Stationary digital breast tomosynthesis system with a multi-beam field emission x-ray source array



Similar documents
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Digital Mammography Update: Design and Characteristics of Current Systems

Experimental study of beam hardening artefacts in photon counting breast computed tomography

Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction

Cone Beam Reconstruction Jiang Hsieh, Ph.D.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Purchasing a cardiac CT scanner: What the radiologist needs to know

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

Digital Breast Tomosynthesis QC Requirements

Fundamentals of Cone-Beam CT Imaging

Mammography. What is Mammography?

Concepts for High-Resolution Low-Dose CT of the Breast

X-ray Imaging Systems

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

X-ray Imaging Systems

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center

Proton tracking for medical imaging and dosimetry

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Development of High Speed Oblique X-ray CT System for Printed Circuit Board

Compressive Sampling Based Interior Reconstruction for Dynamic Carbon Nanotube Micro-CT

Data. microcat +SPECT

Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

X-RAY IMAGING Emerging Digital Technology - CMOS Detectors

Synthetic Sensing: Proximity / Distance Sensors

Industrial X-ray for Nondestructive Testing Unrestricted Siemens AG All rights reserved

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays

CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems

PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.

Experiment 5. Lasers and laser mode structure

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Physics testing of image detectors

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.


Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Development of High-sensitivity and High-resolution Digital Mammography System AMULET

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver /9/07) by M. Jensen and J. E.

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

This document is the author s post-print version of this article, i.e. the final draft version after review. The final document will be printed in

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

CHAPTER 11: Flip Flops

X-ray Tomography in Industrial Metrology

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

Measuring Size and Distance in Digital Mammography

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

Automation of a CT Acquisition: A System-based User-Support. 1. Introduction. Bärbel KRATZ 1, Frank HEROLD 1, Malte KURFISS 1

Performance testing for Precision 500D Classical R/F System

CT Traceability - Prof. Wim Dewulf, Group T - KU Leuven

METHODS FOR PULSED LASER DEPOSITION OF LARGE-AREA FILMS USING MORE THAN ONE TARGET

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

Title : Analog Circuit for Sound Localization Applications

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Agilent AEDB-9140 Series Three Channel Optical Incremental Encoder Modules with Codewheel, 100 CPR to 500 CPR Data Sheet

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

Radiation Strip Thickness Measurement Systems

DICOM Correction Item

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION

INFRARED PARTS MANUAL

Lecture 14. Point Spread Function (PSF)

Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge

Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.

Instruction Sheet

Quality Control of Full Field Digital Mammography Units

III. MEMS Projection Helvetica 20 Displays

The Volumetric Erosion of Electrical Contacts

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

Influence of the Socket on Chip-level ESD Testing

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION

ENGINEERING METROLOGY

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity

I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit

Zero voltage drop synthetic rectifier

Equipment. Lin muyan. Presented by: CocticaUwititonze Cell:+250 (0)

Ion Beam Sputtering: Practical Applications to Electron Microscopy

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

Rigaku XRD-System Instruction Manual v4/19/03. The Krishnan Group/Wilcox 132, University of Washington

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

Automotive Applications of 3D Laser Scanning Introduction

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Transcription:

Stationary digital breast tomosynthesis system with a multi-beam field emission x-ray source array Guang Yang a, Ramya Rajaram b, Guohua Cao a, Shabana Sultana b, Zhijun Liu a, David Lalush c, Jianping Lu a, Otto Zhou a,b,d a Department of Physics & Astronomy, University of North Carolina, Chapel Hill, NC 27599 b Curriculum of Applied & Materials Sciences, University of North Carolina, Chapel Hill, NC 27599 c Joint Department of Biomedical Engineering, University of North Carolina, Chapel Hill, NC 27599 and North Carolina State University, Raleigh, NC 27695 d Lineberger Comprehensive Cancer Center, University of North Carolina, Chapel Hill, NC 27599 ABSTRACT A stationary digital breast tomosynthesis (DBT) system using a carbon nanotube based multi-beam field emission x-ray (MBFEX) source has been designed. The purpose is to investigate the feasibility of reducing the total imaging time, simplifying the system design, and potentially improving the image quality comparing to the conventional DBT scanners. The MBFEX source consists of 25 individually programmable x-ray pixels which are evenly angular spaced covering a 48 field of view. The device acquires the projection images by electronically switching on and off the individual x-ray pixels without mechanical motion of either the x-ray source or the detector. The designs of the x-ray source and the imaging system are presented. Some preliminary results are discussed. Keywords: digital breast tomosynthesis, carbon nanotubes, field emission 1. INTRODUCTION Mammography is currently the most effective screening and diagnostic tool for early detection of breast cancer, and has been attributed to the recent reduction of breast cancer mortality rate 1. However, the nature of the two-dimensional mammogram makes it difficult to distinguish a cancer from overlying breast tissues, and the interpretation can be variable among radiologists. A higher rate of false-positive and false-negative test results exists because the dense tissues interfere with the identification of abnormalities associated with tumors. Digital breast tomosynthesis (DBT) is a threedimensional imaging technique that is designed to overcome this problem. It is a limited angle tomography technique that provides reconstruction planes in the breast using projection images from a limited angular range 2. Several prototype DBT scanners have been manufactured by commercial vendors including GE 3, Hologic 4 and Siemens 5. The system designs are based on the current full-field digital mammography (FFDM) unit. A mammography x-ray tube is used to collect the projection images by moving 10-50 degrees around the object. The reported total scanning time is 7s - 40s depending on the number of views and the thickness of the breast, which is much longer than that of the regular mammography. The long imaging time causes patient motion blur which degrades image quality and make patients uncomfortable. The power of the x-ray source, gantry rotating speed and detector frame rate limit the scanning speed of the current DBT systems as illustrated by the timing diagram of the DBT system in the continues motion mode in Fig. 1. In this case X-ray radiation is only delivered within a small portion (<10%) of the total imaging time. All the prototype DBT systems utilize the standard mammography x-ray tube with a ~300µm x-ray focal spot size. Due to the gantry rotation and mechanical instability, the effective focal spot size during image acquisition is larger than the static value which degrades the image resolution. Two gantry rotation modes have been developed. The GE system uses a stop-and-shoot method. The gantry makes a full stop before taking each projection image. Acceleration/deceleration may cause mechanical instability of the system 5. The extent of this effect on the system resolution has not been fully evaluated. A continuous rotation mode is used in the Siemens and Hologic systems. The gantry keeps a constant rotation speed during the whole imaging process. In this case, the x-ray focal spot size is enlarged along the motion direction. The schematic is shown in Fig. 2. The value of the enlargement depends on the rotation speed, and the exposure time. Medical Imaging 2008: Physics of Medical Imaging, edited by Jiang Hsieh, Ehsan Samei, Proc. of SPIE Vol. 6913, 69131A, (2008) 1605-7422/08/$18 doi: 10.1117/12.770622 2008 SPIE Digital Library -- Subscriber Archive Copy Proc. of SPIE Vol. 6913 69131A-1

Ren et al. 4 reported the x-ray focal spot moves ~1mm in a typical scan. The effect of the focal spot blurring on the system MTF was found to be not significant at this speed especially for thin specimen 4. However it does not lead room for further reduction of the total scanning time, which will require a faster gantry rotation and a larger focal spot blurring. T r Detector T int T exp X-ray output Time 0 T 1 T 2 T n-2 T n-1 T n Fig. 1: Timing diagram of DBT imaging. The total imaging time is determined by n*(t int +T r ), where n is the view number, T int is the detector integration window, and T r is the detector readout time. T exp is the exposure time for each projection image. The total exposure time is less than 10% of the total time in DBT imaging. f Movement of x- ray focal spot θ θ Iso-center Fig. 2: Schematic of the focal spot enlargement of continuous rotation mode. The gantry rotates angle θ between two imaging positions. During the x-ray radiation period, the gantry rotates angle θ and moves f along the arc. f is the additional x-ray focal spot size induced by gantry rotation. The value is determined by the x-ray radiation period and gantry rotation speed. For a typical scan 4, θ=3, θ=0.12, and f=1mm. The purpose of this study is to investigate the feasibility of constructing a stationary DBT system which can increase the scanning speed, simplify the system design and potentially enhance the image quality. The enabling technology is the carbon nanotube (CNT) based multi-beam field emission x-ray (MBFEX) recently developed in our lab 6-8. For demonstration of principle a stationary DBT scanner was designed and constructed at UNC using a spatially distributed MBFEX source with 25 individually controlled x-ray pixels. The device, called Argus, acquires the projection images without mechanical motion of the sources or the detector. In this paper the basic designs of the source and the system are described and the some preliminary results are discussed. Proc. of SPIE Vol. 6913 69131A-2

2. SYSTEM DESCRIPTION 2.1 System geometry A picture of the assembled Argus system is shown in Fig. 3(a). The system is composed of a CNT MBFEX source, a flat panel x-ray detector (Varian Medical System, Paxscan 2520), a control unit, and a computer work station. The system geometry follows the typical values for the regular mammography devices, as shown in Fig. 3(b). The distance between the center of the phantom and the x-ray source is 64.5cm. The source to detector distance is 69.6cm, which leaves a 2.5cm air gap for a normal 5-cm breast phantom. The x-ray pixels are arranged linearly to reduce the system complexity, with even-angular distribution and a 2-degree increment. The total angular coverage of the x-ray source is 48 degrees. In such a design, the distance between the nearest x-ray focal spots varies from 2.5cm to 2.7cm, and the total span of the x-ray source array is 57.5cm. (a) I I ii Detector X-ray source array Phantom (b) X-ray source array Detector Phantom D S-O : 64.5 cm 57.5 cm 48 coverage 2 step D S-D : 69.6 cm 2.5 cm Fig. 3: (a) Picture of the Argus system. The x-ray source array, breast phantom, and detector are shown. (b) Geometry of the Argus system. The 25 x-ray focal spot positions are represented by 25 points. 2.2 X-ray source array The key component of the Argus system is the field emission x-ray source array. The construction of the 25 x-ray pixels is identical. The CAD drawing of one pixel is shown in Fig. 4. The central beams of all the 25 x-ray sources intersect at a Proc. of SPIE Vol. 6913 69131A-3

common iso-center, which is also the center of the phantom. Each x-ray pixel consists of one CNT cathode, one gate electrode, two focusing electrodes, and one anode. The design of each x-ray pixel is similar to that of the field emission micro-focus x-ray source we have published previously9. The cathode is grounded, and all other electrodes are maintained at constant voltages during imaging acquisition. The gate voltage determines the x-ray tube current. Below a threshold, there is no current, and the current increases exponentially with gate voltage once over the threshold. The electron trajectory is controlled by the voltages applied to the two focusing electrodes. The anode is made of molybdenum. The x-ray source array is housed in a customized chamber. A 30µm molybdenum film serves as the x-ray window as well as the filter. The molybdenum anode and filter combination is commonly used for breast imaging10. Up to 40keV high voltage can be applied on the anode. The targeted performance for the source design is that each x-ray pixel can provide 10mA peak current at 200µmx200µm effective focal spot size. Fig. 4: CAD drawing of one x-ray pixel in the array. Each pixel consists of a CNT cathode, a gate electrode, two focusing electrodes, and one molybdenum anode. Fig. 5: An assembled MBFEX source array with 25 individually controllable x-ray pixels which are tilted towards the isocenter. A control unit based on MOSFET (metal oxide semiconductor field-effect transistor) is built to switch the x-ray sources. The basic design of the electronics switching system is similar to the one used in our first demo MBFEX source11. The diagram is shown in Fig. 6. The drains (D), gates (G), and sources (S) of the 25 MOSFET s are connected to the 25 cathodes, 25 TTL trigger signals (provided by the computer board), and a common ground, respectively. When the TTL trigger signal is at low state, the conduction channel between the source and the drain is closed. This causes the CNT cathode potential floating relatively to the common ground, and no electron is emitted from the CNT cathode and no x-ray radiation is generated. When the TTL trigger signal is at high state, the cathode is grounded because of the opened conduction channel. As a result, electrons are extracted by the electrical filed between gate electrode and cathode, and x-ray radiation is produced. The delay time (between switching of the TTL signal and the conduction channel) of the MOSFET is 35-45 ns, which is sufficient considering the tens of milliseconds x-ray exposure period. The 25 x-ray pixels can be switched individually at any given time during the imaging acquisition, which provides great flexibility. Variable resistors are built-in for compensation of the variations in the individual cathode performance. Proc. of SPIE Vol. 6913 69131A-4

0-40Kv V 125KΩ Anode 0-2.5Kv 10KΩ V 0-2.5Kv 10KΩ V Focusing 2 Focusing 1 0-2.5Kv V 10KΩ Gate Cathode G D G D Variable resistor G D MOSFET A S S S Fig. 6: Schematic of the control unit. Through the control unit, each x-ray source can be switched individually by the computer. Computer board 2.3 Detector and image acquisition software Detector trigger 5V 0V T int T r Source 1 trigger Source 2 trigger On Off On Off T exp Source 24 trigger Source 25 trigger On Off On Off Time 0 C 1 C 2 C 23 C 24 C 25 (T total ) Fig. 7: Timing diagram of Argus system. The detector readout and 25 x-ray sources are triggered by TTL signals provided by computer board. The x-ray sources are turned on sequentially, and readouts are triggered after each radiation. A flat panel detector (Varian Medical System, Paxscan 2520) is used for imaging acquisition. The field of view is 19.5cm by 24.4cm, which ensures a full image of the breast. With a 127µm pixel pitch, the total array size is 1536 by 1920. The detector can run under non-binning mode and 2 by 2 binning mode. In the user synchronization mode, the rising edge of a continuous TTL signal triggers the readout of the detector, as illustrated in Fig. 7. The imaging time is determined by the integration window T int and the detector readout time T r. T int is controllable through triggering signal. X-ray radiation is delivered within the integration window T int, and the radiation period is denoted as T exp. The readout time T r depends on the acquisition mode. For the normal and 2 by 2 binning modes, the readout time is 128ms and 32ms, Proc. of SPIE Vol. 6913 69131A-5

respectively. In the Argus system, the 25 projection images are taken sequentially. The total imaging time can be calculated by 25*(T int + T r ). An imaging acquisition software is developed using LabView. The software controls a computer board to trigger both the x-ray pixels and the detector, and also acquires the x-ray images from the detector. The exposure time and tube current of each x-ray source can be adjusted through the software. The software is designed for sequential imaging protocol and can be modified for other imaging sequences. 2.4 Reconstruction program To reconstruct the slice images, an iterative ordered-subset convex (OSC) algorithm 12 based on a maximum-likelihood (ML) model is used. The reconstruction technique applies a shearing method to convert all projection images to a common frame of reference; then uses a pre-computed cone-beam model to project and back-project in the common frame. To reduce the computational load, non-cubic voxels are reconstructed. This technique has been verified on both simulated data and breast phantom images measured from a field emission x-ray source array with limited number of pixels 13. 2.5 Comparison with the prototype systems The designed system performance of the UNC Argus DBT system is listed in Table 1, along with the specifications of 3 prototypes from commercial vendors based on the information from recent scientific publications 3-5. The potential advantages of this system include the followings. 1: The focal spot size of Argus system is 200µm while the values of other systems are 300µm or larger. 2: Stationary design gives less gantry vibration by eliminating the mechanical movement. 3: The exposure time matches the detector integration window. In other words, the utilization of the imaging time is the most efficient. 4: The targeted total scan time (8.8s in binning mode and 11.2s in full-resolution mode, 25 viewing angles) is shorter which can be further reduced by increasing the x-ray tube current which requires relaxing of the focal spot size. Table 1: Specification of Argus and that compares with other prototype DBT systems. UNC: Argus GE: Senographe 2000D Siemens: Mammomat Novation Hologic: Selenia X-ray kvp, ma 25-35kVp, 10mA 25-30kVp, ~130mA ~28kVp, ~180mA 24-39kVp, ~100mA Focal spot size 200µm 300µm 300µm + blur* 300µm + blur* Target/filter Mo/Mo Mo/Mo, Rh/Rh W/Rh (Mo, W)/(Rh, Al) Angle coverage 48 degrees 50 degrees 50 degrees 30 degrees View numbers 25 11 25/49 11 Gantry motion Stationary Step and shoot Continuous rotation Continuous rotation Flat-panel Detector A-silicon Cs:I a-silicon Detector size 19.5 x 24.4 cm pixel pitch: 127µm 18.00 x 23.04 cm pixel pitch: 100µm Direct converter a- selenium 23.9 x 30.5 cm pixel pitch: 85µm Readout time 0.128s/0.032s 0.3s 0.6s/0.3s 0.6s Integration time 0.32s 0.4s 0.2s 1.0s Exposure time 0.32s ~0.1s ~0.03s 0.073s Direct converter a- selenium 24 x 29 cm pixel pitch: 70µm (140µm for DBT) Total scan time ** 11.2s for 25 views 7s for 11 views 20s/39.2s for 25/49 views 18s for 11 views Reconstruction method ordered subsets convex (maximum likelihood) ML-EM FBP: filtered back projection * : Additional focal spot blur due to the gantry movement during exposure. ** : Total scan time = (view number) x (cycle time); cycle time = (readout time) + (integration time). FBP: filtered back projection Proc. of SPIE Vol. 6913 69131A-6

3. PRELIMINARY SYSTEM CALIBRATION 3.1 Spectrum measurement The energy spectrum of the x-ray source was measured at 28keV using a Si-pin photodiode detector from Amptek (XR- 100CR and MCA-8000A). The spectrum is consistent when measured at different locations within the field of view, and from different x-ray pixels. The experimentally measured energy spectrum is shown in Fig. 8, which agrees well with typical Mo/Mo x-ray spectrum 10. Two molybdenum characteristic peaks, one at 17.5keV and the other at 19.6keV, are well recognized. 12000 10000 17.48 8000 Count 6000 4000 2000 19.61 0 0 5 10 15 20 25 30 kev Fig. 8: The experimentally measured energy spectrum of the UNC Argus system at 28keV. 3.2 X-ray tube current Anode current (ma) 4.0 3.5 3.0 2.5 2.0 1.5 1.0 0.5 0.0 400 600 800 1000 1200 1400 Gate voltage (V) Fig. 9: A typical curve of anode current as a function of gate voltage. The threshold value for this x-ray source is about 650V. The emission current from the CNT cathode depends on the electrical field between the gate and the cathode following the known Fowler-Nordheim equation 14. In this particular x-ray source, 72% of the total current passes through the gate electrode and reaches the anode to produce x-ray radiation (also denoted as anode current). Fig. 9 shows the typical anode current versus gate voltage data measured from one pixel in the MBFEX source. Due to the voltage limitation of Proc. of SPIE Vol. 6913 69131A-7

the electronic devices, presently the maximum gate voltage can be applied is 1500V, which limits the anode current to ~4mA, below the targeted value of 10mA. This can be overcome in the future by changing the design and/or optimizing the CNT cathode (when measured in a separate setup, the cathodes fabricated under same conditions can produce over 10mA consistently at higher gate voltages). So far nine of the pixels in the MBFEX source have been characterized. Due to variation in the CNT cathodes, the gate voltages needed to obtain the same current are different. As a reference, the voltages vary from 925V to 1465V for 1mA tube current (Table 2). The voltage difference can be compensated by the variable resistors in the control circuit. With the improvement of fabrication technique and cathode quality control, the variation can be reduced in the future. The current stability was determined by measuring the current of 100 pulsed x-rays at constant voltage. The standard deviation of the current is less than 0.03mA for all pixels tested. Table 2: Gate voltages required to obtain 1mA tube current and the tube current fluctuation at the same gate voltage. X-ray source # Gate voltage (V) Standard deviation of current (ma) 1 1230 0.02 2 925 0.01 3 1230 0.02 4 1015 0.01 5 1300 0.03 6 1070 0.01 7 1160 0.01 8 1465 0.02 9 1030 0.01 3.3 Focal spot measurement The designed x-ray focal spot size is 200x200µm for all 25 x-ray sources. The actual values were measured following the European standard EN12543-5 15. A customized cross wire phantom made of 1mm tungsten wire was fabricated to measure the focal spot size along two orthogonal directions simultaneously. The phantom was placed close to the x-ray source to obtain the large magnification factor. The voltages applied to the two focusing electrodes were first varied to optimize the focal spot size. It was found that the optimal focal spot size is achieved when the two focusing electrodes are at 500V and 1600V, respectively. A typical projection image of the cross phantom is shown in Fig. 10 (a). The line profiles are shown in Fig. 10 (b). 300 Intensity 200 100 X axis Y axis (a) 0 0 20 40 60 80 100 Pixel Number Fig. 10: (a) A typical x-ray projection image of the cross wire phantom. (b) Line profiles of the two wires. X axis is the direction of the x-ray source array, and Y axis is perpendicular to it. (b) Focal spot sizes of 9 x-ray pixels have been measured. As listed in Table 3, the results agree well with the designed specification 0.20 x 0.2mm. The x-ray sources have isotropic focal spot with average value of 0.19mm. Measurements from different x-ray sources are also consistent. Proc. of SPIE Vol. 6913 69131A-8

Table 3: Focal spot size measurement of 9 x-ray pixels. The uncertainty of the measurement is 0.02mm. X-ray source # F x : parallel to x-ray F y : perpendicular to x-ray source array source array 1 0.20mm 0.20mm 2 0.20mm 0.17mm 3 0.18mm 0.19mm 4 0.19mm 0.19mm 5 0.20mm 0.19mm 6 0.19mm 0.17mm 7 0.18mm 0.17mm 8 0.19mm 0.19mm 9 0.18mm 0.19mm 3.4 Geometry calibration The tomosynthesis reconstruction requires precise system geometry parameters. We applied an analytic method based on identification of ellipse parameters 16 for the geometry calibration, which was first established for cone-beam CT calibration. A phantom with two point objects with known distance was machined. The geometry parameters of the 25 x- ray sources were calibrated individually. Six projection images of the phantom (60-degree rotation in-between) were acquired for each pixel. The traces of the two balls form two ellipses on the detector plane. The parameters, including the source-detector distance and x-ray source offset values on the detector plane, can be further calculated based on these elliptical curves. The source-detector distance is calculated to be 69.3cm with 2mm uncertainty. The distances between the x-ray sources are also calculated. The results agree with the design values within 1mm uncertainty. 4. CONCLUSION AND DISCUSSION A stationary digital breast tomosynthesis system with a CNT based distributed x-ray source has been designed and assembled, with comparable geometry to the conventional mammography devices. The system has been fully assembled. The control unit, imaging software, and the calibration procedure have been developed. The preliminary test performed so far shows that the system functions as designed. Detailed system calibration and tomosynthesis imaging test will be performed in the near future. ACKNOWLEDGEMENTS This work was partially supported by grants from NIH-NCI (U54CA119343) and the UNC Lineberger Comprehensive Cancer Center. We acknowledge helpful discussions with Elodia Cole and Etta Pisano. REFERENCES 1 Bassett, L.W. and et al., [Diagnosis of Diseases of the Breast] 2 ed., Elsevier Saunders, Philadelphia, PA (2005) 2 Dobbins, J.T. III and Godfrey, D.J., "Digital x-ray tomosynthesis: current state of the art and clinical potential". Phys. Med. Biol. 48: R65-R106 (2003) 3 Wu, T. and et al., "Tomographic mammography using a limited number of low-dose cone-beam projection images". Med. Phys., 30(3), 365-380 (2003) 4 Ren, B. and et al., "Design and performance of the prototype full field breast tomosynthesis system with selenium based flat panel detector". Proc. SPIE, 5745, 550-561 (2005) Proc. of SPIE Vol. 6913 69131A-9

5 Bissonnette, M. and et al., "Digital breast tomosynthesis using an amorphous selenium flat panel detector". Proc. SPIE, 5745, 529 (2005) 6 Zhang, J. and et al., "Stationary scanning x-ray source based on carbon nanotube field emitters". Appl. Phys. Lett., 86, 184104 (2005) 7 Zhang, J. and et al., "Multiplexing radiography based on carbon nanotube field emission X-ray technology". Proc. SPIE, 6510(65100W) (2007) 8 Zhang, J. and et al., "Multiplexing radiography using a carbon nanotube based x-ray source". Appl. Phys. Lett., 89, 064106 (2006) 9 Liu, Z. and et al., "Carbon nanotube based microfocus field emission x-ray source for microcomputed tomography". Appl. Phys. Lett., 89, 103111 (2006) 10 Beutel, J. and et al., [Handbook of Medical Imaging], SPIE Press, Bellingham, WA (2000) 11 Zhang, J. and et al., "A stationary scanning x-ray source based on carbon nanotube field emitters". Appl. Phys. Lett., 86, 184104 (2005) 12 Kamphuis, C. and Beekman, F.J. "Accelerated iterative transmission CT using an ordered subsets convex algorithm". IEEE Trans. Med. Im., 17, 1101-05 (1998) 13 Lalush, D.S., and et al., "Tomosynthesis reconstruction from multi-beam x-ray sources". Proceedings of 2006 IEEE International Symposium on Biomedical Imaging, 1180-1183 (2006) 14 Gomer, R., [Field Emission and Field Ionization], Harvard University Press, Cambridge, MA (1961) 15 European Committee for Standardization: EN 12543-5, E. Brussels. (1999) 16 Noo, F., and et al., "Analytic method based on identification of ellipse parameters for scanner calibration in conebeam tomography". Phys. Med. Biol., 45, 3489-3508 (2000) Proc. of SPIE Vol. 6913 69131A-10