High Resolution Stellar Spectroscopy with VBT Echelle Spectrometer

Similar documents
WAVELENGTH OF LIGHT - DIFFRACTION GRATING

MY FIRST STEPS IN SLIT SPECTROSCOPY

ING LA PALMA TECHNICAL NOTE No Investigation of Low Fringing Detectors on the ISIS Spectrograph.

Introduction to reflective aberration corrected holographic diffraction gratings

Measuring the Doppler Shift of a Kepler Star with a Planet

Spectrophotometry of Ap Stars

GRID AND PRISM SPECTROMETERS

The Wide Field Cassegrain: Exploring Solution Space

O6: The Diffraction Grating Spectrometer

Alpy guiding User Guide. Olivier Thizy François Cochard

COOKBOOK. for. Aristarchos Transient Spectrometer (ATS)

A NEW LOOK AT RISLEY PRISMS. By Craig Schwarze Senior Systems Engineer OPTRA Inc.

Introduction to CCDs and CCD Data Calibration

Fig.1. The DAWN spacecraft

Efficiency, Dispersion and Straylight Performance Tests of Immersed Gratings for High Resolution Spectroscopy in the Near Infra-red

Development of the Extreme Ultraviolet Spectrometer: EXCEED

Alpy 600 User Guide. Olivier Thizy François Cochard

ATOMIC SPECTRA. Apparatus: Optical spectrometer, spectral tubes, power supply, incandescent lamp, bottles of dyed water, elevating jack or block.

arxiv: v1 [astro-ph.im] 24 Mar 2009

- the. or may. scales on. Butterfly wing. magnified about 75 times.

FTIR Instrumentation

Meteor Spectroscopy First Steps

THE BOHR QUANTUM MODEL

Chapter 6 Telescopes: Portals of Discovery. How does your eye form an image? Refraction. Example: Refraction at Sunset.

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Theremino System Theremino Spectrometer Technology

Extraction Heights for STIS Echelle Spectra

1 Laboratory #5: Grating Spectrometer

Introduction to Fourier Transform Infrared Spectrometry

Non-radial pulsations (NRP) or co-rotating photospheric Helium clouds in ζ Tau?

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

MCAL Spectrophotometry. Spectrophotometry

Alpy calibration User Guide. Olivier Thizy François Cochard

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

Module 13 : Measurements on Fiber Optic Systems

Modern Classical Optics

S2000 Spectrometer Data Sheet

Telescope Types by Keith Beadman

SPECTROSCOPIC OBSERVATIONS OF VENUS FOR ROTATION MADE AT MOUNT WILSON IN 1956

Application Report: Running µshape TM on a VF-20 Interferometer

Understanding Line Scan Camera Applications

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Experiment 5. Lasers and laser mode structure

P R E A M B L E. Facilitated workshop problems for class discussion (1.5 hours)

Physics 441/2: Transmission Electron Microscope

IMPROVED CCD DETECTORS FOR HIGH SPEED, CHARGE EXCHANGE SPECTROSCOPY STUDIES ON THE DIII D TOKAMAK

SSO Transmission Grating Spectrograph (TGS) User s Guide

Understanding astigmatism Spring 2003

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

GLOBAL COLLEGE OF ENGINEERING &TECHNOLOGY: YSR DIST. Unit VII Fiber Optics Engineering Physics

IRON ABUNDANCE DETERMINATION FOR THE SOLAR-LIKE STARS HR4345 AND HR6573

Using Photometric Data to Derive an HR Diagram for a Star Cluster

MEASUREMENT OF END FACE GEOMETRY ON FIBER OPTIC TERMINI...2

Alignement of a ring cavity laser

Interference. Physics 102 Workshop #3. General Instructions

5-Axis Test-Piece Influence of Machining Position

3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

Experiment #5: Qualitative Absorption Spectroscopy

Agilent Cary 60 UV-Vis

X-ray thin-film measurement techniques

Endoscope Optics. Chapter Introduction

6) How wide must a narrow slit be if the first diffraction minimum occurs at ±12 with laser light of 633 nm?

High Resolution Imaging in the Visible from the Ground without Adaptive Optics: New Techniques and Results

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Problem Set 6 UV-Vis Absorption Spectroscopy Express the following absorbances in terms of percent transmittance:

Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR Spectrometer and URA Accessory

Anamorphic imaging with three mirrors: a survey

Light Telescopes. Grade Level: class periods (more if in-depth research occurs)

Light and its effects

Spectroscopy Using the Tracker Video Analysis Program

Plastic Film Texture Measurement With 3D Profilometry

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

First let us consider microscopes. Human eyes are sensitive to radiation having wavelengths between

Classroom Exercise ASTR 390 Selected Topics in Astronomy: Astrobiology A Hertzsprung-Russell Potpourri

SOLSPEC MEASUREMENT OF THE SOLAR ABSOLUTE SPECTRAL IRRADIANCE FROM 165 to 2900 nm ON BOARD THE INTERNATIONAL SPACE STATION

How To Understand Light And Color

Calibration of a High Dynamic Range, Low Light Level Visible Source

ID Objective Requirements Description of Test Date & Examiner 15 Verify that the optics?? OMC

Grazing incidence wavefront sensing and verification of X-ray optics performance

THE MMT MEGACAM Focal Plane Design and Performance

RESULTS FROM A SIMPLE INFRARED CLOUD DETECTOR

WOOD WEAR TESTING USING TRIBOMETER

Rest optical spectroscopy of FeLoBALs and LoBALs at intermediate redshift

Physics 10. Lecture 29A. "There are two ways of spreading light: to be the candle or the mirror that reflects it." --Edith Wharton

Application Note #503 Comparing 3D Optical Microscopy Techniques for Metrology Applications

TECNOTTICA CONSONNI SRL CERTIFIED QUALITY MANAGEMENT SYSTEM COMPANY BY DNV UNI EN ISO 9001:2008

5. The Nature of Light. Does Light Travel Infinitely Fast? EMR Travels At Finite Speed. EMR: Electric & Magnetic Waves

What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher)

An Airborne A-Band Spectrometer for Remote Sensing Of Aerosol and Cloud Optical Properties

Acousto-optic modulator

Copyright by Mark Brandt, Ph.D. 12

Optical Communications

Simplicity. Reliability. Performance

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Spherical Beam Volume Holograms Recorded in Reflection Geometry for Diffuse Source Spectroscopy

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Detecting and measuring faint point sources with a CCD

Physics 41 Chapter 38 HW Key

Transcription:

J. Astrophys. Astr. (2005) 26, 331 338 High Resolution Stellar Spectroscopy with VBT Echelle Spectrometer N. Kameswara Rao, S. Sriram, K. Jayakumar & F. Gabriel Indian Institute of Astrophysics, Bangalore 560 034, India. Abstract. The optical design and performance of the recently commissioned fiber fed echelle spectrometer of 2.34 meter Vainu Bappu Telescope are described. The use of it for stellar spectroscopic studies is discussed. Key words. High resolution spectroscopy. 1. Introduction An efficient high resolution spectrometer has been a long standing need at Vainu Bappu Observatory (that could cover a large region of the spectrum extending from 4000 Å to 1 µm in one exposure) for stellar spectroscopic programmes. The obvious choice that emerged is an echelle grating based spectrometer. The environmental conditions at VBO dictated the efficiency optimization to wavelengths longward of 4000 Å. The final choice was a spectrometer that has transmitting optics configured to operate in Littrow mode that provides a high resolution and a large wavelength coverage simultaneously by using an echelle grating at a large blaze angle (θ of 70 ). The order overlap at the focus is separated by dispersing them in the perpendicular direction by the use of a cross dispersing prism. Prisms are preferred owing to their higher throughput efficiency and the fact that they give rise to a more uniform order spacing. The prism dispersion is greater in the blue where the free spectral range of the echelle grating (i.e., the difference in wavelength between the centers of the successive orders) is less. The major advantage of this type of format is its suitability to image (or record) the spectrum on two-dimensional high quantum efficiency detectors, like CCD. The conventional coude scheme of bringing the star light to a laboratory based spectrometer for VBT requires seven reflections which may cause loss of about 10 per cent light at each reflection. It was realized that there would be much less light loss if transmitted by an optical fiber from the prime focus of the telescope directly to the spectrometer slit. The diameter of the fiber that satisfies the requirement of minimizing the sky light entering the slit while allowing most of the star light, is about 100 µm which corresponds to 2.7 arcsec at the prime focus. This size is slightly larger than the average seeing disc at VBT. There are other advantages as well for using a fiber to link the spectrometer to the telescope like minimizing the image guiding errors, image scrambling thus uniformly illuminating the slit (optics), etc. Thus the final choice was the use of a 45 metre long optical fiber to link the prime focus to the spectrograph slit. In the following account we describe the configuration of the echelle spectrometer. A more detailed account can be found in Rao et al. (2005). 331

332 N. Kameswara Rao et al. Figure 1. Opto-mechanical layout of the spectrometer. F/5 beam emerging from the slit is fed into the collimator by a folding mirror. The collimated beam then passes through the cross disperser (prism) and gets predispersed and illuminates the echelle grating. The dispersed beam from the grating again passes through the prism a second time and finally comes to focus on the CCD plane where the spectrum is recorded. 2. Design and optical layout An echelle spectrometer like the conventional grating spectrometer essentially consists of a slit, collimator, grating, camera and detector. The additional element is a cross disperser prism or a grating. The fundamental consideration is the maximum spectral resolving power based on the astronomical motivation. In our case we chose it to be about 70,000 (4 km s 1 ). Once this is fixed other considerations about the echelle grating, the beam size, the camera, collimator focal length, the detector size, etc., follow. The questions that are considered here are: How to fit most of the spectrum in a given detector real estate (i.e., the CCD size). The extent of the wavelength range (i.e., number of orders). Overlaps (free spectral range, number of lines/mm) and order spacing such that the inter-order background is minimal (i.e., cross-disperser). The intensity variations along the order as well as across the orders should be smooth and less steep (i.e., broad blaze function) so that nearly the same signal to noise could be obtained across an order without saturating in one place and unable to register in other place. Efficiency should be optimal (i.e., throughput should be high) and scattered light minimal. Good sharp lines without tilts and wings (in plane design). Stability and repeatability (environmental, mechanical stability, isolation from the telescope, temperature control and dome vibrations, etc.). Spectral format should be amendable for easy reductions and calibrations. Easy operations. We finally preferred the all transmitting system for the spectrometer with the optical surfaces anti-reflection coated to optimize the throughput rather than reflecting cameras with central obscurations. Following other optimized throughput spectrometers like Sandiford at McDonald Observatory (McCarthy et al. 1993), we opted for Littrow configuration (i.e., the collimator also acts as a camera; the input and output beams trace almost the same path). Again to have less steep light distribution along the orders we preferred to broaden the blaze function by using an echelle grating with more number of grooves/mm. The echelle grating selected has 52.67 gr/mm and blazed at 70 (R2.6).

VBT Echelle Spectrometer 333 1.2 1 0.8 0.6 0.4 0.2 ------: Atlas 0 6550 6555 6560 6565 6570 Figure 2. Spectrum of Arcturus obtained with fiber fed echelle spectrometer at VBT superposed on Kit Peak Atlas spectrum. The expected resolution of the spectrometer with a slit width of 57 µm and a collimated beam diameter of 151 mm is 72,000. The star light after hitting the primary mirror which comes to a focus as an f/3.25 beam, is fed to a 100 µm core fiber located on the optical axis of the telescope. An image acquisition unit containing an intensified camera displays the stellar image and the fiber position on a monitor in the control room from where guiding is done. The light from calibration lamps Thorium Argon for wavelength calibration and Xenon and Tungsten lamps for flat field, is also fed to the fiber through the same unit. The fiber output is collimated and converted to an f/5 converging beam that focuses on to the slit. The image size at the slit is 154 µm. The fiber output can also be used without the slit with a lower resolution for maximum throughput. The light after the slit is reflected into the collimator by a 10 mm folding mirror that sends the beam at an angle of 0.05 to the optical axis of the spectrometer. The collimator is a six element f/5 system designed for a beam size of 151 mm and has a focal length of 755 mm. The first element (also the last element in the double pass mode) corrects for off-axis feeding of the beam. The system further contains a doublet and a triplet, all anti-reflection coated and chromatically corrected to a wavelength range of 4000 Å to 1 µm. All the elements in the collimator is rigidly mounted in a single cylindrical tube with a provision to flush them with dry nitrogen. The collimated beam then passes through an LF5 prism of about 165 mm height and an apex angle of 40 which pre-disperses (in the cross dispersion direction) the light and sends it to the echelle grating. The echelle grating is illuminated by the incident beam fully. The grating is the single

334 N. Kameswara Rao et al. 200 150 100 50 0 0 50 100 150 200 Figure 3. Comparison of equivalent widths in the spectra taken from VBT and Kitt Peak. largest grating available and has a size of 408 208 mm. The dispersed beam retraces the same path and the prism cross-disperses for a second time (thus ensuring enough order separation) and enters the collimator for a second time. Now the collimator acts as a camera and focuses the dispersed spectrum on to a CCD chip. The emergent beam is again slightly off-axis and avoids the reflecting mirror that puts the beam in. The camera has a corrected field of about 60 mm diameter in the focal plane. The spectrum is recorded on a 2048 4096 pixels CCD camera exclusively built for the spectrometer. The CCD is a back illuminated, thinned Marconi chip with square pixels of 15 µm size and is placed in an LN2 cooled dewar. The controller and the associated electronics, and the dewar have all been built in the Indian Institute of Astrophysics laboratories. The slit size of 60 µm covers about four pixels on the CCD. The present CCD chip almost covers half of the area of the two dimensional spectrum format in the focal plane of the camera. The desired wavelength region is recorded either by moving the grating or CCD dewar. All the optical components are mounted on a Milles Griot vibration free table of size 8 4 feet. The opto-mechanical layout of the spectrometer is shown in Fig. 1. The CCD dewar is mounted on a stage with a provision to move in X Y and Z directions, the direction of the optical axis of the collimator. The whole mechanical assembly is placed in an environmentally (temperature, humidity) controlled dark room Coude laboratory. It was realized early that a separate foundation might be required which is isolated from the rest of the telescope building and dome, for the coude laboratory. Thus four pillars have been incorporated in the building in the coude area exclusively for placing the spectrometer. The vibrational free table is placed on these independent foundations to isolate it from the building vibrations.

VBT Echelle Spectrometer 335 Figure 4. Comparison of the spectra of hydrogen deficient R CrB star, V854 Cen obtained with McDonald Observatory 107-inch coude echelle spectrometer with resolution 60,000 in February 1999 (full line) with the spectrum obtained by VBT fiber-fed echelle spectrometer on 13 Jan 2004 in Na ID line region. All the operations of star acquisition, control of various lamp movements, etc., are done from the telescope control room remotely through a PC. The grating and prism movements could also be accomplished remotely through PC commands (encoder displays). The CCD stage movements also can be controlled remotely. Presently the spectrometer operates in two modes with a 60 micron slit ( resolution of 72,000) and without the slit (full fiber at 27,000 resolution) although there is a provision to change the slit width. By narrowing the slit we even obtained more than 100,000 resolution for HR 3117. However, there will be loss of light on the slit jaws whenever the slit is used till an image slicer is built for the system. Presently two other CCD systems are also available to be used with the spectrometer for programmes that do not require large wavelength coverage. A 1024 1024 pixels of 24 µm size CCD system with a read out of 30s and a Pixcellent CCD system with 2048 2048 of 13.5 µm square pixels with the readout time of about 60s. Both the systems have higher quantum efficiency in the red and cover about the same wavelength range. For recording fast time varying spectroscopic phenomena these systems would be more suitable. 3. Some observational studies In the last few months the spectrometer has been extensively used in the studies of stellar elemental abundances (Rao & Reddy 2005). The quality of the spectra obtained

336 N. Kameswara Rao et al. 160 120 120 100 200 180 160 120 100 200 180 160 160 120 100 80 0 0.5 1 phase 0 0.5 1 phase Figure 5. Variations of equivalent widths with period in γ eql. Variations are strong in rare earth lines where as Ca I lines do not show such variations. can be illustrated from a comparison of Arcturus spectrum obtained with VBT and Kitt Peak Atlas (Hinkle et al. 2000). The Fig. 2 shows the spectrum of the star α-boo (Arcturus) in the Hα region obtained with VBT at 72,000 resolution (full line) compared with the spectrum displayed Visible and near-infrared Atlas of the Arcturus spectrum obtained at Kitt Peak at 120,000 resolution. Note the line to line matching and also the depth of the strong Hα line, centered at 6563 Å, in both spectra match very well illustrating the high quality of the spectrum without any scattered light in the VBT spectrometer. Figure 3 shows the comparison of equivalent widths in the spectra to be an agreement within a standard deviation of less than 2 må. The spectrum illustrated in Fig. 4 is of the hydrogen deficient star V854 Cen that has been obtained with VBT echelle (dashed line) compared with the spectrum obtained with 107 inch telescope, McDonald Observatory, in 1999 at a resolution of 60,000 (full line) in the Na ID line region. The spectra are matched to the interstellar components. Note the narrowness of the interstellar lines and terrestrial water vapour lines in VBT spectrum compared to McDonald spectrum due to higher spectral resolution. However the central intensity of the stellar line blend matches in both spectra remarkably.

VBT Echelle Spectrometer 337 1.8 1.6 30 Nov. 2004 23:11UT 22:49 22:27 1.4 21:43 20:58 1.2 20:36 20:13 1.0 23:33 0.8 HR3117 HeI λ6678-200 -30 0 200 400 Helio. rad. velocity (km/s) Figure 6. He I λ 6678 line and its variations with time in HR 3117. The divided profiles are arranged in increasing time sequence starting from the bottom. Note the extra absorption component at radial velocity 29 km s 1 that disappears with time. In the context of the present workshop where variability is the central theme the following examples would illustrate the present capability of the instrument. The roap star HR 1217 has been observed with a time resolution of 1 min at a spectral resolution of 28,000. The variation in the equivalent widths of a few selected lines is illustrated in the period of 12.2 min. Such variations are already well known in the star. The observation have been obtained using 1K 1K EEV CCD camera. More interesting are the observations of the suspected β Cephei star HR 3117 (χ Car). The star is supposed to show light variations with a period of 2 hours 25 minutes (Elst 1979). We have obtained spectra almost every 20 min on 30 November 2004 for a period of 3 1 hours continuously. The spectral resolution achieved is about 65,000 and the S/N 2 ranged from about 108 to 180. During this period interesting changes have been seen in the He I 6678 Å line. As shown in the Fig. 6, the broad He I line extends from 68 km s 1 to 108 km s 1 at the base. The profile shows changes mainly on the blue side becoming from an asymmetrical line to a more symmetrical line. These changes are at 3 to 4 per cent level in relative intensity. The changes are better illustrated by dividing each profile with the last profile in the sequence (and relatively more symmetrical) and the resultant profiles are arranged in a time sequence along with the dividing profile in Fig. 6. It is clear that a narrow (relatively) absorption component that was present initially at radial velocity of 30 km s 1 (the stellar velocity is given as 19.4kms 1 ) slowly disappears in about 2 h 30 min. The equivalent width initially is 12 må. Since the component already exists in our initial spectra and disappears in

338 N. Kameswara Rao et al. 2 1 2 hours suggests the phenomenon, if periodic, would have a period much longer than the canonical light period. (Handler (private communication) says it is not considered as a β Cehei star any more because of lack of detectable light variations in Hipparchus measurements.) The star has been classified as peculiar by Hiltner et al. (1969), a B3 Si star, because of silicon line strengths. The helium lines have been studied by Leone & Lanzafame (1997) and Leone & Catanzaro (1998). Both these sets of authors obtain normal abundance of helium. Leone & Catanzro (1998) determine abundances of other elements that are close to the values of main sequence B stars (including Si). Their spectra obtained on four nights in 1995 at a resolution of 13,000 showed no evidence of spectral variability. They even conclude that HR 3117 is not a peculiar star as reported in general catalogue of Ap and Am stars. Our present observations obtained at much higher resolution do show changes in He I lines. Additional observations are being obtained. The above account describes the present capabilities of the spectrometer. Further developments in improving the throughput and more importantly improving the accuracy of radial velocity measurements are planned. Acknowledgements We would like to thank the organizers of this meeting for providing a nice platform for developing possible collaborations, particularly in the area of instrumentation. We would like to express our appreciation to several colleagues and friends, too numerous to list here, who either participated or offered valuable advice and helped us one way or other in the realizations of this major project. NKR would particularly like to thank Prof. Ramsagar for his kind hospitality during his visit to ARIES. References Elst, E. W. 1979, IAU Inf. Bull. Var. Stars, 1562. Hiltner, W. A., Garrison, R. F., Schild, R. E. 1969, Astrophys. J., 157, 313. Hinkle, K., Wallace, L., Valenti, J., Harmer, D. 2000, ASP Series on Visible and Near-Infrared Atlas of the Arcturus Spectrum 3727 9300 Å, Kitt Peak National Observatory, Astron. Soc. of Pacific, San Francisco. Leone, F., Catanzaro, G. 1998, Astron. Astrophys., 331, 627. Leone, F., Lanzafame, A. L. 1997, Astron. Astrophys., 320, 893. McCarthy, J. Ki., Sandiford, B. A., Boyd, D., Booth, J. 1993, Publ. Astron. Soc. Pac., 105, 881. Rao, N. K., Reddy, B. E. 2005, Mon. Not. R. Astron. Soc., 357, 235. Rao, N. K., Sriram, S., Gabriel, F., Raghavendra Prasad, B., Samson, J. P. A., Jayakumar, K., Srinivasan, R., Mahesh, P. K., Giridhar, S. 2005, Asian J. Phys. (in press).