Instruction Manual Service Program ULTRA-PROG-IR
|
|
|
- Sharleen Whitney Goodwin
- 10 years ago
- Views:
Transcription
1 Instruction Manual Service Program ULTRA-PROG-IR Parameterizing Software for Ultrasonic Sensors with Infrared Interface
2
3 Contents 1 Installation of the Software ULTRA-PROG-IR System Requirements Installation First Installation Installing an Update Uninstalling Specific Features of the Software ULTRA-PROG-IR Displaying Measured Values and Diagnostics in Real Time Chart Recorder and Data Logger Settings at the Ultrasonic Sensor with the Software ULTRA-PROG-IR Settings for Switching Outputs Operating mode of the switching output Switching points Switching hystereses Switching characteristics Delay times Settings for Analog Output Operating mode of the analog output Characteristic Analog range Behavior outside the analog range Settings for Sensing Range Close range Sensing range Blind zones Settings for Control Elements Adjustability with the potentiometer Adjustment range of the potentiometer Teach-In Synchronization and multiplex mode Settings at Evaluation Algorithm Operating mode Temperature compensation Sensitivity Sound cone width Cycle time Evaluation method Average value Filtering measured values Legal Notice... 13
4 1 Installation of the Software ULTRA-PROG-IR 1.1 System Requirements The software ULTRA-PROG-IR is executable on every PC or notebook. For the communication with an IR-Interface UC-18/30GM-IR, a free USB2.0 port is required. For the installation of the software, a free memory of about 5 MB on a storage device is required. The software may be deployed with the following operating systems: Windows XP Service Pack 3 Windows Vista Business Windows 7 (32 bit versions) 1.2 Installation First Installation You can download the software ULTRA-PROG-IR from the Pepperl+Fuchs Internet portal. For installation, start the file setup.exe. Follow the instructions during setup. The installation path is preset to C:\Program Files\ULTRA-PROG-IR. You can change the path during installation. After setup is completed successfully, the software is ready to use Installing an Update Updates are provided at the Pepperl+Fuchs Internet portal. Uninstalling older versions is not necessary before installing an update. Older versions are replaced by the update and are no longer available after the update installation Uninstalling You can use Windows System Control for uninstalling the software ULTRA-PROG-IR. 2 Specific Features of the Software ULTRA-PROG-IR 2.1 Displaying Measured Values and Diagnostics in Real Time Every reading access on the parameters and process data of the sensor is effected in real time. "Real time" refers to the cyclical transmission of the sensor data in line with the measuring cycle of the sensor. Thus, reading accesses to the sensor parameters and its process data do not interrupt the measuring operation of the sensor. The communication via infrared is effected during the breaks of the measuring cycle. A writing access on the sensor can extend the measuring cycle as the changed parameter values have to be saved in the ROM of the sensor. This cyclical reading of measured values in real time allows an optimum diagnostics for adapting the sensor to the application. In addition to the object distance, for example, the signal strength of the sound reflection at the object is transmitted as well. With this information, the object to be detected can be arranged optimally in the application. The following details of the process data are available: Object distance Signal strength of the sound reflection at the object Internal temperature for compensation of the speed of sound Noise level of the internal circuitry for a qualitative evaluation of the application environment
5 2.2 Chart Recorder and Data Logger The chart recorder can record object distances with the associated signal strengths of the sound reflection at the object over time. The recording speed is adjustable in five levels. For measuring cycle times of more than 5 milliseconds, the data logger continuously records measured values without any loss of measuring cycles. The measured values are first filed in the RAM of the PC. After completion of the recording phase, the user can assign a file name and path for the filing on a storage medium. The recording format is CSV (Comma-Separated Values) and can be read and processed with, for example, Microsoft Excel. The expected file size can be calculated using the following formula:
6 file size = size of data packet * recording duration / measuring cycle time + 82 bytes Overview about data packet sizes: Minimal size of a Echo telegram data packet MEASURED VALUE Typical size of a data packet 13 bytes 17 bytes 20 bytes ECHO1 26 bytes 32 bytes 38 bytes ECHO3 38 bytes 51 bytes 70 bytes ECHO5 50 bytes 70 bytes 90 bytes Maximum size of a data packet Example: Recording of the MEASURED VALUE telegram of a UC500 with a measuring cycle time of 14 ms over a duration of one hour: expected file size (typical) = 17 bytes * 3,600,000 ms / 14 ms + 82 bytes = 4,371,511 bytes = 4.17 MB 3 Settings at the Ultrasonic Sensor with the Software ULTRA-PROG-IR 3.1 Settings for Switching Outputs Depending on the connected type of sensor, different tabs are arranged on the user interface of the software ULTRA-PROG-IR. The tab 'Switching Output 1' or 'Switching Output 2' is only available for sensor types with one or two switching outputs and allows settings for the switching outputs of the sensor. The following settings can be made in the tab 'Switching Output 1' or 'Switching Output 2': Operating mode of the switching output Switching output inactive: In this mode, the switching output is inactive, i.e. no switching signal is emitted. 1 switching point: There is one switching point available. The switching point can be set in the whole sensing range. This operating mode is particularly suitable for blanking the background. 2 switching points: There are two switching points available. Both switching points together make up a switching window. The switching window can be set in the whole sensing range. Switching point 1 must be set smaller than switching point 2. If switching point 1 is set higher than switching point 2, the LED associated with the switching output flashes with a frequency of about 5 Hz. In this mode, the switching characteristics are not defined. For the ultrasonic diffuse barrier sensor mode, always two switching points are required. Alarm output: The alarm output does not report the presence of an object, but reports if the signal-to-noise ratio of the echo signal of an object falls below a critical value. Thus, the alarm output sends a warning shortly before the sensor loses the echo signal of an object. The critical value, named alarm amplitude, can be set in the range from %. The setting for the alarm output is only available for switching output 2.
7 3.1.2 Switching points For switching point 1 and switching point 2, a value can be set in millimeter respectively if the respective switching point is active. Switching point 1 must be set smaller than switching point 2. If a switching point can be set with the potentiometer, it cannot be changed with the software ULTRA-PROG-IR Switching hystereses For each active switching point, a corresponding switching hysteresis in millimeter can be set. Make sure that the switching point with its corresponding switching hysteresis does not exceed the sensing range. Furthermore, there are four different types of switching hystereses: Far switching hysteresis: The switching hysteresis is effective if an object moves from a near distance to a far distance. Near switching hysteresis: The switching hysteresis is effective if an object moves from a far distance to a near distance. Symmetrical switching hysteresis: Half of the switching hysteresis is effective if an object moves from a far distance to a near distance. The second half of the switching hysteresis is effective if an object moves from a near distance to a far distance. No switching hysteresis: The switching hysteresis is not effective Switching characteristics Two different switching characteristics are available that correspond to the characteristics of a switch with NO contact or NC contact functionality. If the function NO contact is activated, the switch closes if an object is located in the switching range, i.e. the switching output is switched to supply voltage. If no object is located in the switching range, the switch opens, i.e. no switching signal is output at the switching output. The switching characteristics are exactly vice versa if the function NC contact is activated Delay times It is possible to delay switching on or off the output. Both delay times can be extended by an integer number of measuring cycles. The delay times can be set in the range from measuring cycles.
8 3.2 Settings for Analog Output Depending on the connected type of sensor, different tabs are arranged on the user interface of the software ULTRA-PROG-IR. The tab 'Analog Output' is only available for sensor types with an analog output ma, ma or V and allows settings for the analog output of the sensor. The following settings can be made in the tab 'Analog Output': Operating mode of the analog output Current output ma: The distance of an object in the analog range is output at the analog output in the range ma. This setting is only available for sensors with analog outputs ma or ma. Current output ma: The distance of an object in the analog range is output at the analog output in the range ma. This setting is only available for sensors with analog outputs ma or ma. Voltage output: The distance of an object in the analog range is output at the analog output in the range V. This setting is only available for sensors with analog output V. Analog output inactive: In this mode, the analog output is inactive, i.e. the output is constantly 0 ma or 0 V Characteristic It is possible to set a rising or a falling characteristic. In the case of a rising characteristic, the current or the voltage at the analog output rises with increasing object distances. In the case of a falling characteristic, the current or voltage at the analog output falls with increasing object distances Analog range In the set analog range, the current or voltage of the analog output is output depending on the distance. The start value of the analog range must be smaller than the end value. The analog range can be located anywhere within the sensing range Behavior outside the analog range The behavior of the analog output outside the analog range can be set for the following three modes:
9 No object detected: The behavior of the analog output if no object is detected. Object in front of the analog range: The behavior of the analog output if an object is located in front of the analog range from the sensor's perspective. Object behind the analog range: The behavior of the analog output if an object is located behind the analog range from the sensor's perspective. For every mode, you can define if the maximum or the minimum analog value shall be output at the analog output. The minimal analog value corresponds to 0 ma for sensors with analog output ma, 4 ma for sensors with analog output ma, or 0 V for sensors with analog output V. The maximal analog value corresponds to 20 ma for sensors with analog outputs ma or ma, or 10 V for sensors with analog output V. 3.3 Settings for Sensing Range Depending on the connected type of sensor, different tabs are arranged on the user interface of the software ULTRA-PROG-IR. The tab 'Sensing Range' is available for all sensor types and allows settings for the sensing range of the sensor. The following settings can be made in the tab 'Sensing Range': Close range For direct detection sensors, the close range is caused by the switching of the sensor from transmitting to receiving mode. After the ultrasound has been sent, the ultrasound converter has to die away before ultrasonic echos can be detected. In this time or in this range, objects can be detected only to a limited extent. The size of the close range is defined physically. The indication of the close range is provided for information only and cannot be changed by the user Sensing range The sensing range is the range where objects can be detected. The sensing range starts at the end of the close range and ends at the set value in millimeter. All objects that are located outside the set sensing range are logically blanked out. Only the technical service staff can set the sensing range. The end of the sensing range can be set in the range between the end of the close range and 10,100 mm. However, not all sensors are physically capable of detecting objects up to the maximum possible sensing range. Extending the sensing range usually increases also the cycle time and thus the response time of the sensor Blind zones
10 A blind zone is a range where no objects can be detected. In this range, the sensor is "blind". If an object is located in the blind zone, the sensor acts as if no object is detectable in the sensing range. The blind zones can be defined by the user at any desired range in the close and sensing range. You can set up to two independent blind zones. Blind zones receive a red mark on the software interface. The use of blind zones is particularly suitable for areas where no useful objects occur, but in which disturbing reflections occur due to their mounting. 3.4 Settings for Control Elements Depending on the connected type of sensor, different tabs are arranged on the user interface of the software ULTRA-PROG-IR. The tab 'Adjustment Type' is available for every type of sensor and allows settings for the control elements of the sensor. The following settings can be made in the tab 'Adjustment Type': Adjustability with the potentiometer For potentiometer 1 and potentiometer 2, a range limit can be set that is adjustable with the respective potentiometer. It is also possible to deactivate one or both potentiometers. Depending on the type of sensor, the following range limits can be selected: Switching point 1 of switching output 1 Switching point 2 of switching output 1 Switching point 1 of switching output 2 Switching point 2 of switching output 2 Start of analog range End of analog range Adjustment range of the potentiometer The adjustment range is the range where the selected range limit can be adjusted with the potentiometer. The start of the adjustment range corresponds to the selected range limit at left stop of the potentiometer. The end of the setting range corresponds to the selected range limit at right stop of the potentiometer. The adjustment range always refers to both potentiometers if they are active Teach-In It is possible to use the XI connection of the sensor to teach a range limit. Therefore, an object must be positioned in the desired distance. The object must be located within the sensing range of the sensor. Then, the XI connection is connected to the GND connection of the sensor. The connection must be stable for at least three seconds. After completion of the teach-in process, the green LED lights up for approx. 0.5 seconds. After that the connection of the XI connection can be separated or the teach-in process is repeated after every three seconds.
11 If during the teach-in process no object is located within the sensing range or the object is not detected, no range limit is taught. In this case, the sensor ignores the teach-in signal. Depending on the type of sensor, the following range limits can be taught: Switching point 1 of switching output 1 Switching point 2 of switching output 1 Switching point 1 of switching output 2 Switching point 2 of switching output 2 Start of analog range End of analog range The teach-in function is not activated when the sensor is delivered Synchronization and multiplex mode If the XI connections of two or up to ten sensors are connected, the sensors synchronize. In synchronized mode, all sensors have the same cycle time and send the ultrasonic signals at the same time. For sensors with different cycle times, the common cycle time in synchronized mode is determined by the sensor with the longest cycle time. For sensors with a shorter cycle time than the cycle time in synchronized mode, the initial state outside the actual cycle time is saved. It is possible to delete the initial state during the extended cycle time. In multiplex mode, the connected sensors work with a synchronous cycle time, too, however, the ultrasonic signal is sent alternately. Therefore, an address must be assigned to each sensor and the overall number of sensors must be set. The addresses of the sensors at the same time define the order in which the sensors send the ultrasonic signals. In multiplex mode, the XI connections of the sensors must be connected and then energized altogether at the same time. While the XI connection is set to teach-in, the synchronization and the multiplex mode are both automatically inactive. 3.5 Settings at Evaluation Algorithm Depending on the connected type of sensor, different tabs are arranged on the user interface of the software ULTRA-PROG-IR. The tab "Evaluation" is available for every sensor type and allows settings at the evaluation algorithm of the sensor. The following settings can be made in the tab "Evaluation": Operating mode Sensor locked: In this mode, the sensor does not send any ultrasonic pulses. The object distance that was last measured is saved as long as the sensor is energized. The sensor remains ready for operation, i.e. further settings with the potentiometer or the software ULTRA-PROG-IR can be made. Direct detection sensor: In the operating mode 'Direct detection sensor', the object serves as sound reflector. Thus, the sensing range depends on the reflectivity of the object, i.e. on the condition of the surface and the angle of attack.
12 The presence of an object in the sensing range is displayed by means of a binary signal at the switching output or an analog distance signal at an analog output. Ultrasonic diffuse barrier sensor: In this operating mode, the ultrasound is reflected at a fixed object. This object serves as reference object and is detected by the sensor as long as there is no other object between the sensor and the reference object. Objects that move between the sensor and the reference object, are detected due to the change in the measured distance or due to the missing signal from the reference object. Using this procedure, objects with minor ultrasound reflectivity can be securely detected. Thru-beam sensor: In the operating mode 'Thru-beam sensor', the sensor serves either as transmitter or receiver. Thus, for thrubeam mode at least two sensors are required to create an ultrasonic thru-beam sensor. Transmitter and receiver are mounted opposite each other. To ensure a correct thru-beam mode, transmitter and receiver have to be synchronized, i.e. the XI connections of both sensors have to be connected with each other. If an object impedes the sound propagation along the ultrasound signal path this causes a switching at the receiver. With this procedure, longer distances can be reached compared to the reflectivity principle, because the ultrasound only passes the signal path once. Furthermore, the response behavior decreases considerably as switching between transmitting and receiving mode is not necessary any more Temperature compensation If the temperature compensation is on, the indicated value in the datasheet concerning measurement accuracy applies for the whole temperature range. If the temperature compensation is off, a measurement deviation applies, based on +20 C. With increasing temperatures, a measurement deviation of %/K applies, with decreasing temperatures, the measurement deviation is %/K Sensitivity Using this value given in percent, you can change the sensitivity of the ultrasonic receiver. The smaller the sensitivity is set, the slimmer and shorter the sound cone of the sensor becomes. The sensitivity may be set in the range % Sound cone width This value given in percent indicates the sound cone width of the sensor. The smaller the sound cone width is set, the slimmer the sound cone of the sensor becomes. At settings from %, the maximum effective operating range remains constant. At settings from %, the sound cone becomes both slimmer and shorter Cycle time Here, the cycle time of the sensor is given in milliseconds. The cycle time means the repeating time of the ultrasonic pulse, i.e. after the cycle time has passed, the ultrasound is sent again. The setting of the cycle time is reserved to the technical service staff. Settings between ms are possible. The full range, however, is not available for all sensor types Evaluation method
13 First echo: This evaluation method uses the first echo for evaluation, i.e. the object that is closest to the sensor and reflects ultrasound to the sensor is detected and used for the operation of the outputs. This method is the usual evaluation method. Maximum amplitude: In this method, the echo that has the strongest signal strength concerning the reflected ultrasound is used for the evaluation. Objects with a minor sound reflectivity or objects that are located at the margin of the sound cone are blanked out. This evaluation method is particularly suitable for applications of level measurement as liquids usually feature a very good sound reflectivity. Disturbing edges at the container or reflection of an impeller can be blanked out with this evaluation method Average value Using this setting, past measured values may be rated with a factor. The new measured value respectively gets the factor 1. The higher the average value is set, the more inert is the reaction of the current measured value and thus of the sensor outputs on changes. With this method, variations in the measured values can be easily compensated. Settings ranging from are valid. Averaging is active only for object movements within the sensing range. If an objects moves into the sensing range and if before that, no other object was within the sensing range, the averaging is inactive. In this case, the response delay applies. The same applies if an object moves out of the sensing range and if no other object is located within the sensing range Filtering measured values Filtering measured values for three measuring cycles serves the suppression of noise. Thus, sporadic noise can be filtered if they do not occur longer than two measuring cycles. If the filtering of measured values is switched off, lower response times can be reached. If the sensor is deployed in an environment with possible electromagnetic interferences, the filtering of measured values should not be deactivated. 4 Legal Notice Pepperl+Fuchs GmbH Lilienthalstraße Mannheim Germany Tel: Fax: [email protected] Web:
Technical data. General specifications. Indicators/operating means. 30 Hz Multiplex operation 30 Hz / n, n = number of sensors, n 5
Model Number Single head system Features Parameterization interface for the application-specific adjustment of the sensor setting via the service program ULTRA 000 programmable switch outputs Hysteresis
Capacitive Proximity Sensors Theory of Operation
Capacitive Proximity Sensors Theory of Operation Capacitive proximity sensors are similar to inductive proximity sensors. The main difference between the two types is that capacitive proximity sensors
Automation System TROVIS 6400 TROVIS 6493 Compact Controller
Automation System TROVIS 6400 TROVIS 6493 Compact Controller For panel mounting (front frame 48 x 96 mm/1.89 x 3.78 inch) Application Digital controller to automate industrial and process plants for general
Technical data. General specifications. Indicators/operating means. Electrical specifications Operating voltage U B Power consumption P 0 Interface
Release date: 06-0- 09: Date of issue: 06-0- 009_eng.xml Model Number Single head system Features Parameterization interface for the application-specific adjustment of the sensor setting via the service
crm+25/d/tc/e Extract from our online catalogue: Current to: 2015-10-26
Extract from our online catalogue: crm+25/d/tc/e Current to: 2015-10-26 microsonic GmbH / Phoenixseestraße 7 / 44263 Dortmund / Germany / T +49 231 975151-0 / F +49 231 975151-51 / E [email protected]
Transmitter Interface Program
Transmitter Interface Program Operational Manual Version 3.0.4 1 Overview The transmitter interface software allows you to adjust configuration settings of your Max solid state transmitters. The following
Inwall 4 Input / 4 Output Module
Inwall 4 Input / 4 Output Module IO44C02KNX Product Handbook Product: Inwall 4 Input / 4 Output Module Order Code: IO44C02KNX 1/27 INDEX 1. General Introduction... 3 2. Technical data... 3 2.1 Wiring Diagram...
Logging of RF Power Measurements
Logging of RF Power Measurements By Orwill Hawkins Logging of measurement data is critical for effective trend, drift and Exploring the use of RF event analysis of various processes. For RF power measurements,
Movement/monitoring 1307/1.0 (6325xx)
/monitoring 1307/1.0 (6325xx) General In the following the device will be referred to as the movement detector. The movement block will only switch off when there is no more movement in front of the device
How To Use A Power Supply On A Powerline 2.2 (Ai)
KNX/EIB Product documentation Issue: 05.08.2010 629x1220 Push button sensor 3 comfort 1-gang Push button sensor 3 comfort 2-gang (1+1) Push button sensor 3 comfort 3-gang Push button sensor 3 comfort 4-gang
Tubular Analog model resolution: 0,5 mm (SLOW mode), 1 mm (FAST mode) Right angle Repeatibility: 0,7 mm Vdc
Ultrasonic sensors for high precision detection of clear and transparent objects Clear object detection, inspection on transparent or highly reflective film and liquid level measurement Standard M18 tubular
Various Technics of Liquids and Solids Level Measurements. (Part 3)
(Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,
How To Make An Ultrasonic Sensor
crm+ ultrasonic sensors Microsonic : TR Electronic: Tel: +49 231 975151-0 USA: 248-244-2280 Fax: +49 231 975151-51 Canada: 519-452-1233 www.microsonic.de www.trelectronic.com [email protected] Description
NG360 Inclinometer with digital RS485 communications port and 360 degree measuring range
Inclinometer with digital RS485 communications port and 360 degree measuring range Features integrated 16bit microprocessor RS485 bus output signal no measuring range limitation 0.01 degree resolution
Push button 3gang standard flush-mounted 751633xx
Documentation Product name: Push button 3gang standard Design: device (up) ETS search path: Push button / Push button 3gang / Push button 3gang standard Functional description: The push button 3gang standard
CNC-STEP. "LaserProbe4500" 3D laser scanning system Instruction manual
LaserProbe4500 CNC-STEP "LaserProbe4500" 3D laser scanning system Instruction manual 2 Hylewicz CNC-Technik Siemensstrasse 13-15 D-47608 Geldern Fon.: +49 (0) 2831 133236 E-Mail: [email protected] Website:
Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2. Instruction Manual
Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2 Instruction Manual Firmware version: 2.0.1 Issue: 0.6 Version of the GPRS transmitters configurator: 1.3.6.3 Date of issue: 07.03.2012 TABLE OF CONTENTS
DS1621 Digital Thermometer and Thermostat
Digital Thermometer and Thermostat www.dalsemi.com FEATURES Temperature measurements require no external components Measures temperatures from 55 C to +125 C in 0.5 C increments. Fahrenheit equivalent
Introduction Ericsson Handheld Telephone 1341-B
Ericsson Handheld Telephone 1341-B 2 Contents General 5 The Mobile Telephone Network 6 Base Station and Cell 7 Radio Channels 7 Radio Coverage 8 Transmission Control and Communication 9 Quality Control
Combi B PC software 8213X. Installation and operating instructions
Combi B PC software 8213X Installation and operating instructions EN Table of Contents 1 Important notes... 3 2 Glossary of abbreviations and terms... 4 3 Symbols and special font types used... 5 4 Scope
US-Key New generation of High performances Ultrasonic device
US-Key New generation of High performances Ultrasonic device US-Key connected to a laptop computer US-Key Ultrasound device single channel Features USB2 High Speed connection Ultralow noise preamplifier
Adding Heart to Your Technology
RMCM-01 Heart Rate Receiver Component Product code #: 39025074 KEY FEATURES High Filtering Unit Designed to work well on constant noise fields SMD component: To be installed as a standard component to
User Manual Software for DSL Digital Speed Switch
User Manual Software for DSL Digital Speed Switch Software version from 1.0.1 Baumer Hübner GmbH Max-Dohrn-Str. 2+4 D-10589 Berlin Phone +49 (0)30 690 03-0 Fax +49 (0)30 690 03-104 [email protected]
Push button 2, 3, 5gang with room thermostat (RTR) and display flush-mounted 756627xx, 756637xx, 756657xx
756627xx, 756637xx, 756657xx Documentation Product name: Push button 2-, 3-, 5- gang RTR + display Design: UP (flush-mounting type) ETS search path: Push button / Push button xgang / Push button xgang
Light scene push button 8gang comfort flush-mounted 751688xx
751688xx Documentation Product name: Light scene push button 8gang comfort Design: flush-mounting type (Up) ETS search path: Push button / Push button general / Light scene push button 8gang comfort Functional
Android based Alcohol detection system using Bluetooth technology
For more Project details visit: http://www.projectsof8051.com/android-based-alcohol-detection-system-usingbluetooth-technology/ Code 1435 Project Title Android based Alcohol detection system using Bluetooth
TLK 48 MICROPROCESSOR-BASED DIGITAL ELECTRONIC REGULATOR
TLK 48 MICROPROCESSOR-BASED DIGITAL ELECTRONIC REGULATOR TECHNICAL DATA CARATTERISTICHE MECCANICHE Housing Self-extinguishing plastic, UL 94 V0 Dimensions 48x48 mm DIN depth 98 mm Weight 225 g approx.
Automation and Systems Technology. LABCONTROL TFM/TPM User Manual TroxMConnect Operating Software. 1. Introduction 1. 2. Software Installation 2
1. Introduction 1 2. Software Installation 2 2.1 System Requirements 2 2.2 TroxMConnect Installation 2 2.3 Connection to the PC 3 3. TroxMConnect Description 4 3.1 Starting the Software 4 3.2 Layout 4
Laser diffuse reflection light scanner with background suppression. Dimensioned drawing
Specifications and description HRTL 3B Laser diffuse reflection light scanner with background suppression Dimensioned drawing We reserve the right to make changes DS_HRTL3B_en.fm en 01-2010/12 50114049
Non-invasive CLAMP-ON ultrasonic flowmeter for liquids
Non-invasive CLAMP-ON ultrasonic flowmeter for liquids Model CU100: non-invasive installation flowmeter, with transducers installed outside the pipe Installation in horizontal or vertical pipe Easy and
Technical Manual. FAN COIL CONTROLLER COOLING or HEATING ANALOG or PWM Art. 119914 631001A
COOLING or HEATING ANALOG or PWM Art. 119914 631001A TOTAL AUTOMATION GENERAL TRADING CO. LLC SUITE NO.506, LE SOLARIUM OFFICE TOWER, SILICON OASIS, DUBAI. UAE. Tel. +971 4 392 6860, Fax. +971 4 392 6850
sensors Miniature Sensors - S100 The universal miniature photoelectric sensor
Miniature Sensors - S1 The universal miniature photoelectric sensor Two threaded front mounting holes Two slotted rear mounting holes Anti-tampering sensor (no adjustment) Standard optic functions and
RECOMMENDATION ITU-R F.1113. (Question ITU-R 157/9) b) that systems using this mode of propagation are already in service for burst data transmission,
Rec. ITU-R F.1113 1 RECOMMENDATION ITU-R F.1113 RADIO SYSTEMS EMPLOYING METEOR-BURST PROPAGATION (Question ITU-R 157/9) (1994) Rec. ITU-R F.1113 The ITU Radiocommunication Assembly, considering a) that
Programmable Single-/Dual-/Triple- Tone Gong SAE 800
Programmable Single-/Dual-/Triple- Tone Gong Preliminary Data SAE 800 Bipolar IC Features Supply voltage range 2.8 V to 18 V Few external components (no electrolytic capacitor) 1 tone, 2 tones, 3 tones
Sturdy aluminium housing, anodized in blue Female connector for optical fiber (please order optical fiber separately) Mounting holes
SI-COLO Series SI-COLO2-LWL-HAMP-COL4 - Big measuring range: typ. 25 mm... 40 mm (with optical fiber and attachment optics KL-14 or KL-17) - Large assortment of optical fibers available (reflected or transmitted
Monitoring Network DMN
Monitoring Network DMN User Manual Table of contents Table of contents... 2 1. Product features and capabilities... 3 2. System requirements... 5 3. Getting started with the software... 5 3-1 Installation...
Product Information. Gateway For Connecting EnDat Encoders to PROFIBUS-DP
Product Information Gateway For Connecting EnDat Encoders to PROFIBUS-DP April 2012 PROFIBUS Gateway For Connecting EnDat Encoders Encoders with EnDat interface for connection via gateway All absolute
ABB i-bus EIB / KNX Analogue Input AE/S 4.2
Product Manual ABB i-bus EIB / KNX Analogue Input AE/S 4.2 Intelligent Installation Systems This manual describes the functionality of Analogue Input AE/S 4.2. Subject to changes and errors excepted. Exclusion
OWNERS MANUAL. Status Monitor. for Windows 95, 98, ME, NT 4, 2000 & XP. SIGNALCRAFTERS TECH, INC. www.signalcrafters.com
OWNERS MANUAL Status Monitor for Windows 95, 98, ME, NT 4, 2000 & XP SIGNALCRAFTERS TECH, INC. www.signalcrafters.com 57 Eagle Rock Avenue, East Hanover, NJ 07936 Tel: 973-781-0880 or 800-523-5815 Fax:
W150: Miniature photoelectric switch series with completely integrated electronics
Product group W50 Photoelectric proximity switch BGS Photoelectric proximity switch energ. Photoelectric reflex switch W50: iniature photoelectric switch series with completely integrated electronics Through-beam
Product Information S N O. Portable VIP protection CCTV & Alarm System 2
Product Information S N O Portable VIP protection CCTV & Alarm System 2 G O V E R N M E N T A L S E C U R I T Y S O L U T I VIP KIT Rapid Deployment VIP Protection Kit The VIP KIT has been designed to
USB-500 Series Low-Cost Data Loggers and Accessories
Low-Cost Data Loggers and Accessories Features Stand-alone, remote data loggers and portable logger assistant Measure temperature, humidity, voltage, current, or event/state change 1 or 2 channels Low
Robot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
intelligent solutions
Lichtschranken W 8- Photoelectric Baureihe switches WL - Photoelectric proximity switches, BGS Photoelectric reflex switches Through-beam photoelectric switches W 8-: Integrated stateof-the-art technology
US-SPI New generation of High performances Ultrasonic device
US-SPI New generation of High performances Ultrasonic device Lecoeur Electronique - 19, Rue de Courtenay - 45220 CHUELLES - Tel. : +33 ( 0)2 38 94 28 30 - Fax : +33 (0)2 38 94 29 67 US-SPI Ultrasound device
How To Use An Fsm1
17 Data Storage & Downloading The SFM does not store data on internal memory. Data is stored on Non-Volatile storage: MicroSD cards. Data can be downloaded via USB Cable, wireless Radio Frequency (RF),
FACTORY AUTOMATION MANUAL VISION SENSOR VOS302 / 312
FACTORY AUTOMATION MANUAL VISION SENSOR VOS302 / 312 With regard to the supply of products, the current issue of the following document is applicable: The General Terms of Delivery for Products and Services
FACTORY AUTOMATION INTELLIGENT CLAMP MONITORING FOR MACHINE TOOL SPINDLES WITH INDUCTIVE POSITIONING SYSTEMS
FACTORY AUTOMATION INTELLIGENT CLAMP MONITORING FOR MACHINE TOOL SPINDLES WITH INDUCTIVE POSITIONING SYSTEMS POSITION DETECTION FOR MACHINE TOOL CLAMP OPTIMIZING THE TOOL CLAMPING PROCESS The core of a
E190Q Lecture 5 Autonomous Robot Navigation
E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator
A Review of Security System for Smart Home Applications
Journal of Computer Science 8 (7): 1165-1170, 2012 ISSN 1549-3636 2012 Science Publications A Review of Security System for Smart Home Applications Mohammad Syuhaimi Ab-Rahman and Mohd Ariff Razaly Department
Vehicle data acquisition using CAN By Henning Olsson, OptimumG [email protected]
Vehicle data acquisition using By Henning Olsson, OptimumG [email protected] Introduction: Data acquisition is one of the best tools to increase the understanding of vehicle behavior. One can
Laser Measurement Systems of the LMS100 Product Family
ONLINE HELP Laser Measurement Systems of the LMS100 Product Family In full view in all weathers compact and economical Software version Online Help Laser Measurement Systems of the LMS100 Product Family
DM9368 7-Segment Decoder/Driver/Latch with Constant Current Source Outputs
DM9368 7-Segment Decoder/Driver/Latch with Constant Current Source Outputs General Description The DM9368 is a 7-segment decoder driver incorporating input latches and constant current output circuits
Electrical connection
(I)GSU 14B Ultrasonic Label Fork Dimensioned drawing Part No. 501 09234 4mm 12-30 V DC 4 m/s Ultrasonic forked sensor for universal application Large mouth width, hence also suitable for booklets or fan-fold
crm+130/iu/tc/e Extract from our online catalogue: Current to: 2016-04-21
Extract from our online catalogue: crm+130/iu/tc/e Current to: 2016-04-21 microsonic GmbH / Phoenixseestraße 7 / 44263 Dortmund / Germany / T +49 231 975151-0 / F +49 231 975151-51 / E [email protected]
Time functions. Functions of the relay. instabus EIB Application program description. September 2001. 12 A1 Binary 510D01
September 2 2 A Binary 5D Product family: Product type: Manufacturer: Output Binary output, -fold Siemens Name: Load switch UP 5 Order no.: 5WG 5-2AB Functional description Using the application program
ABB i-bus EIB Shutter Actuators JA/S 2.230.1, JA/S 4.230.1, JA/S 8.230.1, JA/S 4.230.1M, JA/S 8.230.1M, JA/S 4.24.1
Product Manual ABB i-bus EIB Shutter Actuators JA/S 2.230.1, JA/S 4.230.1, JA/S 8.230.1, JA/S 4.230.1M, JA/S 8.230.1M, JA/S 4.24.1 Intelligent Installation Systems Contents Page 1 Introduction.......................................
USB-500/600 Series Low-Cost Data Loggers and Accessories
Low-Cost Data Loggers and Accessories Features Stand-alone, remote data loggers and portable logger assistant Measure temperature, humidity, voltage, current, or event/state change 1 or 2 channels Low
DS1621 Digital Thermometer and Thermostat
www.maxim-ic.com FEATURES Temperature measurements require no external components Measures temperatures from -55 C to +125 C in 0.5 C increments. Fahrenheit equivalent is -67 F to 257 F in 0.9 F increments
ABB i-bus EIB / KNX EIB Monitoring Unit EUB/S 1.1
Product Manual ABB i-bus EIB / KNX EIB Monitoring Unit EUB/S 1.1 Intelligent Installation Systems This manual describes the function of the EIB monitoring module EUB/S 1.1 with the Application program
Series 7500 Torque Sensor
Properties PTO (Power Take-Off) shaft with integrated torque and angle measurement Non-contact measurement system, high robustness Plug & Play solution, no additional electronics required Performance Measurement
Electronic Pressure Switch EDS 300
Electronic Pressure Switch EDS 300 User manual Page 2 of 16 Content 1. Functions of the EDS 300...3 2. Mounting...3 3. Operating keys on the membrane keypad...4 4. Digital display...4 5. Output function...5
L-LAS-TB-CL serie. laser light curtains for inline measuring tasks
L-LAS-TB-CL serie laser light curtains for inline measuring tasks Maximum distance 2.000 mm Measurement range 6 up to 98 mm Resolution up to 4 µm up to 1,5 khz scan rate 0-10 V or 4-20 ma analogue output
MANUAL Device Type Manager
FACTORY AUTOMATION MANUAL Device Type Manager 2D Laser Scanner OBD10M-R2000-4EP-V1V17 With regard to the supply of products, the current issue of the following document is applicable: The General Terms
FLOW CALCULATOR INSTRUCTION MANUAL MESURES BAMOPHOX 759 26-06-2007 759 M1 02 E MES FLOW CALCULATOR 759-02/1
BAMOPHOX 759 E - M FLOW CALCULATOR INSTRUCTION MANUAL MESURES 22, Rue de la Voie des Bans - Z.I. de la Gare - 95100 ARGENTEUIL Tél : (33) 01 30 25 83 20 - Web : www.bamo.fr Fax : (33) 01 34 10 16 05 -
Understanding Alarm Systems
Understanding Alarm Systems A false alarm occurs when an alarm signal designed to elicit an immediate emergency Law Enforcement, Fire, or Medical response is activated, when in fact no emergency exists.
Photoelectric switches WLL 170(T), fibre-optic cables LL 3: flexible solutions with fibre-optic cable systems
WLL 0(T) Photoelectric switches Photoelectric switches with fibre-optic cable Proximity mode Photoelectric switches with fibre-optic cable Through-beam mode Photoelectric switches WLL 0(T), fibre-optic
Long sensing range. Most suitable for conveyor lines and parking lot applications. 10 m 32.808 ft (Red LED)
NX SERIES Compact Multi-voltage Photoelectric Power Supply Built-in Multi-voltage photoelectric sensor usable worldwide Conforming to ow Voltage and EMC Directive Multi-voltage The NX series can operate
T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p
Data Transmission Concepts and terminology Transmission terminology Transmission from transmitter to receiver goes over some transmission medium using electromagnetic waves Guided media. Waves are guided
PTPhoneManager II. User guide
COMMUNICATION AT ITS BEST PTPhoneManager II User guide Communications A member of PEIKER group DOCUMENT PEI TEL Communications GmbH - PTPhoneManager II - 2 - T ABLE OF CONTENTS Table of contents Table
ACTIVE INFRARED MOTION DETECTOR FOR HOUSE SECURITY SYSTEM MIOR MOHAMMAD HAFIIZH BIN ABD. RANI UNIVERSITI MALAYSIA PAHANG
ACTIVE INFRARED MOTION DETECTOR FOR HOUSE SECURITY SYSTEM MIOR MOHAMMAD HAFIIZH BIN ABD. RANI UNIVERSITI MALAYSIA PAHANG ABSTRACT Nowadays, house security system becomes the best solution to overcome house
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
DeviceNet Bus Software Help for Programming an Allen Bradley Control System
FBP FieldBusPlug V7 DeviceNet Bus Software Help for Programming an Allen Bradley Control System DeviceNet Software Help for Programming an Allen Bradley Control System Contents Page General Purpose...
WIRELESS INSTRUMENTATION TECHNOLOGY
BS&B WIRELESS, L.L.C. BS&B WIRELESS, L.L.C. WIRELESS INSTRUMENTATION TECHNOLOGY Printed February 2004 BS&B WIRELESS, L.L.C. 7422-B East 46th Place, Tulsa, OK74145 Phone: 918-622-5950 Fax: 918-665-3904
Single Channel Loop Detector
Single Channel Loop Detector Model - LD100 Series The LD100 is a single channel inductive loop detector designed for parking and access control applications. The detector is connected to an inductive loop
Inwall Room Temperature Unit
Inwall Room Temperature Unit TM11B01KNX TM11B11KNX TM11B21KNX Product Handbook Product: Inwall Room Temperature Unit Order Code: TM11B01KNX TM11B11KNX TM11B21KNX Application Program ETS: TM11B_1KNX Inwall
The Pulse of Automation Product overview Ultrasonic sensors
The Pulse of Automation Product overview Ultrasonic sensors Automation is our world. Perfect application solutions are A willingness to take entrepreneurial risks, a pioneering spirit, and a firm belief
KNX Gateway RGK02-5023E-01
KNX Gateway RGK02-5023E-01 Table of contents Page 1 General Information 1.1 Technical data 5 1.2. Intended Use 6 1.3 Specifications 6 1.4 Compatibility 6 1.5 Scope of delivery 6 2 Information about the
Welcome to the tutorial for the MPLAB Starter Kit for dspic DSCs
Welcome to the tutorial for the MPLAB Starter Kit for dspic DSCs Welcome to this tutorial on Microchip s MPLAB Starter Kit for dspic Digital Signal Controllers, or DSCs. The starter kit is an all-in-one
DREXELBROOK. USonic-R Series. Ultrasonic Level Measurement System. User Friendly
DREXELBROOK A Leader In Level Measurement Solutions USonic-R Series Ultrasonic Level Measurement System User Friendly Set the measurement range directly in inches, feet, millimeters, centimeters, or meters
Shutter Actuator Standard, 4-fold, 230 VAC, MDRC RA/S 4.230.1, GH Q631 0076 R0111. ABB i-bus EIB
, GH Q631 0076 R0111 2CDC 071 240 F0003 The Shutter Actuator Standard is used to control a maximum of four independent 230 VAC drives for positioning shutters, blinds, awnings and other hangings as well
PRESENCE DETECTOR PD-C360i/24 KNX ECO (EP10425844)
In the Master function, the remote control entries will be acknowledged as follows: acknowledged 3 x with the red LED In the Slave function, each detection is acknowledged 2 x with the green LED. 6 TEST
SATEX SA 230. - Manual. Weegtechniek holland b.v. PRECISION BALANCE. Patroonsweg 23-27 3892 DA Zeewolde HOLLAND
- Manual SATEX SA 230 PRECISION BALANCE Weegtechniek holland b.v. Patroonsweg 23-27 3892 DA Zeewolde HOLLAND Tel. + 31 36 522 20 30 Fax. + 31 36 522 20 60 Email [email protected] Website www.satex.nl page
Radio sensor powered by a mini solar cell the EnOcean STM 110 now functions with even less light
Radio sensor powered by a mini solar cell the EnOcean STM 110 now functions with even less light In this issue, we would like to present the EnOcean radio sensor module STM 110 in more detail. The module
Operating instructions Optical distance sensor O1D100 O1D103 704991 / 00 08 / 2012
Operating instructions Optical distance sensor UK O1D100 O1D103 704991 / 00 08 / 2012 Contents 1 Preliminary note4 1.1 Symbols used 4 1.2 Warning signs used 4 2 Safety instructions 4 3 Functions and features
Operating instructions Diffuse reflection sensor. OJ50xx 701396 / 01 07 / 2004
Operating instructions Diffuse reflection sensor OJ50xx 7096 / 0 07 / 004 Contents Preliminary note. Symbols used Function and features Installation. Installation of the supplied mounting fixture 4 4 Electrical
Contents. Document information
User Manual Contents Document information... 2 Introduction... 3 Warnings... 3 Manufacturer... 3 Description... Installation... Configuration... Troubleshooting...11 Technical data...12 Device Scope: PCB
GETTING TO KNOW YOUR TELSTRA PRE-PAID 3G WI-FI
GETTING TO KNOW YOUR TELSTRA PRE-PAID 3G WI-FI LET S GET THIS SHOW ON THE ROAD You must be excited about your brand new Telstra Pre-Paid 3G Wi-Fi. This guide will help you get connected as quickly and
-Helping to make your life betterwww.person-to-person.net
Household Telephone Management System Built on Interceptor ID Technology Owner/Operation Manual Telephone Management System- Model P2P101 Call Receiver - Model P2P301 (Receiver may be sold separately)
USB 3.0* Radio Frequency Interference Impact on 2.4 GHz Wireless Devices
USB 3.0* Radio Frequency Interference Impact on 2.4 GHz Wireless Devices White Paper April 2012 Document: 327216-001 INFORMATION IN THIS DOCUMENT IS PROVIDED IN CONNECTION WITH INTEL PRODUCTS. NO LICENSE,
Impedance 50 (75 connectors via adapters)
VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:
MANUAL FOR RX700 LR and NR
MANUAL FOR RX700 LR and NR 2013, November 11 Revision/ updates Date, updates, and person Revision 1.2 03-12-2013, By Patrick M Affected pages, ETC ALL Content Revision/ updates... 1 Preface... 2 Technical
How To Use An Ams 5812 Pressure Sensor With A Usb Starter Kit
User Guide USB Starter Kit AMS 5812 Phone:+49 (0)6131/91 0730-0 Fax: +49 (0)6131/91 073-30 Internet: E Mail: [email protected] Analog Microelectronics GmbH An der Fahrt 13, D 55124 Mainz May 2012 - Rev.
Short Form Catalogue. Alarm Systems. Reliable Supervision and Control
Short Form Catalogue Alarm Systems Reliable Supervision and Control Alarm Monitors and Indicators The SELCO product range includes a number of alarm monitors and indicator panels for use in numerous applications.
Lab 5 Getting started with analog-digital conversion
Lab 5 Getting started with analog-digital conversion Achievements in this experiment Practical knowledge of coding of an analog signal into a train of digital codewords in binary format using pulse code
WEA-Base. User manual for load cell transmitters. UK WEA-Base User manual for load cell transmitters Version 3.2 UK
WEA-Base User manual for load cell transmitters 1 Contents 1. Technical data... 3 2. Assembly... 4 2.1 Power supply... 4 2.2 Load cells... 4 2.3 RS-485... 4 2.4 Relays... 5 2.5 Digital input... 5 2.6 Analogue
TX GSM SMS Auto-dial Alarm System. Installation and User Manual
TX GSM SMS Auto-dial Alarm System Installation and User Manual Product Features: 1. 16 wireless zones, 3 wired zones alarm system, suitable for small to medium size offices and homes. 2. The system uses
INFRARED PARTS MANUAL
INFRARED PARTS MANUAL PIR325 FL65 GLOLAB CORPORATION Thank you for buying our Pyroelectric Infrared components. The goal of Glolab is to produce top quality electronic kits, products and components. All
Advanced Monolithic Systems
Advanced Monolithic Systems FEATURES Three Terminal Adjustable or Fixed oltages* 1.5, 1.8, 2.5, 2.85, 3.3 and 5. Output Current of 1A Operates Down to 1 Dropout Line Regulation:.2% Max. Load Regulation:.4%