Near Infrared Transmission Spectroscopy in the Food Industry
|
|
|
- Geraldine Patterson
- 10 years ago
- Views:
Transcription
1 Near Infrared Transmission Spectroscopy in the Food Industry Introduction: Near Infrared Spectroscopy is used in many industries including the pharmaceutical, petrochemical, agriculture, cosmetics, chemical and food industries. However in the food industry NIR has an almost universal application. Since food is made mostly from proteins, carbohydrates, fats and water, i.e. >99% by weight, NIR provides a means of measuring these components in almost any food. NIR Analysis: Simple theory. Figure 1. shows part of the electromagnetic radiation spectrum from 300nm to 10000nm. It shows the Visible, NIR and Mid IR spectral regions. When light energy interacts with a material, energy is absorbed at resonant frequencies associated with the atomic and molecular interactions of the material. In the Visible region, energy is absorbed when electrons jump from a lower to a higher energy state or orbit. In this region, chromaphores such as metal chelates, absorb visible light. Likewise Atomic Absorption Spectroscopy also uses this phenomenon to measure metals species as they burn in a flame. Frequency (Wavenumbers) Visible Electronic Shift Near Infrared Overtone and Combination Bands Mid Infrared Fundamental Vibrational Bands Wavelength (Nanometres)
2 In the Mid Infrared region, i.e to 10000nm, energy is absorbed by vibrating molecules at resonant frequencies for each type of vibration, eg, stretching, bending, wagging, twisting etc. In the Mid IR region, all chemical bonds in organic molecules exhibit strong absorption bands. Typically the Mid IR region is suitable for characterizing C-C, C-O, C- N, C-H, O-H, N-H and many other chemical bonds. Sample preparation such as dissolving in a solvent, preparing a nujol mull, drawing film etc, is required in order to get sufficient light through a sample. Mid IR spectroscopy is generally used for non-water bearing materials such petrochemicals, plastics, polymers and chemicals. Mid IR spectroscopy is used mainly for qualitative analysis rather than quantitative analysis. The NIR spectral region, i.e., 720 to 2500nm, is the Overtone and Combination region of the Mid IR region. NIR spectra contain absorbance bands mainly due to three chemical bonds, i.e., C-H (fats, oil, hydrocarbons), O-H (water, alcohol) and N-H (protein). Other chemical bonds may exhibit overtone bands in the NIR region, however they are generally too weak to be considered for use in analysis of complex mixtures such as foods, agricultural product, pharmaceuticals, toiletries, cosmetics, textiles etc. NIR spectra do not have the resolution of the Mid IR spectra however NIR spectra can generally be collected off or through materials without sample preparation as well as it is suitable for measuring high and low water content materials. Whereas Mid IR is mainly a qualitative technique, NIR is mainly a quantitative technique. NIR provides a very rapid means of measuring multiple components in foods, agricultural products, pharmaceuticals, cosmetics, toiletries, textiles and virtually any organic material or compound.
3 Three NIR Regions Figure 2. shows the NIR region from 720 to 2500nm. There are three parts of the NIR spectral region, 1) Reflectance, 2) Transmission and 3) Transflectance. Transflectance Transmission Reflectance 1) Transflectance: 720 to 1100nm. This section is most suited to transflectance through a thick sample, such as, seeds, slurries, liquids and pastes. The absorption bands are due to 3 rd overtones of the fundamental stretch bonds in the Mid IR region. 2) Transmission: 1200 to 1850nm. This section can be used for transmission through liquids and films, as well as diffuse reflectance measurements off samples with high water contents. The absorption bands are due to the 1 st and 2 nd overtones of the fundamental stretch bonds in the Mid IR region 3) Reflectance: 1850 to 2500nm. This section is predominantly used for making diffuse reflectance measurements off ground or solid materials. The absorption bands are due to combination bands, i.e., C-H stretch and bend combination bands.
4 The Transflectance region is of particular interest in the analysis of foods because it is suitable for measuring high moisture and high fat content products including meat, dairy products, jams and conserves, dough and batters. The major advantage of working in this region is that longer pathlength sample cells can be used to collect the NIT spectra. Typically a mm pathlength can be used. This makes sampling easier and allows viscous and non-homogeneous samples to be scanned without further sample processing. A major advantage of measuring in Transflectance as compared with Reflectance is that the spectra represent the variation in components throughout the entire sample, not just the surface. In reflectance, the first 1mm contributes as much as 99% of the spectrum. As such uneven distribution of components in the sample, egg, drying at the surface, or separation of a water or oil layer at a glass window, results in reflectance spectra that do not represent the entire sample. Spectral Collection Modes Transflectance: The Transflectance region was first suggested by Norris as a means of measuring whole cereal grains and oil seeds because the NIR light could penetrate through the grains and oil seeds. Actually the Transflectance mechanism is a combination of reflectance and transmission, as the NIR light reflects off the surface of the seeds as it transmits through the sample to the other side. Figure 4. Schematic of Transflectance Optics
5 Transmission: Transmission spectroscopy is the most common form of spectroscopy. UV-Visible, Mid IR and Atomic Absorption are major analytical techniques used in the analysis of water, metals, biological and organic materials. The basic principle of transmission spectroscopy is that light passes through a clear or transparent sample and energy is absorbed by the chemical components components. The light is not deflected as it passes through the sample. In the NIR transmission region, 1200 to 1850nm, the same principles apply. Figure 5. Schematic of Transmission Optics Reflectance: In Reflectance spectroscopy, light illuminates sample at 0 degree angle. Figure 6. Schematic of Reflectance Optics Light interacts with material and re-radiates Diffuse Reflected energy back into the plane of illumination. The re-radiated light is detected at 45 degree angle, in order to reduce Specular Reflectance.
6 Applications of NIT Spectroscopy for Food Products The following are examples of NIT analyses of food products using the Transflectance region; Raw and Processed Meats: NIT spectra of raw and processed meat can be collected using a 10 mm Squeeze Cell as shown in Figure 7. Figure 8. shows the NIT spectra of minced beef, minced ham, pork sausage, salami and bologna. The differences in density of these samples are shown in the significant baseline variances. By converting the spectra with a Multiplicative Scatter Correction (MSC) algorithm, these differences are reduced. Figure 9. shows the MSC converted spectra of these meat samples. Figure 7. Squeeze Cell used for analyzing meat, cheese and dough. Figure 8. NIT Spectra of various meats, eg, turkey, beef, chicken, pork and sausage mix.
7 Figure 9. Multiplicative Scatter Corrected Spectra of the meat samples As an example of quantitative analysis of meat products, figure 10. shows the plot of NIR Fat vs Soxhlet Extracted Fat. A calibration for fat was developed using 79 samples and a further 21 samples were analysed in duplicate to define the accuracy and precision. The accuracy or Standard Error of Prediction (SEP) was calculated as 0.45% fat and the precision or Standard Deviation of Difference (SDD) was calculated as 0.19%. Calibrations for protein, moisture and salt have also been developed for processed meat samples. Figure 10. Plot of NIR Fat vs Soxhlet Extracted Fat calibration data.
8 Dairy Products: Figure 11. shows the NIT spectra of milk, cheddar cheese, yogurt, milk powder, cream cheese and butter. Except for milk and milk powder, these spectra were also collected using a 10mm pathlength Squeeze Cell. The milk was scanned in a 10mm pathlength glass cuvette and the milk powder was scanned in a 5mm pathlength Powder Cell. Figure 11. NIT Spectra of Milk, Cheese, Milk Powder, Yogurt, Cream Cheese As an example of a quantitative analysis of dairy products, figure 12, shows the plots of the analysis of moisture in sliced cheese samples. The SEP s for moisture = 0.55%. Figure 12. Plot of NIT Fat vs Reference Fat
9 Baked Goods: Figure 13. shows the NIT spectra of cookie dough scanned in a 20mm pathlength Squeeze Cell. Fat, moisture, protein and sugar can be measured in dough, however the ability to rapidly measure fat and moisture in the mixing room, means that the dough will spread correctly when the cookies are laid on the hot oven band. If the fat content is too high then the cookies will spread too much, resulting in out of spec product. The accuracy of the fat and moisture measurements has been shown to be 0.13 and 0.28% respectively. Absorbance Pie Dough Specral Data Wavelength range 720nm nm Figure 13. NIT Spectra of Pie Dough In the baked good industries, NIT can also be used for measuring the incoming flour for protein, moisture and ash, as well as dough stability, dough development and water absorption. Other raw materials used in baking that can be analysed using NIT, include, liquid and powdered eggs, milk and milk powder, shortening, liquid and crystalline sugars and syrups. A very unique application for NIT spectroscopy is the analysis of whole cookies. Manufacturers can measure the protein, moisture, sugar and fat content of whole cookies. The above spectra were collected by placing a whole cookie into a sample cell and lowering into a light beam. The NIR light passed through the cookies, thus giving a measurement representing the outside and inside of the cookie. Figure 14. shows the NIT spectra through several cookie types.
10 Cookies Farmbake Gingernut Premier Butternut FarmbakeChoc 1 P i x e l s ( W a v e l e n g t h s ) Figure 14. NIT Spectra of Whole Cookies Confectionery: Chocolate is the world s most popular confectionery. Figure 15 shows the NIT spectra of a milk chocolate and a dark chocolate using a 5mm pathlength Squeeze Cell. The chocolate samples were softened but not melted. The softened block was compressed between two glass windows and scanned CHOCs1 CHOCs2 CHOCs3 CHOCs4 CHOCs Figure 15 NIT Spectra of Milk and Dark Chocolates Reflectance spectra of chocolate suffer from the problem of drying and uneven distribution of the fat between the surface and the rest of the sample. Transflectance provides a quick method of measuring fat and moisture throughout the entire chocolate sample.
11 Figure 16. shows the NIT spectra of several common types of lollies and candies. Sweets such as jubes or jellies, are made of gelatin, i.e., protein, water and sugar. Since these sweets are easily deformed, they can be measured in a 10mm pathlength Squeeze Cell. Nougats and fondants used in many candy bars are a mixture of fat, sugar and emulsifiers. NIT spectra are easily collected and as such provide manufacturers a rapid means of measuring the consistency of the material before it is used in the final bar. NIT Spectra of Candies and Lollies Absorbance BANANASs1 CARAMELSs1 JELLIESs1 Wavelength(Pixel) Figure 16. NIT Spectra of Lollies or Candies. Grains and Oil Seeds: Figure 17 shows the NIT spectra of wheat, barley, soybeans, corn (maize) and canola seeds. NIT Spectra of Cereals and Oil Seeds Absorbance Barley Canola Wheat Soybean Corn Wavelength(Pixel) Figure 17. NIT Spectra of Cereal Grains and Oil Seeds
12 It is difficult to measure whole seeds in the Transmission and Reflectance regions because of the amount of light absorbed by the sample. However in the Transflectance region, the NIT spectra can be collected using an elongated sample, cell with pathlengths from 8 to 28mm, depending on the size of the seeds or grains. This makes sampling very easy and lends itself to on line measurements. Beverages: Wine, Beer, Spirits and even alcoholic mixer drinks, can be analysed using NIT for alcohol. Figure 18 shows the NIT spectra of several alcoholic beverages. NIT Spectra of Beverages Absorbance Liquer Red Wine White Wine Beer Wavelength(Pixel) Figure 18. NIT Spectra of Beverages Figure 19 shows the linearity and accuracy of measuring alcohol in beer. Alcohol Calibration NIT-38 Alcohol (%) SEC = 0.09% R 2 = Reference Alcohol (%) Figure 19. Plot of NIT Alcohol vs Reference Alcohol
13 Other non-alcoholic beverage scan also be analysed by NIT, such as soy milk, fruit juice and yakult. Figure 20. shows the plot of NIT Protein and Reference Protein in Soy Milk. The SEC is 0.04%. Figure 20. Plot of Protein Results in Soy Milk. Where as Refractometers are a low cost instrument for measuring sugars, NIT is capable of measuring sugar, protein, fat and solids in the same sample.
14 Calibrating NIT Analyser Introduction NIR spectroscopy is a secondary or correlative technique i.e. the spectral data collected is correlated through statistical means to some reference (laboratory) data. Regression Analysis techniques are used to develop calibration models, which are subsequently downloaded into a NIR analyser for use in predicting unknown samples. The following steps presented in the flow chart, outline the procedure in developing a NIT calibration: Flow Chart for Calibration Development Optimise Pathlength Ensure Reproducible Spectra Select Samples (Box Car Distribution) Analyse Samples in Duplicate Scan Samples in Duplicate PLS or MLR Calibration Evaluate Calibration Models Add 5 Low Temp and 5 High Temp Scans Download Calibration To Analyser Test Accuracy, Temperature Stability and Repeatability
15 Calibration Techniques The concentration of a chemical component is proportional to the amount of light absorbed at a frequency specific to that component. This relationship can be expressed as follows; Concentration = Conc Factor x Absorbance Traditional spectroscopy methods such as UV-Visible, use Simple Linear Regression (SLR), sometimes known as Univariate Regression, to estimate the Concentration Factor, CF, which relates the absorbance of a chemical species at a known wavelength. SLR estimates the equation of the line: % Y = b x + b 1 0 Where: Y is the concentration of the constituent of interest X is the absorbance at a specific wavelength b 1 is the Calibration Factor or Slope b 0 is the Intercept or Offset By plotting the absorbance versus concentration, a straight line calibration curve is produced. The slope of the line is the Concentration Factor. Absorbance Sample Absorbance = 0.6 Concentration = 0.035mM Concentration (m M) Figure 21. Plot of Absorbance vs Concentration.
16 Multiple Linear Regression (MLR) In NIR Spectroscopy the spectral bands are commonly overlapping and exhibit some form of baseline correction due to scatter As such, NIR spectra do not follow the same linear relationship between absorbance and concentration as seen in UV and Visible spectroscopy. Multiple Linear Regression (MLR) techniques are require to develop multi term calibration models to compensate for overlapping bands, interferences, matrix effects and scatter. A simple example of MLR is Bi-Modal calibration, where one component is being interfered with by another, eg, tryptophan in the presence of Tyrosine. A simple two-component calibration can be developed. The model would take the following general form. Y=b 0 +b 1 X 1 +b 2 X 2 Where: b 0 =y-intercept b 1 and b 2 =partial regression coefficients x n are the absorbances at each wavelength MLR would be used to estimate the values of b0, b1 and b2. Figure 22. Spectra of Tryptophan and Tyrosine
17 When the sample matrix is complex, ie, a solid sample with multiple components and inconsistent particle size or distribution, then the MLR model becomes more sophisticated and can include up to 20 terms. For example; Y=b 0 +b 1 X 1 +b 2 X 2 +b 3 X 3.. +b 20 X 20 There are two methods for performing MLR Stepwise Forward Stepwise Backward Stepwise Backward Develop a model using all wavelengths Reduce the model, one wavelength at a time until the Standard error of Calibration (SEC) blows out. Used in original filter based instruments (Technicon, Dickey- John Perten). Can easily overfit data set resulting in poor prediction and poor reproducibility. Stepwise Forward Select the first wavelength with the highest correlation. Regression looks for next wavelength, which will increase the correlation and reduce the error. Process is continued until the addition of the next wavelength has no effect or starts to reduce the correlation and increase the error. Best results from derivative spectra (removes baseline effects and spectral data are not intercorrelated). This technique, when used with derivative spectra, usually requires two or three terms in the calibration equation. This reduces the tendency to overfit the data and results in more reliable predictions. (Note, in order to use derivative spectra, a full scanning spectrophotometer is required).
18 Principal Components Regression In Principal Components Regression (PCR), the objective is to isolate Principal Components (PC s) (or vectors of maximum variation) of the spectral data and then use these as terms in a calibration equation. Rather than using a single wavelength to characterise a peak, PCR estimate the shape of the peak or factors affecting the spectra and use these as x terms (spectral reconstruction). Partial Least Squares (PLS) Calibrations PLS regression is very similar to PCR, however in PLS the constituent data is used in the computation of the PC s. Generally PLS is used in preference to PCR in NIR spectroscopy, as PCR tends to incorporate too much redundant information. PLS models also averages the spectra and improve the signal to noise ratio. PLS is defined as a global technique, i.e. the entire spectrum is used to develop a calibration and not just some highly correlated wavelengths. An example of a PLS regression profile is shown in figure 23. Figure 23. Plot of B Coefficients of a PLS Model. The plot represents the weightings or loadings, which are placed on each part of the spectrum. By adding the weighted wavelength readings and adding the regression offset b 0, a value of the constituent is obtained using the following equation: % Y = b0 + b1! 1 + b2! b38! 38
19 Qualitative Analysis Discriminant Analysis is a qualitative tool used to group (or classify) spectra into pre-defined classes. The first step of the process is to perform a Principal Components Analysis (PCA) and plot the most significant PC s in order to identify any clustering of spectra (known as Cluster Analysis). Cluster Analysis is commonly called an unsupervised analysis, as the number of clusters is not known and must be determined by visual inspection of the data. Clusters are usually found using two methods: Correlation techniques- Samples are grouped based on how they correlate to each other. Distance measures- Samples are grouped based on how close they are to each other. One distance of measure is known as the Mahalanobis distance, and is commonly used in statistical methods to define closeness of objects. The Mahalanobis Distance can be visualised in the following diagram. Each wavelength reading, ie. absorbance value in the spectrum, is considered an independent axis with a value equivalent to the absorbance value. The diagram only shows three axes, however in NIR the number of vectors is generally hundreds. The resultant vector, ie, the sum of the three vectors produces a unique vector that represents the entire spectrum. Axis 1 (Abs 1) Resultant Vector Axis 3 (Abs 3) Axis 2 (Abs 2) Figure 24. Example of Mahalanobis Distance Plot
20 Spectra that are similar will have a Resultant Vector with the same angle and length. Normally spectra of even the same sample vary due to packing density, particle size and purity. As such the resultant vectors of spectra from many samples of the same material form an ellipse in space. This is the Cluster that defines the variation in spectra of the specific material. By building a library of spectra for each material, Discriminant Analysis can be used to identify unknown spectra against the library. The closer the resultant vector of the unknown sample is to the Cluster, then the better the fit. The Mahalanobis Distance is the parameter that is calculated to determine how close the unknown spectrum is to each Cluster in the library. If the Mahalanobis Distance is 0, then the spectrum matches perfectly. As such, the smallest Mahalanobis Distance is used to pick which Cluster the unknown spectrum fits best. Discrimination Analysis is used for two tasks: Identification of Unknown Materials Requires library of known materials Matches unknown material to a class defined by the training set Displays top five matches along with the statistical relevance Quantification of Materials or Sameness Comparison of known material to a library sample to measure sameness. The concept of Sameness may be a strange term, but the idea is simple. If a manufacturer buys 10 tonnes of a chemical additive, they want to know if it is the same as the last batch or at least as same as the batches which have proven to be acceptable. Since the NIR spectrum of a material contains information about the chemical composition, as well as the physical characteristics, eg. Particle size and size distribution, compaction, crystallinity etc, then Discriminant Analysis provides a means of inspecting the whole material rather than making a QC assessment on just a couple of chemical tests.
21 Routine NIR Analysis Once a calibration model or models have been developed, they are downloaded into the NIR analyser for use in analysing every day samples. The fundamental assumption in performing MLR or PLS regressions to develop calibration models, is that the calibration set represents all the future variations in the samples to be tested. It also assumes that the environmental conditions are the same for the calibration and the unknown samples. Since it is impossible to include all future sample variations and environmental variations, it is common to add more sample spectra into the calibration set. As such it is important to put aside around 5% of the everyday samples for analysis using the laboratory methods. By comparing the results for these samples against the NIR results, an estimate of the accuracy and stability of the NIR method can be obtained. Plotting control charts of the error between the laboratory and the NIR methods is very useful as it shows whether the errors are random or systematic. An alternative strategy is to use a test sample to check the stability of the NIR analyser on a daily basis. The test sample is analysed each day and if the difference between the tested value and the NIR result is greater repeatability statistic, then a bias adjustment can be made to the calibration. Figure 25. shows an example of calculating a Bias adjustment. Sample Cropscan Ref Diff Plot Cropscan vs Ref Protein Bias = Average Ref Protein% Figure 25. Example of Bias Adjustment
22 Periodically NIR calibrations may require more than a Bias adjustment. A Slope and Bias adjustment can be necessary when the low readings are too high and the high readings are too low and the middle readings are about right. Figure 26. shows an example of where a Slope and Bias adjustment to the calibration are required. Sample Cropscan Ref Diff Bias = Average 0.1 Plot Cropscan vs Ref Protein y = x Ref Protein% Figure 26. Example of Slope and Bias Adjustment The Slope and Intercept of the line of fit between the NIR results and the reference results can be used to adjust the calibration model so that the analyser measures all samples correctly. Temperature Effects NIR analysers are excellent thermometers. As such the sample temperature and the instrument temperature can affect NIR results. To correct for the effects of temperature, one of the following procedures can be used; 1. Add high and low temperature sample and instrument spectra into the calibration. 2. Include a temperature reading into the calibration model. 3. Measure the sample and instrument temperature and adjust the results using a linear calculation.
23 1) Cross Temperature Stabilisation: By scanning 5 samples at both high and low temperatures with the instrument at room temperature and then scanning the same five samples at room temperature but with the instrument at high and low temperatures, the calibration can be stabilised against temperature effects. Add these scans to the calibration set and recalibrate. The additional of the temperature samples will force the software to choose wavelengths that are stabilised against temperature changes. This method is the best for use with grains, powders or solid materials 2) Calibrate Against Temperature: In this method, an extra variable is added into the calibration set, i.e., sample temperature. The calibration procedure then includes temperature into the calibration model. This method requires a thermo couple or thermistor reading placed in the sample, to be read by the instrument. This is difficult to implement in laboratory type analysers but more suited to on line analysers. 3) Linear Temperature Correction: Over narrow temperature ranges, i.e., +/- 10C, the effects of temperature on the predicted results are linear. As such, by measuring the sample temperature, the analyser's computer can make a linear correction. For example, for measuring alcohol in wine the temperature coefficient is approximately -.05% per degree. The predicted result can be corrected by using the simple equation: %Alcohol (temp corrected) = %Alcohol + (20C - Sample Temp)*-.05 This method is ideal for liquids since a thermistor can be placed inside the measuring cell and a quick and reliable temperature reading obtained. Note that all three methods are corrections. As such there is always an error introduced from measuring temperature. The best way to reduce errors due to temperature changes is to maintain the sample and instrument temperatures as close to the temperature used to develop the calibration models.
24 Sample Preparation: Sample presentation is a major source of errors in NIT spectroscopy. The first issue is that the calibration models can be tuned to a very narrow variation in sample packing, grinding, mincing etc. As such, a model may work well at the time of calibration but fail as soon as the sample presentation changes. Just like in the case of temperature, you can add sample presentation variation into the calibration and thereby force the software to choose wavelengths that are less influenced by packing, grinding etc. Alternatively design the sample preparation and presentation procedure to ensure consistency. Weight of the sample is a very important factor in NIT spectroscopy. The amount of light that is absorbed by a sample is proportional to the mass concentration of the components. However most NIT methods do not weigh the sample but simply fill the sample cup. This assumes that the density of the material is always the same and that the operator fills the cup up the same every time. A better method is to weight the sample cup before and after loading and then adjust the predicted results based on mass. By connecting a balance to a NIT analyser and reading the weight of the sample, recording this into the analyser's computer board, then analysing the sample followed by a mass correction to the predicted result will provide far more accurate and reproducible results.
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
Back to Basics Fundamentals of Polymer Analysis
Back to Basics Fundamentals of Polymer Analysis Using Infrared & Raman Spectroscopy Molecular Spectroscopy in the Polymer Manufacturing Process Process NIR NIR Production Receiving Shipping QC R&D Routine
Qualitative NIR Analysis for Ingredients in the Baking Industry
Overview The challenge to all baking companies in today s economy is to operate plants as efficiently as possible, with a focus on quality and keeping costs in check. With regulatory issues becoming more
QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.
QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition
Asian Journal of Food and Agro-Industry ISSN 1906-3040 Available online at www.ajofai.info
As. J. Food Ag-Ind. 008, (0), - Asian Journal of Food and Agro-Industry ISSN 906-00 Available online at www.ajofai.info Research Article Analysis of NIR spectral reflectance linearization and gradient
ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION
Chem 306 Section (Circle) M Tu W Th Name Partners Date ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION Materials: prepared acetylsalicylic acid (aspirin), stockroom samples
Introduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
Infrared Spectroscopy 紅 外 線 光 譜 儀
Infrared Spectroscopy 紅 外 線 光 譜 儀 Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample (nondestructive method). The amount of light absorbed
Infrared Spectroscopy: Theory
u Chapter 15 Infrared Spectroscopy: Theory An important tool of the organic chemist is Infrared Spectroscopy, or IR. IR spectra are acquired on a special instrument, called an IR spectrometer. IR is used
Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare
Chemometric Analysis for Spectroscopy
Chemometric Analysis for Spectroscopy Bridging the Gap between the State and Measurement of a Chemical System by Dongsheng Bu, PhD, Principal Scientist, CAMO Software Inc. Chemometrics is the use of mathematical
Multivariate Chemometric and Statistic Software Role in Process Analytical Technology
Multivariate Chemometric and Statistic Software Role in Process Analytical Technology Camo Software Inc For IFPAC 2007 By Dongsheng Bu and Andrew Chu PAT Definition A system Understand the Process Design
Applications of Near Infrared Spectroscopic Analysis in the Food Industry and. Research
Applications of Near Infrared Spectroscopic Analysis in the Food Industry and Research Written by Rolf Nilsson for the Food Safety Centre, Tasmanian Institute of Agricultural Research, University of Tasmania,
Corona process The new ZEISS spectrometer system for the food industry
Interactive PDF Version 1.0 Start Trust needs reliability quality needs ZEISS The real challenge is to offer a product with consistent quality using raw materials which are themselves subject to ongoing
Symmetric Stretch: allows molecule to move through space
BACKGROUND INFORMATION Infrared Spectroscopy Before introducing the subject of IR spectroscopy, we must first review some aspects of the electromagnetic spectrum. The electromagnetic spectrum is composed
Ultraviolet Spectroscopy
Ultraviolet Spectroscopy The wavelength of UV and visible light are substantially shorter than the wavelength of infrared radiation. The UV spectrum ranges from 100 to 400 nm. A UV-Vis spectrophotometer
The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD
TN-100 The Fundamentals of Infrared Spectroscopy The Principles of Infrared Spectroscopy Joe Van Gompel, PhD Spectroscopy is the study of the interaction of electromagnetic radiation with matter. The electromagnetic
INFRARED SPECTROSCOPY (IR)
INFRARED SPECTROSCOPY (IR) Theory and Interpretation of IR spectra ASSIGNED READINGS Introduction to technique 25 (p. 833-834 in lab textbook) Uses of the Infrared Spectrum (p. 847-853) Look over pages
A Beer s Law Experiment
A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining
Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law
Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
Raman spectroscopy Lecture
Raman spectroscopy Lecture Licentiate course in measurement science and technology Spring 2008 10.04.2008 Antti Kivioja Contents - Introduction - What is Raman spectroscopy? - The theory of Raman spectroscopy
Determining the Structure of an Organic Compound
Determining the Structure of an Organic Compound The analysis of the outcome of a reaction requires that we know the full structure of the products as well as the reactants In the 19 th and early 20 th
2 Spectrophotometry and the Analysis of Riboflavin
2 Spectrophotometry and the Analysis of Riboflavin Objectives: A) To become familiar with operating the Platereader; B) to learn how to use the Platereader in determining the absorption spectrum of a compound
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
Molecular Spectroscopy
Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended
Organic Spectroscopy
1 Organic Spectroscopy Second Year, Michaelmas term, 8 lectures: Dr TDW Claridge & Prof BG Davis Lectures 1 4 highlight the importance of spectroscopic methods in the structural elucidation of organic
About Coffee and Refractometers 2008-2010 Voice Systems Technology, Inc. (VST)
About Coffee and Refractometers 2008-200 Voice Systems Technology, Inc. (VST) www.mojotogo.us.0 Coffee and Refractive Index Refractive Index measurements have been used for process control in the food
FTIR Instrumentation
FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation
ANEXO VI. Near infrared transflectance spectroscopy. Determination of dexketoprofen in a hydrogel. M. Blanco y M. A. Romero
ANEXO VI Near infrared transflectance spectroscopy. Determination of dexketoprofen in a hydrogel M. Blanco y M. A. Romero Enviado para su publicación NEAR INFRARED TRANSFLECTANCE SPECTROSCOPY. Determination
Features of the formation of hydrogen bonds in solutions of polysaccharides during their use in various industrial processes. V.Mank a, O.
Features of the formation of hydrogen bonds in solutions of polysaccharides during their use in various industrial processes. V.Mank a, O. Melnyk b a National University of life and environmental sciences
FT-NIR for Online Analysis in Polyol Production
Application Note: 51594 FT-NIR for Online Analysis in Polyol Production Key Words Acid Number Ethylene Oxide FT-NIR Hydroxyl Value Polyester Polyols Abstract Hydroxyl value and other related parameters
Application Note: Absorbance
Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light
Monograph. NIR Spectroscopy. A guide to near-infrared spectroscopic analysis of industrial manufacturing processes.
Monograph NIR Spectroscopy NIR Spectroscopy A guide to near-infrared spectroscopic analysis of industrial manufacturing processes A guide to near-infrared spectroscopic analysis of industrial manufacturing
Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs
Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs TracePro Opto-Mechanical Design Software s Fluorescence Property Utility TracePro s Fluorescence Property
0 10 20 30 40 50 60 70 m/z
Mass spectrum for the ionization of acetone MS of Acetone + Relative Abundance CH 3 H 3 C O + M 15 (loss of methyl) + O H 3 C CH 3 43 58 0 10 20 30 40 50 60 70 m/z It is difficult to identify the ions
O2PLS for improved analysis and visualization of complex data
O2PLS for improved analysis and visualization of complex data Lennart Eriksson 1, Svante Wold 2 and Johan Trygg 3 1 Umetrics AB, POB 7960, SE-907 19 Umeå, Sweden, [email protected] 2 Umetrics
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
Uses of Derivative Spectroscopy
Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
2 Absorbing Solar Energy
2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could
Industrial Process Monitoring Requires Rugged AOTF Tools
Industrial Process Monitoring Requires Rugged AOTF Tools Dr Jolanta Soos Growth has been rapid in the use of spectroscopic methods to monitor industrial processes, both in production lines and for quality
Indiana's Academic Standards 2010 ICP Indiana's Academic Standards 2016 ICP. map) that describe the relationship acceleration, velocity and distance.
.1.1 Measure the motion of objects to understand.1.1 Develop graphical, the relationships among distance, velocity and mathematical, and pictorial acceleration. Develop deeper understanding through representations
INTERNATIONAL OLIVE COUNCIL
INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 19/Rev. 3 February 2015 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: [email protected]
Concentrations and Dilutions of Food Dyes
Concentrations and Dilutions of Food Dyes Learning Goals: 1. Develop an understanding of the use of volumetric glassware. 2. Prepare a series of dye solutions of known concentrations. 3. Explore the relationship
DETERMINACIÓN DE ESTRUCTURAS ORGÁNICAS (ORGANIC SPECTROSCOPY) IR SPECTROSCOPY
DETERMINACIÓN DE ESTRUCTURAS ORGÁNICAS (ORGANIC SPECTROSCOPY) IR SPECTROSCOPY Hermenegildo García Gómez Departamento de Química Instituto de Tecnología Química Universidad Politécnica de Valencia 46022
Effects of Temperature, Pressure and Water Vapor on Gas Phase Infrared Absorption by CO 2
Effects of Temperature, Pressure and Water Vapor on Gas Phase Infrared Absorption by CO 2 D. K. McDermitt, J. M. Welles, and R. D. Eckles - LI-COR, inc. Lincoln, NE 68504 USA Introduction Infrared analysis
L 202/50 Official Journal of the European Union 31.7.2008 DECISIONS COUNCIL
L 202/50 Official Journal of the European Union 31.7.2008 II (Acts adopted under the EC Treaty/Euratom Treaty whose publication is not obligatory) DECISIONS COUNCIL DECISION No 1/2007 OF THE EC-TURKEY
Science Standard Articulated by Grade Level Strand 5: Physical Science
Concept 1: Properties of Objects and Materials Classify objects and materials by their observable properties. Kindergarten Grade 1 Grade 2 Grade 3 Grade 4 PO 1. Identify the following observable properties
5. The Nature of Light. Does Light Travel Infinitely Fast? EMR Travels At Finite Speed. EMR: Electric & Magnetic Waves
5. The Nature of Light Light travels in vacuum at 3.0. 10 8 m/s Light is one form of electromagnetic radiation Continuous radiation: Based on temperature Wien s Law & the Stefan-Boltzmann Law Light has
Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40
Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Note: there is a second document that goes with this one! 2046 - Absorbance Spectrophotometry - Calibration Curve Procedure. The second document
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
Time out states and transitions
Time out states and transitions Spectroscopy transitions between energy states of a molecule excited by absorption or emission of a photon hn = DE = E i - E f Energy levels due to interactions between
Title: Standard Operating Procedure for Thermo-Miran SapphIRe portable gas analyzer
Procedure No: SOP-039 Revision No: 1.1 January 21, 2011 Page No.: 1 of 11 1. INTRODUCTION AND SCOPE This document will outline specifics with regards to portable sampling analysis utilizing and operating
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
CHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING
CHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING Essential Standard: STUDENTS WILL UNDERSTAND THAT THE PROPERTIES OF MATTER AND THEIR INTERACTIONS ARE A CONSEQUENCE OF THE STRUCTURE OF MATTER,
Colorimetric Determination of Iron in Vitamin Tablets
Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered
Blackbody Radiation References INTRODUCTION
Blackbody Radiation References 1) R.A. Serway, R.J. Beichner: Physics for Scientists and Engineers with Modern Physics, 5 th Edition, Vol. 2, Ch.40, Saunders College Publishing (A Division of Harcourt
Chemistry 111 Lab: Intro to Spectrophotometry Page E-1
Chemistry 111 Lab: Intro to Spectrophotometry Page E-1 SPECTROPHOTOMETRY Absorption Measurements & their Application to Quantitative Analysis study of the interaction of light (or other electromagnetic
Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument
Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument Technical Note INTRODUCTION Direct measurements of nucleic acid samples at OD 260 or protein samples at OD
Current Staff Course Unit/ Length. Basic Outline/ Structure. Unit Objectives/ Big Ideas. Properties of Waves A simple wave has a PH: Sound and Light
Current Staff Course Unit/ Length August August September September October Unit Objectives/ Big Ideas Basic Outline/ Structure PS4- Types of Waves Because light can travel through space, it cannot be
The photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
Lab #11: Determination of a Chemical Equilibrium Constant
Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a
Electromagnetic Radiation (EMR) and Remote Sensing
Electromagnetic Radiation (EMR) and Remote Sensing 1 Atmosphere Anything missing in between? Electromagnetic Radiation (EMR) is radiated by atomic particles at the source (the Sun), propagates through
3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY
DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY Eduard Gilli 1,2 and Robert Schennach 1, 2 1 Graz University of Technology, 8010 Graz, Austria 2 CD-Laboratory for Surface Chemical and Physical
Organic Chemistry Tenth Edition
Organic Chemistry Tenth Edition T. W. Graham Solomons Craig B. Fryhle Welcome to CHM 22 Organic Chemisty II Chapters 2 (IR), 9, 3-20. Chapter 2 and Chapter 9 Spectroscopy (interaction of molecule with
5s Solubility & Conductivity
5s Solubility & Conductivity OBJECTIVES To explore the relationship between the structures of common household substances and the kinds of solvents in which they dissolve. To demonstrate the ionic nature
CHEM 51LB: EXPERIMENT 5 SPECTROSCOPIC METHODS: INFRARED AND NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY
CHEM 51LB: EXPERIMENT 5 SPECTROSCOPIC METHODS: INFRARED AND NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY REACTIONS: None TECHNIQUES: IR, NMR Infrared (IR) and nuclear magnetic resonance (NMR) spectroscopy are
Principle of Thermal Imaging
Section 8 All materials, which are above 0 degrees Kelvin (-273 degrees C), emit infrared energy. The infrared energy emitted from the measured object is converted into an electrical signal by the imaging
DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED
DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED Abstract: There are a number of sources of information that give a mathematical description of the terms used in light scattering. However, these will not
2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY
2nd/3rd Year Physical Chemistry Practical Course, Oxford University 2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY (4 points) Outline Spectrometry is widely used to monitor the progress
where h = 6.62 10-34 J s
Electromagnetic Spectrum: Refer to Figure 12.1 Molecular Spectroscopy: Absorption of electromagnetic radiation: The absorptions and emissions of electromagnetic radiation are related molecular-level phenomena
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
Alignment and Preprocessing for Data Analysis
Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise
MPA. Innovation with Integrity. Multi Purpose Analyzer FT-NIR
MPA Multi Purpose Analyzer Innovation with Integrity FT-NIR Discover the Flexibility of FT-NIR Spectroscopy The MPA is the result of almost 40 years of experience in the engineering and production of FT-IR
UV-Visible Spectroscopy
UV-Visible Spectroscopy UV-Visible Spectroscopy What is UV-Visible Spectroscopy? Molecular spectroscopy that involves study of the interaction of Ultra violet (UV)-Visible radiation with molecules What
Using the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy
Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Introduction The availability and applications of fiber optic sampling in UV/Vis
The accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 130 ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner
Copyright 1999 2010 by Mark Brandt, Ph.D. 12
Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
Module 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
AP Physics 1 and 2 Lab Investigations
AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks
How To Use Lactose In Confectionery
Importance of whey ingredients in confectionery products Andréa Pernot-Barry Technical Director ZDS Central College of the German Confectionery Industry Whey ingredients in confectionery Lactose Whey powder
Fractional Distillation and Gas Chromatography
Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils
Correlation key concepts:
CORRELATION Correlation key concepts: Types of correlation Methods of studying correlation a) Scatter diagram b) Karl pearson s coefficient of correlation c) Spearman s Rank correlation coefficient d)
Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information
1 Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information Instructions for the Operation of the Cary 300 Bio UV-Visible Spectrophotometer See the Thermo OMNIC Help reference on page 49. Ultraviolet-Visible
Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale
Nano-Spectroscopy Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale Since its introduction in the early 80 s, Scanning Probe Microscopy (SPM) has quickly made nanoscale imaging an affordable
Diagrams and Graphs of Statistical Data
Diagrams and Graphs of Statistical Data One of the most effective and interesting alternative way in which a statistical data may be presented is through diagrams and graphs. There are several ways in
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction
Treasure Hunt. Lecture 2 How does Light Interact with the Environment? EMR Principles and Properties. EMR and Remote Sensing
Lecture 2 How does Light Interact with the Environment? Treasure Hunt Find and scan all 11 QR codes Choose one to watch / read in detail Post the key points as a reaction to http://www.scoop.it/t/env202-502-w2
Group Theory and Chemistry
Group Theory and Chemistry Outline: Raman and infra-red spectroscopy Symmetry operations Point Groups and Schoenflies symbols Function space and matrix representation Reducible and irreducible representation
Performance and advantages of qnmr measurements
Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended
Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)
Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based
