1 Table of content 1. PURPOSE PRINCIPLE REQUIREMENTS EQUIPMENT AND MATERIALS REAGENTS PREPARATION OF MEASUREMENT PROCEDURE SAMPLE MEASUREMENT PROCESSING DATA OF MEASUREMENTS CALCULATION OF RESULTS QUALITY CONTROL ERRORS, CALIBRATION AND INTERFERENCES WASTE STREAM AND PROPER DISPOSAL HAZARDS AND PRECAUTIONARY STATEMENTS REFERENCES CONTRIBUTIONS

2 1. PURPOSE This procedure is used to analyze the composition of a biogas sample, specifically the relative quantity of methane and carbon dioxide. 2. PRINCIPLE Gas from either direct samples or gas sampling vials is injected into a gas chromatograph with thermal conductivity detector (GC-TCD). This GC-TCD determines the relative quantity of CH 4, CO 2 and the sum of H 2 and N 2 in function of the different retention times it takes for the sample to migrate through a heated column. By integration of the resulting data, the relative quantity of the main different components (CH 4 and CO 2) is determined. 2

3 3. REQUIREMENTS 3.1. EQUIPMENT AND MATERIALS GC-TCD number 2 (on the right side of number 1) (in room A206) (Brand: Agilent; Type: 6890 Series (GC Systems) Plus +) 3.2. REAGENTS HP GC Autosampler Controller (connection unit to computer) Computer with connection drivers and software for GC controlling and data analysis Agilent J&W Capillary GC column CP7354 Hamilton Gastight Syringe (250 µl) (in the drawer right under the computer) (Brand: Hamilton; Model: GasTight #1725) A series of gas samples in sampling vials to be analyzed Acetone Calibration gas (5% N 2, 5% H 2, 80% CH 4 and 10% CO 2; stored in a small gas bottle next to the GC) 3.3. PREPARATION OF MEASUREMENT Check if the right column is installed in the GC oven (Agilent J&W Capillary GC column CP7354). Check if there is pressure of helium in the helium supply pipe (manometer behind GC number 1). Helium is the carrier gas in this GC-analysis. Boot the computer if it was shut down. This computer is located between the two GC s. The computer should never be shut down after use. Turn on GC2 with the power switch located on the front lower left. Turn on the HP GC Autosampler Controller unit with the power switch located on the front lower left. Run the Bootp shortcut on the desktop of the computer. This links the computer to the controller unit to control the GC and acquisition data 3

4 4. PROCEDURE Fig. 1. GC-TCD (Agilent 6890 Series GC Systems Plus +). Run the GCµTCD shortcut on the desktop to start the software that will acquire data and log it with a preconfigured method. Before measurement of samples the column and detector should be cleaned, running a preconfigured method: 1. Load CLEAN.M from C:\MSDChem\2\METHODS. 4

5 2. Click Sample name in the Instrument control panel to create a log file of the cleaning. This saves the chromatogram of any method to the computer. a. Locate the directory: C:\MSDChem\2\DATA\(in which 2 is the number of the GCunit) b. Create a new directory and name it with the day of the GC-analysis (not the same date!) in the format of YYMMDD in which Y is the number of the year, M for month and D for day. For example: 4 th July 2013 becomes Additionally your name can be added like George. c. Within this directory label the log file CLEAN#.D in which # is the number of the cleaning action on this day. For example CLEAN2.D. 3. Now click START RUN on the computer (do not just click OK ). 4. Press the PREP RUN button at the GC2 panel as soon as not ready LED is longer lit up. 5. Then press the START button on the GC2 panel as soon as the start and run for about 4 minutes. Do not abort this action. After cleaning the GC column will have to cool down before any measurement can be done. 5

6 When cleaning is done, the measurement of samples can begin by loading the method BIOGAS.M from C:\MSDChem\2\METHODS. After the last measurement of a series of samples that have to be compared, method CLEAN.M should be run again as previously described SAMPLE MEASUREMENT Preparing the computer The following steps should be followed before any other gas samples are taken: Make sure the BIOGAS.M method is loaded in the software. This method will run the analysis at a constant 60 C. 6

7 Prepare saving of the log file to C:\MSDChem\2\DATA\ YYMMDD \ with the correct sample name that is to be analyzed next. For example 0703BIOREACTORSOFIE7.D. Do this by clicking Sample name in the Instrument control panel to create a data log file as done before during cleaning Taking the gas sample Prepare the Hamilton gas syringe (250 µl) by reassembling it in case it was taken apart. Before taking any sample, align the valve of the glass Hamilton syringe with the cylinder axis so it is opened. Then rinse the syringe by filling and completely ejecting it 2 or 3 times with air from the room you are in making sure there is nothing but gas inside (no liquids). Finally eject all air and close the valve again. There are two options on how to take gas samples: gas samples can be either taken from sampling gas vials or directly from the reactor. 1. Taking gas sample from a sampling vial a. Turn the second sampling vial (the one with the most gas (± 2 ml)) upside down and pierce the inner ring of the butyl stopper with the gas syringe and aim for the gas headspace. The first sampling vial with the small gas bubbles (± 0.5 ml) are for backup in case something goes wrong with the second one. b. Open the valve of the syringe and rinse the volume of the syringe a couple of times with the content of the gas headspace in the vial. Finally take the gas sample of 250 µl and wait for a few seconds for gas pressure to equalize. c. Close the valve at the top of the syringe and pull out the syringe again in a smooth movement. Place the sampling vial in the correct container again. 7

8 2. Taking direct gas samples from a reactor a. Shake the reactor so that the gas composition is representative for a certain time. b. Pierce the gas syringe all the way down through the septum of the bioreactor (it shouldn t hit the sludge level inside the reactor). c. Open the valve and rinse a few times with gas in the reactor. Finally take the gas sample of 250 µl and wait for a few seconds for pressure to equalize. d. Close the valve at the top of the syringe and take out the syringe again. e. Without opening the valve, pressurize the gas volume from 250 µl to 200 µl while carrying the syringe with the gas to the GC inlet. This makes sure no air tries to enter the syringe, as only gas could escape in case of malfunction Injecting gas sample into GC a. When the correct data saving file name has been entered in the computer and you have the sample ready in the syringe, click START RUN. b. Press the PREP RUN button on the panel of GC2. Wait until the not ready LED is no longer lit up. Carefully bring the gas syringe towards the inlet of the liner of GC2. Hold the syringe downwards near the opening. Do mind the heat of the metal ring on the outside of the liner as you can burn your fingers here. c. Now, without opening the valve of the syringe yet, pressurize the gas in the syringe by reducing the volume from 200 µl to exact 100 µl outside the GC. d. Now quickly but carefully open the valve outside the inlet and then immediately insert the syringe all the way down into the inlet of the liner of GC2 and empty it with a fluent and short movement of the piston from 100 µl to 0 µl. Immediately and simultaneously, press the START button on the GC2 s panel and retract the needle from the inlet. This series of actions should be fluent and identical in timing to the other times when performed and as short as possible. The computer should now have started logging and visualizing the chromatogram until the programmed time of 6 minutes runs out. Do not abort this action. After this time the data is no longer logged, however the GC keeps drawing a chromatogram on the PC screen until the next action is started (for example the run on another sample). e. Wait at least one minute before injecting the next sample as remains of other compounds can still be migrating through the column which is better not included in the next chromatogram as they would be part of the area when integrating. f. Retake all steps from part (Preparing the computer) through and for the amount of samples to be analyzed. 8

9 Finalizing the measurement When all sample measurements are done, check if you see dirt or condensate in the gas syringe. If so, it should be thoroughly cleaned with acetone, disassembled and stored to dry disassembled so acetone can vaporize and escape well. Do not shut down the computer after use; only close the programs you started PROCESSING DATA OF MEASUREMENTS To gain time, this action can be performed on previously acquired data while the next measurement is running, since it is done through a different program. Alternatively, this can be done after all measurements at a later time. Run the GCµTCD Enhanced Data Analysis shortcut on the desktop of the computer to start the software that will enable you to interpret the data. Select the data file (.D ) to be analyzed. Choose Chromatogram in the menu and choose Integrate. Check if the program is able to distinguish 3 peaks at the expected and known succession of retention timings: a) the sum of N 2 and H 2 around 4.23 minutes; b) CH 4 around 4.43 minutes; c) CO 2 around 4.84 minutes. They are cut off by blue lines in the graph. Possibly a fourth peak can be found in front of or at the back of the other peaks, but this can be ignored, as only the ratio of CH 4 and CO 2 is of importance. However if the size of this unknown peak is very large compared to 3 being distinguished, the integration detection of the smallest peak may fail to compute and produce an area of significance. In this case the initial threshold setting should be lowered in steps of 1 unit, until after reintegration a representative area can be computed. This setting can be found in the menu Chromatogram > Signal Integration Parameters. Once changed, choose Integrate again to recompute using the changed parameters. Restore the initial threshold setting to the original setting of 16.5 before loading the next sample data for integration. 9

10 Now choose Chromatogram again in the menu and choose Generate Percent Report. 10

11 Scroll down to the bottom and note this data (or copy it to your personal data file). Data is usually noted up to 4 significant numbers. For example % becomes %. 11

12 o o o Note retention times of the peaks for CH 4 and CO 2 (R.T min column) (this is the top of the retention peak, not the start or end of steep slopes). Time is displayed in digits, so 5 minutes and 30 seconds is written as 5.5. Note the relative percentage in the last column in this report (% of total): - 2 nd peak: CH 4 (around 4.43 minutes) - 3 rd peak: CO 2 (around 4.84 minutes) There is no need to note the peak for N 2 and H 2. In case there is a fourth peak, make sure you read or copy the 2 correct peaks using the retention timing as reference, as data from the fourth peak is irrelevant. Important remark: DO NOT try to open a sample data file with the GCµTCD Enhanced Data Analysis tool or any other program when this file is still being written to by the GCµTCD logging program that controls the GC. Doing so will result in a corrupted.d data file and the sample will have to be measured all over again! Do not shut down the computer after use, only close the programs you started. 5. CALCULATION OF RESULTS It is advised to use a spreadsheet for calculations. Use the calibration factors mentioned earlier (or determine them again yourself using calibration gas of known contents) to multiply the measured area percentage to the correct area percentage. Then calculate the relative quantity of the two peaks (CH 4 and CO 2) with the following formulas by dividing each one through the sum of the two to obtain the relative quantity in the gas sample that is only supposed to consist of these gasses. The final percentages acquired for CH 4 or CO 2 can be divided by the sum of percentages of CH 4 and CO 2 to find the relative quantity of produced gas in this biogas sample. Formulas: ([CH 4 ] area CF CH4 ) 100 [CH 4 ] sample = ([CH 4 ] area CF CH4 ) + ([CO 2 ] area CF CO2 ) [CO 2 ] sample = ([CO 2 ] area CF CO2 ) 100 ([CH 4 ] area CF CH4 ) + ([CO 2 ] area CF CO2 ) With [CH 4 ] sample : concentration of CH 4 in the biogas sample (%v) [CO 2 ] samle : concentration of CO 2 in the biogas sample (%v) 12

13 [CH 4 ] area : area peak percentage of CH 4 (%) [CO 2 ] area : area peak percentage of CO 2 (%) CF CH4 : calibration factor for CH 4 (no unit) CF CO2 : calibration factor for CO 2 (no unit) 6. QUALITY CONTROL A sample of calibration gas should be measured during each sampling run as a quality control. Biogas samples originating from anaerobic digestion batch test should contain 50-90% CH 4 and 10-50% CO ERRORS, CALIBRATION AND INTERFERENCES This GC unit always measures a difference in composition compared to the actual composition as was determined with a calibration gas (consisting of 5 % N 2, 5% H 2, 80% CH 4 and 10% CO 2). The ratio of this measurement to the actual composition was turned into a calibration factor which is applied to all samples to determine the actual composition of gas samples in analysis. Multiply the obtained area percentage with the calibration factors in order to read the correct area percentage of the samples. Table 1 shows an example of calibration factors which are to be determined calibration gas. Table 1: Example of calibration factors for methane and carbon dioxide CH 4 CO 2 Calibration factor WASTE STREAM AND PROPER DISPOSAL Dispose any captured spills or contents of vials in the correct liquid waste stream disposal barrel in the lab (acids). Clean all gas vials and remove their labeling with acetone. 13

14 9. HAZARDS AND PRECAUTIONARY STATEMENTS Mind the heat of the liner opening in which the gas sample is inserted into the GC. 10. REFERENCES APHA, Awwa, WEF, Standard Methods for the Examination of Water and Wastewater, 21st ed. American Public Health Association, Washington DC. Laurent C., Biogaspotentieel van microalgen bacteriën vlokken bij verschillende groeicondities en voorbehandelingen. Annex: Biogas sampling. Master thesis, Ghent University, Campus Kortrijk, 123p. 11. CONTRIBUTIONS SOP written by: Cedric Laurent (master thesis student EnAlgae project). SOP restructured by: Alexandra Lefoulon (internship student EnAlgae project). SOP approved by: Sofie Van Den Hende (staff EnAlgae project). 14

Analysis of Liquid Samples on the Agilent GC-MS

Analysis of Liquid Samples on the Agilent GC-MS Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >

More information

Background Information

Background Information 1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software

More information

Expectations for GC-MS Lab

Expectations for GC-MS Lab Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax. Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative

More information

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note. EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer HT3 Application Note Introduction The analysis of dissolved gases in ground water is important in

More information

Micromass LCT User s Guide

Micromass LCT User s Guide Micromass LCT User s Guide 1) Log on to MassLynx with your username & password. 2) After you have logged in, the MassLynx software will automatically run. 3) After MassLynx has come up, open your project

More information

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated

More information

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

GC-xt. The most flexible GC /GC-MS liquid sample injector

GC-xt. The most flexible GC /GC-MS liquid sample injector GC-xt Prep and Load Platform The most flexible GC /GC-MS liquid sample injector Environmental Foods/Beverages Consumer Products Forensics Petrochemicals/Polymers Pharmaceuticals GC-xt Liquid mode Prep

More information

23 The Thermal Conductivity Detector

23 The Thermal Conductivity Detector 23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing

More information

GAS CHROMATOGRAPHY. DETERMINING THE PERCENT COMPOSITION OF A MIXTURE OF VOLATILE LIQUIDS USING A GAS CHROMATOGRAPH. (Revised: 1-12-93)

GAS CHROMATOGRAPHY. DETERMINING THE PERCENT COMPOSITION OF A MIXTURE OF VOLATILE LIQUIDS USING A GAS CHROMATOGRAPH. (Revised: 1-12-93) DETERMINING THE PERCENT COMPOSITION OF A MIXTURE OF VOLATILE LIQUIDS USING A GAS CHROMATOGRAPH. (Revised: 1-12-93) INTRODUCTION Gas chromatography is a technique used to analyze mixtures. The instrument

More information

Chromatograph Maintenance and Troubleshooting Class # 5050

Chromatograph Maintenance and Troubleshooting Class # 5050 Chromatograph Maintenance and Troubleshooting Class # 5050 Shane Hale Business Development Manager Emerson Process Management Gas Chromatographs Division 5650 Brittmoore Drive Houston, TX 77041 USA Introduction

More information

Identification of Arson Accelerants by Gas Chromatography

Identification of Arson Accelerants by Gas Chromatography Identification of Arson Accelerants by Gas Chromatography Purpose Arson is an insidious crime that annually claims the lives of hundreds of Americans and costs billions of dollars. 1 Scientists and criminal

More information

16.4.3 Lab: Data Backup and Recovery in Windows XP

16.4.3 Lab: Data Backup and Recovery in Windows XP 16.4.3 Lab: Data Backup and Recovery in Windows XP Introduction Print and complete this lab. In this lab, you will back up data. You will also perform a recovery of the data. Recommended Equipment The

More information

On-line Dissolved Gas Analysis

On-line Dissolved Gas Analysis On-line Dissolved Gas Analysis accord. to IEC 567/ASTM 3612 Online-Monitoring of transformers Automatic on-line analysis of 11 gases The analysis is done in 2 steps with high selectivity and accuracy fully

More information

Becton-Dickinson FACSCalibur Flow Cytometer Instructions

Becton-Dickinson FACSCalibur Flow Cytometer Instructions Start-up 1. Turn on flow cytometer by pushing green button on right side of instrument, toward the back. Give it 30 minutes to warm up before collecting data. 2. Turn on monitor. 3. Turn on computer. If

More information

ORGANIC SAMPLE PREPARATION

ORGANIC SAMPLE PREPARATION ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,

More information

Procedure for updating Firmware of EZ4 W or ICC50 W

Procedure for updating Firmware of EZ4 W or ICC50 W Procedure for updating Firmware of EZ4 W or ICC50 W 1. Download the Firmware file for your camera to your PC 2. Download the Leica Camera Configuration program to your PC 3. Install Leica Camera Configuration

More information

Thermal Conductivity Detector

Thermal Conductivity Detector Thermal Conductivity Detector Troubleshooting Tips The thermal conductivity detector (TCD) detects the difference in thermal conductivity between column effluent flow (carrier gas + sample components)

More information

With proper calibration, the upper limit may be extended as needed.

With proper calibration, the upper limit may be extended as needed. While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the

More information

Calibration of Glassware

Calibration of Glassware Calibration of Glassware Introduction Glassware is commonly calibrated using a liquid of known, specific density, and an analytical balance. The procedure is to determine the mass of liquid the glassware

More information

10.3.1.6 Lab - Data Backup and Recovery in Windows XP

10.3.1.6 Lab - Data Backup and Recovery in Windows XP 5.0 10.3.1.6 Lab - Data Backup and Recovery in Windows XP Introduction Print and complete this lab. In this lab, you will back up data. You will also perform a recovery of the data. Recommended Equipment

More information

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in

More information

Small Animal Respiratory Exchange Ratio (RER)

Small Animal Respiratory Exchange Ratio (RER) iworx Physiology Lab Experiment Experiment AMe-1 Small Animal Respiratory Exchange Ratio (RER) Note: The lab presented here is intended for evaluation purposes only. iworx users should refer to the User

More information

Fractional Distillation and Gas Chromatography

Fractional Distillation and Gas Chromatography Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils

More information

Protocol: HPLC (amino acids)

Protocol: HPLC (amino acids) Page 1 of 6 1. Chemicals Composition M MW gram volume (ml) product code 0-pthaldialdehyde C 8H 6O 2 134.13 Sigma P0657 Perchloric acid HClO 4 3.5 100.46 Dilute 70% stock solution 1:1 with demi to obtain

More information

BTU ANALYSIS USING A GAS CHROMATOGRAPH

BTU ANALYSIS USING A GAS CHROMATOGRAPH BTU ANALYSIS USING A GAS CHROMATOGRAPH By Joseph E. Landes Southern Petroleum Laboratories, Inc. Technical Director Hydrocarbon and Field Services 8820 Interchange Drive Houston, Texas 77054 Introduction

More information

GC-2010 Gas Chromatograph INSTRUCTION MANUAL

GC-2010 Gas Chromatograph INSTRUCTION MANUAL 221-40407F Gas Chromatograph INSTRUCTION MANUAL Read this instruction manual thoroughly before using this instrument. Save the manual for future reference. Copyright Shimadzu Corporation 2000. All rights

More information

AIRFREE TECHNIQUE AND SENSITIVE REAGENTS S ECTI O N 1: GLASS W ARE A ND E Q UIP M ENT. A. Using a manifold

AIRFREE TECHNIQUE AND SENSITIVE REAGENTS S ECTI O N 1: GLASS W ARE A ND E Q UIP M ENT. A. Using a manifold AIRFREE TECHNIQUE AND SENSITIVE REAGENTS S ECTI O N 1: GLASS W ARE A ND E Q UIP M ENT Some organic compounds are air sensitive. They can react with the water vapor or oxygen in the air. In order to perform

More information

Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter

Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter INL/EXT-08-15056 Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter J. E. Daw November 2008 The INL is a U.S. Department of Energy National Laboratory operated by Battelle Energy

More information

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2

More information

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* Method 552.1 Revision 1.0 Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* UCT Products: EUQAX156 (quaternary amine with Cl - counter ion, 6 ml cartridge)** CLTTP050 (CLEAN-THRU

More information

EXPERIMENT 12: Empirical Formula of a Compound

EXPERIMENT 12: Empirical Formula of a Compound EXPERIMENT 12: Empirical Formula of a Compound INTRODUCTION Chemical formulas indicate the composition of compounds. A formula that gives only the simplest ratio of the relative number of atoms in a compound

More information

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK. VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS

FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK. VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS PRINCIPLE This method confirms the identity of volatile compounds in biological

More information

Manual IB-3620 Series

Manual IB-3620 Series IB-RD3620SU3 1 IB-3620U3 CONTENT 1. Introduction... 3 1.1 General Information... 3 2. Hardware IB-3620 Series... 4 2.1 LED Indication / Button Front Panel... 4 2.2 Rear View... 5 3. HDD Installation...

More information

Tuning & Mass Calibration

Tuning & Mass Calibration Tuning & Mass Calibration 1 1 The Sample List Sample List Name Project Name 2 The sample list is the top level screen in the TurboMass Gold Software. Data storage is set up in PROJECT files and within

More information

SPE-TT. Tutorial Version 2009-08-11. Bruker BioSpin 2008-2009; File: Tutorial 11 - SPE-TT; Version: 2009-08-11; By: Dr.

SPE-TT. Tutorial Version 2009-08-11. Bruker BioSpin 2008-2009; File: Tutorial 11 - SPE-TT; Version: 2009-08-11; By: Dr. SPE-TT Tutorial Version 2009-08-11 Introduction This document is a step-by-step guide for the usage of a SPE-TT system that elutes samples from SPE cartridges into the liquid handler. - Installation: Software

More information

ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC)

ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC) ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC) 1 Fast Analysis of C1 to C8 hydrocarbons for mud logging applications within 2 minutes using a temperature programmable Micro GC Fusion NEN GAS2015 Debbie

More information

Zetasizer Nano-ZS User Instructions

Zetasizer Nano-ZS User Instructions Zetasizer Nano-ZS User Instructions 1. Activate the instrument computer by logging in to the central login system in the service corridor. If needed, log in to the local instrument computer Username: zetasizer.

More information

VAPOR PRESSURE AS A FUNCTION OF TEMPERATURE. This laboratory covers material presented in section 11.8 of the 9 th Ed. of the Chang text.

VAPOR PRESSURE AS A FUNCTION OF TEMPERATURE. This laboratory covers material presented in section 11.8 of the 9 th Ed. of the Chang text. VAPOR PRESSURE AS A FUNCTION OF TEMPERATURE Objectives: (1) Observe and measure the change in the vapor pressure (dependent variable) as a function of temperature (independent variable). (2) Analyze the

More information

When Completed - Fax to (410) 704-2993

When Completed - Fax to (410) 704-2993 Revised 2/12 Date: / / Hazardous Chemical Waste Disposal Request Page of Requestor: Phone #: Email: Department: Building & Room #: Item # Description of Waste Chemicals Quantity Container Size & Type Comments

More information

Experiment 13: Determination of Molecular Weight by Freezing Point Depression

Experiment 13: Determination of Molecular Weight by Freezing Point Depression 1 Experiment 13: Determination of Molecular Weight by Freezing Point Depression Objective: In this experiment, you will determine the molecular weight of a compound by measuring the freezing point of a

More information

NIH Clinical Center Patient Education Materials Giving a subcutaneous injection

NIH Clinical Center Patient Education Materials Giving a subcutaneous injection NIH Clinical Center Patient Education Materials What is a subcutaenous injection? A subcutaneous injection is given in the fatty layer of tissue just under the skin. A subcutaneous injection into the fatty

More information

Using Excel (Microsoft Office 2007 Version) for Graphical Analysis of Data

Using Excel (Microsoft Office 2007 Version) for Graphical Analysis of Data Using Excel (Microsoft Office 2007 Version) for Graphical Analysis of Data Introduction In several upcoming labs, a primary goal will be to determine the mathematical relationship between two variable

More information

Evaluation copy. Titration of a Diprotic Acid: Identifying an Unknown. Computer

Evaluation copy. Titration of a Diprotic Acid: Identifying an Unknown. Computer Titration of a Diprotic Acid: Identifying an Unknown Computer 25 A diprotic acid is an acid that yields two H + ions per acid molecule. Examples of diprotic acids are sulfuric acid, H 2 SO 4, and carbonic

More information

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point.. Identification of a Substance by Physical Properties 2009 by David A. Katz. All rights reserved. Permission for academic use provided the original copyright is included Every substance has a unique set

More information

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Frank Tontala, Thermo Fisher Scientific, Germering, Germany Technical Note 708 Key Words Chromeleon Chromatography Data System,

More information

Chapter 6. Using the SQL Server

Chapter 6. Using the SQL Server Chapter 6 Using the SQL Server BC30 Using the SQL Server 1 5/2010 Content 1 Installing and setting up the SQL Server... 3 2 Exporting an SQL database... 7 3 Importing an SQL database... 9 4 Opening the

More information

Appendix C. Vernier Tutorial

Appendix C. Vernier Tutorial C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer

More information

CyTOF2 Sample Acquisition: Super Sampler* (CyTOF software version 6.0.626)

CyTOF2 Sample Acquisition: Super Sampler* (CyTOF software version 6.0.626) CyTOF2 Sample Acquisition: Super Sampler* (CyTOF software version 6.0.626) *Amended from Sampler Manual created by Zach Bjornson Because CyTOF computer is on a private network, you will be bombarded with

More information

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen 1. Check the log-book to see if somebody else has already checked the meter for accuracy within the timeframes below. If the meter hasn

More information

EXCEL Tutorial: How to use EXCEL for Graphs and Calculations.

EXCEL Tutorial: How to use EXCEL for Graphs and Calculations. EXCEL Tutorial: How to use EXCEL for Graphs and Calculations. Excel is powerful tool and can make your life easier if you are proficient in using it. You will need to use Excel to complete most of your

More information

GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits

GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits Issue Implementation Page 1 of 7 GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits Scope and Application This method will determine the level of vanillin and ethyl vanillin in distilled

More information

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ)

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ) Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) Closing Date and Time: Wednesday, April 13, 2016 at 5:00 p.m. (Pacific Time) Contact: Aaron Galloway Fax: 541-888-3250

More information

AMD Analysis & Technology AG

AMD Analysis & Technology AG AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds

More information

GETTING STARTED. 6. Click the New Specimen icon.

GETTING STARTED. 6. Click the New Specimen icon. BD FACSDiva Software Basic Experiment Guide Please note, this guide is not intended to take the place of in-person training but is to be used as a reference as needed after completing training with Unified

More information

QUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd

QUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd QUANTITATIVE AMINO ACID ANALYSIS Aurélie Lolia Applications Manager, Biochrom Ltd QUANTITATIVE AMINO ACID ANALYSIS Principles of amino acid analysis Ion exchange chromatography The Biochrom 30 physiological

More information

WARNING! If you get toner in your eyes, wash it out immediately with cool water and consult a doctor.

WARNING! If you get toner in your eyes, wash it out immediately with cool water and consult a doctor. About Toner Cartridges Your printer uses four toner cartridges: black, yellow, magenta, and cyan. Handle the toner cartridges carefully to avoid spilling toner inside the printer or on yourself. The toner

More information

ChromQuest 5.0 Chromatography Data System

ChromQuest 5.0 Chromatography Data System ChromQuest 5.0 Chromatography Data System User Guide for the Surveyor LC CHROM-97257 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. Surveyor and ChromQuest are trademarks

More information

Determination of Caffeine in Beverages HPLC-1

Determination of Caffeine in Beverages HPLC-1 Determination of in s HPLC-1 Introduction This experiment provides an introduction to the application of High Performance Liquid Chromatography (HPLC) to the solution of complex analytical problems. Cola

More information

Transformer Oil Gas Analysis

Transformer Oil Gas Analysis Get on the Analyzer FAST TRACK Transformer Oil Gas Analysis Helping customers protect the environment! 1 Why Analyzers Value to Customers Alleviates Resource Pressure Facilitates Configuration Use Advanced

More information

Standard Operating Procedure

Standard Operating Procedure Standard Operating Procedure Equipment / Procedure: Filename: Assessor s name & date: Reviewer s name & date: Cosmed K4 2012.1201SOP_cosmed_K4 Richard Metcalfe, December 2012 Tom Nightingale, December

More information

SYNSPEC GC s ALPHA 115 and ALPHA 116

SYNSPEC GC s ALPHA 115 and ALPHA 116 SYNSPEC GC s ALPHA 115 and ALPHA 116 Methane/TotalNonMethaneHydroCarbons Operator manual Version January 2010 Synspec b.v. De Deimten 1 9747 AV Groningen Nederland info@synspec.nl CONTENTS CONTENTS...

More information

Setting up a Quantitative Analysis MS ChemStation

Setting up a Quantitative Analysis MS ChemStation Setting up a Quantitative Analysis MS ChemStation Getting Ready 1. Use the tutorial section "Quant Reports" on the MSD Reference Collection CD-ROM that came with your ChemStation software. 2. Know what

More information

The Molar Mass of a Gas

The Molar Mass of a Gas The Molar Mass of a Gas Goals The purpose of this experiment is to determine the number of grams per mole of a gas by measuring the pressure, volume, temperature, and mass of a sample. Terms to Know Molar

More information

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview. Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide

More information

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. Titration Curves PURPOSE To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. GOALS 1 To gain experience performing acid-base titrations with a ph meter. 2

More information

Split / Splitless Injection for Capillary GC

Split / Splitless Injection for Capillary GC Split / Splitless Injection for Capillary GC Speakers John V Hinshaw CHROMacademy GC Dean GC Connections editor Moderator Tony Taylor CHROMacademy Trainer Technical Director, Crawford Scientific Dave Walsh

More information

Mechanical Energy. Mechanical Energy is energy due to position or motion.

Mechanical Energy. Mechanical Energy is energy due to position or motion. Mechanical Energy Mechanical Energy is energy due to position or motion. Position: This means that matter can have energy even though it is not moving. If you knock something off of your kitchen counter,

More information

Best Practices for Troubleshooting Superloader3

Best Practices for Troubleshooting Superloader3 Best Practices for Troubleshooting Superloader3 This document outlines the best practices to troubleshoot common failures on Superloader 3 units. Start by verifying the following: Cables are properly seated

More information

Micro Cryo-Trap Model 971

Micro Cryo-Trap Model 971 Micro Cryo-Trap Model 971 for Liquid CO2 Cooling Scientific Instrument Services, Inc. 1027 Old York Road Ringoes, NJ 08551 Phone (908) 788-5550 Part #951001M revised 10/ /98 Notice The information in this

More information

RULE 1102. DRY CLEANERS USING SOLVENT OTHER THAN PERCHLOROETHYLENE

RULE 1102. DRY CLEANERS USING SOLVENT OTHER THAN PERCHLOROETHYLENE (Adopted January 6, 1978)(Amended August 3, 1979)(Amended July 11, 1980) (Amended August 3, 1990)(Amended December 7, 1990) (Amended November 17, 2000) RULE 1102. DRY CLEANERS USING SOLVENT OTHER THAN

More information

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080 Training on the ZQ Open access MS Questions? Contact Dr. Allis Chien allis@stanford.edu 650-723-0710 0710 STANFORD UNIVERSITY MASS SPECTROMETRY STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD

More information

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C

More information

(S&G 5th ed. Expt 27, 6th, 7th & 8th eds. Expt 26)

(S&G 5th ed. Expt 27, 6th, 7th & 8th eds. Expt 26) Chem 367-2/ Physical Adsorption 63 PHYSICAL ADSORPTION (S&G 5th ed. Expt 27, 6th, 7th & 8th eds. Expt 26) The sample used in this experiment needs to be conditioned for at least two hours before the experiment;

More information

CHROMOCTANE O N LINE ANALYZER

CHROMOCTANE O N LINE ANALYZER GECIL Process GT instruments CHROMOCTANE O N LINE ANALYZER ASTM D6733 DHA and RON and MON on line GC for Reformer unit Automatic Chromatographic system for analysis of naphta feeds and reforming gasolines

More information

Distillation Experiment

Distillation Experiment Distillation Experiment CHM226 Background The distillation process is a very important technique used to separate compounds based on their boiling points. A substance will boil only when the vapor pressure

More information

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14 STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14 OBJECTIVE The objective of this experiment will be the standardization of sodium hydroxide using potassium hydrogen phthalate by the titration

More information

Bringing Innovations to Standards AC Analytical Controls by PAC All In One Biodiesel Analyzer

Bringing Innovations to Standards AC Analytical Controls by PAC All In One Biodiesel Analyzer Bringing Innovations to Standards AC Analytical Controls by PAC All In One Biodiesel Analyzer www.paclp.com Background Biodiesel offers an alternative to fossil diesel fuel & contributes to the reduction

More information

MassLynx 4.1 Getting Started Guide

MassLynx 4.1 Getting Started Guide MassLynx 4.1 Getting Started Guide 71500113203/Revision A Copyright Waters Corporation 2005. All rights reserved. Copyright notice 2005 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IRELAND.

More information

Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.

Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law. Conductivity of strong and weak electrolytes TEC Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law. Principle It is possible to

More information

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab

Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Beckman Coulter DTX 880 Multimode Detector Bergen County Technical Schools Stem Cell Lab Room 213 Beckman Coulter DTX 880 Multimode Detector Information The Beckman Coulter DTX 880 Multimode Detector is

More information

Agilent 7697A Headspace Sampler Data Sheet

Agilent 7697A Headspace Sampler Data Sheet Agilent 7697A Headspace Sampler Data Sheet Overview The Agilent 7697A Headspace Sampler is part of Agilent s line of gas chromtography products. The 7697A based on the architecture of the market-leading

More information

OralChroma OralChroma DataManager

OralChroma OralChroma DataManager OralChroma OralChroma DataManager User s Guide for Windows Version 3.5 Introduction This is a user s manual for the OralChroma DataManager. Copyright Copyright (C) 2008- ABIMEDICAL CORPORATION. All Rights

More information

Biogas. Biology Teams of 2 or 3. Grade 10-12 LEARNING OUTCOMES DESCRIPTION MATERIALS READINESS ACTIVITIES. Science

Biogas. Biology Teams of 2 or 3. Grade 10-12 LEARNING OUTCOMES DESCRIPTION MATERIALS READINESS ACTIVITIES. Science Science Grade 10-12 Lab Teams of 2 or 3 DESCRIPTION Biogas is a fuel which is produced from the breakdown of organic matter. Students will have an opportunity to create their own biogas generators. At

More information

Focus XC. Ultimate Fully Automated Peptide Synthesizer with Sonication and Heating Options

Focus XC. Ultimate Fully Automated Peptide Synthesizer with Sonication and Heating Options Focus XC Ultimate Fully Automated Peptide Synthesizer with Sonication and Heating Options FOCUS XC AUTOMATED PEPTIDE SYNTHESIZER aapptec s Focus XC is a compact, easy to use fully automated peptide synthesizer

More information

Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm

Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm 1. How to prepare straws for pronuclear stages and embryos (use straws with plug, 66.5 mm) Straws are

More information

Spreadsheets and Laboratory Data Analysis: Excel 2003 Version (Excel 2007 is only slightly different)

Spreadsheets and Laboratory Data Analysis: Excel 2003 Version (Excel 2007 is only slightly different) Spreadsheets and Laboratory Data Analysis: Excel 2003 Version (Excel 2007 is only slightly different) Spreadsheets are computer programs that allow the user to enter and manipulate numbers. They are capable

More information

Partner: Jack 17 November 2011. Determination of the Molar Mass of Volatile Liquids

Partner: Jack 17 November 2011. Determination of the Molar Mass of Volatile Liquids Partner: Jack 17 November 2011 Determination of the Molar Mass of Volatile Liquids Purpose: The purpose of this experiment is to determine the molar mass of three volatile liquids. The liquid is vaporized

More information

Quantitative Evaluation of Perfume Fleuressence Samples using the znose

Quantitative Evaluation of Perfume Fleuressence Samples using the znose Quantitative Evaluation of Perfume Fleuressence Samples using the znose Edward J. Staples, Electronic Sensor Technology Electronic Noses Conventional electronic noses (enoses) produce a recognizable response

More information

Support Cleaning Apparatus. Operation, Cleaning, & Safety Manual

Support Cleaning Apparatus. Operation, Cleaning, & Safety Manual Support Cleaning Apparatus Operation, Cleaning, & Safety Manual Revision 1.1 31 May 2014 WARNING: The instructions and precautionary statements contained herein are intended for the Support Cleaning Apparatus

More information

ß-CYCLODEXTRIN SYNONYMS

ß-CYCLODEXTRIN SYNONYMS ß-CYCLODEXTRIN Prepared at the 44th JECFA (1995), published in FNP 52 Add 3 (1995) superseding specifications prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications

More information

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas

More information

ArtisanLink Staining System is an automated special stains slide

ArtisanLink Staining System is an automated special stains slide Technical Tips Tips on using the ArtisanLink Special Staining System Jamie Nowacek, BS, HT(ASCP) CM, QIHC, PMP Dako North America, Inc. Carpinteria, CA, USA ArtisanLink Staining System is an automated

More information

Safe Method: Gas Safety in Catering Establishments

Safe Method: Gas Safety in Catering Establishments Safe Method: Gas Safety in Catering Establishments Safety Point Why? What do you do? Gas equipment and services must only be installed, maintained and repaired by a Gas Safe registered installer. Check

More information

IFC. A new perspective i l ifi ti. p p g p. Single softw FractPAL software plug-in for Agilent ChemStation. Medicinal Chemistry.

IFC. A new perspective i l ifi ti. p p g p. Single softw FractPAL software plug-in for Agilent ChemStation. Medicinal Chemistry. IFC A new perspective i l ifi ti Medicinal Chemistry p p CombiChem Compound Isolation Autopurification Drug Discovery y y q p p p p g p Single softw FractPAL software plug-in for Agilent ChemStation The

More information

POLYVINYL ALCOHOL. SYNONYMS Vinyl alcohol polymer, PVOH, INS No. 1203 DEFINITION DESCRIPTION FUNCTIONAL USES CHARACTERISTICS

POLYVINYL ALCOHOL. SYNONYMS Vinyl alcohol polymer, PVOH, INS No. 1203 DEFINITION DESCRIPTION FUNCTIONAL USES CHARACTERISTICS POLYVINYL ALCOHOL Prepared at the 68 th JECFA (2007) and published in FAO JECFA Monographs 4 (2007), superseding specifications prepared at the 63 rd JECFA (2004) and published in the Combined Compendium

More information