ELECTROCHEMISTRY. Theory and Practice
|
|
|
- Angel Conley
- 9 years ago
- Views:
Transcription
1 ELECTROCHEMISTRY Theory and Practice Dr. Axel Bier Lab Product Application Manager - Electrochemistry For more information, please visit:
2 P a g e 2 CONTENT INTRODUCTION 3 HOW TO MEASURE..3 CALIBRATION...4 GENEARL PH INFORMATION... 5 COMBINED PH ELECTRODE. 6 REFERENCE ELECTRODE...6 LIQUID JUNCTION....7 BUFFERS AND CALIBRATION.. 8 BUFFERS AT VARIOUS TEMPERATURES.8 PRECAUTIONS USING BUFFERS.. 9 TEMPERATURE CORRECTION. 9 PH FAQS Hach Company PO Box 389 Loveland, CO
3 P a g e INTRODUCTION ph in an aqueous solutions is a measure of hydrogen and hydroxide ions. Water molecules dissociate in hydrogen (H + ) and hydroxide (OH - ) ions, H 2 O H + + OH - but the number of ions formed is very small. Water at 25 C contains 1 x10-7 mol/l of hydrogen ions and the same concentration of hydroxide ions, where the concentration (mol/l) of hydrogen ions [H + ] multiplied by the concentration (mol/l) of hydroxide ions [OH - ] is constant: K w = [H + ] [OH - ] <1> K w is the dissociation constant for water and it depends on temperature. Temperature C K w 10 0,2920 * ,4505 * ,6809* ,008 * ,469 * Acids in water increase the [H + ] and, because the product [H + ] [OH - ] must be constant, acids decrease the [OH - ]. Bases increase [OH - ] and decrease [H + ]. For example, suppose an acid is added to water at 25 C and the acid raises the [H + ] to 1.0 x 10-3 mol/l. Because [H + ] [OH - ] must always equal 1.00 x 10-14, [OH - ] will be 1.0 x mol/l. ph is the common way of expressing the hydrogen ion concentration [H + ]. ph is defined as: ph = -log [H + ] <2> In the example above, the hydrogen ion concentration is 1.0 x 10-3 mol/l and the ph is Alternatively, if adding base changes the [H + ] to 1.0 x then the ph is In fact equation <2> is valid for highly diluted solutions only. If concentrated solutions of acids or bases or even salts are used, the hydrogen concentration must be replaced by the ion activity a H+ and the hydroxide concentration by a OH-. The relation between concentration and activity of an ion is a ion = f ion * [ion] <3> where f is the activity coefficient for that ion. The reason for the difference of activity and concentration is that in higher concentrated solutions the ions interact with each other and therefore show a different behaviour than in diluted solutions. That means in higher concentrated solutions the amount of real active ions is lower than expected. This leads to the ph definition ph = -log a H+ <4> 1.2 HOW TO MEASURE ph An electrochemical cell for ph measurement always consists of an indicating electrode whose potential is directly proportional to ph, a reference electrode whose potential is independent of ph, and the aqueous sample to be measured. If all three parts are in contact with each other, a potential can be measured between the indicating electrode and the reference electrode, which depends on the ph of the sample and its temperature. Because of the complexity of a ph measurement, the combination of indicating and reference electrode must be calibrated in advance, to compensate for slight changes over time. See chapter calibration. The relation between measured potential E (mv), ph and temperature (K) is the Nernst equation. 3
4 E (mv) P a g e 4 R T E( T ) E ( T ) log a H <5> M F R T E ( T ) E ( T ) ph <6> M F E(T) : measured potential mv at temperature T (Kelvin) E (T) : constant, standard potential mv at temperature T (Kelvin) R : molar gas constant ( J mol -1 K -1 ) M : factor to convert Ln to Log (2.303) F : Faraday constant C mol -1 T : temperature (Kelvin) Therefore at a given (constant) temperature the potential of a solution depends on the ph only. Equation <6> can be seen as the standard formula for straight lines Y = a + b X, where a is the offset and b is the slope of the line. In case of equation <6> the offset is E (T) and the slope is the R temperature dependent factor - T = -T M F At 25 C the slope gives ( ) = mv/ph slope = -54 T= 5 C slope = -58 T=20 C slope = -62 T=40 C slope (mv/ph) calibration with buffers ph 4.01 ph 7.00 ph ph Figure (1): Influence of temperature changes on the Nernst slope of a ph calibration. 1,3 CALIBRATION The system of ph indicating electrode, reference electrode, ph-meter and lab conditions (stirring / no stirring, thermostated / not thermostated, ambient air humidity / pressure, and so on) is calibrated by placing the electrodes in solutions of known ph (buffers) and measuring the voltage of the cell. Because the cell potential is a linear function of ph (usually in the range of ph 2-11), two calibration points (2 ph buffers) are needed. Most common is the calibration with the IUPAC buffers ph and ph (@ 25 C). If the ph measurements are done in extreme areas of ph < 2 or ph > 11 we recommend to calibrate specifically in the low ph range (buffer ph ph 4.01) or in the high ph range (buffer ph ph 12.45). The calibration parameters are offset (or zero ph) and the slope. Many ph meters calculate the slope as a percentage of the theoretical value, which at 25 C is mv/ph. For example, if the calibration slope is determined to be at 25 o C it would equal 99.3% theoretical. 4
5 P a g e 5 The following table lists theoretical slopes for different temperatures. Temperature C Slope mv/ ph GENERAL In nearly every industrial and scientific application, ph is determined by measuring the potential of an electrochemical cell. Figure 2 shows a simplified diagram of a ph cell. combined ph electrode ph 7.00 refill hole electrolyte reference element liquid junction indicating element temperature sensor ph sensitive glass bulb temperature sensor reference electrode sample ph glass electrode meter Figure (2): combined ph glass electrode Figure (3): ph measurement system A ph measurement system (figure (3) consists of a ph probe, reference probe, temperature sensor, ph meter and the sample to be measured. In most cases the three probes are combined in one electrode (figure (2). When the ph probe is in contact with a solution a potential forms between the ph probe and the reference probe (figure (4). The meter measures the potential and converts it, using the calibration curve parameters, into a ph value. ph glass electrode H + reference electrode sample potential (mv) Figure (4): development of a potential (mv) between ph probe and reference probe: ph cell 5
6 P a g e 6 The ph system (ph cell) has a high internal resistance, therefore the ph meter must have a very high input impedance. Never use a conventional volt meter to measure the potential of a ph electrode. As long as the reference electrode is stable, the ph/potential measurement will vary only with temperature and sample. However, in certain samples the liquid junction plays a role. Based on the type and concentration of ions on both sides of the junction, the liquid junction can add some mv to the measurement potential. 2.2 COMBINED ph ELECTRODE Figure (2) shows the internal components of the ph electrode. The heart of the electrode is a thin bulb of ph-sensitive glass, which is blown onto the end of a length of glass tubing. The ph-sensitive glass (glass membrane) is sealed to the electrode and contains a solution of potassium chloride at ph 7. A silver wire plated with silver chloride contacts the solution. The Ag/AgCl combination in contact with the filling solution sets an internal reference potential. This potential depends on the chloride concentration in the filling solution and as long as this electrolyte concentration is maintained, the electrode potential is constant. Electrolyte Concentration mol/l KCl saturated KCl 3,5 KCl 3 KCl 3 saturated with AgCl NH4Cl 3 Figure (5): most common electrolyte filling solutions for combined ph glass and reference electrodes. Ag wire tip coated with AgCl sat. KCl outer = ph 7 glass inner membrane glass in contact membrane in with the sample contact with KCl Figure (6): working principle of a ph glass membrane As Figure 6 shows, the outside surface of the glass membrane is in contact with the sample being measured, and the inside surface contacts the filling solution. A complex mechanism at each glassliquid interface defines the potential the ph glass electrode. While the inner ph glass / filling solution potential is constant, the outside potentials varies based on the [H + ] in the sample. This equilibrium depends also on temperature. 2.3 REFERENCE ELECTRODE As figure 7 shows, the reference electrode is a silver wire coated with silver chloride in contact with a defined electrolyte solution, see figure (5). In many reference electrodes a gel is used instead of a liquid as the internal filling. These gels also contain KCl to maintain the reference potential and add sufficient conductivity. As described in 2.2 the reference potential is constant as long as the internal electrolyte is constant. The reference potential also varies slightly with temperature. The tube containing all the reference elements and solutions/gel is in contact with the sample to measure through a junction (diaphragm). It is essential to maintain a free flow of ions through the junction. Otherwise the reference electrode will not respond properly to ph changes in the sample. 6
7 P a g e 7 Electrolyte filling Saturated KCl: a saturated solution of KCl, where additional KCl crystals are added, allow the reference system to work in a wide temperature range. At higher temperatures the additional crystals can dissolve and still maintain a saturated solution. At low temperatures more crystals form, but since the solution is already saturated, the KCl concentration is still saturated. In addition if the electrode has to measure over a longer period of time in an aqueous solution, sample liquid can go through the diaphragm inside the probe and dilute the electrolyte. A saturated solution has therefore a higher buffer capacity against dilution. 3 molar KCl: this is the most common electrode filling and available almost everywhere. 3 molar KCl saturated with AgCl: this is a way to longer maintain the Ag/AgCl reference electrode. The additional AgCl slows down the dissolution of the Ag/AgCl element. Gel: the viscosity of the gel filling helps the probe to work under higher pressure conditions than with liquid fillings. In addition the gel usually does not require a refilling or other maintenance from the user. However, because the gel usually cannot be refilled, the lifetime of a gel-filled probe is much shorter than those with liquid filling, approx. 12 months only. 2.4 LIQUID JUNCTION (salt bridge, diaphragm) The salt bridge or diaphragm or liquid junction (see figure (2) and (7) is an integral part of the reference electrode. It provides the electrical connection between the reference electrode and the sample being measured. Salt bridges have a variety of forms. Some are highly porous, where the pores are filled with ions from the filling solution and from the sample. Depending on the mechanical design of the diaphragms, the movement of material from inside the reference half cell into the porous material, and from the sample into the porous material is different. For example using a gel-filled salt bridge only ions are permitted to flow across the porous material and the liquid level of the reference compartment remains constant. Other materials allow liquid to flow out of the reference compartment, requiring periodic refilling. These types of diaphragms minimize contamination of the reference half cell, since ions do not flow back from the sample. Figure (7): variety of diaphragms and there flow behaviour The liquid junction is the most problematic part of the probe. If it clogs due to sample particles going into the porous area or because crystals of the filling solution forming inside the junction, the final reading is false or takes much too long to stabilize. If the electrolyte level inside the reference compartment is low, then another problem can arise where sample solution enters the reference compartment. This dilutes the electrolyte with sample solution and the reference potential is no longer constant. In certain cases there is blockage due to precipitates forming inside the junction. This can occur if Ag + ions come together with S 2- ions to form Ag 2 S precipitate. In samples with S 2- ion, or with biological or medical samples, it is recommended to use a reference probe with double junction, where the outer electrolyte is absolutely free of Ag + ions. 7
8 P a g e BUFFERS AND CALIBRATION Figure (1) shows the ideal ph measurement system, where the potential is zero when the ph is 7, and the slope is T over the entire ph range. A real ph system rarely has at ph 7 a potential of zero, but it is usually between -20 mv and +20 mv. The slope is also not T over the entire range of ph and normally changes at ph < 2 or ph > 11. Because ph systems are not ideal, they must be calibrated before use with ph buffers. These ph buffers have a known ph at different temperatures and cover the whole ph scale. International cooperating institutes (e.g. IUPAC = International Union of Pure and Applied Chemistry, NIST = National Institute of Standards and Technology, DIN = Deutsches Institut für Normung, JSI = Japan Standards Institute) provide the definition of the buffer solutions used for references for ph measurements. All ph measurements are referred to traceable buffer solutions so that those values can be compared worldwide. How are the calibration parameters calculated during a ph calibration? Usually two ph buffers are selected to check the ph probe performance, ph 4.01 and ph The ratio of the measured potentials (E2 - E1) to the difference of ph ( , gives the slope of the straight line. Slope S = E ph 2 2 E1 ph 1 [mv/ph] <7> Equation <7> is valid at the specific temperature of the buffer solutions. Example: ph buffer ph 4.01 and ph 7.00 are used for calibration. In buffer ph 4.01 the ph system measures a potential of 170 mv and in buffer ph 7.0 it measures a potential of -3.1 mv. 173 ( 3.1) Slope S = [mv/ph] 3.1 ph VALUES OF BUFFER SOLUTIONS AT DIFFERENT TEMPERATURES The tables below list the coefficients describing the temperature dependency for several standard buffers.the coefficients A, B, C and D refer to the formula: ph = A / T + B + C T + D T 2 <8> where T is the temperature in Kelvin. 8
9 P a g e 9 Figure (8): ph buffer solutions and the parameters to calculate the ph value for different temperatures. 3.2 PRECAUTIONS USING BUFFERS Because ph buffer solutions are always the reference of our ph measurements, the ph result can only be as good as the ph buffers used for calibration. If the buffers are contaminated or used improperly, the calibration will be false and all following measurements will be wrong. Therefore proper handling, storage and use of buffers is important. 1. Always use ph buffers which bracket the ph of your sample. If the sample has a ph of 6 use buffers ph 4 and ph 7. If the sample has a ph of 7, use buffers with ph 4 and ph 10. If an alkaline sample is measured, calibrate with buffer ph 7 and ph To shorten the stabilization time and to achieve an accurate calibration, make sure the sensor and the buffer are at the same temperature. 3. For greatest accuracy measure samples at close to the same temperature as your calibration buffers. 4. Buffers have limited shelf lives. Do not use a buffer if the expiration date has passed. Store buffers at controlled room temperature and if possible always at the same place. 5. Never return used buffer to the buffer bottle. Discard it. 6. Do not let the buffer bottle open for a longer time (exposure to air). The atmospheric carbon dioxide lowers the ph of alkaline buffers. 7. If a calibration with alkaline buffer gives suspect results first try changing the alkaline buffer. Alkaline buffers are susceptible to contamination by CO Rinse the sensor with deionised water before placing it in the buffer. Remove excess water from the sensor by gently wiping it with a clean tissue. 3.3 TEMPERATURE CORRECTION OF COMPENSATIONS OF THE SAMPLE? The ph of a solution is dependent of temperature (Nernst equation). Buffer solution temperature correction is not the same as sample temperature compensation. Automatic temperature compensation (ATC) is the automatic calculation of ph from the measured potential and temperature, using an adjusted slope to the measured temperature. The change in ph of a solution with temperature is called the solution temperature coefficient (unit is ΔpH/ C). Most standard ph buffers and some chemicals have known solution temperature coefficients (see Figure 8). The temperature variation of standard ph buffers is often printed on their bottles. As long as the solution temperature coefficient is not known for the sample solution, a solution temperature correction can not be done. If necessary the temperature coefficient can be determined manually/empirically. In order to achieve best results ATC is done in most meters (for samples with unknown temperature coefficient). In any case it is recommended to do a calibration and a sample measurement at the same temperature. 9
10 P a g e 10 ph Frequently asked questions ph measurement in samples... QUESTION: What factors need to be taken into account when measuring low ionic strength samples and deionised water? ANSWER: As a general rule, standard ph probes contain a 3 or 3.5 molar KCl electrolyte solution or a gel. As low ionic strength samples and deionised water contain little or no salt, they try to get ions (salt) out of the electrolyte solution. When conventional ph probes are used (e.g. with gel filling), this phenomenon may lead to poor response and unstable readings. For applications such as these, Hach offers ph probes specifically designed for low ionic strength samples. These probes have high reference electrolyte flow rates to insure a free flowing reference junction. When KCl crystals are added, the ionic strength can be increased and the response time improved. It is important to use high purity KCl to avoid influencing the ph. Closed sample containers (flasks) present the advantage of avoiding CO 2 contamination. Use of scrupulously clean equipment and glassware and thorough rinsing before measurement are essential to avoid contamination from previous samples. Prior to sample measurement, the probe must be calibrated by using high-precision ph buffer solutions. QUESTION: What is the best way to measure ph in soil? ANSWER: It is advisable to use a specific ph probe with a strong glass tip and high electrolyte outflow. Mix a 5 g soil sample with 25 g deionised water while stirring carefully. Then let the mixture stand without stirring for 10 minutes to separate the particles from the liquid. Insert the ph probe in such a way that the glass bulb is totally covered by particles, but the diaphragm is not. Wait for a stable reading. QUESTION: What is the best way to measure ph in high temperature samples? ANSWER: Before measuring samples over 80 C, check whether the probe is designed to stand such high temperatures. Gel-filled ph probes can be used up to 80 C but no higher. Hach ph probes have an Ag/AgCl reference system which is ideal for higher temperatures. We recommend ph probes with saturated KCl filling solution. As KCl crystals are soluble at higher temperatures, it is a good idea to ensure excess crystals so that the solution remains saturated. If at all possible lower the temperature of the sample to the temperature of the buffers at calibration. QUESTION: What is the best way to measure ph in high alkaline samples and samples with high salt content? ANSWER: We recommend ph probes with saturated KCl filling solution. As KCl crystals are soluble at higher temperatures, it is a good idea to ensure excess crystals so that the solution remains saturated. Samples with high salt content and ph>12 may be subject to sodium error. Hach/ offers high alkalinity electrodes that minimize sodium error. QUESTION: What is the best way to measure ph in emulsions or fatty solutions? ANSWER: When measuring fats and emulsions, it is important to choose an electrode with the right diaphragm and reference electrolyte and a junction that is easy to clean. We therefore recommend open liquid junctions and sleeve junctions. After the measurement, the probe should be cleaned thoroughly, i.e. any remaining fat or oil must be removed with soap or surface-active agent. 10
11 P a g e 11 Maintenance and storage of the ph electrode QUESTION: How should ph probes be stored best? ANSWER: All Hach ph probes are delivered with a plastic protection cap that can also be used for storage. Pour a few drops of saturated KCl solution into the cap to ensure that the glass membrane is kept hydrated and ready to use. For short-term storage, the ph probe can be placed in a solution of 3.5 molar KCl or ph 4.0 or ph 7.0 ph buffer. Always rinse before use. As a general rule, ph probes (especially reference probes) should never be stored in deionised water. Overnight, the probe should be stored in the corresponding electrolyte solution, usually saturated KCl. For long-term storage (2 weeks or more), the ph probe should be stored with its protection cap filled with storage solution and sealed with Parafilm. QUESTION: What is the best way to remove air bubbles from inside a ph probe? ANSWER: Air bubbles can get trapped in the electrolyte solution of the reference system. This results in unreliable and unstable readings and may make it impossible to achieve a reading. In order to remove air bubbles, hold the probe firmly by the cable and swing around so that the air bubbles move to the upper end of the probe. If air bubbles are trapped inside the solid KCl crystals, heat the electrode tip in warm water (max. 60 C). This will dissolve the crystals and release the air. Afterwards, swing the probe around once more as described above and leave it to cool down. Air bubbles inside the ph glass membrane are normal and cannot be removed. They will not cause problems if they are at the upper end of the probe. QUESTION: Can a ph probe be used straight out of the box? ANSWER: Although Hach ph probes come with a protective cap which is wet inside, the glass membrane can dry out. For best results, we recommend rinsing with deionised water then conditioning the probe in ph 4.0 buffer for at least 2 hours. After further rinsing, it is ready to be calibrated. The normal (quick) response time will be achieved after 24 hours hydration. If measurements are needed before this time, calibrations should be repeated often due to drifting potentials. QUESTION: Do dirty and wet cables have an influence on the ph reading? ANSWER: Due to the extremely small currents which pass through the ph electrode, the cable, plug and connector must be kept clean and dry if reliable measurements are to be obtained. QUESTION: How much electrolyte should the ph probe contain? ANSWER: The level of electrolyte solution should be 1 cm below the filling hole. This is the only way to ensure that the hydrostatic pressure applied to the diaphragm will be high enough to prevent sample passing through the diaphragm into the probe. QUESTION: What causes a diaphragm to get blocked and how can this be remedied? ANSWER: Liquid junctions with fibre or ceramic diaphragms can occasionally get blocked due to KCl crystallisation. Try soaking the electrode in warm tap water to dissolve the KCl crystals that cover the diaphragm. Other types of blockage can occur in the form of a precipitate, for example, silver chloride or silver sulphide. Gentle polishing with abrasive paper and soaking in saturated KCl can help. 11
12 P a g e 12 ph measurement technique QUESTION: What is the maximum cable length between the ph meter and ph probe? ANSWER: All ph electrodes have a high impedance and the mv signal is amplified by the electronics in the ph meter or, with Hach digital electrodes, in the probe head. With analogue electrodes, cable lengths of 3 m can normally be used. Digital probes work with cable lengths of up to 30 m or more, because the digital signal is not impaired by the magnetic fields of motors, e.g. pumps. QUESTION: How does the temperature influence the impedance of the ph glass membrane? ANSWER: The lower the temperature, the higher the impedance of ph glass. For every 10 C decrease in temperature, the glass impedance will increase about 2.5 times, resulting in slow response. For example, if the ph glass impedance is 100 MOhm at 25 C it will increase to 250 MOhm at 15 C. QUESTION: What is the lifetime of a ph probe? ANSWER: The lifetime of a ph electrode depends on several factors including storage conditions, correct maintenance and the type of sample measured. Under normal laboratory conditions, for aqueous samples, the average lifetime is between 12 and 18 months, supposing of course that the electrode is kept clean and kept hydrated during storage. If the probe is used with dirty samples (e.g. stirred solutions with particles), is subjected to mechanical abrasion or used at high temperature or high pressure, the lifetime may be only a few weeks. In hot alkaline solutions, ph probes can be damaged after only a few hours. Regular maintenance helps ph probes keep working efficiently for several years. QUESTION: What is the difference between ph probes with glass body and an epoxy body? ANSWER: Both electrodes have a glass membrane, a diaphragm, a reference system and offer the same measuring quality. However, epoxy electrodes have a limited temperature range of maximum +80 C, while glass bodies can withstand temperatures of C (with an Ag/AgCl reference system). If you are working in the field or in tough conditions, epoxy bodies have the advantage of being more robust and less liable to crack. Electrodes with glass bodies tend to be used in the lab, because they can be cleaned more easily and, unlike epoxy bodies, can withstand organic solvents. ph probes with epoxy bodies are usually less expensive and therefore represent a cost-effective alternative. QUESTION: What is the point of automatic temperature compensation (ATC)? ANSWER: Automatic temperature compensation (ATC) has to do with correcting the ph calibration slope to account for the actual temperature of the sample so the ph electrode gives an accurate reading of the ph of the sample. The ph calibration equation is linear and has a slope value at 25 o C. Any deviation of the actual temperature from 25 o C is compensated in the slope according to the Nernst equation. The ph of the internal solution of the ph bulb is called the isopotential ph. If the ph of the sample is the same as the isopotential ph there is essentially zero potential across the ph membrane. The isopotential ph of most ph electrodes is around ph 7. Temperature has little effect on sample measurement around the isopotential ph and becomes more important as the sample becomes more acidic or basic. ATC becomes important the further the sample ph is away from ph 7. QUESTION: Is it always necessary to perform a 2-point calibration or is a 1-point calibration sufficient? ANSWER: If the last calibration was performed on the same day, a control calibration with one buffer is sufficient. Then only the zero potential is adjusted, the old slope will remain as it was. Otherwise, a 2-point calibration is recommended, because only then an actual probe status can be determined and taken into account for the measurements to come. 12
13 P a g e 13 QUESTION: What is the recipe of IUPAC ph buffers? ANSWER: HCl (ph1.094): 0.1 M HCl, Oxalate (ph 1.679): 0.05 mol/kg KH 3 C 4 O 8, Phthalate (ph 4.005): 0.05 mol/kg KHC 8 H 4 O 4, Acetate (ph 4.650): 0.1/0.1 mol/kg C 2 H 4 O 2 /C 2 H Na, Phosphate (ph 6.865): 0.025/0.025 mol/kg KH 2 PO 4 /Na 2 HPO 4, Phosphate (ph 7.000): approx / mol/kg KH 2 PO 4 /Na 2 HPO 4, Phosphate (ph 7.413): / m KH 2 PO 4 /Na 2 HPO 4, Borate (ph 9.180): 0.01 m Na 2 B 4 O 7, Carbonate (10.012): 0.025/0.025 mol/kg NaHCO 3 /Na 2 CO 3, Ca(OH) 2 (ph 12.45): saturated (at 25 C) and filtered ph measurement: troubleshooting PROBLEM: The ph probe response is slow, tends to drift and results are not reproducible SOLUTION: This may be caused by one of the following: - the glass membrane or the diaphragm is dirty, e.g. oil, fat, paint, dirt. - the ph probe is reaching the end of its lifetime. - a sample of low ionic strength (< 100 µs/cm) is being measured with a conventional rather than a specially designed ph probe. PROBLEM: The ph measurement system can no longer be calibrated by auto-bufferrecognition SOLUTION: Check all parts of the system. First eliminate possible mechanical defects such as loose plugs, damaged cable or probe, too low a level of electrolyte etc. Ensure that fresh ph buffers were used. Buffers are only detected automatically if the mv signal is within a certain range. Old buffers or buffers not meeting DIN/IUPAC specifications cannot be detected by the meter software (AUTOCAL). For instance, ph 6.86 and ph 7.00 buffers are too close together and cannot be distinguished. In order for auto buffer recognition to work properly, the correct type of buffers has first to be selected. Typical combinations of ph buffers are ph 4, 7, 10 or ph 4.01, 6.86, Finally, the ph electrode must be cleaned using special cleaning and conditioning solutions which can be found in the GK Annex Electrode Maintenance Kit from Hach. If the ph probe still does not react normally and a slope of 95% to 102% cannot be achieved during calibration, it should be replaced. PROBLEM: The response time of the ph probe keeps increasing. SOLUTION: If the response time is gradually becoming longer, the diaphragm may be blocked or dirty, or the surface of the ph glass membrane may have fat, oil or paint deposits. In order to clean the diaphragm, use the special RENOVO.N and RENOVO.X cleaning solutions found in the GK Annex Electrode Maintenance Kit. Fat deposits can be cleaned off glass membranes by using surfactant solutions. If the glass membrane does still not react properly, it can gently be etched. Procedure for etching glass membranes The glass bulb is lowered for 1 minute into a 20% ammonium fluoride solution and then for 15 seconds into a 6 molar HCl. It should then be rinsed thoroughly with deionised water and stored for 24 hours in a slightly acidic buffer such as ph QUESTION: What is the usual stabilisation time for a ph probe in ph buffers? ANSWER: At room temperature (20-25 C) and with fresh ph buffer solutions, the stabilisation time should not exceed 1 minute; normally a stable reading should be achieved after 30 seconds. Otherwise the ph electrode should be cleaned and conditioned again. QUESTION: Is it possible to use a ph probe in alcoholic solutions? ANSWER: Yes, but short-term only. Longer exposure to high percentage alcoholic solutions leads to dehydration of the glass membrane which then has to be conditioned again. 13
PN 44-6033/rev. D December 2010. Theory and Practice of ph Measurement
PN 44-6033/rev. D December 2010 Theory and Practice of ph Measurement THEORY AND PRACTICE OF ph MEASUREMENT TABLE OF CONTENTS THEORY AND PRACTICE OF ph MEASUREMENT TABLE OF CONTENTS Section Title Page
ph Measurement and Control
ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
This value, called the ionic product of water, Kw, is related to the equilibrium constant of water
HYDROGEN ION CONCENTRATION - ph VALUES AND BUFFER SOLUTIONS 1. INTRODUCTION Water has a small but definite tendency to ionise. H 2 0 H + + OH - If there is nothing but water (pure water) then the concentration
Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC
Common Problems with Online Water Quality Analyzers Greg Macy Director AquatiPro LLC Objectives Diagnosing Colorimetric Analyzers Identifying Common Problems with Sodium Analyzers Troubleshooting ph Analyzers
SenTix 950 SenTix 980
SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the
Features, Limitations & Potential Errors
MINIMIZING USER ERRORS IN ph MEASUREMENTS Features, Limitations & Potential Errors of the Tools of ph Measurements Advanced Sensor Technologies, Inc. 603 North Poplar Street Orange CA 92868-1011 USA General
A Potentiometric Analysis of Fluoride Ion in Toothpaste
CHEM 311L Quantitative Analysis Laboratory Revision 2.0 A Potentiometric Analysis of Fluoride Ion in Toothpaste In this laboratory exercise, we will analyze a toothpaste sample for its Fluoride Ion (F
CONDUCTIVITY SENSOR. Description D0382. Figure 1. The Conductivity Sensor
CONDUCTIVITY SENSOR Description D0382 Figure 1. The Conductivity Sensor Short Description The Conductivity Sensor can be used to measure either solution conductivity or total ion concentration of aqueous
ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND
#3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric
Lab 7 Soil ph and Salinity OBJECTIVE INTRODUCTION Soil ph active
Lab 7 Soil ph and Salinity OBJECTIVE In this lab you will learn the effect of the concentration of hydrogen ions (ph) and various salts on the fertility of a soil. You will perform some tests which are
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide
Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3
Measuring ph in Water or CaCl2 Using a ph Meter Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 PREPARING THE ph METER FOR USE (Calibration and Buffer Solutions) 4 When is Calibration
Experiment 8 - Double Displacement Reactions
Experiment 8 - Double Displacement Reactions A double displacement reaction involves two ionic compounds that are dissolved in water. In a double displacement reaction, it appears as though the ions are
USEPA Electrode Method Method 8156
ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method
Measuring the Conductivity of Pure Water Using USP<645>
Measuring the Conductivity of Pure Water Using USP Water Analysis Instruments, Thermo Fisher Scientific Application Note 002 Key Words Water conductivity, ultrapure water (UPW), USP, water quality,
ph: Measurement and Uses
ph: Measurement and Uses One of the most important properties of aqueous solutions is the concentration of hydrogen ion. The concentration of H + (or H 3 O + ) affects the solubility of inorganic and organic
The Effects of Temperature on ph Measurement
The Effects of Temperature on ph Measurement Authors: John J. Barron Colin Ashton & Leo Geary Technical Services Department, Reagecon Diagnostics Ltd, Shannon Free Zone, County Clare, Ireland Abstract
9. Analysis of an Acid-Base Titration Curve: The Gran Plot
9. Analysis of an Acid-Base Titration Curve: The Gran Plot In this experiment, you will titrate a sample of pure potassium hydrogen phthalate (Table 10-4) with standard NaOH. A Gran plot will be used to
ph Alkalinity of Water
ph Alkalinity of Water DOC316.52.93085 Based on ISO standard 9963-1:1994 ph-metric Titration 0.4 to 20 mmol/l of Total Alkalinity 1. Introduction Alkalinity of water is its acid-neutralizing capacity.
ph Theory and Practice
ph Theory and Practice - 1 - - when you need to be sure... Preface The importance of ph Many of nature's processes are highly dependent on ph. This is also the case for the chemical reactions which take
ph Measurements of Common Substances
Chem 100 Section Experiment 10 Name Partner s Name Introduction ph Measurements of Common Substances The concentration of an acid or base is frequently expressed as ph. Historically, ph stands for the
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL 430 350/REV A/10-03 MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL CONTENTS Introduction 1 Specification 2 Installation 3 Displays 3-4
High Purity Water Resistivity/ Conductivity Measurement
CONDUCTIVITY MEASUREMENT IN HIGH PURITY WATER SAMPLES below 10 µsiemens/cm High Purity Water Resistivity/ Conductivity Measurement Ultra-Pure Water without any chemical impurities will still have a conductivity
white paper reviews why the chlor-alkali process is so damaging to ph electrodes and discusses a sensor design that has
INGOLD White Paper Leading Process Analytics No More Tears Elevated temperatures, presence of chlorine and high salt concentration throughout the chlor-alkali process make stable and reliable ph measurement
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation Introduction The importance of keeping cooling tower water in proper chemical
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
Acid Dissociation Constants and the Titration of a Weak Acid
Acid Dissociation Constants and the Titration of a Weak Acid One of the most important applications of equilibria is the chemistry of acids and bases. The Brønsted-Lowry acid-base theory defines an acid
Determining the Identity of an Unknown Weak Acid
Purpose The purpose of this experiment is to observe and measure a weak acid neutralization and determine the identity of an unknown acid by titration. Introduction The purpose of this exercise is to identify
To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq)
Weak Acid Titration v120413 You are encouraged to carefully read the following sections in Tro (2 nd ed.) to prepare for this experiment: Sec 4.8, pp 158-159 (Acid/Base Titrations), Sec 16.4, pp 729-43
Operator Quick Guide EC SENSOR
Operator Quick Guide EC SENSOR Revision G - 24/04/2008 General Information About This Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:
PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA
Acid Base Titrations
Acid Base Titrations Introduction A common question chemists have to answer is how much of something is present in a sample or a product. If the product contains an acid or base, this question is usually
EXPERIMENT # 3 ELECTROLYTES AND NON-ELECTROLYTES
EXPERIMENT # 3 ELECTROLYTES AND NON-ELECTROLYTES Purpose: 1. To investigate the phenomenon of solution conductance. 2. To distinguish between compounds that form conducting solutions and compounds that
Ion Selective Electrodes
Ion Selective Electrodes OAKTON has a large selection of Ion Selective Electrodes () to suit a wide variety of applications. Each electrode has a typical response time of 20 to 30 seconds but will vary
Resolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or
Experiment 4 (Future - Lab needs an unknown)
Experiment 4 (Future - Lab needs an unknown) USING A ph TITRATION TO DETERMINE THE ACID CONTENT OF SOFT DRINKS 2 lab periods Reading: Chapter 9, 185-197; Chapter 10, pg 212-218; Chapter 14 pg 317-323,
Volumetric Analysis. Lecture 5 Experiment 9 in Beran page 109 Prelab = Page 115
Volumetric Analysis Lecture 5 Experiment 9 in Beran page 109 Prelab = Page 115 Experimental Aims To prepare and standardize (determine concentration) a NaOH solution Using your standardized NaOH calculate
Experiment 7: Titration of an Antacid
1 Experiment 7: Titration of an Antacid Objective: In this experiment, you will standardize a solution of base using the analytical technique known as titration. Using this standardized solution, you will
Chemistry B11 Chapter 6 Solutions and Colloids
Chemistry B11 Chapter 6 Solutions and Colloids Solutions: solutions have some properties: 1. The distribution of particles in a solution is uniform. Every part of the solution has exactly the same composition
The Theory of ph Measurement
Application Data Sheet ADS 43-002/rev.C November 2010 Theory The Theory of ph Measurement ph measurement has a wide variety of industrial applications in virtually every industry. These range from water
Handbook. Thermo Scientific. ph Electrode Selection Handbook
Handbook Thermo Scientific ph Electrode Selection Handbook Table of Contents ph Theory... 4 ph Electrode Theory... 4 ph versus Temperature Theory... 5 ph Electrode Selection Guide... 6 Thermo Scientific
Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table
Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination
CHEMICAL DETERMINATION OF EVERYDAY HOUSEHOLD CHEMICALS
CHEMICAL DETERMINATION OF EVERYDAY HOUSEHOLD CHEMICALS Purpose: It is important for chemists to be able to determine the composition of unknown chemicals. This can often be done by way of chemical tests.
CYCLE CHEMISTRY ph MEASUREMENT
CYCLE CHEMISTRY ph MEASUREMENT David M. Gray and Edward P. Santini Thornton Associates, Waltham MA Electric Utility Chemistry Workshop Champaign, Illinois May 12-14, 1998 Sponsored by University of Illinois
EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion
EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1 Introduction This lab is designed for you to discover the properties of electrochemical cells. It requires little previous knowledge of electrochemical
105 Adopted: 27.07.95
105 Adopted: 27.07.95 OECD GUIDELINE FOR THE TESTING OF CHEMICALS Adopted by the Council on 27 th July 1995 Water Solubility INTRODUCTION 1. This guideline is a revised version of the original Guideline
Chemistry 51 Chapter 8 TYPES OF SOLUTIONS. A solution is a homogeneous mixture of two substances: a solute and a solvent.
TYPES OF SOLUTIONS A solution is a homogeneous mixture of two substances: a solute and a solvent. Solute: substance being dissolved; present in lesser amount. Solvent: substance doing the dissolving; present
ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.
Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE
UltraBasic Benchtop ph Meter Operation Manual
UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.
ph 200 ph-/redox-/temperature Meter Operating instructions
ph 200 ph-/redox-/temperature Meter Operating instructions 2 Table of Contents Page 1. Scope of delivery...4 2. System description...4 3. Connections...5 4. Display...7 5. Keypad...7 6. Start up...8 7.
Neutralizing an Acid and a Base
Balancing Act Teacher Information Objectives In this activity, students neutralize a base with an acid. Students determine the point of neutralization of an acid mixed with a base while they: Recognize
Separation by Solvent Extraction
Experiment 3 Separation by Solvent Extraction Objectives To separate a mixture consisting of a carboxylic acid and a neutral compound by using solvent extraction techniques. Introduction Frequently, organic
AIRMoN, ph and Conductivity Field Measurement
AIRMoN, ph and Conductivity Field Measurement Items needed: Sample to be measured (sample volume 50 ml) Field Observer Form (FOF) for the sample to be measured conductance cell ph vials, 4 conductivity
User Guide. 9708 Dissolved Oxygen Probes
User Guide 9708 Dissolved Oxygen Probes Introduction The Thermo Scientific Orion 9708 dissolved oxygen probe simplifies measurements of dissolved oxygen, particularly Biochemical Oxygen Demand (BOD).
Inferred ph in Steam Plant Water Chemistry Monitoring
Application Data Sheet ADS 4900-87/rev.B January 2009 Power Industry Inferred ph in Steam Plant Water Chemistry Monitoring INTRODUCTION Inferred ph means ph calculated from straight and cation conductivity.
Bluelab Soil ph Meter Instruction Manual
Instruction Manual www.getbluelab.com Table of Contents 1.0 Information about measuring the ph of soils/media. 2 2.0 Introduction to Bluelab Soil ph Meter 4 3.0 Preparing the Bluelab Soil ph Meter for
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL 970 350/REV A/10-03 MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1 Installation 2 Displays 2-3 Controls 4 Preparation
GA/7 Potentiometric Titration
Rev. 7/99 7-1 INTRODUCTION GA/7 Potentiometric Titration The potentiometric titration is a useful means of characterizing an acid. The ph of a solution is measured as a function of the amount of titrant
6 H2O + 6 CO 2 (g) + energy
AEROBIC RESPIRATION LAB DO 2.CALC From Biology with Calculators, Vernier Software & Technology, 2000. INTRODUCTION Aerobic cellular respiration is the process of converting the chemical energy of organic
Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes
Tech Note Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes Probe Care and Maintenance Periodic cleaning and membrane changes are required to keep your sensors operating properly. Follow
DISSOLVED OXYGEN PROBE
51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe
The role of CO 2 in pool water
The role of CO 2 in pool water A series of e-mail articles from the research group onbalance, January 2006 The role of CO 2 in pool water #1 While some service techs go about their business taking care
The Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:
BOECO PORTABLE PH/ORP/TEMP METER MODEL PT-380 A Simultaneous displays ph or mv and temperature. A Convenient calibration with automatic buffer recognition of European and US buffer sets. Calibration data
SEPARATION OF A MIXTURE OF SUBSTANCES LAB
SEPARATION OF A MIXTURE OF SUBSTANCES LAB Purpose: Every chemical has a set of defined physical properties, and when combined they present a unique fingerprint for that chemical. When chemicals are present
experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration.
81 experiment5 LECTURE AND LAB SKILLS EMPHASIZED Synthesizing an organic substance. Understanding and applying the concept of limiting reagents. Determining percent yield. Learning how to perform a vacuum
Figure 1. A voltaic cell Cu,Cu 2+ Ag +, Ag. gas is, by convention, assigned a reduction potential of 0.00 V.
Voltaic Cells Introduction In this lab you will first prepare a set of simple standard half-cells and then measure the voltage between the half-cells with a voltmeter. From this data you will be able to
Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
Chapter 16: Tests for ions and gases
The position of hydrogen in the reactivity series Hydrogen, although not a metal, is included in the reactivity series because it, like metals, can be displaced from aqueous solution, only this time the
Electrical Conductivity of Aqueous Solutions
Electrical Conductivity of Aqueous Solutions PRE-LAB ASSIGNMENT: Reading: Chapter 4.-4.3 in Brown, LeMay, Bursten & Murphy.. Using Table in this handout, determine which solution has a higher conductivity,.
REDOX AND REFERENCE ELECTRODE USER MANUAL
REDOX AND REFERENCE ELECTRODE USER MANUAL 3 Redox and reference electrode manual Copyright 2012 Unisense A/S Version October 2012 REDOX AND REFERENCE ELECTRODE MANUAL UNISENSE A/S TABLE OF CONTENTS WARRANTY
Care and use guide. Contents. page
Care and use guide Contents page Features Bluelab Combo Meter Introduction Preparing for use Calibration Changing nutrient and temperature display units 6 Measuring hydroponic elements 7 Battery replacement
Trace Dissolved Oxygen Sensor
Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION
Professional Plus. Calibration Tips
Professional Plus Calibration Tips Table of Contents Introduction...1 Calibration Worksheet...1 Temperature...3 Calibration Tips...3 Troubleshooting Tips...3 Conductivity...4 Calibration Tips...4 Troubleshooting
Gel-filled ph Electrode
51935-88 Gel-filled ph Electrode Models 51935-00, 51935-11, 51935-22 Hach Company, 1999-2001. All rights reserved. Printed in the U.S.A. ap/dh 2/99 1ed te/dk 02/01 2ed 2 TABLE OF CONTENTS SPECIFICATIONS...5
MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1
MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry EXPERIMENT #5 THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1 I. PURPOSE OF THE EXPERIMENT In this experiment
Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor
Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process
Chem101: General Chemistry Lecture 9 Acids and Bases
: General Chemistry Lecture 9 Acids and Bases I. Introduction A. In chemistry, and particularly biochemistry, water is the most common solvent 1. In studying acids and bases we are going to see that water
AP FREE RESPONSE QUESTIONS ACIDS/BASES
AP FREE RESPONSE QUESTIONS ACIDS/BASES 199 D A chemical reaction occurs when 100. milliliters of 0.200molar HCl is added dropwise to 100. milliliters of 0.100molar Na 3 P0 solution. (a) Write the two net
HOW TO MAKE STANDARD SOLUTIONS FOR CHEMISTRY
HOW TO MAKE STANDARD SOLUTIONS FOR CHEMISTRY Phillip Bigelow Chemists make two common types of "standard solutions": Molar solutions Normal solutions Both of these solutions are concentrations (or strengths
Experiment 9 Electrochemistry I Galvanic Cell
9-1 Experiment 9 Electrochemistry I Galvanic Cell Introduction: Chemical reactions involving the transfer of electrons from one reactant to another are called oxidation-reduction reactions or redox reactions.
Solubility Product Constant
Solubility Product Constant Page 1 In general, when ionic compounds dissolve in water, they go into solution as ions. When the solution becomes saturated with ions, that is, unable to hold any more, the
Experiment 16-Acids, Bases and ph
Definitions acid-an ionic compound that releases or reacts with water to form hydrogen ion (H + ) in aqueous solution. They taste sour and turn litmus red. Acids react with certain metals such as zinc,
Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
Physical Properties of a Pure Substance, Water
Physical Properties of a Pure Substance, Water The chemical and physical properties of a substance characterize it as a unique substance, and the determination of these properties can often allow one to
Dissolved Oxygen Sensors
Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit
Chemical versus Physical Changes
Chemical versus Physical Changes Permission to Copy - This document may be reproduced for non-commercial educational purposes Copyright 2009 General Electric Company What are physical and chemical changes?
PART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
Extraction: Separation of Acidic Substances
Extraction: Separation of Acidic Substances Chemists frequently find it necessary to separate a mixture of compounds by moving a component from one solution or mixture to another. The process most often
User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
General Chemistry I (FC, 09-10) Lab #3: The Empirical Formula of a Compound. Introduction
General Chemistry I (FC, 09-10) Introduction A look at the mass relationships in chemistry reveals little order or sense. The ratio of the masses of the elements in a compound, while constant, does not
The most common active ingredient used in deodorants is aluminium chlorohydrate. But not all deodorants contain aluminium chlorohydrate:
Engineeringfragrance make a deodorant practical activity 2 student instructions page 1 of 5 chemical compounds The most common active ingredient used in deodorants is aluminium chlorohydrate. But not all
EXPERIMENT 7 Reaction Stoichiometry and Percent Yield
EXPERIMENT 7 Reaction Stoichiometry and Percent Yield INTRODUCTION Stoichiometry calculations are about calculating the amounts of substances that react and form in a chemical reaction. The word stoichiometry
stoichiometry = the numerical relationships between chemical amounts in a reaction.
1 REACTIONS AND YIELD ANSWERS stoichiometry = the numerical relationships between chemical amounts in a reaction. 2C 8 H 18 (l) + 25O 2 16CO 2 (g) + 18H 2 O(g) From the equation, 16 moles of CO 2 (a greenhouse
Q.1 Classify the following according to Lewis theory and Brønsted-Lowry theory.
Acid-base A4 1 Acid-base theories ACIDS & BASES - IONIC EQUILIBRIA 1. LEWIS acid electron pair acceptor H, AlCl 3 base electron pair donor NH 3, H 2 O, C 2 H 5 OH, OH e.g. H 3 N: -> BF 3 > H 3 N BF 3 see
Understanding hydrogen ion concentration (ph)
World Bank & Government of The Netherlands funded Training module # WQ - 06 Understanding hydrogen ion concentration (ph) New Delhi, May 1999 CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas, New
Petri Dish Electrolysis Electrolysis Reactions
elearning 2009 Introduction Petri Dish Electrolysis Electrolysis Reactions Publication No. 95008 Electrolysis is defined as the decomposition of a substance by means of an electric current. When an electric