Physiological Pressure Measurements

Similar documents
Bioinstrumentation. Kuo-Sheng Cheng, Ph.D. Department of Biomedical Engineering National Cheng Kung University

Hemodynamic Monitoring: Principles to Practice M. L. Cheatham, MD, FACS, FCCM

Unsteady Pressure Measurements

huidige en toekomstige hemodynamische bewaking

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Pressure Sensor Based Estimation of Pulse Transit Time

3 rd Russian-Bavarian Conference on Bio-Medical Engineering

Upper Arm Blood Pressure Monitor SC Instructions for use

Section 5.0 : Horn Physics. By Martin J. King, 6/29/08 Copyright 2008 by Martin J. King. All Rights Reserved.

Subminiature Load Cell Model 8417

AgoraLink Agora for Life Science Technologies Linköpings Universitet Kurs i Fysiologisk mätteknik Biofluidflöden

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Direct Arterial Blood Pressure Monitoring Angel M. Rivera CVT, VTS (ECC) Animal Emergency Center Glendale, WI March 2003

OEM MAXNIBP Frequently Asked Questions

Synthetic Sensing: Proximity / Distance Sensors

3. Test Methods for Evaluation of ESCR of Plastics

THE STRAIN GAGE PRESSURE TRANSDUCER

How do you adjust for IOP in a LASIK patient?

How to measure absolute pressure using piezoresistive sensing elements

Digital Systems Ribbon Cables I CMPE 650. Ribbon Cables A ribbon cable is any cable having multiple conductors bound together in a flat, wide strip.

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers

Vascular System The heart can be thought of 2 separate pumps from the right ventricle, blood is pumped at a low pressure to the lungs and then back

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions

INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS

Surface Profilometry as a tool to Measure Thin Film Stress, A Practical Approach. Gianni Franceschinis, RIT MicroE Graduate Student

Introduction to acoustic imaging

Characteristic and use

Endoscope Optics. Chapter Introduction

Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.

Fluid Mechanics: Static s Kinematics Dynamics Fluid

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

AS COMPETITION PAPER 2008

How to measure blood pressure using digital monitors

Soil Suction. Total Suction

Pressure. Pressure. Atmospheric pressure. Conceptual example 1: Blood pressure. Pressure is force per unit area:

Procon Engineering. Technical Document PELR TERMS and DEFINITIONS

A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW ASME Fluids Engineering Division Summer Meeting

The V15VxMR. The V15VxMR offers all these outstanding features. Trackability is the stylus' ability to stay in contact with

AP PHYSICS 2012 SCORING GUIDELINES

RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA

Spring Simple Harmonic Oscillator. Spring constant. Potential Energy stored in a Spring. Understanding oscillations. Understanding oscillations

Transmission Line and Back Loaded Horn Physics

Loudspeaker Parameters. D. G. Meyer School of Electrical & Computer Engineering

Ultrasonic Reference Blocks

Students Manual for the Exam. General Engineering and Electrical Civil Engineering Discipline

Sensors Collecting Manufacturing Process Data

Normal Intracardiac Pressures. Lancashire & South Cumbria Cardiac Network

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

SUSPENSION AND STEERING OVERVIEW

Single-Needle Cross-Over System. Brief description

Comparison of the Response of a Simple Structure to Single Axis and Multiple Axis Random Vibration Inputs

Acceleration levels of dropped objects

MEASURING AND RECORDING BLOOD PRESSURE

RLC Resonant Circuits

INTERFERENCE OF SOUND WAVES

APPLICATION NOTE AP050830

Specification of 5 Wire Analog Touch Panel

BASIC ELECTRONICS AC CIRCUIT ANALYSIS. December 2011

Time Domain and Frequency Domain Techniques For Multi Shaker Time Waveform Replication

Blood pressure measurement devices. December 2013

BIOMEDICAL ULTRASOUND

GMC 2013: Piping Misalignment and Vibration Related Fatigue Failures

Circuits with inductors and alternating currents. Chapter 20 #45, 46, 47, 49

Normal & Abnormal Intracardiac. Lancashire & South Cumbria Cardiac Network

Robot Perception Continued

Harmonic oscillations of spiral springs Springs linked in parallel and in series

Section Four: Pulmonary Artery Waveform Interpretation

Application Note for SDP600 and SDP1000 Series Measuring Differential Pressure and Air Volume with Sensirion s CMOSens technology

Paul Clements, SpR in Anaesthetics, Hope Hospital, Salford, UK. Carl Gwinnutt, Consultant Anaesthetist, Hope Hospital, Salford, UK.

Manufacturing Equipment Modeling

Experiment 1: SOUND. The equation used to describe a simple sinusoidal function that propagates in space is given by Y = A o sin(k(x v t))

Lecture 5 Hemodynamics. Description of fluid flow. The equation of continuity

Forces. Definition Friction Falling Objects Projectiles Newton s Laws of Motion Momentum Universal Forces Fluid Pressure Hydraulics Buoyancy

Lecture 24 - Surface tension, viscous flow, thermodynamics

Instruction Manual Service Program ULTRA-PROG-IR

E190Q Lecture 5 Autonomous Robot Navigation

Figure CPT Equipment

Fitting Guide. Dallas Denver Tulsa

Density (r) Chapter 10 Fluids. Pressure 1/13/2015

Signal to Noise Instrumental Excel Assignment

Manual Blood Pressure Measurement

METHOD STATEMENT HIGH STRIAN DYNAMIC TESTING OF PILE. Prepared by

Micro-Optical Sensor Use in Boundary Layer Flows with Polymers and Bubbles

Pre-lab Quiz/PHYS 224 Magnetic Force and Current Balance. Your name Lab section

Lecture L22-2D Rigid Body Dynamics: Work and Energy

Chapter 11 SERVO VALVES. Fluid Power Circuits and Controls, John S.Cundiff, 2001

MEASUREMENT OF END FACE GEOMETRY ON FIBER OPTIC TERMINI...2

Section 2.0 : Construction and Measurement of a Simple Test Transmission Line

The Next Generation of Cable Technology. A technology primer from NORDX/CDT By, Eric d Allmen

Section Two: Arterial Pressure Monitoring

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:

INTRODUCTION SOME USES OF SPRAY NOZZLES INTRODUCTION TYPES OF SPRAY NOZZLES

Module 13 : Measurements on Fiber Optic Systems

Circulation Stations

Building Design for Advanced Technology Instruments Sensitive to Acoustical Noise

GLOBAL COLLEGE OF ENGINEERING &TECHNOLOGY: YSR DIST. Unit VII Fiber Optics Engineering Physics

ROBOT SYSTEM FOR REMOVING ASBESTOS SPRAYED ON BEAMS

WOOD WEAR TESTING USING TRIBOMETER

Transcription:

Physiological Pressure Measurements TBMT09

today s lecture Literature: Lecture notes/slides! See also Togawa, Tamura, Öberg: Biomedical Transducers and Instruments Ask: Tryckmätning Blom ch 3

Lecture outline Indirect/direct pressure measurement methods Pressure transducers Pressure measurement system

The first invasive blood pressure measurement The first invasive attempt to measure blood pressure was made by Stephen Hales in 1733. From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970

Blood pressure levels Normal blood pressure Hypertension and Hypotension Narrow pressure windows

Pressure measurements Indirect methods Arm-occluding cuff Auscultation Palpation Oscillometric method Automatic methods Aplanation tonometry Direct methods Hydraulic systems Catheter tip systems

Indirect methods

Auscultatory blood pressure measurement Auscultatory blood pressure measurement using a pressure cuff and a stethoscope

Effects of height difference on the measured blood pressure From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997

The reference level in humans From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997

Cuff pressure measurements No sound No sound Fluttering sound Weak noise

Korotkoff sounds From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974

Oscillometric principle of pressure cuff Pressure detection Blood vessel Transducer arm cuff

Automatic Oscillometric blood pressure devices Might be unreliable due to geometry of the cuff, pressure detection algorithms and location of the cuff. Can be used for monitoring but diagnostic value of measurements is questionable

PPG and BP measurements Zero-balance technique with read light transmission (PPG) and a shaker driven pressure cuff keeping the light signal constant. The transducer signal will follow the time course of the blood pressure From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997

Zero-balance technique with light transmission and shaker driven pressure cuff The transmitted red light reflects the blood volume in the finger. The shaker applies a pressure to the finger cuff. By keeping the optical/volume signal constant by varying the shaker signal, the pressure in the cuff will follow instantaneous blood pressure changes

Maximal pulsations reflect maximally relieved blood vessels. The DC component of the PPG signal at this point is used as a set value of the feedback system From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997

Finapres

Micropipettes and BP measurements Zero-balance technique for measuring pressure in blood capillaries using glass micropipettes fill with KCl immersed in the blood. The impedance of the pipette is measured and kept constant by applying a counter pressure from a bellow to the pipette. From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974 The impedance differ very much depending on if the tip of the pipette is filled with high molar KCl or blood/physiological NaCl (se next page). By using a feedback system keeping the impedance constant the pressure applied to the pipette will follow that of the capillaries

Impedance of a micropipette Impedance versus counter pressure of a micropipette filled with 3M KCl and immersed in solutions of various concentrations. With a moderate concentration of this solution (as in the case of blood/physiological saline) there is dramatic impedance change around 5 mmhg making the impedance signal suitable as the control signal of the feedback system. From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974

Tonimetry for intraocular pressuremeasurements Tonimeter with eye microscope

When approaching a flat surface to the eye the cornea will be flattened (applanated). In principle the force on the flat surface is the intraocular pressure times the applanated area of the cornea. Knowing the area, the pressure can be measured from the force. Errors in this measurement are the surface adhesion of the tear fluid on one hand and the bending force of the cornea on the other. With the shown Mackay-Marg tonometer only the central (in this case 1.5 mm) of the flat surface is pressure sensitive. From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974 1. Pressure increases as surface flattens the cornea. 2. Force corresponding to sum of component from intraocular pressure and bending force. 3. Force reflecting intra ocular pressure only since bending force is outside the pressure sensitive surface

Goldman applanation tonimeter With this technique the cornea is applanated to a predetermined area (measured by a split image technique) where in average the forces due to corneal bending and tear fluid adhesion should cancel From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997

Non-contact applanation tonimeter The pneumatic system creates an air jet flow velocity which is controlled by a piston pump. The jet creates a stagnation pressure that displaces/ flattens the corneal surface. The flattened surface is detected by an optical system. Knowing jet velocity and the flattened area the intraocular pressure can be estimated. From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997

Direct methods

The pressure signal One have to measure not only the values of systolic and diastolic pressure but the whole accurate time course of the signal. Requires: 1. Accuracy 2. Bandwidth

Historic background Direct pressure measurement by Fick 1864 From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970

Intra-cardiac measurement First intracardiac measurements by Marey & Chauveau 1860 s From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970

Example of ancient recording Recordings by Marey &Chauveau 1861 From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970

Landois hematogram 1872 From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970

Early pressure transducer Pressure transducer with a dome for connecting catheters and a metal membrane linked to strain-gauge sensing elements (Statham 1940`s) From LA Geddes: The direct and indirect measurement of blood pressure Year Book Medical Publ 1970

Disposable pressure transducer From: T Togawa, T Tamura, PÅ Öberg: Biomedical transducers and instruments, CRC Press, 1997

Clinical pressure transducers Modern pressure transducers with a dome (above) and one where the sensing element is included in the Luer tubing (blow)

Technical development of pressure transducers From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974

Illustration of the piezo-resistivity of silicone. Plotting resistance change as a function of material strain p- and n-type have the opposite characteristics. p-type shows a more linear behavior and is therefore preferred From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974

p-type piezo-resistive elements diffused into silicone material for various geometries Membrane Rod/cantilever From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974

Catheter and Transducer Changing pressure, p Catheter compliance C radius, r Transducer compliance, Co Displacement of the surface and Volume displacment Mass of the fluid Fluid resistance Damping

Hydraulic coupled system From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974

Resistance, inertance, compliance The flow resistance is given by: 8ηl R = π 4 r The Inertance is given by: L ρl = 2 π r The compliance is given by: C C = + 2 0 C

Pressure transfer function The pressure transfer function: which can be written as a second order transfer function: 0 0 0 1 1 ) ( ) ( sc R sl sc s p s p i + + = 2 0 0 2 2 0 0 2 ) ( ) ( ω ω ω + + = s D s s p s p i

Resonance frequency Identification gives us the resonance frequency: f 0 = 2π 1 LC 0 = 2 1 πρ r lc 0 The resonance frequency is proportional to the radius and inversely proportional to the square root of the catheter length and compliance.

Damping The damping is obtained as: D = R C0 4η lc = 3 2 L πρ r 0 The damping depends on the square root of the catheter length and compliance and inversely to the radius to the power of three.

Step response of a second order system The response of a second order system as a function of the damping

Example under-damped system Correctly measured aortic pressure (upper trace) The same signal measured with un under-damped low band-width hydraulic system (lower trace)

Areas of damping and resonance frequency resulting in an accurate recording of the blood pressure signal

Hydraulic parallel damping of a pressure measurement system

Importance of pressure sensor placements 1 P = P + ρv 2 1 P = P ρv 2 P = P 2 2 Depending on the direction of the catheter tip, the measured pressure can consist of both static and dynamic terms. (according to the Bernoulli equation)

A catheter-tip pressure transducer The transducer element is etched in silicon to a pressure sensing membrane

Transducer element A detailed view of the transducer element etched in silicon forming a membrane

Fiber optic pressure transducer From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974

Example of optic sensor

Principles of a fiber optic pressure transducer From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974

Fiber optic pressure transducer From: RC Cobbold. Transducers for biomedical measurements, John Wiley, 1974

Fiber optic pressure transducer The principle behind how a fiber optic pressure transducer works. A fiber (1) emits light within a cone in a certain angle. The other fiber (2) receives light within a similar cone. The overlap between these cones is larger the larger the distance is between the fibers and the reflecting metal membrane. Pressure The surface is close pressure on membrane is high Pressure Distance is larger pressure on membrane is smaller

Fiber optic miniature pressure transducer If the glass-beam is displaced downwards it approaches a glossy silicone corner surface and more of the light is reflected.