AMD Analysis & Technology AG

Similar documents
How To Build An Amd Quas 3 Ar Mass Spectrometer

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Pesticide Analysis by Mass Spectrometry

Electrospray Ion Trap Mass Spectrometry. Introduction

Background Information

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Expectations for GC-MS Lab

Micromass LCT User s Guide

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Analyzing Small Molecules by EI and GC-MS. July 2014

Appendix 5 Overview of requirements in English

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Analysis of Liquid Samples on the Agilent GC-MS

Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

How To Use Gc-Ms

LC-MS/MS for Chromatographers

13C NMR Spectroscopy

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# RFQ)

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

SCION SQ Series GC-MS

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

23 The Thermal Conductivity Detector

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report

A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Purification of reaction mixtures using flash chromatography.

Hardware Manual. AB SCIEX API 4000 LC/MS/MS System

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Waters Core Chromatography Training (2 Days)

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Correlation of the Mass Spectrometric Analysis of Heat-Treated Glutaraldehyde Preparations to Their 235nm / 280 nm UV Absorbance Ratio

CHE334 Identification of an Unknown Compound By NMR/IR/MS

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

F321 THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important are... in the nucleus of an atom

Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis

Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?

Setting up a Quantitative Analysis MS ChemStation

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA

Application of Structure-Based LC/MS Database Management for Forensic Analysis

amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis

An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

Elemental analyzer multi EA 5000

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

m/z

Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

New drugs of abuse? A case study on Phenazepam & Methoxetamine

Analysis of Vegetable Oils by High Performance Liquid Chromatography

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Xcalibur Versions

New Jersey Department of Environmental Protection Division of Remediation Management Response Site Remediation Program

Nuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass

Mass Spectrometry Signal Calibration for Protein Quantitation

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

Discovery of Pesticide Protomers Using Routine Ion Mobility Screening

Signal, Noise, and Detection Limits in Mass Spectrometry

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES

Tuning & Mass Calibration

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Guide to achieving reliable quantitative LC-MS measurements

aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS

Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS

AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY. Andrew Tipler

How To Use An Ionsonic Microscope

Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software

TABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ACKNOWLEDGEMENT ABSTRACT LIST OF TABLES LIST OF FIGURES LIST OF SYMBOLS LIST OF APPENDICES

QGA Quantitative Gas Analyser

All the prepared formulations were subjected for following. evaluation parameters and obtained results were showed in Tables 6.3 &

How To Test For Contamination In Large Volume Water

ß-CYCLODEXTRIN SYNONYMS

Overview. Purpose. Methods. Results

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Mass Spectrometry. Overview

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software

Transcription:

AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds (VOCs) by mass spectrometry (MS) is usually performed by Electron impact (EI) or classical Chemical Ionization techniques (CI). In these cases the matrix gas is introduced by a heated capillary interface and the matrix gas amount used for MS analysis is in the order of 0.3-1 ml/min. Special technologies as proton transfer reaction mass spectrometry (PTR-MS) and related selected ion-flow-tube mass spectrometry (SIFT-MS) play a role, too. For mass spectrometric analysis of polar and non volatile molecules in liquid matrices the established method of choice are API-MS techniques with dedicated interfaces. We have recently reported (Application Note 120329) about a specific ESI-MS system, suitable for the direct analysis of low atomic number elements. In addition we were interested to evaluate the capabilities of this system for the direct analysis of volatile organic compounds in ambient air by the application of Atmospheric Pressure Corona Discharge Ionization (APCDI). The ambient air acts as reagent gas in this case, therefore the term APCI may be used. No GC separation and purge and trap techniques have been applied. B) Summary Analyses have been performed with the experimental version of a new bench-top double focusing mass spectrometer. The dedicated and miniaturized system is based on the original AMD QuAS 3 AR Technology. Emphasis has been placed to the analysis of volatile organic compounds in the gas phase, logically producing mass spectra of low molecular weight compounds. Within the scope of the reported semi-quantification methods no official certified procedures were applied, since the aim here was to achieve orientating results for possible future applications. The existing API Interface useable for ESI-MS or APCI-MS and the mass analyzer are integrated with a multi-stage turbo pumping system. It has been demonstrated again that the special features of a small double focusing mass spectrometer regarding peak shape, mass resolution and abundance sensitivity are of significant advantage for the qualitative and quantitative analysis of ions in the low mass range. Semi-Quantitative Trace Analyses of a number of VOCs (typically used as solvents) in ambient air were performed, exemplarily. Limits of detection (LOD) in the upper ppb range were achieved for a number of these compounds in full scan mode. SIM techniques which would improve sensitivity and extend the LOD to the low ppb range have not been applied because specificity was considered to be more important in this case. The results depend on the ionization efficiency and the production of quasi-molecular ions, respectively on the low number of key fragments which is typical for Chemical Ionization. The influence of scan methodologies including raw data accumulation techniques on the quality of the results and analysis time are also described. The achieved results of APCI-MS applications on VOCs in ambient air indicate the usefulness of the method for trace analysis in gaseous matrices. This may be of interest for environmental or process analysis in a wide range of concentrations. Dependent on the complexity of the gaseous matrix the method may be applied without chromatographic separation (GC). In this context it is of specificity advantage that either mainly quasi-molecular ions or only a few intense key fragment ions are formed. AMD Handelsregister HRB 141560 Amtsgericht Oldenburg USt-ID Nr. DE238516825 Steuer-Nr. 2357/201/12096 Zoll-Nr.: 6602150

Page 2, Application Note 120419 C) Methodology C.1 Mass spectrometer The mass analyzer was a miniaturized dedicated version of the AMD QuAS 3 AR (Quasar070206.pdf) double focusing system, equipped with an API Interface in APCI mode (APCI-MS system). System details are not reported here since an evaluation model was used for the experiments. Miniaturized AMD QuAS 3 AR Magnetic Sector Analyzer Electric Sector 25 cm Array Detector (not yet implemented) Entrance slit 48 cm Fig. 1 Scheme of the ion optics and physical size of the bench-top double focusing MS analyzer Magnetic sector analyzer The miniaturized AMD QuAS 3 AR is based on the original AMD QuAS 3 AR Technology as a multifunctional high performance system and previously described in a system description: SD040617_ AMD Quasar.pdf. The new current miniaturized version is downsized significantly, dedicated to low mass resolution and API-Interfacing only. Multi stage turbo pumping system. However, it incorporates the significant advantages of magnetic sector mass spectrometers for low mass ions regarding resolution and abundance sensitivity Fig. 2 CAD Design of the Miniaturized AMD QuAS 3 AR API-MS system

Page 3, Application Note 120419 C.2 API-MS Interface As a dedicated API-MS (APCI or ESI) bench-top system the interface and the mass analyzer are mounted on a multistage turbo pumping system. The system is fully optimized for highest transmission of low mass ions through the interface to the MS detector and in so far relevant for the applications reported here, specifically dedicated to trace analysis of VOCs in ambient air. ESI Parts The API ion source of the miniaturized AMD QuAS 3 AR dedicated for APCI (ESI)-MS applications consists of an API room including nebulizer assisted sprayer and hot air pipes for desolvation assistance in ESI mode. which are not used for the APCI experiments described here. Corona needle A corona discharge needle is mounted in front of the counter electrode orifice. The multi-stage vacuum interface is integrated with the analyzer multi-stage turbo pumping system. Fig. 3 APCI (ESI)-MS interface front view of the miniaturized API-MS Evaluation Model C.3 Measurement procedure for VOC analysis (example: Ethyl acetate) in ambient air Analysis Details C.3 Air exchange time Ion chromatogram of key fragment (Decline Curve) C 2 H 5 O 2 (C 4 H 8 O 2 +H) + Sample: Ethyl acetate Syringe injection in API room (2 l ambient air volume) at 40 0 C air temperature Ionization: Corona discharge U c +3000 V after injection and evaporation (approximate calculation): c = 9 x10-4 g / 2.5 g = 3.6 10-4 c = 360 ppm : 0.4 l/min from API room Fig. 4 APCI-MS corona discharge mass spectrum of Ethyl acetate in ambient air

Page 4, Application Note 120419 Peak width:12 msec Ion chromatogram of key fragment at mass 61 (Decline Curve) Peak shape at resolution 70 (10% valley) Analysis Details C.3 cont d Scan method: magnet scan, Mass resolution: 70 at 10% valley with fixed slits Scan Speed: 1 sec/dec Measuring time per peak: 12 msec Data acquisition: single scan or raw data accumulation Detector: post acceleration and Channeltron type SEM Data System: AMD Version 12.01.1 Analysis time: Data were recorded over the a time of about 1400 sec until the sample concentration in the API room was below detection limit (see below) Fig.5 APCI-MS peak display of corona discharge mass spectrum of Ethyl acetate in ambient air Analysis Details C.3 cont d Limit of Detection (LOD) MeOH Ion chromatogram of quasi molecular ion of methanol (Decline Curve) Ion chromatogram of key fragment of Ethyl acetate at mass 61 (Decline Curve) Ethyl acetate Sample: Ethyl acetate diluted in Methanol (1:100) Sample amount: 10 nl in 1 µl Methanol Injection in API room as before Ionization: Corona discharge after injection and evaporation (approximate calculation): c = 9 x10-6 g / 2.5 g = 3.6 10-6 c = 3.6 ppm : Fig. 6 APCI-MS corona discharge mass spectrum (single scan) of Ethyl acetate (diluted in Methanol) in ambient air. Methanol yield is about 4 times less than for Ethyl acetate

Page 5, Application Note 120419 background sample Ion chromatogram of key fragment of Ethyl acetate at mass 61 (Decline Curve) Analysis Details C.3 cont d Limit of Detection (LOD) Sample: Ethyl acetate diluted in Methanol 1:500 Sample amount: 2 nl in 1 µl Methanol Ionization: after injection and evaporation (approximate calculation): c = 1.8 x10-6 g /2.5 g = 0.72 10-6 c = 720 ppb Fig. 7 APCI-MS display of key fragment ion of corona discharge mass spectrum of Ethyl acetate in ambient air at limit of detection C.3.1 Short-form summary of the applied methodology and the result for APCI-MS analysis of Ethyl acetate (Fig. 4-7) in ambient air The compound has been used as an example to demonstrate the measurement procedures for APCI-MS trace analysis of a volatile organic compound in ambient air. The rudimentary method used here should give some orientating results for the application of a possible analytical procedure under more sophisticated sample preparation and sample introduction methodology. The counter electrode/nozzle part of the API interface was kept at a temperature of 150 0 C resulting in an ambient air temperature within the API room of about 40 0 C. The volatile compound was injected by a syringe directly into the air of the API room. As described in figures 4 and 5 above an amount of 1 µl Ethyl acetate was injected and evaporated immediately. The API interface air intake of 0.4 l/min diluted the evaporated sample in the ambient air of the API room (2 l volume) by air continuous exchange from outside. The immediate sample concentration was calculated (see fig. 4) by the ratio of evaporated sample quantity to the air quantity in the API room (temperature corrected). According to the decline function about 40% of the sample at higher concentrations may be adsorbed on the inner walls and is desorbing, slowly. In the low ppm or even ppb range a single scan may be sufficient for identification of the compound (see fig 6) in the air. The sensitivity of the method allows detection limits in the low ppb range, if SIM techniques (longer measuring times per ion, higher number of ions but short total analysis time) will be used. However, the individual background signals are the limiting factors. Routine applications will require a controlled introduction (gas pumping) of the air matrix containing the analyte gas sample into the heated counter electrode/nozzle zone via a heated inlet line. The use of external heated volumes for the sample and the clean flushing gas as well as corresponding switch valves will be mandatory for efficient and precise analyses.

Page 6, Application Note 120419 D. APCI-MS trace analysis of Acetonitril (ACN) in ambient air Analysis Details D. background sample Ion chromatogram quasi molecular ion of Acetonitril at mass 42 (Decline Curve) (M+H) + Limit of Detection (LOD) Sample: Acetonitril diluted in Methanol 1:1000 Sample amount: 1 nl in 1 µl Methanol Ionization: (definition ) c = 9 x10-7 g/2.5g = 3.6 10-7 c = 360 ppb Fig. 8 APCI-MS corona discharge mass spectrum (single scan) of Acetonitril (diluted in Methanol) in ambient air. Analysis Details D. cont d Ion chromatogram of mass 42 (background signal) Noise level 3 mv Ion chromatogram quasi molecular ion of Acetonitril at mass 42 (incl. background) Limit of Detection (LOD) Sample: Acetonitril diluted in Methanol 1:1000 Sample amount: 1 nl in 1 µl Methanol Ionization: (definition ) c = 9 x10-7 g/2.5g = 3.6 10-7 c = 360 ppb Fig. 9 Ion chromatograms of Quasi-molecular ion and background signal (S/N ratio = 3:1) of Acetonitril in ambient air at limit of detection (LOD 180 ppb).

Page 7, Application Note 120419 E. APCI-MS trace analysis of Acetone in ambient air Analysis Details E. Ion chromatogram quasi molecular ion of Acetone at mass 59 (Decline Curve) (M+H) + Peak display quasi molecular ion Sample: Acetone Ionization: (definition ) c = 8 x10-4 g/2.5g = 3.2 10-4 c = 320 ppm Dynamic range: see below Fig. 10 APCI-MS corona discharge mass spectrum (single scan) of Acetone in ambient air. Analysis Details E. cont d background signal Sample: Acetone Ionization: (definition ) c = 8 x10-4 g/2.5g = 3.2 10-4 c = 320 ppm Dynamic range: 1:200 (2 mv peak detectable) LOD: about 3 ppm (high background signal) Fig. 11 APCI-MS corona discharge mass spectrum (single scan) of Acetone in ambient air including background signals at quasi molecular ion

Page 8, Application Note 120419 F. APCI-MS trace analysis of Methanol in ambient air Analysis Details F. Ion chromatogram of key fragment at mass 15 (Decline Curve) Ion chromatogram of quasi-molecular ion at mass 33 (Decline Curve) (M+H) + Sample: Methanol Ionization: (definition ) c = 8 x10-4 g/2.5g = 3.2 10-4 c = 320 ppm Dynamic range: 1:100 LOD estimated: < 3 ppm Fig. 12 APCI-MS corona discharge mass spectra of Methanol in ambient air G. APCI-MS trace analysis of Gasoline in ambient air Analysis Details G. Ion chromatogram of key fragment at mass 57 (Decline Curve) single spectrum accumulated spectra Sample: Gasoline Ionization: (definition ) c = 8 x10-4 g/2.5g = 3.2 10-4 c = 320 ppm Dynamic range: 1:200 LOD: Not to be defined for one component Fig.13 APCI-MS corona discharge mass spectra of Gasoline in ambient air. The spectra indicate the mixture of major acyclic hydrocarbons. The method may be useable in this or similar cases for fingerprint analyses but requires further investigations

Page 9, Application Note 120419 H. APCI-MS trace analysis of volatile halogen compounds in ambient air using negative ion detection H.1 APCI-MS trace analysis of Trifluor acedic acid in ambient air Analysis Details H.1 F - CF 3 - (M-H) - no background signals at key sample ion positions Sample:Trifluor acedic acid Ionization:Corona discharge U c - 3000 V for negative ion production (definition ) c = 1.5 x10-3 g/2.5g = 6 10-4 c = 600 ppm LOD: 100 ppm (estimated) Fig. 14 APCI-MS corona discharge negative ion mass spectra of Trifluor acedic acid in ambient air H.2 APCI-MS trace analysis of Dichlormethane in ambient air Analysis Details H.2 Cl 35 - Cl 37 - (M-H) - Sample: Dichlormethane Ionization:Corona discharge U c - 3000 V for negative ion production (definition ) c = 1.3 x10-3 g/2.5g = 5 10-4 c = 500 ppm LOD: 150 ppm (estimated) Fig. 15 APCI-MS corona discharge negative ion mass spectra of Dichlormethane in ambient air.

Page 10, Application Note 120419 H.3 APCI-MS trace analysis of Chloroform in ambient air Chlorine Isotopes accumulated background spectra accumulated sample spectra Analysis Details H.3 Sample: Chloroform Ionization:Corona discharge U c - 3000 V for negative ion production (definition ) c = 1.5 x10-3 g/2.5g = 6 10-4 c = 600 ppm Fig. 16 APCI-MS corona discharge negative ion mass spectra of Chloroform in ambient air. Analysis Details H.3 cont d Cl 35 - Cl 37 - Sample: Chloroform Ionization:Corona discharge U c - 3000 V for negative ion production (definition ) c = 1.5 x10-3 g/2.5g = 6 10-4 c = 600 ppm LOD: 100 ppm (estimated) Fig. 17 APCI-MS detailed view on Chlorine Isotopes of corona discharge negative ion mass spectrum of Chloroform in ambient air.

Page 11, Application Note 120419 H. 4 Short-form summary of APCI-MS trace analysis of volatile halogen compounds in ambient air using negative ion detection The results for demonstration of the applicability of negative ion recording to volatile halogen compounds were achieved under the same rudimentary method as mentioned before for the VOC analyses in positive ion recording mode. Further investigations for clarification of improved limits of detection were beyond the scope of this application note. While for negative ion detection the background is not the limiting factor, the absolute sensitivity achieved here for the method requires further investigations regarding possible improvements. Otherwise the remarks for routine applications mentioned under C.3.1 above are applicable, too.