Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.

Size: px
Start display at page:

Download "Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D."

Transcription

1 Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D. Introduction Screen-film image receptors have been the standard detector used in conventional mammography. New developments in detector technology and computers are altering the landscape of mammography imaging. Full Field Digital Mammography (FFDM) offers the promise of revolutionizing the practice of mammography through its superior dose and contrast performance. Advanced applications made possible through digital imaging, such as automated computer-aided diagnosis, dualenergy and 3D tomosynthesis are expected to further improve diagnostic sensitivity and specificity. This primer describes the technical basis for current and future advances in mammographic detector technology. These include all of the following: Lower dose Improved image quality Computer-aided diagnosis Softcopy review and digital archiving Tomosynthesis and other three-dimensional visualization techniques Reduction in breast compression pressure New flat-panel detectors offer extremely high quantum efficiency and high resolution. These will translate into lower dose and improved image quality mammograms. Digital detectors for mammography can be categorized as indirect or direct conversion detectors. This primer describes these technologies and how direct conversion detectors, in particular seleniumbased detectors, provide the best performance for full field digital mammography systems. Indirect conversion digital detectors utilize a method of imaging s, similar to screen-film, wherein a scintillator absorbs the s and generates a light scintillation. The light scintillation is then detected by an array of photodetectors. However, indirect conversion systems suffer from resolution degradation caused by light spread in the scintillator, and from poor quantum efficiency caused by the use of thin scintillators. Direct conversion digital detectors utilize a direct-conversion method of imaging, wherein the s are absorbed and the electrical signals are created in one step. Systems using amorphous selenium represent the most advanced direct conversion technology for digital mammography. Selenium is an ideal material for a mammography detector, because it has high absorption efficiency (approaching 100 percent at mammographic energies), extremely high intrinsic resolution, low noise and a well-established manufacturing process. The pixel size requirement for FFDM is also addressed in this article. Light blurring in indirect conversion systems interferes with the effective design and performance of systems with very fine resolution. Direct conversion systems, in contradistinction, can have their resolution characteristics optimized independently from their quantum detection efficiency and can thus be designed with an optimum pixel size. The field of view for FFDM systems is very important. In order to be able to image most of the adult female population, the imaging field of view must be similar to the size of the largest screen-film cassette commonly used in screen film imaging, 24 x 30 cm. Detectors that are the size of the smaller cassette, 18 x 24 cm, will require imaging and tiling of many images to cover the breast field of view for larger breasts. For the U.S. demographics, up to 30 percent of all women require the larger field of view. Detector performance is commonly quantified by two metrics: Modulation Transfer Function (MTF) and Detective Quantum Efficiency (DQE). MTF is a measure of resolution and DQE is a measure of dose efficiency.the MTF for screen-film is superior to indirect conversion detectors, while the MTF for direct conversion detectors are superior to both. The DQE of indirect conversion detectors is superior to screen-film, however the DQE of selenium direct conversion detectors is seen to be superior again to both screen-film and indirect conversion detectors. With superior DQE at all spatial frequencies, selenium detectors offer the potential for both improved image quality and lower patient dose. Digital images offer a variety of new and improved applications. Contrast enhancement of the digital image and the wide dynamic range of digital detectors will improve visibility of mammographic features. The digital image will provide image archiving and retrieval advantages over film, and will facilitate the use of computer-aided diagnosis. Systems with high quantum efficiency, especially at increased energies, offer the possibility of decreased breast compression. As is commonly known, painful breast compression is an important factor in undercompliance of women to present for routine breast screening.

2 Future applications such as stereo mammography, breast tomosynthesis, contrast enhanced digital mammography and other imaging modalities, are under investigation. Such advances in technology will provide improved diagnostic information and reduced image confusion from overlapping structures. These three dimensional imaging tasks will benefit from the high quantum efficiency that selenium detectors offer. The digital detectors will also be able to be used for full-field and high resolution stereotactic breast biopsies and diagnostic imaging tasks. It is essential, however, that the full-field digital systems perform well for its primary task, i.e. breast screening. In conclusion, digital detectors offer many advantages compared to screen-film imaging for mammography. It is hoped that the potentials for improved image quality, lower dose and advanced imaging applications will result in improved diagnostic accuracy. Screen-Film Mammography As shown in Figure 1, conventional film systems use intensifying screens to capture s and reduce radiation dose. s that pass through the tissue are collected by phosphor screens. These screens are often constructed of rare earth phosphors such as gadolinium oxysulfide (Gd 2 O 2 S) that output light upon absorption of s.when an is absorbed, the resultant light scintillation creates a number of light photons that spread and illuminate the film in a distribution cloud. Film in close proximity to the screen captures the light photons, and the image is obtained by exposing the film. An important parameter to understand is the thickness of the intensifying screen. Thicker screens capture more s and are therefore more dose efficient and higher speed. However, thicker screens also create more light scatter and blurring of the image. Therefore, it is impossible to offer a screen-film system simultaneously offering the highest possible resolution and lowest possible radiation dose. This trade-off between radiation dose and image quality must be optimized for the specific clinical application. Figure 2 shows the performance trade off of sensitivity versus resolution inherent in the design of screen-film systems. As the screen is made thicker, the cloud of light on the film will increase in size, on average. This reduces the resolution of the system, however the system s sensitivity increases because the thicker phosphor increases the probability that the incoming will be absorbed. Thus screen-film systems have a performance trade off between speed and resolution. Because s are absorbed with a decaying exponential spatial distribution, in a screen-film mammography system the film is placed at the entrance surface of the scintillating screen. While a screen-film system offers several advantages, there are significant disadvantages and limitations of this system. Film does not have a linear sensitivity to photonflux, and there is a narrow range over which it can detect small differences in contrast. In particular, tissue areas of high and low density are often suboptimally imaged. Frequently, the entire image is poorly exposed because of film s stringent requirements for proper exposure, resulting in repeated imaging. Another major problem is film granularity, affecting detective quantum efficiency at high optical densities and visibility of microcalcifications. Film also requires processing time and storage space. Chemical processing of the film adds time to the exam, and the resulting films require a large amount of storage space in patient health records. Film also must be physically transported to the physician for viewing. Digital Mammography Detectors Digital technology offers the potential for several advances in mammography detectors. Because images are captured as a digital signal, electronic transfer and storage of images is possible, eliminating physical storage and distribution required by film. Digital systems offer a large dynamic range of operation, improving visualization of all areas of the breast and increasing exposure latitude. Also, the digital format allows grayscale Light scintillation ~84 µm film Scintillating screen, Gd2O2S Screen/film: low speed, high resolution Screen/film: high speed, low resolution Line spread function Line spread functions Figure 1. Screen-film systems use a scintillating screen to absorb the and generate light photons, which are captured on the film. Figure 2. As the thickness of the screen is increased, its speed increases at the expense of poorer spatial resolution.

3 adjustment to optimize contrast for every imaging task. Softcopy reading, computer-aided diagnosis and three-dimensional imaging offer additional and potentially important opportunities for improvement in mammographic systems. Digital Detector Technology There are two methods of image capture used in digital mammography that represent different generations of technology: indirect conversion and direct conversion. Indirect-Conversion Digital Detectors The earliest digital mammography systems available in the U.S. used indirect conversion detectors. Such detectors use a two-step process for detection, as seen in Figure 3. The first step requires a scintillator layer such as cesium iodide doped with thallium [CsI(Tl)] to capture the energy and convert it to light. An array of thin-film diodes converts light photons to electronic signals that are captured using thin-film transistors. Some systems, like Charge-Coupled Devices (CCDs), use alternative light collection and readout methods. In both systems, the light sensitive imaging function of film has been replaced by digital light imaging. It is in this sense that these can be seen as an evolution of screen film imaging. Similarly to screen-film, light scatter compromises image quality, and there is a performance tradeoff between spatial resolution and radiation sensitivity, as shown in Figure 4. As the scintillator is made thicker, light spread increases resulting in decreased resolution. Because of its columnar structure, CsI(Tl) does not create as much light scatter as other screens. However, compromise between resolution and sensitivity still exists. The placement of the scintillator is more problematic in indirect conversion digital detectors than with screen-film systems. As with film screens, more s are absorbed near the entrance of the scintillation layer than the exit. While film is placed near the entrance side of the scintillator, a photodiode/transistor array is not transparent to s and the array must be placed on the exit surface of the scintillator. This causes degradation in spatial resolution compared to screen-film. Typical thicknesses of CsI(Tl) used in mammography detectors range from 150 to 250 microns, and these indirect conversion digital detectors exhibit light spreading similar to screen-film systems. Examples of indirect conversion detectors are the Fischer Imaging CsI/CCD-based detector and the GE CsI/TFT detector. Direct-Conversion Digital Detectors Direct-conversion digital detectors represent a technological advance, eliminating problems associated with light scatter inherent in indirect conversion systems. In these systems, illustrated in Figure 5, a photoconductor absorbs the s and directly generates the signal (direct conversion). Under the influence of an external electric field, holes (or electrons, depending upon the polarity of the applied field) drift towards a pixel electrode and are collected on a pixel capacitor. Because the electrons and holes travel along the direction of the electric field lines, they move without lateral charge spreading. This results in an exceptionally narrow point spread response, of about 1 micron. The superior photoconductor for use in direct conversion systems is amorphous selenium (a-se). Selenium has a long commercial history in xerography, and its manufacturing processes are well known and optimized. The image quality of xeromammography systems was widely acknowledged, but it suffered from reliability problems as a result of mechanical wear of the plates during toner deposition and crystallization of the selenium during the high-temperature erasure cycles. By depositing selenium on a flat-panel imaging receptor, these problems have been eliminated. In direct-conversion detectors, the response function maintains its sharpness even as the thickness of the photoconductor is increased, so there is no trade off between radiation stopping power and spatial resolution. This is shown in Figure 6. In practice, the photoconductor is made sufficiently thick in order to stop µm Light scintillation Columnar CsI (T1) µm Pixel array Indirect conversion: low speed, high resolution Indirect conversion: high speed, low resolution Pixel array Line spread function Line spread functions Figure 3. Indirect conversion detectors utilize a scintillating layer to absorb the and generate light photons, which are detected by a photodiode array. Figure 4. Indirect conversion detectors generate poorer resolution images as the scintillator thickness is increased.

4 250 µm Line spread function Electron/hole pairs Photoconductor amorphous selenium Pixel array the majority of the incident s, and this can be done without adversely affecting the spatial resolution, an important consideration in mammography with its dual needs of high resolution and low radiation exposure. Using amorphous selenium as the photoconductor, a thickness of 250 microns is adequate to stop more than 95 percent of the s in the mammographic energy range. This is seen in Figure 7. Standard screens for use in film mammography only have about 50 to 70 percent quantum efficiency, and the scintillator CsI(Tl) used in indirect-conversion digital detectors exhibits about 50 to 80 percent quantum efficiency. Systems using amorphous-selenium can achieve almost complete quantum efficiency. Figure 5. Direct-conversion detectors use a photoconductor to absorb the and directly generate the signal. Direct conversion: high speed & resolution Line spread functions Direct conversion: high speed maintained even at higher speed Figure 6. There is no tradeoff between spatial resolution and sensitivity for direct-conversion systems. attenuation mu/rho (cm**2/g) energy (KeV) Figure 7. K-edge for Selenium is just below the diagnostic range for mammography, shown in the area between the two yellow lines. Pixel Design for Digital Detectors Digital detectors require an array of pixels that collect electronic signals. The signals on these pixels are transferred to a computer during a readout sequence. This is known as direct readout, a function of all digital systems, and should not be confused with direct conversion digital detection. Thin film transistor (TFT) arrays are the active electronic readout mechanism commonly used in both direct- and indirect-digital radiography systems. The arrays are typically deposited onto a glass substrate in multiple layers, beginning with readout electronics at the lowest level, followed by charge collector arrays at higher levels. The composition of the top layer depends upon the type of detector. If the system uses indirect conversion detection, both elements and light-sensitive elements are deposited on the top layer. Direct conversion detectors do not require conversion of s to light, so light sensitive elements are not necessary for these systems. Semiconductor arrays for direct conversion detectors are much easier to fabricate than arrays for indirect conversion detectors because selenium-based detectors do not require a photodiode structure on top of the TFT. As a result, these detectors can utilize the same manufacturing processes as large area TFT- LCD arrays, commonly used in computer displays and do not require construction of dedicated manufacturing facilities. Charge-coupled devices (CCDs) are an alternative to TFT arrays in indirect conversion systems. Basic CCD-based systems consist of a series of metal oxide semiconductor capacitors that are fabricated very close together on the semiconductor surface. These systems use fiber optics to capture images from light emitted from scintillators or intensifying screens, but suffer from light loss and added complexity due to the fiber optics. The only commercially available CCD-based system uses a slot scanning technology,whereby a small detector is scanned across the breast to build up the image over several seconds. This is in distinction to the other FFDM systems that acquire the image in a single snapshot.

5 Screen/Film MinR Gd2O2S Indirect Conversion Columnar CsI (T1) µm Direct Conversion amorphous-selenium resolution and system cost. As the pixel size of any digital imaging system is made smaller, the amount of data contained in the image rapidly increases, which increases system costs in terms of data storage, network bandwidth and display capabilities. Fabrication of large-area detectors with very small pixels is an expensive, low-yield process, and integration of such detectors with analog and digital readout electronics and system interconnects pushes the state of the art in microfabrication on several fronts. Line spread functions Figure 8. Line spread functions for screen-film and digital detectors. Screen-film and indirect-conversion detectors have broad presampled line spread functions; direct-conversion detectors have narrower presampled line spread functions. Pixel Size Requirements Mammographic imaging requires the detection and classification of extremely small objects. In particular, microcalcifications can be as small as 100 to 200 microns. Any useful FFDM system must be able to image these smallest microcalcifications of interest. Digital detectors are comprised of arrays of pixels. The smallest feature that can be resolved in any digital imaging system is a function of the pixel size the smaller the pixel the higher the resolution. Figure 8 shows the intrinsic resolution for direct and indirect detectors. In indirect-conversion digital detectors, as the pixel size is decreased, a limit is reached beyond which further reductions in pixel size does not improve resolution. This resolution limit is a function of the light scattering in the scintillator. This limit is about 100 microns in practical indirect conversion scintillators, which is the upper limit for application to digital mammography. CCD based indirect conversion detectors typically have much smaller pixels, however the light spread from the scintillator limits the resolution so the result is an image with more pixels but not superior resolution. In a direct-conversion digital detector, on the other hand, spatial resolution is only limited by the size of the pixel. The size of the pixel in these detectors can be made arbitrarily small (within limits) to make the resolution performance extend to very high spatial frequencies. The ultimate limit to very small pixels is the reduced flux impinging upon the detector. The pixel size for FFDM systems range from about 50 to 100 microns. As explained earlier, the indirect-conversion detectors all have resolution characteristics that are scintillator limited, so the raw pixel size does not accurately reflect the actual resolution characteristics of the image. The pixel size for the Hologic s selenium detector is 70 microns, and because of the design of this detector, this represents its true resolution characteristic. Pixel size design represents a compromise between spatial Field of View Requirements MQSA requirements for mammography facilities require the use of at least two screen-film cassette sizes: 18 x 24 cm and 24 x 30 cm. The rationale for this requirement is the large distribution in compressed breast sizes. A small cassette cannot accommodate large breasts in the field of view, and two overlapping exposures will be required to ensure adequate coverage of all the breast tissue. This is costly, time consuming, and subjects the women to additional radiation dose. On the other hand, the use of a large cassette on smaller women is wasteful of film and processing materials. Digital mammography presents additional field-of-view issues. At the present time, flat-panel detectors are considerably more expensive than screen-film cassettes, precluding the use of more than one receptor size on a given machine. Irrespective of cost, digital detectors are heavy and fragile and do not lend themselves to rapid changeover as do film cassettes. Therefore, digital mammography equipment manufacturers must decide on one optimum detector size, one that is large enough to accommodate most women without requiring double exposures. A digital detector of size 18 x 24 cm (the smaller film cassette size) is inadequate to image approximately 20 to 30 percent of U.S. women. Because of regional differences, some areas of the country might have a larger percentage of women that cannot be screened using the smaller film cassette. If the breast is too large to image on the detector in one exposure, multiple exposures that tile the breast are required. This solution has disadvantages such as requiring multiple additional compressions, additional setup and imaging times, breast regions that suffer from repeated radiation exposure, and the difficulty of radiologist s review of multiple images. Use of a large detector for digital mammography is an important design issue. Issues using a large digital detector to image a small breast does not present the same problems as film. The large image can be cropped to reduce data transfer, display and storage space requirements, and there are no additional material costs. The one technical challenge in the use of a large detector for small breasts is in positioning, however the use of smaller compression paddles in conjunction with the large detector solves this problem. The larger detector can be used to advantage when imaging smaller breasts. The imaging can be performed in a magnification mode, creating the effect of smaller pixels and reducing scattered radiation at the same time. If a system is designed so that breasts

6 that are adequately served by the smaller film cassette size of 18 x 24 cm are imaged in magnification mode to project onto a detector of size 24 x 30 cm, an image of effective pixel size of 56 microns will be achieved for systems with inherent 70 mm pixel sizes. Such a system could offer 56 micron imaging in a screening environment for approximately 80 percent of U.S. women. This is another possibility offered by use of a large digital detector. Characterizing Detector Performance There are parameters that characterize the overall image quality obtained on mammography and other radiography systems. The most important are: Spatial resolution Contrast resolution Signal-to-noise ratio Dose efficiency Dynamic Range Screen-film has a limited dynamic range, which prevents visualization with equal clarity of all breast tissue regions from the chest wall to the skin line. Figure 9 shows the need for improved dynamic range in breast imaging. Digital detectors offer greatly improved performance. It has been shown that for an ideal detector with no inherent detector noise, 3100 gray levels are discernable in a typical mammographic image. Thus a system that provides at least 12 bits of dynamic range will not degrade the underlying information. Digital mammograms will Commercially Available Digital Mammography Detectors The following is a list of full-field-of-view digital mammography detectors commercially available in the U.S. Indirect conversion detectors GE Scintillator CsI(Tl) Pixel size TFT 100 microns Field of view 18 x 23 cm Fischer Imaging Scintillator Pixel size Field of view CsI(Tl) CCD 24/48 microns 22 x 30 cm (scanning) Direct conversion detectors Hologic/Lorad Photoconductor amorphous selenium Pixel size TFT 70 microns Field of view 24 x 29 cm Optical Density markers dense tissue fatty tissue Relative Exposure air skin margin ,000 Figure 9. Characteristic curve of mammographic film illustrating how the display contrast (slope of the curve) is suboptimal in lucent and dense regions of the breast 1. exhibit uniform quality over a range of exposure conditions and breast sizes. Quantum Detection Efficiency quantum detection efficiency is a measure of the percentage of s that hit the detector that are absorbed. Systems with higher quantum efficiency can produce higher quality images, at lower dose. As mentioned earlier, selenium systems stop more than 95 percent of the s in the mammographic energy range. Standard screens for use in film mammography only have about 50 to 70 percent quantum efficiency, and CsI(Tl) used in indirect conversion detectors exhibit about 50 to 80 percent quantum efficiency. Modulation Transfer Function (MTF) Modulation Transfer Function (MTF) and Detective Quantum Efficiency (DQE) provide quantitative measurement of imaging performance. MTF measures spatial resolution, while DQE is a measure of signal-to-noise ratio, contrast resolution and dose efficiency. An imaging system is best characterized by examining corresponding MTF and DQE curves; however, it cannot be adequately described by one number at a single spatial frequency. These measurements are used to determine how well a system captures information over a range of spatial frequencies. Modulation Transfer Function (MTF) is a measure of signal transfer over a range of spatial frequencies and quantifies spatial resolution. Spatial resolution is often measured using a line pair phantom, illustrated in Figure 10. The ultimate resolution limit of any system is determined by its pixel size. For example, a system with a 100-micron pixel cannot adequately resolve spatial frequencies above 5 line pairs/mm (lp/mm). Indirect-conversion methods can scatter light over several pixels, further limiting the effective resolution of the system, more so than indicated by pixel size alone. Direct conversion systems do not suffer from this limitation.

7 As shown in Figure 11, the MTF of a selenium direct conversion detector 2 is superior to those of screen-film 3 and indirectconversion detectors 4. The intrinsic spatial resolution of selenium direct-conversion mammography detectors is superior to those using indirectconversion scintillators. While MTFs for indirect conversion detectors fall dramatically at higher spatial frequencies, the MTF of direct-conversion selenium detectors remain high over a greater range of spatial frequencies. With selenium, there is no lateral movement of the charge through the photoconductor, and its MTF is independent of thickness. Consequently, selenium detectors are very efficient in capturing and converting s to electrical signals. Detective Quantum Efficiency (DQE) Even with a high MTF at high spatial frequencies, small objects can get lost in the noise of the system. Increasing signal and decreasing noise in the system increases visibility of small structures. Detective Quantum Efficiency (DQE) measures signal-to-noise transfer through the system as a function of spatial frequency, and it is a good measure of dose efficiency. Several factors influence DQE, including the amount of absorption, amplitude or strength of the signal profile (measured by MTF) and noise. Figure 12 shows the DQE for film and digital detectors. While film has a high MTF at high spatial frequencies compared to the indirect conversion detectors, the same is not true for DQE. This is one reason why the potential spatial resolution with film is not realized in practice. Film granularity noise limits its achievable DQE at high spatial frequencies. Indirect conversion systems have a superior DQE to screenfilm, especially at the lower spatial frequencies. However, the DQE drops at high spatial frequencies a consequence of the scintillator-induced light blurring. The direct conversion selenium digital mammography detector outperforms both indirect conversion digital and film in terms of DQE. With no signal spreading, and the DQE (and MTF) are governed mainly by the inherent limits of the pixel size. Furthermore, only selenium detectors can be made sufficiently thick to absorb all the incoming s without sacrificing spatial resolution. Direct-conversion systems offer the possibility of similar image quality at lower doses (by as little as 1/2 to 1/3 of the screen-film dose) or improved image quality at the same dose as screen-film. From a clinical point of view, there is the expectation that at equal doses, direct conversion FFDM systems may provide superior diagnostic imaging in terms of sensitivity and specificity, resulting in better cancer detection and lower false positive rates. Figure 10. resolution test pattern to determine imaging resolution performance, in units of line pairs/mm. Presampled MTF 100% 80% 60% 40% 20% 0% Spatial Frequency (lp/mm) Direct Conversion a-selenium TFT 70 µm pixels Indirect Conversion CsI (T1) TFT 1000 µm pixels CsI (T1) CCD 40 µm pixels Screen/film Gd 2 O 2 S OrthoM / MinR Figure 11. Modulation Transfer Function for screen-film (Kodak: OrthoM/MinR), indirect conversion and direct conversion digital mammography detectors. DQE 70% 60% 50% 40% 30% 20% 10% 0% Frequency (lp/mm) Direct Conversion a-selenium TFT 70 µm pixels Indirect Conversion CsI (T1) TFT 1000 µm pixels CsI (T1) CCD 40 µm pixels Screen/film Gd 2 O 2 S OrthoM / MinR Figure 12. Detective Quantum Efficiency for screen-film (Kodak: OrthoM/MinR), indirectconversion and direct-conversion digital mammography detectors.

8 The visibility of phantom details has been evaluated for screen-film and for first and second-generation detectors. 5 The phantom was imaged at 28 kvp at a dose of ~3 mgy. (A high dose was used to emphasize the performance differences.) Results are summarized in Table 1. These results are consistent with predicted performance based on the DQE plots: secondgeneration systems outperform both screen-film and indirect conversion. Figure 13. ACR Mammography Accreditation Phantom, and imaging details. Absorption Percentage 100% 80% 60% 40% 20% 0% 15% 20% 25% 30% X-Ray Energy (KeV) Csl(T1) Gd2O2S a-selenium Figure 14. Percent absorption of incident s for materials used in screen-film: Gd 2 O 2 S 34 mg/cm 2, indirect conversion CsI(Tl) : 73 mg/cm 2 and second generation cselenium: 250 microns thick. ACR Phantom Imaging Performance Figure 13 shows the acrylic phantom used for ACR accreditation and MQSA inspections. The phantom is approximately equivalent in absorption to a 4.2-cm thick compressed breast consisting of 50 percent glandular and 50 percent adipose tissue. The phantom includes appropriate details that range from visible to invisible on a standard mammographic film image. The phantom has fibers with diameters of 1.56, 1.12, 0.89, 0.75, 0.54 and 0.40 mm; specks with diameters of 0.54, 0.40, 0.32, 0.24 and 0.16 mm; and masses with decreasing diameters and thicknesses of 2.00, 1.00, 0.75, 0.50 and 0.25 mm. System Design Considerations There are a variety of system-level design considerations with full field digital mammography equipment. Source The spectrum of energies used in screen-film mammography has been highly optimized. Digital mammographic detectors offer improved performance characteristics, particularly dynamic range, however the issue of the optimum energy is still in the investigational phase. 6 It is possible that higher energies may permit lower dose or higher image quality with digital mammography, particularly for patients with dense breasts. Detectors that have high intrinsic quantum efficiency allow the use of higher energies. Selenium has superior stopping power compared to film screens and CsI(Tl) material, as seen in Figure 14. At typical energies used in mammography (mean energy ~ 20 KeV), selenium s quantum efficiency is optimal. Selenium maintains its high quantum efficiency at higher energies as well. Systems that reduce breast compression will utilize higher energies than currently used, and the pronounced high-energy performance of selenium will become even more important. Scatter Rejection Methods Large area flat-panel detectors are subject to the same deleterious effects of radiation scatter as conventional screen film systems. The preferred method to reduce scatter is to employ the use of radiation anti-scatter grids interposed between the patient and the detector. Anti-scatter Grids There are two common methods of grid design, shown in Figure 15. Standard linear grids are constructed of long thin slats of radio-opaque materials (called septa or laminae) separated by Detector Technology Fibers visible Specks visible Masses visible Kodak MinR 2000 Screen film CCD Indirect conversion a Selenium Direct conversion Table 1. Performance of systems in imaging the ACR Mammography Accreditation Phantom

9 Typical % Improvement Contrast Improvement Factor AVERAGE (50/50) Glandular (70/30) Fatty (30/70) 4% (4.2 cm) ACR Phantom 4.2 cm 4 cm cm % 6.2 cm) 6 cm % (7.2 cm) HTC GRID LINEAR GRID NONGRID (optimized) 7 cm Compressed Breast Thickness Figure 15. Construction of linear and cellular grids. Figure 16. Contrast improvement for linear and high transmission cellular grids. relatively radiolucent spacer materials. These grids effectively reduce scattered photons, but also block some of the primary beam, mainly from absorption in the spacers. An alternative method of grid construction utilizes a crossed air grid, known as a High-Transmission Cellular (HTC) grid. This grid absorbs scattered radiation in two directions, as opposed to the linear grid, which removes scattered radiation only in one direction. This grid absorbs less of the primary beam, because the intersepta material is air. The HTC grid has been shown to have superior contrast improvement (scatter rejection) and Bucky factors (primary beam reduction) to conventional linear grids. 7 For a 7-cm compressed breast, the HTC grid offers a contrast improvement of about 1.7, compared to a linear grid having a contrast improvement of about 1.5. Automatic Exposure Controls Breasts vary in composition and thickness, and there is a need for Automatic Exposure Control (AEC) to optimize image quality and reduce radiation dose for each patient. Because of film s narrow exposure latitude, proper AEC operation is critical. Digital detectors have a much wider dynamic range and consequently are more tolerant of exposure variations. Therefore, digital mammography should have a reduced number of retakes compared to film-screen. Digital mammography systems also offer the possibility of more advanced AEC methods. It is possible to design these systems to acquire a rapid low-dose pre-image, which is analyzed in real time and then used to set the optimum kvp and mas for the exam. Information on breast density can also be measured and used in the AEC algorithm. Breast Compression The need for compressing the breast during mammography imaging will not be eliminated with the use of digital detectors. The relatively large compression pressure used in screen-film mammography is in part due to the limited dynamic range of film, and the use of low energy s. FFDM systems have dynamic range superior to film. Selenium detectors have high quantum efficiency even at higher energies than used today, with the potential of obtaining high quality images with less compression. Large breast compression is used for other reasons, among them reducing overlap of breast structures. This is an area that requires additional study. Output Devices Film images are read as hard copy, with the images viewed on a light box. Initially, digital mammography images have been viewed the same way. The digital image is captured and stored on a computer, then printed on a high-definition laser printer to be viewed analogously to film. Soft copy review via workstation displays offers the true benefits of digital imaging. Digital transmission, easy retrieval and display of previous images and on-line image processing such as windowing, leveling and feature magnification are all advantages of a digital image. Digital mammography places great constraints on the monitor performance, however. The images are large (~10,000,000 pixels, or more) but resolution, speed and light-output characteristics of display monitors are approaching the requirements needed for digital mammograms. Future Applications Digital mammography offers some exciting future applications that are not practical or possible with screen-film. The clinical utility of these applications is under active investigation. Stereotactic Breast Biopsy A full field digital mammography detector can also be used for diagnostic and stereotactic-breast-biopsy imaging. This enables

10 a multi-purpose system ideal for both screening and follow-up imaging tasks. Stereo Mammography Acquisition of binary images, separated by a few degrees, allows the display of mammograms in a stereo fashion, to facilitate the visualization of the three-dimensional characteristics of mammographic features. Computer display of stereo images requires either glasses synchronized to display frame updates, or polarizing glasses. Investigations of this technique 8 show promise in improving three-dimensional depth discrimination in mammograms. Tomosynthesis and Three-Dimensional Imaging These applications utilize acquisition of multiple images from many angles, typically 9-15 about a degree range, with the data processed to generate tomographic slices. These images will exhibit reduced confusion from overlapping structures. Preliminary studies 9 show that these acquisitions can be performed with doses similar to screening mammography, and that the images offer diagnostic information not available with 2D imaging. High quantum efficiency will be important for these applications, to reduce the total dose for the multiple images required. Dual-Energy Imaging In dual-energy imaging, two images are made of the breast at differing energies. Because of the differing attenuation characteristics of glandular tissue, adipose tissue and microcalcifications, processing of the dual images can enhance the visibility of certain structures. The clinical utility of this has been suggested 10, but further research is needed. Dual energy imaging can also be used to quantify fibroglandular breast tissue density. digital image.with a true digital image, the increased dynamic range and access to multiple views and previous studies is expected to improve the accuracy of CAD systems. Conclusions Full field digital mammography offers the potential for significant advances in breast cancer diagnosis, including lower radiation dose, reduced breast compression pressure and improved cancer detection rates. FFDM has advanced rapidly in the past few years, and there are now several systems commercially available for sale in the U.S. These systems offer improvements over screen-film imaging, especially in regards to imaging dynamic range and digital storage and display. The different manufacturer s products can be differentiated by the image receptor s detection technology, its resolution dose and scatter rejection performance, and in the detector s field of view. Andrew Smith, Ph.D., is principal scientist at Hologic, Inc., in Bedford, Mass., where he is involved in research and development in digital radiographic systems. His previous employment was with Digital Scintigraphics Inc., a nuclear medicine company he co-founded. Prior to that, he attended the Massachusetts Institute of Technology, where he received a bachelor s degree and doctoral degree in physics. Computer-Aided Diagnosis The use of computer programs to perform preliminary cancer diagnoses is an area of active research. 11 The digital image is examined by software, and suspicious areas are highlighted for further scrutiny by the radiologist. The challenge for these systems is to find the proper balance between sensitivity and specificity. Increasing sensitivity can result in too many false positives marked on an image, and the radiologist may find this more of an annoyance than help. Conversely, if not enough true positives are marked, the system will be offering little help to the radiologist. The low cancer rate per image in a screening environment makes this task difficult. However, it is expected that eventually these types of systems will become routine, since it is very easy to perform the CAD procedure on the digitally acquired image. Digital mammography CAD systems offer significant advantages compared to film-based CAD systems. Digitizing of the film images is time consuming and yields a less than optimal

11 Glossary a-se amorphous selenium. The most common photoconductor material used in direct conversion detectors. a-si amorphous silicon. The material used to manufacture TFT arrays used in both indirect and direct FFDM image receptors. CsI(Tl) Cesium-iodide with thallium doping. A common scintillator used in indirect conversion detectors. Direct Conversion.....A method of detecting s utilizing a material that directly absorbs an X- ray and generates an electrical signal. DQE Detective Quantum Efficiency. Measure of the square of the output signal/ noise ratio to the input signal/noise ratio, as a function of spatial frequency. FFDM Full Field Digital Mammography. Digital systems offering field of view large enough to image the entire breast. Gd 2 O 2 S Gadolinium oxysulfide. A common phosphor used in film screens and some indirect conversion detectors. HTC Grid High Transmission Cellular grid. Anti-scatter grid design, using a cellular grid construction with air as the inter-septa material. Indirect Conversion....A two step method of detecting s. First, a scintillator is used to generate light photons upon absorption of s, then a light sensitive element is used to convert the light photons into an electrical signal. MQSA Mammography Quality Standards Act, that defines minimum quality standards for mammography equipment, facilities and operators. MTF Modulation Transfer Function. Measure of the system response, as a function of spatial frequency. Scintillator A material that gives off light photons upon absorption of an. Used in screen-film and indirect conversion digital imaging. References 1 Adapted from Yaffe M, Digital Mammography, RSNA Categorical Course in Breast Imaging 1999; pp Gingold E, Lee D, Proc. SPIE Vol. 3977, p , Medical Imaging 2000: Physics of Medical Imaging, James T. Dobbins; John M. Boone; Eds. 3 Bunch P, Proc. SPIE Vol. 2708, p , Medical Imaging 1996: Physics of Medical Imaging, Richard L. Van Metter; Jacob Beutel; Eds. 4 GE Senographe 2000D FDA PMA submission P (1999); Vedantham S, Karellas A, et al., Med Phys Mar;27(3):558-67; Cheung L, Bird R, et al, p in Digital MammographyNijmegen, Kluwer Academic Publishers, Private communication, Zhenxue Jing, Lorad Corp. CCD data from the Lorad LDBI system. 6 Fahrig R. Rowlands J. Yaffe M, imaging with amorphous selenium: optimal spectra for digital mammography, Med Phys April 23(4): de Almeida A,Rezentes P, Barnes G, Mammography grid performance. Radiology 1999; 210: Chan H, Goodsitt M. in Digital Mammography, Medical Physics Publishing, 2000 (in press). 9 Niklason L, Christian B, et al, p in Digital Mammo raphy Nijmegen, Kluwer Academic Publishers, See pg. 390 in Kopans D, Breast Imaging 2nd ed., Lippincott- Raven, See many papers in Digital Mammography, Medical Physics Publishing, 2000 (in press), and also Digital Mammography Nijmegen, Kluwer Academic Publishers, Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations

12 Hologic, Inc. 35 Crosby Drive Bedford, MA U.S.A. T: R-BI-016 March 05

Digital Mammography Update: Design and Characteristics of Current Systems

Digital Mammography Update: Design and Characteristics of Current Systems : Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM [email protected] 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,

More information

Mammography. What is Mammography?

Mammography. What is Mammography? Scan for mobile link. Mammography Mammography is a specific type of breast imaging that uses low-dose x-rays to detect cancer early before women experience symptoms when it is most treatable. Tell your

More information

X-RAY IMAGING Emerging Digital Technology - CMOS Detectors

X-RAY IMAGING Emerging Digital Technology - CMOS Detectors Application Note Case Study Technology Primer White Paper X-RAY IMAGING Emerging Digital Technology - CMOS Detectors Image Sensors X-Ray DETECTORS Scanners Image Processing Custom Solutions In all domains

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

Digital radiography conquers the veterinary world

Digital radiography conquers the veterinary world Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing

More information

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality

More information

Digital Radiography Image Quality: Image Acquisition

Digital Radiography Image Quality: Image Acquisition Digital Radiography Image Quality: Image Acquisition Mark B. Williams, PhD a, Elizabeth A. Krupinski, PhD b, Keith J. Strauss, MS c, William K. Breeden, III, MS d, Mark S. Rzeszotarski, PhD e, Kimberly

More information

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams EPEX Direct-to-Digital System for General Radiography Featuring... Versatility for a complete range of general radiographic exams The Hologic EPEX General Radiography System is a direct-to-digital, fully

More information

Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1

Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1 Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1 Comparison of Amorphous Selenium and Amorphous Silicon Flat-Panel Detectors November 20, 2012 COMPARISON OF FLAT-PANEL DETECTORS 2 Abstract

More information

Overview of Digital Detector Technology

Overview of Digital Detector Technology Overview of Digital Detector Technology J. Anthony Seibert, Ph.D. Department of Radiology University of California, Davis Disclosure Member (uncompensated) Barco-Voxar Medical Advisory Board ALARA (CR

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Page: 1 of 6 Page: 1 of 6

Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,

More information

Chapter 15. Creating the Digital Image. Objectives. Key Terms

Chapter 15. Creating the Digital Image. Objectives. Key Terms Chapter 15 Creating the Digital Image Objectives Describe how a digital image is created and processed. Sequence the steps of image production and evaluation of the digital image. Identify the components

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015) The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015) Table of Contents Stereotactic Breast Biopsy Accreditation Program

More information

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Radiographic Image Production X-ray photons emitted from the x-ray tube interact with the body, exit the patient (exit beam) and interact

More information

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D. Slide 1 Digital Mammography Detector QC Slide 2 Introduction What is FFDM QC and why is it important? Eric A. Berns, Ph.D. What to know before you start [email protected] Overview and compare QC tests

More information

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Development of High-sensitivity and High-resolution Digital Mammography System AMULET

Development of High-sensitivity and High-resolution Digital Mammography System AMULET UDC 616-073.75+618.19-006+771.537 Development of High-sensitivity and High-resolution Digital Mammography System AMULET Kaku IRISAWA*, Takahisa ARAI*, Hajime NAKATA*, Tomonari SENDAI*, Takao KUWABARA*,

More information

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It is intended to replace film/screen

More information

Update on MQSA and ACR Mammography Accreditation

Update on MQSA and ACR Mammography Accreditation MQSA - Who s Who Update on MQSA and ACR Mammography Accreditation The Law: Mammography Quality Standards Act (MQSA) The Regulator: US Food and Drug Administration (FDA) Pamela L. Platt, BSRT(R)(M)(CV)

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes. IAC Ch 41, p.1 641 41.7 (136C) X-ray machines used for stereotactically guided breast biopsy. 41.7(1) Definitions. In addition to the definitions provided in rules 641 38.2(136C), 641 40.2(136C), and 641

More information

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection Digital Mammography Upright Stereo Mammography For 35+ years, screenfilm has been the gold standard for breast cancer detection IN THE BEGINNING Mammography technology has come a long way since the first

More information

Update on ACR Digital Mammography QC Manual

Update on ACR Digital Mammography QC Manual Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual

More information

Concepts for High-Resolution Low-Dose CT of the Breast

Concepts for High-Resolution Low-Dose CT of the Breast RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

Hologic Selenia Dimensions C-View Software Module. October 24, 2012

Hologic Selenia Dimensions C-View Software Module. October 24, 2012 Hologic Selenia Dimensions C-View Software Module October 24, 2012 Introduction and Agenda Peter Soltani, Ph.D. Senior VP & GM, Breast Health Hologic, Inc. Agenda Technology Overview Clinical Overview

More information

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY

ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College

More information

Digital Breast Tomosynthesis QC Requirements

Digital Breast Tomosynthesis QC Requirements Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist

More information

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

Medicare Part B. Mammograms - Updated Billing Guide for Screening and Diagnostic Tests

Medicare Part B. Mammograms - Updated Billing Guide for Screening and Diagnostic Tests Mammograms - Updated Billing Guide for Screening and Diagnostic Tests This article from Medicare B News Issue 223 dated October 21, 2005 is being updated and reprinted to ensure that the Noridian Administrative

More information

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis

Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis David Bernard & Steve Ainsworth Dage Precision Industries Abstract Non-destructive testing during the manufacture of

More information

X-ray (Radiography) - Chest

X-ray (Radiography) - Chest Scan for mobile link. X-ray (Radiography) - Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray produces images of

More information

Patient Prep Information

Patient Prep Information Stereotactic Breast Biopsy Patient Prep Information Imaging Services Cannon Memorial Hospital Watauga Medical Center Table Weight Limits for each facility Cannon Memorial Hospital Watauga Medical Center

More information

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Radiography: 2D and 3D Metrics of Performance Towards Quality Index AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements..

More information

Remote RadEye Product Family

Remote RadEye Product Family Remote RadEye Product Family Overview The Remote RadEye x-ray camera provides the ultimate flexibility in design and product options for your complex imaging applications. Our unique camera design separates

More information

Coronis 5MP Mammo. The standard of care for digital mammography

Coronis 5MP Mammo. The standard of care for digital mammography Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment

More information

Medical imaging monitors specification guidelines

Medical imaging monitors specification guidelines Medical imaging monitors specification guidelines Document details Contact for enquiries and proposed changes If you have any questions regarding this document or if you have a suggestion for improvements,

More information

Indirect X-ray photon counting image sensor with 27T pixels and 15 electrons RMS accurate threshold

Indirect X-ray photon counting image sensor with 27T pixels and 15 electrons RMS accurate threshold Indirect X-ray photon counting image sensor with 27T pixels and 15 electrons RMS accurate threshold B. Dierickx 1,2, B. Dupont 1,3, A. Defernez 1, N. Ahmed 1 1 Caeleste, Antwerp, Belgium 2 Vrije Universiteit

More information

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

How To Compare Mammography To A 3D System

How To Compare Mammography To A 3D System Radiology Advisory Panel Meeting Hologic Selenia Dimensions 3D System with C-View Software Module FDA Review Robert Ochs, PhD Branch Chief Mammography, Ultrasound, and Imaging Software Branch Division

More information

Modern Fluoroscopy Imaging Systems

Modern Fluoroscopy Imaging Systems Modern Fluoroscopy Imaging Systems Eric Gingold, PhD Thomas Jefferson University, Philadelphia, PA SUMMARY Fluoroscopy, or real-time projection X-ray imaging, has been in clinical use since shortly after

More information

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY [email protected]

More information

Performance testing for Precision 500D Classical R/F System

Performance testing for Precision 500D Classical R/F System Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)

More information

Best Practices in Digital Radiography

Best Practices in Digital Radiography WHITE PAPER Best Practices in Digital Radiography Tracy L. Herrmann, M.Ed., R.T.(R); Terri L. Fauber, Ed.D., R.T.(R)(M); Julie Gill, Ph.D., R.T.(R)(QM); Colleen Hoffman, R.T.(R)(M)(CT); Denise K. Orth,

More information

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.

More information

Elements of required physical infrastructures: space, schielding, and patient flow..

Elements of required physical infrastructures: space, schielding, and patient flow.. Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto

More information

Dental Radiography Core Subject. Digital Radiography

Dental Radiography Core Subject. Digital Radiography Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of

More information

2 Absorbing Solar Energy

2 Absorbing Solar Energy 2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could

More information

Encoders for Linear Motors in the Electronics Industry

Encoders for Linear Motors in the Electronics Industry Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy

More information

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture

More information

Physics 441/2: Transmission Electron Microscope

Physics 441/2: Transmission Electron Microscope Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This

More information

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes. Radiologic Technologist Overview The Field - Specialty Areas - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Radiologic technologists

More information

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: [email protected] Cell:+250 (0)7888 44764

Equipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Digital Medical Radiography Diagnostic (DR) system X-ray Equipment Presented by: CocticaUwititonze Email: [email protected] Cell:+250 (0)7888 44764 Lin muyan 1. Goals 2. Introduction 3. X-ray machine

More information

Analog and Digital Systems of Imaging in Roentgenodiagnostics

Analog and Digital Systems of Imaging in Roentgenodiagnostics 8 Analog and Digital Systems of Imaging in Roentgenodiagnostics Dominika Oborska-Kumaszyńska 1 General Radiology, Interventional Radiology and Neuroradiology, University Hospital, Wroclaw, 2 Wolverhampton

More information

A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose

A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose AAPM 2005-Continuing Education Course in Radiographic and Fluoroscopy Physics and Technology A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose Ralph Schaetzing,

More information

Fundamentals of Cone-Beam CT Imaging

Fundamentals of Cone-Beam CT Imaging Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation

More information

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY

BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY N. GEERAERT a,b,c, R. KLAUSZ a, S. MULLER a, I. BLOCH c, H. BOSMANS b a GE Healthcare, Buc, France b Department

More information

Computed Radiography: Acceptance Testing and Quality Control

Computed Radiography: Acceptance Testing and Quality Control Computed Radiography: Acceptance Testing and Quality Control J. Anthony Seibert, Ph.D. University of California Davis Department of Radiology Sacramento, California Introduction CR is the primary means

More information

X-ray (Radiography), Chest

X-ray (Radiography), Chest X-ray (Radiography), Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray makes images of the heart, lungs, airways,

More information

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative

More information

A PACS-Aware DICOM Image Object

A PACS-Aware DICOM Image Object Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry

More information

Q: What differentiates a diagnostic from a screening mammography procedure?

Q: What differentiates a diagnostic from a screening mammography procedure? The following Q&As address Medicare guidelines on the reporting of breast imaging procedures. Private payer guidelines may vary from Medicare guidelines and from payer to payer; therefore, please be sure

More information

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control

More information

A Breakthrough in Breast Specimen Imaging

A Breakthrough in Breast Specimen Imaging BreAST HEALTH SOLUTIONS A Breakthrough in Breast Specimen Imaging Trident Specimen Radiography System Superb Image Quality for Streamlined Workflow The Trident system s sharp, high quality images allow

More information

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Annual MQSA Inspection Questions NMQAAC April 19, 2004 1 1 1 1 1 1 1 1 0 1 0 1 0 1 0 1 Annual MQSA Inspection Questions NMQAAC April 1, 00 1.0 Inspection Information 1.1 Name and Address 1. Equipment Registration.0 Facility Inspection Download.0 Facility Inspections.0

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

Thin Is In, But Not Too Thin!

Thin Is In, But Not Too Thin! Thin Is In, But Not Too Thin! K.V. Ravi Crystal Solar, Inc. Abstract The trade-off between thick (~170 microns) silicon-based PV and thin (a few microns) film non-silicon and amorphous silicon PV is addressed

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices

Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices IT Guidance Documents Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists

More information

Automotive Applications of 3D Laser Scanning Introduction

Automotive Applications of 3D Laser Scanning Introduction Automotive Applications of 3D Laser Scanning Kyle Johnston, Ph.D., Metron Systems, Inc. 34935 SE Douglas Street, Suite 110, Snoqualmie, WA 98065 425-396-5577, www.metronsys.com 2002 Metron Systems, Inc

More information

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Standard Test Method for Classification of Film Systems for Industrial Radiography 1 Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately

More information

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization

More information

Patient Exposure Doses During Diagnostic Radiography

Patient Exposure Doses During Diagnostic Radiography Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University

More information

X- and Gamma Ray Imaging Systems based on CdTe-CMOS Detector Technology

X- and Gamma Ray Imaging Systems based on CdTe-CMOS Detector Technology X- and Gamma Ray Imaging Systems based on dte-mos Detector Technology Konstantinos Spartiotis, Ray Durrant, nssi Leppänen, Henrik Lohman, Olli Männistö, Tuomas Pantsar, Jouni Pyyhtiä, Tom Schulman, Oy

More information

Personalized Breast Screening Service

Personalized Breast Screening Service Frequently Asked Questions WHAT IS BREAST DENSITY? Breasts are made up of a mixture of fibrous, glandular and fatty tissue. Your breasts are considered if you have predominantly fibrous or glandular tissue

More information

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction

More information

Fiber Optics: Engineering from Global to Nanometer Dimensions

Fiber Optics: Engineering from Global to Nanometer Dimensions Fiber Optics: Engineering from Global to Nanometer Dimensions Prof. Craig Armiento Fall 2003 1 Optical Fiber Communications What is it? Transmission of information using light over an optical fiber Why

More information

Aspire FCRm/CRm Fujifilm CR for Mammography

Aspire FCRm/CRm Fujifilm CR for Mammography : May 2013 Fujifilm Medical Systems U.S.A., Inc. 2 CONTENTS FOREWORD... 5 Introduction to Measurements... 6 S value... 7 Overview of Testing... 8 Section A Getting Started... 9 0 Set-up & Baseline Images...10

More information

Using Automatic Exposure Control in Digital Radiography

Using Automatic Exposure Control in Digital Radiography Using Automatic Exposure Control in Digital Radiography A. Kyle Jones, Ph.D. U.T. M. D. Anderson Cancer Center Department of Imaging Physics 2008 AAPM Meeting, Houston, TX Automatic Exposure Control The

More information

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules Abstract J.L. Crozier, E.E. van Dyk, F.J. Vorster Nelson Mandela Metropolitan University Electroluminescence (EL) is a useful

More information

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs TracePro Opto-Mechanical Design Software s Fluorescence Property Utility TracePro s Fluorescence Property

More information

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Janet E. Semmens Sonoscan, Inc. 2149 E. Pratt Boulevard Elk Grove Village, IL 60007 USA Phone: (847)

More information

Silicon-On-Glass MEMS. Design. Handbook

Silicon-On-Glass MEMS. Design. Handbook Silicon-On-Glass MEMS Design Handbook A Process Module for a Multi-User Service Program A Michigan Nanofabrication Facility process at the University of Michigan March 2007 TABLE OF CONTENTS Chapter 1...

More information