COMMISSIONING AND USE OF CONVENTIONAL CT

Similar documents
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Purchasing a cardiac CT scanner: What the radiologist needs to know

Quality Assurance of Radiotherapy Equipment

How To Improve Your Ct Image Quality

INTRODUCTION. A. Purpose

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

Data. microcat +SPECT

IAEA HUMAN HEALTH SERIES

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Brilliance CT 64-channel confi guration

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Fundamentals of Cone-Beam CT Imaging

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Key words: treatment planning, quality assurance, 3D treatment planning

Clinical Rotation 3: PHYS 705 Fall 2015 (Aug. 25, 2015 to Feb. 25, 2016) COURSE INFORMATION

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver /9/07) by M. Jensen and J. E.

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

HDR Brachytherapy 1: Overview of QA. Disclosures: Learning Objectives 7/23/2014

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

SITE IMAGING MANUAL ACRIN 6698

American College of Radiology CT Accreditation Program. Testing Instructions

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

1. Provide clinical training in radiation oncology physics within a structured clinical environment.

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

Mx8000 IDT 16 provides dose control methodologies like DoseRight ACS (Automatic Current Selection) and DoseRight DOM (dynamic dose modulation).

State-of-the-Art Technology in Cardiac CT

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy

AAPM REPORT NO. 39 SPECIFICATION AND ACCEPTANCE TESTING OF COMPUTED TOMOGRAPHY SCANNERS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

A d v a n c e d 1 2 8

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

DICOM Correction Item

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad.

Our Department: structure and organization

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

THE DOSIMETRIC EFFECTS OF

First Three Years After Project Proton Therapy Facility:

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Quality Assurance in Stereotactic. Radiotherapy

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline

Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients

Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

DENTAL Cone beam 3D X-RAY SYSTEM with

IAEA-TECDOC Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems

Accreditation Is Coming

Quality Control of Full Field Digital Mammography Units

Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA USA phone NOMOS

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Page: 1 of 6 Page: 1 of 6

Quality Control and Maintenance Programs

The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

Physics testing of image detectors

Prepublication Requirements

WHERE IN THE WORLD JILL LIPOTI?

QA of intensity-modulated beams using dynamic MLC log files

Low-dose CT for Pulmonary Embolism

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

CTVision System. Accurate Imaging for Precise Treatment.

Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

The power of Multislice CT Philips MX 6000 Dual

MEDICAL IMAGING 2nd Part Computed Tomography

Scripps Proton Therapy Center: Configuration and Implementation

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Automotive Applications of 3D Laser Scanning Introduction

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Concepts for High-Resolution Low-Dose CT of the Breast

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

VIRTUAL SIMULATION FOR RADIATIOTHERAPY TREATMENT USING CT MEDICAL DATA

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

HITACHI Proton Beam Therapy System

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong

Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems

Manual for simulation of EB processing. Software ModeRTL

GE Healthcare. Great treasures in small places. BrightSpeed Elite

Deformable Registration for Image-Guided Radiation Therapy

Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD

Acknowledgements. PMH, Toronto David Jaffray Doug Moseley Jeffrey Siewerdsen. Beaumont Hospital Di Yan Alvaro Martinez. Elekta Synergy Research Group

Transcription:

COMMISSIONING AND USE OF CONVENTIONAL CT Sasa Mutic Department of Radiation Oncology Siteman Cancer Center Mallinckrodt Institute of Radiology Washington University School of Medicine St. Louis, Missouri 63110

Outline Introduction and historical review CT-Technology Technology Multislice Benefits Commissioning Conclusions Disclaimer: Our university uses Philips scanners in radiation therapy and I have easy access to their images which are used in this presentation. This should in no way be interpreted as our endorsement of these products.

CT in Radiation Therapy

CT simulator history 1983 - Goitein and Abrams described multidimensional treatment planning based on CT images Beams eye view (BEV) film created from a divergent projection through the CT study data 1987-1990 1990 - Sherouse et al software analog to conventional simulation Volumetric CT scan represents virtual patient Software functions create a virtual simulator 1990 Nishidai et al and Nagata et al and 1994 Perez et al CT simulator systems with laser beam projecting device 1993-1994 1994 Commercial systems introduced

CT simulator CT scanner with external lasers Flat tabletop Virtual simulation software

CT simulation process 3 4 5 1 2 CT Simulator CT Scanner Virtual Simulation Dose Calculation Treatment Planning System

Technology

CT Simulator Evaluation Task Radiation and patient safety CT dosimetry Evaluation of electromechanical components Evaluation of image quality Solution? AAPM report number 39, AAPM TG53 report AAPM TG66 report www.impactscan.org

Generations 3 rd Generation CT Scanner Fan Beam Rotate/Rotate

Generations 4 th Generation CT Scanner Fan Beam Rotate/Stationary

X-ray Tube Historical Limitation Heat generation - inefficiency Large number of images per study DRR quality Target delineation Rapid study acquisition time Spatial and temporal integrity Motion artifacts Large heat anode storage ability (MHU): 5+ MHU tubes Fast anode cooling rate (MHU/min)

Image Generation Single Slice One Rotation One Image

CT T scanner Single and multi-slice scanning Wider collimator widths Post patient collimation Multiple area detectors 1992 - Elscint CT Twin - first CT scanner capable of simultaneously acquiring more than one transaxial slice 1998 Four major manufacturers introduce scanners capable of scanning 4 slices simultaneously Today - 64+ slice scanners commercially available

Multislice CT 1 2 3 4 One Rotation Multiple Images

Detector Configuration 16-Slice Scanner 4 x 1.5 mm 8x0.75 mm 8x0.75 mm 4 x 1.5 mm 24 detector elements Courtesy Philips Medical Systems, Inc.

Detector Configuration 16-Slice Scanner TUBE 24mm collimation 8 x 3mm collimation 3mm 3mm 3mm 3mm 3mm 3mm 3mm 3mm 3mm slice thickness

CT scanner Multi-slice scanning Faster scan times 4 slice scanner example (8 times faster):» multi: : 0.5 sec/rotation and 4 slices/rotation» single: : 1 sec/rotation and 1 slice/rotation Lower tube heat loading Longer volume covered per rotation Improved temporal resolution - faster scan times Improved spatial resolution thinner slices Decreased image noise more ma available

Resolution The lower limit on slice thickness for most single slice scanners in radiotherapy is 3 mm Often 5 and 8 mm slices were used Multislice scanners can acquire 0.75 mm thick slices with equivalent scan parameters Everything else being equal thinner slice thickness produces better DRRs

DRR Image Quality Everything else being equal, thinner slices produce better images Balance between large amounts of data and image quality 5mm Slices 3mm 0.8mm

DRR Image Quality 5mm Slices 3mm 0.8mm

DRR Image Quality 800 Images 0.8 mm slice thickness

Image Reconstruction Voxel: Volume element representing the slice thickness or depth of the image. 3 Dimensional X Y Z X Y Z

Isotropic Imaging Square isotropic voxels X = Y = Z Sub-millimeter slice thicknes Multi Planar Reconstruction Sagittal Coronal Multi Planar Contouring X Y Z

Multi-planar contouring Axial Sagittal Coronal Resolution the same in all three planes It does not matter which plane is used for contouring Isotropic Resolution

Isotropic Resolution

Data Issues Image viewing speed Network Traffic Archiving Dose calculation Need tools to allow use of data without compromise in treatment planning speed

Multi-slice CT - Speed Single slice scanner 30 seconds Multi slice scanner 2 to 4 seconds Dynamic CT 4D CT 5D CT Gating Tumor motion

Multi-slice CT - Dynamic CT Approximately 7000 images

Multi-slice CT - Dynamic CT

Multi-slice CT- mas mas proportional to number of photons Number of photons affects noise (image quality) Single slice scanner 150 to 300 mas Multi slice scanner up to 2000 mas Increasing mas increases patient dose from a CT scan Increase in dose is a significant concern in diagnostic scanning Dose from a CT scan is insignificant compared to therapy dose Can take advantage of available mas in radiotherapy scanning

Contrast Resolution A measure of a scanners sensitivity, or the ability to discriminate small changes in density Soft tissue contrast Affects ability to contour structures 200 mas 1000 mas

200 mas 300 mas 400 mas 600 mas 700 mas 1000 mas

300 mas 1000 mas

Rows and columns form a matrix. 512x512 768x768 1024x1024 Spatial Resolution

Image Reconstruction Pixel: Picture element representing both the density and area of a small portion of the image. Size is very important! Pixel size = Field of View (FOC) divided by the display matrix. 480/512 = 0.94 mm 300/512 = 0.59 mm Y 480/1024 = 0.47 mm X 300/1024 = 0.29 mm

Resolution Low resolution High resolution

CT scanner Bore size 85 cm Bore Opening 70 cm Bore Opening

CT scanner Bore size Patient Size

CT scanner Bore size Mantle Field

CT Time Line Helical Scanning Commercial CT-simulator Large Bore SS 16 slice sub 0.5 64 slice 85 89 92 94 98 00 01 02 04 Slip Ring Dual slice Scanning 4-slice 0.5 sec 8-slice 0.5 sec Large Bore MS

Paradigm Shift CT scanners used to be made for diagnostic radiology and then modified for radiotherapy There are new CT scanners that were specifically designed for radiotherapy Or they have special features that are designed for radiotherapy

CT Scanner Selection Small Bore vs. Large Bore Patient population Conventional simulator availability The question will (has) become:» How large? (80, 82, or 85 cm) Single Slice vs. Multi Slice DRR quality Image quality 4D CT The question will (has) become:» How many? (2, 4, 6, 10, 16, 40, cone beam)

Commissioning and QA

Geometric Accuracy

Tasks Radiation and patient safety CT dosimetry Evaluation of electromechanical components Evaluation of image quality

Radiation and Patient safety Patient Safety Interlocks Electromechanical Door Interlock CTDI» Definition» Multislice CT Radiation Safety Workload potential pitfall» Significant increase» Shielding design» NCRP 147» Radiation survey

Electromechanical Components X-ray Generator Gantry Alignment Table Alignment/Accuracy Laser Alignment/Accuracy

Electromechanical Components x-ray Generator Need a non-invasive meter kv accuracy Timer accuracy ma linearity HVL measurements

CT Simulator Mechanical Alignemnt

Electromechanical Components Gantry Alignment/Accuracy Gantry tilt accuracy Gantry vertical Imaging plane orthogonal to the couch top Gantry vertical placement reproducibility Especially important for dual purpose scanners

Electromechanical Components Table Alignment/Accuracy Tested with weight Settle Sag Tabletop motion orthogonal/parallel with the imaging plane Table positional accuracy/reproducibility Vertical Longitudinal

Electromechanical Components Laser Alignment/Accuracy Lasers orthogonal/parallel with the imaging plane Lasers spacing Laser positional accuracy Absolute Linearity Reproducibility Coordinate system orientation

Image Quality Indicators Quantitative Phantom Measurements» High Contrast» Low Contrast» Uniformity» Spatial Integrity» Artifacts» Slice thickness» CT # accuracy Qualitative Physician Preferences» Tumor» Normal Structures» DRR/DCR Objects» Workflow» Customized protocols

Image Performance Catphan 500 CTP401 CTP515 CTP445 CTP528 CTP486

Ability of the system to record separate images of small objects that are placed very close together Resolution (High Contrast)

Subject Contrast (Low Contrast) Ability of a system to resolve adjacent objects with small density differences Noise limited

Field of View (FOV) 60 cm

Uniformity

True vs. Extrapolated FOV From impactscan.org report 05071

Evaluation of Extrapolated FOV From impactscan.org report 05071

Image Performance Evaluation

Evaluation of CT Simulation Software What is CT-simulation Software Image input verification (data transfer) Orientation Spatial accuracy Structure delineation Isocenter calculation and movement Output parameters and connectivity

Dynamic DCRs

Dynamic DRR

Conclusions CT will remain the primary imaging modality in radiotherapy CT simulator only departments Large bore multislice CT will become standard scanners for radiotherapy Existing documents (Report #39, tg66, TG53) are a good foundation New CT scanning capabilities/parameters can be evaluated with the existing recommendations