Risk Analysis Fanless Slim Medical Panel PC, models xxxx-onyx-2121dtyxxxxxxx



Similar documents
ISO 14971: Overview of the standard

Risk Assessment for Medical Devices. Linda Braddon, Ph.D. Bring your medical device to market faster 1

Guidance for upgrading to GB (IEC : 2005, MOD)

3.1.1 Full Type Tests & Routine Tests according to Clause 8 2 & Instructions For Installation, Operation & Maintenance

ST Series POWER SUPPLIES USER INSTRUCTIONS

12 SOLAR PHOTOVOLTAIC POWER SUPPLY SYSTEMS by John Ware. PV modules are current-limiting

RISH CON - Hz. Salient Features : Application : Product Features: FREQUENCY TRANSDUCER

Chrome DIN Rail Power Supply 12V 54W 1 Phase (Class II & NEC Class 2) / DRC-12V60W1AG

Current valve. for AC 24 V pulse/pause control of electrical loads up to 30 kw

Advantium 2 Plus Alarm

Installation and safety instructions for AC/DC built-in devices

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

WINGLIKE ELECTRONICS CO.,LTD

EMOSAFE EN-100. Network Isolators for PCB assembly 1 GENERAL DESCRIPTION. Product Data Sheet ENGLISH

NX safety stand-alone

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF

R4D0-FB-IA* FieldBarrier. Features. Assembly. Function. Connection IN + - S OUT Spur 2. Spur 1. Zone 1. Zone 0

PD30ETB20xxIS. Photoelectrics, Background Suppression reflective with IR light. Main features. Description

Model 1756 Test Lead Kit

IEC >>> IEC GUIDE GLOBAL SPECIALIST IN ELECTRICAL AND DIGITAL BUILDING INFRASTRUCTURES

Guide to the electrical parameter classifications of IEC and IEC safety standards

EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM

Handheld portable appliance testers

Calculating Creepage and Clearance Early Avoids Design Problems Later Homi Ahmadi

aseries A13B Mini Bullet Camera User Manual

SECTION DETECTION AND ALARM

Meeting your Electrical Safety Testing & Quality Requirements

Manual Ranging MultiMeter

ELECTRICAL SAFETY RISK ASSESSMENT

PRELIMINARY MANUAL STAGE COLOR 48. Preliminary User Guide. Official version will follow soon on: TABLE OF CONTENTS

Type: EASY719 DC RC Article No.: Ordering information Relay outputs Quantity 6 Power supply V DC 24 V DC. Description

Redundancy Module QUINT-DIODE/40

RM17TE V AC. Main. Product or component type. Product specific application. Relay monitored parameters 250 V DC 5 A DC

Alarm over IP. IRIS Touch Home Installation Manual. Version 1.0 ENGLISH. Now certified and compliant with EN50131, EN50136 Security Grade 4 ATS6

Square D Clipsal DIN-Rail Four-Channel Auxiliary Input Unit

DEVICES OF PROTECTION AND COMMAND

H/L = Power supply. FN = Nominal Frequency. RN = Rated value of output burden. UN = Nominal input voltage. IN = Nominal input current.

Instruction book IQAN-MDM. Publ. no. HY /UK Edition June, 2007

UL TEST REPORT AND PROCEDURE

When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description

Drayton Digistat +2RF/+3RF

Verizon NEBS TM Compliance: Data Center Equipment NEBS Requirements Verizon Technical Purchasing Requirements VZ.TPR.9703 Issue 2, March 2012

Power supply unit Model: PAC-SC50KUA

Chrome DIN Rail Power Supply 24V 30W 1 Phase (Class II & NEC Class 2) / DRC-24V30W1AZ

Battery Charger For Nickel Cadmium and Nickel-Metal Hydride Rechargeable Batteries Model PSN Series

MOUNTING AND OPERATING INSTRUCTIONS

User's Guide. True RMS Industrial Multimeter

323/324/325. Calibration Manual. Clamp Meter

DVI DA2 and DVI DA 4 User Guide

Portable Air Conditioner

Electronic timer CT-AHD.12 OFF-delayed with 1 c/o (SPDT) contact

Current Monitoring Kit

Operating instructions

US Installation Manual Smappee Monitor

Senses SV series industrial monitor user manual

Cross-beam scanning system to detect slim objects. 100 mm in

Server Rack Commissioning

5SV Residual Current Protective Devices

Power Supplies 3. SDN-P DIN Rail Series

USER S MANUAL. MaxPower UPS. Uninterruptible Power System 28-2MAXPO0018

ETC TWO STAGE ELECTRONIC TEMPERATURE CONTROL

MCR1900 Media Converter 19-Slot Chassis

Analog signal converters CC-E I/I Current / current isolators

R.C.C.B. s two-pole LEXIC

UT202A Operating Manual. Contents

Insulation resistance

NOTE: Additional user information for your printer may be available in the "Manuals" section at support.dell.com.

user manual smappee monitor

TMS TANK MANAGEMENT SYSTEM

Application Switchgear for industrial and commercial applications. U e V AC 230/ (per pole)

Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011

5-port / 8-port 10/100BaseTX Industrial Ethernet Switch User Manual

SERVICE MANUAL 12VDC WALL THERMOSTAT AIR CONDITIONING SYSTEMS ROOFTOP UNITS ONLY

EPM3. Phase Sequence and Motor Rotation Tester. Users Manual

Microwave Meter. Instruction Manual

Infrarot-Bewegungsmelder IP44

Online data sheet PET-1CB010N10SVMC PET PRESSURE SENSORS

ABB Stotz-Kontakt GmbH ABB safe&smart Project planning of security systems

SHARP SOLAR Frequently Asked Questions for PV Integrators Revised 05/04/2010

BLACK BOX. T1/E1 Link Extenders OCTOBER 2007 MT195A-T1 MT196A-E1

Emergency lighting units EM powerled

FAQs-Main switchboard design criteria

CPCI-3U-AC-300W-R. Rugged 300W AC Input 3U CompactPCI Power Supply. Electrical Specification for: Telkoor Part Number:

AIR CONDITIONING SYSTEMS

RDJ10RF/SET. Wireless room temperature controller with 24-hour time switch and LCD. Programmable, for heating systems

ADM1TE 5/30A DIN rail single phase two wire energy meter

TM7BDM8B expansion block - TM7 - IP67-8 DI/DO - 24V DC A - M8 connector

What are the basic electrical safety issues and remedies in solar photovoltaic installations?

M7250P. PoE Powered. Gigabit Ethernet Media Converter 1000BASE-TX TO 1000BASE-SX/LX. Installation Guide

Technicaldata General

Molded Case Circuit Breaker Specifications Bulletin Number 140U-D

UM-X Field display for continous level sensors

MEDIUM VOLTAGE CE-BF SWITCHBOARDS. UP TO 40.5 kv. CE - BF - C - en - REV

FLUORESCENT UV- RING LIGHT OPERATING INSTRUCTION

Mercury Helios 2 ASSEMBLY MANUAL & USER GUIDE

Power Supply Unit, Primary Switched, Narrow Design MINI-PS-12-24DC/24DC/1

DRM75A 230V 20/100A DIN rail single phase two wire energy meter

PowerFlex 700S and 700H Frame 12 DC Bus Connector Kit

Climasys Thermal Management System. Schneider Electric Universal Enclosures 2009

Transcription:

Risk Analysis Fanless Slim Medical Panel PC, models xxxx-onyx-2121dtyxxxxxxx Company: Creation Date: 2013-08-22 Abbreviation: Project Number: Project : Version Number: Version : Life Cycle Phase: ONYX-2121DT 1307033002 The models xxxx-onyx-2121dty-xxxxxxx are Fanless Slim Medical Panel PC for medical equipment used. 1.00 Workable version for evaluation purpose Functional design

Table of Content The Following Documents have been Compiled as a Report: Page Project Data 3 Version Data 4 Project Team 7 Definition of the Risk Graph 13 Release Criteria 15 Grey Box 17 Risk Analysis 18 Risk Evaluation Before Measures 26 Risk Graph Before Measures 33 Decision about Measures 34 Risk Control 48 Release Protocol of Measures 54 Evaluation After Measures 120 Risk Graph After Measures 128 Risk Graph Before and After Measures 129 Evaluation of Residual Risk 131 Post production 132 Overview of the Results of the Risk Management 133 2013-08-22 Page 2 of 141

Project Data Project: 1307033002, ONYX-2121DT Version: 1.00 Project Data Abbreviation: ONYX-2121DT Project Number: 1307033002 Project Name: Fanless Slim Medical Panel PC, models xxxx- ONYX-2121DTy-xxxxxxx Project : The models xxxx-onyx-2121dty-xxxxxxx are Fanless Slim Medical Panel PC for medical equipment used. Product Type: EN ISO 14971:2012 Project Type: Product Project Management Project Leader: Substitute: Responsible for the Risk: Mac Lin James Chen Mac Lin 2013-08-22 Page 3 of 141

Version Data Project: 1307033002, ONYX-2121DT Version: 1.00 Version Data Version No.: Derived from Version: Special Version: 1.00 - Version : Workable version for evaluation purpose Freeze Analysis and Evaluation Part: Intended Use: 1. The xxxx-onyx-2121dty-xxxxxxx were Fanless Slim Medical Panel PC with Intel Atom D525 1.8 GHz Dual Core processor-based computer that is designed to serve as a point of care station. It is a PC-based system with 22" true color TFT LCD display, Zero Noise solution; integrated multimedia functions make them the perfect platforms to build comprehensive lifestyle computing applications. There is a VESA compliant for mounting the support arm. 2. The rated input: 24 V dc, 2.91 A (for LCD Monitor); 100-240 V ac, 50/60 Hz, 1.8-0.8 A (for power supply, model GMPU70A-6) 3. The operation temperature/humidity: 10-40 C, 30%-75 % RH and suitable storage from -20 C to 60 C, 10% - 90% RH. 4. This Slim Medical Panel PC intended to be used at any altitudes between 0 to 3,000 m. 5. The Slim Medical Panel PC expected services life was 50,000 hours 6. This Slim Medical Panel PC was judged on the basis of the required creepage and clearances in the Third Edition of the Standard for Medical Electrical Equipment, IEC 60601-1 3rd Ed., Sub. clause 8.9 which covers the end-use product for which the component was designed. 7. This Slim Medical Panel PC was tested on a 20 A branch circuit. If used on a branch circuit greater than this, additional evaluation was necessary. 8. This Slim Medical Panel PC has not been evaluated for patient connected applications. 9. This Slim Medical Panel PC has been evaluated as a Class I, continuous operation, ordinary equipment and has not been evaluated for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide. An additional evaluation shall be made if the power supply is intended for use in other than Class I equipment. 10. This Slim Medical Panel PC shall be installed assembled maintained or repair. 11. This Slim Medical Panel PC incorporates the IEC 60601-1 certified power adapter, model GMPU70A-6 by Eljintek. This power is used in the conditions of overvoltage category II, Pollution 2013-08-22 Page 4 of 141

Possible Misuse: Life Cycle Phase: Degree II. 12. This Slim Medical Panel PC does not deliverer the energy to or extract the energy from the patient. 13. This Slim Medical Panel PC does not deliver the substances to or extract the substances from the patient. 14. No any biological materials processed by the Slim Medical Panel PC for subsequent re -use. 15. This Slim Medical Panel PC does not supply sterile or intended to be sterilized by the user, or other applicable microbiological controls. 16. This Slim Medical Panel PC is not intended to be routinely cleaned and disinfected by the user. 17. This Slim Medical Panel PC is not intended to modify the patient environment. 18. This Slim Medical Panel PC does not take any measurements. 19. This Slim Medical Panel PC is not intended for use in conjunction with medicines or other medical technologies. 20. This Slim Medical Panel PC has no unwanted outputs of energy. 21. This Slim Medical Panel PC is susceptible to environmental influences (ambient, emc). 22. This Slim Medical Panel PC influences the environment (emi). 23. This Slim Medical Panel PC does not provide the essential consumables or accessories. 24. This Slim Medical Panel PC is no need to maintenance and/or calibration. 25. This Slim Medical Panel PC contains software, but the failure of the software does not lead to an unacceptable RISK. 26. This Slim Medical Panel PC has a restricted shelf-life. 27. This Slim Medical Panel PC has no any delayed and/or long-term use effects. 28. This Slim Medical Panel PC does not subject mechanical forces. 29. This Slim Medical Panel PC is not intended for single use. 30. No need to safe decommissioning or disposal of the Slim Medical Panel PC. 31. Installation or use of the Slim Medical Panel PC is required special training. 32. This Slim Medical Panel PC has output wire to connection end-product. 33. This Slim Medical Panel PC is not used by person with special needs. 34. This Slim Medical Panel PC did not use an alarm system. 35. This Slim Medical Panel PC did not hold data critical to patient care. 36. The least favourable working conditions are: maximum normal load and placed in a test corner. 37. The metal enclosure of Slim Medical Panel PC can be touched within 1 sec and plastic enclosure can be touched within 60 sec. Power adapter can be touched within 10 sec. Diagnose use in computed tomography (CT), magnetic resonance imaging (MRI), endoscopy, ultrasound physiotherapy equipment and digital mammography system. Functional design Version Data 2013-08-22 Page 5 of 141

Additional Documents Titel No. Location References IEC Standard IEC Standard IEC Standard ISO Standard Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests Analysis techniques for system reliability - Procedure for failure mode and effects analysis (FMEA) Medical devices - Application of risk management to medical devices IEC 60601-1: 2005 IEC 60601-1-2: 2007 (Modified) IEC 60812: 2006 ISO 14971: 2007 Version Data 2013-08-22 Page 6 of 141

Project Team Project: 1307033002, ONYX-2121DT Version: 1.00 Project Team Project Management Project Leader: Substitute Project Leader: Risk Responsible: Mac Lin James Chen Mac Lin Team Members: see next page 2013-08-22 Page 7 of 141

Chen, James User Data: Project Team Username: James_Chen_Onyx Location: Department: Position: - - QC E-Mail: - Phone: - Fax: - Qualification: Task/Status : Miscellaneous: O025 Risk Manager - Provide guidance for applying the risk management process - Approve Risk Management File and risk review records Rights: Manage Version Data Manage Project Data Define Graph Implement Analysis Implement Evaluation Release Measures Decide about Implementation of Measures Validate Measures Close Versions Manage Project Team Manage Measure Documentation Manage Version Documentation Implement Reevaluation Evaluate Residual Risk Print Reports Manage Post Production Information Manage Measure Details Date, Signature Team Member 2013-08-22 Page 8 of 141

Lin, Eason User Data: Project Team Username: Eason_Lin_Onyx Location: Department: Position: - - Sales E-Mail: - Phone: - Fax: - Qualification: Task/Status : Comment: O056 Product Surveillance Leader (PSL) - Perform as Risk Management Project Team leader for complaint risk assessments - Ensure complaint risk assessment follow this work instruction - Provide input to the determination of probability of occurrence of the hazardous situation(s) for complaint risk assessments Rights: Manage Version Data Manage Project Data Define Graph Implement Analysis Implement Evaluation Release Measures Decide about Implementation of Measures Validate Measures Close Versions Manage Project Team Manage Measure Documentation Manage Version Documentation Implement Reevaluation Evaluate Residual Risk Print Reports Manage Post Production Information Manage Measure Details Date, Signature Team Member 2013-08-22 Page 9 of 141

Lin, Mac User Data: Project Team Username: Mac_Lin_Onyx Location: Department: Position: - - Project Manager E-Mail: - Phone: - Fax: - Qualification: TUV Rh training - General safety requirements for medical appliances acc. IEC/EN 60601-1 UL training - Advanced Workshop on International Regulations of Medical Device UL training - Introduction of ISO14971 and IEC60601-1 Task/Status : Miscellaneous: Rights: Project Manager - Provide guidance regarding hazardous situation identification - Provide the severity and likelihood of harm values and rationales for the hazardous situation - Approve all changes to Risk Management File Manage Version Data Manage Project Data Define Graph Implement Analysis Implement Evaluation Release Measures Decide about Implementation of Measures Validate Measures Close Versions Manage Project Team Manage Measure Documentation Manage Version Documentation Implement Reevaluation Evaluate Residual Risk Print Reports Manage Post Production Information Manage Measure Details Date, Signature Team Member 2013-08-22 Page 10 of 141

Yuan, Andy User Data: Project Team Username: Andy_Yuan_Onyx Location: Department: Position: - - EE E-Mail: - Phone: - Fax: - Qualification: TUV Rh training - General safety requirements for medical appliances acc. IEC/EN 60601-1 UL training - Advanced Workshop on International Regulations of Medical Device UL training - Introduction of ISO14971 and IEC60601-1 Task/Status : Miscellaneous: Rights: MDIC training - Risk Management for Medical Devices Engineering Representative - Identify hazard(s) and hazardous situations - Identify the root cause of the hazard(s) and hazardous situation(s) - Identify mitigations - Determine the probability of occurrence of the hazardous situation(s) and/or root cause as part of the RMPT - Provide standards applicable to manage a hazard or a hazardous situation - Lead the risk/benefit analysis when applicable - Update risk management documentation based on post-production risk activities - Ensure risk assessments for non-conformances following this work instruction - Develop and update the Risk Management File (RMF) for legacy products without an RMF Manage Version Data Manage Project Data Define Graph Implement Analysis Implement Evaluation Release Measures Decide about Implementation of Measures Validate Measures Close Versions Manage Project Team Manage Measure Documentation Manage Version Documentation Implement Reevaluation Evaluate Residual Risk Print Reports Manage Post Production Information Manage Measure Details Date, Signature Team Member 2013-08-22 Page 11 of 141

Chen, Ernest User Data: Project Team Username: Ernest_Chen_Onyx Location: Department: Position: - - ME E-Mail: - Phone: - Fax: - Qualification: TUV Rh training - General safety requirements for medical appliances acc. IEC/EN 60601-1 UL training - Advanced Workshop on International Regulations of Medical Device UL training - Introduction of ISO14971 and IEC60601-1 Task/Status : Miscellaneous: Rights: MDIC training - Risk Management for Medical Devices Engineering Representative - Identify hazard(s) and hazardous situations - Identify the root cause of the hazard(s) and hazardous situation(s) - Identify mitigations - Determine the probability of occurrence of the hazardous situation(s) and/or root cause as part of the RMPT - Provide standards applicable to manage a hazard or a hazardous situation - Lead the risk/benefit analysis when applicable - Update risk management documentation based on post-production risk activities - Ensure risk assessments for non-conformances following this work instruction - Develop and update the Risk Management File (RMF) for legacy products without an RMF Manage Version Data Manage Project Data Define Graph Implement Analysis Implement Evaluation Release Measures Decide about Implementation of Measures Validate Measures Close Versions Manage Project Team Manage Measure Documentation Manage Version Documentation Implement Reevaluation Evaluate Residual Risk Print Reports Manage Post Production Information Manage Measure Details Date, Signature Team Member 2013-08-22 Page 12 of 141

Definition of the Risk Graph Project: 1307033002, ONYX-2121DT Version: 1.00 Defined Graph Definition of the Risk Graph Acceptable ALARP Inacceptable Severity Abbreviation Criterion Example S1 Negligible Inconvenience or temporary discomfort Slight rube faction, pressure sores, hot fingers, short-time failure of single functions which can, however, be corrected by the user without expenditure S2 Minor Results in temporary injury or impairment not requiring professional medical intervention S3 Serious Results in injury requiring professional medical intervention S4 Critical Results in permanent impairment or life-threatening injury Slight cuts, minor scalds, unintentional switch-off, reversible malfunction in normal operation Cuts, moderate scalds Scalds, cuts, electric shocks, longerterm and lasting failure of functions 2013-08-22 Page 13 of 141

Abbreviation Criterion Example S5 Catastrophic Results in patient death. Severe scalds, heavy electric shocks, death, complete destruction of the device, severe damage to the environment (damage caused by fire or by water) Probability Abbreviation Criterion Example P1 Improbable < 1/1,000,000 Event is not reproducible or cannot be replicated under the specified operating conditions of the device. P2 Remote <1/100,000 and >= 1/1,000,000 Event occurs infrequently during the life of the product under the specified operating conditions of the device. P3 Occasional <1/10,000 and >= 1/100,000 Event occurs occasionally during the life of the product under the specified operating conditions of the device. P4 Probable < 1/1000 and >= 1/10,000 Event occurs proable during the life of the product under the specified operating conditions of the device. P5 Frequent >= 1/1,000 Event can be replicated repeatedly and/or is likely to occur regularly or many times during the life of the product under the specified operating conditions of the device. Definition of the Risk Graph Defined Risk Graph s Acceptable ALARP Inacceptable Acceptable As low as reasonably possible Not acceptable 2013-08-22 Page 14 of 141

Release Criteria Project: 1307033002, ONYX-2121DT Version: 1.00 Release Criteria Required Documentation Complete risk management documentation including the complete release report of single action. QD2-010 Risk Management Procedure s When is a Risk Acceptable? The criteria for risk acceptability were based upon applicable national or regional regulations and relevant International Standards, and takes into account available information such as the generally accepted state of the art and known stakeholder concerns. Risk = Severity x Probability Results: 1... 6 (exclude 5): Acceptable; 5, 8... 25: Not acceptable - Definition Risk Graph with the mentioned parameters - QD2-010 Risk Management Procedure s Criteria of the QM System QD2-001 Verification Planning QP2-002 Design Validation QD2-010 Risk Management Procedure QQ2-002 Retention Time of Quality Records QQ1-001 Quality Manual s Review Plan Annual Review as Review Plan QQ2-010 QQ2-006 Preparation and Management of the Technical Documentation QQ2-001 501 Management QQ1-001 Quality Manual s Procedure to Prove the Effectiveness of Risk Management Recognition of all functions, hazards and their causes, assessment of the individual risks, introduction, implementation and enabling of all individual measures decided to be necessary for the implementation, reassessment of the risks and acceptance of the residual risks. QD2-010 Risk Management Procedure s 2013-08-22 Page 15 of 141

Limits of acceptance for the procedure of evidence Method is only acceptable when either all non-acceptable risks have been eliminated or satisfactory reasons have been given for the release in the enabling sheet. Release Criteria QD2-010 Risk Management Procedure s Procedure for Verification Checking the risk management on correctness and completeness before it is enabled by the technical manager. QD2-010 Risk Management Procedure s Limit of Acceptance for Verification Exceptions from the procedure only with the written approval of the entire management. QD2-010 Risk Management Procedure s Procedure for Validation Checking the results of the user studies (focus groups) QP2-002 Design Validation and taking them into consideration, then: regular market investigations and updates of the risk analysis QD2-010 Risk Management Procedure s Limit of Acceptance for Validation Exceptions from the procedure only with the written approval of the entire management. QD2-010 Risk Management Procedure s 2013-08-22 Page 16 of 141

Grey Box 2013-08-22 Page 17 of 141

Risk Analysis Project: 1307033002, ONYX-2121DT Version: 1.00 Risk Analysis Function: F-01 Connect to Mains Hazard Cause Effect on Measure 2013-08-22 Page 18 of 141

Hazard Cause H-01 Electromagnetic energy C-01 Short circuit of any or all insulation that does not comply with the requirements of IEC 60601-1, 3rd Ed., Clause. 8.8 C-02 Short circuit of any or all CREEPAGE DISTANCES or AIR CLEARANCES that do not comply with the requirements of IEC 60601-1, 3rd Ed., Clause 8.9 C-03 Open circuit of any or all earth connections that do not comply with the requirements of IEC 60601-1, Clause 8.6, including any functional earth connection. C-04 Short circuit of any one insulation that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.8 C-05 Short circuit of any one CREEPAGE DISTANCE or AIR CLEARANCE that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.9 Effect on Measure M-01 The limits specified in IEC 60601-1, Clause 8.4 is not be exceeded for ACCESSIBLE PARTS Risk Analysis M-01 The limits specified in IEC 60601-1, Clause 8.4 is not be exceeded for ACCESSIBLE PARTS M-01 The limits specified in IEC 60601-1, Clause 8.4 is not be exceeded for ACCESSIBLE PARTS M-01 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. M-01 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. 2013-08-22 Page 19 of 141

Hazard Cause C-06 Open circuit of any one PROTECTIVE EARTH CONDUCTOR or internal PROTECTIVE EARTH CONNECTION that complies with the requirements of IEC 60601-1, Clause 8.6 C-07 All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION. C-08 The reliability of components that are used as MEANS OF PROTECTION are not assessed for the conditions of use in the PC. C-09 Storage (humidity preconditioning treatment) Effect on Measure M-01 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. M-01 All components, including wiring shall be used in accordance with their specified ratings. M-02 All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION are fixed in place by UL R/C nonchemical bonding compound. M-01 They shall comply with the applicable safety requirements of a relevant IEC, UL or ISO standard; M-02 Where there is no relevant IEC or ISO standard, the requirements of IEC 60601-1, 3rd Ed. were applied. M-01 The humidity preconditioning treatment is performed in a humidity cabinet containing air with a relative humidity of 93 % ± 3 % @ 25 C 48 hr C-10 Transport M-01 The strength of packing carton / fixed frame. The transportation stimulation for the verification. All connections were mechanically secure. C-11 Components of PC, unwanted movement of which could result in an unacceptableb RISK, is not mounted securely to prevent such movement. C-12 Too high leakage current, causing a death or serious injury to a patient. M-01 Said components were secured in place on pcb by UL Recognized non-chemical bonding compound. M-01 Meet IEC 60601-1, Clause 8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS requirements Risk Analysis 2013-08-22 Page 20 of 141

Hazard Cause C-13 The LEAKAGE CURRENTS from, to or between ACCESSIBLE PARTS other than PATIENT CONNECTIONS C-14 ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. C-15 Internal parts, other than contacts of plugs, connectors and socketoutlets, that can be touched by the test pin shown in IEC 60601-1, Figure 8. C-16 Accidental detachment of conductors and connectors where breaking free could lead to a HAZARDOUS SITUATION C-17 The resistance to heat is not retained by all types of insulation, including insulating partition walls, during the EXPECTED SERVICE LIFE of the PC. C-18 Internal cables and wiring is not adequately protected against contact with a moving part or Effect on Measure M-01 Meet IEC 60601-1, Clause 8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS requirements Risk Analysis M-01 Supplied by SELV power adapter and DC/AC inverter meet LCC requirements. M-01 The suitable enclosure shall be provided. M-01 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. M-01 Shall compliance with the following tests: resistance to moisture, etc. (see IEC 60601-1, Clause 11.6); dielectric strength (see IEC 60601-1, Clause 8.8.3); and mechanical strength (see IEC 60601-1, Clause 15.3). M-02 Ball pressure test - according to IEC 60601-1, Clause 8.8.4.1, b) requirements M-01 Internal cables and wiring is fixed in place by cable tie to protect against contact with a moving part or from friction at sharp corners and edges where damage to insulation 2013-08-22 Page 21 of 141

Hazard Cause from friction at sharp corners and edges where damage to insulation. C-19 Electromagnetic compatibility of PC. Effect on Measure M-01 Meet IEC 60601-1-2 requirements C-20 Packing M-01 The permissible environmental conditions for transport and storage were marked on the outside of the packaging. C-21 Wirings and connectors of Panel PC Monitor supply is not secured or insulated that accidental detachment can result in a HAZARDOUS SITUATION. C-22 Internal cables and wiring were not adequately protected against contact from friction at sharp corners and edges where damage to insulation could result in a HAZARDOUS SITUATION. H-02 Thermal Energy C-01 Excessive temperatures in PC and/or power supply. M-01 All wirings are tied together by cable tie to prevent accidental detachment M-01 The internal wirings were secured in place on PCB by UL R/C nonchemical bonding component. C-23 Cleaning M-01 The equipment complied with the requirements of IEC 60601-1, Clause 11.6. C-02 ENCLOSURES have no the strength and rigidity necessary to avoid a fire. M-01 When PC is operated in worst-case NORMAL USE including the maximum ambient operating temperature, the parts shall not reach temperatures exceeding the values given in IEC 60601-1, Table 22 and Table 23. M-01 Compliance with the mechanical strength tests for ENCLOSURES (see IEC 60601-1, Clause 15.3) Risk Analysis 2013-08-22 Page 22 of 141

Function: F-02 Enclosure Hazard Cause Effect on Measure H-01 Mechanical energy C-01 Falling M-01 Meet IEC 60601-1, Clause 9.4.2 requirements. Risk Analysis C-02 Suspended masses M-01 Meet IEC 60601-1 Clause 9.8.2 and Table 21 requirements. C-03 Sharp edges and corners M-01 Rounding of sharp edges and corners. H-02 Thermal Energy C-01 Excessive temperatures in enclosure M-01 When PC is operated in worst-case NORMAL USE including the maximum ambient operating temperature, the parts shall not reach temperatures exceeding the values given in IEC 60601-1, Table 22 and Table 23. Function: F-03 Rear Bracket Hazard Cause Effect on Measure H-01 Mechanical energy C-01 Crack M-01 Meet IEC 60601-1 Clause 9.8.2 and Table 21 requirements. C-02 Suspended masses M-01 Meet IEC 60601-1 Clause 9.8.2 and Table 21 requirements. C-03 Instability M-01 Meet IEC 60601-1, Clause 9.4.2 requirements. Function: F-04 Signal input Hazard Cause H-01 Electromagnetic energy C-01 ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. Effect on Measure M-01 Supplied by SELV power adapter and DC/AC inverter meet LCC requirements. 2013-08-22 Page 23 of 141

Function: F-05 Signal output Hazard Cause Effect on Measure H-01 Acoustic energy (sound) C-01 earphone M-01 Meet IEC 60950-1, Annex NAD Acoustic tests requirements Risk Analysis Function: F-06 User control and indicator Hazard Cause H-01 Operational C-01 Switches used to control power to PC or its parts, including switches, did not have their "on" and "off" positions Effect on Measure M-01 Indicated by an adjacent indicator light M-02 Marked with symbols IEC 60417 Function: F-07 Information Hazard Cause H-01 Labeling C-01 PC did not marke with the name or trademark of Onyx and with a MODEL OR TYPE REFERENCE C-02 Incomplete instructions for use C-03 Inadequate description of performance characteristics C-04 Inadequate specification of intended use. C-05 Inadequate disclosure of limitations Effect on Measure M-01 PC marked the name or trademark of Onyx with model no. M-01 Meet IEC 60601-1, Clause 7 requirements. M-01 Meet IEC 60601-1, Clause 7 requirements. M-01 Meet IEC 60601-1, Clause 7 requirements. M-01 Meet IEC 60601-1, Clause 7 requirements. 2013-08-22 Page 24 of 141

Hazard Cause H-02 Operating instructions C-01 Inadequate specification of accessories to be used with the medical device. Effect on Measure M-01 Meet IEC 60601-1, Clause 7 requirements. Risk Analysis C-02 Inadequate specification of pre-use checks. M-01 Meet IEC 60601-1, Clause 7 requirements. C-03 Over-complicated operating instructions M-01 Meet IEC 60601-1, Clause 7 requirements. H-03 Specification of service and maintenance C-01 Installation instruction M-01 Meet IEC 60601-1, Clause 7 requirements. Function: F-08 Connection to an external d.c. power source Hazard Cause H-01 Thermal Energy C-01 When a connection with the wrong polarity is made Effect on Measure M-01 Provided a connector that has special housing to prevent a connection with the wrong polarity is made Function: F-09 Display Hazard Cause Effect on Measure H-01 Radiation energy C-01 Visible Light M-01 The intensity of visible light from monitor can be controlled by the user. The highest intensity achievable cannot produce detrimental biological effects. Function: F-10 BIOS (Basic Input/Output System - real time clock) Hazard Cause Effect on Measure H-01 Mechanical energy C-01 Lithium battery explosion M-01 Battery circuits is designed so that discharging at a rate exceeding the battery manufacturer s recommendations, and unintentional charging, are prevented. 2013-08-22 Page 25 of 141

Risk Evaluation Before Measures Project: 1307033002, ONYX-2121DT Version: 1.00 Function: F-01 Connect to Mains Hazard Cause Severity Probability Region of Risk Comment Risk Evaluation Before Measures 2013-08-22 Page 26 of 141

Hazard Cause H-01 Electromagnetic energy C-01 Short circuit of any or all insulation that does not comply with the requirements of IEC 60601-1, 3rd Ed., Clause. 8.8 C-02 Short circuit of any or all CREEPAGE DISTANCES or AIR CLEARANCES that do not comply with the requirements of IEC 60601-1, 3rd Ed., Clause 8.9 C-03 Open circuit of any or all earth connections that do not comply with the requirements of IEC 60601-1, Clause 8.6, including any functional earth connection. C-04 Short circuit of any one insulation that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.8 C-05 Short circuit of any one CREEPAGE DISTANCE or AIR CLEARANCE that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.9 C-06 Open circuit of any one PROTECTIVE EARTH CONDUCTOR or internal PROTECTIVE EARTH CONNECTION that complies with the requirements of IEC 60601-1, Clause 8.6 C-07 All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION. C-08 The reliability of components that are used as MEANS OF PROTECTION are not assessed for the conditions of use in the PC. C-09 Storage (humidity preconditioning treatment) Severity Probability Region of Risk Serious Occasional Inacceptable Serious Occasional Inacceptable Serious Occasional Inacceptable Critical Occasional Inacceptable Critical Remote Inacceptable Critical Occasional Inacceptable Serious Remote Acceptable Critical Remote Inacceptable Serious Occasional Inacceptable Comment Risk Evaluation Before Measures 2013-08-22 Page 27 of 141

Hazard Cause Severity Probability Region of Risk C-10 Transport Serious Remote Acceptable C-11 Components of PC, unwanted movement of which could result in an unacceptableb RISK, is not mounted securely to prevent such movement. C-12 Too high leakage current, causing a death or serious injury to a patient. C-13 The LEAKAGE CURRENTS from, to or between ACCESSIBLE PARTS other than PATIENT CONNECTIONS C-14 ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. C-15 Internal parts, other than contacts of plugs, connectors and socket-outlets, that can be touched by the test pin shown in IEC 60601-1, Figure 8. Serious Remote Acceptable Critical Occasional Inacceptable Critical Occasional Inacceptable Critical Occasional Inacceptable Serious Remote Acceptable Comment Risk Evaluation Before Measures C-16 Accidental detachment of conductors and connectors where breaking free could lead to a HAZARDOUS SITUATION Serious Occasional Inacceptable C-17 The resistance to heat is not retained by all types of insulation, including insulating partition walls, during the EXPECTED SERVICE LIFE of the PC. Serious Occasional Inacceptable C-18 Internal cables and wiring is not adequately protected against contact with a moving part or from friction at sharp corners and edges where damage to insulation. Serious Occasional Inacceptable C-19 Electromagnetic compatibility of PC. Critical Occasional Inacceptable C-20 Packing Serious Occasional Inacceptable 2013-08-22 Page 28 of 141

Hazard Cause C-21 Wirings and connectors of Panel PC Monitor supply is not secured or insulated that accidental detachment can result in a HAZARDOUS SITUATION. C-22 Internal cables and wiring were not adequately protected against contact from friction at sharp corners and edges where damage to insulation could result in a HAZARDOUS SITUATION. Severity Probability Region of Risk Serious Remote Acceptable Serious Remote Acceptable C-23 Cleaning Minor Remote Acceptable H-02 Thermal Energy C-01 Excessive temperatures in PC and/or power supply. C-02 ENCLOSURES have no the strength and rigidity necessary to avoid a fire. Serious Occasional Inacceptable Critical Remote Inacceptable Comment Risk Evaluation Before Measures Function: F-02 Enclosure Hazard Cause Severity Probability Region of Risk H-01 Mechanical energy C-01 Falling Serious Occasional Inacceptable C-02 Suspended masses Serious Occasional Inacceptable C-03 Sharp edges and corners Minor Probable Inacceptable H-02 Thermal Energy C-01 Excessive temperatures in enclosure Serious Occasional Inacceptable Comment 2013-08-22 Page 29 of 141

Function: F-03 Rear Bracket Hazard Cause Severity Probability Region of Risk H-01 Mechanical energy C-01 Crack Serious Occasional Inacceptable C-02 Suspended masses Serious Remote Acceptable C-03 Instability Serious Occasional Inacceptable Comment Risk Evaluation Before Measures Function: F-04 Signal input Hazard Cause H-01 Electromagnetic energy C-01 ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. Severity Probability Region of Risk Critical Remote Inacceptable Comment Function: F-05 Signal output Hazard Cause Severity Probability Region of Risk H-01 Acoustic energy (sound) C-01 earphone Serious Occasional Inacceptable Comment Function: F-06 User control and indicator Hazard Cause H-01 Operational C-01 Switches used to control power to PC or its parts, including switches, did not have their "on" and "off" positions Severity Probability Region of Risk Serious Occasional Inacceptable Comment 2013-08-22 Page 30 of 141

Function: F-07 Information Hazard Cause H-01 Labeling C-01 PC did not marke with the name or trademark of Onyx and with a MODEL OR TYPE REFERENCE Severity Probability Region of Risk Minor Occasional Acceptable C-02 Incomplete instructions for use Critical Occasional Inacceptable C-03 Inadequate description of performance characteristics Serious Occasional Inacceptable C-04 Inadequate specification of intended use. Critical Occasional Inacceptable C-05 Inadequate disclosure of limitations Critical Occasional Inacceptable H-02 Operating instructions C-01 Inadequate specification of accessories to be used with the medical device. C-02 Inadequate specification of pre-use checks. H-03 Specification of service and maintenance Serious Remote Acceptable Minor Occasional Acceptable C-03 Over-complicated operating instructions Minor Occasional Acceptable C-01 Installation instruction Serious Occasional Inacceptable Comment Risk Evaluation Before Measures Function: F-08 Connection to an external d.c. power source Hazard Cause H-01 Thermal Energy C-01 When a connection with the wrong polarity is made Severity Probability Region of Risk Serious Occasional Inacceptable Comment 2013-08-22 Page 31 of 141

Function: F-09 Display Hazard Cause Severity Probability Region of Risk H-01 Radiation energy C-01 Visible Light Minor Occasional Acceptable Function: F-10 BIOS (Basic Input/Output System - real time clock) Hazard Cause Severity Probability Region of Risk H-01 Mechanical energy C-01 Lithium battery explosion Critical Occasional Inacceptable Comment Comment Risk Evaluation Before Measures 2013-08-22 Page 32 of 141

Risk Graph Before Measures Project: 1307033002, ONYX-2121DT Version: 1.00 Both Conditions Operating Condition Risk Graph Before Measures Failure Mode Acceptable ALARP Inacceptable 2013-08-22 Page 33 of 141

Decision about Measures Project: 1307033002, ONYX-2121DT Version: 1.00 Measure: RC-001 All components, including wiring shall be used in accordance with their specified ratings. Implement: Decision about Measures Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION. Effective Measure: RC-002 All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION are fixed in place by UL R/C non-chemical bonding compound. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION. Effective Measure: RC-003 All wirings are tied together by cable tie to prevent accidental detachment Implement: 2013-08-22 Page 34 of 141

Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Wirings and connectors of Panel PC Monitor supply is not secured or insulated that accidental detachment can result in a HAZARDOUS SITUATION. Effective Measure: RC-004 Ball pressure test - according to IEC 60601-1, Clause 8.8.4.1, b) requirements Decision about Measures Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy The resistance to heat is not retained by all types of insulation, including insulating partition walls, during the EXPECTED SERVICE LIFE of the PC. Effective Measure: RC-005 Battery circuits is designed so that discharging at a rate exceeding the battery manufacturer s recommendations, and unintentional charging, are prevented. Implement: Function Hazard Cause Effectiveness Assessment of the Measure BIOS (Basic Input/Output System Mechanical energy Lithium battery explosion Effective - real time clock) 2013-08-22 Page 35 of 141

Measure: RC-006 Compliance with the mechanical strength tests for ENCLOSURES (see IEC 60601-1, Clause 15.3) Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Thermal Energy ENCLOSURES have no the strength and rigidity necessary to avoid a fire. Effective Decision about Measures Measure: RC-007 Indicated by an adjacent indicator light Implement: Function Hazard Cause Effectiveness Assessment of the Measure User control and indicator Operational Switches used to control power to PC or its parts, including switches, did not have their "on" and "off" positions Effective Measure: RC-008 Internal cables and wiring is fixed in place by cable tie to protect against contact with a moving part or from friction at sharp corners and edges where damage to insulation Implement: 2013-08-22 Page 36 of 141

Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Internal cables and wiring is not adequately protected against contact with a moving part or from friction at sharp corners and edges where damage to insulation. Effective Measure: Marked with symbols IEC 60417 RC-009 Decision about Measures Implement: Function Hazard Cause Effectiveness Assessment of the Measure User control and indicator Operational Switches used to control power to PC or its parts, including switches, did not have their "on" and "off" positions Effective Measure: RC-010 Meet IEC 60601-1 Clause 9.8.2 and Table 21 requirements. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Enclosure Mechanical energy Suspended masses Effective Rear Bracket Mechanical energy Crack Effective Rear Bracket Mechanical energy Suspended masses Effective 2013-08-22 Page 37 of 141

Measure: RC-011 Meet IEC 60601-1, Clause 7 requirements. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Information Labeling Incomplete instructions for use Effective Information Labeling Inadequate description of performance characteristics Effective Information Labeling Inadequate specification of intended use. Effective Information Labeling Inadequate disclosure of limitations Effective Information Operating instructions Inadequate specification of accessories to be used with the medical device. Effective Information Operating instructions Inadequate specification of preuse checks. Effective Information Operating instructions Over-complicated operating instructions Effective Information Specification of service and maintenance Installation instruction Effective Decision about Measures Measure: RC-012 Meet IEC 60601-1, Clause 8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS requirements Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Too high leakage current, causing a death or serious injury to a patient. Effective 2013-08-22 Page 38 of 141

Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy The LEAKAGE CURRENTS from, to or between ACCESSIBLE PARTS other than PATIENT CONNECTIONS Effective Measure: RC-013 Meet IEC 60601-1, Clause 9.4.2 requirements. Decision about Measures Implement: Function Hazard Cause Effectiveness Assessment of the Measure Enclosure Mechanical energy Falling Effective Rear Bracket Mechanical energy Instability Effective Measure: Meet IEC 60601-1-2 requirements RC-014 Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Electromagnetic compatibility of Effective PC. Measure: RC-015 Meet IEC 60950-1, Annex NAD Acoustic tests requirements Implement: 2013-08-22 Page 39 of 141

Function Hazard Cause Effectiveness Assessment of the Measure Signal output Acoustic energy (sound) earphone Effective Measure: RC-016 PC marked the name or trademark of Onyx with model no. Implement: Decision about Measures Function Hazard Cause Effectiveness Assessment of the Measure Information Labeling PC did not marke with the name or trademark of Onyx and with a MODEL OR TYPE REFERENCE Effective Measure: RC-017 Provided a connector that has special housing to prevent a connection with the wrong polarity is made Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connection to an external d.c. Thermal Energy When a connection with the wrong Effective power source polarity is made Measure: RC-018 Rounding of sharp edges and corners. Implement: 2013-08-22 Page 40 of 141

Function Hazard Cause Effectiveness Assessment of the Measure Enclosure Mechanical energy Sharp edges and corners Effective Measure: RC-019 Said components were secured in place on pcb by UL Recognized non-chemical bonding compound. Implement: Decision about Measures Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Components of PC, unwanted movement of which could result in an unacceptableb RISK, is not mounted securely to prevent such movement. Effective Measure: RC-020 Shall compliance with the following tests: resistance to moisture, etc. (see IEC 60601-1, Clause 11.6); dielectric strength (see IEC 60601-1, Clause 8.8.3); and mechanical strength (see IEC 60601-1, Clause 15.3). Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy The resistance to heat is not retained by all types of insulation, including insulating partition walls, during the EXPECTED SERVICE LIFE of the PC. Effective 2013-08-22 Page 41 of 141

Measure: RC-021 Supplied by SELV power adapter and DC/AC inverter meet LCC requirements. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. Effective Signal input Electromagnetic energy ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. Effective Decision about Measures Measure: RC-022 The equipment complied with the requirements of IEC 60601-1, Clause 11.6. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Cleaning Effective Measure: RC-023 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. Implement: 2013-08-22 Page 42 of 141

Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Short circuit of any one insulation that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.8 Effective Connect to Mains Electromagnetic energy Short circuit of any one CREEPAGE DISTANCE or AIR CLEARANCE that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.9 Effective Connect to Mains Electromagnetic energy Open circuit of any one PROTECTIVE EARTH CONDUCTOR or internal PROTECTIVE EARTH CONNECTION that complies with the requirements of IEC 60601-1, Clause 8.6 Effective Connect to Mains Electromagnetic energy Accidental detachment of conductors and connectors where breaking free could lead to a HAZARDOUS SITUATION Effective Decision about Measures Measure: RC-024 The humidity preconditioning treatment is performed in a humidity cabinet containing air with a relative humidity of 93 % ± 3 % @ 25 C 48 hr Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Storage (humidity preconditioning Effective treatment) 2013-08-22 Page 43 of 141

Measure: RC-025 The intensity of visible light from monitor can be controlled by the user. The highest intensity achievable cannot produce detrimental biological effects. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Display Radiation energy Visible Light Effective Decision about Measures Measure: RC-026 The internal wirings were secured in place on PCB by UL R/C non-chemical bonding component. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Internal cables and wiring were not adequately protected against contact from friction at sharp corners and edges where damage to insulation could result in a HAZARDOUS SITUATION. Effective Measure: RC-027 The limits specified in IEC 60601-1, Clause 8.4 is not be exceeded for ACCESSIBLE PARTS Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Short circuit of any or all insulation that does not comply with the requirements of IEC 60601-1, 3rd Ed., Clause. 8.8 Effective 2013-08-22 Page 44 of 141

Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Short circuit of any or all CREEPAGE DISTANCES or AIR CLEARANCES that do not comply with the requirements of IEC 60601-1, 3rd Ed., Clause 8.9 Effective Connect to Mains Electromagnetic energy Open circuit of any or all earth connections that do not comply with the requirements of IEC 60601-1, Clause 8.6, including any functional earth connection. Effective Decision about Measures Measure: RC-028 The permissible environmental conditions for transport and storage were marked on the outside of the packaging. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Packing Effective Measure: RC-029 The strength of packing carton / fixed frame. The transportation stimulation for the verification. All connections were mechanically secure. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Transport Effective 2013-08-22 Page 45 of 141

Measure: RC-030 The suitable enclosure shall be provided. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy Internal parts, other than contacts of plugs, connectors and socketoutlets, that can be touched by the test pin shown in IEC 60601-1, Figure 8. Effective Decision about Measures Measure: RC-031 They shall comply with the applicable safety requirements of a relevant IEC, UL or ISO standard; Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy The reliability of components that are used as MEANS OF PROTECTION are not assessed for the conditions of use in the PC. Effective Measure: RC-032 When PC is operated in worst-case NORMAL USE including the maximum ambient operating temperature, the parts shall not reach temperatures exceeding the values given in IEC 60601-1, Table 22 and Table 23. Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Thermal Energy Excessive temperatures in PC Effective and/or power supply. 2013-08-22 Page 46 of 141

Function Hazard Cause Effectiveness Assessment of the Measure Enclosure Thermal Energy Excessive temperatures in Effective enclosure Measure: RC-033 Where there is no relevant IEC or ISO standard, the requirements of IEC 60601-1, 3rd Ed. were applied. Decision about Measures Implement: Function Hazard Cause Effectiveness Assessment of the Measure Connect to Mains Electromagnetic energy The reliability of components that are used as MEANS OF PROTECTION are not assessed for the conditions of use in the PC. Effective 2013-08-22 Page 47 of 141

Risk Control Project: 1307033002, ONYX-2121DT Version: 1.00 Risk Control No. Measure RC-001 All components, including wiring shall be used in accordance with their specified ratings. RC-002 All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION are fixed in place by UL R/C nonchemical bonding compound. RC-003 All wirings are tied together by cable tie to prevent accidental detachment RC-004 Ball pressure test - according to IEC 60601-1, Clause 8.8.4.1, b) requirements RC-005 Battery circuits is designed so that discharging at a rate exceeding the battery manufacturer s recommendations, and unintentional charging, are prevented. State of Work Validation Responsible Implementation Planning Actual Values Effort Supervision Begin End Begin End Planned Is Release Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/2/2013 6/18/2013 Yes Yes Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/3/2013 7/17/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/12/2013 7/5/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/6/2013 6/28/2013 Yes Yes Completed Yuan, Andy Lin, Eason 5/1/2013 8/31/2013 5/8/2013 6/18/2013 Yes Yes 2013-08-22 Page 48 of 141

No. Measure RC-006 Compliance with the mechanical strength tests for ENCLOSURES (see IEC 60601-1, Clause 15.3) RC-007 Indicated by an adjacent indicator light RC-008 Internal cables and wiring is fixed in place by cable tie to protect against contact with a moving part or from friction at sharp corners and edges where damage to insulation RC-009 Marked with symbols IEC 60417 RC-010 Meet IEC 60601-1 Clause 9.8.2 and Table 21 requirements. RC-011 Meet IEC 60601-1, Clause 7 requirements. RC-012 Meet IEC 60601-1, Clause 8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS requirements RC-013 Meet IEC 60601-1, Clause 9.4.2 requirements. State of Work Validation Responsible Implementation Planning Actual Values Effort Supervision Begin End Begin End Planned Is Release Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/13/2013 6/25/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/14/2013 6/18/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/3/2013 6/20/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/2/2013 6/25/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/8/2013 7/5/2013 Yes Yes Completed Lin, Eason Chen, Ernest 5/1/2013 8/31/2013 5/6/2013 7/5/2013 Yes Yes Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/6/2013 7/15/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/9/2013 6/27/2013 Yes Yes Risk Control 2013-08-22 Page 49 of 141

No. Measure RC-014 Meet IEC 60601-1-2 requirements RC-015 Meet IEC 60950-1, Annex NAD Acoustic tests requirements RC-016 PC marked the name or trademark of Onyx with model no. RC-017 Provided a connector that has special housing to prevent a connection with the wrong polarity is made RC-018 Rounding of sharp edges and corners. RC-019 Said components were secured in place on pcb by UL Recognized nonchemical bonding compound. RC-020 Shall compliance with the following tests: resistance to moisture, etc. (see IEC 60601-1, Clause 11.6); dielectric strength (see IEC 60601-1, Clause 8.8.3); and mechanical strength (see IEC 60601-1, Clause 15.3). State of Work Validation Responsible Implementation Planning Actual Values Effort Supervision Begin End Begin End Planned Is Release Completed Yuan, Andy Lin, Eason 5/1/2013 8/31/2013 5/10/2013 6/19/2013 Yes Yes Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/3/2013 7/8/2013 Yes Yes Completed Lin, Eason Chen, Ernest 5/1/2013 8/31/2013 5/8/2013 6/21/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/7/2013 6/25/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/13/2013 6/25/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/24/2013 6/28/2013 Yes Yes Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/17/2013 7/8/2013 Yes Yes Risk Control 2013-08-22 Page 50 of 141

No. Measure RC-021 Supplied by SELV power adapter and DC/AC inverter meet LCC requirements. RC-022 The equipment complied with the requirements of IEC 60601-1, Clause 11.6. RC-023 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. RC-024 The humidity preconditioning treatment is performed in a humidity cabinet containing air with a relative humidity of 93 % ± 3 % @ 25 C 48 hr RC-025 The intensity of visible light from monitor can be controlled by the user. The highest intensity achievable cannot produce detrimental biological effects. RC-026 The internal wirings were secured in place on PCB by UL R/C non-chemical bonding component. State of Work Validation Responsible Implementation Planning Actual Values Effort Supervision Begin End Begin End Planned Is Release Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/5/2013 7/22/2013 Yes Yes Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/7/2013 7/11/2013 Yes Yes Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/3/2013 6/28/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/8/2013 6/26/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/7/2013 7/4/2013 Yes Yes Completed Chen, Ernest Yuan, Andy 5/1/2013 8/31/2013 5/8/2013 6/25/2013 Yes Yes Risk Control 2013-08-22 Page 51 of 141

No. Measure RC-027 The limits specified in IEC 60601-1, Clause 8.4 is not be exceeded for ACCESSIBLE PARTS RC-028 The permissible environmental conditions for transport and storage were marked on the outside of the packaging. RC-029 The strength of packing carton / fixed frame. The transportation stimulation for the verification. All connections were mechanically secure. RC-030 The suitable enclosure shall be provided. RC-031 They shall comply with the applicable safety requirements of a relevant IEC, UL or ISO standard; RC-032 When PC is operated in worst-case NORMAL USE including the maximum ambient operating temperature, the parts shall not reach temperatures exceeding the values given in IEC 60601-1, Table 22 and Table 23. State of Work Validation Responsible Implementation Planning Actual Values Effort Supervision Begin End Begin End Planned Is Release Completed Yuan, Andy Lin, Eason 5/1/2013 8/31/2013 5/8/2013 7/9/2013 Yes Yes Completed Chen, Ernest Lin, Eason 5/1/2013 8/31/2013 5/8/2013 7/18/2013 Yes Yes Completed Chen, Ernest Lin, Eason 5/1/2013 8/31/2013 5/9/2013 7/15/2013 Yes Yes Completed Chen, Ernest Lin, Eason 5/1/2013 8/31/2013 5/2/2013 7/2/2013 Yes Yes Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/8/2013 7/12/2013 Yes Yes Completed Yuan, Andy Chen, Ernest 5/1/2013 8/31/2013 5/7/2013 7/10/2013 Yes Yes Risk Control 2013-08-22 Page 52 of 141

No. Measure RC-033 Where there is no relevant IEC or ISO standard, the requirements of IEC 60601-1, 3rd Ed. were applied. State of Work Validation Responsible Implementation Planning Actual Values Effort Supervision Begin End Begin End Planned Is Release Completed Lin, Mac Yuan, Andy 5/1/2013 8/31/2013 5/6/2013 6/27/2013 Yes Yes Risk Control 2013-08-22 Page 53 of 141

Release Protocol of Measures Project: 1307033002, ONYX-2121DT Version: 1.00 Measure Number: Measure Text: Measure Status: RC-001 All components, including wiring shall be used in accordance with their specified ratings. Completed Release Protocol of Measures Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Planning: Begin: 5/1/2013 End: 8/31/2013 Actual Values: Begin: 5/2/2013 End: 6/18/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: All components, including wiring shall be used in accordance with their specified ratings. Where safety is involved, components shall comply either with the requirements of IEC 60601-1 or with the safety aspects of the relevant IEC component standards. Functions Hazards Causes Connect to Mains Electromagnetic energy All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 54 of 141

Procedure for Evidence of Efficency of Measures: Evaluation and testing of components shall be conducted as follows: - a component that has been demonstrated to comply with a standard harmonized with the relevant IEC component standard shall be checked for correct application and use in accordance with its rating. It shall be subjected to the applicable tests of IEC 60601-1 as part of the equipment with the exception of those tests which are part of the relevant IEC component standard; - a component that has not been demonstrated to comply with a relevant standard as above shall be checked for correct application and use in accordance with its specified rating. It shall be subjected to the applicable tests of IEC 60601-1, as part of the equipment, and to the applicable tests of the component standard, under the conditions occurring in the equipment; - where no relevant IEC component standard exists, or where components are used in circuits not in accordance with their specified ratings, the components shall be tested under the conditions occurring in the equipment. The number of samples required for test is, in general, the same as required by an equivalent standard. QD2-001, QD2-004, QD2-005 Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: QD2-001, QD2-004, QD2-005 Procedure of Verification: QD2-001, QD2-004, QD2-005 Limits of Acceptance for Verification: QD2-001, QD2-004, QD2-005 Procedure of Validation: QD2-001, QD2-004, QD2-005 Limits of Acceptance for Validation: QD2-001, QD2-004, QD2-005 2013-08-22 Page 55 of 141

Measure Number: Measure Text: Measure Status: RC-002 All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION are fixed in place by UL R/C non-chemical bonding compound. Completed Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Release Protocol of Measures Planning: Begin: 5/1/2013 End: 8/31/2013 Actual Values: Begin: 5/3/2013 End: 7/17/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION are fixed in place by UL R/C non-chemical bonding compound. Conductors shall be provided with a means (for example, barriers or fixing), or be so terminated, that they and their terminators (for example, ring terminals and flat quick-connect terminals) cannot, in normal use, become so displaced that CLEARANCES or CREEPAGE DISTANCES are reduced to less than the values specified in IEC 60601-1. Functions Hazards Causes Connect to Mains Electromagnetic energy All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 56 of 141

Procedure for Evidence of Efficency of Measures: Having used the unjointed probe, which was in the shape of the standard test finger and which incorporated a force gauge, a steady force of 10 N for internal parts was applied for a period of 5 seconds. QD2-001, QM2-011 Limits of Acceptance for Procedure of Evidence: During the application of the test force, clearances behind earthed or unearthed conductive accessable parts were not reduced to a level that would result in an energy hazard. QD2-001, QM2-011 Release Protocol of Measures Procedure of Verification: See above QD2-001, QM2-011 Limits of Acceptance for Verification: See above QD2-001, QM2-011 Procedure of Validation: See above QD2-001, QM2-011 Limits of Acceptance for Validation: See above QD2-001, QM2-011 2013-08-22 Page 57 of 141

Measure Number: Measure Text: Measure Status: RC-003 All wirings are tied together by cable tie to prevent accidental detachment Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/12/2013 End: 7/5/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Tied together by cable tie. Conductors shall be provided with a means (for example, barriers or fixing), or be so terminated, that they and their terminators (for example, ring terminals and flat quick-connect terminals) cannot, in normal use, become so displaced that CLEARANCES or CREEPAGE DISTANCES are reduced to less than the values specified in IEC 60601-1. Functions Hazards Causes Connect to Mains Electromagnetic energy Wirings and connectors of Panel PC Monitor supply is not secured or insulated that accidental detachment can result in a HAZARDOUS SITUATION. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: Having used the unjointed probe, which was in the shape of the standard test finger and which incorporated a force gauge, a steady force of 10 N for internal parts was applied for a period of 5 seconds. QD2-001, QM2-011 2013-08-22 Page 58 of 141

Limits of Acceptance for Procedure of Evidence: During the application of the test force, clearances behind earthed or unearthed conductive accessible parts were not reduced to a level that would result in an energy hazard. QD2-001, QM2-011 Procedure of Verification: See above QD2-001, QM2-011 Release Protocol of Measures Limits of Acceptance for Verification: See above QD2-001, QM2-011 Procedure of Validation: See above QD2-001, QM2-011 Limits of Acceptance for Validation: See above QD2-001, QM2-011 2013-08-22 Page 59 of 141

Measure Number: Measure Text: Measure Status: RC-004 Ball pressure test - according to IEC 60601-1, Clause 8.8.4.1, b) requirements Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/6/2013 End: 6/28/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Compliance is checked by subjecting the part to the ballpressure test according to IEC 60695-10-2 and IEC 60601-1, Clause 8.8.4.1, b) Thermoplastic parts on which parts at HAZARDOUS VOLTAGE are directly mounted shall be resistant to abnormal heat. Functions Hazards Causes Connect to Mains Electromagnetic energy The resistance to heat is not retained by all types of insulation, including insulating partition walls, during the EXPECTED SERVICE LIFE of the PC. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 60 of 141

Procedure for Evidence of Efficency of Measures: A sample of each external part of insulating material indicated in the table was tested using the test apparatus shown in Fig. 21 of IEC 60601-1, 3rd Edition. (In this apparatus, a steel ball of 5 mm diameter exerts a force of 20 N when pressed against a surface). The apparatus and a fire (or sand) brick were placed in an air-circulating oven. Once the apparatus and brick attained the oven temperature, the sample was placed in the oven on top of the brick. The steel ball was then positioned such that the apparatus was balanced on the surface of the sample. After 1 h, it was removed and the diameter of the impression made by the ball was measured. The specified oven temperature was the higher of the following: a. the sum of the specified ambient and the temperature rise of relevant part during the Temperature Test; AND b. 75 C ± 2 C (test per Clause 8.8.4.1a), OR 125 C ± 2 C (test per Clause 8.8.4.1b) IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: The results were less 2 mm. Cracks were not visible to the naked eye. IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 8.8.4.1 2013-08-22 Page 61 of 141

Measure Number: Measure Text: Measure Status: RC-005 Battery circuits is designed so that discharging at a rate exceeding the battery manufacturer s recommendations, and unintentional charging, are prevented. Completed Responsible: Implementation: Yuan, Andy Supervision: Lin, Eason Release Protocol of Measures Planning: Begin: 5/1/2013 End: 8/31/2013 Actual Values: Begin: 5/8/2013 End: 6/18/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Battery circuits is designed so that discharging at a rate exceeding the battery manufacturer s recommendations, and unintentional charging, are prevented. Equipment containing batteries shall be designed to reduce the risk of fire, explosion and chemical leaks under normal conditions and after a single fault in the equipment, including a fault in circuitry within the equipment battery pack. Functions Hazards Causes BIOS (Basic Input/Output System - Mechanical energy Lithium battery explosion real time clock) Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 62 of 141

Procedure for Evidence of Efficency of Measures: The non-rechargeable battery cell was subjected to the test detailed below. The unit was connected to rated input voltage and operated at maximum normal load. The following conditions were imposed one at a time. 1) Unintentional charging of a non-rechargeable battery (pack or cell). The battery was charged while briefly subjected to the simulation of any single component failure that was likely to occur in the charging circuit and that would result in unintentional charging of the battery. To minimize testing time, the failure was chosen that causes the highest charging current. The battery was then charged for a single period of 7 h with that simulated failure in place. 2) Excessive discharging rate for any battery (pack or cell). The battery was subjected to rapid discharge by open-circuiting or short-circuiting any current-limiting or voltage-limiting components in the load circuit of the battery under test. 3) Reversing - The connections of the battery were reversed and the equipment was energized. The battery was examined for emission of flames, toxic gases, molten metal, risk of explosion. IEC 60601-1, Clause 8.8.3 and 15.4.3 Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: These tests shall not result in any of the following: - chemical leaks caused by cracking, rupturing or bursting of the battery jacket, if such leakage could adversely affect required insulation; or - spillage of liquid from any pressure relief device in the battery, unless such spillage is contained by the equipment without risk of damage to the insulation or harm to the USER; or - explosion of the battery, if such explosion could result in injury to a USER; or - emission of flame or expulsion of molten metal to the outside of the equipment ENCLOSURE. After completion of the tests, the equipment is subjected to the electric strength tests of IEC 60601-1, Clause 8.8.3 IEC 60601-1, Clause 8.8.3 and 15.4.3 Procedure of Verification: IEC 60601-1, Clause 8.8.3 and 15.4.3 Limits of Acceptance for Verification: IEC 60601-1, Clause 8.8.3 and 15.4.3 Procedure of Validation: IEC 60601-1, Clause 8.8.3 and 15.4.3 Limits of Acceptance for Validation: IEC 60601-1, Clause 8.8.3 and 15.4.3 2013-08-22 Page 63 of 141

Measure Number: Measure Text: Measure Status: RC-006 Compliance with the mechanical strength tests for ENCLOSURES (see IEC 60601-1, Clause 15.3) Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/13/2013 End: 6/25/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The suitable plastic enclosure is provided. The PC or its parts shall have adequate mechanical strength and shall not result in an unacceptable RISK due to moulding stress or when subjected to mechanical stress caused by pushing, impact, dropping, and rough handling. Functions Hazards Causes Connect to Mains Thermal Energy ENCLOSURES have no the strength and rigidity necessary to avoid a fire. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: Compliance is checked by application of the tests in Table 28. The tests are not applied to handles, levers, knobs, the face of cathode ray tubes (see 9.5.2), or to transparent or translucent covers of indicating or measuring devices unless with the handle, lever, knob, or cover removed there is an unacceptable RISK of electric shock. IEC 60601-1, Clause 15.3 2013-08-22 Page 64 of 141

Limits of Acceptance for Procedure of Evidence: There was no cracking of the enclosure. There were no live parts that became accessible. IEC 60601-1, Clause 15.3 Procedure of Verification: Force (15.3.2) - A sample of the unit was subjected to inward directed force of 250 N plus or minus 10 N for a period of 5 seconds applied to different parts of the enclosure via a test tool providing contact over a circular plane surface 30 mm in diameter as indicated in Table 15.3. Note: not applied to the bottom of a product having a mass of greater than 18 kg. Impact (15.3.3) - The impact resistance of an enclosure was tested by application of an impact conveyed via a solid steel 50 mm (approx) ball of mass 500 g plus or minus 25 g, fall freely from 1.3 meter height. The unit was rigidly supported and the impact was applied to the locations indicated in Table 15.3. IEC 60601-1, Clause 15.3 Release Protocol of Measures Limits of Acceptance for Verification: IEC 60601-1, Clause 15.3 Procedure of Validation: IEC 60601-1, Clause 15.3 Limits of Acceptance for Validation: IEC 60601-1, Clause 15.3 2013-08-22 Page 65 of 141

Measure Number: Measure Text: Measure Status: RC-007 Indicated by an adjacent indicator light Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/14/2013 End: 6/18/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The colours of indicator lights and their meanings shall comply with Table 2 of IEC 60601-1. It is important for an OPERATOR and SERVICE PERSONNEL to be able to determine the functional status of ME EQUIPMENT. In NORMAL USE, the OPERATOR needs to be able to distinguish between ME EQUIPMENT in stand-by and ME EQUIPMENT in a fully functional state. Functions Hazards Causes User control and indicator Operational Switches used to control power to PC or its parts, including switches, did not have their "on" and "off" positions Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: QD2-001, QM2-011 IEC 60601-1, Clause 7.8.1 and 15.4.4 2013-08-22 Page 66 of 141

Limits of Acceptance for Procedure of Evidence: QD2-001, QM2-011 IEC 60601-1, Clause 7.8.1 and 15.4.4 Procedure of Verification: QD2-001, QM2-011 IEC 60601-1, Clause 7.8.1 and 15.4.4 Release Protocol of Measures Limits of Acceptance for Verification: QD2-001, QM2-011 IEC 60601-1, Clause 7.8.1 and 15.4.4 Procedure of Validation: QD2-001, QM2-011 IEC 60601-1, Clause 7.8.1 and 15.4.4 Limits of Acceptance for Validation: QD2-001, QM2-011 IEC 60601-1, Clause 7.8.1 and 15.4.4 2013-08-22 Page 67 of 141

Measure Number: Measure Text: Measure Status: RC-008 Internal cables and wiring is fixed in place by cable tie to protect against contact with a moving part or from friction at sharp corners and edges where damage to insulation Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Release Protocol of Measures Planning: Begin: 5/1/2013 End: 8/31/2013 Actual Values: Begin: 5/3/2013 End: 6/20/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The internal wirings were secured in place on PCB by UL R/C non-chemical bounding component. Wireways shall be smooth and free from sharp edges. Wires shall be protected so that they do not come into contact with burrs, cooling fins, moving parts, etc., which could cause damage to the insulation of conductors. Holes in metal, through which insulated wires pass, shall have smooth well-rounded surfaces or shall be provided with bushings. Functions Hazards Causes Connect to Mains Electromagnetic energy Internal cables and wiring is not adequately protected against contact with a moving part or from friction at sharp corners and edges where damage to insulation. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 68 of 141

Procedure for Evidence of Efficency of Measures: QD2-001, QD2-010, QM2-011, QD2-009 QQ1-001 Limits of Acceptance for Procedure of Evidence: QD2-001, QD2-010, QM2-011, QD2-009 QQ1-001 Release Protocol of Measures Procedure of Verification: QD2-001, QD2-010, QM2-011, QD2-009 QQ1-001 Limits of Acceptance for Verification: QD2-001, QD2-010, QM2-011, QD2-009 QQ1-001 Procedure of Validation: QD2-001, QD2-010, QM2-011, QD2-009 QQ1-001 Limits of Acceptance for Validation: QD2-001, QD2-010, QM2-011, QD2-009 QQ1-001 2013-08-22 Page 69 of 141

Measure Number: Measure Text: Measure Status: RC-009 Marked with symbols IEC 60417 Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/2/2013 End: 6/25/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: - marked with symbols IEC 60417, - indicated by an adjacent indicator light Comment: Switches used to control power to ME EQUIPMENT or its parts, including mains switches, shall have their "on" and "off" positions Functions Hazards Causes User control and indicator Operational Switches used to control power to PC or its parts, including switches, did not have their "on" and "off" positions Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: Compliance with the requirements is checked by inspection and by application of the tests and criteria in 7.1.2 and 7.1.3 of IEC 60601-1. IEC 60601-1, Clause 7.4.1, 7.4.2, 7.8.1 and 15.4.4 2013-08-22 Page 70 of 141

Limits of Acceptance for Procedure of Evidence: The colours of indicator lights and their meanings shall comply with Table 2 of IEC 60601-1. IEC 60601-1, Clause 7.4.1, 7.4.2, 7.8.1 and 15.4.4 Procedure of Verification: IEC 60601-1, Clause 7.4.1, 7.4.2, 7.8.1 and 15.4.4 Release Protocol of Measures Limits of Acceptance for Verification: IEC 60601-1, Clause 7.4.1, 7.4.2, 7.8.1 and 15.4.4 Procedure of Validation: IEC 60601-1, Clause 7.4.1, 7.4.2, 7.8.1 and 15.4.4 Limits of Acceptance for Validation: IEC 60601-1, Clause 7.4.1, 7.4.2, 7.8.1 and 15.4.4 2013-08-22 Page 71 of 141

Measure Number: Measure Text: Measure Status: RC-010 Meet IEC 60601-1 Clause 9.8.2 and Table 21 requirements. Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/8/2013 End: 7/5/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The construction of the support, suspension or actuation system shall be designed based upon Table 21 of IEC 60601-1 and the TOTAL LOAD. Where ME EQUIPMENT parts are designed to support loads, then a mechanical fault could not constitute an unacceptable RISK. Functions Hazards Causes Enclosure Mechanical energy Suspended masses Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: The suspension system parts indicated in the table below (chain, cable, band, spring, belt, jack screw nut, pneumatic, hydraulic hoses, or the like) employed to support a load, were defeated by convenient means (cut or removed suddenly), thereby causing the maximum normal load to fall from the most adverse position permitted by the construction of the equipment and to activate a safety device. The travel after such defeat was measured as well as other observed results that could affect the risk of possible injury. IEC 60601-1, Clause 9.8.2 2013-08-22 Page 72 of 141

Limits of Acceptance for Procedure of Evidence: There was no evidence of damage to a safety catch or other restraining means that would affect its ability to perform its intended function. IEC 60601-1, Clause 9.8.2 Procedure of Verification: IEC 60601-1, Clause 9.8.2 Release Protocol of Measures Limits of Acceptance for Verification: IEC 60601-1, Clause 9.8.2 Procedure of Validation: IEC 60601-1, Clause 9.8.2 Limits of Acceptance for Validation: IEC 60601-1, Clause 9.8.2 2013-08-22 Page 73 of 141

Measure Number: Measure Text: Measure Status: RC-011 Meet IEC 60601-1, Clause 7 requirements. Completed Responsible: Implementation: Lin, Eason Supervision: Chen, Ernest Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/6/2013 End: 7/5/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: For Ward Terminal to be well designed, its markings and ACCOMPANYING DOCUMENTS should be clear, consistent, and help to reduce potential use error. Additional markings are permitted, provided that they do not give rise to misunderstanding. Where symbols are used, they shall conform to ISO 7000 or IEC 60417 where appropriate symbols exist. Sufficient information shall be provided to the USER concerning any condition necessary to ensure that, when used as prescribed by the manufacturer, the equipment is unlikely to present a hazard within the meaning of IEC 60601-1. Functions Hazards Causes Information Labeling Incomplete instructions for use Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: Each specified marking was rubbed by hand, without undue pressure, first for 15 s with a cloth rag soaked with distilled water, then for 15 s with a cloth rag soaked in methylated spirit at ambient temperature, and then for 15 s with a cloth rag soaked with isopropyl alcohol. IEC 60601-1, 3rd Edition, Clause 7.1.3b 2013-08-22 Page 74 of 141

Limits of Acceptance for Procedure of Evidence: The markings worked and did not curl at the edges. The markings were clearly readable. IEC 60601-1, 3rd Edition, Clause 7.1.3b Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 7.1.3b Release Protocol of Measures Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 7.1.3b Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 7.1.3b Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 7.1.3b 2013-08-22 Page 75 of 141

Measure Number: Measure Text: Measure Status: RC-012 Meet IEC 60601-1, Clause 8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS requirements Completed Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/6/2013 End: 7/15/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Meet IEC 60601-1, Clause 8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS requirements The electrical isolation providing protection against electric shock shall be of such quality that currents flowing through it are limited to the values specified in IEC 60601-1, Clause 8.7.3. Functions Hazards Causes Connect to Mains Electromagnetic energy Too high leakage current, causing a death or serious injury to a patient. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 76 of 141

Procedure for Evidence of Efficency of Measures: Testing was conducted after thermal stabilization and was repeated Humidity Preconditioning. Following the humidity conditioning (see Humidity Conditioning Test for further details), the equipment was located in an environment with a temperature approximately equal to t where t is the temperature of the humidity cabinet and a relative humidity between 45% and 65% and was conducted 1 hour after the end of the humidity preconditioning treatment. In Normal Condition and in the specified Single Fault Conditions (IEC 60601-1, 3rd Ed., 8.7.2) With the Equipment energized in stand-by and fully operating and with any switch in the Mains Part in any position. With the equipment connected to a supply source with the highest rated supply frequency and 110% of the highest rated Mains Voltage as indicated in the RESULTS table. IEC 60601-1, 3rd Edition, Clause 8.7 Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: IEC 60601-1, 3rd Ed. Table 3 and Table 4 IEC 60601-1, 3rd Edition, Clause 8.7 Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 8.7 Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 8.7 Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 8.7 Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 8.7 2013-08-22 Page 77 of 141

Measure Number: Measure Text: Measure Status: RC-013 Meet IEC 60601-1, Clause 9.4.2 requirements. Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/9/2013 End: 6/27/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The equipment did not overbalance. PC or its parts shall not overbalance when placed in any transport position of NORMAL USE on a plane inclined at an angle of 10 from the horizontal plane. Functions Hazards Causes Enclosure Mechanical energy Falling Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: 10 Incline (9.4.2.1) - The equipment was placed in any normal use transport position on a plane inclined 10 from the horizontal. Prior to testing the equipment was prepared as noted in accordance with Clause 9.4.2.2. IEC 60601-1, 3rd Edition, Clause 9.4.2 Limits of Acceptance for Procedure of Evidence: The equipment did not overbalance. IEC 60601-1, 3rd Edition, Clause 9.4.2 2013-08-22 Page 78 of 141

Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 9.4.2 Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 9.4.2 Release Protocol of Measures Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 9.4.2 Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 9.4.2 2013-08-22 Page 79 of 141

Measure Number: Measure Text: Measure Status: RC-014 Meet IEC 60601-1-2 requirements Completed Responsible: Implementation: Yuan, Andy Supervision: Lin, Eason Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/10/2013 End: 6/19/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Meet IEC 60601-1-2 requirements Electromagnetic emission requirements are necessary for the protection of: - safety services (e.g. police, fire and ambulance communications); - other ME EQUIPMENT and ME SYSTEMS; - non-me EQUIPMENT (e.g. computers); - telecommunications (e.g. radio/tv, telephone, radionavigation). Functions Hazards Causes Connect to Mains Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Electromagnetic energy Electromagnetic compatibility of PC. Procedure for Evidence of Efficency of Measures: IEC 60601-1-2 EMC test report 2013-08-22 Page 80 of 141

Limits of Acceptance for Procedure of Evidence: EMC test report Procedure of Verification: EMC test report Release Protocol of Measures Limits of Acceptance for Verification: EMC test report Procedure of Validation: EMC test report Limits of Acceptance for Validation: EMC test report 2013-08-22 Page 81 of 141

Measure Number: Measure Text: Measure Status: RC-015 Meet IEC 60950-1, Annex NAD Acoustic tests requirements Completed Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/3/2013 End: 7/8/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The equipment having acoustic outputs at communication receivers and similar devices used for voice Telecommunication, regardless of transmission medium (e.g., TELECOMMUNICATION NETWORK, CABLE DISTRIBUTION NETWORK, wireless network), shall demonstrate that the equipment complies with the appropriate acoustics requirements related to short duration impulses and long duration disturbances as described in NAD.3 and NAD.4. of IEC 60950-1. These requirements apply to equipment that is intended to be connected directly or indirectly to a TELECOMMUNICATION NETWORK or CABLE DISTRIBUTION SYSTEM or other transmission medium, and that contains an earpiece or receiver that is held against or in the ear. The effect on human hearing of impulsive noise or of disturbances that are less than 0,5 s in duration shall be evaluated under NAD.3. The effect of longer disturbances, such as those that might be produced during tone-type dialing, shall be evaluated under NAD.4. Functions Hazards Causes Signal output Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Acoustic energy (sound) earphone 2013-08-22 Page 82 of 141

Procedure for Evidence of Efficency of Measures: The headset of the equipment was placed according to the general setup page and operating for the exchange of calls (such as talking state with the handset raised). A short duration peak applied to the tip and ring by applying digital codes to the receive input of the device. This may be performed by using an analog test set to drive a reference codec, or by use of a digital code generator. The analog level shall be set to switch between the maximum positive and the maximum negative values for the reference codec. The switching rate shall be through the range of 100 Hz to 3400 Hz for narrow band and 100 Hz to 6800 Hz for wideband. The duration of the ON codes shall be a number of complete cycles approximating but exceeding 500 ms. The ON codes must be followed by a quiet interval of at least 500 ms before repeating the codes. For equipment not intended to be directly connected to the Telecommunication Network, such as equipment used with specialized PBX or Key systems and the like, the short duration peak was applied between the tip and ring terminals of the system. The equipment was checked for self generated acoustic impulses such as those produced by operation of the hook switch or by dialing. IEC 60950-1, Annex NAD Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: The peak acoustic pressure did not exceed 136 db ERP IEC 60950-1, Annex NAD Procedure of Verification: IEC 60950-1, Annex NAD Limits of Acceptance for Verification: IEC 60950-1, Annex NAD Procedure of Validation: IEC 60950-1, Annex NAD Limits of Acceptance for Validation: IEC 60950-1, Annex NAD 2013-08-22 Page 83 of 141

Measure Number: Measure Text: Measure Status: RC-016 PC marked the name or trademark of Onyx with model no. Completed Responsible: Implementation: Lin, Eason Supervision: Chen, Ernest Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/8/2013 End: 6/21/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Marked the name or trademark and with model no. Identification Functions Hazards Causes Information Labeling PC did not marke with the name or trademark of Onyx and with a MODEL OR TYPE REFERENCE Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: Meet IEC 60601-1 3rd Ed. Clause 7.2.2 requirements IEC 60601-1 3rd Ed. Clause 7.2.2 Limits of Acceptance for Procedure of Evidence: Meet IEC 60601-1 3rd Ed. Clause 7.2.2 requirements IEC 60601-1 3rd Ed. Clause 7.2.2 2013-08-22 Page 84 of 141

Procedure of Verification: Meet IEC 60601-1 3rd Ed. Clause 7.2.2 requirements IEC 60601-1 3rd Ed. Clause 7.2.2 Limits of Acceptance for Verification: Meet IEC 60601-1 3rd Ed. Clause 7.2.2 requirements IEC 60601-1 3rd Ed. Clause 7.2.2 Release Protocol of Measures Procedure of Validation: Meet IEC 60601-1 3rd Ed. Clause 7.2.2 requirements IEC 60601-1 3rd Ed. Clause 7.2.2 Limits of Acceptance for Validation: Meet IEC 60601-1 3rd Ed. Clause 7.2.2 requirements IEC 60601-1 3rd Ed. Clause 7.2.2 2013-08-22 Page 85 of 141

Measure Number: Measure Text: Measure Status: RC-017 Provided a connector that has special housing to prevent a connection with the wrong polarity is made Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/7/2013 End: 6/25/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Provided a connector that has special housing to prevent a connection with the wrong polarity is made. No HAZARDOUS SITUATION, other than absence of ESSENTIAL PERFORMANCE, shall develop when a connection with the wrong polarity is made. Functions Hazards Causes Connection to an external d.c. Thermal Energy When a connection with the wrong power source polarity is made Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: QD2-001, QD2-009 and QD2-010 QQ1-001 Limits of Acceptance for Procedure of Evidence: QD2-001, QD2-009 and QD2-010 QQ1-001 2013-08-22 Page 86 of 141

Procedure of Verification: QD2-001, QD2-009 and QD2-010 QQ1-001 Limits of Acceptance for Verification: QD2-001, QD2-009 and QD2-010 QQ1-001 Release Protocol of Measures Procedure of Validation: QD2-001, QD2-009 and QD2-010 QQ1-001 Limits of Acceptance for Validation: QD2-001, QD2-009 and QD2-010 QQ1-001 2013-08-22 Page 87 of 141

Measure Number: Measure Text: Measure Status: RC-018 Rounding of sharp edges and corners. Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/13/2013 End: 6/25/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Rounding of sharp edge and corner. Rough surfaces, sharp corners and edges of Ward Terminal that could result in an unacceptable RISK. Functions Hazards Causes Enclosure Mechanical energy Sharp edges and corners Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: Accessible edges or corners of the unit, except those required for proper functioning of the equipment, were subjected to this test. Surfaces described below were judged using the Sharp Edge Tester. The curved face of the tester head was covered with three layers of tape: bottom layer 1.6 mm black vinyl foam tape; middle layer 0.8 mm white vinyl foam tape; outer layer 0.08 mm (3 mil) Teflon tape skived. The tape-covered head of the tester was positioned on the edge to check for sharpness. The arm of the tester was calibrated so that a 6.7 N force was present at the center of the head when the arm was between stops. This force was applied on the sharp edge. The tester was immediately moved 50.5 mm along the edge and then back to its starting position without removing the tester so that total distance traveled was 101 mm. The time of travel was no longer than 5 seconds nor less than 2 seconds. IEC 60601-1, Clause 9.3 2013-08-22 Page 88 of 141

Limits of Acceptance for Procedure of Evidence: The edges and corners were considered to be rounded and smooth. IEC 60601-1, Clause 9.3 Procedure of Verification: IEC 60601-1, Clause 9.3 Release Protocol of Measures Limits of Acceptance for Verification: IEC 60601-1, Clause 9.3 Procedure of Validation: IEC 60601-1, Clause 9.3 Limits of Acceptance for Validation: IEC 60601-1, Clause 9.3 2013-08-22 Page 89 of 141

Measure Number: Measure Text: Measure Status: RC-019 Said components were secured in place on pcb by UL Recognized nonchemical bonding compound. Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/24/2013 End: 6/28/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Said components were secured in place on pcb by UL Recognized non-chemical bonding compound. Components of power supply, unwanted movement of which could result in an unacceptable RISK, shall be mounted securely to prevent such movement. Functions Hazards Causes Connect to Mains Electromagnetic energy Components of PC, unwanted movement of which could result in an unacceptableb RISK, is not mounted securely to prevent such movement. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: Having used the unjointed probe, which was in the shape of the standard test finger and which incorporated a force gauge, a steady force of 10 N for internal parts was applied for a period of 5 seconds. QD2-001, QD2-010, QM2-011, QD2-009 2013-08-22 Page 90 of 141

Limits of Acceptance for Procedure of Evidence: During the application of the test force, clearances behind earthed or unearthed conductive accessible parts were not reduced to a level that would result in an energy hazard. QD2-001, QD2-010, QM2-011, QD2-009 Procedure of Verification: QD2-001, QD2-010, QM2-011, QD2-009 Release Protocol of Measures Limits of Acceptance for Verification: QD2-001, QD2-010, QM2-011, QD2-009 Procedure of Validation: QD2-001, QD2-010, QM2-011, QD2-009 Limits of Acceptance for Validation: QD2-001, QD2-010, QM2-011, QD2-009 2013-08-22 Page 91 of 141

Measure Number: Measure Text: Measure Status: RC-020 Shall compliance with the following tests: resistance to moisture, etc. (see IEC 60601-1, Clause 11.6); dielectric strength (see IEC 60601-1, Clause 8.8.3); and mechanical strength (see IEC 60601-1, Clause 15.3). Completed Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Release Protocol of Measures Planning: Begin: 5/1/2013 End: 8/31/2013 Actual Values: Begin: 5/17/2013 End: 7/8/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Shall compliance with the following tests: resistance to moisture, etc. (see IEC 60601-1, clause 11.6); dielectric strength (see IEC 60601-1, clause 8.8.3); and mechanical strength (see IEC 60601-1, clause 15.3). The resistance to heat shall retained by all types of insulation, including insulating partition walls, during the EXPECTED SERVICE LIFE of the PC. Functions Hazards Causes Connect to Mains Electromagnetic energy The resistance to heat is not retained by all types of insulation, including insulating partition walls, during the EXPECTED SERVICE LIFE of the PC. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 92 of 141

Procedure for Evidence of Efficency of Measures: A sample of each external part of insulating material indicated in the table was tested using the test apparatus shown in Fig. 21 of IEC 60601-1, 3rd Edition. (In this apparatus, a steel ball of 5 mm diameter exerts a force of 20 N when pressed against a surface). The apparatus and a fire (or sand) brick were placed in an air-circulating oven. Once the apparatus and brick attained the oven temperature, the sample was placed in the oven on top of the brick. The steel ball was then positioned such that the apparatus was balanced on the surface of the sample. After 1 h, it was removed and the diameter of the impression made by the ball was measured. The specified oven temperature was the higher of the following: a. the sum of the specified ambient and the temperature rise of relevant part during the Temperature Test; AND b. 75 C ± 2 C (test per Clause 8.8.4.1a), OR 125 C ± 2 C (test per Clause 8.8.4.1b) IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: The results were less 2 mm. Cracks were not visible to the naked eye. IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 8.8.4.1 Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 8.8.4.1 2013-08-22 Page 93 of 141

Measure Number: Measure Text: Measure Status: RC-021 Supplied by SELV power adapter and DC/AC inverter meet LCC requirements. Completed Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/5/2013 End: 7/22/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Supplied by SELV power adapter and DC/AC inverter meet LCC requirements ACCESSIBLE PARTS (output connector) shall not exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. Functions Hazards Causes Connect to Mains Electromagnetic energy ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: A short of the operational or basic insulation indicated in the table was made. The voltage present at the accessible parts/circuitry was measured, recorded and compared to 42.4 V pk ac and 60 V dc. QD2-001, QD2-010, QM2-011 and QD2-009 2013-08-22 Page 94 of 141

Limits of Acceptance for Procedure of Evidence: The voltage did not exceed 42.4 V peak ac or 60 V dc. QD2-001, QD2-010, QM2-011 and QD2-009 Procedure of Verification: QD2-001, QD2-010, QM2-011 and QD2-009 Release Protocol of Measures Limits of Acceptance for Verification: QD2-001, QD2-010, QM2-011 and QD2-009 Procedure of Validation: QD2-001, QD2-010, QM2-011 and QD2-009 Limits of Acceptance for Validation: QD2-001, QD2-010, QM2-011 and QD2-009 2013-08-22 Page 95 of 141

Measure Number: Measure Text: Measure Status: RC-022 The equipment complied with the requirements of IEC 60601-1, Clause 11.6. Completed Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/7/2013 End: 7/11/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The equipment complied ]with the requirements of IEC 60601-1, Clause 11.6. The monitor and their parts, including ACCESSORIES, shall be capable of withstanding, without damage or deterioration of safety provisions, the cleaning or disinfection PROCESSES specified in the instructions for use. Functions Hazards Causes Connect to Mains Electromagnetic energy Cleaning Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: The equipment was cleaned/disinfected as specified in the instructions for use. After an adequate cooling and drying period, the unit was subjected to Leakage Current and Dielectric Voltage Withstand Tests as noted below. The effect of multiple cleanings/disinfections over the expected service life of the equipment was considered. IEC 60601-1, 3rd Edition, Clause 11.6) 2013-08-22 Page 96 of 141

Limits of Acceptance for Procedure of Evidence: The equipment complied with the requirements of IEC 60601-1, Clause 11.6. The equipment did not show signs of wetting of uninsulated live parts. There were no dielectric breakdown and/or leakage current test non-compliances. IEC 60601-1, 3rd Edition, Clause 11.6) Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 11.6) Release Protocol of Measures Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 11.6) Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 11.6) Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 11.6) 2013-08-22 Page 97 of 141

Measure Number: Measure Text: Measure Status: RC-023 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. Completed Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/3/2013 End: 6/28/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Meet clause 13 of IEC 60601-1, 3rd Ed. requirements. HAZARDOUS SITUATIONS and fault conditions. Functions Hazards Causes Connect to Mains Electromagnetic energy Short circuit of any one insulation that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.8 Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 98 of 141

Procedure for Evidence of Efficency of Measures: The device was operated until ultimate results were obtained. All final results were recorded. If required, final temperatures were measured and recorded. The normal conditions were also applied in the least favorable combination. - The unit was positioned in the same way as for the Temperature Test - Failure of Components - The components were short circuited or opened, one at a time. - The load current was then increased by 5 % with the supply voltage maintained at its original value. When steady thermal conditions were again established, the load was again increased. These steps were continued until the overload protector operated or until thermal stability was reached. - For each fault condition, the Dielectric Voltage Withstand Test was repeated immediately following the conclusion of the test. IEC 60601-1, 3rd Edition, Clause 13 Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: 1. Self-resetting thermal cutouts did not operate. 2. The setting of any thermostat, thermal cutout or overload release did not change due to heating, vibration or other causes. 3. The measured temperatures did not exceed those allowable. 4. In each case the hazardous situations described in Clauses 13.1.2 to 13.1.4 of IEC 60601-1, 3rd ED. did not occur. 5. See Table 11 of IEC 60601-1, 3rd ED. for Temperatures obtained during the indicated Abnormal Operation tests. IEC 60601-1, 3rd Edition, Clause 13 Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 13 Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 13 Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 13 Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 13 2013-08-22 Page 99 of 141

Measure Number: Measure Text: Measure Status: RC-024 The humidity preconditioning treatment is performed in a humidity cabinet containing air with a relative humidity of 93 % ± 3 % @ 25 C 48 hr Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/8/2013 End: 6/26/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The humidity preconditioning treatment is performed in a humidity cabinet containing air with a relative humidity of 93 % ± 3 % @ 25 C, 48 hr Storage (humidity preconditioning treatment) Functions Hazards Causes Connect to Mains Electromagnetic energy Storage (humidity preconditioning treatment) Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 100 of 141

Procedure for Evidence of Efficency of Measures: PC or its parts shall be set up complete. Covers used during transport and storage are detached. This treatment is applied only to those power supply parts which are influenced by the climatic conditions that are simulated by the test. Parts that can be detached without the use of a TOOL are detached but are treated simultaneously with the major part. ACCESS COVERS that can be opened or detached without the use of a TOOL are opened and detached. A humidity chamber having relative humidity of 90 to 96 % was set to any convenient Temperature T between 20 and 32 C. The unit, before being placed in the chamber was brought to a temperature between T and T + 4 C and kept at this temperature a minimum of 4 h. Then the sample was placed in the humidity chamber and kept there for the time specified in Table 5.7 of IEC 60601-1, 3rd Ed. (minimum 48 h). During this time the temperature of the air in the chamber was maintained at T ± 2 C. Immediately after the treatment and while still in the chamber, the sample was subjected to Dielectric Voltage Withstand Test. Then the unit was removed from the chamber, placed in normal environment (temperature approximately T, humidity 45-65%). Immediately following the humidity preconditioning treatment, the power supply was subjected to a repeat of the dielectric withstand and leakage current tests. IEC 60601-1, 3rd Edition, Clause 5.7 Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: There was no dielectric breakdown IEC 60601-1, 3rd Edition, Clause 5.7 Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 5.7 Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 5.7 Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 5.7 Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 5.7 2013-08-22 Page 101 of 141

Measure Number: Measure Text: Measure Status: RC-025 The intensity of visible light from monitor can be controlled by the user. The highest intensity achievable cannot produce detrimental biological effects. Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Release Protocol of Measures Planning: Begin: 5/1/2013 End: 8/31/2013 Actual Values: Begin: 5/7/2013 End: 7/4/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Computer monitors emit electromagnetic radiation in a broad range of frequencies. Although the electromagnetic radiation field intensities are measurable, they are not substantially above background office levels. The effect on health from the use of monitors has been studied for many years, and the research continues. The present consensus of experts is that there are no data to suggest a health risk from exposure to the electromagnetic fields associated with the use of monitors. PC users have expressed concerns about the possible health effects from the electromagnetic radiation that monitors produce. The concerns generally center around adverse pregnancy outcome (spontaneous abortion or birth defects). It has been postulated that these effects are due to electromagnetic radiation exposure from monitors. The biological effects of exposure to VLF and ELF fields similar to those produced by monitors have been studied for many years, and the research continues today. Cancer and reproductive outcomes are the topics studied most intensively. Current available scientific information does not identify a health risk from exposure to the electromagnetic fields associated with the use of computer monitors. Functions Hazards Causes Display Radiation energy Visible Light 2013-08-22 Page 102 of 141

Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: https://ehs.mit.edu/site/content/radiation-computer-monitors https://ehs.mit.edu/site/content/radiation-computermonitors Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: Procedure of Verification: Limits of Acceptance for Verification: Procedure of Validation: Limits of Acceptance for Validation: 2013-08-22 Page 103 of 141

Measure Number: Measure Text: Measure Status: RC-026 The internal wirings were secured in place on PCB by UL R/C nonchemical bonding component. Completed Responsible: Implementation: Chen, Ernest Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/8/2013 End: 6/25/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The internal wirings were secured in place on PCB by UL R/C non-chemical bonding component. Internal cables and wiring shall adequately protected against contact from friction at sharp corners and edges where damage to insulation could result in a HAZARDOUS SITUATION. Functions Hazards Causes Connect to Mains Electromagnetic energy Internal cables and wiring were not adequately protected against contact from friction at sharp corners and edges where damage to insulation could result in a HAZARDOUS SITUATION. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: Having used the unjointed probe, which was in the shape of the standard test finger and which incorporated a force gauge, a steady force of 10 N for internal parts was applied for a period of 5 seconds. QD2-001, QD2-010, QM2-011 and QD2-009 2013-08-22 Page 104 of 141

Limits of Acceptance for Procedure of Evidence: During the application of the test force, clearances behind earthed or unearthed conductive accessable parts were not reduced to a level that would result in an energy hazard. QD2-001, QD2-010, QM2-011 and QD2-009 Procedure of Verification: QD2-001, QD2-010, QM2-011 and QD2-009 Release Protocol of Measures Limits of Acceptance for Verification: QD2-001, QD2-010, QM2-011 and QD2-009 Procedure of Validation: QD2-001, QD2-010, QM2-011 and QD2-009 Limits of Acceptance for Validation: QD2-001, QD2-010, QM2-011 and QD2-009 2013-08-22 Page 105 of 141

Measure Number: Measure Text: Measure Status: RC-027 The limits specified in IEC 60601-1, Clause 8.4 is not be exceeded for ACCESSIBLE PARTS Completed Responsible: Implementation: Yuan, Andy Supervision: Lin, Eason Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/8/2013 End: 7/9/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The insulation that is relied upon as a MEANS OF PROTECTION, including REINFORCED INSULATION shall be subject to testing. Insulation Functions Hazards Causes Connect to Mains Electromagnetic energy Short circuit of any or all insulation that does not comply with the requirements of IEC 60601-1, 3rd Ed., Clause. 8.8 Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: A short of the operational or basic insulation indicated in the table was made. The voltage present at the accessible parts/circuitry was measured, recorded and compared to 42.4 V pk ac and 60 V dc. QD2-001, QD2-010, QM2-011, QD2-009 IEC 60601-1, 3rd Edition, Clause 8.4.2 and 8.11.1 2013-08-22 Page 106 of 141

Limits of Acceptance for Procedure of Evidence: The voltage did not exceed 42.4 V peak ac or 60 V dc. QD2-001, QD2-010, QM2-011, QD2-009 IEC 60601-1, 3rd Edition, Clause 8.4.2 and 8.11.1 Procedure of Verification: QD2-001, QD2-010, QM2-011, QD2-009 IEC 60601-1, 3rd Edition, Clause 8.4.2 and 8.11.1 Release Protocol of Measures Limits of Acceptance for Verification: QD2-001, QD2-010, QM2-011, QD2-009 IEC 60601-1, 3rd Edition, Clause 8.4.2 and 8.11.1 Procedure of Validation: QD2-001, QD2-010, QM2-011, QD2-009 IEC 60601-1, 3rd Edition, Clause 8.4.2 and 8.11.1 Limits of Acceptance for Validation: QD2-001, QD2-010, QM2-011, QD2-009 IEC 60601-1, 3rd Edition, Clause 8.4.2 and 8.11.1 2013-08-22 Page 107 of 141

Measure Number: Measure Text: Measure Status: RC-028 The permissible environmental conditions for transport and storage were marked on the outside of the packaging. Completed Responsible: Implementation: Chen, Ernest Supervision: Lin, Eason Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/8/2013 End: 7/18/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The permissible environmental conditions for transport and storage were marked on the outside of the packaging Packing Functions Hazards Causes Connect to Mains Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Electromagnetic energy Packing Procedure for Evidence of Efficency of Measures: QD2-001, QM2-007 QQ1-001 Limits of Acceptance for Procedure of Evidence: QD2-001, QM2-007 QQ1-001 2013-08-22 Page 108 of 141

Procedure of Verification: QD2-001, QM2-007 QQ1-001 Limits of Acceptance for Verification: QD2-001, QM2-007 QQ1-001 Release Protocol of Measures Procedure of Validation: QD2-001, QM2-007 QQ1-001 Limits of Acceptance for Validation: QD2-001, QM2-007 QQ1-001 2013-08-22 Page 109 of 141

Measure Number: Measure Text: Measure Status: RC-029 The strength of packing carton / fixed frame. The transportation stimulation for the verification. All connections were mechanically secure. Completed Responsible: Implementation: Chen, Ernest Supervision: Lin, Eason Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/9/2013 End: 7/15/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: The strength of packing carton / fixed frame. The transportation stimulation for the verification. All connections were mechanically secure. Transport Functions Hazards Causes Connect to Mains Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Electromagnetic energy Transport Procedure for Evidence of Efficency of Measures: QD2-001 and QM2-007 QQ1-001 Limits of Acceptance for Procedure of Evidence: QD2-001 and QM2-007 QQ1-001 2013-08-22 Page 110 of 141

Procedure of Verification: QD2-001 and QM2-007 QQ1-001 Limits of Acceptance for Verification: QD2-001 and QM2-007 QQ1-001 Release Protocol of Measures Procedure of Validation: QD2-001 and QM2-007 QQ1-001 Limits of Acceptance for Validation: QD2-001 and QM2-007 QQ1-001 2013-08-22 Page 111 of 141

Measure Number: Measure Text: Measure Status: RC-030 The suitable enclosure shall be provided. Completed Responsible: Implementation: Chen, Ernest Supervision: Lin, Eason Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/2/2013 End: 7/2/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Compliance is checked by the mechanical strength tests for ENCLOSURES (see 15.3). ENCLOSURES shall have the strength and rigidity necessary to avoid a fire that could occur as a result of a total or partial collapse caused by reasonably foreseeable misuse. Functions Hazards Causes Connect to Mains Electromagnetic energy Internal parts, other than contacts of plugs, connectors and socket-outlets, that can be touched by the test pin shown in IEC 60601-1, Figure 8. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 112 of 141

Procedure for Evidence of Efficency of Measures: - Force (15.3.2) - A sample of the unit was subjected to inward directed force of 250 N plus or minus 10 N for a period of 5 seconds applied to different parts of the enclosure via a test tool providing contact over a circular plane surface 30 mm in diameter as indicated in Table 15.3. Note: not applied to the bottom of a product having a mass of greater than 18 kg. - Impact (15.3.3) - The impact resistance of an enclosure was tested by application of an impact conveyed via a solid steel 50 mm (approx) ball of mass 500 g plus or minus 25 g, fall freely from 1,3 meter height. The unit was rigidly supported and the impact was applied to the locations indicated in Table 15.3. IEC 60601-1, 3rd Edition, Clause 15.3 Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: There was no cracking of the enclosure. There were no live parts that became accessible IEC 60601-1, 3rd Edition, Clause 15.3 Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 15.3 Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 15.3 Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 15.3 Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 15.3 2013-08-22 Page 113 of 141

Measure Number: Measure Text: Measure Status: RC-031 They shall comply with the applicable safety requirements of a relevant IEC, UL or ISO standard; Completed Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/8/2013 End: 7/12/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: They shall comply with the applicable safety requirements of a relevant IEC, UL or ISO standard; The reliability of components that are used as MEANS OF PROTECTION shall be assessed for the conditions of use in the monitor and power supply. Functions Hazards Causes Connect to Mains Electromagnetic energy The reliability of components that are used as MEANS OF PROTECTION are not assessed for the conditions of use in the PC. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy 2013-08-22 Page 114 of 141

Procedure for Evidence of Efficency of Measures: Evaluation and testing of components shall be conducted as follows: - a component that has been demonstrated to comply with a standard harmonized with the relevant IEC component standard shall be checked for correct application and use in accordance with its rating. It shall be subjected to the applicable tests of IEC 60601-1 as part of the equipment with the exception of those tests which are part of the relevant IEC component standard; - a component that has not been demonstrated to comply with a relevant standard as above shall be checked for correct application and use in accordance with its specified rating. It shall be subjected to the applicable tests of IEC 60601-1, as part of the equipment, and to the applicable tests of the component standard, under the conditions occurring in the equipment; - where no relevant IEC component standard exists, or where components are used in circuits not in accordance with their specified ratings, the components shall be tested under the conditions occurring in the equipment. The number of samples required for test is, in general, the same as required by an quivalent standard. QD2-001, QD2-010, QM2-011, QD2-009 and IEC 60601-1, Clause 4.8 Release Protocol of Measures Limits of Acceptance for Procedure of Evidence: QD2-001, QD2-010, QM2-011, QD2-009 and IEC 60601-1, Clause 4.8 Procedure of Verification: QD2-001, QD2-010, QM2-011, QD2-009 and IEC 60601-1, Clause 4.8 Limits of Acceptance for Verification: QD2-001, QD2-010, QM2-011, QD2-009 and IEC 60601-1, Clause 4.8 Procedure of Validation: QD2-001, QD2-010, QM2-011, QD2-009 and IEC 60601-1, Clause 4.8 Limits of Acceptance for Validation: QD2-001, QD2-010, QM2-011, QD2-009 and IEC 60601-1, Clause 4.8 2013-08-22 Page 115 of 141

Measure Number: Measure Text: Measure Status: RC-032 When PC is operated in worst-case NORMAL USE including the maximum ambient operating temperature, the parts shall not reach temperatures exceeding the values given in IEC 60601-1, Table 22 and Table 23. Completed Responsible: Implementation: Yuan, Andy Supervision: Chen, Ernest Release Protocol of Measures Planning: Begin: 5/1/2013 End: 8/31/2013 Actual Values: Begin: 5/7/2013 End: 7/10/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: When PC and power supply is operated in worst-case NORMAL USE including the maximum ambient operating temperature, the parts shall not reach temperatures exceeding the values given in IEC 60601-1, Table 22 and Table 23. Excessive temperatures in PC and/or power supply. Functions Hazards Causes Connect to Mains Thermal Energy Excessive temperatures in PC and/or power supply. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: The sample was placed in a test corner. The test corner consisted of two walls at right angles, and a floor, all of dull black painted plywood of 20 mm thickness. The linear dimensions of the test corner were at least 115 percent of the linear dimensions of the unit. The unit was positioned as near to the walls as possible and to the floor. The power supply is operated until THERMAL STABILITY is reached. IEC 60601-1, 3rd Edition, Clause 11 2013-08-22 Page 116 of 141

Limits of Acceptance for Procedure of Evidence: IEC 60601-1, 3rd Edition, Clause 11 Procedure of Verification: IEC 60601-1, 3rd Edition, Clause 11 Release Protocol of Measures Limits of Acceptance for Verification: IEC 60601-1, 3rd Edition, Clause 11 Procedure of Validation: IEC 60601-1, 3rd Edition, Clause 11 Limits of Acceptance for Validation: IEC 60601-1, 3rd Edition, Clause 11 2013-08-22 Page 117 of 141

Measure Number: Measure Text: Measure Status: RC-033 Where there is no relevant IEC or ISO standard, the requirements of IEC 60601-1, 3rd Ed. were applied. Completed Responsible: Implementation: Lin, Mac Supervision: Yuan, Andy Planning: Begin: 5/1/2013 End: 8/31/2013 Release Protocol of Measures Actual Values: Begin: 5/6/2013 End: 6/27/2013 Effort: planned: Is: Validation: Validated: Validated by: Chen, James Validated on: 7/19/2013 Release: Released: Released by: Lin, Mac Release on: 7/19/2013 Decided Individual Measures: Comment: Where there is no relevant IEC or ISO standard, the requirements of IEC 60601-1 3rd Ed. were applied. The reliability of components that are used as MEANS OF PROTECTION are not assessed for the conditions of use in the PC. Functions Hazards Causes Connect to Mains Electromagnetic energy The reliability of components that are used as MEANS OF PROTECTION are not assessed for the conditions of use in the PC. Implementation Team: Chen, James Lin, Mac Lin, Eason Chen, Ernest Yuan, Andy Procedure for Evidence of Efficency of Measures: QD2-001, QD2-010, QM2-011 and QD2-009 IEC 60601-1, Clause 4.8 2013-08-22 Page 118 of 141

Limits of Acceptance for Procedure of Evidence: QD2-001, QD2-010, QM2-011 and QD2-009 IEC 60601-1, Clause 4.8 Procedure of Verification: QD2-001, QD2-010, QM2-011 and QD2-009 IEC 60601-1, Clause 4.8 Release Protocol of Measures Limits of Acceptance for Verification: QD2-001, QD2-010, QM2-011 and QD2-009 IEC 60601-1, Clause 4.8 Procedure of Validation: QD2-001, QD2-010, QM2-011 and QD2-009 IEC 60601-1, Clause 4.8 Limits of Acceptance for Validation: QD2-001, QD2-010, QM2-011 and QD2-009 IEC 60601-1, Clause 4.8 2013-08-22 Page 119 of 141

Evaluation After Measures Project: 1307033002, ONYX-2121DT Version: 1.00 Function: F-01 Connect to Mains Evaluation After Measures Hazard Cause Severity Probability Risk Risk-Benefit- Analysis Comment 2013-08-22 Page 120 of 141

Hazard Cause H-01 Electromagnetic energy C-01 Short circuit of any or all insulation that does not comply with the requirements of IEC 60601-1, 3rd Ed., Clause. 8.8 C-02 Short circuit of any or all CREEPAGE DISTANCES or AIR CLEARANCES that do not comply with the requirements of IEC 60601-1, 3rd Ed., Clause 8.9 Severity Probability Risk Serious Remote Acceptable Serious Remote Acceptable Risk-Benefit- Analysis Comment Evaluation After Measures C-03 Open circuit of any or all earth connections that do not comply with the requirements of IEC 60601-1, Clause 8.6, including any functional earth connection. Serious Remote Acceptable C-04 Short circuit of any one insulation that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.8 Critical Improbable Acceptable C-05 Short circuit of any one CREEPAGE DISTANCE or AIR CLEARANCE that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.9 Critical Improbable Acceptable C-06 Open circuit of any one PROTECTIVE EARTH CONDUCTOR or internal PROTECTIVE EARTH CONNECTION that complies with the requirements of IEC 60601-1, Clause 8.6 Critical Improbable Acceptable C-07 All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION. Serious Improbable Acceptable 2013-08-22 Page 121 of 141

Hazard Cause C-08 The reliability of components that are used as MEANS OF PROTECTION are not assessed for the conditions of use in the PC. C-09 Storage (humidity preconditioning treatment) Severity Probability Risk Critical Improbable Acceptable Serious Improbable Acceptable C-10 Transport Serious Remote Acceptable C-11 Components of PC, unwanted movement of which could result in an unacceptableb RISK, is not mounted securely to prevent such movement. Serious Remote Acceptable Risk-Benefit- Analysis Comment Evaluation After Measures C-12 Too high leakage current, causing a death or serious injury to a patient. Critical Improbable Acceptable C-13 The LEAKAGE CURRENTS from, to or between ACCESSIBLE PARTS other than PATIENT CONNECTIONS Critical Improbable Acceptable C-14 ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. Critical Improbable Acceptable C-15 Internal parts, other than contacts of plugs, connectors and socketoutlets, that can be touched by the test pin shown in IEC 60601-1, Figure 8. Serious Improbable Acceptable C-16 Accidental detachment of conductors and connectors where breaking free could lead to a HAZARDOUS SITUATION Serious Improbable Acceptable 2013-08-22 Page 122 of 141

Hazard Cause C-17 The resistance to heat is not retained by all types of insulation, including insulating partition walls, during the EXPECTED SERVICE LIFE of the PC. C-18 Internal cables and wiring is not adequately protected against contact with a moving part or from friction at sharp corners and edges where damage to insulation. Severity Probability Risk Serious Remote Acceptable Serious Remote Acceptable Risk-Benefit- Analysis Comment Evaluation After Measures C-19 Electromagnetic compatibility of PC. Critical Improbable Acceptable C-20 Packing Serious Improbable Acceptable C-21 Wirings and connectors of Panel PC Monitor supply is not secured or insulated that accidental detachment can result in a HAZARDOUS SITUATION. Serious Improbable Acceptable C-22 Internal cables and wiring were not adequately protected against contact from friction at sharp corners and edges where damage to insulation could result in a HAZARDOUS SITUATION. Serious Remote Acceptable C-23 Cleaning Minor Improbable Acceptable 2013-08-22 Page 123 of 141

Hazard Cause H-02 Thermal Energy C-01 Excessive temperatures in PC and/or power supply. C-02 ENCLOSURES have no the strength and rigidity necessary to avoid a fire. Severity Function: F-02 Enclosure Probability Risk Serious Remote Acceptable Critical Improbable Acceptable Risk-Benefit- Analysis Comment Evaluation After Measures Hazard Cause Severity Probability Risk H-01 Mechanical energy C-01 Falling Serious Remote Acceptable C-02 Suspended masses Serious Remote Acceptable C-03 Sharp edges and corners Minor Remote Acceptable H-02 Thermal Energy C-01 Excessive temperatures in enclosure Serious Remote Acceptable Risk-Benefit- Analysis Comment Function: F-03 Rear Bracket Hazard Cause Severity Probability Risk H-01 Mechanical energy C-01 Crack Serious Remote Acceptable C-02 Suspended masses Serious Remote Acceptable C-03 Instability Serious Remote Acceptable Risk-Benefit- Analysis Comment 2013-08-22 Page 124 of 141

Function: F-04 Signal input Hazard Cause H-01 Electromagnetic energy C-01 ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. Severity Probability Risk Critical Improbable Acceptable Risk-Benefit- Analysis Comment Evaluation After Measures Function: F-05 Signal output Hazard Cause Severity Probability Risk H-01 Acoustic energy (sound) C-01 earphone Serious Remote Acceptable Risk-Benefit- Analysis Comment Function: F-06 User control and indicator Hazard Cause H-01 Operational C-01 Switches used to control power to PC or its parts, including switches, did not have their "on" and "off" positions Severity Probability Risk Serious Remote Acceptable Risk-Benefit- Analysis Comment Function: F-07 Information Hazard Cause Severity Probability Risk Risk-Benefit- Analysis Comment 2013-08-22 Page 125 of 141

Hazard Cause H-01 Labeling C-01 PC did not marke with the name or trademark of Onyx and with a MODEL OR TYPE REFERENCE C-03 Inadequate description of performance characteristics C-04 Inadequate specification of intended use. C-05 Inadequate disclosure of limitations H-02 Operating instructions C-01 Inadequate specification of accessories to be used with the medical device. C-02 Inadequate specification of preuse checks. C-03 Over-complicated operating instructions H-03 Specification of service and maintenance Severity Probability Risk Minor Remote Acceptable C-02 Incomplete instructions for use Critical Improbable Acceptable Serious Remote Acceptable Critical Improbable Acceptable Critical Improbable Acceptable Serious Improbable Acceptable Minor Remote Acceptable Minor Remote Acceptable C-01 Installation instruction Serious Remote Acceptable Risk-Benefit- Analysis Comment Evaluation After Measures Function: F-08 Connection to an external d.c. power source Hazard Cause H-01 Thermal Energy C-01 When a connection with the wrong polarity is made Severity Probability Risk Serious Improbable Acceptable Risk-Benefit- Analysis Comment Function: F-09 Display Hazard Cause Severity Probability Risk Risk-Benefit- Analysis Comment 2013-08-22 Page 126 of 141

Hazard Cause Severity Probability Risk H-01 Radiation energy C-01 Visible Light Minor Remote Acceptable Risk-Benefit- Analysis Function: F-10 BIOS (Basic Input/Output System - real time clock) Hazard Cause Severity Probability Risk H-01 Mechanical energy C-01 Lithium battery explosion Critical Improbable Acceptable Risk-Benefit- Analysis Comment Comment Evaluation After Measures 2013-08-22 Page 127 of 141

Risk Graph After Measures Project: 1307033002, ONYX-2121DT Version: 1.00 Both Conditions Operating Condition Risk Graph After Measures Failure Mode Acceptable ALARP Inacceptable 2013-08-22 Page 128 of 141

Risk Graph Before and After Measures Project: 1307033002, ONYX-2121DT Version: 1.00 Before Measures Both Conditions After Measures Both Conditions Risk Graph Before and After Measures Operating Condition Operating Condition Acceptable ALARP Inacceptable 2013-08-22 Page 129 of 141

Failure Mode Failure Mode Risk Graph Before and After Measures Acceptable ALARP Inacceptable 2013-08-22 Page 130 of 141

Evaluation of Residual Risk Project: 1307033002, ONYX-2121DT Version: 1.00 Identification of Known or Predictable Hazards and Evaluation of the Risks: 1. All reasonable hazards and their cause were identified and the resulting risks have been evaluated. All of the data have been documented in the enclosed tables before the required measures for the risk diminution have been defined and a final evaluation has been carried out. 2. The aim of this analysis is to improve the security of the product. The assessment is based on the companies' security policy which was implemented for the product in the definition of the risk graph. There areas were defined: "broadly acceptable" and "intolerable". Evaluation of Residual Risk 3. The definite acceptance of the product can only be carried out when the evaluation are located mainly in the "broadly acceptable" region. 4. The product is not marketable according to the safety philosophy of our company if it does contain evaluations in the "intolerable" region after implementation of the measures. Remaining Residual Risk, Summarizing Evaluation In this risk analysis the "present state" of the device has been evaluated. Having carried out the evaluation of the product, the risk of appearance of a not acceptable fault is quite low. The product construction, its function as well as the application does - under normal conditions- not represent any risk for the user. 2013-08-22 Page 131 of 141

Post production Project: 1307033002, ONYX-2121DT Version: 1.00 Post production 2013-08-22 Page 132 of 141

Overview of the Results of the Risk Management Project: 1307033002, ONYX-2121DT Version: 1.00 Function: F-01 Connect to Mains Hazard H-01 Electromagnetic energy Cause C-01 Short circuit of any or all insulation that does not comply with the requirements of IEC 60601-1, 3rd Ed., Clause. 8.8 C-02 Short circuit of any or all CREEPAGE DISTANCES or AIR CLEARANCES that do not comply with the requirements of IEC 60601-1, 3rd Ed., Clause 8.9 C-03 Open circuit of any or all earth connections that do not comply with the requirements of IEC 60601-1, Clause 8.6, including any functional earth connection. C-04 Short circuit of any one insulation that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Risk Before Effect on Measures S3 / P3 Inacceptable S3 / P3 Inacceptable S3 / P3 Inacceptable S4 / P3 Inacceptable Measure Implemense Measures Relea- Risk After M-01 The limits specified in IEC Yes Yes S3 / P2 60601-1, Clause 8.4 is not be Acceptable exceeded for ACCESSIBLE PARTS M-01 The limits specified in IEC 60601-1, Clause 8.4 is not be exceeded for ACCESSIBLE PARTS Yes Yes M-01 The limits specified in IEC 60601-1, Clause 8.4 is not be exceeded for ACCESSIBLE PARTS Yes Yes M-01 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. Yes Yes S3 / P2 Acceptable S3 / P2 Acceptable S4 / P1 Acceptable Overview of the Results of the Risk Management 2013-08-22 Page 133 of 141

Hazard Cause Clause 8.8 C-05 Short circuit of any one CREEPAGE DISTANCE or AIR CLEARANCE that complies with the requirements for one MEANS OF PROTECTION as specified in IEC 60601-1, Clause 8.9 C-06 Open circuit of any one PROTECTIVE EARTH CONDUCTOR or internal PROTECTIVE EARTH CONNECTION that complies with the requirements of IEC 60601-1, Clause 8.6 C-07 All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION. C-08 The reliability of components that are used as MEANS OF PROTECTION are not assessed for the conditions of use in the PC. Effect on Risk Before Measures S4 / P2 Inacceptable S4 / P3 Inacceptable S3 / P2 Acceptable S4 / P2 Inacceptable Measure Release Implement M-01 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. Yes Yes M-01 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. Yes Yes M-01 All components, including wiring Yes Yes shall be used in accordance with their specified ratings. M-02 All components, including wiring, the failure of which could result in a HAZARDOUS SITUATION are fixed in place by UL R/C nonchemical bonding compound. Yes Yes M-01 They shall comply with the Yes Yes applicable safety requirements of a relevant IEC, UL or ISO standard; M-02 Where there is no relevant IEC or Yes Yes ISO standard, the requirements of IEC 60601-1, 3rd Ed. were applied. Risk After Measures S4 / P1 Acceptable S4 / P1 Acceptable S3 / P1 Acceptable S4 / P1 Acceptable Overview of the Results of the Risk Management 2013-08-22 Page 134 of 141

Hazard Cause C-09 Storage (humidity preconditioning treatment) Risk Before Effect on Measures S3 / P3 Inacceptable C-10 Transport S3 / P2 Acceptable C-11 Components of PC, unwanted movement of which could result in an unacceptableb RISK, is not mounted securely to prevent such movement. C-12 Too high leakage current, causing a death or serious injury to a patient. C-13 The LEAKAGE CURRENTS from, to or between ACCESSIBLE PARTS other than PATIENT CONNECTIONS C-14 ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. C-15 Internal parts, other than contacts of plugs, connectors and socketoutlets, that can be touched by the test pin S3 / P2 Acceptable S4 / P3 Inacceptable S4 / P3 Inacceptable S4 / P3 Inacceptable S3 / P2 Acceptable Measure Implemense Measures Relea- Risk After M-01 The humidity preconditioning Yes Yes S3 / P1 treatment is performed in a Acceptable humidity cabinet containing air with a relative humidity of 93 % ± 3 % @ 25 C 48 hr M-01 The strength of packing carton / Yes Yes S3 / P2 fixed frame. The transportation Acceptable stimulation for the verification. All connections were mechanically secure. M-01 Said components were secured in place on pcb by UL Recognized non-chemical bonding compound. Yes Yes M-01 Meet IEC 60601-1, Clause 8.7 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS requirements Yes Yes M-01 Meet IEC 60601-1, Clause 8.7 Yes Yes LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS requirements M-01 Supplied by SELV power adapter Yes Yes and DC/AC inverter meet LCC requirements. M-01 The suitable enclosure shall be Yes Yes provided. S3 / P2 Acceptable S4 / P1 Acceptable S4 / P1 Acceptable S4 / P1 Acceptable S3 / P1 Acceptable Overview of the Results of the Risk Management 2013-08-22 Page 135 of 141

Hazard Cause shown in IEC 60601-1, Figure 8. C-16 Accidental detachment of conductors and connectors where breaking free could lead to a HAZARDOUS SITUATION C-17 The resistance to heat is not retained by all types of insulation, including insulating partition walls, during the EXPECTED SERVICE LIFE of the PC. C-18 Internal cables and wiring is not adequately protected against contact with a moving part or from friction at sharp corners and edges where damage to insulation. C-19 Electromagnetic compatibility of PC. Effect on Risk Before Measures S3 / P3 Inacceptable S3 / P3 Inacceptable S3 / P3 Inacceptable S4 / P3 Inacceptable C-20 Packing S3 / P3 Inacceptable Measure Release Implement M-01 The HAZARDOUS SITUATIONS in IEC 60601-1, Clause 13.1.2 to 13.1.4 (inclusive) do not occur in the PC and power adapter. Yes Yes M-01 Shall compliance with the Yes Yes following tests: resistance to moisture, etc. (see IEC 60601-1, Clause 11.6); dielectric strength (see IEC 60601-1, Clause 8.8.3); and mechanical strength (see IEC 60601-1, Clause 15.3). M-02 Ball pressure test - according to IEC 60601-1, Clause 8.8.4.1, b) requirements Yes Yes M-01 Internal cables and wiring is fixed Yes Yes in place by cable tie to protect against contact with a moving part or from friction at sharp corners and edges where damage to insulation M-01 Meet IEC 60601-1-2 Yes Yes requirements M-01 The permissible environmental conditions for transport and storage were marked on the outside of the packaging. Yes Yes Risk After Measures S3 / P1 Acceptable S3 / P2 Acceptable S3 / P2 Acceptable S4 / P1 Acceptable S3 / P1 Acceptable Overview of the Results of the Risk Management 2013-08-22 Page 136 of 141

Hazard Cause C-21 Wirings and connectors of Panel PC Monitor supply is not secured or insulated that accidental detachment can result in a HAZARDOUS SITUATION. C-22 Internal cables and wiring were not adequately protected against contact from friction at sharp corners and edges where damage to insulation could result in a HAZARDOUS SITUATION. Risk Before Effect on Measures S3 / P2 Acceptable S3 / P2 Acceptable C-23 Cleaning S2 / P2 Acceptable H-02 Thermal Energy C-01 Excessive temperatures in PC and/or power supply. C-02 ENCLOSURES have no the strength and rigidity necessary to avoid a fire. S3 / P3 Inacceptable S4 / P2 Inacceptable Measure Implemense Measures Relea- Risk After M-01 All wirings are tied together by Yes Yes S3 / P1 cable tie to prevent accidental Acceptable detachment M-01 The internal wirings were secured Yes Yes in place on PCB by UL R/C nonchemical bonding component. M-01 The equipment complied with the requirements of IEC 60601-1, Clause 11.6. Yes Yes M-01 When PC is operated in worstcase Yes Yes NORMAL USE including the maximum ambient operating temperature, the parts shall not reach temperatures exceeding the values given in IEC 60601-1, Table 22 and Table 23. M-01 Compliance with the mechanical Yes Yes strength tests for ENCLOSURES (see IEC 60601-1, Clause 15.3) S3 / P2 Acceptable S2 / P1 Acceptable S3 / P2 Acceptable S4 / P1 Acceptable Overview of the Results of the Risk Management Function: F-02 Enclosure Hazard Cause Effect on Risk Before Measures H-01 Mechanical energy C-01 Falling S3 / P3 Inacceptable Measure Implemense Measures Relea- Risk After M-01 Meet IEC 60601-1, Clause 9.4.2 Yes Yes S3 / P2 requirements. Acceptable 2013-08-22 Page 137 of 141

Hazard Cause Effect on Risk Before Measures C-02 Suspended masses S3 / P3 Inacceptable C-03 Sharp edges and corners S2 / P4 Inacceptable H-02 Thermal Energy C-01 Excessive temperatures in enclosure S3 / P3 Inacceptable Measure Release Implement M-01 Meet IEC 60601-1 Clause 9.8.2 Yes Yes and Table 21 requirements. M-01 Rounding of sharp edges and Yes Yes corners. M-01 When PC is operated in worstcase NORMAL USE including the maximum ambient operating temperature, the parts shall not reach temperatures exceeding the values given in IEC 60601-1, Table 22 and Table 23. Yes Yes Risk After Measures S3 / P2 Acceptable S2 / P2 Acceptable S3 / P2 Acceptable Overview of the Results of the Risk Management Function: F-03 Rear Bracket Hazard Cause Effect on Risk Before Measures H-01 Mechanical energy C-01 Crack S3 / P3 Inacceptable C-02 Suspended masses S3 / P2 Acceptable Measure Implemense Measures Relea- Risk After M-01 Meet IEC 60601-1 Clause 9.8.2 Yes Yes S3 / P2 and Table 21 requirements. Acceptable M-01 Meet IEC 60601-1 Clause 9.8.2 Yes Yes and Table 21 requirements. S3 / P2 Acceptable C-03 Instability S3 / P3 Inacceptable Function: F-04 Signal input M-01 Meet IEC 60601-1, Clause 9.4.2 Yes Yes requirements. S3 / P2 Acceptable Hazard Cause Effect on Risk Before Measures Measure Release Implement Risk After Measures 2013-08-22 Page 138 of 141

Hazard H-01 Electromagnetic energy Cause C-01 ACCESSIBLE PARTS exceed 42.4 V peak a.c. or 60 V d.c. in NORMAL CONDITION or in SINGLE FAULT CONDITION. Risk Before Effect on Measures S4 / P2 Inacceptable Function: F-05 Signal output Hazard H-01 Acoustic energy (sound) Cause Effect on Risk Before Measures C-01 earphone S3 / P3 Inacceptable Measure Implemense Measures Relea- Risk After M-01 Supplied by SELV power adapter Yes Yes S4 / P1 and DC/AC inverter meet LCC Acceptable requirements. Measure Implemense Measures Relea- Risk After M-01 Meet IEC 60950-1, Annex NAD Yes Yes S3 / P2 Acoustic tests requirements Acceptable Overview of the Results of the Risk Management Function: F-06 User control and indicator Hazard Cause H-01 Operational C-01 Switches used to control power to PC or its parts, including switches, did not have their "on" and "off" positions Effect on Risk Before Measures S3 / P3 Inacceptable Measure Implemense Measures Relea- Risk After M-01 Indicated by an adjacent indicator Yes Yes S3 / P2 light Acceptable M-02 Marked with symbols IEC 60417 Yes Yes Function: F-07 Information Hazard Cause H-01 Labeling C-01 PC did not marke with the name or trademark of Onyx and with a MODEL OR TYPE REFERENCE C-02 Incomplete instructions for use Effect on Risk Before Measures S2 / P3 Acceptable S4 / P3 Inacceptable Measure Implemense Measures Relea- Risk After M-01 PC marked the name or Yes Yes S2 / P2 trademark of Onyx with model no. Acceptable M-01 Meet IEC 60601-1, Clause 7 Yes Yes requirements. S4 / P1 Acceptable 2013-08-22 Page 139 of 141

Hazard H-02 Operating instructions Cause C-03 Inadequate description of performance characteristics C-04 Inadequate specification of intended use. C-05 Inadequate disclosure of limitations C-01 Inadequate specification of accessories to be used with the medical device. C-02 Inadequate specification of pre-use checks. Effect on Risk Before Measures S3 / P3 Inacceptable S4 / P3 Inacceptable S4 / P3 Inacceptable S3 / P2 Acceptable S2 / P3 Acceptable Measure Implemense Measures Relea- Risk After M-01 Meet IEC 60601-1, Clause 7 Yes Yes S3 / P2 requirements. Acceptable M-01 Meet IEC 60601-1, Clause 7 Yes Yes requirements. M-01 Meet IEC 60601-1, Clause 7 Yes Yes requirements. M-01 Meet IEC 60601-1, Clause 7 Yes Yes requirements. M-01 Meet IEC 60601-1, Clause 7 Yes Yes requirements. S4 / P1 Acceptable S4 / P1 Acceptable S3 / P1 Acceptable S2 / P2 Acceptable Overview of the Results of the Risk Management C-03 Over-complicated operating instructions S2 / P3 Acceptable M-01 Meet IEC 60601-1, Clause 7 Yes Yes requirements. S2 / P2 Acceptable H-03 Specification of service and maintenance C-01 Installation instruction S3 / P3 Inacceptable M-01 Meet IEC 60601-1, Clause 7 Yes Yes requirements. S3 / P2 Acceptable Function: F-08 Connection to an external d.c. power source Hazard Cause H-01 Thermal Energy C-01 When a connection with the wrong polarity is made Effect on Risk Before Measures S3 / P3 Inacceptable Measure Implemense Measures Relea- Risk After M-01 Provided a connector that has Yes Yes S3 / P1 special housing to prevent a Acceptable connection with the wrong polarity is made Function: F-09 Display Hazard Cause Effect on Risk Before Measures Measure Release Implement Risk After Measures 2013-08-22 Page 140 of 141

Hazard Cause Effect on Risk Before Measures H-01 Radiation energy C-01 Visible Light S2 / P3 Acceptable Measure Implemense Measures Relea- Risk After M-01 The intensity of visible light from Yes Yes S2 / P2 monitor can be controlled by the Acceptable user. The highest intensity achievable cannot produce detrimental biological effects. Function: F-10 BIOS (Basic Input/Output System - real time clock) Hazard Cause Effect on Risk Before Measures H-01 Mechanical energy C-01 Lithium battery explosion S4 / P3 Inacceptable Measure Implemense Measures Relea- Risk After M-01 Battery circuits is designed so Yes Yes S4 / P1 that discharging at a rate Acceptable exceeding the battery manufacturer s recommendations, and unintentional charging, are prevented. Overview of the Results of the Risk Management 2013-08-22 Page 141 of 141