners and How to Use Them
Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types
More informationQUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
More informationUltrasound - Vascular
Scan for mobile link. Ultrasound - Vascular Vascular ultrasound uses sound waves to evaluate the body s circulatory system and help identify blockages and detect blood clots. A Doppler ultrasound study
More informationImpact Echo Scanning Technology for Internal Grout Condition Evaluation in Post-Tensioned Bridge Ducts
Impact Echo Scanning Technology for Internal Grout Condition Evaluation in Post-Tensioned Bridge Ducts Yajai Tinkey, Ph.D. Olson Engineering, Wheat Ridge, Colorado Larry D. Olson, P.E. Olson Engineering,
More informationAutomotive Applications of 3D Laser Scanning Introduction
Automotive Applications of 3D Laser Scanning Kyle Johnston, Ph.D., Metron Systems, Inc. 34935 SE Douglas Street, Suite 110, Snoqualmie, WA 98065 425-396-5577, www.metronsys.com 2002 Metron Systems, Inc
More informationDose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography
More informationPercutaneous Abscess Drainage
Scan for mobile link. Percutaneous Abscess Drainage An abscess is an infected fluid collection within the body. Percutaneous abscess drainage uses imaging guidance to place a thin needle through the skin
More informationPreparation iagnostic Medical Sonographer Overview"
Diagnostic Medical Sonographer Overview The Field - Preparation - Specialty Areas - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Diagnostic imaging
More information1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.
CHAPTER 3: OSCILLOSCOPE AND SIGNAL GENERATOR 3.1 Introduction to oscilloscope 1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. 2. The graph show signal change
More informationAdvancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications
Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Janet E. Semmens Sonoscan, Inc. 2149 E. Pratt Boulevard Elk Grove Village, IL 60007 USA Phone: (847)
More informationMRI QA Technologist s Tests
AAPM 55th Annual Meeting MRI QA Technologist s Tests Guang Jia, PhD Associate Professor Department of Physics Louisiana State University Background Outline QC importance Technologist s role QC test tips
More informationPicture archiving and communication systems (PACS) and guidelines on diagnostic display devices
IT Guidance Documents Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists
More informationCoronis 5MP Mammo. The standard of care for digital mammography
Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment
More informationHow can I be sure it will be performed to a high standard?
FREQUENTLY ASKED QUESTIONS (About ultrasound scans at Derriford) What is an Ultrasound Scan? How does it work? What are the dangers? What are the advantages? Who will perform my scan? How can I be sure
More informationIrradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
More informationHow To Use An Ultrasound For Medical Research
Scan for mobile link. General Ultrasound What is General Ultrasound Imaging? Ultrasound is safe and painless, and produces pictures of the inside of the body using sound waves. Ultrasound imaging, also
More informationNDT to Identify Concrete Bridge Deck Deterioration
NDT to Identify Concrete Bridge Deck Deterioration Nenad Gucunski Rutgers University Department of Civil and Environmental Engineering Center for Advanced Infrastructure and Transportation (CAIT) April
More informationDental Radiography Core Subject. Digital Radiography
Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of
More informationDoppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19
Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and
More informationFirst floor, Main Hospital North Services provided 24/7 365 days per year
First floor, Main Hospital North Services provided 24/7 365 days per year General Radiology (X-ray) Fluoroscopy Ultrasound (Sonography) Nuclear Medicine P.E.T. imaging Computed Tomography (CT scan) Magnetic
More informationName: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography
Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,
More informationApplication of Digital Radiography for the Detection and Classification of Pneumoconiosis
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational
More informationVarious Technics of Liquids and Solids Level Measurements. (Part 3)
(Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,
More informationMoving Forward What does this mean for the Medical Physicist and the Imaging Community?
Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical
More informationForensic Science: The Basics. Microscopy
Forensic Science: The Basics Microscopy Chapter 6 Jay A. Siegel,Ph.D. Power point presentation by Greg Galardi, Peru State College, Peru Nebraska Presentation by Greg Galardi, Peru State College CRC Press,
More informationCT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
More informationSAR ASSOCIATED WITH THE USE OF HANDS-FREE MOBILE TELEPHONES
SAR ASSOCIATED WITH THE USE OF HANDS-FREE MOBILE TELEPHONES S.J. Porter, M.H. Capstick, F. Faraci, I.D. Flintoft and A.C. Marvin Department of Electronics, University of York Heslington, York, YO10 5DD,
More informationThe Basics of Scanning Electron Microscopy
The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single
More informationIntroduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.
Slide 1 Digital Mammography Detector QC Slide 2 Introduction What is FFDM QC and why is it important? Eric A. Berns, Ph.D. What to know before you start eric.berns@gmail.com Overview and compare QC tests
More informationQuantitative Imaging In Clinical Trials Using PET/CT: Update
Quantitative Imaging In Clinical Trials Using PET/CT: Update Paul Kinahan, Robert Doot Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Supported by RSNA Quantitative
More informationBasics of Ultrasound Imaging
2 Basics of Ultrasound Imaging Vincent Chan and Anahi Perlas Introduction... Basic Principles of B-Mode US... Generation of Ultrasound Pulses... Ultrasound Wavelength and Frequency... Ultrasound Tissue
More informationMulti-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014
Multi-Channel Radiochromic Film Dosimetry Adapted from A.Micke Spain, April 2014 Single Channel Film Dosimetry Calibration Curve X=RR ave = R ave (D) D R = D R (R ave ) Color channels X=RGB D X = D( X
More informationLIST OF CONTENTS CHAPTER CONTENT PAGE DECLARATION DEDICATION ACKNOWLEDGEMENTS ABSTRACT ABSTRAK
vii LIST OF CONTENTS CHAPTER CONTENT PAGE DECLARATION DEDICATION ACKNOWLEDGEMENTS ABSTRACT ABSTRAK LIST OF CONTENTS LIST OF TABLES LIST OF FIGURES LIST OF NOTATIONS LIST OF ABBREVIATIONS LIST OF APPENDICES
More informationCURRICULUM VITAE. Kalpana M. Kanal, Ph.D., DABR
May 31, 2005 ADDRESS: PERSONAL: CURRICULUM VITAE Kalpana M. Kanal, Ph.D., DABR Department of Radiology University of Washington Medical Center 1959 NE Pacific Avenue Seattle, WA 98195-7115 (206) 616-1966
More informationTechnical data. General specifications. Indicators/operating means. 30 Hz Multiplex operation 30 Hz / n, n = number of sensors, n 5
Model Number Single head system Features Parameterization interface for the application-specific adjustment of the sensor setting via the service program ULTRA 000 programmable switch outputs Hysteresis
More informationA software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department
More informationLecture 14. Point Spread Function (PSF)
Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect
More informationPerformance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
More information1. General Description
. General Description The is a Color Active Matrix with an integral Cold Cathode Fluorescent Lamp(CCFL) back light system. The matrix employs asi Thin Film Transistor as the active element. It is a transmissive
More informationVivid S6. Cardiovascular ultrasound system. imagination at work. GE Healthcare
GE Healthcare Vivid S6 Cardiovascular ultrasound system imagination at work Taking care of everyone. Strong performance and innovative miniaturization make the Vivid S6 a highly desirable system for a
More information