Feedback Control in an Atomic Force Microscope Used as a Nano-Manipulator M. Hrouzek

Similar documents
Lecture 4 Scanning Probe Microscopy (SPM)

Scanning Probe Microscopy

STM and AFM Tutorial. Katie Mitchell January 20, 2010

1 Introduction. 1.1 Historical Perspective

The Design and Characteristic Study of a 3-dimensional Piezoelectric Nano-positioner

INTRODUCTION TO SCANNING TUNNELING MICROSCOPY

MEMS mirror for low cost laser scanners. Ulrich Hofmann

It has long been a goal to achieve higher spatial resolution in optical imaging and

Multi-mode Atomic Force Microscope (with High Voltage Piezo Force Microscope and +/ Oe Variable Field module.)

Atomic Force Microscope

Tecniche a scansione di sonda per nanoscopia e nanomanipolazione 2: AFM e derivati

Keysight Technologies How to Choose your MAC Lever. Technical Overview

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale

Scanning Near-Field Optical Microscopy for Measuring Materials Properties at the Nanoscale

Non-Contact Vibration Measurement of Micro-Structures

5. Scanning Near-Field Optical Microscopy 5.1. Resolution of conventional optical microscopy

Surface Profilometry as a tool to Measure Thin Film Stress, A Practical Approach. Gianni Franceschinis, RIT MicroE Graduate Student

Precision Work on the Human Eye

ATOMIC FORCE MICROSCOPY

7/3/2014. Introduction to Atomic Force Microscope. Introduction to Scanning Force Microscope. Invention of Atomic Force Microscope (AFM)

Force Spectroscopy with the Atomic Force Microscope

Fast-scanning near-field scanning optical microscopy. using a high-frequency dithering probe

Piezoelectric Scanners

Synthetic Sensing: Proximity / Distance Sensors

Physics 441/2: Transmission Electron Microscope

Self-Mixing Laser Diode Vibrometer with Wide Dynamic Range

Experiment 5. Lasers and laser mode structure

High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering

(Nano)materials characterization

Atomic Force Microscopy. Long Phan Nanotechnology Summer Series May 15, 2013

Precision Work on the Human Eye

A METHOD OF PRECISE CALIBRATION FOR PIEZOELECTRICAL ACTUATORS

BIOMEDICAL ULTRASOUND

CREOL, College of Optics & Photonics, University of Central Florida

Basic principles and mechanisms of NSOM; Different scanning modes and systems of NSOM; General applications and advantages of NSOM.

Resolution Enhancement of Imaging Chips with Piezo Technology

Vibration Suppression R&D at University of British Columbia. Tom Mattison UBC

Tecniche a scansione di sonda per nanoscopia e nanomanipolazione: STM, AFM e derivati

Atomic Force Microscopy Observation and Characterization of a CD Stamper, Lycopodium Spores, and Step-Height Standard Diffraction Grating

A Practical Guide to Free Energy Devices

Nano Meter Stepping Drive of Surface Acoustic Wave Motor

AFM-kit. Development of a kit for building from scratch an educational (but decently performing - i.e. usable) AFM.

- particle with kinetic energy E strikes a barrier with height U 0 > E and width L. - classically the particle cannot overcome the barrier

ENS 07 Paris, France, 3-4 December 2007

ATOMIC FORCEMICROSCOPYASTOOL INCELLBIOLOGICAL RESEARCH FORGROUNDBASEDANDIN-FLIGHTSTUDIES

Waves - Transverse and Longitudinal Waves

Calibration of AFM with virtual standards; robust, versatile and accurate. Richard Koops VSL Dutch Metrology Institute Delft

Lecture 6 Scanning Tunneling Microscopy (STM) General components of STM; Tunneling current; Feedback system; Tip --- the probe.

Current control of piezoelectric actuators with power loss compensation

AFM (Atomic Force Microscope) Instructions

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Whitepaper. Image stabilization improving camera usability

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706

Self-Mixing Differential Laser Vibrometer

19 th INTERNATIONAL CONGRESS ON ACOUSTICS MADRID, 2-7 SEPTEMBER 2007

The accelerometer designed and realized so far is intended for an. aerospace application. Detailed testing and analysis needs to be

Regularities of Signal and Sensitivity Variation of a Reflection Fiber Optopair Sensor Dependent on the Angle between Axes of Fiber Tips

Piezoelectric Driven Non-toxic Injector for Automated Cell Manipulation

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

Safakcan Tuncdemir 1, William M. Bradley *2. 1. Introduction

Scanning Near Field Optical Microscopy: Principle, Instrumentation and Applications

OPEN LOOP CONTROL OF FLEXIBLE BEAM PERIODIC MOTION VIA FREQUENCY RESPONSE ANALYSIS

Motion and Positioning

Introduction to Fourier Transform Infrared Spectrometry

Acoustic GHz-Microscopy: Potential, Challenges and Applications

Atomic Force Microscope and Magnetic Force Microscope Background Information

DETECTION OF SUBSURFACE DAMAGE IN OPTICAL TRANSPARENT MATERIALSS USING SHORT COHERENCE TOMOGRAPHY. Rainer Boerret, Dominik Wiedemann, Andreas Kelm

Department of Aerospace Engineering Indian Institute of Science Bangalore

The Basics of Scanning Electron Microscopy

Robot Perception Continued

Encoders for Linear Motors in the Electronics Industry

ENGINEERING METROLOGY

Optical Measurement Techniques for Dynamic Characterization of MEMS Devices

Acousto-optic modulator

Fiber Optics: Engineering from Global to Nanometer Dimensions

Laser-induced surface phonons and their excitation of nanostructures

Module 13 : Measurements on Fiber Optic Systems

An Electromagnetic Micro Power Generator Based on Mechanical Frequency Up-Conversion

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm

Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator

Raman Spectroscopy Basics

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications

Practical Application of Industrial Fiber Optic Sensing Systems

Active Vibration Isolation of an Unbalanced Machine Spindle

Raman spectroscopy Lecture

Challenges in DWDM System Spectral Analysis By Laurent Begin and Jim Nerschook

Welcome to this presentation on LED System Design, part of OSRAM Opto Semiconductors LED 101 series.

Tube Control Measurement, Sorting Modular System for Glass Tube

Alternative Linear Motion Systems. Iron Core Linear Motors

Accurate Air Flow Measurement in Electronics Cooling

Section 5.0 : Horn Physics. By Martin J. King, 6/29/08 Copyright 2008 by Martin J. King. All Rights Reserved.

AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE

Usage of Carbon Nanotubes in Scanning Probe Microscopes as Probe. Keywords: Carbon Nanotube, Scanning Probe Microscope

State Newton's second law of motion for a particle, defining carefully each term used.

A More Efficient Way to De-shelve 137 Ba +

A. Ricci, E. Giuri. Materials and Microsystems Laboratory

Microscopie à force atomique: Le mode noncontact

Cover Page. The handle holds various files of this Leiden University dissertation.

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

Blackbody radiation derivation of Planck s radiation low

Transcription:

Feedback Control in an Atomic Force Microscope Used as a Nano-Manipulator M. Hrouzek This paper offers a concise survey of the most commonly used feedback loops for atomic force microscopes. In addition it proposes feedback control loops in order to minimize the effect of thermal noise on measurements of weak forces, and to improve the manipulability of the AFM. Growing requirements to study and fabricate systems of ever-shrinking size mean that ever-increasing performance of instruments like atomic force microscopes (AFM) is needed. A typical AFM consists of a micro-cantilever with a sharp tip, a sample positioning system, a detection system and a control system. Present day commercial AFMs use a standard PI controller to position the micro-cantilever tip at a desired distance from the sample. There is still a need for studies showing the optimal way to tune these controllers in order to achieve high closed-loop positioning performance. The choice of other controller structures, more suitable for dealing with the robustness/performance compromise can also be a solution. Keywords: automatic control, atomic force microscopy, thermal noise. 1 Description of AFM Atomic Force Microscopy (AFM) is an aspect of Scanning Force Microscopy, and is capable of measuring the interaction forcebetweenthesampleandasharptipmountedontheend of a weak cantilever, see Fig. 1. The cantilever is made of silicon and usually has a rectangular or triangular shape. Its length does not exceed 600 m. Two main AFM operation modes are used: contact mode and dynamic mode. In contact mode the cantilever is in full contact with the surface, and the interaction force (displacement of the lever) is determined by measuring the static deflection of the cantilever. This technique is capable of detecting displacements on atomic scale resolution (0.1 200 nm). The dynamic detection mode can be divided into techniques called the tapping mode and the non- (pseudo)- contact mode [1]. In these detection modes, the cantilever is driven, usually at its resonant frequency, with an amplitude typically less than 10 nm. The cantilever driver is usually a piezo-electric element, but many experiments have been performed with electrostatic, magnetic, thermo-optic and acoustic coupling drivers. The driver is attached to the head of the microscope and the chip with the cantilever and tipisassembledontopofthedriver. In the tapping mode, the tip is touches the surface in each period of cantilever movement at its maximum deflection to- Fig. 1: Schema of an Atomic Force Microscope. Stage positioning is controlled by two loops (X-Y) displayed in the lower part of the picture. The driving and head positioning loops are at the top. Their only input is a signal from the photo detector that measures the deflection of the cantilever. Czech Technical University Publishing House http://ctn.cvut.cz/ap/ 65

Acta Polytechnica Vol. 45 No. 4/2005 Czech Technical University in Prague wards the surface. There is direct mechanical contact between the tip and the surface. Another way to scan soft samples is with the use of a non-contact technique. The cantilever vibrates very close to the surface, but does not come into direct contact with it. Attractive long distance interaction forces affect the effective spring constant of the cantilever, and the cantilever shifts its resonance frequency. This technique is used for measuring weak forces, such as the van der Waals force, and for scanning soft samples. Accuracy in measuring the cantilever displacement is very important in order to achieve good AFM resolution. Many techniques have been developed for this measurement, based on various principles, including capacitive and optical techniques as well as interferometry, optical beam deflection and laser diode detection. Each technique has specific advantages and disadvantages. A method widely used by many manufacturers of AFMs is the optical beam deflection technique, shown in Fig. 1. Its function is based on a sensitive photodetector which is able to detect very small laser beam displacements. The photodiode receiving the reflected laser beam from the cantilever is divided into two sections, A and B. Due to the macroscopic length of the reflected light path, any deflection causes a magnified displacement of the reflected laser spot on the photodiode. The relative amplitudes of the signals from the two segments of the photodiode change in response to the motion of the spot. The difference signal (A B) is very sensitive to cantilever deflection. 2 Feedback control in the AFM Modern AFMs include many feedback loops to control different blocks of this complicated instrument. Some of the most exciting and still widely open areas in AFM feedback control are: 2.1 Loop controlling position of the AFM stage on the X-Y axis Many experiments have been performed with various techniques for moving the stage with the scanned sample with high precision and speed. Over many years, most AFM manufacturers and scientists have preferred piezo-electric stack actuators. These devices offer fine position capabilities achieving sub-angstrom resolution and high-speed manipulation. Piezo electric actuators have disadvantages in nonlinear properties such as hysteresis and creep, which become more significant with the increasing need for fast movement given by the actuators, see Fig. 2. This illustrates the need for a better and more accurate description of the piezo elements and for improving the feedback control systems by robust design techniques. Surface physicists have used AFM mainly for scanning purposes in the past. Recently, AFM has been used in many different fields of biology, medicine, material science, nanomanipulation and nano-lithography. These extended uses involve the need for an improvement in the accuracy and repeatability of movements with the tip. In addition, the possibilities of using AFMs as manipulators on an atomic scale rather than as classical scanners are being explored. The difference between these two approaches is displayed in Fig 3. The part on the left is a simple movement across the sample in straight lines, scanning. On the other side, we see the Fig. 2: Sketch of the hysteresis of a piezo stack actuator for slow and fast movement with the stage. Increasing speed of manipulation usually leads to problems with accuracy of positioning, due to the rising influence of nonlinearities such as hysteresis. movement that is required to manipulate a small object on the surface. The cantilever has to reach the desired position, then approachthesurfaceandmakeamovementwhichisnotnecessarily straight but can be of any shape. While manipulating an particle, it is very important to control the applied force by the tip on to the manipulated object. Currently used regulators usually have a simple proportional-integral structure, which offers zero steady state regulation error but is not able to achieve the desired output in a short period of time. These regulators usually experience tracking problems that become significant with rising speed. The bandwidth limitations and non robust control are not satisfactory. Present-day feedback control loops for the x and y Fig. 3: Scanning approach on the left, manipulation approach on the right 66 Czech Technical University Publishing House http://ctn.cvut.cz/ap/

Fig. 4: Feedback control loop that can be used to control the position of the sample mounted on the stage of the AFM axis have to be able to perform very complex movements at very high speed, rather than simply dragging the cantilever across the sample in straight lines with a frequency of a few hertz. This area is now open to new techniques and new ideas on how to control the nonlinearities of systems coming from piezo hysteresis in order to achieve the desired performance, see [2]. Fig. 4 displays the feedback control loops that are employed in the stage positioning system. Fig. 5: Frequency spectrum of a freely vibrating cantilever driven by thermal noise (Brownian motion). The spectrum displays the first two resonance frequencies. Measurements on the NanoTec AFM Czech Technical University Publishing House http://ctn.cvut.cz/ap/ 67

Acta Polytechnica Vol. 45 No. 4/2005 Czech Technical University in Prague Regulators have to be capable of treating nonlinearities of the controlled systems by using internal inverse models. With increasing requirements for accuracy of stage positioning by the piezo actuator, the problem of measuring distances on a smaller scale is becoming apparent. Most applications use a Linear Variable Differential Transformer (LVDT), which offers the best sensitivity. Position sensors experience problems with exceedingly high levels of electric and thermal noise, which lower the resolution. This problem is another field in which of feedback control might be applied. An observer based regulator could achieve better signal noise separation than current systems without any feedback. 2.2 Loop controlling the position of the AFM head on the Z axis and cantilever excitation The description given below focuses on the non-contact operating mode of the AFM. The non-contact dynamic technique is very sensitive to thermal noise, but is frequently used for nondestructive measurement of soft biological samples. The cantilevers are designed and manufactured small and soft enough not to destroy the surface of the measured sample. A very soft cantilever is sensitive to thermal excitation, which appears in the system as the largest source of noise [3]. Thermal noise significantly affects sensitivity, and limits possible improvements in resolution. Fig. 5 displays the amplitude spectrum measured on the cantilever without any artificial excitation. The spectrum shows little excitation across all frequencies, and large noise excitations are visible at the resonance frequencies of the cantilever. Some techniques already exist for improving sensitivity by eliminating the thermal noise of the cantilever. An interesting technique for improve is Noise Squeezing [4]. The cantilever is of course exposed to many other noises, such as electrostatic noise, electromagnetic noise and air turbulence. The influence of these noises can be eliminated by appropriate mechanical and electrical construction of the AFM. Only thermal noise cannot be attenuated by changes in the construction of the microscope. Some applications enable the sample and microscope to be put in a thermal bath at a very low temperature, which reduces the thermal excitation. However, there are many applications that have to be at room temperature. By applying a certain feedback control loop, the noise can be eliminated and the sensitivity of the instrument can be increased. Scanning force microscopes are usually controlled by simple proportional-integral regulators which are not able to give the best possible results. Fig. 6 shows control loops affecting the sensitivity of the AFM. The regulators used in these loops need to be designed by optimal control techniques to achieve better sensitivity. The control system consists of two main parts: the positioning loop and the driving loop. The positioning loop works at lower frequencies, and its task is to keep the head in a position that ensures a constant signal from the tip-surface interaction. It simultaneously measures the surface without losing any information. In scanning force microscopy this constant signal is known as the set point, and ensures that the cantilever vibrates at its maximum harmonic deflection without hard tapping on the surface. The driving loop operates at the higher frequencies and keeps the cantilever excited at its resonant frequency and at the same amplitude despite the contact (van der Waals attractive and mechanical repulsive interactions) with the surface of the sample. This is usually done by the piezoelectric element (driver) harmonically vibrating at the desired frequency (slightly off the resonance frequency) with a very small amplitude. Fig. 6: Feedback control loops of the head positioning and the cantilever drive 68 Czech Technical University Publishing House http://ctn.cvut.cz/ap/

The third additional regulator is responsible for thermal noise attenuation. Application of this loop is very complicated, because the controller can easily destroy all measured signals by applying excessively large attenuating signals. Its function can be similar to Cold Damping techniques, known from cooling of nano-mechanical resonators or cavities [5]. Themaingoalofthistechniqueistostabilizethepositionof the mechanical part so that its displacement will be minimized and it appears virtually colder. In our case this is not fully true. The displacement caused by thermal noise needs to be eliminated, but the displacement caused by the driver or interaction forces needs to remain undisturbed, or less the measured signal will be changed. Another possible design is an observer based regulator, which would be able to approximate the influence of the thermal noise and interaction forces [6]. Both of these approaches face problems with determining which part of the signal is unwanted noise. 3 Conclusion Atomic Force Microscopy is a new field for applying advanced control systems capable of improving the properties of the instrument. Presently used control systems are very simple, but ensure functionality. More sophisticated control system will enable Atomic Force Microscopes to be applied as nano-manipulators on an atomic scale, with improved sensitivity. 4 Acknowledgments This work was supported by research project J22/98:26 22000 12 Research in Informatics and Control Systems. References [1] Stark,M.,Stark,R.W.,Heckl,W.M.,Guckenberger,R.: Spectroscopy of the Anharmonic Cantilever Oscillations in Tapping-Mode Atomic-Force Microscopy. Applied Physics Letters, November 2000. [2] Salapaka, S., Sebastian, A., Cleveland, J. P., Salapaka, M. V.: High Bandwidth Nanopositioner: A Robust Control Approach. Review of Scientic Instruments, September 2002. [3] Cleland,A.N.,Roukes,M.L.: NoiseProcessesinNanomechanical Resonators. Journal of Applied Physics, 2002. [4] Rugar, D., Grutter, P.: Mechanical Parametric Amplication and Thermomechanical Noise Squeezing. Physical Review Letters, August 1991. [5] Pinard,M.,Cohadon,P.F.,Briant,T.,Heidmann,A.: Full Mechanical Characterization Of A Cold Damped Mirror. Physical Review A, December 2000. [6] Sebastian, A., Sahoo, D. R., Salapaka, M. V.: An Observer Based Sample Detection Scheme for Atomic Force Microscopy. In Proceedings of the 42 nd IEEE Conference on Decision and Control, Maui, Hawaii, USA, December 2003. Michal Hrouzek e-mail: mhx@seznam.cz Laboratoire d Automatique de Grenoble INP/UJF Grenoble, France and Dept. of Control and Instrumentation FEEC, BUT Brno, Czech Republic Czech Technical University Publishing House http://ctn.cvut.cz/ap/ 69