Introduction. Predictive Maintenance by Oil Analysis
|
|
|
- Nigel Erick McKinney
- 10 years ago
- Views:
Transcription
1 Spectroil Q 100, the Latest in a Nearly 30 Year Evolution of Oil Analysis Spectrometers Daniel P. Anderson Spectro Inc./Qinetiq North America, Littleton, MA, 01460, USA Introduction The basis of modern oil analysis is the use of optical emission spectroscopy (OES) to measure the ppm (parts per million) levels of wear metals, contaminants and additives in oil samples. Whatever else, an oil lab may measure, a multi-elemental analysis is the core of in-service oil analysis. This paper reviews the origins of OES as well as its application to in-service oil analysis resulting in the introduction of the Spectroil Q 100 Oil Analysis Spectrometer, the latest in a steady progression of smaller, lighter, more versatile instruments. Predictive Maintenance by Oil Analysis Spectrometric oil analysis is applicable to any closed loop lubricating system, such as those found in gas turbines, diesel and gasoline engines, transmissions, gearboxes, compressors and hydraulic systems. In practice, an oil sample is periodically taken from a system. The spectrometer analyzes the sample for trace levels of metals worn from moving parts as well as contamination and additive element levels. The resulting data, when compared to previous analyses and allowable limits, may indicate a sound mechanism showing only normal wear, or it may point out a potentially serious problem in its early stages. With this advanced warning, steps may be taken to correct the situation before serious damage or injury occurs. Spectrometric oil analysis works because fine particles are generated by relative motion of metallic parts in an oil-wetted system. The lubricating oil may be thought of as a diagnostic medium because the oil carries with it the particles generated by the wear contact. Abnormal wear modes such as corrosion, abrasion, severe wear, spalling, etc., cause an increase in the concentration of wear metals in the oil. Contaminants are detected and lubricant mix-ups or badly degraded lubricants are identified by the concentration of additive elements. Multi-element analysis, coupled with knowledge of the materials of construction, often allows identification of a specific component in distress. Cost Avoidance by Detecting Lubricant Mix-Up A serious recurring problem in maintenance procedures is the use of an incorrect lubricant. A condition monitoring program can readily identify such problems through the analysis of the lubricant additive package and lubricant physical property analysis. Lubricant mix-ups often occur when an oil system is "topped off" to replace the oil that has been lost due to use or leakage. Usually a small amount of incorrect oil in a large closed loop system presents few immediate problems. This is, however, not the case in Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 1 of 15
2 certain diesel engines as illustrated by this example 1. Table 1 is a summary of the last four spectrometric oil analyses for a medium speed diesel engine from a locomotive. Only the data for the most relevant elements are shown. Table 1, Spectrometric Results in ppm for an EMD Medium Speed Diesel Locomotive Fe Cu Ag Mg P Zn 30-Sep Dec Mar Jun The data clearly show that after the first two samples, an incorrect oil was used to top-off the reservoir. The three additive metals magnesium (Mg), phosphorus (P), and zinc (Zn) appear in the third analysis and increase in the fourth, a clear indication that the oil formulation has changed. In this type of engine, incorrect oil, which contains a zinc based additive package, can result in severe wear problems. Several components, such as wrist pin bearings have silver coatings that corrode and wear in the presence of zinc. The early stages of the corrosive action cause by the zinc additive are indicated by the increase in the iron, copper and silver wear metals. A recommendation based on the analysis was made to drain and flush the system and to observe correct top-off oil requirements. This particular fault would not have manifested itself by any other condition monitoring technique, such as vibration analysis, thermography, ultrasound, or performance monitoring. Without oil analysis, the wear problem could have resulted in a bearing failure and a major overhaul costing over $150,000. How is spectrometric data such as these obtained? Origins of Spectroscopy Anders Jonas Ångström is credited with being one of the first people to observe discreet bands of color in hot gases. He found it convenient to chart spectral bands in distances of m. This distance became know as the Ångström (Å) which put the Swedish scientist into a pretty select group. It s not everyone that gets a scientific unit named after them, although others come to mind like Faraday, Coloumb, Stoke, Ampere, Watt, Poise, Joule, Henry, to name several. However, what s impressive is the Ångström is a unit of length, one of the fundamental metrics of nature, not some fancy unit that needs to combine several units like distance, time and force in a complicated way to come up with something useful. So, it seems spectroscopy got a lot of respect right from the start! Around the same time in the mid-1800 s, two scientists, Gustav Kirchhoff and Robert Bunsen, were doing experiments together. Most people remember Bunsen because the Bunsen burner was named after him, but he and Kirchoff were doing things a lot more serious such as describing the fixed lines of the solar spectrum, discovering new elements, and advancing understanding of optics and electrical engineering. An easy demonstration of elemental identification by spectral emission is to introduce various elements into the flame of a Bunsen burner. A little table salt and yellow appears from Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 2 of 15
3 the sodium, some lithium salt gives a red flame from the lithium and copper compounds gave an intense green color. It wasn t until many years later, when 20 th century physicists, like Rutherford, Bohr, Einstein, Planck, Pauli, Schrodinger and Heisenberg worked out quantum mechanics and the behavior of atoms, that the existence of spectral lines and emission colors could be theoretically explained. Principles of Spectroscopy Spectroscopy is a technique for detecting and quantifying the presence of elements in a material. Spectroscopy utilizes the fact that each element has a unique atomic structure. When subjected to the addition of energy, each element emits light of specific wavelengths, or colors. Since no two elements have the same pattern of spectral lines, the elements can be differentiated. The intensity of the emitted light is proportional to the quantity of the element present in the sample allowing the concentration of that element to be determined. Figure 1 Shell Model of the Sodium Atom Figure 1 provides a simplified view of the events that take place in the spectroscopic process. To begin with, electrons surround the nucleus of the atom in well-defined shells. The first shell, the most inner shell, accommodates only two electrons, the next shell holds eight and the third shell also holds eight electrons. In the case of sodium, there is only one shell in the outer orbital. Figure 1a shows a sketch of an atom in its ground state. When an atom is cool, prior to excitation, the electrons in the atomic structure of each element revolve in their lowest energy or ground state. During excitation, whether it be by the heat of a flame, an electric arc or some other means, the atoms get hot and the electrons surrounding the nucleus absorb energy and rise to a higher energy state. The excited electrons are temporarily forced away from the nucleus of the atom into a higher, unstable orbit, as in Figure 2b. After reaching this unstable state, the electrons release this absorbed energy as they very quickly return to the ground or stable state. The energy released has a specific value corresponding to the particular electron transition that has occurred in the excited atom. The energy is given off in the form of light, as in Figure 1c. The light has a specific frequency or wavelength (frequency is inversely proportional to wavelength) determined by the energy of the electron in Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 3 of 15
4 transition. Since many transitions of different energy are possible for complicated atoms which have many electrons, light of many different wavelengths is emitted. If this light is dispersed by using a dispersing element, such as a prism, Figure 2, a line spectrum will result. Figure 2 - Light dispersed by a prism These spectral lines are unique to the atomic structure of only one element. For the hydrogen atom, with only one proton (positive charge of 1) in the nucleus and one electron (negative 1), the spectrum is fairly simple, See Figure 3. Figure 3 - Emission Spectrum of Hydrogen Figure 4 - Emission Spectrum of Iron On the other hand, the spectrum of iron with atomic number 26 (the nucleus has a positive charge of 26) is much more complicated with many emission lines in the visible spectrum corresponding to the many possible electronic transitions that may occur. See Figure 4. Figures 3 and 4 are showing only lines in the visible spectrum, which runs from about 400 nm (purple) to 700 nm (red). There are many more lines, often the most intense for some elements, in the ultra-violet region of the spectrum. These cannot be seen as colors, but can be seen as dark lines on photographic film The intensity of the spectral lines is proportional to the concentration of the element present in the sample. If more than one element is present in the sample, spectral lines of distinctively different wavelengths will appear for each element. These lines must be separated in order to identify and quantify the elements present in the sample. Usually only one spectral line among many possible choices is chosen to determine the concentration of a certain element. This line will be chosen for its intensity and freedom Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 4 of 15
5 from interference from spectral lines of other elements. To accomplish this, an optical system is required. Atomic Absorption Spectrscopy (AAS) One thing not yet mentioned is that when atoms of a certain type are present and they are exposed to polychromatic light (light of all wavelengths), they will absorb light at exactly those wavelengths at which they would emit light when they are excited. This process is a lot more efficient if the atoms are already hot. One type of spectrometer, namely an atomic absorption (AA) spectrometer, is built around this principle. Some sample is introduced into a flame. The flame needs to be as hot as possible. Then light of the element being analyzed is directed through the flame. This is done with special lamps, that contain excited atoms of the element being analyzed, so that they emit wavelengths characteristic of that element. If none of the element is present in the sample, the amount of light going through the flame and being measured at the detector is maximum. As the concentration of the element of interest increases, the detector signal decreases. Oxygen-acetylene and nitrous oxide-acetylene fuel 2 are often chosen for AA spectrometers. These flames achieve temperatures around 3000 C. As hot as that may be, it s inadequate for certain hard to excite elements. AA spectrometers are capable of doing a good job analyzing typical elements of interest in oil samples, but they are rarely used for this purpose because they analyze only one element at a time. Optical Emission Spectroscopy (OES) Spectrometers that look at the multitude of spectral lines from a heated (or excited ) sample, are called optical emission spectrometers. All optical emission spectrometers consist of three main components, these components are: 1. Excitation Source - introduces energy to the sample. 2. Optical System - separates and resolves the resulting emission from that excitation into its component wavelengths. 3. Readout System - detects and measures the light that has been separated into its component wavelengths by the optical system and presents this information to the operator in a usable fashion. Excitation Source One typical method used in the excitation source in modern spectrometers is an electric discharge. The source is designed to impart the energy generated in an arc or spark to the sample. For oil analysis spectrometers, a large electric potential is set up between a disc and rod electrode with the oil sample in the gap between them. An electric charge stored by a capacitor is discharged across this gap creating a high temperature electric arc which vaporizes a portion of the sample forming a plasma. A plasma is a hot, highly ionized gas Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 5 of 15
6 which emits intense light. The light given off as a result of this process contains emissions from all the elements present in the sample. These emissions can now be separated into individual wavelengths and measured using a properly designed optical system. Temperatures in the 5000 to 6000 C range are achieved and hard to excite elements emit enough light to be readily detected. Since the early days of spectroscopic oil analysis, oil has been sparked (or burned ) between a rotating carbon disc electrode and a carbon rod electrode. The sample is placed in a sample cap, the disc is partially immersed in the oil sample and the disc rotates as the burn proceeds. See Figure 5. This requires about 2 or 3 ml of sample depending on the exact cap being used. A fresh disc and a newly sharpened rod are required for each sample to eliminate sample carryover. This method is called rotating disc electrode (RDE) optical emission spectroscopy (OES), or combining the two, RDE- OES. Alternatively, one often sees it referred to as RDE-AES, for rotating disc electrode atomic emission spectroscopy. Figure 5 RDE Spectrometer Sample Stand Showing Oil Sample Being Burned Optical System The purpose of the optical system in a spectrometer is to separate the light coming from the plasma into the discrete wavelengths of which it is comprised. Most people are familiar with the phenomenon illustrated in Figure 2, in which a prism is used to separate a beam of white light into a spectrum of different colors or wavelengths. The same principle is employed in the spectrometer, except that instead of a prism, an optical device called a diffraction grating is used to separate the discreet wavelengths. The diffraction grating is a concave mirror with very fine lines on its surface that causes incident polychromatic light to be separated into component wavelengths. Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 6 of 15
7 Figure 6 shows the major components of an oil analysis spectrometer using a polychromator optic based on the Rowland Circle concept. Light from the excitation process (burn) exits the fiber optic cable and passes through the entrance slit and is concentrated on the diffraction grating by a lens. The entrance slit introduces light made up of all the elements present in the oil sample and defines the shape of the spectral lines at the focal curve after it is diffracted by the grating. The purpose of the grating is to separate (diffract) this light into its component wavelengths. The spectral lines can be photographed or electronically quantified by photomultiplier tubes (PMT s) or charge coupled devices (CCD s). Figure 6 Schematic of an Rotating Disc Electrode Optical Emission Spectrometer for Oil Analysis An important consideration when designing a spectrometer is the region of the spectrum where the wavelengths of interest occur. Many elements emit light in the visible region of the spectrum. However, there are elements that emit mainly in the Far Ultra Violet (FUV) region of the spectrum. This is significant because FUV radiation does not transmit well through air; rather, it is mostly absorbed. For the optic to see FUV spectral lines, it is necessary for the optical system to be mounted in a vacuum chamber or otherwise filled with gas transparent to FUV light, so that the emitted light can reach the grating, be diffracted, and then be detected at the focal curve. Thus, a sealed chamber and a vacuum pump or gas supply system become part of the system. History and Evolutionary Improvements Sometime after World War II, the Denver and Rio Grande Railroad 3, now defunct, began analyzing diesel locomotive engine oil by looking at the spectral lines emitted by an inservice oil sample when excited by a strong electric arc using carbon electrodes. Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 7 of 15
8 In those days, emission spectrometers were huge instruments because the diffraction gratings that separated the light from the hot sample into its component wavelengths were not nearly as good as they are today. The essential feature of a diffraction grating is that it has very fine lines of equal spacing ruled onto a reflective surface. In the early days, diffraction gratings were made by ruling machines that were capable of scribing very fine, regularly spaced lines onto an appropriate substrate. Since early diffraction gratings did not do a great job of separating spectral lines, the distance between the diffraction grating and the focal curve of the spectrometer, where the spectral lines are in focus, was very large so that sufficient separation could be achieved between adjacent spectral lines. Early emission spectrometers had focal lengths around three meters. We have one veteran employee who worked on spectrographs in the early days. Occasionally, he would crawl into the optic of one of these big instruments, then have the panels closed to look for stray light leaks. Today, and for the last 30 years or so, holographic diffraction gratings have been used that have superior dispersion and light throughput. The fine spacing on the reflective surface is made by photolithographic techniques created from a holographic interference pattern. The optic of the spectrometer can therefore be a lot smaller and still achieve adequate dispersion. Another inconvenience faced by early spectroscopists was that there was no easy way to measure the intensity of spectral lines at the focal curve of the spectrometer optic. Early spectrometers used glass photographic plates or photographic film along the focal curve of the optic to record spectral lines. See Figure 7. This is a idealized sketch to give an idea of the concept. In actuality, there would be hundreds, even thousands of spectral lines on a typical plate or film. First, the analyst had to determine if he was looking at the correct line correspondingly to the element of interest. Then, to determine concentration in the sample, the analyst needed to compare the density of the spectral line with reference photographs made from samples of known concentration. The analyst was able to report a probable range of concentration, but not exact numbers. For example, he might have been able to say < 5 ppm, 5 to 10 ppm, 10 to 20 ppm, 20 to 30ppm, etc. Figure 7 Photographic Film Used to Record Spectral Lines Early optical emission instruments were called spectrographs from the graph in photographs. A big advancement in the practical application of spectroscopy was the development by Walter Baird and his colleagues at Baird Corporation in the late 1950 s of the direct reading spectrometer. Instead of using film to record spectra, photomultiplier tubes were placed behind exit slits at just the right locations along the Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 8 of 15
9 focal curve to collect light from the emission lines of interest. In prior practice, only a few of the most important wear elements, such as Fe, Cu, Pb and Al, were measured with a spectrograph because of the tedious labor involved. With the new photomultiplier tube direct reader, as many as twenty or so elements could be measured with no additional burn time. It s not really correct to call today s spectrometers, spectrographs or to call spectrometric oil analysis, spectrographic oil analysis. It s hardly the worst mistake, but people familiar with spectrometers might think you still work for the Denver and Rio Grande! In 1960, not very long after the introduction of the photomultiplier tube equipped direct reading optical emission spectrometer, the first commercial oil analysis laboratory, Analysts Inc., was founded 3. The U.S. Navy had been funding research into wear metal analysis as a means of predicting aircraft engine failures. This led to the establishment of the Joint Oil Analysis Program (JOAP), a technical support and service center for all branches of the military. The JOAP manages a correlation program to ensure that all the oil analysis spectrometers enrolled in the program give similar results. Currently, there are nearly 600 U.S. military Spectro spectrometers enrolled in the JOAP correlation program. Figure 8 A Direct Reading Oil Analysis Spectrometer from the 1970 s Figure 8 shows a Baird FAS-2C spectrometer from the 1970 s, many of which were purchased by the U.S. military. These were deployed at various military bases throughout the world and were also installed on aircraft carriers. This was still a rather large instrument with a one meter optic (the distance from the grating to the focal curve) and weighing about 800 lbs (364 kg). With original factory crating it weighed 1400 lbs (636 kg). It was hardly a field mobile instrument! Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 9 of 15
10 The next advance in spectrometers applied to oil analysis was the development of the Spectroil Jr., the first field-mobile oil analysis spectrometer. Quite a few of these were purchased by the military, both U.S. and overseas. One important innovation that allowed the Spectroil Jr to be transported without losing focus between the emission source and the optic was the use of fiber optic cable to carry light from the source to the entrance slit of the optic. A certain number of Spectroil Jr s were deployed in the first Gulf War in 1991 and from that experience, the Spectroil M was designed from the frame up as a military fieldmobile oil analysis spectrometer. The Spectroil M was somewhat smaller and lighter than the Spectroil Jr, had automatic voltage and frequency selection and had various protections against shock, vibration, operator error, electro-magnetic radiation (into and out of) and dust ingression. One notable feature was the use of a heat exchanger mounted on the rear of the instrument to provide instrument cooling rather than the conventional approach of a filter and a fan. Upon the conclusion of the first Gulf War, a number of the Spectroil Jr s found their way back to our factory for service and the filter/fan approach in the Spectroil Jr s did not stop fine dust from being deposited throughout these instruments. The heat exchanger in the Spectroil M provides much better protection in dusty environments. The next big improvement to the Spectroil M, in 1998, was the design of a solid-state excitation source to drive the electric discharge between the disc and rod electrodes. This made the source frequency far more stable, especially with changes in atmospheric pressure, or if input power was switched from 50 to 60 Hz and vice-versa. This resulted in improved repeatability, greater long-term stability and also allowed water, coolant and other highly conductive samples to be run without shorting out the source ignition circuit. CCD Optics The most recent improvement to the Spectroil M was the introduction of a spectrometer optic using CCD detectors at the focal plane rather than using exit slits and photomultiplier tubes. A CCD detector is a semiconductor chip which is sensitive to light. The Spectroil M optic has 15 such CCD detectors which are mounted on the focal plane to cover the wavelength range of the analytical program. The chips are rectangular in shape and are subdivided into a grid of 2048 pixels. When light from the excitation process by way of the entrance slit and grating strikes a pixel, it generates a small electrical charge which is stored for later read-out. The size of the charge increases cumulatively as more light strikes the surface, and the greater, or brighter the light, the greater the charge. The CCD s are connected to electronic circuitry for power, control and read-out. The clear advantage of CCD light detectors is that the signal at any place along the focal curve can be measured so that one optic can be used for any combination of elements and ranges. Multiple spectral lines can be used for a single element and very closely spaced spectral lines can be readily measured without concern for the proximity of exit slits and photomultiplier tubes. New elements can be added later without hardware modification. Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 10 of 15
11 A further important mobility advantage is that changes in atmospheric pressure (i.e., altitude) can be easily accommodated. Readout System The readout system of the Spectroil M is controlled by an industrial grade processor and software. A clocking circuit and amplifier periodically reads the charge on a CCD chip and converts it from an analog to digital (ADC) signal to measure the light that has fallen on a pixel. The charge accumulated on a pixel is converted to an arbitrary number defined as intensity units. At the end of the analysis, the total intensities for each element are compared to calibration curves stored in memory and are converted to the concentration of the element present in the sample. Concentration is usually expressed in parts per million (ppm). This information is displayed on a video screen, or can be printed out on a printer. Once the analysis is completed and the results recorded, the system is ready for the next analysis. The analysis results may be left on the screen, stored on the hard disk, or can be sent to an external computer. User friendly software enables simple operation with automatic recalibration, while providing statistical functions for averaging and repeatability. Spectroil Q 100 The Spectroil Q 100, Figure 9, is Spectro s newest and completely solid-state spectrometer designed specifically for the analysis of oil samples. It measures trace quantities of elements dissolved or suspended as fine particles in mineral or synthetic petroleum based products using the time-tested and reliable rotating disc electrode (RDE) technique. Figure 8 The New Spectroil Q100 Oil Analysis Spectrometer With a modern look, the Spectroil Q 100 uses the same technology and design concepts that Spectro Incorporated has applied over the years to the Spectroil M family of spectrometers. All Spectro oil analysis instruments benefit from this technology, providing innovative features, ease of use and ruggedness that have made them the Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 11 of 15
12 standard workhorse instrument at most commercial oil analysis laboratories throughout the world that require the rapid analysis of wear metals, contaminants and additives in lubricants. The Spectroil Q 100 fulfills the requirements of ASTM D6595 Standard Method for Determination of Wear Metals and Contaminants in Used Lubricating Oils or Hydraulic Fluids by Rotating Disc Electrode Atomic Emission Spectrometry. The Spectroil Q 100 standard configuration has 22 elements, contaminants and additive; see Table 2. Additional elements can also be added in the field at any time. Table 2, Standard Configuration of the Spectroil Q 100 Wear Metals Contaminants Additives Aluminum Boron Barium Cadmium Calcium Boron Chromium Potassium Calcium Copper Silicon Chromium Iron Sodium Copper Lead Magnesium Manganese Molybdenum Nickel Silver Tin Titanium Vanadium Zinc Magnesium Molybdenum Phosphorus Silicon Zinc The Spectroil Q 100 has a low cost of ownership because of its high sample throughput. An experienced operator can comfortably process 45 to 50 samples an hour. The Q 100 requires not special utilities, such as bottled gas or cooling water. It requires only electric power. A built-in exhaust fan is capable of venting the fumes generated when the sample is burned up to 20 feet (6 meters) through standard (USA) 4 inch (10 cm) flexible exhaust duct. The Q 100 provides overall ease of operation requiring minimal operator training. The Q 100 is continuously available to perform analyses. Warm-up and standardization with calibration standards is not necessary for performing a measurement. The Q 100 sample stand has been redesigned to make changing electrodes, inserting the sample cap, gapping the electrodes and routine cleaning easier than ever before. The Spectroil Q 100 is the ideal spectrometer for the in-service oil analysis laboratory because technology has made it smaller and the inclusion of an innovative CCD optical system has made it more flexible for even the most exotic in-service oil analysis applications. An external computer with current Windows software controls the system. Additional capabilities to analyze engine coolants and water are available as options. Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 12 of 15
13 Data is easily downloaded to other programs, such as LIMS (laboratory information management system), or the very versatile web-based SpectroTrack software from Spectro. LIMS software manages laboratory operations, analytical data, and maintenance recommendations and histories, see Figure 9. Figure 9 SpectroTrack Analysis Recommendation Maintenance Feedback Who does Oil Analysis? Commercial oil analysis and machine condition monitoring have experienced rapid growth in the last decade. The market is more diverse than the military application, and machine condition monitoring based on oil analysis can be found in all of the following industries: Commercial Laboratories Oil Companies Chemical Processing Electric Power Generating Companies Railroads Refineries Construction Equipment Manufacturers and Dealers Airlines Mining Operations Public Transportation Companies Steel Mills Manufacturing Marine Fleets Formula One Racing Teams Commercial laboratories typically receive samples from a large variety of oil wetted systems. Applications are not limited to a few systems but include almost anything that uses oil as a lubricant. Typical requirements include generators, turbines, gear boxes, Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 13 of 15
14 engines, compressors, transmissions and even hydraulic systems. The Spectroil Q100 can process all of these samples types with essentially no carryover from one type to another. Benefits of Oil Analysis Typical advantages of an oil analysis program include: Reduced Maintenance Costs - By detecting a failure mode from its early stages, money is saved by: - preventing total loss of the equipment, - limiting the amount of secondary damage. For example, if a bearing which is wearing abnormally is replaced before damage is done to the shaft it supports, a great deal of money is saved. Increased Equipment Availability - Monitoring of equipment will prevent unexpected failures and unscheduled downtime. Improved Safety - In some equipment, most notably single engine aircraft and helicopters, mechanical failures can be life threatening. Extended Oil Drain Interval - With a well planned oil analysis program, oil change intervals can often be extended giving savings in labor, equipment availability and oil consumed. Longer Equipment Life - One of the side benefits of an oil analysis program is improved cleanliness and physical condition of the lubricant. More attention is paid to contaminant levels so that filters are more likely to be in proper working order and better housekeeping habits are practiced around the equipment. Since ingestion of outside contaminants is one of the principal causes of wear, cleaner oil means extended overall equipment lifetime. Conclusion The Spectroil Q100 is the latest member of a family of spectrometers with a long lineage that permit the rapid, convenient and reliable determination of the elemental content of wear metals, additive and contaminants in oil samples in support of a predictive maintenance program by oil analysis. References 1) Malte Lukas & Daniel P. Anderson, Machine and Lubricant Condition Monitoring for Extended Equipment Lifetimes and Predictive Maintenance at Power Plants, Proceedings of Power-Gen International, Jakarta, Indonesia, ) %20Flame%20for%20Atomic%20Absorption Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 14 of 15
15 3) Jack Poley, 21 st Century Oil Analysis, article in TLT magazine published by the STLE, March, ) Spectroil Q 100, Evolution of Oil Analysis Spectrometers Page 15 of 15
OVERVIEW OF ROTATING DISC ELECTRODE (RDE) OPTICAL EMISSION SPECTROSCOPY FOR IN-SERVICE OIL ANALYSIS
WHITE PAPER OVERVIEW OF ROTATING DISC ELECTRODE (RDE) OPTICAL EMISSION SPECTROSCOPY FOR IN-SERVICE OIL ANALYSIS Spectro Scientific Introduction The basis of modern oil analysis is the use of optical emission
New Developments and Functional Enhancements in RDE Used Oil Analysis Spectrometers
New Developments and Functional Enhancements in RDE Used Oil Analysis Spectrometers by: Malte Lukas, Daniel P. Anderson, and Robert J. Yurko Spectro Incorporated 160 Ayer Road Littleton, MA 01460-1103
Flame Tests & Electron Configuration
Flame Tests & Electron Configuration INTRODUCTION Many elements produce colors in the flame when heated. The origin of this phenomenon lies in the arrangement, or configuration of the electrons in the
ILLUSTRATIVE EXAMPLE: Given: A = 3 and B = 4 if we now want the value of C=? C = 3 + 4 = 9 + 16 = 25 or 2
Forensic Spectral Anaylysis: Warm up! The study of triangles has been done since ancient times. Many of the early discoveries about triangles are still used today. We will only be concerned with the "right
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
The Phenomenon of Photoelectric Emission:
The Photoelectric Effect. The Wave particle duality of light Light, like any other E.M.R (electromagnetic radiation) has got a dual nature. That is there are experiments that prove that it is made up of
5. The Nature of Light. Does Light Travel Infinitely Fast? EMR Travels At Finite Speed. EMR: Electric & Magnetic Waves
5. The Nature of Light Light travels in vacuum at 3.0. 10 8 m/s Light is one form of electromagnetic radiation Continuous radiation: Based on temperature Wien s Law & the Stefan-Boltzmann Law Light has
petroleum technologies
petroleum technologies TABLE OF CONTENTS BASICS OF OIL ANALYSIS p. 4-5 Introduction p. 4 Trend Oil Analysis p. 4 Engine, Transmission & Hydraulic Systems p. 5 Oil Analysis Measure p. 5 Why Oil Analysis?
Electrons in Atoms & Periodic Table Chapter 13 & 14 Assignment & Problem Set
Electrons in Atoms & Periodic Table Name Warm-Ups (Show your work for credit) Date 1. Date 2. Date 3. Date 4. Date 5. Date 6. Date 7. Date 8. Electrons in Atoms & Periodic Table 2 Study Guide: Things You
How to read your Oil Analysis Report
How to read your Oil Analysis Report A better understanding Petroleum Technologies Group, LLC. 4665 Broadmoor S.E. Ste#15 Grand Rapids, MI 49512 Phone # 616/698 9399 Fax # 616/698 9395 www.oil lab.com
Adapter Catalog MOBILE METAL ANALYZER. SPECTRO isort and SPECTROTEST
Adapter Catalog MOBILE METAL ANALYZER SPECTRO isort and SPECTROTEST Testing Adapters for Mobile Metal Analyzers A full range of metal analyzer products for onsite metal analysis tasks SPECTRO isort SPECTROTEST
Light as a Wave. The Nature of Light. EM Radiation Spectrum. EM Radiation Spectrum. Electromagnetic Radiation
The Nature of Light Light and other forms of radiation carry information to us from distance astronomical objects Visible light is a subset of a huge spectrum of electromagnetic radiation Maxwell pioneered
Oil Analysis. It is important that the sampling be done in the following manner: 1.
Oil Analysis Introduction Oil analysis was originally used by the military and the railroad as a preventative maintenance program. It has expanded today to the fi eld of study called Tribology. Oil analysis
Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
Atomic Structure: Chapter Problems
Atomic Structure: Chapter Problems Bohr Model Class Work 1. Describe the nuclear model of the atom. 2. Explain the problems with the nuclear model of the atom. 3. According to Niels Bohr, what does n stand
ATOMIC SPECTRA. Apparatus: Optical spectrometer, spectral tubes, power supply, incandescent lamp, bottles of dyed water, elevating jack or block.
1 ATOMIC SPECTRA Objective: To measure the wavelengths of visible light emitted by atomic hydrogen and verify the measured wavelengths against those predicted by quantum theory. To identify an unknown
TIME OF COMPLETION NAME SOLUTION DEPARTMENT OF NATURAL SCIENCES. PHYS 3650, Exam 2 Section 1 Version 1 October 31, 2005 Total Weight: 100 points
TIME OF COMPLETION NAME SOLUTION DEPARTMENT OF NATURAL SCIENCES PHYS 3650, Exam 2 Section 1 Version 1 October 31, 2005 Total Weight: 100 points 1. Check your examination for completeness prior to starting.
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
ALS TRIBOLOGY Engine Coolant Reference Guide
ALS TRIBOLOGY Engine Coolant Reference Guide Engine coolants are a mixture of, inhibitors, and water. Each formula is designed for specific protection and engine requirements. Mixing different coolants
LEAP SM Oil Analysis Program Saves Money with Consistent, Accurate Monitoring
LEAP SM Oil Analysis Program Saves Money with Consistent, Accurate Monitoring Safely extends oil drain intervals Partnering the Lubrication Engineers Oil Analysis Program with extended service interval
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development
USED LUBE OIL ANALYSIS & ANALYTICAL FERROGRAPHY
USED LUBE OIL ANALYSIS & ANALYTICAL FERROGRAPHY Paul Goldman MRT Laboratories 305 Nebraska Ave South Houston, TX 77587 713-944-8381 Oil Analysis Provides Four Important Pieces of Information 1) Condition
Technical Training Guide Turbine Oil Condition Monitoring
Technical Topic Technical Training Guide Turbine Oil Condition Monitoring Introduction Modern steam and gas turbines subject the turbine lubricant to ever greater demands. Higher temperatures are encountered
3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
Name Date Class ELECTRONS IN ATOMS. Standard Curriculum Core content Extension topics
13 ELECTRONS IN ATOMS Conceptual Curriculum Concrete concepts More abstract concepts or math/problem-solving Standard Curriculum Core content Extension topics Honors Curriculum Core honors content Options
Oil Analysis Handbook
Predictive Equipment Maintenance Oil Analysis Handbook HB0001 Copyright 2014 Spectro Scientific. All rights reserved. 107 Oil Analysis Handbook for Predictive Equipment Maintenance Oil Analysis Handbook
In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the
In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the discussion on the major components of the x-ray machine. Today
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
The photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law
Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine
FS Lubricants. Oil Analysis Program
FS Lubricants Oil Analysis Program What can an oil analysis do for you? 1. Establish safe and proper drain intervals. 2. Provide a reduction in unforeseen breakdowns. 3. Reduce down time. 4. Minimize the
COURSE 1020 OIL ANALYSIS WHY ATTEND
OIL ANALYSIS WHY CHOOSE US? 1 2 3 ICML 4 PERSONALIZED 5 6 OIL 7 ISO9001-2008 8 YEARS 9 EXPERTISE ADVANCED INVESTIGATIONS for sources of contamination, wear and oil degradation through a wide range of tests
3/5/2015. Oil Analysis Tests. Lubricating Oils. Six Functions of Lubricating Oil
Oil Analysis Tests Oil analysis is a series of laboratory tests used to evaluate the condition of used lubricants and equipment components. By studying the results of the used oil analysis tests, a determination
Cat S O SSM. Services. Understanding Your Results. Understanding Your: S O S Oil Reports S O S Oil Results S O S Coolant Reports S O S Coolant Results
Cat S O SSM Services Understanding Your Results Understanding Your: S O S Oil Reports S O S Oil Results S O S Coolant Reports S O S Coolant Results Contents Overview of S O S Oil Reports...4-5 Overview
Electromagnetic Radiation (EMR) and Remote Sensing
Electromagnetic Radiation (EMR) and Remote Sensing 1 Atmosphere Anything missing in between? Electromagnetic Radiation (EMR) is radiated by atomic particles at the source (the Sun), propagates through
How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
Theremino System Theremino Spectrometer Technology
Theremino System Theremino Spectrometer Technology theremino System - Theremino Spectrometer Technology - August 15, 2014 - Page 1 Operation principles By placing a digital camera with a diffraction grating
Oil Analysis Program ISO 9001: 2008 Certified.
il Analysis Program IS 91 28 Certified. Page 1 / 6 25548 Code Powertek Berhad Name Address Make GE Engine Turbine Gas Model Frame 9E il type / Test code 814 911 912 66 662 663Viscosity SHELL TURB CC 32
Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
GRID AND PRISM SPECTROMETERS
FYSA230/2 GRID AND PRISM SPECTROMETERS 1. Introduction Electromagnetic radiation (e.g. visible light) experiences reflection, refraction, interference and diffraction phenomena when entering and passing
WAVES AND ELECTROMAGNETIC RADIATION
WAVES AND ELECTROMAGNETIC RADIATION All waves are characterized by their wavelength, frequency and speed. Wavelength (lambda, ): the distance between any 2 successive crests or troughs. Frequency (nu,):
Copyright 1999 2010 by Mark Brandt, Ph.D. 12
Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits
X Ray Flourescence (XRF)
X Ray Flourescence (XRF) Aspiring Geologist XRF Technique XRF is a rapid, relatively non destructive process that produces chemical analysis of rocks, minerals, sediments, fluids, and soils It s purpose
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
How To Understand Light And Color
PRACTICE EXAM IV P202 SPRING 2004 1. In two separate double slit experiments, an interference pattern is observed on a screen. In the first experiment, violet light (λ = 754 nm) is used and a second-order
Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:
Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T
THE BOHR QUANTUM MODEL
THE BOHR QUANTUM MODEL INTRODUCTION When light from a low-pressure gas is subject to an electric discharge, a discrete line spectrum is emitted. When light from such a low-pressure gas is examined with
Cathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect
Objectives: PS-7.1 Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect Illustrate ways that the energy of waves is transferred by interaction with
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,
DO PHYSICS ONLINE FROM QUANTA TO QUARKS QUANTUM (WAVE) MECHANICS
DO PHYSICS ONLINE FROM QUANTA TO QUARKS QUANTUM (WAVE) MECHANICS Quantum Mechanics or wave mechanics is the best mathematical theory used today to describe and predict the behaviour of particles and waves.
What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher)
What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher) Introduction: There is more to a color than a name. Color can tell us lots of information. In this lab you will use a spectrophotometer
MODERN ATOMIC THEORY AND THE PERIODIC TABLE
CHAPTER 10 MODERN ATOMIC THEORY AND THE PERIODIC TABLE SOLUTIONS TO REVIEW QUESTIONS 1. Wavelength is defined as the distance between consecutive peaks in a wave. It is generally symbolized by the Greek
PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.
PHYS 222 Spring 2012 Final Exam Closed books, notes, etc. No electronic device except a calculator. NAME: (all questions with equal weight) 1. If the distance between two point charges is tripled, the
PUMPED Nd:YAG LASER. Last Revision: August 21, 2007
PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow
Making the Most of SáOá SM Services
MAINTENANCE PRODUCTS AND SERVICES Making the Most of SáOá SM Services With S O S Services, you don t just get test results, you get accurate analysis from the people who know your Cat equipment best. Your
Proper planning can prevent severe damage in the event of a failure.
BEST PRACTICES Mike Johnson & Matt Spurlock Fourth of a five-part series on oil analysis (April, May, July, November, December) Proper planning can prevent severe damage in the event of a failure. This
World-Class Oil Sampling It is Possible
www.lubecontrol.com.au World-Class Oil Sampling It is Possible Lube Control Pty Ltd Providing lubrication solutions! Unit 5/53 Norfolk Rd, Marion, SA 5043 t: (08) 8298 5563 f: (08) 8298 6253 e: [email protected]
Blackbody Radiation References INTRODUCTION
Blackbody Radiation References 1) R.A. Serway, R.J. Beichner: Physics for Scientists and Engineers with Modern Physics, 5 th Edition, Vol. 2, Ch.40, Saunders College Publishing (A Division of Harcourt
EXAMPLE EXERCISE 4.1 Change of Physical State
EXAMPLE EXERCISE 4.1 Change of Physical State State the term that applies to each of the following changes of physical state: (a) Snow changes from a solid to a liquid. (b) Gasoline changes from a liquid
X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
Raman Spectroscopy Basics
Raman Spectroscopy Basics Introduction Raman spectroscopy is a spectroscopic technique based on inelastic scattering of monochromatic light, usually from a laser source. Inelastic scattering means that
Ultrasound Condition Monitoring
Ultrasound Condition Monitoring Whitepaper Alan Bandes UE Systems, Inc. Abstract: Instruments based on airborne/structure borne ultrasound technology offer many opportunities for reducing energy waste
Analytical Tools to Detect and Quantify Large Wear Particles in Used Lubricating Oil
Analytical Tools to Detect and Quantify Large Wear Particles in Used Lubricating Oil M. LUKAS Spectro Inc., Littleton, MA, U.S.A. D. P. ANDERSON Spectro Inc. Littleton, MA, U.S.A. SUMMARY Machine condition
S O SSM. Services. Cat. Management Guide. Accurate Analysis Improved Cost Control Optimized Performance
S O SSM Services Cat Management Guide Accurate Analysis Improved Cost Control Optimized Performance Contents S O S SM Services introduction... 3 Making the most of S O S Services...4-5 Good equipment health
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control
By-Pass Filters: Taking Your Fleet the Extra Mile
T By-Pass Filters: Taking Your Fleet the Extra Mile Donald S. Culpepper Gulf Coast Filters, Inc. here has been an industry-wide push over the last few years to extend oil drain intervals on fleet equipment.
18.2 Comparing Atoms. Atomic number. Chapter 18
As you know, some substances are made up of only one kind of atom and these substances are called elements. You already know something about a number of elements you ve heard of hydrogen, helium, silver,
Indiana's Academic Standards 2010 ICP Indiana's Academic Standards 2016 ICP. map) that describe the relationship acceleration, velocity and distance.
.1.1 Measure the motion of objects to understand.1.1 Develop graphical, the relationships among distance, velocity and mathematical, and pictorial acceleration. Develop deeper understanding through representations
Longer lifetime for hydraulic oil
Longer lifetime for hydraulic oil In many applications hydraulic oils and lubricating oils are exchanged with new oil after only some thousand hours of operation. Changing the oil causes high costs and
Introduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
CHAPTER 6 INSTRUMENTATION AND MEASUREMENTS 6.1 MEASUREMENTS
CHAPTER 6 INSTRUMENTATION AND MEASUREMENTS 6.1 MEASUREMENTS Atmospheric electricity is a field that is very easy to get into because it does not require a large capital investment for measuring equipment.
Using the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
DIESEL FUEL CONDITIONING
COD FILTRATION DIESEL FUEL CONDITIONING REMOVE WATER & PARTICLES TO EXTEND FUEL INJECTOR LIFE & INCREASE COMBUSTION ENGINE FUEL EFFICIENCY Remove free water to saturation point with high single pass efficiency
High-end stationary vacuum spectrometer. Innovation with Integrity OES
Q8 MAGELLAN High-end stationary vacuum spectrometer Innovation with Integrity OES The Evolution of Perfection Engineered for ultimate sparc spectrometry Next Step to perfection The genes are right: Q8
SSO Transmission Grating Spectrograph (TGS) User s Guide
SSO Transmission Grating Spectrograph (TGS) User s Guide The Rigel TGS User s Guide available online explains how a transmission grating spectrograph (TGS) works and how efficient they are. Please refer
47374_04_p25-32.qxd 2/9/07 7:50 AM Page 25. 4 Atoms and Elements
47374_04_p25-32.qxd 2/9/07 7:50 AM Page 25 4 Atoms and Elements 4.1 a. Cu b. Si c. K d. N e. Fe f. Ba g. Pb h. Sr 4.2 a. O b. Li c. S d. Al e. H f. Ne g. Sn h. Au 4.3 a. carbon b. chlorine c. iodine d.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
Vehicle Fire Protection. An Ownerʼs Manual for ANSUL Fire Suppression/Detection Systems
Vehicle Fire Protection An Ownerʼs Manual for ANSUL Fire Suppression/Detection Systems 002792 Off-road vehicles do have fires Statistics show that off-road vehicles burn with alarming frequency. These
MCAL Spectrophotometry. Spectrophotometry
MCAL Spectrophotometry Instruments include: Cary 50 UV-vis Spectrophotometer Eclipse Spectrofluorometer HPLC Diode Array and Fluorescence ICP-OES with CCD detection Spectrophotometry The instruments all
HAZARD RISK ASSESSMENT
HAZARD RISK ASSESSMENT Mt Grace Batchelor Magnesium Project A hazard risk assessment was undertaken to identify and assess the consequences of unplanned incidents that may pose safety, environmental risks,
Infrared Spectroscopy: Theory
u Chapter 15 Infrared Spectroscopy: Theory An important tool of the organic chemist is Infrared Spectroscopy, or IR. IR spectra are acquired on a special instrument, called an IR spectrometer. IR is used
Experiment 10. Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado
1 Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado Introduction Some radioactive isotopes formed billions of years ago have half- lives so long that they are
Physics 30 Worksheet # 14: Michelson Experiment
Physics 30 Worksheet # 14: Michelson Experiment 1. The speed of light found by a Michelson experiment was found to be 2.90 x 10 8 m/s. If the two hills were 20.0 km apart, what was the frequency of the
Unit 3 Study Guide: Electron Configuration & The Periodic Table
Name: Teacher s Name: Class: Block: Date: Unit 3 Study Guide: Electron Configuration & The Periodic Table 1. For each of the following elements, state whether the element is radioactive, synthetic or both.
SECTION 23 81 03 - PACKAGED ROOFTOP AIR CONDITIONING UNITS NON-CUSTOM
SECTION 23 81 03 - PACKAGED ROOFTOP AIR CONDITIONING UNITS NON-CUSTOM PART 1 - GENERAL 1.1 SUMMARY A. Section Includes: 1. Packaged rooftop air conditioning unit (5 tons and smaller). 2. Roof curb. 1.2
2. John Dalton did his research work in which of the following countries? a. France b. Greece c. Russia d. England
CHAPTER 3 1. Which combination of individual and contribution is not correct? a. Antoine Lavoisier - clarified confusion over cause of burning b. John Dalton - proposed atomic theory c. Marie Curie - discovered
Chem 1A Exam 2 Review Problems
Chem 1A Exam 2 Review Problems 1. At 0.967 atm, the height of mercury in a barometer is 0.735 m. If the mercury were replaced with water, what height of water (in meters) would be supported at this pressure?
Factory owners must ensure the boiler is:
Factory owners must ensure the boiler is: * Registered with the Boilers and Pressure Vessels Division, Labour Department * Examined by an appointed examiner and has a valid certificate of fitness * Supervised
Oil and Coolant Circulating Heating System. Model - OCSM
Oil and Coolant Circulating Heating System Model - OCSM Installation & Operation Manual 216280-000 REV 2 Identifying Your System The HOTSTART heating system is designed to heat fluids for use in marine
Spectrometric Oil Analysis An Untapped Resource for Condition Monitoring
Spectrometric Oil Analysis An Untapped Resource for Condition Monitoring Dahunsi Olurotimi Akintunde Mechanical Engineering Department, Federal University of Technology Akure, Ondo State, Nigeria Email:
UVC LEDs for. Environmental Monitoring
UVC LEDs for Environmental Monitoring Environmental monitoring relies extensively on molecular spectroscopy for tracking air quality, water and wastewater quality, and detecting hazardous substances. Advances
Experiment #12: The Bohr Atom. Equipment: Spectroscope Hydrogen and Helium Gas Discharge Tubes, Holder, and Variac Flashlight
Experiment #12: The Bohr Atom Purpose: To observe the visible spectrum of hydrogen and helium and verify the Bohr model of the hydrogen atom. Equipment: Spectroscope Hydrogen and Helium Gas Discharge Tubes,
WHAT IS INFRARED (IR) THERMOGRAPHY
WHAT IS INFRARED (IR) THERMOGRAPHY IR Thermography is the technique of producing pictures called from the invisible thermal radiation that objects emit. This is a non-contact means of identifying electrical
MCQ - ENERGY and CLIMATE
1 MCQ - ENERGY and CLIMATE 1. The volume of a given mass of water at a temperature of T 1 is V 1. The volume increases to V 2 at temperature T 2. The coefficient of volume expansion of water may be calculated
Symmetric Stretch: allows molecule to move through space
BACKGROUND INFORMATION Infrared Spectroscopy Before introducing the subject of IR spectroscopy, we must first review some aspects of the electromagnetic spectrum. The electromagnetic spectrum is composed
Raising the Bar in FT-IR: Getting the Most Out of On-site Oil Condition Monitoring
Technical Note Raising the Bar in FT-IR: Getting the Most Out of On-site Oil Condition Monitoring FT-IR Spectroscopy Introduction Lubrication is a vital component of machinery and, as with any component,