Routine Ultrasound Equipment Tests for Quality Assurance. Contents

Size: px
Start display at page:

Download "Routine Ultrasound Equipment Tests for Quality Assurance. Contents"

Transcription

1 Routine Ultrasound Equipment Tests for Quality Assurance James A. Zagzebski, Ph.D. Nick Rupert, M.S. Ryan DeWall, M.S. Tomy Varghese, Ph.D. Dept. of Medical Physics University of Wisconsin, Madison, WI, USA Contents Elements of a Routine Test Program What system components are considered most important in routine testing? Concentrate on transducers! Steps to carry out tests on a regular basis Activity of Ultrasound Lab Accreditation Groups Basic QA methods Phantoms and tests 1

2 Routine QA Program Guidelines available from US facility accreditation bodies (ACR; AIUM) Emphasis is given to safety, cleanliness, and very basic performance tests. Issues related to image performance, such as spatial resolution, are not included (except ACR breast). Traditional parameters such as distance measurements are believed by many to be unnecessary. Includedi n some protocols, but only briefly. Routine QA (ACR General US Program) System sensitivity and/or penetration capability 1999: Do for 2 probes (?) Image Uniformity Look for element dropout Photography and other hard copy recording 2010: workstation monitor display Low contrast object detectability Assurance of electrical and mechanical safety and sterility Verify measurement accuracy during program initiation 2

3 American College of Radiology Requirements: General Sonography ACR physics committee developing this standard based it on measurements using a phantom. A phantom was selected (fall,1998) but turned out to be much more costly than originally promised. Physics Committee chose to leave the choice of phantoms to the user ACR staff published the following: Using a phantom will be helpful in responding to questions about low contrast detectability. However, the use of a phantom is optional at this time. Questions relating to characteristics associated with system sensitivity and image uniformity may be answered without the use of a phantom or test object. American Institute of Ultrasound in Medicine: General US QC 2008 Originally called QA in the Clinic Sonographers and clinicians helped draft Outlines what to do Sonographers Physicists/engineers Limited information on methodology Requires a phantom Phantom left to users See 3

4 American Institute of Ultrasound in Medicine: General US QC American Institute of Ultrasound in Medicine: General Sonography (physicist, engineer) Transducer Uniformity Phantom Electronic probe tester Sensitivity, Maximum depth of visualization into a phantom Distance measurement accuracy Target detection and imaging Focal targets such as simulated cysts or low contrast objects Choice of phantoms left to users Image display fidelity Cables OK, air filters clean, mechanical and electrical inspection Minimum frequency Annual for physicists/engineer tests Program includes daily, weekly checks by sonographers 4

5 Water-based gels Currently used materials: Advantages: Speed of sound = 1540 m/s Attenuation proportional to frequency Backscatter Disadvantages: Subject to desiccation Must be kept in containers Requires scanning window Currently used TM materials: Solid, non-water-based materials (urethane) Advantages: Not subject to desiccation No need for scanning window Produce tissue-like backscatter Disadvantages: C= m/s Attenuation not proportional to frequency (~f 1.6 ) Surface easily damaged if not cleaned regularly to remove gels 5

6 Maximum Depth of Visualization Considered by many as a good overall check of the integrity of the system FOV at 18 cm (or set to match the phantom) Output power (MI) at max Transmit focus at deepest settings Gains, TGC for visualization to the maximum distance possible Maximum Depth of Visualization How far can you see the speckle pattern in the material? 6

7 Objective Maximum Depth of Visualization Gibson, Dudley, Griffith, A computerized Quality Control System for B-mode Ultrasound, Ultrasound Med Biol 27:1697 : , 1711, Shi, Al-Sadah, Mackie, Zagzebski, Signal to Noise Ratio Estimates on Ultrasound Depth of Penetration (abstract only), Medical Physics 30: : 11367, Gorny, Tradup, Bernatz, Stekel, and Hangiandreou, Evaluation of an Automated Depth of Penetration Measurement ofr the Purpose of Ultrasonic Scanner Comparison, (abstract only), J. Ultrasound Med 23: S76, MPV Compute mean pixel value vs depth for phantom (signal) and for noise only (noise) Depth where signal/noise = 1.5 =DOP SNR -DOP tracks well Observer-DOP 16 Observers-DOP & SNR-DOP comparision (ATL C5-2) Depth (cm) (Al-Sadah, UW-Madison) Observer's DOP DOP using SNR'= Mechanical Index MI 7

8 Routine QA: Image Uniformity Considered to be the most impor- tant and useful test! Ideally: No evidence of element dropout No vertical shadows No loss of sensitivity near edges Uniformity Turn off spatial compounding, speckle reduction Use 1 tx focus Search for shadows emanating from transducer Common in new and old probes! 8

9 Coupling to a flat surface phantom scanning window with curvilinear transducers Coupling to a flat surface phantom scanning window with curvilinear transducers Solution 1: rock transducer from side to side 9

10 Coupling to a flat surface phantom scanning window with curvilinear transducers Solutside to side Coupling to a flat surface phantom scanning window with curvilinear transducers 10

11 Coupling to a flat surface phantom scanning window with curvilinear transducers Solution 2: Use a liquid TM material Coupling to a flat surface with curvilinear transducers 1 Sensitivity 0.8 Volts p-p Elements Electronic Probe test Solution 2: Use a liquid TM material (pseudo liquid TM slurry; Madsen et al.) 11

12 Coupling to a flat surface phantom scanning window with curvilinear transducers Solution 3: Use a phantom having concave windows (Goodsitt et al) Visualizing 1-2 element dropouts Use persistence; translate transducer. 12

13 Visualizing 1-2 element dropouts Use persistence; translate transducer. Visualizing 1-2 element dropouts Record Image loops while translating the transducer; Compress a 100 frame loop into 1 image (averages speckle) 1 image Ultrasonix SonixTouch Enables masking off individual elements Elements 95 and 96 masked 13

14 Visualizing 1-2 element dropouts Record Image loops while translating the transducer; Compress a 100 frame loop into 1 image (averages speckle) Average image Ultrasonix SonixTouch Enables masking off individual elements Elements 95 and 96 masked Visualizing 1-2 element dropouts Use persistence; translate transducer Use Record Image loops while translating the transducer; compress a 100 frame loop into 1 image (averages speckle) Phased array transducer performance Need electronic probe tester Transducer test modes on the ultrasound instrument are possible; manufacturers need to be encouraged to provide these Future 2-D probes, probes that do part of the beam forming within the probe housing 14

15 Gray Scale Photography, Workstation Fidelity Important for monitor on machine to be set up properly to view all echo levels available and entire gray bar pattern. Set up during acceptance testing Take steps to avoid the meat hook effect (mark or inscribe contrast and brightness controls) Workstation monitors and/or hardcopy devices should be matched to the machine monitor. AAPM, July, 2010 Check hardcopy, workstation displays. Are all gray bar transitions visible? 15

16 SMPTE Test Pattern Available on most scanners 0% to 100% gray bar pattern Squares for detecting geometric distortion Are all gray transitions visible? Distance Measurement Accuracy: Vertical Actual 2.5 Measure 2.46 error 1.5% Acceptable 16

17 Distance Measurement Accuracy: Horizontal Actual 1.0 Measure 1.01 error <1% Acceptable Routine QA (ACR General US Program) Distance Measurement Accuracy tests as needed Necessary? ( Scanner is a transducer tied to a computer. ) May be important for specific uses Images reregistered from 3-D data sets Workstation measurements Radiation seed implants 17

18 ACR Recommended QC for Breast US Maximum Depth of visualization Vertical and horiz measurement accuracy Semiannually Semiannually Service eng./physicist Service eng./physicist Hardcopy Recording Semiannually Service eng./physicist Uniformity Semiannually Service eng./physicist Elec-mechanical condition Semiannually Service eng./physicist Anechoic void perception Semiannually Service eng./physicist Ring down Semiannually Service eng./physicist Lateral Resolution Semiannually Service eng./physicist QC Checklist Semiannually Service eng./physicist Adherence to infection control procedures After each biopsy Sonographer Clean transducers After each biopsy Sonographer Vertical and horiz measurement accuracy Quarterly Sonographer Grey-scale photography Quarterly Sonographer Dead Zone? No suitable targets for dead zone testing, even in phantoms that advertise deadzone tragets! 18

19 Anechoic Sphere Imaging, a better choice Anechoic Sphere Imaging 19

20 Image of a phantom is useful for qualitative comparisons! Conventional Spatial Compounding Images obtained during routine Breast QC testing, 3/2010 Image of a phantom is useful! Conventional Spatial Compounding Images obtained 1 month later, after a software change; 20

21 Routine QA: Cleanliness, safety Transducer Inspection Delaminations Frayed cables Proper cleaning Console Air filters Viewing monitor, keyboard clean Lights, indicators Wheels, wheel locks Proper cleaning Summary Setting up, maintaining an equipment QA program is straight forward The ACR listed procedures are a useful, basic QA program Directed by physicist or lab personnel Integrated effort including lab and technical staff Requires a Phantom Closely correlates with AIUM list of factors to test Transducer uniformity is a frequent fault in today s scanning machines Computational methods can be developed for objective tests 21

22 References Goodsitt et al, Real-time B-mode ultrasound quality control test procedures, Medical Physics 1998; 25: J Zagzebski, US quality assurance with phantoms. In Categorical Course in Diagnostic Radiology Physics: CT and US Cross-Sectional Imaging, Edited by L. Goldman and B. Fowlkes, 2000, Oak Brook, IL: Radiological Society of North America, pp J Zagzebski and J Kofler, Ultrasound Equipment Quality Assurance, in Quality Management in the Imaging Sciences, ed. By J Papp, 2002, St. Louis, Mosby, pp QA Manual for Gray Scale Ultrasound Scanners, 1995, American Institute of Ultrasound in Medicine, Laurel, MA. D Groth et al, Cathode ray tube quality control and acceptance program: initial results for clinical PACS displays, Radiographics 2001, 21:

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim- QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound

More information

Diagnostic Ultrasound Performance Testing & Accreditation

Diagnostic Ultrasound Performance Testing & Accreditation Diagnostic Ultrasound Performance Testing & Accreditation Nicholas J Hangiandreou, PhD Associate Professor of Medical Physics Department of Radiology Mayo Clinic and Foundation Rochester, Minnesota hangiandreou@mayo.edu

More information

Ultrasound Phantoms: B-Mode, Doppler and Others. Heather M. Pierce, B.S. Norfolk, VA

Ultrasound Phantoms: B-Mode, Doppler and Others. Heather M. Pierce, B.S. Norfolk, VA Ultrasound Phantoms: B-Mode, Doppler and Others Heather M. Pierce, B.S. Norfolk, VA Overview Design of ultrasound phantoms Types of phantoms available Use of phantoms in a QA/QC program Advantages and

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors Health NYC Quality Assurance Guidelines for Primary Diagnostic Monitors June 2015 Content I. Overview... 1 II. Definitions... 1 III. Acceptance Testing Requirements... 4 IV. Bi-weekly PDM Testing... 4

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals

More information

Scanning Acoustic Microscopy Training

Scanning Acoustic Microscopy Training Scanning Acoustic Microscopy Training This presentation and images are copyrighted by Sonix, Inc. They may not be copied, reproduced, modified, published, uploaded, posted, transmitted, or distributed

More information

Breast MRI Quality Control

Breast MRI Quality Control Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Discuss the importance of breast MRI quality control (QC). Provide an overview of the new ACR Breast MRI Accreditation

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

Joon-Il Choi, MD, Seung Eun Jung, MD, Pyo Nyun Kim, MD, Sang Hoon Cha, MD, Jae Kwan Jun, MD, Hoo-Yeon Lee, MD, Eun-Cheol Park, MD

Joon-Il Choi, MD, Seung Eun Jung, MD, Pyo Nyun Kim, MD, Sang Hoon Cha, MD, Jae Kwan Jun, MD, Hoo-Yeon Lee, MD, Eun-Cheol Park, MD ORIGINAL RESEARCH Quality Assurance in Ultrasound Screening for Hepatocellular Carcinoma Using a Standardized Phantom and Standard Clinical Images A 3-Year National Investigation in Korea Joon-Il Choi,

More information

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen.

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen. Scan for mobile link. Ultrasound - Abdomen Ultrasound imaging of the abdomen uses sound waves to produce pictures of the structures within the upper abdomen. It is used to help diagnose pain or distention

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

Ultrasonic Wave Propagation Review

Ultrasonic Wave Propagation Review Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic

More information

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 Version 3.1 Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 MQSA Quality Control Manual for Monochrome Monitors for Mammography Foreword Purpose of this Document This document is for

More information

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950 HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

Checklist for the Qualification of Digital Detector Array Systems

Checklist for the Qualification of Digital Detector Array Systems Paper No. 010-12 Checklist for the Qualification of Digital Detector Array Systems 1 February 2012 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction

More information

Robot Perception Continued

Robot Perception Continued Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart

More information

WHERE IN THE WORLD JILL LIPOTI?

WHERE IN THE WORLD JILL LIPOTI? WHERE IN THE WORLD IS JILL LIPOTI? HELLO FROM NEW JERSEY CRCPD - National Symposium on Fusion Imaging and Multimodalities February 18-20, 2004 Kansas City, Missouri New Jersey s Requirements As They Pertain

More information

Pipeline External Corrosion Analysis Using a 3D Laser Scanner

Pipeline External Corrosion Analysis Using a 3D Laser Scanner Pipeline Technology Conference 2013 Pipeline External Corrosion Analysis Using a 3D Laser Scanner Pierre-Hugues ALLARD, Charles MONY Creaform, www.creaform3d.com 5825 rue St-Georges, Lévis (QC), Canada,

More information

Accreditation Is Coming

Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt

More information

The promise of ultrasonic phased arrays and the role of modeling in specifying systems

The promise of ultrasonic phased arrays and the role of modeling in specifying systems 1 modeling in specifying systems ABSTRACT Guillaume Neau and Deborah Hopkins This article illustrates the advantages of phased-array systems and the value of modeling through several examples taken from

More information

Performance testing for Precision 500D Classical R/F System

Performance testing for Precision 500D Classical R/F System Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)

More information

MQSA Quality Control Manual for Monochrome Displays for Mammography

MQSA Quality Control Manual for Monochrome Displays for Mammography MQSA Quality Control Manual for Monochrome Displays for Mammography Ver1.0 Applicable Models LMD-DM50, LMD-DM30 Sony Corporation - 1 - MQSA Quality Control Manual for Monochrome Displays for Mammography

More information

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions

The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions The American College of Radiology Breast Ultrasound Accreditation Program: Frequently Asked Questions Table of Contents APPLICATION - GENERAL... 1 MOVED FACILITIES AND UNITS... 3 EQUIPMENT... 4 PERSONNEL...

More information

Medical imaging monitors specification guidelines

Medical imaging monitors specification guidelines Medical imaging monitors specification guidelines Document details Contact for enquiries and proposed changes If you have any questions regarding this document or if you have a suggestion for improvements,

More information

ASSESSMENT REPORT. Program: AAS Diagnostic Medical Sonography Cardiac Vascular Tract

ASSESSMENT REPORT. Program: AAS Diagnostic Medical Sonography Cardiac Vascular Tract ASSESSMENT REPORT Date Submitted: February 9, 2015 School: Engelstad School of Health Sciences Program: AAS Diagnostic Medical Sonography Cardiac Vascular Tract Submitted By: Tracy Lopez, Program Director

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,

More information

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp

More information

Automated Optical Inspection is one of many manufacturing test methods common in the assembly of printed circuit boards. This list includes:

Automated Optical Inspection is one of many manufacturing test methods common in the assembly of printed circuit boards. This list includes: What is AOI? Automated Optical Inspection is one of many manufacturing test methods common in the assembly of printed circuit boards. This list includes: Test methods for electronic assemblies: - FT (Functional

More information

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity

More information

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT

ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

Synthetic Sensing: Proximity / Distance Sensors

Synthetic Sensing: Proximity / Distance Sensors Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,

More information

A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)

A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time) chapter 2 Pulsed Ultrasound In diagnostic ultrasound imaging, short bursts, or pulses, of acoustic energy are used to create anatomic images. Continuous wave sound cannot create anatomic images. Analogy

More information

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL

More information

Update on ACR Digital Mammography QC Manual

Update on ACR Digital Mammography QC Manual Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual

More information

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy Microscopy from Carl Zeiss AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy New Dimensions in Performance AxioCam MR from Carl Zeiss Both

More information

Module 13 : Measurements on Fiber Optic Systems

Module 13 : Measurements on Fiber Optic Systems Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)

More information

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,

More information

Encoded Phased Array Bridge Pin Inspection

Encoded Phased Array Bridge Pin Inspection Encoded Phased Array Bridge Pin Inspection James S. Doyle Baker Testing Services, Inc. 22 Reservoir Park Dr. Rockland, MA 02370 (781) 871-4458; fax (781) 871-0123; e-mail jdoyle@bakertesting.com Product

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany ABSTRACT: PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany Today, in-line inspection tools are used routinely

More information

BIOMEDICAL ULTRASOUND

BIOMEDICAL ULTRASOUND BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA N. Zarrinpanjeh a, F. Dadrassjavan b, H. Fattahi c * a Islamic Azad University of Qazvin - nzarrin@qiau.ac.ir

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study

Quality control of CT systems by automated monitoring of key performance indicators: a two-year study JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Quality control of CT systems by automated monitoring of key performance indicators: a two-year study Patrik Nowik, a Robert Bujila,

More information

Staff Doses & Practical Radiation Protection in DEXA

Staff Doses & Practical Radiation Protection in DEXA Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /

More information

Understanding Digital Modalities: System Integration and Use

Understanding Digital Modalities: System Integration and Use Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information

More information

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes. IAC Ch 41, p.1 641 41.7 (136C) X-ray machines used for stereotactically guided breast biopsy. 41.7(1) Definitions. In addition to the definitions provided in rules 641 38.2(136C), 641 40.2(136C), and 641

More information

AIUM Ultrasound Practice Accreditation. Paula S. Woletz, MPH, RDMS, RDCS American Institute of Ultrasound in Medicine Laurel, MD

AIUM Ultrasound Practice Accreditation. Paula S. Woletz, MPH, RDMS, RDCS American Institute of Ultrasound in Medicine Laurel, MD AIUM Ultrasound Practice Accreditation Paula S. Woletz, MPH, RDMS, RDCS American Institute of Ultrasound in Medicine Laurel, MD Accreditation: A voluntary peer review process that Assesses the strengths

More information

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture

More information

Digital Breast Tomosynthesis QC Requirements

Digital Breast Tomosynthesis QC Requirements Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist

More information

Scanners and How to Use Them

Scanners and How to Use Them Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types

More information

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative

More information

Ultrasound - Vascular

Ultrasound - Vascular Scan for mobile link. Ultrasound - Vascular Vascular ultrasound uses sound waves to evaluate the body s circulatory system and help identify blockages and detect blood clots. A Doppler ultrasound study

More information

Impact Echo Scanning Technology for Internal Grout Condition Evaluation in Post-Tensioned Bridge Ducts

Impact Echo Scanning Technology for Internal Grout Condition Evaluation in Post-Tensioned Bridge Ducts Impact Echo Scanning Technology for Internal Grout Condition Evaluation in Post-Tensioned Bridge Ducts Yajai Tinkey, Ph.D. Olson Engineering, Wheat Ridge, Colorado Larry D. Olson, P.E. Olson Engineering,

More information

Automotive Applications of 3D Laser Scanning Introduction

Automotive Applications of 3D Laser Scanning Introduction Automotive Applications of 3D Laser Scanning Kyle Johnston, Ph.D., Metron Systems, Inc. 34935 SE Douglas Street, Suite 110, Snoqualmie, WA 98065 425-396-5577, www.metronsys.com 2002 Metron Systems, Inc

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

Percutaneous Abscess Drainage

Percutaneous Abscess Drainage Scan for mobile link. Percutaneous Abscess Drainage An abscess is an infected fluid collection within the body. Percutaneous abscess drainage uses imaging guidance to place a thin needle through the skin

More information

Preparation iagnostic Medical Sonographer Overview"

Preparation iagnostic Medical Sonographer Overview Diagnostic Medical Sonographer Overview The Field - Preparation - Specialty Areas - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Diagnostic imaging

More information

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. CHAPTER 3: OSCILLOSCOPE AND SIGNAL GENERATOR 3.1 Introduction to oscilloscope 1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. 2. The graph show signal change

More information

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Janet E. Semmens Sonoscan, Inc. 2149 E. Pratt Boulevard Elk Grove Village, IL 60007 USA Phone: (847)

More information

MRI QA Technologist s Tests

MRI QA Technologist s Tests AAPM 55th Annual Meeting MRI QA Technologist s Tests Guang Jia, PhD Associate Professor Department of Physics Louisiana State University Background Outline QC importance Technologist s role QC test tips

More information

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices IT Guidance Documents Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists

More information

Coronis 5MP Mammo. The standard of care for digital mammography

Coronis 5MP Mammo. The standard of care for digital mammography Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment

More information

How can I be sure it will be performed to a high standard?

How can I be sure it will be performed to a high standard? FREQUENTLY ASKED QUESTIONS (About ultrasound scans at Derriford) What is an Ultrasound Scan? How does it work? What are the dangers? What are the advantages? Who will perform my scan? How can I be sure

More information

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department

More information

How To Use An Ultrasound For Medical Research

How To Use An Ultrasound For Medical Research Scan for mobile link. General Ultrasound What is General Ultrasound Imaging? Ultrasound is safe and painless, and produces pictures of the inside of the body using sound waves. Ultrasound imaging, also

More information

NDT to Identify Concrete Bridge Deck Deterioration

NDT to Identify Concrete Bridge Deck Deterioration NDT to Identify Concrete Bridge Deck Deterioration Nenad Gucunski Rutgers University Department of Civil and Environmental Engineering Center for Advanced Infrastructure and Transportation (CAIT) April

More information

Dental Radiography Core Subject. Digital Radiography

Dental Radiography Core Subject. Digital Radiography Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

First floor, Main Hospital North Services provided 24/7 365 days per year

First floor, Main Hospital North Services provided 24/7 365 days per year First floor, Main Hospital North Services provided 24/7 365 days per year General Radiology (X-ray) Fluoroscopy Ultrasound (Sonography) Nuclear Medicine P.E.T. imaging Computed Tomography (CT scan) Magnetic

More information

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,

More information

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational

More information

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Various Technics of Liquids and Solids Level Measurements. (Part 3) (Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,

More information

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Moving Forward What does this mean for the Medical Physicist and the Imaging Community? Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical

More information

Forensic Science: The Basics. Microscopy

Forensic Science: The Basics. Microscopy Forensic Science: The Basics Microscopy Chapter 6 Jay A. Siegel,Ph.D. Power point presentation by Greg Galardi, Peru State College, Peru Nebraska Presentation by Greg Galardi, Peru State College CRC Press,

More information

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of

More information

SAR ASSOCIATED WITH THE USE OF HANDS-FREE MOBILE TELEPHONES

SAR ASSOCIATED WITH THE USE OF HANDS-FREE MOBILE TELEPHONES SAR ASSOCIATED WITH THE USE OF HANDS-FREE MOBILE TELEPHONES S.J. Porter, M.H. Capstick, F. Faraci, I.D. Flintoft and A.C. Marvin Department of Electronics, University of York Heslington, York, YO10 5DD,

More information

The Basics of Scanning Electron Microscopy

The Basics of Scanning Electron Microscopy The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single

More information

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D. Slide 1 Digital Mammography Detector QC Slide 2 Introduction What is FFDM QC and why is it important? Eric A. Berns, Ph.D. What to know before you start eric.berns@gmail.com Overview and compare QC tests

More information

Quantitative Imaging In Clinical Trials Using PET/CT: Update

Quantitative Imaging In Clinical Trials Using PET/CT: Update Quantitative Imaging In Clinical Trials Using PET/CT: Update Paul Kinahan, Robert Doot Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Supported by RSNA Quantitative

More information

Basics of Ultrasound Imaging

Basics of Ultrasound Imaging 2 Basics of Ultrasound Imaging Vincent Chan and Anahi Perlas Introduction... Basic Principles of B-Mode US... Generation of Ultrasound Pulses... Ultrasound Wavelength and Frequency... Ultrasound Tissue

More information

Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014

Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014 Multi-Channel Radiochromic Film Dosimetry Adapted from A.Micke Spain, April 2014 Single Channel Film Dosimetry Calibration Curve X=RR ave = R ave (D) D R = D R (R ave ) Color channels X=RGB D X = D( X

More information

LIST OF CONTENTS CHAPTER CONTENT PAGE DECLARATION DEDICATION ACKNOWLEDGEMENTS ABSTRACT ABSTRAK

LIST OF CONTENTS CHAPTER CONTENT PAGE DECLARATION DEDICATION ACKNOWLEDGEMENTS ABSTRACT ABSTRAK vii LIST OF CONTENTS CHAPTER CONTENT PAGE DECLARATION DEDICATION ACKNOWLEDGEMENTS ABSTRACT ABSTRAK LIST OF CONTENTS LIST OF TABLES LIST OF FIGURES LIST OF NOTATIONS LIST OF ABBREVIATIONS LIST OF APPENDICES

More information

CURRICULUM VITAE. Kalpana M. Kanal, Ph.D., DABR

CURRICULUM VITAE. Kalpana M. Kanal, Ph.D., DABR May 31, 2005 ADDRESS: PERSONAL: CURRICULUM VITAE Kalpana M. Kanal, Ph.D., DABR Department of Radiology University of Washington Medical Center 1959 NE Pacific Avenue Seattle, WA 98195-7115 (206) 616-1966

More information

Technical data. General specifications. Indicators/operating means. 30 Hz Multiplex operation 30 Hz / n, n = number of sensors, n 5

Technical data. General specifications. Indicators/operating means. 30 Hz Multiplex operation 30 Hz / n, n = number of sensors, n 5 Model Number Single head system Features Parameterization interface for the application-specific adjustment of the sensor setting via the service program ULTRA 000 programmable switch outputs Hysteresis

More information

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department

More information

Lecture 14. Point Spread Function (PSF)

Lecture 14. Point Spread Function (PSF) Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect

More information

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Performance evaluation and quality assurance of Varian enhanced dynamic wedges JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department

More information

1. General Description

1. General Description . General Description The is a Color Active Matrix with an integral Cold Cathode Fluorescent Lamp(CCFL) back light system. The matrix employs asi Thin Film Transistor as the active element. It is a transmissive

More information

Vivid S6. Cardiovascular ultrasound system. imagination at work. GE Healthcare

Vivid S6. Cardiovascular ultrasound system. imagination at work. GE Healthcare GE Healthcare Vivid S6 Cardiovascular ultrasound system imagination at work Taking care of everyone. Strong performance and innovative miniaturization make the Vivid S6 a highly desirable system for a

More information