PHYSICS 176 UNIVERSITY PHYSICS LAB II. Experiment 4. Alternating Current Measurement

Similar documents
EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

Lab 1: The Digital Oscilloscope

Electrical Resonance

Beginners Guide to the TDS 210 and TDS 220 Oscilloscopes

RC Circuits and The Oscilloscope Physics Lab X

Lab E1: Introduction to Circuits

Oscilloscope, Function Generator, and Voltage Division

Using an Oscilloscope

AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz

EXPERIMENT NUMBER 8 CAPACITOR CURRENT-VOLTAGE RELATIONSHIP

Figure 1: Multiple unsynchronized snapshots of the same sinusoidal signal.

Experiment 2 Diode Applications: Rectifiers

ε: Voltage output of Signal Generator (also called the Source voltage or Applied

FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER

User Manual. TDS 200-Series Digital Real-Time Oscilloscope

= V peak 2 = 0.707V peak

Fundamentals of Signature Analysis

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.

Annex: VISIR Remote Laboratory

Ph 3504 Nuclear Magnetic Resonance and Electron Spin Resonance

ANADOLU UNIVERSITY DEPARTMENT OF ELECTRICAL AND ELECTRONICS ENGINEERING

Lab 3 Rectifier Circuits

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.

ECEN 1400, Introduction to Analog and Digital Electronics

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement

Physics 120 Lab 6: Field Effect Transistors - Ohmic region

Experiment #11: LRC Circuit (Power Amplifier, Voltage Sensor)

USER S MANUAL. Hantek6022BE. V1.0.3

SERIES-PARALLEL DC CIRCUITS

CHAPTER 11: Flip Flops

DSO-1062D/DSO-1102D/DSO-1202D different bandwidths. Digital Oscilloscope User Manual

INTERFERENCE OF SOUND WAVES

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b

User s Manual of OWON Colour Digital Storage Oscilloscope

EE 1202 Experiment #4 Capacitors, Inductors, and Transient Circuits

Inductors in AC Circuits

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT

Op-Amp Simulation EE/CS 5720/6720. Read Chapter 5 in Johns & Martin before you begin this assignment.

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE Test Method for AC to DC Power Supplies

AC Direct Off-Line Power Supplies

TESTS OF 1 MHZ SIGNAL SOURCE FOR SPECTRUM ANALYZER CALIBRATION 7/8/08 Sam Wetterlin

Reading assignment: All students should read the Appendix about using oscilloscopes.

LABORATORY 2 THE DIFFERENTIAL AMPLIFIER

RIGOL Data Sheet. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Product Overview. Applications. Easy to Use Design

DCM555 - Data Communications Lab 8 Time Division Multiplexing (TDM) Part 1 - T1/DS1 Signals

Dash 18X / Dash 18 Data Acquisition Recorder

Lesson 3 DIRECT AND ALTERNATING CURRENTS. Task. The skills and knowledge taught in this lesson are common to all missile repairer tasks.

[F/T] [5] [KHz] [AMP] [3] [V] 4 ) To set DC offset to -2.5V press the following keys [OFS] [+/-] [2] [.] [5] [V]

Electromagnetic Induction: Faraday's Law

RC & RL Transient Response

Basic oscilloscope operation

The full wave rectifier consists of two diodes and a resister as shown in Figure

MATERIALS. Multisim screen shots sent to TA.

Dash 8Xe / Dash 8X Data Acquisition Recorder

U1602A Handheld Oscilloscopes, 20 MHz

User Manual. Digital Storage Oscilloscope DSO25 / DSO100. Version No.: V 1.6

RLC Series Resonance

Reading: HH Sections , (pgs , )

RIGOL Data Sheet. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Product Overview. Applications. Easy to Use Design

Electronic WorkBench tutorial

Embest DSO2300 USB Oscilloscope

Lab Exercise 1: Acoustic Waves

Table of Contents. SAFETY INSTRUCTION...6 Safety Symbols... 6 Safety Guidelines... 7 Power cord for the United Kingdom... 9

CONSTRUCTING A VARIABLE POWER SUPPLY UNIT

UT4000. Series USER MANUAL. Digital Storage Oscilloscope

Experiment 5. Strain Gage Measurements

AC CIRCUITS - CAPACITORS AND INDUCTORS

Lab #9: AC Steady State Analysis

PropScope USB Oscilloscope

Lab 3 - DC Circuits and Ohm s Law

Q1. The graph below shows how a sinusoidal alternating voltage varies with time when connected across a resistor, R.

Lecture - 4 Diode Rectifier Circuits

AMERICAN NATIONAL STANDARD

Frequency Response of Filters

User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706

Laboratory Manual for AC Electrical Circuits

Power Supplies. 1.0 Power Supply Basics. Module

Precision Diode Rectifiers

DSO8000E SERIES HANDHELD OSCILLOSCOPE

Table of Contents. The Basics of Electricity 2. Using a Digital Multimeter 4. Testing Voltage 8. Testing Current 10. Testing Resistance 12

Dayton Audio is proud to introduce DATS V2, the best tool ever for accurately measuring loudspeaker driver parameters in seconds.

Automotive Sensor Simulator. Automotive sensor simulator. Operating manual. AutoSim

Instruction Manual and Experiment Guide for the PASCO scientific Model WA-9307A D 6/94 FOURIER SYNTHESIZER PASCO scientific $7.

Electricity & Electronics 5: Alternating Current and Voltage

How To Test A Computer With A Powerline 2.5 (Powerline) And Powerline (Powerplant) (Powerboard) (Awn) (Ios) (Mini Computer) (Microphone) (Wireless) (

Electronics. Discrete assembly of an operational amplifier as a transistor circuit. LD Physics Leaflets P

css Custom Silicon Solutions, Inc.

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

Lab 4 - Data Acquisition

Technical Datasheet Scalar Network Analyzer Model MHz to 40 GHz

Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist.

ANALYTICAL METHODS FOR ENGINEERS

OPTICAL FIBERS INTRODUCTION

EE 255 ELECTRONICS I LABORATORY EXPERIMENT 2 POWER SUPPLY DESIGN CONSIDERATIONS

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

Parallel Plate Capacitor

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

Transcription:

PHYSICS 176 UNIVERSITY PHYSICS LAB II Experiment 4 Alternating Current Measurement Equipment: Supplies: Oscilloscope, Function Generator. Filament Transformer. A sine wave A.C. signal has three basic properties: amplitude, frequency, and phase. The measurement of amplitude and frequency with an oscilloscope is illustrated below. The vertical and horizontal axes can be set to various voltage/div and time/div scales respectively. The peak-to-peak amplitude of the sine wave is the voltage V pp between the top of the wave and the bottom. Its value is found from the trace on the oscilloscope screen as follows: V pp = (peak-to-peak amplitude in div) x (volts/div) The period of the sine wave is time T between successive peaks of the wave. Its value is found from the trace on the oscilloscope screen as follows: T = (distance between peaks in div) x (time/div) The frequency of the sine wave signal is the inverse of the period: f = 1/T Hertz (1 Hertz = 1 second -1 ) The mathematical expression for the sine wave signal is usually in terms of the peak amplitude V p = V pp /2 and the angular frequency ω = 2πf. Signal voltage = V p sin (ωt + θ) The phase angle θ is given in radians (2π radians = 360 ). The phase angle describes how one sine wave signal is shifted relative to another. The dual trace oscilloscope allows two signals to be displayed on the screen simultaneously and this permits a measurement of their relative phases. The measurement of phase will be considered in more detail in Experiment 5. 1

A. Oscilloscope Operation The front panel of TDS 1002 oscilloscope is given below. It is divided into easy-to-use functional areas. Vertical controls: Cursor 1 and Cursor 2: Position the waveform vertically. Ch1 Menu, Ch2 Menu: Display the vertical menu selections and toggles the display of the channel waveform on/off. Math Menu: Displays the waveform math operations menu. Volts/Div: Selects calibrated scale factors. Horizontal controls: Position: Adjusts the horizontal position of the waveform. Horiz. menu: Displays the Horizontal menu. Set to zero: Sets the horizontal position to zero. Sec/Div: Selects the horizontal time/div (scale factor) for the window time base. Trigger controls: Used to set up the trigger level. Menu and Control buttons: Save/Recall: Displays the Save/Recall menu Measure: Displays the automated measurements menu Acquire: Displays the acquire menu Cursor: Displays the cursor menu Utility: Displays the utility menu Auto Set: Automatically sets the oscilloscope controls to produce a usable display of the input signal. This is your panic button. Punch it in anytime something goes wrong and you do not know why Default set up: Recalls the factory set up. Single Seq: Acquires a single waveform and then stops. Run/Stop: Continuously acquires waveforms. Print: Starts print operations. The BNC connectors labeled Ch1 and Ch2 are two inputs. Each input is an amplifier which takes the input signal and uses it to drive the electron beam up and down. To use channel 1 only, punch in the button Ch 1; similarly for channel 2. Because there are two input channels, the TDS 1002 is called a dual trace oscilloscope. However, there is only one electron beam, which is shared by the two channels. 2

The electron beam appears on the screen left to the controls. It sweeps across the screen with a speed selected with the Sec/Div control. Each channel input has AC-GND-DC coupling selected by pushing the Ch1/Ch2 menu buttons: AC: A blocking capacitor is in series (internally) with the input so that the trace on the screen shows only the A.C. signal part of the input. GND: The scope trace is a horizontal line which represents the level of zero volts D.C. (ground level). DC: The trace on the screen is shifted relative to the GND level by an amount equal to any D.C. voltage input. There may be an A.C. voltage (signal) superimposed on the D.C. voltage. If the input signal is periodic (which is the usual case), you want the horizontal sweep of the electron beam to start at the same point of the signal pattern each time; otherwise the trace of the signal on the screen will not appear stationary. This is accomplished by triggering the sweep of the electron beam with a chosen level of the input signal (Trigger controls). In order to find the trace initially (when the scope is first turned on), push the AutoSet button. The sweep is then triggered by the power company s 60 Hz line voltage. In order to trigger on a displayed signal use the Trigger controls. B. The Filament Transformer In the days when vacuum tube circuits were common, the tube filament was heated (to boil off electrons form the tube cathode) by passing an A.C. current through it. The A.C. voltage across the filament was derived from the secondary winding of a filament transformer. The A.C. voltage across the primary winding of the transformer was the power company's line voltage. When an A.C. voltage v(t) = V p sinωt is applied to a resistor R, an A.C. current i(t) = (V p /R)sinωt flows through the resistor. The instantaneous heating effect of this current in terms of heat dissipated/second (joules/second) is the instantaneous power p(t) = i 2 (t) R = (V p 2 /R) sin 2 ωt The time average power is P av = ½ (V p 2 /R) since the time average of sin 2 ωt is ½. A.C. voltages are often referred to by their root-mean-square values V rms =.707 V p 3

Then the time average power, in terms of the root-mean-square voltage is P av = V 2 rms/r which has the same form as the power formula for D.C. (P = I 2 R = V 2 /R). The root-mean-square value of the power company's line voltage is about 110 volts. The root-mean-quare value of the filament transformer output is typically 6.3 volts (or sometimes 12.6 volts). C. The Function Generator The function generator produces square waves, triangle waves, and sine waves shown below: In this Lab you will be using BK PRECISION Model 4017A function generator. The front panel of the generator is given in the next page. To operate the generator: Turn on the Power switch. Select the waveform of the generated signal by pushing the corresponding Function switch. Push Range Selector switch to select the frequency range at your desire. Adjust the frequency of the signal by turning the Coarse/Fine Frequency Controls. The amplitude of the output signal is controlled continuously with operating the Output Level Control (use it to generated waveforms with amplitudes from 1 V to 10 V). If the -20 db Switch is also engaged, the output amplitude is reduced by a factor of 20dB (use it to generated waveforms with amplitudes from 100 mv to 1 V) The DC OFFSET control may be used to offset the waveform above or below ground (0 volts) by a D.C. voltage in the range ±10 volts. 4

1. POWER On/OFF Switch. 2. RANGE Switch. Seven ranges from 1 Hz to 10MHz 3. FUNCTION Switch. Selects sine, square, or triangle waveforms. 4. OUTPUT LEVEL Control: up to ±20 Vpp (open circuit) or ±10 V pp into 50 Ω. 5. DC OFFSET Control: up to ±10 Vpp (open circuit) or ±5 V pp into 50 Ω. 6. OUTPUT Jack outputs the generated waveform. 7. TTL/CMOS Jack. TTL or CMOS square wave, selected by switch (15), is output at this jack. 8. CMOS LEVEL Control. 9. VCG/SWEEP Jack. 10. DUTY CYCLE Control. 11. -20 db Switch. When engaged, the signal at the OUTPUT jack is attenuated by -20dB. 12. SWEEP INT/EXT Switch. 13. SWEEP LIN/LOG Switch. 14. DC OFFSET Switch. When engaged, enables operation of the DC OFFSET control (5). 15. CMOS LEVEL Switch. 16. SWEEP WIDTH Control. 17. DUTY CYCLE Switch. 18. SWEEP TIME Control. 19. FINE FREQUENCY Control. 20. COARSE FREQUENCY Control. 21. COUNTER DISPLAY. Displays frequency of internally generated waveform. 22. GATE LED. Indicates when the frequency counter display is updated. 23. Hz and KHz LED. Indicates whether the counter is reading in Hz or khz 5

D. The Circuit Ground When you measure signals it is a good practice to have the function generator, the oscilloscope and your circuit at a common ground. It is essential that you identify the ground connections on the oscilloscope and on the function generator output cable. On one end of the function generator cable there is a silver BNC connector which plugs into the output of the function generator box. The other end of this cable may again be terminated with a BNC connector or have two leads: one with a red collar and one with a black collar. The lead with the red collar carries the signal produced by the function generator, while the lead with the black collar is the ground. When an end with two leads is used connect the ground lead to the ground of the circuit you are building The oscilloscope probes also have two leads: one carries the signal and the other (black collar with an alligator type clip) is the ground. Always connect this end to the ground of the circuit you are building. The grounds must always be connected to the same point of any circuit. When building circuits it is useful to use a ground bar so that ground is easily identified. The ground bar is a silver metal rod about ten inches long with two brass connectors to fasten it to the pegboard. The filament transformer also plugs into a wall socket, but the primary winding has no D.C. connection with the secondary winding. Therefore the transformer secondary winding is D.C. isolated from the ground in the primary winding. This is shown by the schematic diagram for the transformer below: A voltage is induced in the secondary winding when the current in the primary winding changes (Faraday's law of induced emf). 6

Setting Up Your Oscilloscope 1. Turn the oscilloscope On. 2. Examine the scope. (You should see a line "trace" in the display window). Locate the controls and examine their function and various menus. 3. Attach function generator coaxial cable (output) to oscilloscope input (Channel 1). 4. Turn on the function generator and set the following: Function = (Square Wave) Frequency = x1k A thousand 's are sent out every second Amplitude: 2 V 5. To activate the oscilloscope and display the signal push the AUTOSET button and select Channel 1. 6. Observe: 7. Adjust VOLTS/DIV and SEC/DIV knobs so that you understand their effect on the signal. 8. Examine the rising and falling edges of the square wave signal. Push the corresponding menu buttons (just on the right of the display window) to display only the edges on the screen. 9. Measure the rise time and the amplitude of the signal using moving cursors. To activate the cursors push the CURSOR button. Push the TYPE option button and select Time/Voltage. Use the Cursor1 and Cursor 2 knobs to move the cursors (appearing as broken lines). The position of the cursors are displayed/digitized on the screen 10. Completely disassemble and then begin lab. 7

Procedure - Experiment 4 A. First Time Oscilloscope Operation 1. Connect the channel 1 input to Probe COMP (located at the bottom right corner of the oscilloscope). This connection provides a 5 V, 1 khz square wave which is used to check the oscilloscope probe for capacitive compensation. If the signal which you see is not a clean square wave, the oscilloscope probe needs compensation and you should contact your laboratory instructor. 2. If you do not see any trace on the oscilloscope screen check the following: a) Is the oscilloscope plugged in? Is the scope turned on? b) Is the intensity high enough? c) Are the Position controls properly adjusted? d) Is the Triggering Mode set on Auto? 3. Position the square wave trace so that it is centered. 4. Adjust the Volts/Div and Sec/Div controls to give a good picture of the square wave. Set both the probes and the scope to 1X. Keep this setting during all measurements. 5. Measure the peak-to-peak amplitude of the square wave. 6. Measure the period T of the square wave. Compute the frequency f. 7. Switch channel 1 to GND. Use the Position control to superimpose the GND line on the central line of the screen mesh. Then switch channel 1 to DC and note the position of the square wave trace relative to the GND line. Does the square wave contain a D.C. component? B. The Filament Transformer 1. When the transformer is plugged in, do not allow its secondary leads to touch. Contact of the secondary leads puts a short circuit across the transformer output. Since a short circuit represents effectively zero resistance, a very large current would be drawn and the result would be to overheat the transformer windings. 2. Connect the channel 1 probe across the entire secondary; connect the channel 2 probe across 1/2 the secondary by using the center-tap. BE SURE THAT THE GROUNDS OF THE TWO PROBES ARE CONNECTED TOGETHER. If you have any doubts about the ground connections, check with your instructor before you make any connections. A ground on one end of the secondary and another ground on the other end of the secondary represents a short circuit across the secondary and will cause serious overheating of the transformer. 3. Measure the peak-to-peak values of the two sine waves. 4. Compute the peak values of the sine wave amplitudes. 5. Compute the root-mean-square values of the two sine waves. 6. Measure the frequency of the sine wave, that is, measure the period T and then compute the frequency f. You should find the frequency to be about 60 Hz. 8

C. The Function Generator 1. Set the function generator for a 1 volt p-p, 1 khz signal. 2. Connect the function generator cable to the channel 1 oscilloscope probe. 3. With the channel 1 input on AC, measure the peak-to-peak voltage of the 1 khz signal (nominally 1 volt). Note the following: (a) setting of the Volts/Div control, and (b) number of divisions (centimeters) peak-to-peak. 4. Measure the frequency of the 1 khz signal. Note the following: (a) setting of the Sec/Div dial, and (b) number of divisions (centimeters) between successive peaks. 5. Switch channel 1 to GND (be sure that the Triggering Mode is Auto) to establish the ground level on the scope screen. Use the Position control to put the ground level in the center of the screen. 6. Switch channel 1 to DC and determine the maximum D.C. offset voltages (both positive and negative) that can be added to the 1 khz signal voltage. D. Oscilloscope Practice Your lab partner should adjust every knob on the oscilloscope and function generator. Then you should try to obtain this signal. Now it is your turn. Adjust all the knobs and ask your partner to find these two signals: Signal 1 Signal 2 9

Data Sheet - Experiment 4 Name Section # A. First Time Oscilloscope Operation 1. Peak-to-peak amplitude of square wave V pp = mv = volts 2. Period of square wave T = msec = sec Frequency f = Hz 3. Sketch a picture of the square wave showing the ground level. B. The Filament Transformer 1. Peak-to-peak amplitudes V pp (1) = V pp (2) = 2. Peak amplitudes V p (1) = V p (2) = Root-mean-square V rms (1)= V rms (2) = amplitudes 3. Period and frequency T = sec f = Hz C. The Function Generator 1. Peak-to-peak amplitude VOLTS/DIV setting = Number of div. peak-to-peak = 2. Frequency TIME/DIV setting = Number of div. between peaks = 3. Maximum positive D.C. offset voltage = Volts Maximum negative D.C. offset voltage = Volts D. Oscilloscope Practice Show the signals to your instructor! 10