Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint
|
|
|
- Clifford Thomas
- 9 years ago
- Views:
Transcription
1 Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint Tomohiro Sahara, Hiroshi Tsutsui, Shigehiro Hashimoto, Shuichi Mochizuki, Kenichi Yamasaki, Hideo Kondo, Teruo Miyazaki Department of Biomedical Engineering, Osaka Institute of Technology, Osaka, , Japan and Masayasu Motozuka, Kazuo Okuyama Division of Radiological Technology, Osaka City University Hospital, Osaka, Japan and Yuichi Inoue Department of Radiology, Osaka City University Medical School, Osaka, Japan ABSTRACT An inductively coupled wireless radio frequency (RF) coil has been developed to improve the quality in the magnetic resonance imaging. The wireless coils of a Helmholtz type and a birdcage type were manufactured, and the resonance frequency was tuned at 64 MHz. Several magnetic resonance images were taken on a phantom, on a rabbit s head and on a human knee joint with the RF coils. The signal to noise ratio (SNR) has been compared at the acquired images. The experimental results show that the SNR is improved with application of the inductive coupling between the RF coils. Keywords: Biomeasurement, Magnetic Resonance Imaging, Radio Frequency Coil, Signal to Noise Ratio, Inductive Coupling 1. INTORDUCTION Magnetic Resonance Imaging (MRI) is a computed tomography, which being used a nuclear magnetic resonance phenomenon [1]. It is useful tool for diagnostic medical imaging, since it is a non-invasive tool to study both anatomy and function of human body. The radio frequency coils, known as RF (radio frequency) resonators or RF probes, are the key components in a magnetic resonance scanner [2]. They pick up the radio frequency waves which being radiated inside the human body in the strong static magnetic field. The RF coils have been studied to improve their capabilities by many researchers. Arrayed divided coils were designed to decrease noise in the Magnetic Resonance Image, for example, because noise increases with the size of the coil [3]. In the previous study, an Archimedean spiral coil for MRI was designed and the electric character
2 was simulated [4, 5]. In the RF coil, an electric noise increases with its dimension, which decreases the signal to noise ratio (SNR). To maintain the quality of magnetic resonance images, several factors have been investigated in many trials. The RF coils are classified into two types: a surface type and a volume type. The surface coil has an advantage to increase local resolution, where the volume coil has an advantage to get the uniform signal field. The volume coil has a cylindrical shape to generate a uniform magnetic field and is called birdcage coil. The uniformity is disturbed, however, when an asymmetrical subject like a knee is inserted. The inductive coupling in radio frequency probe is one of the factors to improve the image [6-10]. In the present study, a wireless coil is inserted in the birdcage coil to improve the SNR in MRI. An inductively coupled wireless RF coil with the birdcage coil has been developed in whole-body magnetic resonance (MR) scanners in the present study. 2. METHODS Inductive Coupling Dual Coils System An inductive coupling dual coils system was designed (Figs. 1 & 2) [11]. The system includes two coils: a birdcage coil and a wireless coil of a cylindrical type or of a Helmholtz type. Secondary probe Primary probe Fig. 1: Inductively coupling RF probe primary secondary Fig. 2: Equivalent circuit of dual coils. Fig. 3: Animal test. Secondary probe Primary probe Animal Test The RF probe of the cylinder with 280mm diameter and with 280 mm height was applied for the primary coil, which is conventionally used for human head. To generate a uniform magnetic field at resonance, a wireless probe of the secondary coil was manufactured in a cylindrical type, which is the same kind of shape as the primary coil. The coil consists of eight elements along the lateral surface of the plastic cylinder. The secondary coil has dimensions of 120 mm diameter and 100 mm height to fit rabbit s head (Figs. 3 & 4). The secondary coil was located at the centre of the primary coil. Magnetic resonance images were taken on a rabbit s head under anesthesia in RF probe with and without the wireless secondary probe, respectively. T1-weighted and T2-weighted images at spin-echo sequence (TR/TE=4000 ms/96 ms) were compared between two types of RF coil arrangements. Resolution of the image
3 was 0.39 mm 0.39 mm 3.0 mm voxel. Phantom Study The frequency of resonance was adjusted to 64 MHz, which corresponds to 1.5 teslas of a MR system (Siemens Magnetom Vision) (Fig. 6). Fig. 6 shows MR system with the volume coil, which was applied to the phantom. To minimize the shift of resonance frequency with the inserted wireless coil, the reactance component is neutralized with capacitance of high voltage proof. Several magnetic resonance images were taken on a cylindrical phantom of copper sulfate of 50 mm diameter and 150mm height (Fig. 7). The signal to noise ratio (SNR) of the image was compared about the position of the coils: a birdcage alone, longitudinal and vertical. Fig. 8 shows two positions in the phantom study: longitudinal and vertical. Human Knee Another coil of a Helmholtz type (Fig. 5) was manufactured in the dimension of 120 mm diameter to be applied to the human knee, where the cylindrical type is hard to fit. Magnetic resonance images were taken on a human knee of a volunteer. Signal to Noise Ratio The methodology to calculate SNR (R) from I (the mean intensity of the signal in the region of interest) and B (the standard deviation of the mean intensity in the background) in equation (1) is based on the National Electric Manufactures Association (NEMA) standards. Fig. 4: Secondary probe in animal test. Fig. 6: Magnetic resonance scanner. Main Static fields Secondary probe Secondary probe Parallel Main Static fields Vertical Fig. 5: Two types of position between primary and secondary probes: parallel and vertical.
4 t L L Fig. 7: Helmholtz coil (wireless coil) L=120 mm, t=10 mm. R = I / B (1) The standard deviation in the background is calculated on the image composed from subtraction signal of two images with the same parameter. 3. RESULTS Fig. 8: Rabbit s head; T2-weighted image, top; T1-weighted image, bottom; without wireless coil, left; with wireless coil, right Fig. 8 shows the image of the sagittal section of rabbit. The experimental results on rabbit s head show that SNR with the secondary wireless coil is higher than that without the wireless coil (Table 1). Fig. 9 shows the proton density image of the longitudinal section of the phantom column at each position of coils in the phantom experiment. Table 2 shows SNR calculated from each image of phantom. The experimental results with the phantom show that SNR with the additional wireless coil winded around the phantom is two times higher than that without the wireless coil (one coil). Fig. 10 shows the proton density image of the human knee in each position of coils. Fig. 11 shows five regions to measure SNR in the image around the articular cartilage of the knee. Fig. Table 1: Signal to noise ratio in animal test Fig. 9: Phantom study: parallel, left; vertical, right: phantom, bottom right. 12 shows SNR calculated from the image at each region in three positions of coils. The SNR Head with wireless Head alone Improvement T1weighted image T2weighted image
5 Table 2: Signal to noise ratio in phantom test. Phantom Head alone Pallarel Vertical SNR results in proton images of human knee show that SNR in the vertical position is almost 60 double that with the birdcage alone in most of Vertical regions. 50 The results show that the developed inductively coupled coil improves the signal to noise ratio. 4. DISCUSSION Signal to noise ratio Birdcage alone Horizontal The articular cartilage of the knee joint is easy to be degenerated with aging. Taking the magnetic resonance image of the articular Number of region (I 1 -I 5 in Fig.6) B (A) (B) (C) Fig. 10: Proton density images of knee. (A) birdcage alone, (B) wireless vertical, (C) wireless horizontal. I 1 I 2 Fig. 11: Regions to measure SNR in proton density image of knee I 3 I 5 I 4 Fig. 12: Signal to noise ratio in each region (I1-I5 in Fig. 11) cartilage, however, is not easy, because the area includes only small amount of proton. Maximizing the signal to noise ratio of the radio frequency probe is effective to improve the quality of the image. Helmholtz coil is selected for the wireless coil to get uniform magnetic field among the space between two circles of the coil. The electric current in the birdcage coil increases by inductive coupling with the wireless coil, which improves signal detection. The variation of resonance frequency with inductive coupling between coils should be considered, because it affects SNR. The additional inductively coupled wireless radio frequency coil can be simply applied to conventional radio frequency coils to improve the quality of magnetic resonance images. The volume ratio of the subject to the inner space surrounded by the coil is called filling factor. Maximized filling factor increases signal to noise ratio in magnetic resonance image with the volume type of the radio frequency coil. Experimental results on
6 rabbit s head show that a small cylindrical secondary coil is effective to increase the filling factor. A wireless coil of flexible material should be easy to adjust the subject and improve the filling factor. Moderate inductive coupling between primary and secondary coils generates resonance, and increase electric current in the primary coil, which might improve detection efficiency. 5. CONCLUSION The experimental results show that the developed inductively coupled Wireless Radio Frequency coils are effective to improve the signal to noise ratio on magnetic resonance images and to be applied to the human knee joint. 6. ACKNOWLEDGMENT This work was supported in part by a Grant-in-Aid for Scientific Research from the Japanese Ministry of Education, Culture, Sports, Science and Technology. REFERENCES [1] P. C. Lauterbur, Image formation by induced local interactions: examples employing nuclear magnetic resonance, Nature, Vol. 242, 1973, pp [2] C-N. Chen, D. I. Hoult, Biomedical Magnetic Resonance Image Technology, Institute Physics Publishing, [3] P.B. Roemer, W.A. Edelstein, C.E. Hayes, S.P. Souza, O.M. Mueller, The NMR Phased Array, Magn Reson Med, Vol. 16, 1990, pp [4] T.Sahara, S.Hashimoto, H.Tsutsui, Y.Morita, K.Yamasaki, H.Kondo, T.Furukawa, W.Ikeda, M.Motozuka, K.Okuyama and Y.Inoue, Development of radio frequency pick-up Archimedean spiral coil for magnetic resonance imaging, Proceedings of 11th International Conference on Information Systems Analysis and Synthesis & 2nd International Conference on Cybernetics and Information Technologies, Systems and Applications, vol. 1, 2005, pp [5] T. Sahara, S. Hashimoto, H. Tsutsui, Y. Morita, S. Mochizuki, K. Yamasaki, H. Kondo, Simulation on Newly Designed Radio Frequency Pick-Up Archimedean Spiral Coil for Magnetic Resonance Imaging, Proc. 10th World Multiconference on Systemics Cybernetics and Informatics, Vol. 4, 2006, pp [6] A. Staubert, M. Vester, V.M. Tronnier, W. Renz, C.R. Wirtz, M.M. Bonsanto, S. Kunze, Interventional MRI-guided brain biopsies using inductively coupled surface coils, Magn Reson Med, Vol. 43, 2000, pp [7] DI. Hoult, B. Tomanek, Use of mutually inductive coupling in probe design, Concepts in Magnetic Resonance (Magnetic Resonance Engineering) Vol. 15, 2002, pp [8] H. Graf, P. Martirosian, F. Schick, M. Grieser, M.E. Bellemann, Inductively Coupled RF Coils for Examinations of Small Animals and Objects in Standard Whole-Body MR Scanners, Medical Physics, Vol. 30, No. 6, 2003, pp [9] M. Bilgen, Inductively-Overcoupled Coil Design for High Resolution Magnetic Resonance Imaging, BioMedical Engineering OnLine, Vol. 5, No. 3, [10] T. Sahara, S. Hashimoto, H. Tsutsui, S. Mochizuki, K. Yamasaki, H. Kondo, E. Yamada, K. Akazawa, K. Okuyama and Y. Inoue, Development of Inductively Coupled Wireless Radio Frequency Coil for Magnetic Resonance Scanners, Proc Inaugural IEEE International Conference on Digital Ecosystems and Technologies (IEEE-DEST 2007), 2007, pp [11] K. Henney, Radio Engineering Handbook, 5th ed., New York, McGraw-Hill, 1959.
GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time
Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs
E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE
E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter
Breast MRI Quality Control
Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Discuss the importance of breast MRI quality control (QC). Provide an overview of the new ACR Breast MRI Accreditation
Design of an U-slot Folded Shorted Patch Antenna for RF Energy Harvesting
Design of an U-slot Folded Shorted Patch Antenna for RF Energy Harvesting Diponkar Kundu, Ahmed Wasif Reza, and Harikrishnan Ramiah Abstract Novel optimized U-slot Folded Shorted Patch Antenna (FSPA) is
NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706
(revised 4/21/03) NUCLEAR MAGNETIC RESONANCE Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706 Abstract This experiment studies the Nuclear Magnetic Resonance of protons
Impedance Matching and Matching Networks. Valentin Todorow, December, 2009
Impedance Matching and Matching Networks Valentin Todorow, December, 2009 RF for Plasma Processing - Definition of RF What is RF? The IEEE Standard Dictionary of Electrical and Electronics Terms defines
5 Factors Affecting the Signal-to-Noise Ratio
5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed
Exposure Assessment of a High-Resonance Wireless Power Transfer System under the Misaligned Condition
, pp.11-15 http://dx.doi.org/10.14257/astl.2015.116.03 Exposure Assessment of a High-Resonance Wireless Power Transfer System under the Misaligned Condition SangWook Park 1, ByeongWoo Kim 2, BeomJin Choi
ELECTRON SPIN RESONANCE Last Revised: July 2007
QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron
BARE PCB INSPECTION BY MEAN OF ECT TECHNIQUE WITH SPIN-VALVE GMR SENSOR
BARE PCB INSPECTION BY MEAN OF ECT TECHNIQUE WITH SPIN-VALVE GMR SENSOR K. Chomsuwan 1, S. Yamada 1, M. Iwahara 1, H. Wakiwaka 2, T. Taniguchi 3, and S. Shoji 4 1 Kanazawa University, Kanazawa, Japan;
An equivalent circuit of a loop antenna.
3.2.1. Circuit Modeling: Loop Impedance A loop antenna can be represented by a lumped circuit when its dimension is small with respect to a wavelength. In this representation, the circuit parameters (generally
The performance improvement by ferrite loading means - increasing, - increasing of ratio, implicitly related to the input impedance.
3.2.3. Ferrite Loading Magnetic ferrite loading can enhance a transmitting signal as high as 2 to 10 db for MHz [Devore and Bohley, 1977]. There is an optimum frequency range where ferrite loading is beneficial.
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
Numerical Modelling of E-M Occupational Exposures associated with MRI
Numerical Modelling of E-M Occupational Exposures associated with MRI Jeff Hand and Yan Li Imaging Sciences Dept, Imperial College London, Hammersmith Hospital Campus Physical Agents Directive (EMFs) 2004/40/EC
Edmund Li. Where is defined as the mutual inductance between and and has the SI units of Henries (H).
INDUCTANCE MUTUAL INDUCTANCE If we consider two neighbouring closed loops and with bounding surfaces respectively then a current through will create a magnetic field which will link with as the flux passes
45. The peak value of an alternating current in a 1500-W device is 5.4 A. What is the rms voltage across?
PHYS Practice Problems hapters 8- hapter 8. 45. The peak value of an alternating current in a 5-W device is 5.4 A. What is the rms voltage across? The power and current can be used to find the peak voltage,
THE ANSI/IEEE RF SAFETY STANDARD AND ITS RATIONALE. Om P. Gandhi and Gianluca Lazzi Department of Electrical Engineering Salt Lake City, Utah 84112
Introduction THE ANSI/IEEE RF SAFETY STANDARD AND ITS RATIONALE Om P. Gandhi and Gianluca Lazzi Department of Electrical Engineering Salt Lake City, Utah 84112 In 1960, the American Standards Association
THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider
THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI Douglas C. Noll* and Walter Schneider Departments of *Radiology, *Electrical Engineering, and Psychology University
Introduction to Nuclear Magnetic Resonance Spectroscopy
Introduction to Nuclear Magnetic Resonance Spectroscopy Dr. Dean L. Olson, NMR Lab Director School of Chemical Sciences University of Illinois Called figures, equations, and tables are from Principles
Current Probes, More Useful Than You Think
Current Probes, More Useful Than You Think Training and design help in most areas of Electrical Engineering Copyright 1998 Institute of Electrical and Electronics Engineers. Reprinted from the IEEE 1998
Analysis of Broadband Slot Cut Semi-Circular Microstrip Antennas
Analysis of Broadband Slot Cut Semi-Circular Microstrip Antennas Amit A. Deshmukh EXTC, DJSCOE Vile Parle (W), Mumbai, India Ankita R. Jain EXTC, DJSCOE Vile Parle (W), Mumbai, India Apurva A. Joshi EXTC,
Particle Surface Area
Particle Surface Area TM Introducing the Acorn Area TM The Acorn Area TM is a revolutionary instrument designed to measure the surface area of nanoparticles dispersed in a liquid. This patented nuclear
Simulation and Design of Printed Circuit Boards Utilizing Novel Embedded Capacitance Material
Simulation and Design of Printed Circuit Boards Utilizing Novel Embedded Capacitance Material Yu Xuequan, Yan Hang, Zhang Gezi, Wang Haisan Huawei Technologies Co., Ltd Lujiazui Subpark, Pudong Software
MINIMALLY INVASIVE THERMAL THERAPY FOR CANCER TREATMENT BY USING THIN COAXIAL ANTENNAS
MINIMALLY INVASIVE THERMAL THERAPY FOR CANCER TREATMENT BY USING THIN COAXIAL ANTENNAS K. Ito 1, K. Saito 1, T. Taniguchi 2, S. Okabe 2, and H. Yoshimura 1 1 Faculty of Engineering, Chiba University, Chiba,
Software for Design NMR Probes Using the Shielded Split Ring and the Shielded Symmetrical Band Resonators
Software for Design NMR Probes Using the Shielded Split Ring and the Shielded Symmetrical Band Resonators Nasreddine Benahmed University of Tlemcen, Algeria ABSTRACT This article presents a software (NMR
Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment
Fluid Structure Interaction VI 3 Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment J. Hengstler & J. Dual Department of Mechanical and Process
GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R
GE 3.0T 3.0T WRIST Invivo 8CH Wrist Coil Sequence Ax T2 Cor PD Cor PDFS Cor T1 Cor PD (Small FOV) FOV (mm) 80 80 80 80 40 Matrix 384x224 384x256 320x256 384x320 320x192 Phase Direction RL RL RL RL RL #
Nuclear Magnetic Resonance
Nuclear Magnetic Resonance Practical Course M I. Physikalisches Institut Universität zu Köln May 15, 2014 Abstract Nuclear magnetic resonance (NMR) techniques are widely used in physics, chemistry, and
Single Transistor FM Transmitter Design
Single Transistor FM Transmitter Design In telecommunications, frequency modulation (FM) conveys information over a carrier wave by varying its frequency. FM is commonly used at VHF radio frequencies for
Effects of Mobile Phone Radiation onto Human Head with Variation of Holding Cheek and Tilt Positions
Effects of Mobile Phone Radiation onto Human Head with Variation of Holding Cheek and Tilt Positions M. R. Iqbal-Faruque* 1, N. Aisyah-Husni 2, Md. Ikbal-Hossain 1, M. Tariqul-Islam 2 and N. Misran 2 1
Human Exposure to Outdoor PLC System
1602 PIERS Proceedings, Marrakesh, MOROCCO, March 20 23, 2011 Human Exposure to Outdoor PLC System Vicko Doric 1, Dragan Poljak 1, and Khalil El Khamlichi Drissi 2 1 University of Split, Croatia 2 Blaise
Broadband Slotted Coaxial Broadcast Antenna Technology
Broadband Slotted Coaxial Broadcast Antenna Technology Summary Slotted coaxial antennas have many advantages over traditional broadband panel antennas including much smaller size and wind load, higher
RFID Receiver Antenna Project for 13.56 Mhz Band
RFID Receiver Antenna Project for 13.56 Mhz Band Fatih Eken TE 401 Microwave Course Term Project, Fall 2004 Supervised by Asst. Prof. İbrahim Tekin Telecommunication Program in Faculty of Engineering and
a leap ahead in analog
EMV Contactless Payment Systems based on AS3911 Overview and System Simulations Giuliano Manzi, PhD Mannheim, May 23 25, 2012 CST EUROPEAN USER CONFERENCE 2012 a leap ahead in analog OUTLINE AS3911 OVERVIEW
Standex-Meder Electronics. Custom Engineered Solutions for Tomorrow
Standex-Meder Electronics Custom Engineered Solutions for Tomorrow RF Reed Relays Part II Product Training Copyright 2013 Standex-Meder Electronics. All rights reserved. Introduction Purpose Designing
Operating Frequency Selection for Loosely Coupled Wireless Power Transfer Systems with Respect to RF Emissions and RF Exposure Requirements
Operating Frequency Selection for Loosely Coupled Wireless Power Transfer Systems with Respect to RF Emissions and RF Exposure Requirements Jagadish Nadakuduti, Lin Lu, Paul Guckian Qualcomm Technologies,
ENS 07 Paris, France, 3-4 December 2007
ENS 7 Paris, France, 3-4 December 7 FRICTION DRIVE SIMULATION OF A SURFACE ACOUSTIC WAVE MOTOR BY NANO VIBRATION Minoru Kuribayashi Kurosawa, Takashi Shigematsu Tokyou Institute of Technology, Yokohama
Applications in EMC testing. Outline. Antennas for EMC Testing. Terminology
Antennas for EMC Testing Zhong Chen ETS-Lindgren 1301 Arrow Point Drive Cedar Park, TX 78613 [email protected] Outline EMC Terms and Definitions Typical EMC Antennas Calibration of EMC Antennas
Radiated Emission and Susceptibility
Radiated Emission and Susceptibility Tzong-Lin Wu, Ph.D. EMC Lab Department of Electrical Engineering National Taiwan University Differential-Mode v.s. Common-mode Currents 1 Differential-Mode v.s. Common-mode
Design of an Auxiliary Power Distribution Network for an Electric Vehicle
Design of an Auxiliary Power Distribution Network for an Electric Vehicle William Chen, Simon Round and Richard Duke Department of Electrical & Computer Engineering University of Canterbury, Christchurch,
Understanding SMD Power Inductors. Application Note. July 2011
Understanding SMD Power Inductors July 2011 Application Note Power inductors play an important role in voltage conversion applications by yielding lower core losses. They are also used to store energy,
32-Channel Head Coil Imaging at 3T
32-Channel Head Coil Imaging at 3T Thomas Benner Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA
EE302 Lesson 14: Antennas
EE302 Lesson 14: Antennas Loaded antennas /4 antennas are desirable because their impedance is purely resistive. At low frequencies, full /4 antennas are sometime impractical (especially in mobile applications).
CONCEPT-II. Overview of demo examples
CONCEPT-II CONCEPT-II is a frequency domain method of moment (MoM) code, under development at the Institute of Electromagnetic Theory at the Technische Universität Hamburg-Harburg (www.tet.tuhh.de). Overview
Musculoskeletal MRI Technical Considerations
Musculoskeletal MRI Technical Considerations Garry E. Gold, M.D. Professor of Radiology, Bioengineering and Orthopaedic Surgery Stanford University Outline Joint Structure Image Contrast Protocols: 3.0T
Tag Tuning/RFID. Application Note. Tag Tuning. Introduction. Antenna Equivalent Circuit
Tag Tuning Introduction RFID tags extract all of their power to both operate and communicate from the reader s magnetic field. Coupling between the tag and reader is via the mutual inductance of the two
Basic Wire Antennas. Part II: Loops and Verticals
Basic Wire Antennas Part II: Loops and Verticals A loop antenna is composed of a single loop of wire, greater than a half wavelength long. The loop does not have to be any particular shape. RF power can
Basic Principles of Magnetic Resonance
Basic Principles of Magnetic Resonance Contents: Jorge Jovicich [email protected] I) Historical Background II) An MR experiment - Overview - Can we scan the subject? - The subject goes into the magnet -
Clamp Filters that Suppress Emission Noise Provide Immunity Against Surge Noise
TDK EMC Technology Product Section Clamp Filters that Suppress Emission Noise Provide Immunity Against Surge Noise TDK Shonai Corporation Satoru Saito Reduce Emission Noise from Cables Even if an electronic
FDA Guidelines for Magnetic Resonance Equipment Safety
FDA Guidelines for Magnetic Resonance Equipment Safety Loren A. Zaremba, Ph.D. Center for Devices and Radiological Health Food and Drug Administration Outline I. Introduction II. Static Magnetic Field
Selecting IHLP Composite Inductors for Non-Isolated Converters Utilizing Vishay s Application Sheet
VISHAY DALE www.vishay.com Magnetics Selecting IHLP Composite Inductors for Non-Isolated Converters INTRODUCTION This application note will provide information to assist in the specification of IHLP composite
An extended EMC study of an electrical powertrain for transportation systems
European Association for the Development of Renewable Energies, Environment and Power Quality (EA4EPQ) International Conference on Renewable Energies and Power Quality (ICREPQ 12) Santiago de Compostela
Changes PN532_Breakout board
Changes PN532_Breakout board Document: Changes PN532_Breakout board Department / Faculty : TechnoCentrum - Radboud University Nijmegen Contact: René Habraken Date: 17 May 2011 Doc. Version: 1.0 Contents
MEASUREMENT SET-UP FOR TRAPS
Completed on 26th of June, 2012 MEASUREMENT SET-UP FOR TRAPS AUTHOR: IW2FND Attolini Lucio Via XXV Aprile, 52/B 26037 San Giovanni in Croce (CR) - Italy [email protected] Trappole_01_EN 1 1 DESCRIPTION...3
AN2866 Application note
Application note How to design a 13.56 MHz customized tag antenna Introduction RFID (radio-frequency identification) tags extract all of their power from the reader s field. The tags and reader s antennas
Nano Meter Stepping Drive of Surface Acoustic Wave Motor
Proc. of 1st IEEE Conf. on Nanotechnology, Oct. 28-3, pp. 495-5, (21) Maui, Hawaii Nano Meter Stepping Drive of Surface Acoustic Wave Motor Takashi Shigematsu*, Minoru Kuribayashi Kurosawa*, and Katsuhiko
Nuclear Magnetic Resonance
Nuclear Magnetic Resonance NMR is probably the most useful and powerful technique for identifying and characterizing organic compounds. Felix Bloch and Edward Mills Purcell were awarded the 1952 Nobel
There are at least six ways to go about loading a short vertical monopole.
With a mobile installation and most base locations, the HF seems to pose problems especially using the top band. 160 meters is also known as the top band. It is the only frequency group in the MF band
The Dedicated MRI Factbook. Esaote MRI: A world of difference
The Dedicated MRI Factbook Esaote MRI: A world of difference Esaote Dedicated MRI Measures of Excellence Creativity in Healthcare is Esaote s motto which represents the philosophy that has been adopted
Coupling Magnetic Signals to a SQUID Amplifier
SQUID Application Note 105-0 Coupling Magnetic Signals to a SQUID Amplifier Matching the effective inductances of the Pickup Coil and the Input Coil to detect and couple magnetic flux maximizes the sensitivity
WAVEGUIDE-COAXIAL LINE TRANSITIONS
WAVEGUIDE-COAXIAL LINE TRANSITIONS 1. Overview Equipment at microwave frequencies is usually based on a combination of PCB and waveguide components. Filters and antennas often use waveguide techniques,
Investigation of Magnetic Resonance Imaging Effects when Using Bone Conduction Implants
Investigation of Magnetic Resonance Imaging Effects when Using Bone Conduction Implants Master of Science Thesis in Biomedical Engineering, MPBME Karl-Johan Fredén Jansson Department of Signals and Systems
Electromagnetic Field Safety
SAR A Basic Restriction for Electromagnetic Field Safety David Baron, PE Consulting Engineer Electromagnetic Field Safety [email protected] NOTE. CENAM is not responsible for the content of this document.
A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS
A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS Joseph J. Stupak Jr, Oersted Technology Tualatin, Oregon (reprinted from IMCSD 24th Annual Proceedings 1995) ABSTRACT The
National Laboratory of Antennas and Microwave Technology Xidian University Xi an, Shaanxi 710071, China
Progress In Electromagnetics Research, PIER 76, 237 242, 2007 A BROADBAND CPW-FED T-SHAPE SLOT ANTENNA J.-J. Jiao, G. Zhao, F.-S. Zhang, H.-W. Yuan, and Y.-C. Jiao National Laboratory of Antennas and Microwave
FDTD Analysis of Site Free Space VSWR in Test Site Used for Disturbance Measurement above 1 GHz
九 州 工 業 大 学 学 術 機 関 リポジトリ Title FDTD analysis of site free space VS disturbance measurement above 1 GHz Author(s) Kuwabara N; Midori M; Kawabata M Issue Date 9-8 URL http://hdl.handle.net/8/368 9 IEEE.
Magnetic Field of a Circular Coil Lab 12
HB 11-26-07 Magnetic Field of a Circular Coil Lab 12 1 Magnetic Field of a Circular Coil Lab 12 Equipment- coil apparatus, BK Precision 2120B oscilloscope, Fluke multimeter, Wavetek FG3C function generator,
Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator
Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator Kenji MAWATARI, Koich SAMESHIMA, Mitsuyoshi MIYAI, Shinya MATSUDA Abstract We developed a new inkjet head by applying MEMS
6 J - vector electric current density (A/m2 )
Determination of Antenna Radiation Fields Using Potential Functions Sources of Antenna Radiation Fields 6 J - vector electric current density (A/m2 ) M - vector magnetic current density (V/m 2 ) Some problems
PHY114 S11 Term Exam 3
PHY4 S Term Exam S. G. Rajeev Mar 2 20 2:0 pm to :45 pm PLEASE write your workshop number and your workshop leader s name at the top of your book, so that you can collect your graded exams at the workshop.
104 Practice Exam 2-3/21/02
104 Practice Exam 2-3/21/02 1. Two electrons are located in a region of space where the magnetic field is zero. Electron A is at rest; and electron B is moving westward with a constant velocity. A non-zero
Amplifier for Small Magnetic and Electric Wideband Receiving Antennas (model AAA-1B)
Amplifier for Small Magnetic and Electric Wideband Receiving Antennas (model AAA-1B) 1. Description and Specifications Contents 1.1 Description 1.2 1.2 Specifications 1.3 1.3 Tested parameters in production
Spatio-Temporal Mapping -A Technique for Overview Visualization of Time-Series Datasets-
Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.603-608 (2011) ARTICLE Spatio-Temporal Mapping -A Technique for Overview Visualization of Time-Series Datasets- Hiroko Nakamura MIYAMURA 1,*, Sachiko
Experimental Study of Free Convection Heat Transfer From Array Of Vertical Tubes At Different Inclinations
Experimental Study of Free Convection Heat Transfer From Array Of Vertical Tubes At Different Inclinations A.Satyanarayana.Reddy 1, Suresh Akella 2, AMK. Prasad 3 1 Associate professor, Mechanical Engineering
Real-life Applications of ICNIRP Guidelines to Various Human EMF Exposure Issues
Technical Forum Overview and Latest Development of Standards in Human Exposure to Electromagnetic Fields (EMF) Real-life Applications of ICNIRP Guidelines to Various Human EMF Exposure Issues Dr. Brian
Functional neuroimaging. Imaging brain function in real time (not just the structure of the brain).
Functional neuroimaging Imaging brain function in real time (not just the structure of the brain). The brain is bloody & electric Blood increase in neuronal activity increase in metabolic demand for glucose
510(k) Summary. This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.
MAR, 2009 / \MLECH MRI-TECH Sp. z o.o., ul. Zielihska 3, 31-227 Krak6w, Poland MRI-TECH Canada, Inc. 206-3820 Cambie Street Vancouver BC V5Z 2X7 510(k) Summary This summary of 510(k) safety and effectiveness
Design Optimization of Printed Circuit Board Embedded Inductors through Genetic Algorithms with Verification by COMSOL
Design Optimization of Printed Circuit Board Embedded Inductors through Genetic Algorithms with Verification by COMSOL Mickey P. Madsen *, Jakob D. Mønster, Arnold Knott and Michael A.E. Andersen Technical
Reproducibility of T1 and T2 Relaxation Times Calculated from Routine MR Imaging Sequences: Phantom Study
277 Reproducibility of T1 and T2 Relaxation Times Calculated from Routine MR Imaging Sequences: Phantom Study Bent O. Kjos 1 Richard L. Ehman Michael Brant-Zawadki Measurement of nand T2 relaxation times
NMR Techniques Applied to Mineral Oil, Water, and Ethanol
NMR Techniques Applied to Mineral Oil, Water, and Ethanol L. Bianchini and L. Coffey Physics Department, Brandeis University, MA, 02453 (Dated: February 24, 2010) Using a TeachSpin PS1-A pulsed NMR device,
Candidate Number. General Certificate of Education Advanced Level Examination June 2014
entre Number andidate Number Surname Other Names andidate Signature General ertificate of Education dvanced Level Examination June 214 Physics PHY4/1 Unit 4 Fields and Further Mechanics Section Wednesday
DL-QRP-AG Lambda/2 no Counterpoise: Fuchs Antenna matching unit
DL-QRP-AG Lambda/2 no Counterpoise: Fuchs Antenna matching unit QRPproject Molchstr. 15 12524 Berlin http://www.qrpproject.de Telefon: +49(30) 85 96 13 23 e-mail: [email protected] Handbucherstellung:
Application Note. So You Need to Measure Some Inductors?
So You Need to Measure Some nductors? Take a look at the 1910 nductance Analyzer. Although specifically designed for production testing of inductors and coils, in addition to measuring inductance (L),
Prepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
Specific Absorption Rate (SAR) in the head of Tablet user's
7th Latin American Workshop On Communications - 2015 Specific Absorption Rate (SAR) in the head of Tablet user's Juliana Borges Ferreira Post-graduation program in Electrical Engineering Federal University
ES250: Electrical Science. HW7: Energy Storage Elements
ES250: Electrical Science HW7: Energy Storage Elements Introduction This chapter introduces two more circuit elements, the capacitor and the inductor whose elements laws involve integration or differentiation;
4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:
4 SENSORS The modern technical world demands the availability of sensors to measure and convert a variety of physical quantities into electrical signals. These signals can then be fed into data processing
Electronic Supplementary Information
Electronic Supplementary Material (ESI) for Physical Chemistry Chemical Physics. This journal is the Owner Societies 2016 Electronic Supplementary Information Achieving High Resolution and Controlling
Human Exposure Limits
Human Exposure Limits Session 3 0 Version December 2014 Learning objectives In this session we will: Learn about the international exposure limits for workers and the public Learn about methods for assessing
Pillbox Antenna for 5.6 GHz Band Dragoslav Dobričić, YU1AW [email protected]
Pillbox Antenna for 5.6 GHz Band Dragoslav Dobričić, YU1AW [email protected] Introduction The pillbox or cheese antenna is made of two parallel plates which are connected to the narrow strip of parabolic
1. The diagram below represents magnetic lines of force within a region of space.
1. The diagram below represents magnetic lines of force within a region of space. 4. In which diagram below is the magnetic flux density at point P greatest? (1) (3) (2) (4) The magnetic field is strongest
Physics 25 Exam 3 November 3, 2009
1. A long, straight wire carries a current I. If the magnetic field at a distance d from the wire has magnitude B, what would be the the magnitude of the magnetic field at a distance d/3 from the wire,
Enhanced Stripline Scanning Array B.M. Cahill and J.C. Batchelor
Enhanced Stripline Scanning Array B.M. Cahill and J.C. Batchelor This paper is a postprint of a paper submitted to and accepted for publication in IET Microwaves, Antennas and Propagation and is subject
Chapter 4. LLC Resonant Converter
Chapter 4 LLC Resonant Converter 4.1 Introduction In previous chapters, the trends and technical challenges for front end DC/DC converter were discussed. High power density, high efficiency and high power
Physics 221 Experiment 5: Magnetic Fields
Physics 221 Experiment 5: Magnetic Fields August 25, 2007 ntroduction This experiment will examine the properties of magnetic fields. Magnetic fields can be created in a variety of ways, and are also found
WIRELESS MAGNETIC CONTACT
WIRELESS MAGNETIC CONTACT The magnetic contact wireless MAG HCS is a sensor able to detect opening or closing doors or windows and transmits via radio a alarm signal. It is composed of two distinct elements:
ACR AAPM TECHNICAL STANDARD FOR DIAGNOSTIC MEDICAL PHYSICS PERFORMANCE MONITORING OF MAGNETIC RESONANCE IMAGING (MRI) EQUIPMENT
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College