Metrological features of a beta absorption particulate air monitor operating with wireless communication system



Similar documents
Certificate: / 29 April 2014

PRODUCT CONFORMITY CERTIFICATE

CERTIFICATE of Product Conformity (QAL1 )

TÜV RHEINLAND ENERGIE UND UMWELT GMBH

MAC-ENC-PRO-057 AIR QUALITY MONITORING PROGRAM

Ambient Air Quality Division Air Quality and Noise Management Bureau Pollution Control Department

AMBIENT AIR QUALITY MONITORING SYSTEM. Air Quality Monitoring... Easier. Faster. Better.

WEATHER STATION FOR SOLAR FARM MONITORING

1.0 What Are the Purpose and Applicability of Performance Specification 11?

TransCity Legacy Way Air Monitoring Network

Ramon 2.2 Radon-Monitor

Gas Custody Transfer Calibration

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland

Air Pollution Control Division. Technical Services Program

RECOMMENDATION ITU-R F (Question ITU-R 157/9) b) that systems using this mode of propagation are already in service for burst data transmission,

New Automatic Radiation Monitoring Network in Slovenia

TÜV RHEINLAND IMMISSIONSSCHUTZ UND ENERGIESYSTEME GMBH

The design of the Romanian national air quality monitoring network

Energy savings in commercial refrigeration. Low pressure control

Overview. What is EMR? Electromagnetic Radiation (EMR) LA502 Special Studies Remote Sensing

Trace Gas Exchange Measurements with Standard Infrared Analyzers

RESULTS FROM A SIMPLE INFRARED CLOUD DETECTOR


Current and Future Methods for Calibration of Black Carbon Real-Time Mass Instruments

SCADA and Monitoring for Solar Energy Plant

NSW Air Quality Monitoring Network: Quality Assurance Procedures

DOE-HDBK Module 2.18 Air Sampling Equipment. Instructor s Guide

REMOTE TELEMETRY UNITS FOR CONTINUOUS WATER QUALITY MONITORING: AN EXAMPLE FROM SILVERSTREAM LANDFILL Mike Millar, Geotechnics Ltd ABSTRACT

2 Absorbing Solar Energy

Instruction Manual Service Program ULTRA-PROG-IR

Grant Agreement No SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS

Overview of the IR channels and their applications

Application Note: Absorbance

Non-invasive CLAMP-ON ultrasonic flowmeter for liquids

FAAST. Fire Alarm Aspiration Sensing Technology

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

Hands on an infrared gas analyzer (IRGA) to measure photosynthesis

CHAPTER 6 INSTRUMENTATION AND MEASUREMENTS 6.1 MEASUREMENTS

A Demonstration Plant of a Liquid Desiccant Air Conditioning Unit for Drying Applications

SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL

Environment-Laboratory Ambient Conditions

INFLUENCE OF MEASUREMENT SYSTEM QUALITY ON THE EVALUATION OF PROCESS CAPABILITY INDICES

RADIATION PROTECTION. Alpha-Beta Particulates Monitor LB 150 D-R. ABPD Particulates Monitor combined with Alpha Energy Range Discrimination (AERD)

Department of Engineering Enzo Ferrari University of Modena and Reggio Emilia

GP2Y0A02YK0F. Distance Measuring Sensor Unit Measuring distance: 20 to 150 cm Analog output type GP2Y0A02YK0F

Coarse PM Methods Evaluation Study

How To Calculate Global Radiation At Jos

National Environment Protection (Ambient Air Quality) Measure

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION

Labkotec SET. Labkotec separator alarm system protecting our environment. Measure your success

Results of Airborne Monitoring by the Ministry of Education, Culture, Sports, Science and Technology and the U.S. Department of Energy

Gravimetric determination of pipette errors

AN ESTIMATION OF A PASSIVE INFRA-RED SENSOR'S PROBABILITY OF DETECTION

CALIBRATION PRINCIPLES

Wireless Data Logger Monitoring Systems

Ambient Air Monitoring at Deployment Locations in the Middle East

QFP910. Water detector. Building Technologies. CE1N2732en. Synco living

Experiment 10. Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado

Certification system of air filter quality (CAF mark)

Measurement of RF Emissions from a Final Coat Electronics Corrosion Module

Continuous, remote monitoring and logging of electromagnetic fields

REAL-TIME DUST AND AEROSOL MONITORING

Laboratory Manual for Air Quality Sampling and Analysis

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA

MODEL KS /2,5/1

Iophic Smoke Detection Technology. An Introduction. Universal Security Instruments. Over 40 Years of Residential & Commercial Safety Innovations

Copyright PEOPLECERT Int. Ltd and IASSC

Standard Hydraulic Piping Flushing Procedure

Index-Velocity Rating Development (Calibration) for H-ADCP Real-Time Discharge Monitoring in Open Channels

The prediction of flue gas emissions from the combustion in the industrial tubular heaters

Turnkey Solutions for Radiation Protection in PET Laboratories R A D I A T I O N P R O T E C T I O N

DETERMINATION OF THE HEAT STORAGE CAPACITY OF PCM AND PCM-OBJECTS AS A FUNCTION OF TEMPERATURE. E. Günther, S. Hiebler, H. Mehling

Gas Smart Metering System

High Volume PM10/TSP Calibration/Verification/Audit Guidance. Issued: 12/15/14 - FINAL

LLB-OCMFM 2000 PORTABLE DOPPLER AREA-VELOCITY FLOWMETER

PERFORMANCE EVALUATION OF SOLAR ENERGY DEVICES BY USING A DATA LOGGING SYSTEM

Monitoring of air pollution spread on the car-free day in the city of Veszprém

FIRE ALARM SYSTEM. reliable fire detection and alarm signalling system

How To Understand Algebraic Equations

Principle of Thermal Imaging

TR SUPPLEMENTARY REQUIREMENTS FOR THE ACCREDITATION OF CONTINUOUS AMBIENT AIR QUALITY MONITORING STATIONS

Impacts of large-scale solar and wind power production on the balance of the Swedish power system

- 1 - BISC 367 Plant Physiology Laboratory SFU

Online Infrared Flue Gas Analyzer Gasboard 3000

Indian Research Journal of Extension Education Special Issue (Volume I), January,

RADIATION MONITORING NETWORK OF THE CZECH REPUBLIC

HEATED INLET SAMPLE CONDITIONING EFFECTS ON REMOTE DUST MONITORING

Wireless power meter monitoring with power theft detection and intimation system using GSM and Zigbee networks

MEASUREMENT DATA ACQUISITION SYSTEM UTILISING THE GSM PHONE SYSTEM

Wireless Soil Moisture and Weather Monitoring System Caipos

Transcription:

NUKLEONIKA 2008;53(Supplement 2):S37 S42 ORIGINAL PAPER Metrological features of a beta absorption particulate air monitor operating with wireless communication system Adrian Jakowiuk, Piotr Urbański, Edward Świstowski, Bronisław Machaj, Jan Pieńkos, Ewa Kowalska Abstract. A system for measurements of particulate matter in the ambient air, employing beta absorption instrument AMIZ-2007 with wireless communication facilities based on GPRS technology, is presented. Uncertainty of measurements caused by the counting statistics was analyzed and it was found that at least 3 h of sampling time is needed to achieve coefficient of variation lower than 20% for the average concentration of particulate matter in the ambient air exceeding 10 μg/m 3. Application of a C-14 beta ray source instead of Pm-147 improves sensitivity of the measurement ca. two times. Some results of on-line operation of the system with PM10 and PM2.5 samplers are also shown. Key words: ambient air beta-ray absorption particulate matter wireless communication Introduction A. Jakowiuk, P. Urbański, E. Świstowski, B. Machaj, J. Pieńkos, E. Kowalska Department of Radioisotope Instruments and Methods, Institute of Nuclear Chemistry and Technology, 16 Dorodna Str., 03-195 Warsaw, Poland, Tel.: +48 22 5041125, Fax: +48 22 8111917, E-mail: a_jakowiuk@ichtj.waw.pl Received: 3 November 2008 Accepted: 24 November 2008 Metrological features of a beta absorption particulate air monitor [2] were already considered in the past by some authors [4, 8]. However, recent instruments are totally different from those produced in the past. Implementation of new informatics and communication technologies caused that the modern instruments for particulate monitoring are fully automated and wirelessly controlled devices, operating in large networks, and the results of measurements are instantly presented on the web sites [3, 5, 6]. Moreover, new requirements from the end users appeared. It is expected now that these instruments have to be capable of measuring not only the overall dust suspended in the ambient air, but also its fractions of grain size (aerodynamic diameter) below 10 μm (with PM 10 sampling head [7]) or even 2.5 μm (with PM2.5 probe). Thus, more detailed analysis of the factors influencing uncertainty of the measurement, especially at low concentration of the particulate matter in the air, seems to be necessary. In this paper, a system for the measurement of particulate matter in the ambient air, employing beta absorption instrument AMIZ-2007 with wireless communication facilities based on GPRS technology, is presented. Also analysis of the measurement uncertainty

S38 is carried out and examples of the application of the whole system are shown. A. Jakowiuk et al. An AMIZ-2007 system is designed for automatic measurements of airborne dust concentration in the ambient air. It is also equipped with meteorological sensors measuring temperature, atmospheric pressure and relative humidity of the ambient air. Optionally, the monitor can be fitted with wind speed and wind direction sensors. The principle of the operation is based on the measurements of mass of dust deposited on a fiber glass filter tape, from the known volume of the air. The air volume is determined by pumping period (sampling time), while the air flow is kept constant. The mass of the collected dust is measured basing on absorption of beta rays. Operation of the instrument is controlled by a microprocessor system, which apart from the controlling measuring cycle, processes measured signal, computes dust concentrations and stores the results in a local memory. A block diagram of the AMIZ-2007 monitor is shown in Fig. 1. In case the AMIZ-2007 operates as an independent individual unit (not in monitoring network), then it is programmed with a local keyboard, and the stored results can be transmitted to an external computer through a USB port. In case the instrument operates in a monitoring network, then the transmission of results and the remote communication with an external computer is realized in a wireless manner by means of the mobile phone network GSM and the internet. Airborne dust monitoring system AMIZ-2007 Calibration models The principle of operation of the radiometric particulate monitors is based on the measurements of the mass per area of the dust collected on the filter band using the absorption of beta rays. Since the mass per area of the clean filter is not homogeneous and varies ca. from 6.5 to 8 mg/cm 2, it is necessary to perform two measurements of the count numbers, one for the clean filter (N 0 ) and the second for the filter with the collected dust (N 1 ). The mass per area of the collected dust may be found from the formula (1): 1 N0 (1) m = ln k N 1 where k is a coefficient, whose numeric value for the logarithmic model (1) may be close to beta-ray absorption coefficient [2]. However, due to the fact, that mass of the collected dust is much smaller than the mass of a clean filter, it may be assumed that the difference between the measured count number N 0 N 1 can also be proportional to m, so the linear calibration model can be proposed: 1 (2) m = ( N 0 N 1 ) k 1 where k 1 is a proportionality coefficient. Relation between the filter mass m and the measured differences in the collected count number N 0 N 1 for the linear model (2) and for two radioisotope sources Pm-147 and C-14 is shown in Fig. 2. Correlation coefficient R was about 99.86% for the linear and 99.70% for logarithmic model, so it can be assumed that the metrological features of both the models are almost the same. The results presented in Fig. 2 show that sensitivity of the measurement, represented by the value of the coefficient k l, is about two times better for beta rays from C-14 than from the Pm-147 radioactive source. Generally, mass of the deposited particulate matter is within 100 to 300 μg/cm 2. Fig. 1. Block diagram of AMIZ-2007 system. 1 air pump with temperature sensor; 2 motor for filter movement; 3 beta-ray source; 4 measuring head; 5 air inlet; 6 air filter tape; 7 GM beta-ray detector. Fig. 2. Relation between the mass per area deposited on the filter vs. number of accumulated count difference for two beta-ray sources C-14 (E β = 156 kev) and Pm-147 (E β = 224 kev).

Metrological features of a beta absorption particulate air monitor operating with wireless... S39 Uncertainty of measurements exp( km) N1 (7) wlog = s kzavtq Relative uncertainty w log for the logarithmic model was computed using an approximate formula for variance of the count ratio derived in [3]. Having numerical values of the instrument parameters, one can compute the value of relative uncertainty w as a function of sampling time T and for various concentration of the particulate matter in air Z av (Fig. 3). The computations were performed for one particular instrument fitted with the samplers PM10 or PM2.5 and the C-14 or Pm-147 radioactive sources. The flow rate q was 1 m 3 /h and the active measurement area about 1 cm 2. It appeared that the computed value w does not depend on the applied calibration model; almost the same value was obtained for the logarithmic as well as linear model. However, the counting uncertainty w is strongly influenced by the energy of used beta radiation. For the C-14 source, emitting beta rays of maximal energy 156 kev, (halving mass m 1/2 = 2.7 mg/cm 2 ) [7] the uncertainty is almost two times lower than that for beta rays from Pm-147 of maximal energy 224 kev and halving mass m 1/2 = 4.5 mg/cm 2. Taking also into account that the radioactive half-life for Pm-147 is 2.6 y, whereas for C-14 it is 5730 y, the superiority of C-14 over Pm-147 is evident. The chart of uncertainty computed individually for every manufactured gauge can help the end users of the beta absorption particulate monitors in proper setting instrument parameters. It should be mentioned that the considered here uncertainty of measurements pertains only counting statistics and other factors influencing its value were not taken into account in computations. To examine repeatability of the AMIZ-2007, two instruments were installed in the same place and par- The mass per area of the dust m deposited on the filter can be expressed as: ZavTq (3) m = s T where: Zav = 1 T z( t) dt mean value of the dust concentration in air [μg/m 3 ] over the sampling period T [h]; z(t) instant value of the dust concentration; q air flow rate [m 3 /h]; s active surface area of the filter [cm 2 ]. Coefficient of variation of the mass per area of the deposited on the filter σ m /m for the linear and logarithmic models are, respectively to equal: σ σ m ( N0 N1) (4) = s m Z Tq (5) where: σ 2 (N0 N1) variance of the count difference; σ 2 ln(n0/n1) variance of the logarithm of count ratio. Since the mass of deposited dust is proportional to the measured dust concentration, its variation coefficient, which can be considered as a measure of relative uncertainty w = (σ Z av/z av ) = σ m /m, can be easily computed for both the models: (6) 0 σ av σ m ln( N0/ N1) = m ZavTq w N + N s 0 1 lin = kz l avtq s Fig. 3. Calculated relation between the relative uncertainty of measurement (coefficient of variation) vs. sampling time T for various average concentrations of the particulate matter, (a) with Pm-147, (b) with C-14.

S40 Fig. 4. Particulate matter concentration Z av measured with two separate instruments fitted with PM10 samplers and two different beta-ray sources: C-14 and Pm-147 (1 14 March 2008, to the north of Warsaw). A. Jakowiuk et al. in Fig. 4. It can be seen that the mean difference is about 0.4 μg/m 3 which is very close to zero and the standard deviation of differences is about 6 μg/m 3. This means, that 90% of results obtained in the above experiment differs less than ± 12 μg/m 3. The sampling time was 3 h and was chosen taking into account expected uncertainty of measurements (Fig. 3). Similar results were obtained for the PM2.5 sampler. A similar test was carried out for the two instruments with the same beta-ray source, but with the two different samplers: PM2.5 and PM10. Results of that test for 67 successive measurements for a sampling time of 3 h are shown in Fig. 5. Generally, the results obtained using the PM10 sampler are higher than those with PM2.5, but in a few cases lower values are observed. It seems that such an effect can be explained by the relatively high uncertainty of measurements at low concentration of the particulate matter. In the considered test, the mean value of differences between results obtained for both the samplers was about 5.8 μg/m 3 and the standard deviation about 6 μg/m 3. Operation in monitoring network Fig. 5. Particulate matter concentration measured with two separate instruments fitted with PM10 and PM2.5 samplers and with the same beta-ray source Pm-147 (19 27 February 2008, to the north of Warsaw). allel measurements were carried out. Obtained results of mean concentrations of the particulate matter (Z av 1, Z av 2) measured with both the instruments in Warsaw, and the difference between them (Z av 1 Z av 2) are plotted The network operates under control of software installed on the main computer of the monitoring center. Data transmission and communication with the measuring station is carried out in the two configurations: Communication network uses only the GSM system, Communication network uses both the GSM and internet system (Fig. 6). In the first configuration, all information is wirelessly transmitted only to the main computer, whereas in the second the direct communication with the internet is also provided. In this case, all facilities offered by the internet can be employed. The results obtained from Fig. 6. Diagram of the wireless communication network.

Metrological features of a beta absorption particulate air monitor operating with wireless... S41 Fig. 7. Results of medium term measurements of the particulate matter with two separate instruments fitted with different samplers PM10 and PM2.5. Sampling time 3 h. The horizontal line shows the permissible level (50 μg/m 3 ) of daily average particulate matter concentration. Fig. 8. Results of long term (3 months) measurements of the particulate matter (averaged for every day) with two separate instruments fitted with different samplers PM10 and PM2.5. Sampling time 3 h. The horizontal line shows the permissible level (50 μg/m 3 ) of daily average particulate matter concentration. the particulate monitor can be stored in a web page available only for the authorized users. All results: concentration of the particulate matter as well as meteorological data (temperature, pressure humidity, wind speed and direction) are stored in a file (in the text, Word and Excel format). There is also other information to be available, such as operation status, measurement mode, air pump temperature and others. Some of these parameters can be remotely changed from the monitoring centre, and others, as. e.g. faulty operation of the measuring system can be sent back from the place of installation of the gauge to the central computer and/or to the chosen mobile phone. To demonstrate operation of the monitoring system, two examples of the obtained results from AMIZ-2007 are presented in Figs. 7 and 8. Two monitors, one with the PM10 sampler, the second with the PM 2.5 were installed in an air conditioned kiosk at Otwock, a town about 30 km to the south of Warsaw, where a limit value of PM10 is exceeded [1]. The sampling time 3 h, Y scale automatically adjusted to measured dust concentration. The results presented in Fig. 7 were obtained during a relatively short term of 22 days (from 10 April to 1 May 2008). In Fig. 8, there are shown the daily averaged data for a period of 99 days (from 10 April to 19 July 2008).

S42 Beside the particulate concentration, meteorological data are also available on the same web site. The system was installed in a few places where it operates under control of the institution responsible for environmental protection. Conclusions A complex system for the remote wireless monitoring of the particulate matter concentration in the ambient air was developed. The concentration monitor is operating on the beta-ray absorption principle and can be used either with the PM10 as well as with the PM2.5 sampler. Results of particulate matter concentration obtained with the PM10 sampler are generally higher than those with the PM2.5. However, in some cases, especially at low concentration of particulate mater the opposite situation can happen. This can be explained by a relatively high uncertainty of measurements occurring at low concentration of the measured particulate matter. Analysis of the uncertainty of measurements caused by the counting statistics shows that at least a 3 h sampling time is needed to achieve a coefficient of variation lower than 20% for beta rays from Pm-147 and 10% when the C-14 beta source is applied. Derived formulae allow to asses duration of the sampling time for various values of the particulate mass in the air to get the required relative uncertainty of measurements. Application of the C-14 beta ray source instead of the Pm-147 source improves almost two times the sensitivity of the measurement. Generally, it seems, that the developed wireless monitoring system can deliver a huge amount of important data, which can be useful for the environmental A. Jakowiuk et al. protection services to find and identify the origin and sources of ambient air pollution. Acknowledgment. This study was supported by the R&D Grant no. R 14 007 02 from the Polish Ministry of Science and Higher Education. References 1. Directive 2008/50/EC of the European Parliament and of the Council of 21 May 2008 on ambient air quality and cleaner air for Europe 2. ISO (2000) International Standard. Ambient air measurements of the mass of particulate matter on a filter medium beta-ray absorption method. ISO 10473:2000(E) 3. Johansen GA, Jackson P (2007) Radioisotope gauges for industrial process measurements, John Wiley & Sons, Ltd, Chichester 4. Naimpour MA, Rahimi H, Amini AA, Machaj B (1993) Measurements of airborne dust concentration in Tehran employing beta absorption. Nukleonika 32;2:67 78 5. PN-EN 12341:2006. Air quality Determination of the PM 10 fraction of suspended particulate matter Reference method and field test procedure to demonstrate reference equivalence of measurement methods 6. PN-EN 14907:2006. Ambient air quality Standard gravimetric measurement method for the determination of the PM2.5 mass fraction of suspended particulate matter 7. Putman JL (1969) Radioisotope techniques in process control and analysis. In: Non-destructive testing. Harwell Post-Graduate Series, Oxford University Press, pp 126 131 8. Urbański P, Machaj B, Harasimczuk J, Strzałkowski J (1992) Optimization of performance of airborne dust monitors. Nukleonika 37;3:55 65