ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS
|
|
|
- Susan Lucas
- 9 years ago
- Views:
Transcription
1 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner CSIRO Minerals, PMB 5, Menai 2234, NSW, Australia. [email protected] An XRF analyser suitable for measuring up to 15 liters of slurry has been tested in the laboratory. A range of slurries consisting of approximately 40 wt% solids and low concentrations of tantalum were obtained and measured. The accuracy of the analyser has been determined to be ±2 ppm (1 sigma). Improvements are planned that will significantly increase the speed and accuracy of the analyser and facilitate its adaptation to online / byline monitoring of mineral process streams. INTRODUCTION CSIRO Minerals is developing methods for the on-stream measurement of parts-permillion (ppm) concentrations of heavy metals in process slurries using K-shell X-ray fluorescence (XRF). As part of this ongoing development, an investigation was carried out on mineral slurries containing approximately 100 ppm of tantalum (Tantalum concentrations are reported as tantalum oxide (Ta2O5) as a proportion of the dry solids mass). MEASUREMENT OF SLURRIES WITH THE CSIRO ANALYSER Sample provision and third-party analysis CSIRO Minerals was supplied with eight samples from a tantalum production facility. Each sample consisted of at least 10 liters of slurry, containing wt% solids. The solids content was determined by drying the samples in an oven at 110 C for at least 2 days and weighing the residue. After each slurry had been dried and sub-sampled for analytical XRF analysis, the remaining dry ore was weighed and reconstituted with water for measurement in the CSIRO analyser. The original set of only eight slurries was considered too small for multivariate analysis. Therefore, additional samples were prepared based on this set. Firstly, the tantalum concentrations of two of the samples were increased in 5 ppm intervals by the addition of Ta2O5 powder. Secondly, iron oxide (Fe2O3) was added to one sample to increase its iron content to 2, 4 and finally 8 wt%. These additions were made to investigate the effects of changing matrix composition on the Ta measurement. In this way, the core set of 8 samples was expanded to 22 samples suitable for measurement.
2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -
3 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume The tantalum content of all samples was measured independently by three separate facilities using differing techniques, namely laboratory XRF analysis, neutron activation analysis (NAA) and ICP. A Grubbs [1] intercomparison analysis was performed and the most accurate measurement technique was determined to be the XRF analytical measurements with an estimated repeatability error of 1.6 ppm. The ICP measurements were found to have systematic errors related to the difficulty of completely dissolving all the tantalum into solution, whereas the NAA results showed relatively large statistical errors and hence neither were used in any further analysis. The CSIRO XRF Analyser The CSIRO XRF slurry analyzer is illustrated in figure 1. Two counter-rotating mixers are used to ensure slurry uniformity. The rig is designed to allow a large volume of slurry (approximately 10 liters) to be measured. The radiation source is a 37 MBq (1 mci) Co- 57 radioisotope source. The resulting scattered and fluorescent X-rays are transmitted through a polycarbonate window and are detected by a high resolution X-ray detector. The characteristic tantalum K X-rays are used to estimated the tantalum content rather than the L X-rays. The K X-rays are more suitable for two reasons: Firstly, they are more highly penetrating allowing a much larger volume to be analysed; and secondly they are also much less sensitive to matrix and particle size effects. Slurry sample Mixers Polycarbonate window Incident γ-ray Scattered X-ray Collimator X-ray detector in vacuum housing γ-ray source Figure 1. Schematic drawing of laboratory rig. A high-purity, planar germanium detector (Canberra 1000 mm2 LEGe) was used to detect the scattered and fluorescent X-rays, read-out using a Aptec 5004 combination shaping amplifier/adc/mca card. This was operated at a modest count rate (limited by the intensity of the source) of approximately 30k c/s with a corresponding dead time of 25%. Each measurement took approximately 5½ hours.
4 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume Uniformity of the Slurry During Measurement The most difficult problem faced in obtaining repeatable measurements was ensuring that the slurry was homogeneous. In particular, it was not sufficient just to lift all the solids off the bottom of the tank if a density profile remained where the larger particles were found near the bottom of the tank. This is important since the rig is more sensitive to material within a few centimeters of the source and detector. In the slurry samples used, the tantalum is not uniformly distributed with respect to particle size in fact there are systematic variations! These introduce significant errors in tantalum determination if the slurry is not well mixed. Calculation of Tantalum Concentration and Solids Mass Fraction Raw XRF Spectra Figure 2 shows a typical X-ray spectrum for a slurry containing approximately 100 ppm of tantalum. Each channel is 100 ev wide. The broad peak centered at 85 kev is due to Compton scattered source γ-rays, with the smaller peaks at channels 122 and kev due to Raleigh scattered source γ-rays. 4 x e l n a h pe rc ts C oun Channel number (100 ev bins) Figure 2. An X-ray spectrum obtained from measurement of a slurry containing approximately 100 ppm of Ta2O5. Figure 3 is an enlarged view of the spectrum in the vicinity of the tantalum Ka X-ray peaks. The tantalum Ka2 and Ka1 peaks are centered on channels 563 and 575.
5 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume x e l n a h pe rc ts C oun Channel number (100 ev bins) Tantalum Determination Figure 3. An enlarged view of the measured XRF spectrum in the vicinity of the tantalum Ka peaks. A four term linear regression analysis was used to relate the tantalum content of each sample to features observed in the X-ray spectrum. The terms used were the number of counts in the tantalum peaks, the background in the vicinity of the peaks, the number of counts in the Compton scattering peak, and a small term necessary to correct for a monotonic drift in the electronics with respect to time. To avoid over-fitting on the relatively small data set used, the 22 samples measured were divided into a training set (consisting of 21 points) used to determine the regression coefficients, and a test set (1 point) used to evaluate the performance of the calibration. The calibration/training process was cycled through all 22 of the samples. The presence of tungsten at 0 10 ppm levels was revealed by the NAA data. We were able to correct for any interference satisfactorily by estimating the background as the sum of three different windows two of which straddled the tungsten Ka1 peak. Solids Fraction Determination To determine the solids loading, a multivariate linear regression analysis was used to relate information on the shape of the Compton scattering peak to the solids fraction. The Compton peak shape is expected to vary with solids fraction for two reasons: 1. Doppler broadening of Compton scattered X-rays depends on the atomic number (Z) of the element from which the X-ray scatters, with higher Z elements having broader Compton scattering peaks. Compton profile analysis [2] has been shown to provide
6 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume good discrimination between the lightest elements (such as H and O) and heavier elements present in the tantalum ore. 2. The proportion of multiply to singly Compton scattered photons decreases with increasing solids loading fraction due to photoelectric absorption. This measurement technique has the advantage that no additional apparatus is needed and that the on-line mass measurement is performed over almost an identical region of slurry to which the tantalum concentration is being measured. RESULTS Tantalum Concentration Figure 4 shows a plot of the CSIRO calculated tantalum versus the analytical XRF laboratory determined values. The horizontal error bars represent the uncertainty of the analytical determination, and the vertical error bars on each point are the combined CSIRO and analytical XRF error. Figure 4. Plot of tantalum concentration for the CSIRO feasibility rig versus the analytical XRF concentration. The RMS error of 2.3 ppm (1 sigma) is the root mean square error between the CSIRO and analytical XRF analyses. It is a measure of the total error including the estimated 1.6 ppm repeatability error of the analytical analysis. Correcting for this results in an estimated error for the CSIRO analyzer of 1.7 ppm. There is always a degree of uncertainty in apportioning errors between two measurements and a more conservative claim would be that our data is consistent with a 1 standard deviation error of approximately 2 ppm.
7 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume The 2.3 ppm RMS error between the analytical XRF results and the CSIRO results is better than the agreement between analytical XRF and either the neutron activation analysis (about 3 ppm) or the ICP (>10 ppm) results. Whilst our results are very encouraging, it should be noted that they have been obtained on a fairly limited suite of measurements, based on just 8 discrete samples. Further measurements are required to properly evaluate other factors contributing to the analyzer accuracy, notably matrix and particle size effects. Solids Mass Fraction The on-line determinations of the solids mass fraction were compared with the true mass fractions, calculated from the masses of solid and water used to reconstitute the slurries. The measurement accuracy for the solid fraction determination was found to be about 0.3 wt%, with a repeatability error of 0.16 wt%. For a typical solids loading of about 40 wt%, this leads to a 0.7% relative error in the determination of the Ta2O5 content, expressed as a fraction of the dry mass. CONCLUSION Measurements undertaken by CSIRO Minerals at the Lucas Heights Laboratory have shown that the current XRF rig is capable of accurately measuring the tantalum content of slurries. An accuracy of approximately 2 ppm was obtained for a measurement time of 5½ hours. Several improvements to the analyzer are planned. These include: Using a strong radioisotope source together with a high-speed digital X-ray processor to increase the X-ray counting rate; Altering the geometry of the analyzer to measure slurry passing through a pipeline. This improves the signal-to-background ratio for the tantalum X-rays and minimizes the sensitivity of the analyzer to the spatial distribution of the solids fraction. It is estimated that it will be possible to achieve a statistical error on the measurement of Ta2O5 of approximately 1 ppm with a 20 minute counting time. In principle, the same technique can be used to determine other high Z elements to similar accuracy. ACKNOWLEDGEMENTS The authors would like to thank Steve Rainey and Ivan Kekic for assistance in the design and construction of the slurry analyser. REFERENCES [1] Grubbs F, 1948, On Estimating Precision of Measuring Instruments and Product Variability, J Amer. Statistical Assoc., 43, [2] Tickner J and Roach G, Characterization of coal and minerals using Compton Profile Analysis, to appear in Nuclear Instruments and Methods B
EDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
GAMMA-RAY SPECTRA REFERENCES
GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.
CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY
243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms
Quantifying Amorphous Phases. Kern, A., Madsen, I.C. and Scarlett, N.V.Y.
Kern, A., Madsen, I.C. and Scarlett, N.V.Y. This document was presented at PPXRD - Pharmaceutical Powder X-ray Diffraction Symposium Sponsored by The International Centre for Diffraction Data This presentation
Introduction to Geiger Counters
Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists
Introduction to the Monte Carlo method
Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw
Spectral distribution from end window X-ray tubes
Copyright (C) JCPDS-International Centre for Diffraction Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg 2 1 FORTEX - E.N.S.A.I.S. 24, bld de la Victoire,
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 1 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 367
Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies
Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies (Ci), where 1 Ci = 3.7x10 10 disintegrations per second.
Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist
Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist University of Puerto Rico Mayagüez Campus Department of Geology X-ray / Microanalysis Laboratory F - 304
Coating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems
Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Ronald M. Keyser, Timothy R. Twomey, Sam Hitch ORTEC 801 South Illinois Avenue Oak Ridge, TN, 37831
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF
Quantitative Analysis Software for X-Ray Fluorescence XRF-FP XRF-FP is a full-featured quantitative analysis package for XRF APPLICATIONS X-Ray Fluorescence Thin-film Analysis RoHS/WEEE Analysis Teaching
Radiation Detection and Measurement
Radiation Detection and Measurement June 2008 Tom Lewellen [email protected] Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron
Atomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
AN INTRODUCTION TO THE RADOS XRF ORE SORTER. RS Fickling Metanza Mineral Processors (Pty) Ltd
6 th Southern African Base Metals Conference 2011 AN INTRODUCTION TO THE RADOS XRF ORE SORTER RS Fickling Metanza Mineral Processors (Pty) Ltd 1. ABSTRACT RADOS has developed ore sorters in Russia over
Figure 1. A typical Laboratory Thermometer graduated in C.
SIGNIFICANT FIGURES, EXPONENTS, AND SCIENTIFIC NOTATION 2004, 1990 by David A. Katz. All rights reserved. Permission for classroom use as long as the original copyright is included. 1. SIGNIFICANT FIGURES
The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün
N.B. A shorter version of this article was published in the ICOM-CC Ethnographic Conservation Newsletter, Number 30, January 2009, pp. 5-8. The Limitations of Hand-held XRF Analyzers as a Quantitative
MASS DEFECT AND BINDING ENERGY
MASS DEFECT AND BINDING ENERGY The separate laws of Conservation of Mass and Conservation of Energy are not applied strictly on the nuclear level. It is possible to convert between mass and energy. Instead
Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy
Information technology (IT) for teaching X- and gamma-ray transport: the computer codes MUPLOT and SHAPE, and the web site dedicated to photon transport Jorge E. Fernández Laboratory of Montecuccolino
X-Ray Fluorescence - Data Collection and Analysis
XRF User Manual 1 X-Ray Fluorescence - Data Collection and Analysis Matthias Zeller, Jerry Hunter, Sept. 19 th 2008 A) Data Collection Logging In: The login and password to log into the program as a normal
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
Balance Accuracy. FlexiWeigh
FlexiWeigh The balance is an integral part of the FlexiWeigh workstation. The role and limitations of the balance are important to understand in this environment. This technical note describes the different
XRF MAPPING: NEW TOOLS FOR DISTRIBUTION ANALYSIS
Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 19 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42
FTIR Instrumentation
FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system
EDS system Most common X-Ray measurement system in the SEM lab. Major elements (10 wt% or greater) identified in ~10 secs. Minor elements identifiable in ~100 secs. Rapid qualitative and accurate quantitative
How To Understand Light And Color
PRACTICE EXAM IV P202 SPRING 2004 1. In two separate double slit experiments, an interference pattern is observed on a screen. In the first experiment, violet light (λ = 754 nm) is used and a second-order
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.
X-RAY DIFFRACTION IMAGING AS A TOOL OF MESOSTRUCTURE ANALYSIS
Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 241 X-RAY DIFFRACTION IMAGING AS A TOOL OF MESOSTRUCTURE ANALYSIS ABSTRACT J. Fiala, S. Němeček Škoda
AP Physics 1 and 2 Lab Investigations
AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
Appendix A. An Overview of Monte Carlo N-Particle Software
Appendix A. An Overview of Monte Carlo N-Particle Software A.1 MCNP Input File The input to MCNP is an ASCII file containing command lines called "cards". The cards provide a description of the situation
2 Absorbing Solar Energy
2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine
Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror
Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum
Lectures about XRF (X-Ray Fluorescence)
1 / 38 Lectures about XRF (X-Ray Fluorescence) Advanced Physics Laboratory Laurea Magistrale in Fisica year 2013 - Camerino 2 / 38 X-ray Fluorescence XRF is an acronym for X-Ray Fluorescence. The XRF technique
Introduction to X-Ray Powder Diffraction Data Analysis
Introduction to X-Ray Powder Diffraction Data Analysis Center for Materials Science and Engineering at MIT http://prism.mit.edu/xray An X-ray diffraction pattern is a plot of the intensity of X-rays scattered
BORON IN GLASS DETERMINATION USING WDXRF
269 ABSTRACT BORON IN GLASS DETERMINATION USING WDXRF Alexander Seyfarth Bruker AXS Inc., 5465 East Cheryl Parkway, Madison, Wisconsin 53711-5373 The application of wavelength-dispersive XRF to the analysis
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING
X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING Brian L. Riise and Michael B. Biddle MBA Polymers, Inc., Richmond, CA, USA Michael M. Fisher American Plastics Council, Arlington, VA, USA X-Ray Fluorescence
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
Table 1 DENSITY Density Solids Percent By Volume
Willard Says Useful stuff to know about hydraulic dredging. The Discharge Pipeline Slurry Density Slurry density is: Poorly understood. The ratio of the weights of a volume of slurry relative to an identical
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy Upp, Daniel L.; Keyser, Ronald M.; Gedcke, Dale A.; Twomey, Timothy R.; and Bingham, Russell D. PerkinElmer Instruments, Inc. ORTEC,
ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY. Donald L. Horrocks
ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY Donald L. Horrocks Nuclear Systems Operations Beckman Instruments, Inc. SmithKlirie Beckman Irvine,
Effects of Temperature, Pressure and Water Vapor on Gas Phase Infrared Absorption by CO 2
Effects of Temperature, Pressure and Water Vapor on Gas Phase Infrared Absorption by CO 2 D. K. McDermitt, J. M. Welles, and R. D. Eckles - LI-COR, inc. Lincoln, NE 68504 USA Introduction Infrared analysis
ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy
Equipment Needed from ORTEC GEM10P4/CFG-PV4/DWR-30 Coaxial Detector System (Includes detector, cryostat, dewar, preamplifier, and 12-ft. cable pack); typical specifications: 10% relative efficiency, 1.75
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics
Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics
X Ray Flourescence (XRF)
X Ray Flourescence (XRF) Aspiring Geologist XRF Technique XRF is a rapid, relatively non destructive process that produces chemical analysis of rocks, minerals, sediments, fluids, and soils It s purpose
Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)
Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
Tutorial 4.6 Gamma Spectrum Analysis
Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning
Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:
Gamma Rays OBJECT: To understand the various interactions of gamma rays with matter. To calibrate a gamma ray scintillation spectrometer, using gamma rays of known energy, and use it to measure the energy
Chapter Test B. Chapter: Measurements and Calculations
Assessment Chapter Test B Chapter: Measurements and Calculations PART I In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1.
Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector
International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector
The Basic Principles of Sieve Analysis
The Basic Principles of Sieve Analysis Introduction Many natural and manufactured materials occur in a disperse form, which means that they consist of differently shaped and sized particles. The particle
Application Note: Absorbance
Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
The photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE
Portable X-ray fluorescence Spectroscopy Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE OBJECTIVES Background of the method Features of the instrument Applications
The Mainz LXe TPC MC simulations for a Compton scattering experiment
The Mainz LXe TPC MC simulations for a Compton scattering experiment Pierre Sissol Johannes Gutenberg Universität Mainz 12 November 2012 1 / 24 Outline 1 Dark Matter 2 Principle of a dual-phase LXe TPC
Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July 2001. Reprinted from Photonics Spectra, Laurin Publishing
Reprint (R22) Avoiding Errors in UV Radiation Measurements By Thomas C. Larason July 2001 Reprinted from Photonics Spectra, Laurin Publishing Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1
Using the PDF for material identification using elemental data. from XRF and SEM EDS.
XRF and SEM EDS Using the PDF for material identification using elemental data from XRF and SEM EDS. XRF and SEM EDS What? The Powder Diffraction File contains data on pure solid state compounds of well
Metals in Soil Analysis Using Field Portable X-ray Fluorescence
Metals in Soil Analysis Using Field Portable X-ray Fluorescence A guideline to using portable XRF according to EPA Method 6200, basic overview of the technique of x-ray fluorescence (XRF), appropriate
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
Determination of Insoluble Solids in Pretreated Biomass March 2008 Material
National Renewable Energy Laboratory Innovation for Our Energy Future A national laboratory of the U.S. Department of Energy Office of Energy Efficiency & Renewable Energy Determination of Insoluble Solids
1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number
1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number A. 3(x - x) B. x 3 x C. 3x - x D. x - 3x 2) Write the following as an algebraic expression
New Portable X-Ray Diffraction/X-Ray Fluorescence Instrument (XRD/XRF)
1 New Portable X-Ray Diffraction/X-Ray Fluorescence Instrument (XRD/XRF) Introduction The primary goal of most analyses of art objects is to identify the material composition of the object. This may lead
Application Note # EDS-10 Advanced light element and low energy X-ray analysis of a TiB 2 TiC SiC ceramic material using EDS spectrum imaging
Quantitative analysis Ceramics sample Peak deconvolution EDS map Phase analysis Application Note # EDS-10 Advanced light element and low energy X-ray analysis of a TiB 2 TiC SiC ceramic material using
Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods
Short report Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods A. Chaparian 1*,M.A. Oghabian 1, V. Changizi
SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
- particle with kinetic energy E strikes a barrier with height U 0 > E and width L. - classically the particle cannot overcome the barrier
Tunnel Effect: - particle with kinetic energy E strikes a barrier with height U 0 > E and width L - classically the particle cannot overcome the barrier - quantum mechanically the particle can penetrated
ABSORPTION OF BETA AND GAMMA RADIATION
ABSORPTION OF BETA AND GAMMA RADIATION The purpose of this experiment is to understand the interaction of radiation and matter, and the application to radiation detection and shielding Apparatus: 137 Cs
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
Florida Method of Test for TESTING OF GROUND TIRE RUBBER Designation: FM 5-559
Florida Method of Test for TESTING OF GROUND TIRE RUBBER Designation: FM 5-559 1 SCOPE 1.1 This method is used to determine the physical requirements of ground tire rubber for use in asphalt rubber using
DESCRIPTIVE STATISTICS. The purpose of statistics is to condense raw data to make it easier to answer specific questions; test hypotheses.
DESCRIPTIVE STATISTICS The purpose of statistics is to condense raw data to make it easier to answer specific questions; test hypotheses. DESCRIPTIVE VS. INFERENTIAL STATISTICS Descriptive To organize,
OLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
CHI-SQUARE: TESTING FOR GOODNESS OF FIT
CHI-SQUARE: TESTING FOR GOODNESS OF FIT In the previous chapter we discussed procedures for fitting a hypothesized function to a set of experimental data points. Such procedures involve minimizing a quantity
X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
STATS8: Introduction to Biostatistics. Data Exploration. Babak Shahbaba Department of Statistics, UCI
STATS8: Introduction to Biostatistics Data Exploration Babak Shahbaba Department of Statistics, UCI Introduction After clearly defining the scientific problem, selecting a set of representative members
Vacuum Evaporation Recap
Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)
Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know
Analysis of Soil Samples Using a Portable X-Ray Fluorescence Spectrometry (XRF)
Auckland Wellington Christchurch PATTLE DELAMORE PARTNERS LTD Analysis of Soil Samples Using a Portable X-Ray Fluorescence Spectrometry (XRF) Marlborough District Council solutions for your environment
Quality Control on Aerospace Components Using Handheld X-ray Fluorescence (XRF)
International Workshop SMART MATERIALS, STRUCTURES & NDT in AEROSPACE Conference NDT in Canada 2011 2-4 November 2011, Montreal, Quebec, Canada Quality Control on Aerospace Components Using Handheld X-ray
An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry
Limits to Detection/Measurement Ionization is fundamentally a Probabilistic Process Just like radioactive decay So is transmission through the analyzer There is an intrinsic statistical uncertainty Proportional
Guideline for Control Surveys by Differential Levelling
Guideline for Control Surveys by Differential Levelling Special Publication 1 Intergovernmental Committee on Surveying and Mapping (ICSM) Permanent Committee on Geodesy (PCG) 24 September 2014 Intergovernmental
Introduction: Growth analysis and crop dry matter accumulation
PBIO*3110 Crop Physiology Lecture #2 Fall Semester 2008 Lecture Notes for Tuesday 9 September How is plant productivity measured? Introduction: Growth analysis and crop dry matter accumulation Learning
99.37, 99.38, 99.38, 99.39, 99.39, 99.39, 99.39, 99.40, 99.41, 99.42 cm
Error Analysis and the Gaussian Distribution In experimental science theory lives or dies based on the results of experimental evidence and thus the analysis of this evidence is a critical part of the
Sample preparation for X-ray fluorescence analysis
Technical articles Sample preparation for X-ray fluorescence analysis III. Pressed and loose powder methods Gakuto Takahashi* 1. Introduction There are two main sample preparation techniques for measurement