Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Size: px
Start display at page:

Download "Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography"

Transcription

1 Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error, off-center error, offfocus error, and upside-down error on grid cutoff; as well as, how each affects radiographic brightness, visibility of detail, and radiographic quality. Learning Objectives After completing this lab, you should be able to: 1. Use the laboratory equipment properly. 2. Set up the control console and ceiling tube mount correctly. 3. Function effectively in group work. 4. Perform the experiment independently. 5. Evaluate the effects of grid positioning on radiographic density. 6. Describe the relationship between grid positioning and visibility of detail. 7. Summarize the grid positioning error s relationships to radiographic quality. 8. Predict the effect of changes in grid position on radiographic density and radiographic quality. Materials Needed Five 14 x 17 inch 200 RS analog image receptors Five 14 x 17 inch computed radiography (CR) image receptors Focused grid (12:1 grid ratio, inch focal range) Knee phantom Set of lead numbers Lead rubber sheets RS2: Page 1 Lab 3

2 Experimental Procedure 1 Computed Radiography (CR) Procedure (Knee AP technique) Only exposures 1 thru 10 use CR image receptors. Instructions for Exposure 2 1. Place a 14 x 17 inch CR image receptor on the tabletop lengthwise. 2. Cover the image receptor with a 14 x 17 inch focused grid that has a inch focal range. 3. Align the grid s lead strips so they are parallel to the table s long axis (lengthwise). 4. Place the knee phantom on top of the grid, in the AP position. 5. Tape the appropriate lead markers onto the image receptor. (The room, side and exposure number should be labeled on all radiographs.) 6. Center the knee to the middle of the image receptor. 7. Direct the central ray perpendicular to the center of the grid at 40 SID. 8. Collimate the field size to the anatomy of interest.. 9. Expose the CR image receptor using the technique in the worksheet provided. RS2: Page 2 Lab 3

3 Instructions for Exposure 3 1. Place an unexposed 14 x 17 inch CR image receptor on the tabletop lengthwise. 2. Set up the knee and focused grid in the exact same orientation as exposure one. 3. Direct the central ray 21º cephalic to the center of the grid. 4. Use a tape measure to verify the SID needed to compensate for the central ray angle. 5. Collimate the field size to the anatomy of interest. 6. Expose the CR image receptor using the technique in the worksheet provided. (Rule of Thumb: To maintain the correct SID, the distance from the x-ray tube to the tabletop must be reduced by 1 inch for every 7 degrees of angulation. Use a tape measure to verify the change in SID. Instructions for Exposure 4 1. Place an unexposed 14 x 17 inch CR image receptor on the tabletop lengthwise. 2. Set up the knee and focused grid in the exact same orientation as exposure one. 3. Direct the central ray 30º transverse to the center of the grid. 4. Use a tape measure to verify the SID needed to compensate for the central ray angle. 5. Collimate the field size to the anatomy of interest. 6. Expose the CR image receptor using the technique in the worksheet provided. (Rule of Thumb: To maintain the correct SID, the distance from the x-ray tube to the tabletop must be reduced by 1 inch for every 7 degrees of angulation. Use a tape measure to verify the change in SID. RS2: Page 3 Lab 3

4 RS2: Page 4 Lab 3

5 Instructions for Exposure 5 7. Place an unexposed 14 x 17 inch CR image receptor on the tabletop crosswise. 8. Align the grid s lead strips so they are perpendicular to the table s long axis (crosswise). 9. Direct the central ray 30º transverse to the center of the grid. 10. Use a tape measure to verify the SID needed to compensate for the central ray angle. 11. Collimate the field size to the anatomy of interest. 12. Expose the CR image receptor using the technique in the worksheet provided. (Rule of Thumb: To maintain the correct SID, the distance from the x-ray tube to the tabletop must be reduced by 1 inch for every 7 degrees of angulation. Use a tape measure to verify the change insid RS2: Page 5 Lab 3

6 Instructions for Exposure 6 1. Place an unexposed 14 x 17 inch CR image receptor on the tabletop lengthwise. 2. Set up the pixie and focused grid in the exact same orientation as exposure one. 3. Direct the central ray perpendicular to the RIGHT EDGE (5 off the center) of the grid at 40 SID. 4. Collimate the field size to the anatomy of interest. 5. Expose the CR image receptor using the technique in the worksheet provided. RS2: Page 6 Lab 3

7 Instructions for Exposure 7 1. Place a 14 x 17 inch CR image receptor on the tabletop lengthwise. 2. Cover the image receptor with a 14 x 17 inch focused cap grid that has a inch focal range. 3. Place the grid so the side marked TUBE SIDE is not facing the tube; the grid must be upside down. 4. Align the grid s lead strips so they are parallel to the table s long axis (lengthwise). 5. Place the kneeon top of the grid, in the AP position. 6. Tape the appropriate lead markers onto the image receptor. (The room, side and exposure number should be labeled on all radiographs.) 7. Center the knee to the middle of the image receptor. 8. Direct the central ray perpendicular through the center of the grid at 40 SID. 9. Collimate the field size to the anatomy of interest. 10. Expose the CR image receptor using the technique in the worksheet provided. RS2: Page 7 Lab 3

8 Instructions for Exposure 8 1. Place a 14 x 17 inch CR image receptor on the tabletop lengthwise. 2. Cover the image receptor with a 14 x 17 inch focused cap grid that has a inch focal range. 3. Align the grid s lead strips so they are parallel to the table s long axis (lengthwise). 4. Place the pixie on top of the grid, in the AP position. 5. Tape the appropriate lead markers onto the image receptor. (The room, side and exposure number should be labeled on all radiographs.) 6. Center the knee to the middle of the image receptor. 7. Direct the central ray perpendicular to the center of the grid at 20 SID. 8. Collimate the field size to the anatomy of interest. 9. Expose the CR image receptor using the technique in the worksheet provided. RS2: Page 8 Lab 3

9 Instructions for Exposure Place a 14 x 17 inch CR image receptor on the tabletop lengthwise. 11. Cover the image receptor with a 14 x 17 inch focused cap grid that has a inch focal range. 12. Align the grid s lead strips so they are parallel to the table s long axis (lengthwise). 13. Place the pixie on top of the grid, in the AP position. 14. Tape the appropriate lead markers onto the image receptor. (The room, side and exposure number should be labeled on all radiographs.) 15. Center the knee to the middle of the image receptor. 16. Direct the central ray perpendicular to the center of the grid at 80 (or max) SID. 17. Collimate the field size to the anatomy of interest. 18. Expose the CR image receptor using the technique in the worksheet provided. RS2: Page 9 Lab 3

10 Instructions for Exposure Place an unexposed 14 x 17 inch CR image receptor on the tabletop lengthwise. 20. Set up the pixie and focused grid in the exact same orientation as exposure one. 21. Direct the central ray perpendicular to the center of the grid at 40 SID. 22. Tilt the grid and the image receptor approximately 35⁰. 23. Collimate the field size to the anatomy of interest. 24. Expose the CR image receptor using the technique in the worksheet provided. RS2: Page 10 Lab 3

11 Radiation Detectors Density Selector Radiation Detectors Density Selector Technique Worksheet A Rooms A, B, C, D Computed Radiography System Use AEC for image 1 Manual Technique Mode lgm kv ma Time Grid SID mas sec cap Use film No. 1 as the standard image for comparison. Adjust technique as necessary to obtain lgm Use manual Technique for images 2 to 10 lgm kv ma Time Grid SID mas cap cap cap cap cap cap 40 RS2: Page 11 Lab 3

12 cap cap cap 40 RS2: Page 12 Lab 3

13 Experimental Procedure 2 Direct Digital (DDR) Procedure (Knee AP technique) Exposures 11 thru 17 use wireless digital image receptors. Instructions for Exposure Place a 14 x 17 inch wireless digital image receptor in the bucky. 2. Place the knee phantom on top of the table. 3. Tape the appropriate lead markers onto the image receptor. (The room, side and exposure number should be labeled on all radiographs.) 4. Center the knee to the middle of the image receptor. 5. Direct the central ray perpendicular to the center of the grid in the bucky at 40 SID. 6. Collimate the field size to the anatomy of interest.. 7. Expose the wireless digital image receptor using the technique in the worksheet provided. Instructions for Exposure 13 RS2: Page 13 Lab 3

14 1. Place an unexposed 14 x 17 inch wireless digital image receptor in the bucky lengthwise. Place an unexposed 14 x 17 inch wireless digital image receptor in the bucky lengthwise. 2. Set up the knee and focused grid in the exact same orientation as exposure one. 3. Direct the central ray 21º cephalic to the center of the grid. 4. Use a tape measure to verify the SID needed to compensate for the central ray angle. 5. Collimate the field size to the anatomy of interest. 6. Expose the wireless digital image receptor using the technique in the worksheet provided. (Rule of Thumb: To maintain the correct SID, the distance from the x-ray tube to the tabletop must be reduced by 1 inch for every 7 degrees of angulation. Use a tape measure to verify the change in SID. Instructions for Exposure 14 RS2: Page 14 Lab 3

15 1. Place an unexposed 14 x 17 inch wireless digital image receptor in the bucky lengthwise. 2. Place an unexposed 14 x 17 inch wireless digital image receptor in the bucky lengthwise. 3. Set up the knee in the exact same orientation as exposure one. 4. Direct the central ray 30º transverse to the center of the grid. 5. Use a tape measure to verify the SID needed to compensate for the central ray angle. 6. Collimate the field size to the anatomy of interest. 7. Expose the wireless digital image receptor using the technique in the worksheet provided. (Rule of Thumb: To maintain the correct SID, the distance from the x-ray tube to the tabletop must be reduced by 1 inch for every 7 degrees of angulation. Use a tape measure to verify the change in SID. RS2: Page 15 Lab 3

16 Instructions for Exposure Place an unexposed 14 x 17 inch wireless digital image receptor in the bucky lengthwise. 2. Set up the knee the exact same orientation as exposure one. 3. Direct the central ray perpendicular to the RIGHT EDGE (5 off the center) of the grid at 40 SID. 4. Collimate the field size to the anatomy of interest. 5. Expose the wireless digital image receptor using the technique in the worksheet provided. RS2: Page 16 Lab 3

17 Instructions for Exposure Place an unexposed 14 x 17 inch wireless digital image receptor in the bucky lengthwise. 2. Set up the knee the exact same orientation as exposure one. 3. Center the knee to the middle of the image receptor. 4. Direct the central ray perpendicular to the center of the grid at 20 SID. 5. Collimate the field size to the anatomy of interest. 6. Expose the wireless digital image receptor using the technique in the worksheet provided. Instructions for Exposure 17 RS2: Page 17 Lab 3

18 1. Place an unexposed 14 x 17 inch wireless digital image receptor in the bucky lengthwise. 2. Set up the knee the exact same orientation as exposure one. 3. Direct the central ray perpendicular to the center of the grid at maximum SID (>44 ). 4. Collimate the field size to the anatomy of interest. 5. Expose the wireless digital image receptor using the technique in the worksheet provided. RS2: Page 18 Lab 3

19 Radiation Detectors Density Selector Radiation Detectors Density Selector Technique Worksheet B Rooms A, B, C, & D Use AEC for image 1 AECTechnique Mode set it to Table 40-inch AEC Technique Mode Room A set it to 200 Rooms B,C,D set it to L EI TEI DI kv ma Time Grid SID mas C 0 bucky 40 Use film No. 1 as the standard image for comparison. Adjust technique as necessary to obtain DI < 0.5 Use manual Technique for images 12 to 17 EI TEI DI kv ma Time Grid SID mas N/A N/A bucky N/A N/A bucky N/A N/A bucky N/A N/A bucky N/A N/A bucky N/A N/A bucky max RS2: Page 19 Lab 3

20 RS2: Page 20 Lab 3

21 Worksheet A Computed Radiography lgm Number Briefly describe the overall brightness, contrast and noise level of each image. Describe appearance of artifacts RS2: Page 21 Lab 3

22 RS2: Page 22 Lab 3

23 10 RS2: Page 23 Lab 3

24 Worksheet A Computed Radiography EI TEI DI Briefly describe the overall brightness, contrast and noise level of each image. Describe appearance of artifacts RS2: Page 24 Lab 3

25 RS2: Page 25 Lab 3

26 RS2: Page 26 Lab 3

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Performance testing for Precision 500D Classical R/F System

Performance testing for Precision 500D Classical R/F System Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)

More information

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality

More information

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:

More information

Geometric Optics Converging Lenses and Mirrors Physics Lab IV

Geometric Optics Converging Lenses and Mirrors Physics Lab IV Objective Geometric Optics Converging Lenses and Mirrors Physics Lab IV In this set of lab exercises, the basic properties geometric optics concerning converging lenses and mirrors will be explored. The

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

Page: 1 of 6 Page: 1 of 6

Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique

More information

The Exposure Index and Its Standardization

The Exposure Index and Its Standardization The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany ulrich.neitzel@philips.com Content The Exposure Index (EI) What is it? What is it good

More information

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION

More information

Experiment 3 Lenses and Images

Experiment 3 Lenses and Images Experiment 3 Lenses and Images Who shall teach thee, unless it be thine own eyes? Euripides (480?-406? BC) OBJECTIVES To examine the nature and location of images formed by es. THEORY Lenses are frequently

More information

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Alzen,, G., Beiderwellen,, K., Volz,, M., Berthold,, L.D.,

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Required RS Training Info

Required RS Training Info C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness

More information

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Annual MQSA Inspection Questions NMQAAC April 19, 2004 1 1 1 1 1 1 1 1 0 1 0 1 0 1 0 1 Annual MQSA Inspection Questions NMQAAC April 1, 00 1.0 Inspection Information 1.1 Name and Address 1. Equipment Registration.0 Facility Inspection Download.0 Facility Inspections.0

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes

More information

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950 HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

2012 LACC Clinical Obligations & Grading System. Attendance Requirement

2012 LACC Clinical Obligations & Grading System. Attendance Requirement 2012 LACC Clinical Obligations & Grading System Attendance Requirement The Radiology program has a Monday through Friday schedule during the fall, winter, spring and summer semesters. Students are required

More information

Bitewing Radiography B.E. DIXON. B.D.S., M.Sc., D.P.D.S.

Bitewing Radiography B.E. DIXON. B.D.S., M.Sc., D.P.D.S. Bitewing Radiography B.E. DIXON B.D.S., M.Sc., D.P.D.S. Main Indications Detection of Dental Caries Monitoring progression of caries Assessment of existing restorations Assessment of Periodontal status

More information

TS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 Instructions

TS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 Instructions TS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 ENG Instructions Thank you for purchasing a Canon product. Canon s TS-E lenses are tilt-shift lenses designed for EOS cameras. The tilt-shift mechanism enables

More information

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams EPEX Direct-to-Digital System for General Radiography Featuring... Versatility for a complete range of general radiographic exams The Hologic EPEX General Radiography System is a direct-to-digital, fully

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D. Slide 1 Digital Mammography Detector QC Slide 2 Introduction What is FFDM QC and why is it important? Eric A. Berns, Ph.D. What to know before you start eric.berns@gmail.com Overview and compare QC tests

More information

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses.

3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. 1 3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. (A) Any person may apply to the director for approval to conduct an educational program for

More information

Digital Breast Tomosynthesis QC Requirements

Digital Breast Tomosynthesis QC Requirements Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist

More information

BASIC QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY

BASIC QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY AAPM REPORT No. 4 BASIC QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY American Association of Physicists in Medicine FOREWORD This document, Basic Quality Control in Diagnostic Radiology, is the fourth in a

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

Bergen Community College Division of Health Professions/ Radiography Program. Course Syllabus and Course of Study for Lecture

Bergen Community College Division of Health Professions/ Radiography Program. Course Syllabus and Course of Study for Lecture Course Information: Course title: Advanced Imaging Equipment and Patient Care Course Code: RAD 184 Credits: 2.0 (30 lecture hours) Pre-requisites: RAD 276 - Imaging Equipment RAD 281 - Radiography II RAD

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

Ion Beam Sputtering: Practical Applications to Electron Microscopy

Ion Beam Sputtering: Practical Applications to Electron Microscopy Ion Beam Sputtering: Practical Applications to Electron Microscopy Applications Laboratory Report Introduction Electron microscope specimens, both scanning (SEM) and transmission (TEM), often require a

More information

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment

More information

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC

More information

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

A PACS-Aware DICOM Image Object

A PACS-Aware DICOM Image Object Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp

More information

COLLEGE ON MEDICAL PHYSICS

COLLEGE ON MEDICAL PHYSICS COLLEGE ON MEDICAL PHYSICS (30 August - 22 September 2004) Miramare - Trieste, Italy Directors: A. Benini, P. Sprawls and S. Tabakov Local Organizer: L. Bertocchi FINAL PROGRAMME (all lectures were held

More information

ABHES BULLETIN. Subject: Final Radiologic Technology Standards Effective January 1, 2013

ABHES BULLETIN. Subject: Final Radiologic Technology Standards Effective January 1, 2013 ACCREDITING BUREAU OF HEALTH EDUCATION SCHOOLS 7777 Leesburg Pike, Suite 314 N. Falls Church, Virginia 22043 Tel. 703/917.9503 Fax 703/917.4109 E-Mail: info@abhes.org ABHES BULLETIN To: From: ABHES-Accredited

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

Checklist for the Qualification of Digital Detector Array Systems

Checklist for the Qualification of Digital Detector Array Systems Paper No. 010-12 Checklist for the Qualification of Digital Detector Array Systems 1 February 2012 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Abdomen X-Ray (AXR) Collimation is ideally from diaphragms to lower border of the symphysis pubis and the lateral skin margins.

Abdomen X-Ray (AXR) Collimation is ideally from diaphragms to lower border of the symphysis pubis and the lateral skin margins. Abdomen X-Ray (AXR) Collimation is ideally from diaphragms to lower border of the symphysis pubis and the lateral skin margins. LMP of child-bearing age female patients should be checked. 1. Acute abdomen

More information

Cutting Edge Digital Radiography Technology in Daily NDT Use

Cutting Edge Digital Radiography Technology in Daily NDT Use Cutting Edge Digital Radiography Technology in Daily NDT Use Ron PINCU, Eli DAYAN, Ofra KLEINBERGER Vidisco Ltd. Or-Yehuda, Israel 60375 Tel: +972 (3) 533-3001 Email: ndt@vidisco.com ABSTRACT Digital Radiography

More information

Reflection and Refraction

Reflection and Refraction Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,

More information

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,

More information

RL HW / RL HW+ / RL HGW / RL HV / RL HVPW/RL HVPW-G

RL HW / RL HW+ / RL HGW / RL HV / RL HVPW/RL HVPW-G Auto-Levelling Rotary Laser Level RL HW / RL HW+ / RL HGW / RL HV / RL HVPW/RL HVPW-G 77-496 / 77-429 / 77-439 / 77-497 / 77-427/ 77-441 Please read these instructions before operating the product Auto-Levelling

More information

Integrated Management & Educational Consultancy

Integrated Management & Educational Consultancy Integrated Management & Educational Consultancy Services MANDEC Radiology Update Seminar for The Dental Team Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email:admin@imecs.freeserve.co.uk

More information

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES 1. PROTECTIVE SCREENS (a) Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) Typically, 15 kg m -2 (1.3 mm) sheet lead or its shielding

More information

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

Measurement of Charge-to-Mass (e/m) Ratio for the Electron Measurement of Charge-to-Mass (e/m) Ratio for the Electron Experiment objectives: measure the ratio of the electron charge-to-mass ratio e/m by studying the electron trajectories in a uniform magnetic

More information

DR Retrofit Kit 1417 WL Configuration Integrated PACS

DR Retrofit Kit 1417 WL Configuration Integrated PACS DR Retrofit Kit 1417 WL Configuration Integrated PACS Description: The DR Retrofit Kit 1417WL - Configuration Integrated PACS makes the use of CR Cassettes/Imaging Plates or the use of analogue x-ray films,

More information

RESOURCE MANUAL FOR COMPLIANCE TEST PARAMETERS DIAGNOSTIC X-RAY SYSTEMS. Reprinted. July 15, 1999. Diagnostic Devices Branch

RESOURCE MANUAL FOR COMPLIANCE TEST PARAMETERS DIAGNOSTIC X-RAY SYSTEMS. Reprinted. July 15, 1999. Diagnostic Devices Branch RESOURCE MANUAL FOR COMPLIANCE TEST PARAMETERS OF DIAGNOSTIC X-RAY SYSTEMS Reprinted July 15, 1999 Diagnostic Devices Branch Division of Enforcement I Office of Compliance 2098 Gaither Road Rockville,

More information

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act

More information

Best Practices in Digital Radiography

Best Practices in Digital Radiography WHITE PAPER Best Practices in Digital Radiography Tracy L. Herrmann, M.Ed., R.T.(R); Terri L. Fauber, Ed.D., R.T.(R)(M); Julie Gill, Ph.D., R.T.(R)(QM); Colleen Hoffman, R.T.(R)(M)(CT); Denise K. Orth,

More information

Thin Lenses Drawing Ray Diagrams

Thin Lenses Drawing Ray Diagrams Drawing Ray Diagrams Fig. 1a Fig. 1b In this activity we explore how light refracts as it passes through a thin lens. Eyeglasses have been in use since the 13 th century. In 1610 Galileo used two lenses

More information

33-10-06-03. General requirements. 1. Administrative Controls.

33-10-06-03. General requirements. 1. Administrative Controls. 33-10-06-03. General requirements. 1. Administrative Controls. a. Registrant. The registrant shall be responsible for directing the operation of the x-ray systems which have been registered with the department.

More information

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of

More information

Parts List. The HyperStar Lens Assembly includes three pieces: HyperStar Lens Secondary Mirror Holder Counterweight

Parts List. The HyperStar Lens Assembly includes three pieces: HyperStar Lens Secondary Mirror Holder Counterweight The HyperStar Lens allows CCD imaging at f/2.3 with compatible Celestron 9.25 Schmidt-Cassegrain telescopes and compatible CCD cameras. These instructions show the proper methods for installing the HyperStar

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

Basic Radiographic Procedures

Basic Radiographic Procedures Basic Radiographic Procedures The GPT Method G. Patrick Thomas, Jr., DC, DACBR Copyright 2002 G. Patrick Thomas, Jr., DC, DACBR PO Box 1000 Blue Springs, MO 64013-1000 gptdacbr@juno.com Table of Contents

More information

The Radiologic Technology Program at Trenholm University

The Radiologic Technology Program at Trenholm University H. Councill Trenholm State Technical College 243 Location: Trenholm Campus Program Information The Radiology Technology program at TrenholmState will provide students with the necessary training to gain

More information

Video surveillance camera Installation Guide

Video surveillance camera Installation Guide Video surveillance camera Installation Guide TV7085 TV7086 TV7087 TV7088 14 1. Preface Dear Customer, Thank you for purchasing this Eyseo digital surveillance camera. You made the right decision in choosing

More information

American College of Radiology Mammography Accreditation Program

American College of Radiology Mammography Accreditation Program American College of Radiology Mammography Accreditation Program Objectives Overview of the ACR Mammography Accreditation Program Interrelationship between mammography accreditation, FDA certification and

More information

Radiographic Procedure Manual for Examinations of the Knee, Hand, Pelvis and Lower Limbs. Osteoarthritis Initiative: A Knee Health Study

Radiographic Procedure Manual for Examinations of the Knee, Hand, Pelvis and Lower Limbs. Osteoarthritis Initiative: A Knee Health Study Radiographic Procedure Manual for Radiographic Procedure Manual for Examinations of the Knee, Hand, Pelvis and Lower Limbs Osteoarthritis Initiative: A Knee Health Study SYN RC W H E R E S C I E N C E

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

RADIATION SAFETY STANDARD

RADIATION SAFETY STANDARD Radiation Safety Act 1999 RADIATION SAFETY STANDARD NM004:2010 Standard for radiation apparatus used to carry out industrial radiography Preface Under section 17 of the Radiation Safety Act 1999, a possession

More information

WARNING: We recommend that you consult the Safety, Regulatory and Technical Specification User Guide before using the CS 2200.

WARNING: We recommend that you consult the Safety, Regulatory and Technical Specification User Guide before using the CS 2200. CS 2200 User Guide Notice Congratulations on your purchase of the CS 2200. Thank you for your confidence in our products and we will do all in our power to ensure your complete satisfaction. The User Guide

More information

7.0, 4 Wire Touch Panel

7.0, 4 Wire Touch Panel 7.0, 4 Wire Touch Panel Product Specification PRODUCT NUMBER DTS408-0700-00 Product Mgr Quality Mgr Engineering Document Control Date: Date: Date: Date: Product No. DTS408-0700-00 REV. B Page 1 / 10 TABLE

More information

ViZion Installation Guide

ViZion Installation Guide ViZion Installation Guide v2.0 1 ViZion Installation Guide Table of Contents Inventory Hardware Setup Understanding Synchronization Cable Begin by taking an inventory of the required equipment DR Unit

More information

Unarmored Variable Frequency Drive (VFD) Cable Termination Guide

Unarmored Variable Frequency Drive (VFD) Cable Termination Guide Authorized Distributor Unarmored Variable Frequency Drive (VFD) Cable Termination Guide Be Certain with Belden A Step-by-Step Look at the Connection/Termination of Unarmored VFD Cables Terminating Unarmored

More information

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film. INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

More information

The DC Motor. Physics 1051 Laboratory #5 The DC Motor

The DC Motor. Physics 1051 Laboratory #5 The DC Motor The DC Motor Physics 1051 Laboratory #5 The DC Motor Contents Part I: Objective Part II: Introduction Magnetic Force Right Hand Rule Force on a Loop Magnetic Dipole Moment Torque Part II: Predictions Force

More information

QUALITY CONTROL RECOMMENDATIONS FOR DIAGNOSTIC RADIOGRAPHY VOLUME 3 RADIOGRAPHIC OR FLUOROSCOPIC MACHINES

QUALITY CONTROL RECOMMENDATIONS FOR DIAGNOSTIC RADIOGRAPHY VOLUME 3 RADIOGRAPHIC OR FLUOROSCOPIC MACHINES CRCPD Publication 01-6 $20.00 QUALITY CONTROL RECOMMENDATIONS FOR DIAGNOSTIC RADIOGRAPHY VOLUME 3 RADIOGRAPHIC OR FLUOROSCOPIC MACHINES July 2001 Published by Conference of Radiation Control Program Directors,

More information

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Standard Test Method for Classification of Film Systems for Industrial Radiography 1 Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately

More information

MAGNETIC CARD READER DESIGN KIT TECHNICAL SPECIFICATION

MAGNETIC CARD READER DESIGN KIT TECHNICAL SPECIFICATION MAGNETIC CARD READER DESIGN KIT TECHNICAL SPECIFICATION Part Number: 99821002 Rev 21 FEBRUARY 2011 REGISTERED TO ISO 9001:2008 1710 Apollo Court Seal Beach, CA 90740 Phone: (562) 546-6400 FAX: (562) 546-6301

More information

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015) The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015) Table of Contents Stereotactic Breast Biopsy Accreditation Program

More information

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Digital Medical Radiography Diagnostic (DR) system X-ray Equipment Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Lin muyan 1. Goals 2. Introduction 3. X-ray machine

More information

AN INNOVATIVE & CUTTING-EDGE DIGITAL RADIOGRAPHY SYSTEM

AN INNOVATIVE & CUTTING-EDGE DIGITAL RADIOGRAPHY SYSTEM SAMSUNG SMART DIGITAL RADIOGRAPHY 2012 Samsung Electronics All rights reserved. Samsung Electronics reserves the right to modify the design, packaging, specifications and features shown herein, without

More information

SECURITY CAMERA INSTRUCTION MANUAL ENGLISH VERSION 1.0 LBC5451. www.lorextechnology.com

SECURITY CAMERA INSTRUCTION MANUAL ENGLISH VERSION 1.0 LBC5451. www.lorextechnology.com SECURITY CAMERA INSTRUCTION MANUAL ENGLISH VERSION 1.0 LBC5451 www.lorextechnology.com Contents 1x Camera and mounting stand 1x Power adapter 1x Mounting kit 1x Allen key 1x Screw cap 1x 60 ft. (18m)

More information

TECHNICAL BULLETIN. Meritor WABCO Cab Leveling Valves and Chassis Leveling Valves. How the Cab Leveling and Chassis Leveling Valves Work

TECHNICAL BULLETIN. Meritor WABCO Cab Leveling Valves and Chassis Leveling Valves. How the Cab Leveling and Chassis Leveling Valves Work Revised 02-00 TECHNICAL BULLETIN Meritor WABCO Cab Leveling Valves and Chassis Leveling Valves This technical bulletin covers both cab and chassis leveling valves manufactured by Meritor WABCO. While the

More information

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational

More information

CLINICAL PERFORMANCE OBJECTIVES

CLINICAL PERFORMANCE OBJECTIVES CLINICAL PERFORMANCE OBJECTIVES CLINICAL COMPETENCY CLINICAL PERFORMANCE CRITERIA 1. Evaluation of Requisition: Student was able to: a. Identify procedures to be performed b. Recall the patient s age,

More information

Manual for simulation of EB processing. Software ModeRTL

Manual for simulation of EB processing. Software ModeRTL 1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended

More information

SAFELIGHT FILTERS AND DARKROOM LAMPS

SAFELIGHT FILTERS AND DARKROOM LAMPS TECHINCAL INFORMATION SAFELIGHT FILTERS AND DARKROOM LAMPS A RANGE OF HIGH QUALITY FILTERS AND LAMPS FOR DARKROOM ILLUMINATION DESCRIPTION ILFORD manufacture a range of high quality Safelight Colour Use

More information

Polarization of Light

Polarization of Light Polarization of Light References Halliday/Resnick/Walker Fundamentals of Physics, Chapter 33, 7 th ed. Wiley 005 PASCO EX997A and EX999 guide sheets (written by Ann Hanks) weight Exercises and weights

More information

Magnetic Fields and Their Effects

Magnetic Fields and Their Effects Name Date Time to Complete h m Partner Course/ Section / Grade Magnetic Fields and Their Effects This experiment is intended to give you some hands-on experience with the effects of, and in some cases

More information

Radiologic Technology (RAD, CLE) courses are open only to Radiologic Technology majors.

Radiologic Technology (RAD, CLE) courses are open only to Radiologic Technology majors. RADIOLOGIC TECHNOLOGY (A.A.S Degree) Director: Prof. Virginia Mishkin, M.S., R.T. (R) (M) (QM) A radiologic technologist is a skilled professional who provides a specialized health care service. This rewarding

More information