Info Note UV-VIS Nomenclature and Units

Similar documents
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Using the Spectrophotometer

Copyright by Mark Brandt, Ph.D. 12

Chemistry 111 Lab: Intro to Spectrophotometry Page E-1

Application Note: Absorbance

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

MCAL Spectrophotometry. Spectrophotometry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

2 Spectrophotometry and the Analysis of Riboflavin

Problem Set 6 UV-Vis Absorption Spectroscopy Express the following absorbances in terms of percent transmittance:

Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information

Upon completion of this lab, the student will be able to:

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

Reaction Stoichiometry and the Formation of a Metal Ion Complex

where h = J s

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law


Lab #11: Determination of a Chemical Equilibrium Constant

SPECTROPHOTOMETRIC MEASUREMENTS TECHNIQUES FOR FERMENTATION PROCESS BASE THEORY FOR UV-VIS SPECTROPHOTOMETRIC MEASUREMENTS

Experiment #5: Qualitative Absorption Spectroscopy

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

UV-Visible Spectroscopy

Molecular Spectroscopy

Ultraviolet Spectroscopy

Chem 131A: Absorbance of Riboflavin

3 - Atomic Absorption Spectroscopy

Colorimetry Extinction coefficient (ε) Lambda max (λ max ) Qualitative vs. quantitative analysis

A Beer s Law Experiment

Raman spectroscopy Lecture

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher)

Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument

VWR SPECTROPHOTOMETERS

UV-Vis spectrophotometers

Back to Basics Fundamentals of Polymer Analysis

SPECTROSCOPY. Light interacting with matter as an analytical tool

Introduction to Fourier Transform Infrared Spectrometry

Measuring Protein Concentration through Absorption Spectrophotometry

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

INTERNATIONAL OLIVE COUNCIL

Raman Spectroscopy Basics

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry

Hellma. Where precision becomes an art. Calibration standards for spectrophotometers by Hellma

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

5.33 Lecture Notes: Introduction to Spectroscopy

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION

Time out states and transitions

FTIR Instrumentation

Laboratory 5: Properties of Enzymes

Uses of Derivative Spectroscopy

2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY

Reaction of Blue Food Dye with Bleach

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.

instruments Analytical Instruments for Science

The photoionization detector (PID) utilizes ultraviolet

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Science In Action 8 Unit C - Light and Optical Systems. 1.1 The Challenge of light

Current Staff Course Unit/ Length. Basic Outline/ Structure. Unit Objectives/ Big Ideas. Properties of Waves A simple wave has a PH: Sound and Light

Infrared Spectroscopy: Theory

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F)

Chemistry 118 Laboratory University of Massachusetts Boston Beer s Law

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

NITRIC OXIDE and NITROGEN DIOXIDE 6014

Organic Spectroscopy. UV - Ultraviolet-Visible Spectroscopy. !! nm. Methods for structure determination of organic compounds:

Indiana's Academic Standards 2010 ICP Indiana's Academic Standards 2016 ICP. map) that describe the relationship acceleration, velocity and distance.

Lab 2. Spectrophotometric Measurement of Glucose

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

2 Absorbing Solar Energy

Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V.

Atomic and Nuclear Physics Laboratory (Physics 4780)

Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer. NAME: Lab Section: Date: Sign-Off:

COLORIMETER. Description 0358BT. Figure 1. The Colorimeter

HUMBOLDT-UNIVERSITÄT ZU BERLIN MATHEMATISCH-NATURWISSENSCHAFTLICHE FAKULTÄT I

Nucleic Acid Purity Assessment using A 260 /A 280 Ratios

Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July Reprinted from Photonics Spectra, Laurin Publishing

Measuring Manganese Concentration Using Spectrophotometry

University of California at Santa Cruz Electrical Engineering Department EE-145L: Properties of Materials Laboratory

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

P R E A M B L E. Facilitated workshop problems for class discussion (1.5 hours)

Title: Standard Operating Procedure for Thermo-Miran SapphIRe portable gas analyzer

Optical or Acoustic Process measurement of liquids Turbidity Colour Oil in Water Water in Oil Oil on Water Process UV- Photometry

Ces deux analyseurs permettent le contrôle de la qualité de l eau et des effluents.

Chemistry 111 Laboratory Experiment 7: Determination of Reaction Stoichiometry and Chemical Equilibrium

Flame Direct Attach UV-VIS Integrated Sampling System Installation and Operation Instructions

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

TOPIC 5 (cont.) RADIATION LAWS - Part 2

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs

APPLIED. Measuring Caffeine Concentration. Application Summary. Analytes: caffeine (C 8 H 10 N 4 O 2

4. Molecular spectroscopy. Basel, 2008

Energy. Mechanical Energy

Solar Energy. Outline. Solar radiation. What is light?-- Electromagnetic Radiation. Light - Electromagnetic wave spectrum. Electromagnetic Radiation

CHEM 161: Beer s Law and Analysis of a Sports Drink

Chapter 28: High-Performance Liquid Chromatography (HPLC)

Phenolphthalein-NaOH Kinetics

Transcription:

Info Note 804: Ultraviolet-visible spectroscopy or ultraviolet-visible spectrophotometry (UV/VIS) involves the spectroscopy of photons in the UV-visible region. It uses light in the visible and adjacent near ultraviolet (UV) and near infrared (NIR) ranges. In this region of the electromagnetic spectrum, molecules undergo electronic transitions. This technique is complementary to fluorescence spectroscopy, in that fluorescence deals with transitions from the excited state to the ground state, while absorption measures transitions from the ground state to the excited state. UV/VIS is based on absorbance. In spectroscopy, the absorbance A is defined as: (1) where I is the intensity of light at a specified wavelength λ that has passed through a sample (transmitted light intensity) and I 0 is the intensity of the light before it enters the sample or incident light intensity. Absorbance measurements are often carried out in analytical chemistry, since the absorbance of a sample is proportional to the thickness of the sample and the concentration of the absorbing species in the sample, in contrast to the transmittance I / I 0 of a sample, which varies exponentially with thickness and concentration. The Beer-Lambert law is used for concentration determination. The term absorption refers to the physical process of absorbing light, while absorbance refers to the mathematical quantity. Also, absorbance does not always measure absorption: if a given sample is, for example, a dispersion, part of the incident light will in fact be scattered by the dispersed particles, and not really absorbed. Absorbance only contemplates the ratio of transmitted light over incident light, not the mechanism by which light intensity decreases. Despite this fact, absorbance can still be used to determine concentrations (of particles) in such cases. Although absorbance does not have true units, it is quite often reported in "Absorbance " or AU. - 1 -

Optical density, or OD, is the absorbance per unit length, i.e., the absorbance divided by the thickness of the sample, although it is sometimes used as a synonym for the absorbance with a base-10 logarithm. Optical density is given as: (2) Where: L = the distance that light travels through the sample (the sample thickness), measured in cm A λ = the absorbance at wavelength λ T= the per-unit transmittance I 0 = the intensity of the incident light beam I= the intensity of the transmitted light beam Although absorbance does not have true units, it is quite often reported in "Absorbance " or AU. Accordingly, optical density is measured in ODU, which are equivalent to AU cm 1. The higher the optical density, the lower the transmittance. Optical density times 10 is equal to a transmission loss rate expressed in decibels per cm, e.g., an optical density of 0.3 corresponds to a transmission loss of 3 db per cm. Beer-Lambert law The method is most often used in a quantitative way to determine concentrations of an absorbing species in solution, using the Beer-Lambert law:, (3) where A is the measured absorbance, I 0 is the intensity of the incident light at a given wavelength, I is the transmitted intensity, L the pathlength through the sample, and c the concentration of the absorbing species. For each species and wavelength, ε is a constant known as the molar absorptivity or extinction coefficient. This constant is a fundamental - 2 -

molecular property in a given solvent, at a particular temperature and pressure, and has units of 1 / M * cm or often AU / M * cm. The absorbance and extinction ε are sometimes defined in terms of the natural logarithm instead of the base-10 logarithm. The Beer-Lambert Law is useful for characterizing many compounds but does not hold as a universal relationship for the concentration and absorption of all substances. Ultraviolet-visible spectrophotometer The instrument used in ultraviolet-visible spectroscopy is called a UV/VIS spectrophotometer. It measures the intensity of light passing through a sample (I), and compares it to the intensity of light before it passes through the sample (I o ). The ratio I / I o is called the transmittance, and is usually expressed as a percentage (%T). The absorbance, A, is based on the transmittance: A = log(%t) (4) The basic parts of a spectrophotometer are a light source (often an incandescent bulb for the visible wavelengths, or a deuterium arc lamp in the ultraviolet), a holder for the sample, a diffraction grating or monochromator to separate the different wavelengths of light, and a detector. The detector is typically a photodiode or a CCD. Photodiodes are used with monochromators, which filter the light so that only light of a single wavelength reaches the detector. Diffraction gratings are used with CCDs, which collects light of different wavelengths on different pixels. - 3 -

Diagram of a single-beam UV/vis spectrophotometer. A spectrophotometer can be either single beam or double beam. In a single beam instrument, all of the light passes through the sample cell. I o must be measured by removing the sample. This was the earliest design, but is still in common use in both teaching and industrial labs. Samples for UV/Vis spectrophotometry are most often liquids, although the absorbance of gases and even of solids can also be measured. Samples are typically placed in a transparent cell, known as a cuvette. Cuvettes are typically rectangular in shape, commonly with an internal width of 1 cm. (This width becomes the path length, L, in the Beer-Lambert law.) Test tubes can also be used as cuvettes in some instruments. The best cuvettes are made of high quality quartz, although glass or plastic cuvettes are common. (Glass and most plastics absorb in the UV, which limits their usefulness to visible wavelengths.) Instrument Measurement range Any real measuring instrument has a limited range over which it can accurately measure absorbance. An instrument must be calibrated and checked against known standards if the readings are to be trusted. Many instruments will become non-linear (fail to follow the Beer- Lambert law) starting at approximately 2 AU (~1% Transmission). It is also difficult to accurately - 4 -

measure very small absorbances that are very close to zero absorbance. The theoretical best accuracy for most instruments is in the range near 1 AU. When possible, the path length or concentration should be adjusted to achieve readings near this range. It is extremely difficult to achieve a large instrument range, between 0 and 6 AU. Ultraviolet-visible spectrum An ultraviolet-visible spectrum is essentially a graph of light absorbance versus wavelength in a range of ultraviolet or visible regions. Such a spectrum can often be produced directly by a more sophisticated spectrophotometer, or the data can be collected one wavelength at a time by simpler instruments. Wavelength is often represented by the symbol λ. Similarly, for a given substance, a standard graph of the extinction coefficient (ε) vs. wavelength (λ) may be made or used if one is already available. Such a standard graph would be effectively "concentration-corrected" and thus independent of concentration. For the given substance, the wavelength at which maximum absorbance in the spectrum occurs is called λ max, pronounced "Lambda-max". Applications in Gold Nanoparticles UV/Vis spectroscopy is routinely used in the quantitative determination of gold nanoparticle solutions. The Beer-Lambert law states that the absorbance of a solution is directly proportional to the solution's concentration. Thus UV/VIS spectroscopy can be used to determine the concentration of a solution. It is necessary to know how quickly the absorbance changes with concentration. This can be taken from references (tables of molar extinction coefficients), or more accurately, determined from a calibration curve. We use UV/VIS spectroscopy for characterization of our gold nanoparticles. With it we can determine size, concentration, binding events, and of course, optical absorption characteristics. - 5 -

- 6 -