Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System
|
|
|
- Elfrieda Bradford
- 9 years ago
- Views:
Transcription
1 Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System Xiang He, 1 Xiaohong Chen, 1 Adrian M. Taylor 2 and Hua-fen Liu 1 1 AB SCIEX, Redwood City, California, USA; 2 AB SCIEX, Concord, Ontario, Canada Overview The AB SCIEX TripleTOF 4600 liquid chromatography tandem mass spectrometry (LC/MS/MS) system is a hybrid quadrupole/ time-of-flight (QqTOF) instrument with rapid speed in acquisition of highly sensitive full scan MS and MS/MS spectra with high resolution and mass accuracy. An integrated multiplex system [MPX -2 SP High Throughput System] has been specifically designed to increase the throughput for LC/MS/MS forensic drug screening analysis. The combination of the MPX -2 SP High Throughput System and TripleTOF 4600 LC/MS/MS System meets the needs of a high throughput forensic screening workflow. The workflow allows qualitative screening and confirmation in a single automated run with highest confidence based on accurate mass product ion specificity, mass spectral library matching as well as mass error, LC retention time and isotopic ratio pattern. The workflow also automates report generation and facilitates subsequent quantitation when needed. Introduction Synthetic cannabinoids (K2) are a major health problem and a rapid, sensitive screening test for K2/metabolites is needed. Immunoassays are available, but lack flexibility for adding new analytes. LC/MS/MS offers increased sensitivity and specificity for identifying K2/metabolites and ability to add newly emerging analytes in a forensic screening panel with ease. TripleTOF 4600 high resolution accurate mass systems are capable of full MS range data collection and simultaneous Information Dependent Acquisition (IDA) product ion scan confirmation, providing an ideal platform for screening specified targets and also unregulated new analogues. Recent developments in LC multiplexing further improves throughput and makes LC/MS/MS more effective for screening applications. PeakView software with the MasterView add-in was used for targeted data processing. The MasterView manages large lists of compounds and performs extracted ion chromatogram (XIC) AB SCIEX TripleTOF 4600 System calculations, both targeted and non-targeted peak finding operations and library searching. It allows the ability to review results based upon retention times, accurate mass, isotopic pattern and MS/MS library searching. The goal here is to demonstrate a fast and reliable LC/MS/MS based high throughput screening (HTS) method for K2/metabolites on the AB SCIEX TripleTOF 4600 mass spectrometer and MPX -2 High Throughput System. Specifically, we aim for optimal sensitivity and specificity, and achieve injection-to-injection cycle time around 2 minutes. An integrated multiplex system has been specifically designed to synchronize two LC systems and a mass spectrometer, allowing injection of samples into two LC streams in parallel (Figure 1). The overlapping LC runs and efficient use of MS detection, achievable with the MPX -2 High Throughput System, is shown p 1
2 inject to result in an overall higher throughput system for LC-MS/MS analyses. path, and system state for both LC streams) in the Status Pane (Figure 3). idle acquire from stream 1 acquire from stream 2 wait for sync signal to stream N send sync signal Minimizes MS idle time during LC column equilibration Improves LC/MS throughput Figure 1. Multiplex LC system Designed to Support Two Parallel LC Streams into a Single MS Experimental Conditions Figure 2. MPX -2 High Throughput System Configuration and Module Arrangement Used in this Application. Sample Preparation Urine calibrator samples were prepared by spiking K2/metabolites standards into blank urine at various concentrations. Urine calibrator and urine samples were both spiked with internal standards with optional hydrolysis. Urine samples were diluted 5-fold or 10-fold in 50% MeOH/H2O. The mixtures were centrifuged before being transferred to autosampler vials for LC/MS/MS injection. Multiplex Software Configuration and Operation An integrated multiplex LC/MS-MS System was used, consisting of an AB SCIEX TripleTOF 4600 LC/MS/MS system, two Shimadzu Prominence XR LC systems, a pump containing a four solvent selection valve for sample loading and 3 switching valves for flow path control (Figure 2). The two chromatographic channels were not independent as they share a single high pressure loading pump which provides additional flexibility for injection and wash solvent composition. All hardware modules are controlled by Analyst TF 1.6 software with MPX TF Driver 1.0 add-on. Precise timing of the switching valves allowed each LC stream to perform interleaved injection and LC gradient elution. MPX Driver provides an easy-to-use interface for the user to modify the system configuration in the Settings Pane and to create or update a multiplex LC method in the Method Pane. The information entered in both Settings and Method panes allows the software to calculate the throughput and have it displayed. The interface also allows monitoring the real-time acquisition and system status (including pressure, flow rate, temperature, flow Figure 3. MPX TF Driver 1.0 interface with exemplary LC configuration and method. For this fast screening method a Restek Ultra Biphenyl (50 x 2.1 mm, 5 µm) was used, with a 3.5 minute gradient of water and methanol with 5 mm NH 4FA and a 1.5 minute acquisition window (Figure 3). Multiplex data acquisition for a batch of samples is automated by the MPX TF Driver 1.0. For the particular LC method used in multiplexed mode, compared to single-stream data acquisition, two-fold increase in throughput was achieved and injection-toinjection cycle time was 2.09 min (Figure 4). In comparison, current sample throughput of regular immunoassay screening for synthetic cannabinoids was roughly 2 minutes. p 2
3 Injection to Injection Times Single Stream 4 min 9 sec MPX-2 2 min 5 sec Figure 4. Increasing throughput by 2 times. Demonstration of injection-to-injection cycle time reduction (from 4.15 to 2.09 min) using both LC streams in an interleaved fashion with MPX TF Driver 1.0 interface. Results and Discussion This workflow adopts a simple dilute-and-shoot sample preparation procedure. It is known that some K2 metabolites can be converted to their phase-2 metabolites such as glucuronide or sulfate conjugates, therefore, enzyme hydrolysis can be optionally performed to release the parent analytes/phase-1 metabolites. With hydrolysis, sample preparation will be extended but data processing will be simplified as fewer analytes need to be monitored. Further, after hydrolysis, all unconjugated analytes share similar hydrophobicity which shortens analytes elution window and improves the throughput of the workflow. Figure 5 shows the extracted ion chromatograms of all analytes that elute during a very short 1.5 minute data window. Mass Spectrometric Conditions For mass calibration of the AB SCIEX TripleTOF 4600 system, an on-column injection of an AB SCIEX APCI calibration solution via the LC autosampler was performed. This automatic calibration workflow was based on a multiple-mass mixture and specific retention time was assigned to each mass covering a wide mass range for both TOF-MS and TOF-MS/MS modes calibration. The duration of each calibration was 3 min and the frequency of this calibration can be user-defined (e.g. once in several hours or longer). The AB SCIEX TripleTOF 4600 system was operated with DuoSpray source and Electrospray (ESI) probe. The data acquisition method consisted of a TOF-MS full scan and several IDA-TOF-MS/MS product ion scans using 35 V Collision Energy (CE) with 15 V (Collision Energy Spread) CES used for each candidate ion in each data acquisition cycle. An inclusion list consisting of >40 pairs of m/z and retention times was used for IDA-TOF-MS/MS product ion scans. 150 msec TOF-MS and 35 msec TOF-MS/MS accumulation time were used. This workflow allows quantitation (using TOF-MS primarily) and confirmation (using IDA-TOF-MS/MS) in the single run. Further, the TOF-MS data allows for retrospective data analysis on a later date, making it possible for data interrogation on the compounds that were not targeted in the first round. The flexibility in acquiring IDA-TOF-MS/MS product ion scan without an inclusion list makes it possible to acquire complementary confirmative MS/MS data in a non-targeted fashion, ideal for added analytes at later times. Figure 5. Extracted ion chromatograms of all analytes that elute during a 1.5-min data window. Using MasterView for Targeted Qualitative Screening and Identification MasterView add-in is an integral part of PeakView. It is the main software for viewing, analyzing, processing, and finally reporting the screening results. An XIC list needs to be constructed to extract ions from the data, and subsequently data is analyzed using multiple processing settings such intensity and signal-to-noise threshold; and confidence settings such as mass error, retention time, isotope matching, library searching and formula finding (Figure 6). Acquired MS/MS spectra were searched against an accurate mass forensic toxicology LC/MS/MS library. p 3
4 Figure 7. Report file size comparison between reports generated using various number of confidence points. Figure 6. PeakView software with MasterView ; the top pane shows an XIC table loaded for target drug screening and identification Figure 6 shows all targeted analytes that were identified from a spiked urine sample (25 ng/ml) using information of the isotopic pattern of the detected molecular ion, retention time, accurate mass and MS/MS spectral searching against an accurate mass forensic toxicology LC/MS/MS library. The accurate mass TOF- MS/MS product ion scan of a drug compound is shown as a mirror view with the matched library spectrum for easy comparison. The bottom right panel of Figure 5 shows an example of the spectral comparison of library versus urine sample (isobaric mixture of two JWH-019 and 120 metabolites). The sample spectrum is the top blue trace and the library is the grey trace. In screening applications, both false positive and false negative hits need to be kept to a minimum so it is critical to utilize all possible screening criteria for data review. In terms of false positives, for instance, when only mass accuracy and retention time were used for data review and report generation, it was observed that the sheer file size doubled the report generated with all criteria enforced (mass accuracy, retention time, isotope ration difference and library score) for a 10-sample batch (Figure 7). On the other hand, when all criteria were strictly in place, it was inevitable that some false negatives were created. Figure 7 shows a situation in which one spiked K2 analyte JWH-200 was categorized as negative based on its elevated isotope ratio difference due to a co-eluting substance that shares similar molecular weight as JWH-200 s second isotopic peak. However, with >90% library matching score, finding of JWH-200 in this particular sample was shown to be a true positive. This example highlighted the importance of using MS/MS library matching in data review. In our testing, it was frequently observed that the isotope ratio patterns of many analytes were altered in real urine samples and having MS/MS data proved to be critical. Figure 8. Interference caused isotopic ratio change, but MS/MS confirmed identity and eliminated possible false negative caused by elevated isotope ratio difference for JWH-200 in neat solution. From Screening to Quantitation After sample data was processed in MasterView, the m/zs and retention times were determined for all target analytes, along with the related extraction windows for both m/z and retention time. Such results can be linked with MultiQuant software and used for quantitation. Figure 9 shows a calibration curve for JWH-018 that was generated in MultiQuant TM software. The process was fast, reliable and flexible. p 4
5 Figure 9. Calibration curve of JWH-018 in human urine. Reporting MasterView has flexible reporting features showing the compound identifications using both high resolution and accurate mass MS and MS/MS information. Customizable report templates can be used to generate word, csv text or PDF reports (Figure 10). Figure 10. Flexible Reporting Features in MasterView Conclusion The MPX-2 High Throughput system with the AB SCIEX TripleTOF 4600 high resolution accurate mass system has enabled a fast forensic screening, around 2-min injection to injection time, to screen for synthetic cannabinoids/metabolites utilizing multiple points of selectivity to ensure confident ID of components in urine samples. The workflow is simple, highly flexible in adding new analytes, and has superior sensitivity and specificity than immunoassay-based screening approaches. p 5
6 For Research Use Only. Not for use in diagnostic procedures AB SCIEX. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. p 6
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS
Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics
Bruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS
Bruker ToxScreener TM A Comprehensive Screening Solution for Forensic Toxicology Innovation with Integrity UHR-TOF MS ToxScreener - Get the Complete Picture Forensic laboratories are frequently required
DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM
DRUG METABLISM Drug discovery & development solutions FR DRUG METABLISM Fast and efficient metabolite identification is critical in today s drug discovery pipeline. The goal is to achieve rapid structural
A Streamlined Workflow for Untargeted Metabolomics
A Streamlined Workflow for Untargeted Metabolomics Employing XCMS plus, a Simultaneous Data Processing and Metabolite Identification Software Package for Rapid Untargeted Metabolite Screening Baljit K.
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,
Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants
Metabolomic Profiling of Accurate Mass LC- Data to Identify Unexpected Environmental Pollutants André Schreiber 1, David Cox 1, Nadia Pace 1, Christopher Borton 2 1 AB SCIEX, Concord, ntario, Canada; 2
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with the sensitivity and selectivity
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs
AB SCIEX TOF/TOF 4800 PLUS SYSTEM Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM It s just what you expect from the industry leader. The AB SCIEX 4800 Plus MALDI TOF/TOF
Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification
Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Overview LightSight Software for Metabolite Identification is a complete environment
Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
Authenticity Assessment of Fruit Juices using LC-MS/MS and Metabolomic Data Processing
Authenticity Assessment of Fruit Juices using LC-MS/MS and Metabolomic Data Processing Lukas Vaclavik 1, Ondrej Lacina 1, André Schreiber 2, and Jana Hajslova 1 1 Institute of Chemical Technology, Prague
MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis
MarkerView Software 1.2.1 for Metabolomic and Biomarker Profiling Analysis Overview MarkerView software is a novel program designed for metabolomics applications and biomarker profiling workflows 1. Using
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs
Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan
MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics
MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics With Unique QTRAP and TripleTOF 5600 System Technology Targeted peptide quantification is a rapidly growing application
Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
A Navigation through the Tracefinder Software Structure and Workflow Options. Frans Schoutsen Pesticide Symposium Prague 27 April 2015
A Navigation through the Tracefinder Software Structure and Workflow Options Frans Schoutsen Pesticide Symposium Prague 27 April 2015 Kings day in The Netherlands 1 Index Introduction Acquisition, Method
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report
PRODUCT BULLETIN Cliquid ChemoView 3.0 Software for Routine Screening and Quantitation Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report KEY FEATURES Secure user login that
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.
Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
Waters Core Chromatography Training (2 Days)
2015 Page 2 Waters Core Chromatography Training (2 Days) The learning objective of this two day course is to teach the user core chromatography, system and software fundamentals in an outcomes based approach.
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
Agilent 6000 Series LC/MS Solutions CONFIDENCE IN QUANTITATIVE AND QUALITATIVE ANALYSIS
Agilent 6000 Series LC/MS Solutions CONFIDENCE IN QUANTITATIVE AND QUALITATIVE ANALYSIS AGILENT 6000 SERIES LC/MS SOLUTIONS UNRIVALED ACHIEVEMENT AT EVERY LEVEL The Agilent 6000 Series of LC/MS instruments
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
WATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS
EKSIGENT EKSPERT NANOLC 4 Unmatched flexibility for low flow LC/MS EKSIGENT EKSPERT NANOLC 4 Focused ekspertise In 23, Eksigent changed the way scientists looked at nanoscale chromatography with the introduction
ProteinPilot Report for ProteinPilot Software
ProteinPilot Report for ProteinPilot Software Detailed Analysis of Protein Identification / Quantitation Results Automatically Sean L Seymour, Christie Hunter SCIEX, USA Pow erful mass spectrometers like
AxION edoor. Web-Based, Open-Access Mass Spectrometry Software
AxION edoor Web-Based, Open-Access Mass Spectrometry Software 2 AxION edoor is the open-access software solution for today s fast-paced laboratory. It is designed to operate and optimize the management
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
TECHNICAL NOTE Toxicology Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with
Analyst 1.6 Software. Software Reference Guide
Analyst 1.6 Software Software Reference Guide Release Date: August 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment.
Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening
Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening Application Note Forensics Author Erin Shonsey Director of Research Alabama Department of Forensic Science Abstract
Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software
Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software Abstract Presented here is a case study of a walk-up liquid chromatography
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
NUVISAN Pharma Services
NUVISAN Pharma Services CESI MS Now available! 1st CRO in Europe! At the highest levels of quality. LABORATORY SERVICES Equipment update STATE OF THE ART AT NUVISAN CESI MS Now available! 1st CRO in Europe!
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
HRMS in Clinical Research: from Targeted Quantification to Metabolomics
A sponsored whitepaper. HRMS in Clinical Research: from Targeted Quantification to Metabolomics By: Bertrand Rochat Ph. D., Research Project Leader, Faculté de Biologie et de Médecine of the Centre Hospitalier
Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests
Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests AB SCIEX TOF/TOF 5800 System with DynamicExit Algorithm and ProteinPilot Software for Robust Protein
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
Instrumentation. Sample pretreatment
Analysis of Propofol (2,6-Diisopropylphenol) and its Metabolites in One Injection using the DuoSpray Ion Source AB SCIEX QTRAP 500 LC-MS/MS System and Gerstel, Inc. MultiPurpose Sampler Adrian M. Taylor,
Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow
Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow Abstract There is high demand in clinical research and forensic toxicology for comprehensive,
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
Analyst 1.6.2 Software. Release Notes
RUO-IDV-03-0236-A Release Date: July 2013 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment. This document is copyright protected
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide
amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS
amazon SL Setting New Standards in Performance, Simplicity and Value Innovation with Integrity Ion Trap Best-In-Class Ion Trap Mass Spectrometer for Routine Analysis The amazon SL entry-level system is
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
Application of Structure-Based LC/MS Database Management for Forensic Analysis
Cozette M. Cuppett and Michael P. Balogh Waters Corporation, Milford, MA USA Antony Williams, Vitaly Lashin and Ilya Troisky Advanced Chemistry Development, Toronto, ntario, Canada verview Application
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
Thermo Scientific ExactFinder Software
mass spectrometry Thermo Scientific ExactFinder Software Workflow software for routine targeted and general unknown screening Unifies qualitative and quantitative high-resolution accurate mass workflows
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method
Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method Introduction During the last decade, the complexity of samples
Analyst 1.6.1 Software. Advanced User Guide
This document is provided to customers who have purchased AB Sciex equipment to use in the operation of such AB Sciex equipment. This document is copyright protected and any reproduction of this document
Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit
Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit Fast, simple and cost effective analysis Many areas of biochemical research and
Tutorial 9: SWATH data analysis in Skyline
Tutorial 9: SWATH data analysis in Skyline In this tutorial we will learn how to perform targeted post-acquisition analysis for protein identification and quantitation using a data-independent dataset
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
Opiates in Urine by SAMHSA GC/MS
application Note Gas Chromatography/ Mass Spectrometry Author Timothy D. Ruppel PerkinElmer, Inc. Shelton, CT 06484 USA Opiates in Urine by SAMHSA GC/MS Introduction The United States Department of Health
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
AppNote 4/2014. Prep-and-Shoot : The Completely Automated Workfl ow for the Hydrolysis and Analysis of Urine Samples by LC/MS/MS.
AppNote 4/2014 Prep-and-Shoot : The Completely Automated Workfl ow for the Hydrolysis and Analysis of Urine Samples by LC/MS/MS. Oscar G. Cabrices, Fred D. Foster Gerstel, Inc., 701 Digital Dr. Suite J,
Investigating Biological Variation of Liver Enzymes in Human Hepatocytes
Investigating Biological Variation of Liver Enzymes in Human Hepatocytes MS/MS ALL with SWATH Acquisition on the TripleTOF Systems Xu Wang 1, Hui Zhang 2, Christie Hunter 1 1 AB SCIEX, USA, 2 Pfizer, USA
Analysis of Phthalate Esters in Children's Toys Using GC-MS
C146-E152 Analysis of Phthalate Esters in Children's Toys Using GC-MS GC/MS Technical Report No.4 Yuki Sakamoto, Katsuhiro Nakagawa, Haruhiko Miyagawa Abstract As of February 29, the US Consumer Product
New drugs of abuse? A case study on Phenazepam & Methoxetamine
New drugs of abuse? A case study on Phenazepam & Methoxetamine Presenter: Nadia Wong Co authors: Dr Yao Yi Ju & Alex Low Xuan Kai Analytical Toxicology Laboratory Clinical & Forensic Toxicology Unit Applied
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT
AppNote 1/2014 Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine Fred D. Foster, John R. Stuff, Edward A. Pfannkoch GERSTEL, Inc., 701 Digital Dr. Suite J,
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS Mikaël LEVI Shimadzu France Welcome Seminar Barcelona 2013 Interest in Steroids assay by LC-MS/MS Steroids measurement is essential
A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System
A Generic LC-MS Method for the Analysis of Multiple of Drug of Abuse Classes with the Thermo Scientific Exactive TM System Kent Johnson Fortes lab, Wilsonville Oregon List of drug of abuse candidates for
Alignment and Preprocessing for Data Analysis
Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise
Urine Drug Testing. Why drug test? Set the Standard!
Urine Drug Testing Urine drug testing is a valuable tool for the physician when making treatment decisions. Patients and physicians benefit from safe and effective treatment of pain but there are always
INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS
AxION iqt GC/MS/MS INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE IDENTIFY WHAT YOU RE LOOKING FOR (AND EVEN WHAT YOU RE NOT) The GC/MS/MS Solution for the Work You Do The ideal solution for industrial,
Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS
Therapeutic Drug Monitoring of Antiretroviral Drugs with PLC-M Ursula Gutteck-Amsler, Katharina M. Rentsch Abstract Prospective and retrospective studies have provided some evidence of the clinical and
Mass Frontier 7.0 Quick Start Guide
Mass Frontier 7.0 Quick Start Guide The topics in this guide briefly step you through key features of the Mass Frontier application. Editing a Structure Working with Spectral Trees Building a Library Predicting
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
Analyst 1.6 Software. Getting Started Guide
Analyst 1.6 Software Getting Started Guide Release Date: August 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment. This
Mass Frontier Version 7.0
Mass Frontier Version 7.0 User Guide XCALI-97349 Revision A February 2011 2011 Thermo Fisher Scientific Inc. All rights reserved. Mass Frontier, Mass Frontier Server Manager, Fragmentation Library, Spectral
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System Olaf Scheibner, Maciej Bromirski, Thermo Fisher Scientific, Bremen, Germany
AppNote 9/2013. Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides KEYWORDS ABSTRACT
AppNote 9/2013 Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides Oscar G. Cabrices Gerstel, Inc., 701 Digital Dr. Suite J, Linthicum, MD 21090, USA
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,