General Purpose ph/orp Sensor
|
|
|
- Daniella Leonard
- 10 years ago
- Views:
Transcription
1 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 General Purpose ph/orp Sensor Specifications Figure Insertion/Submersion Sensor Sensor Type: General purpose 3900 Measured Range: ph AccuGlass 0-14 ORP to mv Percent Linearity Over ph Range: Range Linearity % % % % % % % Operating Temperature: -10 to 100 C (14 to 212 F) Automatic temperature compensation -10 to 100 C (14 to 212 F) Maximum Pressure: 790 kpa [abs] (100 psig) at 100 C (212 F) CRN rating: 60 psig up to 100 C (212 F). Materials of Construction: stainless steel, glass, Teflon, polyphenylene sulfide (PPS), and EPDM Materials of ORP: Platinum Process Connections: Front facing: 3 4 and 1 MNPT Rear facing: 1 MNPT Integral Cable: 32 ft (10m) cable with integral SMART preamp; 15 ft (4.7m) cable without preamp sensor VP8 Cable: use XX, 2.5 ft (.8m) to 100 ft (31m) Weight/Shipping Weight: 0.45 kg/0.9 kg (1 lb/2 lb) OPERATIONAL NOTE The 3900 Sensor responds to changes in ph at a minimum conductivity of 0.1mS/cm in deionized water. Sample flow rate must be controlled to 2 gph (7.6L/hr). The offset is approximately -0.2 ph after 3 months at ambient temperature. CAUTION: SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsibility of the user. WARNING Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process temperature is lowered to a safe level!
2 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 CAUTION The solution used during calibration is an acid and should be handled with care. Follow the directions of the acid manufacturer. Wear the proper protective equipment. Do not let the solution come in contact with skin or clothing. If contact with skin is made, immediately rinse with clean water. ATEX DIRECTIVE: Special Conditions for safe use 1. All ph/orp sensors have a plastic enclosure which must only be cleaned with a damp cloth to avoid the danger due to a build up of an electrostatic charge. 2. All ph/orp sensor models are intended to be in contact with the process fluid and may not meet the 500V r.m.s. a.c. test to earth. This must be taken into consideration at installation. Storage 1. It is recommended that electrodes be stored in their original shipping containers until needed. 2. Do not store at temperatures below -10 C (14 F). 3. Electrodes should be stored with a protective cap containing KCl solution (PN ). 4. For overnight storage, immerse the sensor in tap water or 4 ph buffer solution. 5. ph glass electrodes slowly deteriorate in storage. There is no specific expiration date. However, the calibration procedure described below should be followed to determine that the sensor calibrates properly. Electrode preparation 1. Remove electrode from shipping container. 2. Remove the protective boot covering the electrode bulb. 3. Rinse away salt film with clean water, then shake the electrode so that the internal solution fills the bulb, thus removing any air trapped there. Installation For sensor orientation, see Figure 2. For sensor dimensions, see Figure 3. For mounting options, see Figures 4-7. For wiring, see Figures For additional wiring information, please visit our website at Wrap the sensor threads with six or seven (6 7) turns of Teflon tape to prevent leakage. Do not over tighten the sensor into it s receptacle. Hand tighten the sensor, and then tighten one or two (1 2) turns with a wrench. Two point buffer calibration Select two stable buffer solutions, preferably ph and 10.0 (ph buffers other than ph 4.0 and ph 10.0 can be used as long as the ph values are at least two ph units apart). NOTE: A ph 7 buffer solution reads a mv value of approx. zero, and ph buffers read approximately ± 59.1 mv for each ph unit above or below ph 7. Check the ph buffer manufacturer specifications for millivolt values at various temperatures since it may affect the actual value of the buffer solution mv/ph value. 1. Immerse sensor in the first buffer solution. Allow sensor to equilibrate to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/ transmitter. 2. Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water. 3. Repeat steps 1 and 2 using the second buffer solution. 4. The theoretical slope value, according to the Nernst equation for calculating ph, is approximately 59.1 mv/ph. Over time the sensor will age, both in the process and in storage, which will result in reduced slope values. To ensure accurate readings, it is recommended that the electrode be replaced when the slope value falls below 47 to 49 mv/ph. Recommended ph Sensor Standardization For maximum accuracy, the sensor can be standardized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sensor s slope but will simply adjust the analyzer s reading to match that of the known process ph.
3 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Maintenance for ph electrode Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement. 1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent. 2. To remove scale deposits, soak electrodes for 1 to 5 minutes in a 5% hydrochloric acid solution. 3. Temperature effect on life expectancy: If glass electrode life expectancy is 100% at 25 C (77 F), then it will be approximately 25% at 80 C (176 F), and approximately 10% at 100 C (212 F). CAUTION Hydrochloric acid is toxic and highly corrosive. Avoid skin contact, wear protective gloves. Use only in a well ventilated area. Do not inhale fumes. In case of an accident, consult a doctor immediately. ORP Calibration 1. After making an electrical connection between the sensor and the instrument, obtain a standard solution of saturated quinhydrone PN R508-8OZ (475mV). This can also be made quite simply by adding a few crystals of quinhydrone to either ph 4 or ph 7 buffer. Quinhydrone is only slightly soluble, therefore only a few crystals will be required. 2. Immerse the sensor in the standard solution. Allow 1-2 minutes for the ORP sensor to stabilize. 3. Standardize the instrument to the solution value shown in the table, right. The resulting potentials, measured with a clean platinum electrode and Figure 2. Sensor Orientation saturated KCl/AgCl reference electrode, should be within +/- 20 millivolts of the value shown in the table below. Solution temperature must be noted to ensure accurate interpretation of results. The ORP value of saturated quinhydrone solution is not stable over long periods of time. Therefore, these standards should be made up fresh each time they are used. 4. Remove the sensor from the buffer, rinse, and install in the process. Maintenance for ORP Electrode Electrodes should respond rapidly. Sluggishness, offsets, and erratic readings are indicators that the electrodes may need cleaning or replacement. 1. To remove oil deposit, clean the electrode with a mild non-abrasive detergent. 2. To remove scale deposits, soak electrodes for 1 to 5 minutes in a 5% hydrochloric acid solution. 3. ORP (metallic) electrodes should be polished with moistened baking soda. Figure 3. Sensor Dimensions 3/4 IN MNPT 1 IN MNPT WRENCH FLATS 1.30 IN ACROSS 1 IN MNPT TEMP COMP SOLN GND REFERENCE JUNCTION ph ELECTRODE
4 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 3. Sensor Dimensions 3/4 IN MNPT 1 IN MNPT WRENCH FLATS 1.30 IN ACROSS 1 IN MNPT TEMP COMP SOLN GND REFERENCE JUNCTION ph ELECTRODE Figure 4. Typical Flow Through insertion installation using PN Pipe Tee (sensor must be installed at least 10 above the horizon) Horizontal Pipe Tee (PN ) Pressure/Temperature Ratings psig (kpa [abs]) F ( C) 150 (1136) 150 (65) 128 (984) 160 (71) 102 (805) 170 (77) 80 (653) 180 (82) 57 (494) 200 (93) 48 (432) 210 (99) Figure 5. Low Flow Cell PN INCH MILLIMETER Inlet and outlet connections are stainless steel and take 1/4-inch OD tubing. Flow cell is polycarbonate with 1/4-inch FNPT fittings. Wetted materials: Body & Nut Polyester/Polycarbonate Fittings 316 SST Seals Silicone Flow cell ratings: Temperature: F (0-70 C) Max. Pressure: 90 PSIG (721 kpa [abs]) Flow rate: 2-5 GPH ( LPH) 4
5 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 6. Jet Spray Cleaner PN (sensor must be installed at least 10 above the horizon) The Jet Spray Cleaner eliminates routine, manual sensor maintenance by cleaning the sensor with water or compressed air. Flow through the cleaner can be controlled by a solenoid valve. NOTE: The Jet Spray Cleaner can be used with Handrail Mounting Assembly (PN , not shown) or can be mounted through conduit as shown below. Jet Spray Cleaner shown with 3900VP ph sensor Figure 7. Low Flow Panel PN SQP10077-LQD Low Flow Panel Specifications Inlet flow Inlet pressure Temperature 3-80 gph ( L/hr) 3 65 psig ( kpa abs)* F (0-50 C) *The minimum inlet pressure is required to open a check valve, which prevents the flow cell from draining if sample flow is lost. Removing the check valve lowers the inlet pressure requirement to a few feet of water head. 5
6 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 NOTE For additional wiring information on this product, including sensor combinations not shown here, please refer to either our online wiring programs or the Manual DVD enclosed with each product. 1056, 1057, 56, 5081, 6081, 54e, and XMT: and sensors with SMART preamps: Wiring_Diagram.aspx 1055: Figure and 3900VP-01 to 54epH/ORP Figure and 3900VP-02 to 54epH/ORP 6
7 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 10. Sensor Wiring 3900VP-01 gray cable to 54epH/ORP Figure 11. Sensor Wiring 3900VP-02 gray cable to 54epH/ORP 7
8 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure and 3900VP-01 to Xmt Figure and 3900VP-02 to Xmt 8
9 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 14. Sensor Wiring 3900VP-01 with gray cable to XMT Figure 15. Sensor Wiring 3900VP-02 with gray cable to Xmt 9
10 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure and 3900VP-01 to 1055 Figure 17. Two-Input and 3900VP-01 to
11 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure and 3900VP-02 to 1055 Figure 19. Two-Input and 3900VP-02 to
12 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 20. Wiring with Remote Junction Box and Preamp (PN ) to and 3900VP-02 Figure pH/pH Wiring through Remote Junction Boxes to and 3900VP-02 12
13 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure Sensors, Preamplifier in Junction Box ID (inner drain) (No Connection Green White White/Red Red White/Gray Gray Blue Clear Orange 3900 Sensors, Preamplifier in Junction Box For and 3900VP-02 sensors Non-preamp sensors Blue Cable ID (inner drain) Connect to instrument ground or as shown 3 White (RTD Return) 3 RTD RTN /RTN 4 White/Red (RTD Sense) 4 RTD SENSE /SNS 5 Red (RTD in) 5 RTD IN 6 White/Gray (Reference Shield) 6 Ref SHLD / SHLD /GUARD 7 8 Gray (Reference in) 7 REF / REF IN Blue (Solution Ground) 8 GND/SOL GND 9 9 White/Black ( mv/ph Shield) ph SHLD / SHLD / GUARD 10 Black ( mv/ph in) 10 ph/ph in TB 1 BNC No Connection TB 2 Brown (-5VDC) Green (+5 VDC) Extension Cable PN or shown Extension Cable PN can also be used; extend color scheme from sensor or VP8 cable instead of using shown wiring scheme. -5VDC / -V SNSR +5VDC / +V SNSR Rosemount Analytical intruments: 56 / 1056 / 1057 / 1066 / 6081 Remote Junction Box PN or Sensor Head Junction Box PN (Both J-boxes include Preamplifier Board PN ) epH/ORP XMT 13
14 General Purpose ph/orp Sensor October 2014 Figure Sensors, Preamplifier in Sensor Instruction Sheet LIQ_MAN_ABR_3900 Green White White/Red Red White/Gray Gray Blue Clear Orange 3900 Sensors, Preamplifier in Sensor For and 3900VP-01 sensors Preamplifier in sensor Blue Cable Green Connect to instrument ground or as shown White (RTD Return) RTD RTN /RTN White/Red (RTD Sense) Red (RTD in) RTD SENSE /SNS RTD IN Gray (Reference in) Blue (Solution Ground) Clear ( mv/ph Shield) Orange ( mv/ph in) White/Gray (-5V) ID (inner drain) (+5V) Ref SHLD / SHLD /GUARD REF / REF IN GND/SOL GND ph SHLD / SHLD / GUARD ph/ph in -5VDC / -V SNSR +5VDC / +V SNSR Rosemount Analytical intruments: 56 / 1056 / 1057 / 1066 / epH/ORP , 3400HTVP-70, 10 XMT
15 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure Sensors, Preamplifier in Sensor 3900 Sensors, Preamplifier in Sensor, Junction Box used for extending cable For and 3900VP-02 sensors Preamplifier in sensor Blue Cable Green White White/Red Red Gray Blue Clear (Orange shield) Orange White/Gray ID (inner drain) TB TB 2 Remote Junction Box PN (with Extension board for point-to-point wiring) Inner Drain White (RTD Return) White/Red (RTD Sense) Red (RTD in) White/Gray (No connection) Gray (Reference in) Blue (Solution Ground) White/Black (mv/ph Sheild) Black ( mv/ph in) Brown (-5VDC) Green (+5 VDC) Extension Cable PN or shown. Connect to instrument ground or as shown Extension Cable PN can also be used; White/gray is replacement wire color for Brown wire. RTD RTN /RTN RTD SENSE /SNS RTD IN Ref SHLD / SHLD /GUARD REF / REF IN GND/SOL GND ph SHLD / SHLD / GUARD ph/ph in -5VDC / -V SNSR +5VDC / +V SNSR Rosemount Analytical intruments: 56 / 1056 / 1057 / 1066 / epH/ORP XMT 15
16 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure and 3900VP-01 to Models 5081, 4081 or 3081 Figure and 3900VP-02 to Models 5081, 4081 or
17 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 10. Wiring for and 3900VP-01 to 56, 1056 & 1057 ph Figure 11. Wiring for and 3900VP-02 to 56, 1056 & 1057 ph 17
18 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 26. Sensor Wiring 3900VP-01 with gray cable to 1056/56 Figure 27. Sensor Wiring 3900VP-02 with gray cable to 1056/56 18
19 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 28. Sensor Wiring 3900VP-01 with gray cable to 1057 Figure 29. Sensor Wiring 3900VP-01 with gray cable to
20 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 30. ph/orp Sensor Wiring 1066/3900VP-01 (blue cable refer to manual for recommended cable order) Figure 31. ph/orp Sensor Wiring 1066/3900VP-02 (gray cable refer to manual for recommended cable order) 20
21 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 32. ph/orp Sensor Wiring 1066/3900VP-01 (refer to manual for recommended cable order) Figure 33. ph/orp Sensor Wiring 1066/3900VP-01 (blue cable refer to manual for recommended cable order) 21
22 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure 34. ph/orp Sensor Wiring 1066/3900VP-02 (refer to manual for recommended cable order) Figure 35. ph/orp Sensor Wiring 1066/3900VP-02 (blue cable refer to manual for recommended cable order) 22
23 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 36. Sensor Wiring 3900VP-01 with gray cable to 5081 Figure 37. Sensor Wiring 3900VP-01 with gray cable to
24 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Figure and 3900VP-01 to 6081-P Figure and 3900VP-02 to 6081-P 24
25 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Figure 40. Sensor Wiring 3900VP-01 with gray cable to 6081 Figure 41. Sensor Wiring 3900VP-02 with gray cable to
26 General Purpose ph/orp Sensor October 2014 Instruction Sheet LIQ_MAN_ABR_3900 Ordering Information The 3900/3900VP ph/orp Sensor is housed in a PSS plastic body with built-in solution ground for advanced diagnostics. A Pt-100 RTD is offered for temperature compensator. The sensor is available with an integral cable or VP8 connector. Junction box kits with preamplifiers can be ordered separately if an analyzer/ transmitter with an integral preamplifier is not located within 15 ft. of the sensor. SMART preamplifier (-01) is the standard preamplifier option, and is compatible with models XMT, 3081, 5081, 6081, 54e, 1055 and 1056 analyzers/transmitters ph/orp Sensor - New Generation general purpose sensor Code Preamplifier Option 01 Preamplifier 02 No preamplifier Code Measuring Electrode 10 General Purpose Low resistivity GPLR ph Glass (0-14pH) 12 Platinum ORP EXAMPLE VP ph/orp Sensor - New Generation general purpose sensor with VP Code Preamplifier Option 01 Preamplifier 02 No preamplifier Code Measuring Electrode 10 General Purpose Low resistivity GPLR ph Glass (0-14pH) 12 Platinum ORP EXAMPLE 3900VP
27 Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 Accessories Part Number Description ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector ft. cable with mating VP8 connector Jet cleaner Insertion adapter Low flow cell Junction Box with Preamplifier for Models 54e, 3081, 4081, 5081, XMT, 1055, CPVC flow through Tee, 1-1/2 in. NPT process connection Sensor handrail assembly Buffer solution, 4.01 ph, 16oz Buffer solution, 6.86 ph, 16oz Buffer solution, 9.18 ph, 16oz Extension cable, gray Extension cable, blue SQP10077-LQD Low Flow Control Panel 27
28 LIQ_MAN_ABR_3900 Rev. E October 2014 facebook.com/emersonrosemountanalytical AnalyticExpert.com twitter.com/raihome youtube.com/user/rosemountanalytical 8 Credit Cards for U.S. Purchases Only. Emerson Process Management 2400 Barranca Parkway Irvine, CA USA Tel: (949) Fax: (949) RosemountAnalytical.com 2014 Rosemount Analytical, Inc. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
General Purpose ph/orp Sensors
General Purpose ph/orp Sensors Product Data Sheet LIQ_PDS_3900 / Rev. F December 2014 General Purpose ph/orp Sensors SMART preamplifier-stores calibration and other data. Extended sensor life provided
ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.
Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE
ENDURANCE Conductivity Sensors
Instruction Manual 400 and 400VP LIQ_MAN_ABR_400_400VP December 2013 ENDURANCE Conductivity Sensors For additional information, please refer visit our website at www.rosemountanalytical.com. CAUTION Sensor/Process
Submersion/Insertion Toroidal Sensor
Instruction Sheet PN 51A-228/rev.P December 2010 Model 228 Submersion/Insertion Toroidal Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS
Submersion/Insertion Toroidal Sensor
Instruction Manual 228 Sensor LIQ_MAN_ABR_228 May 2014 PN 51A-228 Submersion/Insertion Toroidal Sensor For additional information, please visit our website at www.rosemountanalytical.com CAUTION SENSOR/PROCESS
Trace Dissolved Oxygen Sensor
Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION
Toroidal Conductivity Sensor
Instruction Sheet PN 51A-/rev.C June 2012 Toroidal Conductivity Sensor For additional information, please visit our website at www.emersonprocess.com/rosemountanalytical.com SPECIFICATIONS Wetted Materials:
Ozone Sensor. Model 499A OZ CAUTION CAUTION. Instruction Sheet PN 51A-499AOZ/rev.F January 2007
Instruction Sheet PN 51A-499AOZ/rev.F January 2007 Model 499A OZ Ozone Sensor For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit our website at www.emersonprocess.com/raihome/liquid/.
For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/.
Instruction Sheet PN 51A-499AOZ/rev.H September 2010 Model 499A OZ Ozone Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS APPLICATION
Sanitary and CIP Toroidal Sensor
Instruction Sheet PN 51A-225/rev.I December 2010 Model 225 Sanitary and CIP Toroidal Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS
Dissolved Oxygen Sensors
Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit
PART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
Digital inductive conductivity transmitter
Digital inductive conductivity transmitter Optimal solution for conductivity measurements in difficult fluids (polluted, dirty,...) PEEK/PPA version for CIP applications Large range of process connections
TB2CS Conductivity sensor 2-Electrode. Superior accuracy in low level conductivity measurement ranges
Data sheet TB2CS Conductivity sensor 2-Electrode Superior accuracy in low level conductivity measurement ranges Measurement range from 0 to 199.9 µs/cm and 0 to 19.99 ms/cm Resolution to 0.001 µs/cm in
PRESSURE AND TEMPERATURE SENSORS for use with the BURNERLOGIX Burner Management System and PPC4000 Series Fuel/Air Ratio Controller
BLZPTS-1 NOVEMBER 19, 2013 TS350 TS752 BLPS PRESSURE AND TEMPERATURE SENSORS for use with the BURNERLOGIX Burner Management System and PPC4000 Series Fuel/Air Ratio Controller DESCRIPTION The BLPS, TS350
Instruction Manual PN 51-381/rev.E March 2010 Model 381. ph/orp Sensor
Instruction Manual PN 51-381/rev.E March 2010 Model 381 ph/orp Sensor ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests its products to meet
ph Measurement and Control
ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components
Multi-Parameter Electrochemical / Optical Water Quality System
Product Data Sheet PDS 71-WQS/rev.H September 2011 Model WQS Multi-Parameter Electrochemical / Optical Water Quality System COMPLETE SYSTEM INCLUDES sensors, cable, and instruments. MEASURES ph, ORP, Conductivity,
Combination ph/redox (ORP) Sensors
Data Sheet SS/AP00 Issue 4 Combination ph/redox (ORP) Sensors AP00 Series Long life large PTFE junction provides resistance to fouling double junction with gelled electrolyte provides better-defined junction
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
OXIGENO DISUELTO. Dissolved Oxygen. Install & forget. Sistema autolimpiable. Autoclean system
Sistema autolimpiable Dissolved Oxygen Autoclean system Applications: - water treatment - activated sludge - de-nitrification - fish pond Install & forget For special applications, the D.O monitor can
Powers Controls TH 192 HC Heating/Cooling Room Thermostat
Powers Controls TH 192 HC Heating/Cooling Room Thermostat Technical Instructions Document No. 155-066P25 TH 192-2 50 60 70 80 70 TH0356R1 60 80 POWERS Description The TH 192 HC thermostats are proportional
TWO-WIRE TRANSMITTERS FOR ph, ORP, CONDUCTIVITY, OXYGEN, OZONE, AND CHLORINE
Product Data Sheet 71-5081/rev.K April 2006 Model 5081 TWO-WIRE TRANSMITTERS FOR ph, ORP, CONDUCTIVITY, OXYGEN, OZONE, AND CHLORINE Model 5081 Family of Two-wire Transmitters CHOICE OF COMMUNICATION PROTOCOL:
MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter
Thermal Mass meter and Temperature Transmitter Description The MTI10 insertion thermal mass flowmeter and the MTL10 in-line thermal mass flowmeter provide accurate mass flow measurement of clean, dry gases
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
Sanitary Type Electronic Flow Switches
Industries Pharmaceutical Sanitary Type Electronic Flow Switches Food and Beverage Biotechnology Semiconductor Water Treatment S Series Ameritrol, Inc. Instruments and Controls Features No Moving Parts
HG600/HG800. Hydro-Power Unit. Installation and Operation Guide
HG600/HG800 Hydro-Power Unit Installation and Operation Guide Copyright 04 HydroSpin, All rights reserved. The data contained in this document is proprietary to HydroSpin. It is disclosed to the receiving
Powers Controls Free Energy Band TH 193 HC Heating/Cooling Room Thermostat
Document No. 155-068P25 TH 193-4 Powers Controls Free Energy Band TH 193 HC Heating/Cooling Room Thermostat 50 60 70 80 70 TH0356R1 60 80 POWERS Description The TH 193 HC thermostats are proportional dual
Powers Controls TH 192 S Single Temperature Room Thermostat
Technical Instructions Document No. 155-065P25 TH 192-1 Powers Controls TH 192 S Single Temperature Room Thermostat 50 60 70 80 TH0355R1 70 80 POWERS Description The TH 192 S thermostats are proportional
Self Purging Liquid Trap (SPLT ) User s Manual
Self Purging Liquid Trap (SPLT ) User s Manual High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can be released unexpectedly and with extreme force. High
VOYAGER 570G. 744A Sprayer Control
VOYAGER 570G 744A Sprayer Control U S E R M A N U A L U S E R M A N U A L Table of Contents CHAPTER 1 - INTRODUCTION...1 SYSTEM CONFIGURATIONS...1 KIT CONTENTS...3 CONTROL HOUSING ASSEMBLY...5 CHAPTER
BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS
BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS PLEASE READ COMPLETE INSTRUCTIONS BEFORE YOU BEGIN LEAVE INSTALLATION INSTRUCTIONS AND USER'S GUIDE WITH OWNER ALL ELECTRIC WIRING AND PLUMBING MUST BE DONE
User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
Electronic Flow Switches
Industries Electronic Flow Switches Petrochemical Refining Oil Production Water Treatment Pharmaceutical Food and Beverage Pulp and Paper Power Production Gas Processing Mining Biotechnology Semiconductor
Level Plus. Magnetostrictive Liquid-Level Sensors with Temposonics Technology. M-Series Model MR Transmitter with Analog Output.
Level Plus Magnetostrictive Liquid-Level Sensors with Temposonics Technology M-Series Model MR Transmitter with Analog Output Data Sheet SENSORS Document Part Number 550677 Revision J Features 4 to 20
BLUE RIBBON CORP. BC001 Birdcage Installation Manual
Product Overview: This manual is applicable for the BC001 Birdcage Submersible Level Transmitters. If the product you have has a different prefix part number than listed above please contact the factory
Replacement parts. WM97+ gas-fired water boiler Boiler Manual. OBTAIN PARTS ONLY THROUGH WEIL-McLAIN THE BOILER CONTAINS CERAMIC FIBER MATERIALS
Replacement parts DO NOT SERVICE THE BOILER WITHOUT A WM97+ MAINTENANCE KIT AVAILABLE Failure to adhere to these guidelines can result in severe personal injury, death or substantial property damage. The
Baumann 87000 Flexsleev Sanitary Control Valve
Product Bulletin D103345X012 Baumann 87000 Flexsleev Sanitary Control Valve The Baumann 87000 control valve (figures 1 and 2) is excellent for throttling high purity liquid or gaseous media commonly found
Product Data. Model 575 Series Submersible Level Transmitters ELECTRONIC PRESSURE MEASUREMENT PRODUCTS DESCRIPTION FEATURES APPLICATIONS
EPP - 1 DECRIPTION The odel 575 is used in general water/wastewater applications. The transmitter is offered in a standard non-clogging snub nose end, flush diaphragm or protected flush diaphragm with
Stainless Steel Single and Dual Circulation Kits
Instruction Sheet P/N 160780 01 Stainless Steel Single and Dual Circulation Kits Introduction The single and dual high-pressure circulation kits allow you to vary and control the circulation rate of coating
UltraBasic Benchtop ph Meter Operation Manual
UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.
2 WAY (2/2) AND 3 WAY (3/2) BRASS AND STAINLESS STEEL SOLENOID VALVES
2 WAY (2/2) AND 3 WAY (3/2) BRASS AND STAINLESS STEEL SOLENOID VALVES DIRECT ACTING Brochure N-289 SPECIFICATIONS Port Size: 1/8 and1/4 Solenoid Valve for air, oil, water and other fluids Flow Direction:
Introduction. Safety Guidelines. Part Numbers. Operation and Maintenance Manual. Operation and Maintenance Manual
Operation and Maintenance Manual Introduction Read all instructions thoroughly. Installation of the OilTector must comply with all Federal, State and Local Codes, Regulations and Practices. The OilTector
Instruction Manual LIQ-MAN-5081P/rev.J January 2015. Model 5081-P. Two-Wire ph/orp Transmitter
Instruction Manual LIQ-MAN-5081P/rev.J January 2015 Model 5081-P Two-Wire ph/orp Transmitter ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests
Models SFP33 & SFP32 Pressure Transmitters
SFP33 Flow-Through Pressure Transmitter SFP32 Single-Ended Pressure Transmitter DS-PR-SFP-eng April, 2016 Data Sheet Models SFP33 & SFP32 Pressure Transmitters Pressure Brooks Seal Free Pressure Measurement
Mobrey Ultrasonic. MSP900SH Level and MSP900FH Flow Transmitters. Product Data Sheet February 2015 IP2032, Rev BA
Mobrey Ultrasonic MSP900SH Level and MSP900FH Flow Transmitters Product Data Sheet February 2015 IP2032, Rev BA Non-contacting measurement with no moving parts Fast and simple to install and configure
In-Line Electronic Flow Switches
In-Line Electronic Flow Switches Industries Petrochemical Refining Oil Production Water Treatment Pharmaceutical Food and Beverage Pulp and Paper Power Production Gas Processing Mining Biotechnology Semiconductor
How To Use A Flowmeter
INLINE flowmeter for continuous flow measurement Economic integration in pipe systems without any additional piping 3-wire frequency pulse version to directly interface with PLC s (both PNP and NPN) Connection
USER MANUAL. 10 GPM Small Single Line Cryogenic Coupling
JC Carter LLC LNG Couplings USER MANUAL 10 GPM Small Single Line Cryogenic Coupling Models 50E722 Nozzle and 50E721 Receptacle User s Manual and Installation Guide! Warning: Read the User s Manual and
Micro Motion Model D and DL Coriolis Flow and Density Meters
Product Data Sheet PS-00388, Rev. F April 2013 Micro Motion Model D and DL Coriolis Flow and Density Meters Micro Motion Model D and DL sensors provide flow and density measurement for liquids, gases,
WEL Series ph/orp ELECTRODES
Iwaki America Inc. WEL Series ph/orp ELECTRODES Instruction Manual Five Boynton Road Hopping Brook Park Holliston, MA 01746 TEL: 508-429-1110 Web: www.walchem.com Notice 2016 WALCHEM, An Iwaki America
A99B Series Temperature Sensors
FANs 125, 121, 930, 930.5 Product/Technical Bulletin A99 Issue Date 0615 A99B Series Temperature s The A99B Series Temperature s are passive PTC (Positive Temperature Coefficient) sensors. The A99B sensors
Combustible Gas Transmitter 65-2400RK-05
Combustible Gas Transmitter 65-2400RK-05 Overview This detection insert describes the combustible gas transmitter. This insert also describes how to install, start up, maintain, and calibrate the transmitter.
OxyGuard. Dissolved Oxygen Probe
OxyGuard Dissolved Oxygen Probe Standard type USER MANUAL D02 Standard Probe manual gb 0903 CONTENTS 1. DESCRIPTION 1.1 The OxyGuard Probe... 1 1.3 Which kind of transmitter can be used with the OxyGuard
Baumann 24003 3 Way Bronze or Stainless Steel Control Valve
Product Bulletin D1039X012 24003 Valve Baumann 24003 3 Way Bronze or Stainless Steel Control Valve 3 Way Rugged Industrial Control Valve For Mixing and Diverting Applications The Baumann 24003 3 way control
ExStik TM ph Waterproof Meters
User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL
ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:
BOECO PORTABLE PH/ORP/TEMP METER MODEL PT-380 A Simultaneous displays ph or mv and temperature. A Convenient calibration with automatic buffer recognition of European and US buffer sets. Calibration data
Instruction Manual. Image of SP-3015 & SP-3815. Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function
Important Safeguards READ ALL INSTRUCTIONS BEFORE USE. Instruction Manual Automatic Dispensing Hot Water Pot with Reboil Function Image of SP-3015 & SP-3815 SP-3015: 3.0L SP-3815: 3.8L SP-3017: 3.0L (Stainless
Instruction Manual. Model FU20 ph/orp Combination sensor. IM 12B6J3-02E-E 7th edition
Instruction Manual Model FU20 ph/orp Combination sensor 7th edition TABLE OF CONTENTS 1. General............................................................ 1 1-1. Introduction........................................................
OPERATING INSTRUCTIONS
PIPELINE INSPECTION COMPANY LTD. OPERATING INSTRUCTIONS Wet Sponge Holiday Detectors 670,673, and MSRB Wet Sponge Holiday Detectors Portable and In-Plant Detectors Table of Contents General Information.......................3
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation Introduction The importance of keeping cooling tower water in proper chemical
How To Use A Cdm250 Digital Multimeter
User Manual CDM250 Digital Multimeter 070-6736-03 Copyright Tektronix, Inc. 1987. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this
SenTix 950 SenTix 980
SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the
3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE
4 3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE A. Changing the Control Board B. Replacing a Heating Element C. Thermistor Replacement D. High Limit Switch Replacement E. Level Detector Replacement
OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S
Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and
20000068 WIRELESS REMOTE ASSEMBLY, 24VDC, 1 TRANS,1 REC
20000068 WIRELESS REMOTE ASSEMBLY, 24VDC, 1 TRANS,1 REC Fitment to MP-25 620 CR 4841, Haslet, TX 76052 Ph 817.439.1108 Fax 817.636.5675 www.machine-technologies.com Kit contents 1 Transmitter 2 button,
Installation Instructions / Warranty
Installation Instructions / Warranty Vertical Spray Bidet Fitting Rough 1322X181 Citterio 39224xx0 Citterio 39225xx0 Starck 10220xx0 Steel 35220800 Uno 38220xx0 Terrano 37224xx0 Terrano 37225xx0 Carlton
Y7230A CO 2 Alarm System
Y7230A CO 2 Alarm System FOR CO 2 GAS DETECTION PRODUCT DATA FEATURES APPLICATION Y7230A CO 2 alarm system is used to detect gas leaking from CO 2 beverage dispensing systems and posts an alarm to occupants
Installation Instructions 6028.801
DAZZLE Installation Instructions 08.80 Spread Lavatory Faucet with Speed Connect Drain* Congratulations on purchasing your American Standard faucet with Speed Connect drain, a feature found only on American
450TOC Analyzer. Portable, Real-time, Continuous Total Organic Carbon Measurement. 450TOC Analyzer
450 Analyzer 450 Analyzer Fast, Continuous Measurement Compliant with USP, Ch P, EP and JP USB Data Logging and Printer Output Robust Portable Design Portable, Real-time, Continuous Total Organic Carbon
Installing a Probe. Typical Installation
for hazardous locations. Prior to any installation in a Classified Hazardous Location, verify installation methods by the Control Drawing referenced on the product s name tag 2. The Feed Through Assembly
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA &
CEILING CASSETTE FAN COIL UNIT
CEILING CASSETTE FAN COIL UNIT Operation and Installation Manual MHCFC4W-04, 08, 12, 16 Four way ceiling cassette fan coil unit Please read this manual before using the fan coil unit. Installation and
Flow Sensors. Convenient, Reliable and Economical Flow Measurement. THORNTON Leading Pure Water Analytics
THORTO Leading Pure Water nalytics Economical Paddlewheel Hi-Purity PF Ultrasonic Vortex Insertion-type Vortex Convenient, Reliable and Economical Flow Measurement Paddlewheel Reliable and Economical Inline
USE &CARE GUIDE. Remote Faucet Pump System. See Important Safeguards on page 2
Remote Faucet Pump System USE &CARE GUIDE See Important Safeguards on page 2 An exclamation point within an equilateral triangle is intended to alert user to the presence of important operating and maintenance
SDX Submersible Depth Transmitter User Manual
SDX Submersible Depth Transmitter User Manual January 2011 USER INFORMATION Stevens makes no warranty as to the information furnished in these instructions and the reader assumes all risk in the use thereof.
The Theory of ph Measurement
Application Data Sheet ADS 43-002/rev.C November 2010 Theory The Theory of ph Measurement ph measurement has a wide variety of industrial applications in virtually every industry. These range from water
OPL BASIC. Dosing System for Professional Laundry machines. Contents
OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page
ATL Fuel Level Sender Probes
T E C H N I C A L S P E C I F I C A T I O N The ATL EL-AD-151 (Resistance Output) and EL-AD-152 (Voltage Output) Fuel Level Senders are highly advanced sensors for continuously measuring the contents of
Professional Plus. Calibration Tips
Professional Plus Calibration Tips Table of Contents Introduction...1 Calibration Worksheet...1 Temperature...3 Calibration Tips...3 Troubleshooting Tips...3 Conductivity...4 Calibration Tips...4 Troubleshooting
AT500 Magnetostrictive Level Transmitter. Compact magnetostrictive liquid level transmitter for direct insertion K-TEK Products
Data sheet DS/AT500-EN Rev. M AT500 Magnetostrictive Level Transmitter Compact magnetostrictive liquid level transmitter for direct insertion K-TEK Products Features Mounts from Top of Tank High Resolution
Installation Instructions
Installation Instructions Built-In Dishwasher If you have questions, call 1-800-4-MY-HOME (1-800-469-4663) or visit our website at: www.sears.com BEFORE YOU BEGIN Read these instructions completely and
Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes
Tech Note Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes Probe Care and Maintenance Periodic cleaning and membrane changes are required to keep your sensors operating properly. Follow
RPS 8000. High Accuracy Resonant Pressure Sensor. GE Measurement & Control Solutions. Features:
GE Measurement & Control Solutions RPS 8000 High Accuracy Resonant Pressure Sensor For over 35 years, Druck has manufactured precision pressure sensors with a capability to meet critical applications in
Resolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or
Arecont Vision H.264 Color or Day/Night SurroundVideo Series Installation Manual
0 P a g e H.264 Color or Day/Night SurroundVideo Installation Manual Inside the box: A. Arecont Vision SurroundVideo camera B. Mounting template C. RJ45 female to female coupler D. Hex key E. Security
Technical Update TAA.TU.11093 Rev. 1
http://www.gambro.com/en/usa_tech/ 800-525-2623 303-222-6500 Technical Update TAA.TU.11093 Rev. 1 Effective: 05 APR 2013 CO# 13084 Product: Subject: From: Phoenix Dialysis System Required Electrical Safety
Advantium 2 Plus Alarm
ADI 9510-B Advantium 2 Plus Alarm INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501 Harpers Road Virginia
Pressure transmitter For general industrial applications Model A-10
Electronic pressure measurement Pressure transmitter For general industrial applications Model A-10 WIKA data sheet PE 81.60 for further approvals see page 9 Applications Machine building Shipbuilding
G9MXE. Product Specifications
Up to 96% AFUE, Single Stage, ECM Gas Furnace EASIER TO SELL Up to 96% AFUE, all models all positions California NOx approved Select models earn the ENERGY STAR (See below, indicated with check mark.)
Wind Direction Smart Sensor (S-WDA-M003)
(S-WDA-M003) The Wind Direction smart sensor is designed to work with HOBO Stations. The smart sensor has a plug-in modular connector that allows it to be added easily to a HOBO Station. All sensor parameters
RAY-MAX Integrated Solar Power Strip
RAY-MAX Integrated Solar Power Strip 600008, 600009, 600010, 600208, 600209, 600210 Owner s Manual NEXTRONEX, INC. Revision Date: 10/27/14 Contents 1. Safety Instructions... 3 2. General Equipment Warnings...
OEM-EP Pressure Controllers
Pressure Controllers Typical Applications Carrier Gas Pressure Control Air over Liquid Control Mass Spectrometer Process Gas Supply Pressure Control Miniature ler The Miniature ler converts a variable
AKRAPOVIC SLIP-ON EXHAUST SYSTEM for the HONDA CB1000R
Installation instructions: *502131* AKRAPOVIC SLIP-ON EXHAUST SYSTEM for the HONDA CB1000R Congratulations on purchasing an Akrapovic exhaust system. Please read these installation instructions carefully.
CW-4622 INSTALLATION INSTRUCTIONS CB-4622 RISER KIT TO SUIT BLACK LEVER EXPOSED VALVES & BATH/SHOWER VALVES
INSTALLATION INSTRUCTIONS CB-4622 RISER KIT TO SUIT BLACK LEVER EXPOSED VALVES & BATH/SHOWER VALVES CW-4622 RISER KIT WITH WHITE LEVER DIVERTER TO SUIT EXPOSED VALVES & BATH/SHOWER VALVES CB-4622 CW-4622
POWERS Controls No. 4 Pneumatic Damper Actuator
POWERS Controls No. 4 Pneumatic Damper Actuator Technical Instructions Document No. 155-032P25 AP 331-2 Actuator Assembly 331-2929 Typical Actuator Assembly 331-2904 Typical Actuator Assembly 331-3000