Routine quality control tests for fullfield digital mammography systems Equipment report 1303: fourth edition

Similar documents
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

MQSA Quality Control Manual for Monochrome Displays for Mammography

Digital Breast Tomosynthesis QC Requirements

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Quality Control and Maintenance Programs

INTRODUCTION. A. Purpose

Quality Control of Full Field Digital Mammography Units

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Performance testing for Precision 500D Classical R/F System

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

THE ROYAL AUSTRALIAN AND NEW ZEALAND COLLEGE OF RADIOLOGISTS

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Update on ACR Digital Mammography QC Manual

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

Checklist for the Qualification of Digital Detector Array Systems

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

DICOM Correction Item

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Physics testing of image detectors

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Measuring Size and Distance in Digital Mammography

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

The Exposure Index and Its Standardization

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Medical imaging monitors specification guidelines

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

Aspire FCRm/CRm Fujifilm CR for Mammography

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices

Lecture 14. Point Spread Function (PSF)

How To Track Life Expectancy In England

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Scanners and How to Use Them

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

CONSOLIDATED GUIDANCE ON STANDARDS FOR THE NHS BREAST SCREENING PROGRAMME

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

Geometric Optics Converging Lenses and Mirrors Physics Lab IV

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Display screen equipment (DSE) workstation checklist

Mammography. What is Mammography?

Troubleshooting Guide for Digital Printing on Creative Papers

Coronis 5MP Mammo. The standard of care for digital mammography

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

Guide To Creating Academic Posters Using Microsoft PowerPoint 2010

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

EPSON SCANNING TIPS AND TROUBLESHOOTING GUIDE Epson Perfection 3170 Scanner

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Digital Mammography Update: Design and Characteristics of Current Systems

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

WHITE PAPER. Methods for Measuring Flat Panel Display Defects and Mura as Correlated to Human Visual Perception

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

American College of Radiology Mammography Accreditation Program

Holes & Selective Laser Sintering

Quality Assurance Guidance

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014

Protocol for Microscope Calibration

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

Digital Mammogram National Database

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

Data Processing Flow Chart

Fundamentals of Cone-Beam CT Imaging

Thin Lenses Drawing Ray Diagrams

CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems

Digital Image Requirements for New Online US Visa Application

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION

Certification protocol for breast screening and breast diagnostic services

Protocol for ordering, storing and handling vaccines. March 2014

How To Improve Your Ct Image Quality

Mammography Applications for MAMMOMAT 1000/3000 Nova

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

Digitization of Old Maps Using Deskan Express 5.0

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Purchasing a cardiac CT scanner: What the radiologist needs to know

Transcription:

Routine quality control tests for fullfield digital mammography systems Equipment report 1303: fourth edition October 2013

About the NHS Cancer Screening Programmes The national office of the NHS Cancer Screening Programmes is operated by Public Health England. Its role is to provide national management, coordination, and quality assurance of the three cancer screening programmes for breast, cervical, and bowel cancer. About Public Health England Public Health England s mission is to protect and improve the nation s health and to address inequalities through working with national and local government, the NHS, industry and the voluntary and community sector. PHE is an operationally autonomous executive agency of the Department of Health. www.gov.uk/phe Lead authors: Gill Baxter Vivienne Jones Vivien Milnes Jenny Oduko Vivien Phillips Sarah Sellars Zoe Vegnuti Crown copyright 2013 PHE publications gateway number: You may re-use this information (excluding logos) free of charge in any format or medium, under the terms of the Open Government Licence v2.0. To view this licence, visit OGL or email psi@nationalarchives.gsi.gov.uk. Where we have identified any third party copyright information you will need to obtain permission from the copyright holders concerned. Any enquiries regarding this publication should be sent to sarah.sellars@phe.gov.uk 2

Document lnformation Title Routine quality control tests for fullfield digital mammography systems Policy/document type Equipment report 1303 Electronic publication date October 2013 Version 4 Superseded publications Versions 1-3 Review date None Author/s Owner Document objective (clinical/healthcare/social questions covered) Population affected Target audience Gill Baxter Vivienne Jones Vivien Milnes Jenny Oduko Vivienne Phillips Sarah Sellars Zoe Vegnuti Comments may be sent to Sarah Sellars, sarah.sellars@phe.gov.uk in readiness for review. Updates previous guidance, removing a section on CR and updating tests to reflect current experience and new guidance. Women screened and assessed within the NHSBSP Radiographers, Assistant Practitioners and Physicists working in the NHSBSP 3

Contents Contents 4 Executive summary 5 1. Introduction 6 2. Monitor checks 8 2.1 Daily checks on acquisition and reporting monitors 8 2.2 Monthly test of acquisition and reporting monitors 8 2.3 Optional monthly test 9 3. System checks with Perspex blocks 10 3.1 Daily system check 10 3.2 Weekly check of contrast-to-noise ratio (CNR) 11 3.3 Weekly artefact and uniformity check 12 3.4 Monthly AEC thickness check 14 4. Weekly image quality tests 16 4.1 Method for weekly image quality test 16 5. Detector flat-field calibration 17 6. Monthly mechanical safety and function tests 18 7. Analysis of repeat images 19 8. Printer checks 20 8.1 Daily printer checks using test pattern 20 8.2 Printer checks following a software upgrade 20 9. Tests after mobile units are moved 21 10. Tests after an engineer s visit 21 11. Tests to be performed after software changes to the imaging chain, including rulers and callipers 21 12. Small field digital mammography systems used for biopsy, including stereo 22 13. Specimen cabinets 22 Appendix 1: Examples of artefacts 23 Appendix 2: Design of test object 24 Appendix 3: Examples of SNR and CNR calculations 25 Appendix 4: Example data collection forms 27 4

Executive summary This document, Routine quality control tests for full field digital mammography systems, offers guidance for the quality control processes that ensure that all mammography equipment meets NHSBSP standards. It is designed to complement European and national NHSBSP guidance documents, as well as the guidance from manufacturers, to provide a full range of tests and checks and to ensure that equipment is working within agreed standards and parameters and functioning safely. The tests covered comprise: monitor checks system checks with Perspex blocks weekly image quality tests detector flat-field calibration monthly mechanical safety and function tests analysis of repeat images printer checks tests to be run after mobile units are moved tests to be run after an engineer s visit tests to be run after a software change tests for small field digital mammography systems used for biopsy tests for specimen cabinets 5

1. Introduction Following a recommendation in the document Improving Outcomes: A Strategy for Cancer 1, direct digital (DR) mammography systems are being used for routine mammography in the NHS Breast Screening Programme (NHSBSP). Routine quality control is essential to ensure that the equipment meets NHSBSP standards and is performing as expected. This guide describes the recommended routine QC tests that should be undertaken by radiographic staff. It was originally based on the European protocol for the quality control of the physical and technical aspects of mammography screening 2, but also incorporates current knowledge and understanding of digital systems. Guidance on commissioning and the routine testing by physicists of full field digital mammography systems is given in NHSBSP Equipment Report 0604, version 3. 3 Baseline values need to be established at installation in conjunction with the local physicist i and must be re-established if conditions are changed. For many tests it does not matter whether unprocessed ii images or processed (clinical) images are used. Manufacturers and the local physicist should be consulted as to which type of image is appropriate for each individual piece of equipment. The same type of image should be used each time a test is carried out. Manufacturers tests can be used as long as they are equivalent to those described in this document. Local physicists must be consulted for advice, and test protocols agreed and accepted after commissioning. The routine tests for the calibration or maintenance of systems recommended by the manufacturer of the X-ray set, workstation, or printer should be added to the local test protocol. Some systems have built-in tests for the detector and/or display, and it is hoped that manufacturers will continue to develop such automated QC systems. The local physicist should be asked to advise whether built-in tests are suitable for use in place of the tests described in this document. The results of all tests must be recorded on paper, or electronically on spreadsheets (the latter is preferable) to facilitate data analysis and auditing. If results (for example baseline images) are kept on the X-ray set then these are likely to be removed when i This is the person involved in provision of medical physics expert advice to the specific NHSBSP screening centre. ii Different manufacturers have different terms for unprocessed images e.g. raw (GE), QC-raw (Siemens), flatfield (Hologic). 6

software is upgraded. Relevant images iii must therefore be stored securely for a minimum of 8 months. Baseline images iv must be stored permanently. All quantitative and qualitative data generated by routine tests and observations should be recorded. The routine QC tests for DR systems are summarised in Table 1. This guidance should be used in conjunction with Routine Quality Assurance guidelines for mammography (NHSBSP Publication no 63). 4 Screening Programmes must consider the interface intercompatibility of the different parts of the imaging chain e.g. workstations and modalities from different suppliers. Table 1 Recommended routine QC tests for DR systems Frequency Test Section Daily Checks on acquisition and reporting 2.1 monitors Daily System check 3.1 Daily Printer checks using test pattern 8.1 Weekly Check of contrast-to-noise ratio 3.2 Weekly Image quality tests 4 Weekly Artefact and uniformity check 3.3 Monthly AEC thickness check 3.4 Monthly Test of acquisition and reporting monitors 2.2 Monthly Mechanical safety and function checks 6 As Detector flat-field calibration 5 required As Repeat analysis 7 required As Printer checks following software upgrade 8.2 required As Check after mobile unit moves 9 required As Check after engineer s visit 10 required As required Check after software or any other changes to the imaging chain including rulers/callipers 11 iii This should be decided in consultation with the Superintendent and local physicist. iv These are test images taken by the mammographer when the system was first commissioned or when the baselines are reset after significant changes 7

2. Monitor checks Monitor checks should be performed on both the acquisition and reporting v monitors under recommended vi working conditions, as agreed with the local physicist. 2.1 Daily checks on acquisition and reporting monitors The following method must be followed: check for obvious faults such as dirty screens, artefacts (see Appendix 1) for CRT monitors only: check for flicker, distortion, and whether text and lines on the screen are sharp and straight check general condition clean if necessary (follow the supplier s instructions) An additional optional test is to display a test pattern or standard clinical mammogram and check its appearance. A record must be kept of all checks. Problems must be noted, and action taken to correct them. 2.2 Monthly test of acquisition and reporting monitors The test uses the TG18-QC test pattern 5 shown in Figure 1 (this is the preferred pattern). Refer to the supplier or local physicist for advice on how to display this. The following method must be followed: check that the room brightness is as recommended, with no glare from other monitors, light boxes, or windows clean the monitors (follow the supplier s instructions) display the TG18-QC pattern on each monitor in turn examine the image carefully under working conditions, and check that: o there are no significant reflections on the monitor o borders are completely visible o lines are straight o the active display area is centred on the screen o the 5% square is visible within the larger 0% square (area A) o the 95% square is visible within the larger 100% square (area B) v The term acquisition monitor is used here to denote a monitor used by the mammographer when performing the mammogram (also known as a review or secondary monitor). The term reporting monitor is used to denote a monitor in the workstation used by the film reader when reporting the mammogram (also known as a diagnostic or primary monitor). vi Ambient light should be less than 20 lux (LCD monitors) or 10 lux (CRT monitors) for primary display devices. 6 8

o each grey scale step from 0% to 100% can be distinguished from the adjacent squares (see dotted arrows) o the text on the pattern is sharp and in focus record the results for each monitor Area A Area B Figure 1 TG18-QC test pattern 2.2.1 Remedial level If the system fails any of the above checks then take action to correct the problem. 2.3 Optional monthly test Send a TOR(MAM) image (or equivalent image quality phantom image) to all workstations to which clinical images are sent from the X-ray unit being tested. Score the image on all monitors at each workstation (see section 4). These images should be stored to enable future review. 2.3.1 Remedial level Significant variation between TOR(MAM) scores on different workstations (this would be decided locally) or significant deterioration of scores on all workstations from previous months values. 9

3. System checks with Perspex blocks These tests will detect changes in the performance of the X-ray set or the image receptor. See Appendix 2 for suggested details of the test object design. 3.1 Daily system check The following method must be followed: position the test object on the unit (see Appendix 2). Always use a dedicated QC paddle. Add further thicknesses of Perspex if necessary to give a total thickness of 4.0 or 4.5 cm compress to a consistent compression force or thickness (chosen so as to give the same kv, target and filter every day in normal circumstances). If using thickness, compression should not be greater than a set value, for example not more than 60 newtons, to avoid possible damage to the plastic paddle select either a processed or an unprocessed image; the same type of image should always be used expose using automatic exposure control record post-exposure factors (kv, target/filter, mas) record an indicator of dose to detector (e.g. displayed dose, mean pixel value) inspect the image for artefacts and variations in the noise pattern using a narrow window width, for example of about 20% of window level. (See Appendix 1 for a discussion of the type of artefacts that may be seen) draw a ROI as shown in Figure 2 and record the mean (M) and standard deviation (SD) of the pixel value, or the SNR as given by the system divide M by SD to calculate the signal-to-noise ratio (SNR) as in Equation 1 (see Appendix 3 for an example of the calculation): SNR M SD (1) 10

Mid-line Al square ROI Chest wall edge Consistent fixed distance Figure 2 ROI for SNR measurement. 3.1.1 Remedial level mas baseline ± 10% (provided kv and target/filter are the same as for the baseline measurement) Detector dose indicator baseline ± 10% SNR baseline ± 10% If any of the above levels are exceeded then action must be taken to correct the problem. If the problem persists after checking, action must be taken, in line with local protocols, before the equipment is put back in use. 3.2 Weekly check of contrast-to-noise ratio (CNR) The following method must be followed: use the image of the test object from the daily test (see 3.1) draw two ROIs, as shown in Figure 3 record the mean (M1) and standard deviation (SD) of the pixel value in ROI 1 record the mean (M2) of the pixel value in ROI 2 subtract M2 from M1 and divide by the standard deviation of the pixel value (SD) to calculate the contrast-to-noise ratio (CNR) as shown in Equation 2 (see Appendix 3 for an example of the calculation) CNR M1 M2 SD (2) 11

Mid-line Al square ROI 1 Chest wall edge ROI 2 Consistent fixed distance Figure 3 ROIs for CNR measurement 3.2.1 Remedial level CNR baseline ± 10% If this level is exceeded then action must be taken to correct the problem. If the problem persists after checking, action must be taken, in line with local protocols, before the equipment is put back in use. 3.3 Weekly artefact and uniformity check Uniformity should be visually checked on a weekly basis using an image of the maximum field size, with Perspex covering the whole field. Before testing, clean any dust from the top surface of the paddle, the breast table and the Perspex. Set a narrow window width of about 20% of window level to show up any areas of nonuniformity. Magnify or zoom the image electronically and inspect it in a systematic fashion to look for artefacts such as faulty clusters of pixels or areas of unusually low noise (where the background mottle appears blurred or smoother than other areas of the image). Repeat the uniformity check for all other target/filter combinations used clinically, in case there is any debris in the system or damage to one of the filters. The following method must be followed: position the test object on the unit compress to a consistent thickness or compression (if exposing under AEC) or use a Perspex block at the tube head 12

select either a processed or unprocessed image, the same image type should always be used expose using automatic exposure control, or manual exposure if following the manufacturer s or local physicist s recommendations record post-exposure factors (kv, target/filter, mas). Inspect for uniformity and artefacts. Set a narrow window width of about 20% of window level to show up any areas of non-uniformity. Magnify or zoom the image electronically and inspect it in a systematic fashion to look for artefacts, such as faulty clusters of pixels or areas of unusually low noise (where the background mottle appears blurred or smoother than other areas of the image) expose and record exposure factors for alternative target/filter combinations and inspect the images (manual exposures may be required in order to test the range of combinations available) for the first image only, draw ROIs as shown in Figure 4 record the mean pixel value (Mcentre) in the central ROI record the mean pixel values in the other ROIs and find the one that is most different from Mcentre call this Medge calculate the maximum percentage deviation from the central value using Equation 3: MaxDev Mcentre Medge 100 Mcentre (3) ROI ROI Midway between chest wall and nipple edges Mid-line Central ROI (avoiding aluminium square) Chest wall edge ROI ROI Figure 4 ROIs for uniformity check 13

3.3.1 Remedial level >10% maximum deviation from the value at the centre or value specified by the manufacturer. If an artefact is observed, first check the Perspex, breast table, grid and paddle for marks, scratches or debris then repeat the test with the test object in a different orientation, if possible. If the artefact remains visible then seek further advice from the manufacturer or local physicist. 3.3.2 Suspension level If no improvements are achievable then this should be discussed with the local physicist 3.4 Monthly AEC thickness check This should be checked with at least three thicknesses of Perspex, covering the clinical range used and including the test object described in Appendix 2. The range of thicknesses used must ensure that all clinically used target/filters are tested. Ensure that the aluminium square is always positioned at the same height. The following method must be followed: position the Perspex blocks on the unit compress to a consistent force or thickness expose using automatic exposure control record post-exposure factors (kv, target/filter, mas) record an indicator of dose to the detector (e.g. displayed dose, mean pixel value) inspect image for artefacts and variations in the noise pattern using a narrow window setting (high contrast), see Appendix 1 measure SNR and compare with baseline values (see section 3.1) measure CNR and compare with baseline values (see section 3.2) repeat for the other thicknesses 3.4.1 Remedial level mas Baseline for that thickness ± 10% (provided kv and target/filter are the same as for the baseline measurement) Detector dose indicator Baseline for that thickness ± 10% SNR Baseline for that thickness ± 10% CNR Baseline for that thickness ± 10% 14

No disturbing artefacts should be visible. If any of the levels are exceeded, action should be taken to correct the problem. If the problem persists after checking, action must be taken, in line with local protocols, before the equipment is put back in use. 15

4. Weekly image quality tests 4.1 Method for weekly image quality test A suitable test object, such as the TOR (MAM) on 3 cm of Perspex, should be used under AEC control. Since this test object has features similar to a breast, a clinical image processing algorithm should be used. The TOR (MAX) and TOR (MAS) are not suitable for this test. The following method must be followed: compress to a consistent force or thickness record post-exposure factors (target, filter, mas, kv) record an indicator of dose to detector send the image to a designated reporting workstation (the same workstation should always be used) examine each image, using standard window width and level and image processing, and view under standard conditions. Ensure that the image is displayed at the same size that it would be displayed at clinically score according to the agreed documented instructions When evaluating the results, inter-observer variations should be taken into account. 4.1.2 Remedial level Significant change from baseline (this should be decided locally). 16

5. Detector flat-field calibration Some DR detectors may have a non-uniform response (for example, due to variations in sensitivity, or faulty pixels). Also, there are non-uniformities in the X-ray beam due to the anode heel effect and X-ray beam divergence. DR systems correct for these inherent non-uniformities by a process of flat-fielding. Flat-field correction maps are obtained using a standard beam attenuator (usually a Perspex block) for one or more exposure conditions (e.g. different target/filter combinations and focal spot sizes). Some systems require the user to carry out this flat-fielding process periodically, and it is therefore included here, although it is not strictly a QC test. On other systems the check is carried out by the service engineer at routine service visits. The following method must be followed: carry out the flat field calibration according to the manufacturer s protocols. Ensure that the top surface of the paddle is clean to avoid incorporating the image of any dense particles in the calibration record and initial that the procedure has been performed 17

6. Monthly mechanical safety and function tests The safety and function of the system must be checked on a monthly basis. It is recommended that a local checklist is drawn up for each system to identify relevant features to be checked (for example items that are safety-critical, or areas known to be prone to faults). This should be based on the guidance in Quality assurance guidelines for mammography including radiographic control (NHSBSP Publication number 63) 4, plus additional items specific to the local system, for example: environmental checks (some digital systems are particularly sensitive to environmental conditions, such as temperature and humidity) checks relating to the reporting workstation (ergonomics) Keep a record of all checks carried out, and note any problems and the action taken to have them corrected. 18

7. Analysis of repeat images Early experience of changing from analogue to digital has shown that the number of repeat images may increase initially. A log of all repeat examinations must therefore be kept and regularly audited. The design of the PACS/digital systems should allow for repeat and reject analysis. Relevant repeat and recall data must be collected and input onto the breast screening IT system (NBSS). NHSBSP guidance on collecting, monitoring and reporting repeat examinations must be followed. 7 The NHSBSP minimum standard for repeat and recall examinations is < 3% of total examinations. 19

8. Printer checks When a printer is installed, the installation engineer should ensure that the hard copy matches the soft copy image. Hard copy quality can be checked subjectively by using a standard mammography test object, such as TOR(MAM). 8.1 Daily printer checks using test pattern On each day that a printer is used, a printer check must be carried out using standard viewing conditions. Print the TG18-QC test pattern (see section 2.2) or the manufacturer-supplied test pattern and perform the following checks: geometrical distortion: check that the image is printed without geometrical distortion; the borders should be completely visible and straight lines should be straight contrast visibility: in the TG18-QC test pattern, the 5% and 95% squares should be clearly visible printer artefacts: check the test pattern for printer artefacts (see Appendix 1); no disturbing artefacts should be visible if a densitometer is available, measure densities and compare with baseline values. 8.2 Printer checks following a software upgrade After software changes or an upgrade, it may be advisable to print both a test pattern and a clinical image to confirm that the hard copy remains similar to the soft copy display. 20

9. Tests after mobile units are moved When mobiles are moved to new sites appropriate tests should be performed on the mammography equipment, as agreed with the local physicist. 10. Tests after an engineer s visit There should be clear handover procedures to follow after an engineer has performed routine preventative maintenance or as a call-out. Appropriate tests should be performed, as agreed with the local physicist, after an engineer s visit or changes to any parts of the imaging chain (including software), e.g. workstations. 11. Tests to be performed after software changes to the imaging chain, including rulers and callipers Relevant checks, including displayed image quality and the accuracy of rulers/callipers, must be undertaken at commissioning and at appropriate intervals thereafter e.g. change of software, changes of monitors. These checks should be made in both contact and magnification modes. Further advice will be sought on standard testing and this document will be updated in due course. 21

12. Small field digital mammography systems used for biopsy, including stereo There are some small field systems in use in the NHSBSP. The medical physics service should perform tests on small field digital systems on a six-monthly basis and whenever a new digital detector is installed, according to the guidance provided in Commissioning and routine testing of small field digital mammography systems (NHSBSP Report 01/09). 8 The tests that must be run include checks of the detector, the display monitor, and the hard copy system as for full field systems. It is recommended that local protocols are developed for more frequent testing by radiographic staff, in conjunction with the local physicist and taking account of the equipment manufacturer s recommendations. The tests that are appropriate will depend on the system, but prone table systems require the full QC protocol. These system tests should be agreed with the local physicist. Where a lateral arm facility is used, testing should be undertaken according to the equipment manufacturer s guidance and the recommendations for full field digital testing. Needle positioning accuracy should be tested according to Quality assurance guidelines for mammography including radiographic quality control (NHSBSP Publication 63) 4 and all of the recommended tests for full field digital attachments should be followed. 13. Specimen cabinets These should be set up and calibrated as per the manufacturer s instructions to ensure accuracy of images. An occasional check that calcifications are clearly visible must be recorded. 22

Appendix 1: Examples of artefacts To inspect an image of a uniform test object on a monitor, adjust the window width (WW) to about 20% of the window level (WL) and use magnification if required. Detector artefacts: DR systems faulty individual pixels, clusters of pixels or lines of pixels may be observed. They may be always black, always white, or randomly fluctuating, and they may or may not disappear after flat fielding. Their significance depends on how many pixels are involved and where they are located. This should be discussed with the local physicist in scanning-type DR systems, faulty pixels may give rise to linear artefacts perpendicular to the scan direction loss of resolution (blurring) may occur in one or both directions, in one part of the detector or all over. This may be seen as a subtle change in the background noise pattern. The image will appear smoother (less noisy) where it is blurred ghosting/image retention is characterised by a faint image of a previously imaged breast or test object, or of collimation to the smaller field size when using the whole detector fine lines may appear, close together, on part of the image Monitor artefacts: cathode ray tube (CRT) distortion may occur, possibly due to interference from other electrical devices note that some monitors have one or more fine horizontal black lines, which form part of the monitor s calibration system. These are not artefacts Monitor artefacts: flat panel display (FPD) faulty pixels may appear Printer artefacts if an artefact is only seen on hard copy, then it is caused by the printer banding and streaking may occur a fine line in the direction of film travel may be seen pick off (damage to the film) 23

Appendix 2: Design of test object Perspex block(s) are used, with 0.2 mm of aluminium foil to provide contrast. Blocks of size 18 x 24 cm are suitable; blocks of other sizes or shapes may be used if the local physicist confirms that they provide suitable coverage of the AECs of all the X- ray sets in use. A suitably sized piece of aluminium should be used, with one edge on the midline. This needs to be fixed to the Perspex to prevent movement, but uneven layers of glue must be avoided. The height of the aluminium above the breast platform is not critical, and it can be sandwiched between Perspex layers. Mid-line Aluminium square Chest wall edge Consistent fixed distance 24

Appendix 3: Examples of SNR and CNR calculations The SNR and CNR are usually calculated in a spreadsheet. SNR example Mid-line Al square ROI Chest wall edge Consistent fixed distance Baseline value of SNR is 50 for this example. 10% of 50 is 5. Lower remedial level = 50-5 = 45 Upper remedial level = 50+5 = 55 Measurement 1: Mean pixel value in ROI = M = 397.1 Standard deviation of pixel values in ROI = SD = 8.1 SNR M SD 397.1 49.0 8.1 This is between the lower and upper remedial levels, so this measurement passes the test. Measurement 2: Mean pixel value in ROI = M = 405.2 Standard deviation of pixels values in ROI = SD = 6.3 SNR M SD 405.2 64.3 6.3 This is above the upper remedial level, so this measurement fails the test. 25

CNR example Mid-line Al square ROI 1 Chest wall edge ROI 2 Consistent fixed distance Baseline value of CNR is 7.7 for this example. 10% of 7.7 is 0.77. Lower remedial level = 7.7-0.77 = 6.93 Upper remedial level = 7.7+0.77 = 8.47 Measurement 1: Mean pixel value in ROI 1 (Perspex) = M1 = 386.7 Standard deviation of pixel values in ROI 1 = SD = 7.5 Mean pixel value in ROI 2 (Perspex plus aluminium) = M2 = 328.3 CNR M1 -M2 SD 386.7-328.3 7.5 58.4 7.8 7.5 This is between the lower and upper remedial levels, so this measurement passes the test. Measurement 2: Mean pixel value in ROI 1 (Perspex) = M1 = 386.7 Standard deviation of pixel values in ROI 1 = SD = 7.5 Mean pixel value in ROI 2 (Perspex plus aluminium) = M2 = 341.2 CNR M1 -M2 SD 386.7-341.2 7.5 45.5 7.5 6.1 This is below the lower remedial level, so this measurement fails the test. 26

Appendix 4: Example data collection forms Ideally the QC results should be directly entered into spreadsheets, which facilitates error-free calculations, enables automatic checks against tolerances and can provide graphical display of the data to assist with trend analysis. Acceptable ranges could be entered into the spread sheet to help with analysis; also the compression force to be used could be recorded. Example results sheets follow. Daily and weekly system checks with Perspex blocks (sections 3.1, 3.2) Unit/Location, Model, Identifying Code Perspex (cm) Perspex size (cm) Compression paddle Compression force Exposure mode Processing Daily Weekly Date Detector temp ( C) kv Target/ Filter mas Pixel value (PV1) Standard deviation (sd1) SNR (PV1 / sd1) Pixel value (PV2) CNR (PV1-PV2) / sd1 Initial Any visible artefacts? Comments 27

Example results charts from sample datasets, including limiting tolerance 125 mas 120 115 110 105 100 95 90 85 80 04 Oct 11 24 Oct 11 13 Nov 11 03 Dec 11 23 Dec 11 12 Jan 12 650 Pixel value (PV1) 600 550 500 450 400 04 Oct 11 24 Oct 11 13 Nov 11 03 Dec 11 23 Dec 11 12 Jan 12 SNR 85 80 75 70 65 60 55 50 45 40 04 Oct 11 24 Oct 11 13 Nov 11 03 Dec 11 23 Dec 11 12 Jan 12 Pixel value (PV1) 700 600 500 400 300 200 100 0 04 Oct 11 24 Oct 11 13 Nov 11 03 Dec 11 23 Dec 11 12 Jan 12 28

Example results charts for weekly uniformity test (section 3.3) Unit/Location, Machine, Identifying Code C 1 2 Perspex (cm) Perspex size (cm) Compression paddle Compression force 24x30 Exposure mode Processing W E 3 4 5 ROIs Mean pixel value Date kv TargetFilter mas ROI 1 ROI 2 Centre (ROI 3) ROI 4 ROI 5 Any visible artefacts? Comments Initial Max deviation from centre (Limit 10%) Acceptable? 29

Monthly AEC thickness check (section 3.4) Unit/Location, Machine, Identifying Code Add range of tolerances 2 cm 4.5 cm 7 cm Date k V Target -filter ma s PV 1 sd 1 PV 2 SN R CN R k V Target -filter ma s PV 1 sd 1 PV 2 SN R CN R k V Target -filter ma s PV 1 sd 1 PV 2 SN R CN R Initia l Comment 30

References 1. Improving Outcomes: A Strategy for Cancer. London: Department of Health, January 2011 2. European protocol for the quality control of the physical and technical aspects of mammography screening. Part 2b: Digital mammography. In: European guidelines for quality assurance in breast cancer screening and diagnosis, 4th edition. Luxembourg: European Commission, 2006. 3. Commissioning and routine testing of full field digital mammography systems (NHSBSP Equipment Report 0604, Version 3). Sheffield: NHS Cancer Screening Programmes, 2010. 4. Quality assurance guidelines for mammography including radiographic quality control (NHSBSP Publication No. 63). Sheffield: NHS Cancer Screening Programmes, 2006. 5. Assessment of display performance for medical imaging systems. American Association of Physicists in Medicine (AAPM). Online report 03, 2005 (available at www.aapm.org/pubs/reports/or_03.pdf). 6. Supplement to the European Guidelines 4 th Edition. Nijmegen: EUREF, 2011 7. Collecting, monitoring and reporting repeat examinations (NHSBSP Good Practice Guide No 4, Version 2). Sheffield: NHS Cancer Screening Programmes 2006 8. Commissioning and routine testing of small field digital mammography systems (NHSBSP Equipment Report 0705). Sheffield: NHS Cancer Screening Programmes, 2007 31