Win Scale YSW03. Operating Instructions. Verifiable PC driver program with verifiable display and alibi memory for scales and weighing platforms
|
|
|
- Emory Jenkins
- 10 years ago
- Views:
Transcription
1 Operating Instructions Win Scale YSW03 Verifiable PC driver program with verifiable display and alibi memory for scales and weighing platforms
2 Intended use M Note: With regard to operation and user programming, special attention must be paid to the sections marked with the M symbol. The Sartorius Win Scale application is a type-approved driver program for a verifiable display and alibi memory used to store values measured by one or several balances, scales or weighing platforms (weighing modules). Communication between a PC and the balance, scale or weighing platform takes place over a serial RS-232/RS-485 data interface. The devices must be suitable, set up and certified for the transmission protocol xbpi (extended Balance Processor Interface) or SBI (Sartorius Balance Interface), which are used for communication. In the configuration of xbpi for some model series, it is possible to show the verified display and operating settings on the screen of the PC. No additional operating terminal for the weighing platform is then required. In the configuration of SBI for other model series, the program is suitable for secure storage of the verified display values on the PC (alibi memory). The display and operating settings of the scale then continue to be used. It is therefore possible, for example, to connect scales of the FB, FC, FD, LA and LP series, isi-industrial terminals and IS weighing platforms. If you use Sartorius Win Scale in conjunction with an IS weighing platform, an isi industrial display and operator terminal is not necessary. Several devices (up to 31) can be connected via a special network ( xbus network ) in which each device is identifiable by a unique address. The Sartorius Win Scale application is used: to emulate a verifiable display and operator terminal (a verifiable window ) on the PC. The verifiable window is placed on the screen in a location selected freely by the user and cannot be overwritten there by any other application. for verifiable storage of data (in the alibi memory; please see the section entitled Alibi Memory above on the right). as an interface between the user, other software processes running in parallel (DDE clients: Dynamic Data Exchange) and the scale or weighing platform. Example: Data exchange between the scale or IS weighting platform and the PC takes place over a special, verifiable communication library that is adapted to the external device. Data from the device is stored in alibi memory, which provides reliable protection against manipulation. The alibi memory enables the user to retrieve and further process verifiable data directly from the balance, scale or weighing platform using his own software. Furthermore, the user can take advantage of the range of commands available to him to initiate various functions on the balance, scale or IS platform (e.g. taring/ zeroing the display, starting calibration/ adjustment, accessing or changing the balance or scale operating menu). On verifiable devices, the range of commands is limited accordingly. Alibi memory The determination of the mass of an object for trade purposes is subject to calibration and verification requirements. If, for example, an invoice indicating the mass of an object is generated via a connected electronic data processing device (the number of intermediate stations does not matter here), calibration and verification regulations apply all the way to the printout of that invoice or any other record. The connected electronic data processing device can be a PC or computer network, for which it would be unrealistic to demand type approval certificates or verification. However, in order to comply with the requirements of this legally regulated area, the concept of alibi memory was created. This is a type-approved, protected memory. Data to be further processed in transactions subject to calibration requirements is stored in this memory, where it cannot be manipulated. The data can be reaccessed on command at any time. When the data is processed, it is only necessary to indicate an ID that subsequently allows for unique retrieval of the stored data. It is the responsibility of the operator to ensure that this takes place. The archiving duration of the data is regulated at the national level. Compliance with the established periods is also the responsibility of the operator. Since this may become problematic for large amounts of data and long storage times, the Sartorius memory concept permits the stored data to be securely exported and reimported as required. M Note: Windows 95, Windows 98, Windows 2000, Windows NT, Windows XP, MS Office, MS Excel and MS Word are registered trademarks of Microsoft Inc. All other names are trademarks of the corresponding companies. Hotline $ Please direct technical questions on design, specifications and installation to your operating partner or directly to Sartorius, at the hotline: Telephone: +49 (0) 551 / Fax: +49 (0) 551 / Explanation of symbols The following symbols are used in the text: Instructions for correct operation $ Special instructions for exceptional cases > Description of the consequences of an operating step! Warnings M Legal metrology Legal and copyright information This documentation may not be reproduced or distributed either in whole or in part without the explicit written permission of Sartorius AG. This documentation is intended only for use by the buyer. Transfer to third parties, whether free or in exchange for payment, is not permitted. The software on the included CD-ROM is the property of Sartorius AG. The software may not be copied or modified, it may not be decompiled or modified through assimilation. The buyer may only use the software for his own purposes and may not make the software available to third parties either for use for free or in exchange for payment. If you encounter problems with the included software, please contact the software distributor. Sartorius AG reserves the right to deliver updated software for this product. No liability is accepted for software installed before the purchase of this product. The buyer is liable for misuse of the software.
3 Contents Instructions for the Use of Sartorius Win Scale in Legal Metrology M Page 2 Intended Use M 2 Alibi Memory M 3 Instructions for the Use of Sartorius Win Scale in Legal Metrology 4 System Overview M 4 Software System M 4 xbus Network 5 SBI 5 Installation 5 System Requirements 5 Hardware Requirements 5 Software Requirements 5 Scope of Delivery 5 Installing Sartorius Win Scale 6 Program Installation 9 First Use 10 Configuring the Communication Library 10 Changing the Device Name 13 Creating the Program Icon 14 Connecting Devices to the PC 14 Connecting the FB/FC series to the PC 14 Connecting the LA/LP series to the PC 14 Connecting an isi Operating Terminal/FD Series to the PC 14 Connecting an IS Weighing Platform to the PC 15 Operation M 15 Loading Sartorius Win Scale M 16 Information in the Display M 17 Exiting Sartorius Win Scale 17 Selecting a Device 17 Taring a Device 17 Zeroing a Device 18 Writing the Weight to Alibi Memory M 18 Accessing the Service Menu M 19 Changing the Service Menu Password 20 Reading Information from the Connected Device 21 Reading Program Information M 22 Defining the Communication Library 24 Configuring Communication with the Device 26 Defining DDE Connection Parameters 27 Configuring the Alibi MemoryM 28 Performing the Function Test of the Alibi Memory 30 Reading the Alibi Memory and Exporting Data 32 Configuring the Display Position 33 Configuring the Display Colors 34 Communication with External Programs (DDE Clients) 34 DDE Interface 34 Data Exchange 34 Creating the Data Request to the DDE Server (Request String) 34 Implementation Principles for User Software 35 DDE Application Example 1 36 DDE Application Example 2 40 DDE Functions 41 License Agreement 42 Error Messages 42 Frequently Occurring Errors and Troubleshooting 43 Test Certificate M M M M M M Sartorius Win Scale can be used in legal metrology either as an additional device or as a display and operating terminal (for balances and scales without an integrated display). In both cases, calibration by a duly authorized representative of Sartorius AG or calibration authorities is required. Subsequent installation of Sartorius Win Scale for use in conjunction with a balance or scale that has already been calibrated requires recalibration by the calibration authorities. The section entitled Information in the Display in the Operation section describes the inspection of the display required during calibration to ensure that the data is presented correctly. In addition, the mandatory inspection of correct storage of the weight values subject to calibration in the alibi memory is explained under Function Test of the Alibi Memory in the section entitled Accessing the Service Menu in the Operation section. The mandatory comparison of the software ID with a target value is described under Reading Program Information in the section entitled Accessing the Service Menu in the Operation section. When using the Sartorius Win Scale application, it is imperative that you follow the rules listed below: The display window of Sartorius Win Scale may not be imitated. This means that no similar display window can be generated, since this could lead to confusion with the display window of Sartorius Win Scale. The display window of Sartorius Win Scale in XBPi mode may not be hidden. The application has been designed so that the display window reappears in the foreground of the screen at least after any change in the readout from a connected weighing instrument. It is therefore up to the operator to ensure that no attempts are made to place the display window in the background, to hide parts of the window or to falsify the window with outside information. - In SBI mode, no verifiable readout may be shown in the display window; for this reason, hiding of the window is permitted. In addition to inspection of the findings with regard to the information displayed on the monitor (when using as a verifiable display), a function test is sufficient for the verifiable memory. - The following function test of the verifiable memory is sufficient: Performance of 5 weighings with saving of the data in each case (e.g. press the printout button on the scale). The weighing results should be recorded for subsequent checking. Inspection of the stored weighing results for agreement with the previously recorded values: Click on S in monitor display to access the service menu. In the Password dialog, enter the user or system password and confirm with OK. In the Service Menu dialog, select Verifiable Memory and confirm with OK. In the Display Under Day File dialog, select the desired day and under Search enter the desired number under Verifiable ID, if necessary. The user must check the integrity of the legal metrology software modules every day. This is accomplished by accessing the software information (for this purpose, please refer to the sections entitled Accessing the Service Menu and Reading Program Information in the Operation section). A weight value may only be used and/or further processed in a verified context if it has been stored in the alibi memory. The storage time for legal metrology data depends on the application. It is the responsibility of the operator to ensure compliance. The operator is responsible for providing storage space for the alibi memory on the local hard drive. Configuration of the ring storage of the alibi memory must be performed by the operator (see here the section Operation Configuring the Alibi Memory ). This also applies to measures for backing up the data. The alibi memory is protected against manipulation by checksums. Nevertheless, the operator may not attempt to bypass this protection by analyzing or manipulating the files. The computer used must conform to EEC Council Directive 89/336/EEC and must bear the CE symbol. All connected peripheral devices subject to calibration requirements (keyboard, monitor, etc) must also bear this certification. The PC platform must fulfill the minimum requirements specified under System Requirements in the Installation section.
4 System Overview M Software System The figure below shows the individual components of Sartorius Win Scale as a server in conjunction with a balance, scale or weighing platform and an external software application (client). Externally connected devices (balances, scales or weighing platforms) communicate with the help of modularly designed, device-specific drivers known as verifiable communication libraries. Each library contains a defined range of functions (SARTOCOM protocol) as a software interface to a non-verifiable application and specific commands for communication with the external device. When the application is running, an appropriate verifiable communication library is loaded for the particular external device with which the program is to communicate. All program-specific settings are stored in a configuration file. When the application is closed, a copy is made of this configuration file. If the original configuration file is destroyed, a valid configuration file can be restored from this copy. xbus Network All devices equipped with a serial RS-485 data interface and set to the xbpi communication mode can be linked together in an xbus network (see the figure on the left). In this special Sartorius network, up to 31 devices (balances, scales or IS weighing platforms) can be connected together. On scales or balances (such as the FB/FC or LA/LP series), the standard RS- 232 data interface must be replaced with an RS-485 interface. This is available as an accessory for devices of the FB/FC series (YDO01F and YDO02F for scales and balances with high loads 12 kg) and for the LA/LP series upon request in the Fast Factory department. To connect an xbus network or an individual device equipped with an RS-485 data interface (e.g. an IS weighing platform) to a PC equipped with a standard RS-232 interface, use one of the two options described below: Installation of an RS-485 interface card in the PC. For this purpose, Sartorius offers the W&T PC card (order no.: 62PC-Y014). This is a dual interface that can be installed in any commercially available PC with a standard AT bus. A detailed description is included with the card. Use of the interface converter (see figure on page 5). The RS-232 data interface on the converter is connected to the RS-232 interface on the PC via cable (25-pin D-Sub female connector at both ends). The wiring diagram of the RS-485 interface on the converter and all other details are included in the accompanying operating instructions. Because of the individual cable lengths used for data transmission according to the RS-485 standard, there is no standard cable for this purpose. If you operate IS weighing platforms in the xbus network, they must have their own power supply. If necessary, the IS weighing platform must be converted (exception: IS weighing platforms that are controlled by an isi industrial terminal and scales of the FD series). For information on connecting IS weighing platforms, consult the section Connecting an IS Weighing Platform to the PC under Connecting Devices to the PC. To ensure unique identification in the network, each component must have a unique bus address that is stored in the device. The addresses may have the values 1 to 31. The data exchange within the network takes place in binary. Cyclical data output from a device is not possible in bus operation mode. A detailed description of the xbus can be found in the Protocol and Function Description of the xbpi Interface of the isi Operating Terminals. In addition to the EC type approval of the balance or scale, the associated test certificate with appendix must be presented for calibration. It must be determined whether the verifiable program is operating properly. The actual value of the software ID is calculated and displayed by loading the menu item Information, Software in the service menu. The saved set value (CB for software version / 8228A94E for software version ) is also displayed. If both values match, the software has been loaded properly and the computer may be used for purposes subject to calibration requirements. Computer with Windows95/Windows NT-compatible operating system Monitor Balance/scale or weighing platform Sartorius WinScale Verifiable window Verifiable communication library User interface Alibi memory (data with verifiable ID) Nonverifiable application Printer (data with verifiable ID) Computer Device 1 Device 2... Device RS-232 converter RS-485 xbus network interface interface
5 SBI If weighing platforms are equipped with an RS-232 data interface, they can be addressed within the SBI protocol. In SBI mode, data can be received from up to three connected devices. Note: In legal metrology, only the connection of a single balance or scale is permitted in SBI mode! No multi-range scales or multi-division scales may be connected when using the SBI protocol. In addition to the EC type approval of the balance or scale, the associated test certificate with appendix must be presented for calibration. It must be determined whether the verifiable program is operating properly. The actual value of the software ID is calculated and displayed by loading the menu item Information, Software in the service menu. The saved set value (8228A94E version for SBI) is also displayed. If both values match, the software has been loaded properly and the computer may be used for purposes subject to calibration requirements. If the recorded data should be stored in the alibi memory, the configuration data must first be entered based on the information on the test certificate: in case of calibration, the calibration value e must be entered. Combined tare/null or blockprint setting, indication of d for scales d not equal to e. For balances and scales with combined taring and zeroing function, Combined tare/null must be selected, for balances and scales with separate taring and zeroing functions, Blockprint must be selected. Serial number, decimal places, etc. are entered in the configuration dialog windows since this data is not automatically collected (as in XBPI mode). Installation System Requirements Hardware Requirements: Computer: IBM-compatible PC, Processor: Pentium or equivalent, Clock speed > 133 MHz, RAM > 32 MB, Hard drive space > 1 GB CE certification according to the EMC Directive 89/336/EEC Display: Diagonal display size > 10, Windows setting Small font Resolution: Maximum 800x600 pixels for a diagonal size of <12 Maximum 1024x768 pixels for a diagonal size of <15 Maximum 1152x864 pixels for a diagonal size of <17 At least 256 colors in all resolutions Data interface: RS-232/RS-485 Software Requirements: Operating system: MS Windows 95, MS Windows 98, MS Windows 2000, MS Windows NT, MS Windows XP All standard drivers of the respective operating system are used. These do not change the verifiable data. Transmission protocol: xbpi or SBI Note: MS Windows 98, MS Windows 2000, MS Windows NT, MS Windows XP are registered trademarks of Microsoft Inc. Scope of Delivery The program and all additionally required files (e.g. initialization file, configuration file and help file) are delivered on a CD. Installing Sartorius Win Scale Note: Exit all Windows programs before installing the Sartorius Win Scale application. The hard disk of the computer (i.e. the C:\ drive) must have sufficient storage capacity to accommodate the program (> 3.5 MB).
6 Program Installation Insert the CD into the CD drive of the computer. Start Setup.exe by double-clicking on it. > This starts the installation program from the CD. > The initial screen of the installation program for the Sartorius Win Scale application is displayed. Click the button [Next]. > The software license agreement for the program is displayed. The user must accept the license agreement for this program. You can view the rest of the text by sliding the scrollbar with the mouse pointer or by clicking on the (Arrow) button with the mouse. Clicking on the button [Cancel] will stop the installation, [Back] will return you to the previous screen. Accept the license agreement by clicking on the radio button. Click the button [Next].
7 To continue the installation: Click the button [Next]. > The installation program allows the user to configure the installation according to his own requirements. > A directory tree and the directory in which Sartorius Win Scale and all accompanying, necessary files will be saved are displayed. Clicking on the button [Next] confirms these settings. > Clicking on the button [Browse] allows the user select the drive (in the field Drives ) and to select the directory within the directory tree (in the field Directories ) in order to install the program where desired. By editing the entry in the Path field, the user can directly enter the desired path. Note: If you do not select the standard installation directory C:\Program Files\Sartorius\Sartorius Win Scale as the installation path, you will receive the error message ERR APP 13 the first time you access the program. You can alleviate this error by setting the correct directory path for the communication library (for more information, please refer to the section entitled Configuring the Communication Library at the end of the Installation section). Enter the installation path as desired. To continue the installation: Click the button [OK]. > The actual installation procedure begins now. Select the language version, e.g. English, and click the button to confirm.
8 > You have now made all the necessary entries for installation of Sartorius Win Scale. The actual installation process takes place automatically. The installation program automatically copies all necessary files into the target directory. > At the end of the installation procedure, you can choose to have the ReadMe file displayed and/or start the Sartorius Win Scale application. Click on the check boxes with the mouse to select or unselect the individual options. By default, Show ReadMe File is checked and Start Sartorius Win Scale is unchecked (application will not start). > The ReadMe file is displayed on the screen. The user can print out the file by selecting the Print option under the File menu. To close the ReadMe file, either click on the x button in the upper righthand corner of the window or select the Exit option under the File menu. > If the option Start Sartorius Win Scale was selected, the application is started. If both options were selected, the application is started and the ReadMe file is displayed at the same time. If neither of the two options was selected, the installation process ends with the Setup completed window. This completes the installation of Sartorius Win Scale. Select the options and click the button [Finish]. This completes the installation process. > The program searches for a connection to the connected balance or scale. A virtual display is shown on the screen. If no device is connected to the serial interface COM1 (default setting at program start, see also the section Configuring Communication with the Device in the Operation section), the following error message appears: ERR DLL -6 ( cannot connect to scale ). If a balance, scale or weighing platform is connected to the COM1 interface and properly set to operate in the xbpi communication mode (see the section entitled Connecting Devices to the PC ), a weight readout will appear in the virtual display and operator terminal.
9 Clicking on the button changes the setting from xbpi to SBI. First Use Load the Sartorius Win Scale application to test whether the program was installed correctly. It is not necessary to connect a device to the PC for this purpose. To access the program from the Windows platform, select Start and Programs. > In the list of programs, the application is shown with an icon and the name Sartorius Win Scale. The exact location in the menu tree depends on the path selected during installation (e.g. for the standard path: Sartorius AG Sartorius WinScale ) Click on this name with the mouse to start the application. > The program performs an integrity test. Once this has been successfully completed, the verifiable window appears as a virtual display and operator terminal in the upper left-hand corner of the screen. If a balance, scale or weighing platform is connected to the COM1 interface and properly set to operate in the xbpi communication mode (see the section entitled Connecting Devices to the PC ), a weight readout will appear in the virtual display and operator terminal. For more information, refer to the sections entitled Operation and Error Messages. (see the section Error Messages ). Exit the application by clicking on the on-screen button x or X on the keyboard. or by entering When prompted as to whether you are sure that you want to exit the application, click on the Yes button to quit or enter y or Y on the keyboard. > This exits the Sartorius Win Scale application, and the virtual display and operator terminal disappears from the screen (see also the section entitled Exiting Sartorius Win Scale in the Operation section).
10 Configuring the Communication Library If the error message ERR APP 13 ( Integrity test of the communication library (DLL) has failed ) appears in the virtual display and operator terminal, the path assigned to this library in the program must be updated. This error occurs because the user selected a different directory path during installation rather than the standard installation directory C:\Program Files\Sartorius AG\Sartorius Win Scale. You can correct this error as follows (see also the sections entitled Accessing the Service Menu and Defining the Communication Library in the Operation section, which provide detailed descriptions): Access the service menu: Click on the S button with the mouse or enter s or S on the keyboard. When the selection screen for the user status and the password appears, select System for the user status by clicking on that choice with the mouse and enter the accompanying password (the preset password after installation is WINSCALE ). Click on the OK button to confirm. In the Service menu selection screen, click on the Communication library button. In the Communication library selection screen, click on the... button located to the right of the current path for the communication library, which is shown as Communication library. In the Browse dialog, set the correct path for the communication library. The filename is srvind3_3_0_1.dll. If this setting is not correct, enter the correct access code in the field labeled Access code. Click on the Update button. The updated communication library is displayed in the Current configuration box. Confirm all changes by clicking on the OK button. The program then returns to the service menu. Click the OK button to exit the service menu. The communication library has now been set correctly. If no device is connected to the COM1 interface, the error message ERR DLL -6 will appear on the display. If a device is properly connected, the weight value appears in the virtual display and operator terminal. If other error messages arise, consult the Error Messages section. Changing the Device Name (XBPI Mode) During the installation of Sartorius Win Scale, the communication parameters for the connected device in XBPI mode are set to the following defaults: ID name: xbpi Baud rate: 9600 Protocol: srvind 3301 Parity: ka Address: 0 Handshake: RTS COM Port: 1 Data bits: 8 The protocol name srvind 3301 is linked to the associated communication library. It can be changed in the service menu under Communication library (see the section Defining the Communication Library in the Operation section). The ID name srvind 3301 is user-definable. If you would like to change this ID, proceed as follows (see also the sections entitled Accessing the Service Menu and Configuring Communication with the Device in the Operation section, where this setting is described in detail). Changing the Device Name (SBI Mode) During the installation of Sartorius Win Scale, the communication parameters for the connected device in SBI mode are set to the following defaults: ID name: sbi1 Baud rate: 9600 Protocol: srvsbi 3_6_0_6 Parity: ka Address: 0 Handshake: RTSXONXOFF COM Port: 1 Data bits: 8 The protocol name srvsbi 3_6_0_6 is linked to the associated communication library. It can be changed in the service menu under Communication library (see the section Defining the Communication Library in the Operation section). The ID name srvsbi 3_6_0_6 is user-definable. If you would like to change this ID, proceed as follows (see also the sections entitled Accessing the Service Menu and Configuring Communication with the Device in the Operation section, where this setting is described in detail).
11 Access the service menu: Click on the S button with the mouse or enter s or S on the keyboard. When the selection screen for the user status and the password appears, select System for the user status by clicking on that choice with the mouse and enter the accompanying password (the preset password after installation is WINSCALE ). Confirm by clicking on the OK button. In the Service menu selection screen, click on the Weighing platform button.
12 The Weighing platform selection screen appears. The configuration data for the weighing platform is displayed. If operating mode SBI is selected, additional settings can be configured via the SBI Setting button that subsequently becomes active. Enter the ID for the weighing platform. Select whether or not it is a verified device. The result of a calibration/ adjustment process can be printed out. With the Blockprint setting, the results are collected and printed out as a blockprint upon request. Consult the configurations according to the respective operating instructions for the weighing platform. Based on the respective operating instructions, select whether the combined tare/null setting can be selected. The buttons can be pressed separately, for example, on the film keypad of laboratory scales and balances. Confirm the entered data with OK. Settings for XBPI: See image Save the new ID in the program by clicking on the Update button and confirm the change by clicking on the OK button. The program then returns to the service menu. Click the OK button to exit the service menu. Click on the button or press n or N on the keyboard. A selection screen appears with the options SBI or XBPI. Click on your selection (e.g.: SBI or XBPI) and confirm the selection by clicking the OK button. Reselect the device (see also the section Selecting a Device in the Operation section). The new device ID now appears in the corresponding field of the virtual display and operator terminal. Select the new ID in this list by clicking on it.
13 Creating the Program Icon You can assign an icon to the Sartorius Win Scale application on the desktop of the PC as follows: Using the right mouse button, double-click on a free area of the PC desktop. A context menu is displayed. If you select New Shortcut, the Create Shortcut window appears on screen. Click on the Browse... button, set the directory path for the application WinScale.exe and confirm the entry by clicking on the Open button (or double click on WinScale.exe ). In the Command Line window, add the file name WinScale.ini after the application name WinScale.exe (you must leave a space before the word WinScale, i.e.:...\winscale.exe WinScale.ini). Confirm the new command line by clicking on the Next button. Assign a name to the shortcut, e.g. Win Scale and create the shortcut by clicking on the Finish button. > The icon for the new shortcut to the Sartorius Win Scale application appears on the PC desktop:. You can start the Sartorius Win Scale application by double clicking on this icon. Note: Once the program has been installed, both passwords (i.e. for the User and System user profiles) have the default setting WINSCALE. To protect the program from unauthorized changes to relevant system settings, change these passwords immediately after starting up the application and give each user profile a unique password. Record the passwords and store them in a safe place. If you lose the system password, you must reinstall the program. This completes all installation procedures for the Sartorius Win Scale application.
14 Connecting Devices to the PC Connecting the FB/FC series to the PC Scales of the FB/FC series are equipped with a built-in RS-232 data interface. The data output port (12-pin, round socket) is connected to the PC via cable YCC01-03ISM5. If the PC has a 9- pin connector rather than a 25-pin connector, you must also connect a adapter cable between the PC and the interface cable. Set the balance to operate in the xbpi communication mode. For more information, refer to the installation and operating instructions for the particular balance. Connecting the LA/LP series to the PC Scales of the LA/LP series are equipped with a built-in RS-232 data interface. The data output port (25-pin D-Sub female) is connected to the PC via cable If the PC has a 9-pin connector rather than a 25-pin connector, you must also connect a adapter cable between the PC and the interface cable. Set the balance to operate in the xbpi communication mode. For more information, refer to the installation and operating instructions for the particular balance. Connecting an isi Operating Terminal/FD Series to the PC The isi industrial terminal must be equipped with a serial RS-232 interface. Scales of the FD Series come standard with an RS-232 interface. The data output port (12-pin, round socket) is connected to the PC via cable YCC01-03ISM5. If the PC has a 9-pin connector rather than a 25-pin connector, you must also connect a adapter cable between the PC and the interface cable. Set the RS-232 interface ( COM interface) to operate in the XBPI-232 communication mode by accessing the system settings. For more information, refer to the installation and operating instructions for the isi industrial terminal or the FD series scale. Connecting an IS Weighing Platform to the PC IS weighing platforms are equipped at the factory with an RS-485 interface and, in general, do not have their own power supply. The weighing platforms are operated in conjunction with an isi industrial terminal, and power is supplied via the connecting cable. To directly connect a weighing platform (or several weighing platforms in an xbus network see the section entitled xbus Network for more information) to the PC, the weighing platform must be retrofitted with its own power supply. For this purpose, you will need an installation kit for an external power supply (for line voltage or mains supply of 220V: order no ) and a YAS03IS upgrade kit. To facilitate data exchange between the PC and the weighing platform, you must do one of the following: Use an RS-232/RS-485 interface converter, model , which is connected to the PC via cable The pin assignment for connecting a weighing platform is included in the installation and operating instructions for the interface converter. Install an RS-485 interface card in the PC. For example, the W&T PC card (order no. 62PC-Y014) is suitable for this purpose. Additional information on this can be found in the section xbus Network in the System Overview section.
15 Operation M Loading Sartorius Win Scale Requirements: M The connection between the device and the PC must be properly established (see also Connecting a Device to the PC ). The application must be installed (see also Installing Sartorius Win Scale ). Double-click on the Win Scale icon on the PC desktop. > This starts the application. An integrity test of the program file (.exe file ) is performed first. If this fails, a corresponding error message appears. The application is closed immediately. This implies that the program file has been manipulated. In this case, notify your Sartorius dealer, your nearest Sartorius Service Center or the International Technical Support Unit based in Goettingen, Germany, since you will not be able to run the program and it is, therefore, unusable. If one of the error messages shown on the left is displayed when you start the application, Sartorius Win Scale will also close immediately. In these cases, the application was not installed correctly or the file shown in the error message could not be found. To correct this error: Reinstall the application (see the section entitled Installing Sartorius Win Scale for more information). If the error continues to occur, contact your Sartorius dealer, your nearest Sartorius Service Center or the International Technical Support Unit based in Goettingen, Germany. If the self-test of the application is completed successfully, the verifiable window will appear on screen as the virtual display and operator terminal. The first time Sartorius Win Scale is run after reinstallation, the verifiable window is positioned in the upper left-hand corner of the screen. Otherwise, the window remains in the position defined by the user the last time the application was accessed (see also the section entitled Configuring the Display Position ). The virtual display and operator terminal of the Sartorius Win Scale application is shown in a fixed area on the screen where it cannot be overwritten. It cannot be hidden by any other application, and it returns to the foreground of the display after each change to a measured value. The figure on the left shows a practical example of the Sartorius Win Scale virtual display and operator terminal, which cannot be overwritten, together with another application (e.g. MS Excel).
16 If the application starts correctly but the communication library necessary for communication with the connected weighing instrument is missing, the error message ERR APP 13 is displayed. The communication library can be selected in the service menu (see the section entitled Accessing the Service Menu for more information). If the display window is empty, a weighing instrument was not selected. In this case, access the service menu, select the appropriate device and click on the on-screen button to confirm (see the section entitled Accessing the Service Menu for more information). If other error messages are displayed, consult the section entitled Error Messages. You can access on-line help by pressing the [F1] key Information in the Display 1 Zero-indicating device (zero indication) 2 Maximum capacity 3 Minimum capacity 4 Measured value (weight) 5 Verification scale interval e * 6 ID of the auxiliary indicating devices / Scale interval d * 7 Weight unit / stability symbol 8 Scale interval d 9 Net symbol 10 Caution triangle 11 Busy symbol 12 Exit application button 13 Gross value 14 Service menu button 15 Store measured value in alibi memory button 16 Tare value 17 Zero-setting button 18 Tare button 19 Device ID 20 Device address 21 PC serial interface number 22 Select device button 23 + or - sign * Not displayed on devices in a non-verifiable configuration. Note M When using the Sartorius Win Scale application in legal metrology, make certain that the metrology line (maximum capacity (2), minimum capacity (3), verification scale interval e (5) and, if present, scale interval d (8)) (depending on the design of the scale or balance, see the respective operating instructions) contains the correct data. For weighing platforms with a scale interval d not equal to the verification scale interval e, the additional display position d must be labeled (this does not apply to weighing platforms with e = 1 mg and d < 0.1 mg). This must be ensured before calibration of the scale. Set the required configuration according to the respective operating instructions for the weighing platform. The on-screen buttons are activated either by clicking on them with the mouse or by entering the letter or number indicated on the buttons using the PC keyboard. Both upper-case and lower-case letters can be entered. The same is true for the buttons in the selection windows and information windows that are displayed while the program is running. These buttons can be activated either by clicking on them with the mouse or by holding down the [ALT] key and pressing the underlined letter shown on the on-screen button on the keyboard (shortcut).
17 button (12) or enter X on the key- Exiting Sartorius Win Scale Using the mouse, click on the board. > The application prompts you to confirm whether you would like to quit the application. Click on the Yes button or enter Y on the keyboard. > The Sartorius Win Scale application closes, and the virtual display and operator terminal disappears from the screen. Selecting a Device Using the mouse, click on the button (22) or enter N on the keyboard. > A selection screen that allows you to choose the desired weighing instrument is displayed. Select a device from the list by clicking on it with the mouse or by pressing the cursor buttons and on the keyboard, and confirm the selection by clicking the OK button. > The Sartorius Win Scale application establishes the link to the selected weighing instrument. The device ID appears in the field (20) of the virtual display and operator terminal. If you would like to connect to a device not listed on the selection screen, access the service menu using the System user status and select the desired device from the list of available weighing instruments via the Weighing Platform menu (for more information, see the section entitled Configuring Connection with the Device under Accessing the Service Menu ). Taring a Device Use the mouse to click on the T button (19) or enter T on the keyboard. > The weighing instrument is tared, and the readout (4) is set to zero. The net symbol (9) appears in the virtual display and operator terminal and indicates that the measured value (4) displayed is a net value. The tare value appears in the tare value display (16), and the gross value is shown in the gross value display (13). If you attempt to tare the device with a gross value 0, the error message ERR 09 appears in the virtual display and operator terminal. The function is not executed. Zeroing a Device Using the mouse, click on the button (17) or enter 0 on the keyboard. > This zeroes the display the weight value measured (4), tare display (16) and gross value display (13) are all set to zero. If the weight (4) displayed was a net value before the display was zeroed, the net symbol (9) disappears. If you attempt to zero the device outside of its zeroing range, the error message ERR 08 appears in the virtual display and operator terminal. The function is not executed.
18 Writing the Weight to Alibi Memory M Note: You must configure the alibi memory before using it for the first time. This process is described under Accessing the Service Menu Configuring the Alibi Memory. Once the alibi memory has been configured, you must check it to make sure it functions properly (for more information, refer to the section entitled Function Test of the Alibi Memory under Accessing the Service Menu ). Click on the E button (15) with the mouse or enter E on the keyboard. > The weight is stored in the alibi memory. If you have not yet configured the alibi memory, a corresponding error message will be displayed. In this case, access the service menu using the System user status and configure the alibi memory using the Settings menu. See here the section Accessing the Service Menu Configuring the Alibi Memory. Minimize The display can be minimized on the screen with the M button on your keyboard; the display then disappears from the screen. Note: Only possible with the SBI setting! Accessing the Service Menu M Use the mouse to click on the S (14) button or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. 3 user profiles are available for selection: Normal status, in which no password is required. The user can access general system information, such as the following: Information about the currently selected device (selection of the device with the button (22)) Information about the software Sartorius Win Scale (e.g. software version and software ID for inspecting the software modules relevant to verification) User status (requires the user password). The user can also: Read the data stored in the alibi memory and export this data Read the data in the directory Archived Day Files and export this data Read the data in the directory Altered Day Files and export this data System status (requires the system password). The user has access to all settings in the service menu. Select the user status ( User or System ) by clicking on the desired status with the mouse or by using the cursor buttons - and on the keyboard, enter the associated password and confirm by clicking on the OK button. Note: The password is case-sensitive, so be aware of upper-case and lower-case letters when making your entry. If an incorrect password is entered, an error message will be displayed. You have the option of re-entering the password at this time. If you only wish to view general system information (normal status, which does not require a password), just click on the OK button without entering a password. In this case, the user status displayed is meaningless.
19 > The service menu appears on screen. This menu is divided into 5 sections ( Weighing Platform Communication, DDE Communication, Information, Verifiable Memory (alibi memory) and Display ). One or several on-screen buttons appear in each section. Note: Depending on the user status selected, the on-screen buttons that cannot be activated have different lettering to distinguish them from the others in accordance with the Microsoft Windows conventions. The figure on the left shows the service menu when the user status User is selected, and the figure below it shows the service menu for the System user status. To activate an on-screen button, either click on it with the mouse or enter [ALT] + the the keyboard key for the underlined letter shown on the button (shortcut). To exit the service menu: Click on the OK button or enter [ALT] + O on the keyboard. Changing the Service Menu Password Purpose: Limiting the right to read, set or change data relevant to security and verification in the service menu to very select groups of properly trained users. Use the mouse to click on the S (14) button or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. Select the user status for which you would like to change the password ( User or System ) by clicking on it with the mouse or by using the cursor buttons and on the keyboard, then click the button Change.
20 > An entry screen for changing the password appears. The password for a particular user status can only be changed if you know the currently valid password. First enter the old password, then enter the new password and confirm it by re-entering the new password. Confirm the password change by clicking on the OK button. Note: The password is case-sensitive, so be aware of upper-case and lower-case letters when making your entry. Reading Information from the Connected Device Purpose: Reading identifying data, such as the model name, serial number and metrological data, from the selected weighing instrument. If necessary, select a weighing instrument (see the section entitled Selecting a Device for more information). To access the service menu: Use the mouse to click on the S (14) button or enter S on the keyboard. A password is not required (see the section entitled Accessing the Service Menu ). > The service menu appears on screen. In the Information section, click on the Weighing platform button or enter [ALT] + F on the keyboard. > The screen Weighing Platform Information appears with the desired information. To exit the screen: Click on the OK button. You can also retrieve this information by selecting the configuration menu for the connected weighing instrument ( Weighing Platform Settings screen). To do so, access the service menu using the System user status, click on the Weighing platform button in the Communication weighing platform section of the menu, select the desired weighing instrument from the list, and click on the Test button. This procedure is described in the section entitled Configuring Communication with the Device.
21 Reading Program Information M Purpose: Reading information about the program, such as software version, software ID, DLL version and PTB test certificate number. Inspection of the software modules relevant to verification When you access the program information, the version ID of each function relevant to verification in the Sartorius Win Scale application is checked and a total checksum is determined from this data via checksum calculation (CRC: Cyclic Redundancy Check). This allows the user to compare the calculated actual checksum to the displayed target checksum. To access the service menu: Use the mouse to click on the button (14) or enter S on the keyboard. A password is not required (see the section entitled Accessing the Service Menu ). > The service menu appears on screen. Enter the password In the Information section, click on the Software button or [ALT] + S on the keyboard. > The screen Software Information appears with the desired information. M The actual and target checksums are indicated for the software ID. The integrity of the software modules relevant to verification is only ensured when these two values are identical. To exit the screen: Click on the OK button.
22 Defining the Communication Library Purpose: Selecting the verifiable communication library necessary for communication with the connected weighing instrument (or instruments). In order to work with Sartorius Win Scale, at least one verifiable communication library must exist. This library functions as a driver for the connected weighing instrument and must, therefore, correspond to the device with which communication is to be established. The communication library (e.g. srvind3301.dll ) is a separate file, not an integral component of the Sartorius Win Scale software module (see also Software System in the System Overview section), but it is included in the delivery. The supplied file enables communication with the most important Sartorius instruments via the xbpi protocol; please consult with your Sartorius dealer, the Sartorius Service Center or the International Technical Support Unit based in Goettingen, Germany, for special applications. For licensing reasons, an access code for each communication library is delivered separately; this code functions as a key to enable use of the corresponding library. The following steps are necessary to define the communication library: Selection of a protocol ID Assignment of the verifiable communication library linked to this protocol ID (entry of the file name with directory path) Entry of the corresponding access code Updating of the library directory To access the service menu: Use the mouse to click on the S (14) button or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. Select the System user status by clicking on it. Enter the system password and confirm by clicking on the OK button. > The service menu appears on screen. Click on the Communication library button in the Communication weighing platform section of the menu or enter [ALT] + L on the keyboard.
23 > The Communication Library Settings screen is displayed. Enter the desired protocol ID (e.g. srvind3301.dll ). The protocol ID is user-definable and enables the application to use the communication library linked to this name. Enter the file name of the communication library with the complete directory path. To do this, click on... button, select the required file in the Browse screen, (e.g. srvind3301.dll ) and confirm by clicking on the Open button. The file then appears in the Communication library window. Enter the access code. The access code is delivered separately with the product. Update the communication library: Click on the Update button or enter [ALT] + U on the keyboard. If a communication library already exists, you will be asked to confirm that you wish to update the library. > The selected communication library is displayed in the Current configuration box. If one or more libraries are already listed here, the new library will be placed at the top of the list. To exit the screen and confirm changes to the settings: Click on the OK button. To cancel without changing the settings: Click on the Cancel button. Changing the current communication library (if several communication libraries exist): Select the new communication library in the Current configuration box by clicking on it with the mouse. Update the communication library by clicking on the Update button. Answer the prompt with Yes. > The currently valid library (protocol ID, file name with directory path and access code) is redisplayed in the Assign name section. The list of available libraries in the Current Configuration box is reordered so that the current configuration is listed first.
24 Removing a Communication Library (If multiple communication libraries exist): Select the communication library to be removed in the Current Configuration box by clicking on it with the mouse. Click on the Remove button or enter [ALT] + R on the keyboard to remove the communication library. Answer the prompt with Yes. > The currently valid library (protocol ID, file name with directory path and access code) is redisplayed in the Assign name section. The deleted communication library is removed from the list of all existing communication libraries in the Current configuration box. If no communication libraries exist at this point, the error message ERR APP 01 will appear in the virtual display and operator terminal. For the program to function properly, a communication library must first be defined again. Configuring Communication with the Device Purpose: Setting all parameters necessary for communication with a weighing instrument, such as the device ID, communication protocol, device address, COM port on the PC and other interface parameters. Proper communication with the weighing instrument can be tested. To establish communication between the Sartorius Win Scale application and an external device, the following settings are required for the application and the external device: The Sartorius Win Scale application must use the verifiable communication library that corresponds to the external device. A name (protocol ID) that establishes the necessary connection is assigned to the library in the selection screen Communication Library Settings (see also the section Defining the Communication Library ). The device address, connection of the serial interface on the PC (COM port) and the interface parameters must be adapted for communication with the external device. The external device must be set to the transmission protocol configured in the Sartorius Win Scale application (see the section entitled Connecting Devices to the PC ). The external device must be correctly configured in order to select it for communication with Sartorius Win Scale via the button (22) in the virtual display and operator terminal. You can configure these settings in the Weighing Platform Settings screen, which is explained below.
25 To access the service menu: Use the mouse to click on the S (14) button or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. Select the System user status by clicking on it. Enter the system password and confirm by clicking on the OK button. > The service menu appears on screen. Click on the Weighing platform button in the Communication weighing platform section of the menu or enter [ALT] + W on the keyboard. > The Weighing Platform Settings screen is displayed (see figure on the next page). Enter the device ID in the ID field. The device ID is user-definable. Once you have selected a device by clicking on the button (22) in the virtual display and operator terminal, this name appears in the field above the on-screen button (marked as (19) in the figure in the section entitled Information in the Display ). Note: Each new entry in the ID name field opens a new set of parameters (= communication path). Several sets of parameters can exist for the same device. However, this makes little sense in actual use because all parameters (communication protocol and all interface parameters) must be identical for the same device. Select the communication protocol for data transfer between the PC and the device in the Protocol field. Clicking on the button opens a selection window in which you can click on the desired protocol. This protocol name establishes the link to the verifiable communication library that supports communication with the external device (see the section entitled Defining the Communication Library ). Enter the device address in the Address field. The device address is a number between 1 and 31 (for an individual device, the address 0 is also acceptable) that is assigned only to that device. With the help of configuration programs, such as YAD01IS, a particular address can be assigned to a device. If you leave the Address field blank, the address will be interpreted as 0. Enter the number of the serial interface on the PC (COM port) used for data exchange between the PC and the device in the COM Port field. If, for example, the device is connected to the COM1 interface, enter 1 in the COM Port field. Entry of other interface parameters, such as baud rate, parity, handshake, data bits, and stop bits is optional. If these fields are left blank, ka (no entry) appears in each field. If data is entered into these fields, however, it must agree with the configuration of the device interface. Otherwise, data cannot be transmitted properly between the PC and the external device. To transfer the new communication parameters to the PC: Click on the Update button with the mouse or enter [ALT] + U on the keyboard. If you have changed an existing configuration (the entry in the ID name field was not changed), you must answer a prompt to confirm the updates. > The new set of parameters appears in the Current configuration list. This list is arranged alphabetically according to the entries in the ID name field. To exit the screen and confirm changes to the settings: Click on the OK button. To cancel without changing the settings: Click on the Cancel button.
26 Changing a communication parameter set (if multiple parameter sets exist): Select the set of parameters to be changed in the Current configuration box by clicking on it. The communication parameters that have been valid up to now are displayed in the corresponding fields and can be changed. If you change the data in the ID name field in the process, this will be interpreted by the program as a new set of parameters (see the note on the previous page). Update the parameter set by clicking on the Update button. Answer the prompt with Yes. > The changed parameter set is displayed again in the Current configuration list. Removing a set of communication parameters (if several sets of parameters exist): Select the set of parameters to be removed in the Current configuration box by clicking on it. Click on the Remove button or enter [ALT] + R on the keyboard to remove the parameter set. Answer the prompt with Yes. > The deleted set of parameters is removed from the list of all existing sets of communication parameters under Current configuration. If no sets of parameters exist at this point, the error message ERR APP 01 will appear in the virtual display and operator terminal. This error message disappears as soon as proper communication to an external device is re-established. Testing the connection to a device: Select the connection to be tested by clicking on it in the Current configuration box. > The corresponding communication parameters are displayed in the applicable fields. Click on the Test button or enter [ALT] + T on the keyboard. > The Weighing Platform Information screen is displayed. If a connection to a particular device has been established, the applicable identifying data for that device, such as the model name, serial number and metrological data, appear on this screen. If the connection could not be established (e.g. due to faulty communication parameter settings),??? appears in the Model and Serial Number fields. To exit the screen: Click on the OK button. Defining DDE Connection Parameters Purpose: Selecting the communication parameters necessary for Dynamic Data Exchange (DDE) between a DDE server program and a DDE client program (see section Communication with External Programs (DDE Clients) ). As a DDE server program, the Sartorius Win Scale application can receive and execute queries and requests from an external program (DDE client program) (e.g. requests to read a net value at stability and provide this data to the client program or to tare the weighing instrument). The DDE server program makes information available for retrieval without actively seeking communication with other software processes. DDE client programs can establish DDE communication with a server application using the DDE communication parameters made available by the server. These parameters are defined by the three terms Service, Topic and Item : The Service parameter is the name that the server program recognizes when a DDE client program attempts to establish a DDE connection. The Topic parameter describes the type of data source. The Item parameter identifies the actual data requested by the DDE client. The entries for the DDE communication parameters are userdefinable in the DDE server. This section describes only the entry of connection parameters. For further information on the DDE interface and application examples, refer to the section entitled Communication with External Programs (DDE Clients). To access the service menu: Use the mouse to click on the S (14) button or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. Select the System user status by clicking on it. Enter the system password and confirm by clicking on the OK button. > The service menu appears on screen. Click on the Connection parameters button in the Communication DDE section or enter [ALT] + B on the keyboard.
27 > The Connection Parameters Settings screen is displayed. Enter the Service, Topic and Item parameters in the respective fields. These must also be defined in the client application to be served by Sartorius Win Scale. To confirm the settings: Click on the OK button. If you do not want to save the settings: Click on the Cancel button. Configuring the Alibi Memory M Purpose: Defining a directory path and storage period for data in the alibi memory. Data stored in the alibi memory is protected against manipulation by a security mechanism (CRC: Cyclic Redundancy Check). The security mechanism is designed to satisfy the special statutory requirements that apply to the archiving of verifiable data. Data stored in this memory is considered verifiable for the entire period for which it is stored in the memory. The first time you run Sartorius Win Scale after installing the program, you must define a directory path under which files will be stored and protected in the alibi memory. The directory ( ESP is generally selected as the name for the alibi memory) must be created on the local hard disk of the PC being used. It may not be located on a diskette or a network drive. Each time data is stored in the alibi memory, the program checks whether sufficient storage capacity exists under the defined directory path. As a ring buffer, the alibi memory is set up with a defined storage capacity allocated in days. Once the defined time period expires, the oldest day file is moved from the alibi memory to the archive directory WINSCALE-ESP-DEL- FILES ( archived day files ). The files moved in this way are protected against manipulation with checksums in the same manner as the files in the ring buffer. If an attempt is made to alter a file stored in the alibi memory or in the archive directory, the file is identified by the program as a manipulated file and removed from the alibi memory or archive directory. These files are stored in the directory WINSCALE-ESP-CRC-ERROR. To access the service menu: Use the mouse to click on the S (14) button or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. Select the System user status by clicking on it. Enter the system password and confirm by clicking on the OK button. > The service menu appears on screen. Click on the Settings button in the Verifiable Memory (alibi memory) section of the menu or enter [ALT] + E on the keyboard.
28 > The Settings screen is displayed. Enter the time period for data to be stored in the ring buffer (1 to 30,000 days) in the Time frame field. The time period is dependent on the legal requirements of your country. Enter the path for the Verifiable Memory directory in the Memory path field. The default setting for the path is C:\ESP. Note: The alibi memory (ring buffer) must be located on the local hard disk. If this is not the case, a corresponding error message will be displayed, and you will be given the option of correcting the error (see the figure below on the left). If you do not make a correction, the default C:\ESP will be used as the path. Note: The paths for archived data ( WINSCALE-ESP-DEL-FILES ) and altered data ( WINSCALE-ESP-CRC-ERROR ) are created as subdirectories of the alibi memory (ESP). To save the configuration: Click on the OK button or enter [ALT] + O on the keyboard. To cancel the configuration: Click on the Cancel button or enter [ALT] + C on the keyboard. Function Test of the Alibi Memory M Purpose: Testing the alibi memory to ensure error-free function. Immediately after it has been configured, the alibi memory must be tested to ensure error-free function and compliance with the applicable legal requirements. This is done by performing several weighing operations, storing the resulting data in the alibi memory, accessing the stored data and checking to ensure that it was stored correctly (for more information, see also the sections entitled Writing the Weight to Alibi Memory and Reading the Alibi Memory and Exporting Data ). Perform a weighing operation and store the weight in the alibi memory. To do this, click on the E button (15) with the mouse or enter E on the keyboard (see Writing the Weight to the Alibi Memory ). Repeat the procedure several times with various weights. To access the service menu and read the data in the Alibi memory (see also Accessing the Service Menu Reading the Alibi Memory and Exporting Data ): Click on the S button (14) with the mouse or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. Click on User to select this status. You can also select the System user status instead, which carries higher privileges. Enter the password for the particular user status selected, and click on the OK button to confirm.
29 > The service menu appears on screen. Select the alibi memory by clicking on the Verifiable memory button in the Verifiable memory section of the menu or by entering [ALT] + V on the keyboard. > The Display screen appears. To select a date: Clicking on the button opens a selection window in which you can click on the desired date. Note: If you do not first choose a date when selecting a data record, you will be alerted by a corresponding error message. Select the desired data record by clicking on it with the mouse in the Data record box. As an alternative method, enter the verifiable ID for this data record. The ID appears in every data record as the first group of numbers before the date. > The selected data record, which appears in compressed format in the Data record box, is shown in detail in the Contents section of the window. Check the displayed data record to determine if it has a verifiable ID and a checksum. These must be displayed in their respectively designated fields. The checksum must be labeled with OK:. The actual checksum can be seen at the bottom of the data record display after the remarks, and the whole checksum can be viewed by sliding the scroll bar to the right side of the window (see figure).
30 Check the data records of all test weighing operations in the same manner. All data records must have a unique verifiable ID and a correct checksum. If errors occur, notify your Sartorius dealer, the nearest Sartorius Service Center or the International Technical Support Unit based in Goettingen, Germany. In this case, the Sartorius Win Scale application may not be used in legal metrology. Reading the Alibi Memory and Exporting Data Purpose: Reading out data stored in the 3 memories for weights. The data records can be selected and exported for further processing. In the process, a copy of the current day file is created in ASCII format. The Sartorius Win Scale application features 3 storage areas for saving weights. They provide a graduated hierarchy of protection to safeguard the integrity of the data (see also the section entitled Configuring the Alibi Memory ): Verifiable Memory directory: Ring buffer with defined capacity (1 to 30,000 days) Archived Day Files directory: Contains all data records that were moved out of the Verifiable Memory directory due to their age in relation to the current date Altered Day Files directory: Contains all files that were originally in either the Verifiable Memory directory or the Archived Day Files directory but were moved out of these directories because the content of the files was changed (e.g. the files were edited). You can view data stored in these directories by selecting the respective memory area in the service menu. For further processing by other applications, such as EXCEL, they must first be exported. In the process, the data is presented in ASCII format and can therefore be edited. To access the service menu: Use the mouse to click on the S (14) button or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. Click on User to select this status. You can also select the System user status instead, which carries higher privileges. Enter the password for the particular user status selected, and click on the OK button to confirm. > The service menu appears on screen. Select the desired memory by clicking on the respective button in the Verifiable memory section of the menu: Ring buffer (alibi memory): Click on the Verifiable memory button or enter [ALT] + I on the keyboard. Archived day files: Click on the Archived Day Files button or enter [ALT] + A on the keyboard. Altered day files: Click on the Altered Day Files button or enter [ALT] + T on the keyboard.
31 > The Display screen appears. To select a date: Clicking on the button opens a selection window in which you can click on the desired date. Note: If you do not first choose a date when selecting a data record, you will be alerted by a corresponding error message (see figure on the left below). > In the Data record box, all data records for the date selected are listed. Note: Only data records that were generated within the storage period are located in the Verifiable Memory directory. All other data records are located in the archive directory. This directory can be reached from the service menu by clicking on the Archived Day Files button. All data records that have been changed are stored in the Altered Day Files directory, regardless of the date they were created. This directory can be reached from the service menu by clicking on the Altered Day Files button. Therefore, if you cannot find a certain data record in a particular memory, or if this memory is empty, you should search the other two memories as well. Select the desired data record by clicking on it with the mouse in the Data record box. As an alternative method, enter the verifiable ID for this data record. The ID appears in every data record as the first group of numbers before the date. > The selected data record, which appears in compressed format in the Data record box, is shown in detail in the Contents section of the window. Exporting data for further processing: Click on the Export button or enter [ALT] + E on the keyboard. > The Export screen is displayed. Select the storage medium (hard disk, diskette,...), target directory and file name. The name of the data record with the file extension asc is suggested as the file name, since the exported file is in ASCII format. Next, click on the Save button to save the data record under the specified path. The program then returns to the Display screen. To cancel data export: Click on the Cancel button, and the program returns to the Display screen. To exit the screen: Click on the OK button or enter [ALT] + O on the keyboard. The program returns to the Service menu screen.
32 Configuring the Display Position Purpose: Positioning the virtual display and operator terminal on the screen. The Sartorius Win Scale application shows a virtual display and operator terminal on screen. This display cannot be overwritten by another application and moves into the foreground following every weight change registered by the connected weighing instrument. The position of this display on screen is user-definable so that applications running in parallel can be conveniently accessed as well. You must have system privileges to adjust the position of the display. To access the service menu: Use the mouse to click on the S (14) button or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. Select the System user status by clicking on it. Enter the system password and confirm by clicking on the OK button. > The service menu appears on screen. Click on the Position button in the Display section of the menu or enter [ALT] + P on the keyboard. > The Position Settings screen is displayed. In this screen, the current monitor resolution in pixels and the current display position are shown. Changing the position of the display: If you would like the display to appear in one of the four corners of the screen, click on the desired area with the mouse. To position the display in another location on the screen: Click on the display position User-defined and indicate the exact location by entering the coordinates of the upper left corner of the display. The current monitor resolution in pixels defines the maximum possible values for the X and Y coordinates. To save the new setting: Click on the OK button or enter [ALT] + O on the keyboard. If you do not wish to change the setting: Click on the Cancel button or enter [ALT] + C on the keyboard. > The program returns to the Service menu screen.
33 Configuring the Display Colors Purpose: Selecting a color profile for showing the virtual display and operator terminal on the screen. You can select from 2 different color profiles for showing the virtual display and operator terminal: 3-D color display Display in gray tones The color profile is normally selected by the user. The display in gray tones is especially useful for monochrome monitors. You must have system privileges to adjust the colors of the display. To access the service menu: Use the mouse to click on the S (14) button or enter S on the keyboard. > A selection screen appears for entering the user status and corresponding password. Select the System user status by clicking on it. Enter the system password and confirm by clicking on the OK button. > The service menu appears on screen. Click on the Colors button in the Display section of the menu or enter [ALT] + C on the keyboard. > The Colors Settings screen is displayed. The currently configured color profile is indicated in this screen: Profile 1: 3-D color display Profile 2: Display in gray tones To change the color profile: Click on the desired color profile or enter the respective number on the keyboard, then click on the OK button. To exit without changing the color profile: Click on the Cancel button. > The program returns to the Service menu screen.
34 Communication with External Programs (DDE Clients) DDE Interface The Sartorius Win Scale application can be used as an interface between a user, one or more software processes running in parallel and one or more devices used to exchange data with the software process or processes. This type of data exchange is referred to as Dynamic Data Exchange (DDE). In this case, Sartorius Win Scale functions as a DDE server, and the individual user process is referred to as the DDE client. The Sartorius Win Scale DDE server makes information available for retrieval without actively seeking communication with the user (DDE client) processes. DDE client programs can establish the DDE connection to the server application program using the DDE communication parameters made available by the server. These communication parameters are usually defined as Service, Topic and Item. The Service parameter contains an ID that the server recognizes when a DDE client attempts to establish DDE communication. The Topic parameter describes the type of data source. The Item parameter contains a variable identifying the actual data requested by the DDE client. DDE communication parameters in the Sartorius Win Scale DDE server are user-definable (see the section entitled Defining DDE Connection Parameters in the Operation section). The following default settings are valid at the start of the application: Service: = WinScale Topic: = AktWaage Item: = AktDaten read out automatically and, therefore, the DDE client s request to be answered by the DDE server. An alternative solution would be the explicit reading out of the Item variable with the DDE command Request. Creating the Data Request to the DDE Server (Request String) The request string consists of the key word sarto_cmd(...). The ID for the command to be executed is placed in the parentheses. If the command contains multiple parameters, these are separated by semicolons (e.g. sarto_cmd(p1;p2;p3) with the parameters P1, P2, P3). Example: Request for the current net weight with a stability check. The weight should be stored in the verifiable memory with the label Test_weighing. Required request string: sarto_cmd(w;esp;test_weighing) (see also the table DDE Functions ) The parameters of the command have the following meanings: Parameter 1: Command ID: W Request weight Parameter 2: Command ID: ESP Transfer weight to alibi memory Parameter 3: Transfer variable: Test_weighing Transferred weight labeled in the alibi memory with the comment Test_weighing Data Exchange The user program (DDE client) and the Sartorius Win Scale DDE server communicate by exchanging strings. First the DDE client sends a request string to the DDE server. The server then extracts the command parameter(s) from this string, executes the command and creates a response string for the DDE client. The request string is case-insensitive, so upper-case and lower-case letters have no bearing on the correct interpretation of the command. The following application example illustrates the fundamental concept of data exchange between the user program and the DDE server. The details of the DDE implementation depend on the development environment used to create the implementation. Task: The DDE client requests the current weight. The data exchange between the DDE client and the DDE server includes the following steps: The connection is established by the DDE client in accordance with the DDE communication parameters initialized in the DDE server (e.g.: Service = WinScale, Topic = AktWaage, Item = AktDaten). The DDE client requests the current weight with the DDE command Execute. For this purpose, the string sarto_cmd(w) is sent to the DDE server (see also the section entitled DDE Functions ). The Sartorius Win Scale DDE server decodes the DDE client request, from which it creates an internal command according to the internal SARTOCOM command set. The standardized SARTOCOM software interface can communicate with the individual device-specific communication libraries. The command is executed, and the current weight is sent back to the application. The application places this response in the Item variable via the DDE software interface so it can be read by the DDE client. Depending on how the DDE communication has been implemented in the DDE client, a change to the Item variable triggers an event, for example, that causes the variable to be Implementation Principles for User Software The software described in the following sections (application examples 1 and 2) was developed using the development system Microsoft Visual Basic. Menu structures and screen layouts were designed according to the standards of Microsoft Windows 95 and Windows NT and were primarily designed in the interests of user-friendliness. Windows 95, Windows 98, Windows 2000, Windows NT, Windows XP, MS Office, MS Excel and MS Word are registered trademarks of Microsoft Inc.
35 DDE Application Example 1 In the application example illustrated here, the following functions can be carried out: Establishment of the connection to Sartorius Win Scale Requesting of the net weight without stability check with device status information Note: The following source code is provided as an example for the fundamental structure of a DDE application. Sartorius assumes no liability for the error-free performance of the program. Form: Button: Button: Label: HerstellenVerbBtn AnfordernWertBtn Anzeige Source code: Private Sub HerstellenVerBtn_Click() Cancel DDE link Anzeige.LinkMode = 0 Set DDE connection parameters Anzeige.LinkTopic = WinScale AktWaage Establish DDE link Anzeige.LinkMode = 1 End Sub Private Sub AnfordernWertBtn_Click() Set DDE timeout Anzeige.LinkTimeout = Request and display weight data Anzeige.LinkExecute sarto_cmd(ws) End Sub After you run the program, a window with two on-screen buttons and a status line is displayed (see figure above on the left). Clicking on the Establish Link to Win Scale button creates a communication path from the DDE client to the Sartorius Win Scale DDE server. The DDE server is now ready to receive commands from the DDE client. The text in the status line changes from No link to Sartorius Win Scale - OK. If you click on the Request weight button, the command Request current net weight without stability check with device status information is sent to the DDE server. As soon as this data has been made available by the server, the DDE client retrieves it and displays it in the status line.
36 DDE Application Example 2 In the application example illustrated here, the following functions can be carried out: Establishment of the connection to Sartorius Win Scale A weight is automatically requested and saved in an Excel spreadsheet Termination of the automatic request for weights Showing and hiding Win Scale Reading device data (model name, serial number, maximum capacity, minimum capacity and tare memory). All data is saved in an Excel spreadsheet. Note: This program requires Excel - version ³ V7.0 (Office2000 ). Excel and Office2000 are registered trademarks of Microsoft Inc. Note: The following source code is provided as an example for the fundamental structure of a DDE application. Sartorius assumes no liability for the error-free performance of the program. Option Explicit Dim Anzahl$ Dim WithEvents ExcelObjekt As Excel.Application Attribute ExcelObjekt.VB_VarHelpID = -1 Dim WithEvents ExcelTabelle As Excel.Workbook Attribute ExcelTabelle.VB_VarHelpID = -1 Dim co As Object Object variable for diagram Dim Z Task ID of Win Scale
37 Private Sub AnfordernWertBtn_Click() Set DDE timeout Anzeige.LinkTimeout = 650 Delete any old values in list Lst_werte.Clear Show Excel ExcelObjekt.Visible = True ExcelObjekt.WindowState = xlnormal Timer reads out scale in cycles Timer1.Enabled = True Disable button AnfordernWertBtn.Enabled = False Enable cancel button Btn_stop.Enabled = True Prevent changes to number txt_anzahl.enabled = False Anzahl = txt_anzahl Prevent changes to interval Txt_Interval.Enabled = False End Sub Private Sub btn_hide_click() Hide Win Scale Lbl(4).LinkExecute sarto_cmd(sleep) Lbl(4).LinkRequest End Sub Private Sub btn_show_click() Unhide Win Scale Lbl(4).LinkExecute sarto_cmd(awake) Lbl(4).LinkRequest End Sub Private Sub Btn_stop_Click() Cancel reading out of values Timer1.Enabled = False Btn_stop.Enabled = False AnfordernWertBtn.Enabled = True txt_anzahl = Anzahl txt_anzahl.enabled = True Txt_Interval.Enabled = True End Sub Private Sub Cmd_Click(Index As Integer) Dim s$ Select Case Index Case 0 Model name Lbl(Index).LinkExecute sarto_cmd(si;m) Lbl(Index).LinkRequest Extract the desired information from the received string s$ = Lbl(Index).Caption s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) Lbl(Index).Caption = s Case 1 Serial number Lbl(Index).LinkExecute sarto_cmd(si;n) Lbl(Index).LinkRequest Extract the desired information from the received string s$ = Lbl(Index).Caption s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) Lbl(Index).Caption = s Case 2 Maximum capacity Lbl(Index).LinkExecute sarto_cmd(bi;h;r) Lbl(Index).LinkRequest Extract the desired information from the received string s$ = Lbl(Index).Caption s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) Lbl(Index).Caption = Left$(s, InStr(1, s, ; ) - 1) + Right(s, Len(s) - InStr(1, s, ;, 1)) Case 3 Minimum capacity Lbl(Index).LinkExecute sarto_cmd(bi;l) Lbl(Index).LinkRequest Extract the desired information from the received string s$ = Lbl(Index).Caption s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) Lbl(Index).Caption = Left$(s, InStr(1, s, ; ) - 1) + Right(s, Len(s) - InStr(1, s, ;, 1))
38 Case 4 Tare Lbl(Index).LinkExecute sarto_cmd(tt;s) Lbl(Index).LinkRequest End Select End Sub Private Sub Command1_Click() End the program Unload frmddeclient End Sub Private Sub Form_Load() Z = 0 On Error GoTo NotFound Z = Shell(App.Path + \binewt\winscale winscale.ini ) Start Win Scale. On Error Resume Next Create Excel as object Set ExcelObjekt = GetObject(, Excel.Application ) If ExcelObjekt Is Nothing Then Set ExcelObjekt = CreateObject( Excel.Application ) If ExcelObjekt Is Nothing Then MsgBox Cannot load Excel! Program terminating!, vbcritical End Else ExcelObjekt.Visible = True Open Excel spreadsheet Set ExcelTabelle = ExcelObjekt.Workbooks.Add (xlwbatworksheet) End If End If On Error GoTo 0 Application on screen Show AppActivate frmddeclient.caption Exit Sub NotFound: MsgBox The WINSCALE application must be stored in the + App.Path + \binewt directory! + vbcrlf + Cannot run program!, vbcritical End End Sub Private Sub Form_Unload(Cancel As Integer) On Error Resume Next If Not ExcelObjekt Is Nothing Then Exit Excel ExcelObjekt.Worksheets( Sheet1 ).Activate ExcelObjekt.Quit End If If Z <> 0 Then Lbl(4).Caption = Lbl(4).LinkMode = vblinknone Lbl(4).LinkTimeout = 6 Lbl(4).LinkTopic = WinScale AktWaage Lbl(4).LinkItem = AktDaten Lbl(4).LinkMode = vbautomatic Exit Win Scale Lbl(4).LinkExecute sarto_cmd(quit) End If End Sub Private Sub HerstellenVerbBtn_Click() Dim i% Cancel DDE link Anzeige.LinkMode = vblinknone Set DDE connection parameters Anzeige.LinkTopic = WinScale AktWaage Anzeige.LinkItem = AktDaten Establish DDE link Anzeige.LinkMode = vblinkautomatic Label for information For i = 0 To Lbl.Count - 1 Lbl(i).Caption = Lbl(i).LinkMode = vblinknone Lbl(i).LinkTimeout = 650 Lbl(i).LinkTopic = WinScale AktWaage Lbl(i).LinkItem = AktDaten Lbl(i).LinkMode = vblinkmanual Next i AnfordernWertBtn.Enabled = True tn_hide.enabled = True btn_show.enabled = True Frame1.Enabled = True End Sub
39 Private Sub Timer1_Timer() Dim s$, Cmd$, still$, Wert$, Einheit$, Stamp$ If Val(txt_Anzahl) > 0 Then On Error GoTo noconnection Anzeige.LinkExecute sarto_cmd(ws;) txt_anzahl = Trim(Str(Val(txt_Anzahl) - 1)) Extract the desired information from the received string s$ = Anzeige.Caption s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) Cmd = Left$(s, InStr(1, s, ;, 1) - 1) s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) still = Left$(s, InStr(1, s, ;, 1) - 1) s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) Wert = Left$(s, InStr(1, s, ;, 1) - 1) s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) Einheit = Left$(s, InStr(1, s, ;, 1) - 1) s$ = Right(s, Len(s) - InStr(1, s, ;, 1)) Stamp = s If ExcelObjekt Is Nothing Then Else Write value to Excel spreadsheet ExcelObjekt.Worksheets ( Sheet1 ).Activate ExcelObjekt.Worksheets ( Sheet1 ).Range ( A + Trim(Str(Val(Anzahl) - Val(txt_Anzahl)) )).Activate ExcelObjekt.ActiveCell.Value = Val(Wert) End If Insert value with stability attribute into list Lst_werte.AddItem still + + Wert + + Einheit Else Insert diagram ExcelObjekt.Worksheets ( Sheet1 ).Range( A1:A + Trim(Anzahl) ).Activate Does a diagram already exist? If co Is Nothing Then No, create new diagram Set co = Sheets( Sheet1 ).ChartObjects.Add(70, 20, 320, 300) End If Show diagram co.chart.chartwizard Source:=ExcelObjekt.Worksheets ( Sheet1 ).Range( A1:A + Trim(Anzahl)), Gallery:=xlLine Should values be read out in cycles? If Chk_cycle.Value = 0 Then No, operation has ended Timer1.Enabled = False Btn_stop.Enabled = False AnfordernWertBtn.Enabled = True txt_anzahl = Anzahl txt_anzahl.enabled = True Txt_Interval.Enabled = True Else Yes, start operation at the beginning txt_anzahl = Anzahl Lst_werte.Clear End If End If DoEvents Exit Sub noconnection: Timer1.Enabled = False Btn_stop.Enabled = False AnfordernWertBtn.Enabled = True txt_anzahl = Anzahl txt_anzahl.enabled = True Txt_Interval.Enabled = True MsgBox No link, vbcritical End Sub Private Sub Txt_Interval_LostFocus() If Val(Txt_Interval) <= 99 Then MsgBox Value too small (<=99)!, vbcritical Txt_Interval = 100 Else Timer1.Interval = Val(Txt_Interval) End If End Sub
40 DDE Functions Description Command Response Access in verified (const char*outdata) (char*indata) mode permitted 1 Request current net weight without stability check sarto_cmd (W) String Yes Request current net weight with stability check. Save the weight as verifiable data with a comment sarto_cmd (W;ESP;ABC) String Yes Request current net weight without stability check with device status information sarto_cmd (WS) String Yes Tare or zero (combined tare) without stability check (immediately) sarto_cmd (T;D) String No Tare or zero (combined tare) after stability reached; x = Timeout [sec] x >0: Check comes back if command successful or timeout expired x <0: Check comes back immediately sarto_cmd (T;S;x) String Yes 2 Read out tare value (total of zero and tare registers) sarto_cmd (T;R) String Yes Functions like sarto_cmd (T...), but always tare, no zeroing. During reading, only the tare register is read (only for xbpi driver) sarto_cmd (TT;...) String Yes 2 Functions like sarto_cmd (TT...), but the application tare register 1 is used (only for xbpi driver) sarto_cmd (T1;...) String Yes 2 Functions like sarto_cmd (TT...), but the application tare register 2 is used (only for xbpi driver) sarto_cmd (T2;...) String Yes 2 The display is zeroed and all tare registers are cleared. This is only possible in the range of ±2% from the maximum of the zero point (for xbpi driver only) sarto_cmd (TZ;...) String Yes Preset tare (application tare 1) (subtract x weight units from tare) sarto_cmd (TP;W;x) String Yes 2 Read out contents of the preset tare memory sarto_cmd (TP;R) String Yes Alternative form for the functions sarto_cmd (TP...) sarto_cmd (TP1;..) String Yes Functions like sarto_cmd (TP1;..), but for application tare 2 sarto_cmd (TP2;..) String Yes Lock device keys sarto_cmd (KB;L) String Yes Release device keys sarto_cmd (KB;F) String Yes Prompt to press device key sarto_cmd (KB;G) String Yes Activate one of the 3 weight unit levels (n = 1, 2, 3) sarto_cmd (US;W;n) String Yes Ask which weight unit level is active sarto_cmd (US;R) String Yes Start calibration/adjustment function that is active in the menu sarto_cmd (BC) String No Continue calibration/adjustment (save new calibration weight in device) (only for xbpi driver) sarto_cmd (BC;J) String No Perform external calibration/adjustment with default weight (only for xbpi driver) sarto_cmd (BC;D) String No Perform external calibration/adjustment with automatic weight detection (only for xbpi driver) sarto_cmd (BC;A) String No Perform external calibration/adjustment with user defined (user) weight (only for xbpi driver) sarto_cmd (BC;U) String No Adjust linearity with default linearity points (only for xbpi driver) sarto_cmd (BC;L) String No Adjust linearity with user-defined linearity points (only for xbpi driver) sarto_cmd (BC;M) String No Set preload (only for xbpi driver) sarto_cmd (BC;S) String Yes Clear preload (only for xbpi driver) sarto_cmd (BC;C) String Yes Cancel an active function (all calibration/adjustment functions and tare functions) sarto_cmd (C;BC) String Yes Cancel a tare function sarto_cmd (C;T) String Yes Query whether a function is active (all calibration/adjustment functions and tare functions, if not xbpi driver) sarto_cmd (S;BC) String Yes Query whether a tare function is active sarto_cmd (S;T) String Yes Query: Maximum capacity in range r sarto_cmd (BI;H;r) String Yes Query: Minimum capacity sarto_cmd (BI;L) String Yes Query: Maximum tare value sarto_cmd (BI;T) String Yes Query: Scale interval in range r (r = 0: current range) sarto_cmd (BI;D;r) String Yes
41 Description Command Response Access in verified (const char*outdata) (char*indata) mode permitted 1 Query: Verification scale interval in range r (r = 0: current range) sarto_cmd (BI;E;r) String Yes Read parameter setting p (see xbpi function 0x55) sarto_cmd (BM;R;p) String Yes Read setting option p, i.e. lock flag block (see xbpi function 0x54) sarto_cmd (BM;O;p) String Yes Set parameter setting p (see xbpi function 0x56) sarto_cmd (BM;W;p;n) String Yes Reconfigure a weighing system parameter p (p = M, W, T, A, S, C, U1, U2) sarto_cmd (BS;W;p;n) String Yes 3 Query the setting of a weighing system parameter sarto_cmd (BS;R;p) String Yes Query configurable options of a parameter (bit mask) sarto_cmd (BS;O;p) String Yes Configurable/readable parameters are: Ambient conditions (filter mode) sarto_cmd (BS;W;M;4) String Yes Stability range sarto_cmd (BS;R;W) String Yes Weight unit, level 1 sarto_cmd (BS;R;U1) String Yes Weight unit, level 2 sarto_cmd (BS;R;U2;3) String Yes Menu locking sarto_cmd (BS;W;X;1) String Yes Permanently save device configuration data sarto_cmd (SC;S) String Yes Reset device configuration data to factory settings sarto_cmd (SC;R) String Yes Reconfigure device sarto_cmd (SD;C) String Yes Restart device sarto_cmd (SD;R) String Yes Read the model name of the device sarto_cmd (SI;M) String Yes Read the serial number of the device sarto_cmd (SI;N) String Yes Read the software release of the device driver program sarto_cmd (SI;S) String Yes Read the version number of the current communication library sarto_cmd (SI;D) String Yes Read the version number of the software application sarto_cmd (SI;SW) String Yes Information on active connection sarto_cmd (SI;AC) String Yes External command to device sarto_cmd (XC;cc) String No Hide application 4 sarto_cmd (SLEEP) String Yes Show application sarto_cmd (AWAKE) String Yes Exit application sarto_cmd (QUIT) String Yes 1 If you are using a weighing instrument certified for legal metrology (weighing platform, scale or balance), all functions not permitted in the verifiable mode will generate an error message if they are attempted. These functions will be blocked by the weighing instrument. 2 Tare conditions: 1. No access for negative net value 2. If T(P)2 set, then T(P)1 locked 3. If T(P)1 set, then T(P) locked 3 Non-permissible options are blocked on the weighing instrument. 4 If a weight is requested via a DDE command, the display returns to the foreground. License Agreement The purchaser receives a non-exclusive right of use to the software and is only allowed to make a backup copy of the program. It is otherwise forbidden to make copies of the program or parts of the program or to duplicate the program or parts of it in any other way, including the data temporarily stored in the working memory. The purchaser may not run this program on more than one computer at a time. The purchaser may transfer this right of use to a third party. In this case, the purchaser must physically delete all existing copies of the program that he possesses after giving the program to a third party.
42 Error Messages There are 3 classes of error messages that are displayed in the virtual display and operator terminal: Errors that occurred in the application Errors that occurred in the communication library Errors that occurred in the weighing instrument Errors in the application: Key: ERR: ERROR APP: APPLICATION NR: ERROR NUMBER 01: No link to weighing system due to lack of or faulty communication 02: Reserved 03: Communication library cannot be loaded 04: Service menu is active 05: DDE request cannot be processed. The desired function may not be available. 06 Weight readout is unstable, and the function has been canceled 07 No verifiable memory path defined 08 Data record could not be saved. The directory may not exist. 09 Data record could not be saved. Possibly insufficient storage capacity. 10 Weighing instrument indicates that load is too light (Low) 11 Weighing instrument indicates that load is too heavy (High) 12 No weight available (NAN) 13 Integrity test of the communication library (DLL) failed Errors in the communication library: Key: ERR: ERROR NR: ERROR NUMBER -1: Write string is too long -2: Cannot connect to server -3: Cannot initialize DDE interface -4: Does not start with COM -5: Cannot receive open status from server -6: Cannot connect to scale -7: Read timeout on COM port -8: Read timeout on COM port -9: Internal error communication Client Server -10: Internal error communication Client Server -11: Internal error communication Client Server -12: Internal error communication Client Server -13: Internal error communication Client Server -14: Internal error communication Client Server -15: Open status from server is too short -16: Open status from server signals -17: Automatic server start was not possible -18 Automatic server start not yet supported for 16-bit -19: COM port cannot be opened, not available or used by another application -20: Read timeout in DDE interface -21: No success on writing to DDE server -22: No success on loading DLL server -23: No success on loading DLL server function calls -24: DLL server function calls illegally used -25: Connection successful, but scale unknown -26: Problems with serial communication Frequently Occurring Errors and Troubleshooting Error Cause Solution CRC error: Application terminated Changes were made to the program code of the application. The application cannot be started in this state. No day file available for selection All files present in the alibi memory are displayed. If no files are shown for selection, the alibi memory is probably empty or the path to the directory is not configured correctly. Configuration file not defined Connection parameter unknown in testing of the weighing platform configuration Weighing platform selection list is empty upon device selection using Select device button No connection to a weighing platform could be established. No weighing instrument has been configured. Errors that occurred in the weighing instrument: Key: ERR: ERROR NR: ERROR NUMBER 8: Zeroing for gross value outside of zeroing range 9: Taring for gross value : Internal error code in weighing instrument Reinstall the application. If the error continues to occur, contact your Sartorius dealer or the Sartorius Service Center (see section Installation ) Check the directory path for the alibi memory and reconfigure, if necessary. Data not present in the alibi memory may have been moved to the archive memory or may be located in the memory for altered data. Reinstall the application (see section Installation ). Inspect the communication settings and the connection between the device and the PC. Configure the device using the configuration menu (see section Configuring Communication with the Device ).
43
44
45
46
47
48 Sartorius AG Weender Landstrasse Göttingen Telephone Fax Copyright by Sartorius AG, Göttingen, BR Germany. Reprinting or translation, even in excerpts, is not permitted without written permission from Sartorius AG. All rights reserved by Sartorius AG in accordance with copyright law. The information and figures contained in these instructions correspond to the version date specified below. Sartorius AG reserves the right to make changes to the technology, features, specifications and design of the equipment without notice. Version date: September 2008, Sartorius AG, Göttingen Printed in Germany. Printed on paper bleached without chlorine. JW W3A000.WZA KT Publication No.: WWZ6007-d09082
Sartorius YAD01IS Win
T I F I E D C E R T I F I E D Sartorius YAD01IS Win Service Software for Windows Program Description, Version 1.00 2 Contents Intended Purpose...5 Operating Design...6 YAD01IS Software...7 Installation...7
SA-9600 Surface Area Software Manual
SA-9600 Surface Area Software Manual Version 4.0 Introduction The operation and data Presentation of the SA-9600 Surface Area analyzer is performed using a Microsoft Windows based software package. The
Sartorius ProControl@Informer
Operating Instructions Sartorius ProControl@Informer Evaluation Software for SPCfWin (Sartorius ProControl for Windows) 98646-002-87 Intended Use Sartorius ProControl@Informer (SPC@Informer) is a software
PARTNER ACS R4.0 Remote Administration R4.0. Getting Started
PARTNER ACS R.0 Remote Administration R.0 Getting Started 8-6-66 700080 Issue May 00 Copyright 00, Avaya Inc. Document 8-6-66 All Rights Reserved 700080 Printed in USA Issue May 00 Notice Every effort
File Management Utility. T u t o r i a l
File Management Utility T u t o r i a l Contents System Requirements... 2 Preparing Files for Transfer to GlobalMark... 2 Application Launch... 2 Printer Setup... 2 Communication Status... 4 Communication
MGC WebCommander Web Server Manager
MGC WebCommander Web Server Manager Installation and Configuration Guide Version 8.0 Copyright 2006 Polycom, Inc. All Rights Reserved Catalog No. DOC2138B Version 8.0 Proprietary and Confidential The information
Korosi Monitoring System (KMS) Software Manual
PT. KOROSI SPECINDO Corrosion Monitoring System Eng. Doc. No.: KS-MNL-KMS Korosi Monitoring System (KMS) Software Manual Revision Date Prepared Approved Description 0 23/05/2008 ES JH Issued for approval
DUKANE Intelligent Assembly Solutions
PC Configuration Requirements: Configuration Requirements for ipc Operation The hardware and operating system of the PC must comply with a list of minimum requirements for proper operation with the ipc
Orbit PCI Mk 2 Network Card. User Manual. Part No. 502566 Issue 4
Orbit PCI Mk 2 Network Card User Manual Part No. 502566 Issue 4 Information in this document is subject to change without notice. Companies, names and data used in examples herein are fictitious unless
GUARD1 PLUS SE Administrator's Manual
GUARD1 PLUS SE Administrator's Manual Version 4.4 30700 Bainbridge Road Solon, Ohio 44139 Phone 216-595-0890 Fax 216-595-0991 [email protected] www.guard1.com i 2010 TimeKeeping Systems, Inc. GUARD1 PLUS
User s Guide for version 5 Page
Network Management Interface Catalog VPi 32.15E For use with these ATS products: 7000 & 4000 Series (Group 5 Controller) Series 300 (Group 1 Controller), and 940, 962, 436, 434, 447, 448 (Group 6A/7A Control
Control Technology Corporation CTC Monitor User Guide Doc. No. MAN-1030A Copyright 2001 Control Technology Corporation All Rights Reserved Printed in USA The information in this document is subject to
footscan SCSI system Installation guide: balance software
footscan SCSI system Installation guide: balance software Doc nr RS-PRD-00137-01 PN 80000061 Page 1 Information in this guide is subject to change without notice. No part of this document may be reproduced
EPSON Scan Server & EPSON TWAIN Pro Network
EPSON Scan Server & EPSON TWAIN Pro Network EPSON Scan Server & EPSON TWAIN Pro Network SCANNER UTILITY PROGRAMS All rights reserved. No part of this publication may be reproduced, stored in a retrieval
SHARP Digital Signage Software Pro PN-SS05 OPERATION MANUAL
SHARP Digital Signage Software Pro PN-SS05 Version 4.1 OPERATION MANUAL Contents Introduction... 2 Precautions on Use...2 Trademarks...2 How to Read this Manual...3 Definitions...3 Installing/Launching...
SECURITY SYSTEM MANAGEMENT SOFTWARE FOR WINDOWS. Quick Start Instructions
TM SECURITY SYSTEM MANAGEMENT SOFTWARE FOR WINDOWS Quick Start Instructions Introduction WinLoad for Windows was developed to simplify and speed up the task of programming Paradox control panels. Remote
AccXES Account Management Tool Administrator s Guide Version 10.0
AccXES Account Management Tool Administrator s Guide Version 10.0 701P41531 May 2004 Trademark Acknowledgments XEROX, AccXES, The Document Company, and the identifying product names and numbers herein
Data Analysis Software
TRAKPRO Data Analysis Software User s Manual November 1996 P/N 1980200 Rev. D TSI Incorporated TRAKPRO Data Analysis Software User s Manual November 1996 P/N 1980200 Rev. D SHIP TO: TSI Incorporated 500
Remote Management System
RMS Copyright and Distribution Notice November 2009 Copyright 2009 ARTROMICK International, Inc. ALL RIGHTS RESERVED. Published 2009. Printed in the United States of America WARNING: ANY UNAUTHORIZED
Handheld LPC Utility Software for Windows. Operation Manual
Handheld LPC Utility Software for Windows MODEL S388-70 Operation Manual Read this manual carefully and understand the warnings described in this manual before operating the product. Keep this manual handy
History of Revisions. Ordering Information
No part of this document may be reproduced in any form or by any means without the express written consent of II Morrow Inc. II Morrow, Apollo, and Precedus are trademarks of II Morrow Inc. Windows is
Newton Backup Utility User s Guide. for the Windows Operating System
Newton Backup Utility User s Guide for the Windows Operating System K Apple Computer, Inc. 1995 Apple Computer, Inc. All rights reserved. Under the copyright laws, this manual may not be copied, in whole
Device configurator DRC200
Operating manual 42/49-27 EN Engineer IT Device configurator DRC200 R&C Process Data Management Software Impressum Device configurator DRC200 Operating manual Document No. 42/49-27 EN Date of issue: 11.02
When upgrading your TAPIT software make sure that all call accounting data has been exported before you begin the installation process.
When upgrading your TAPIT software make sure that all call accounting data has been exported before you begin the installation process. Registration Settings: SERIAL NUMBER: COMPUTER ID: REGISTRATION NUMBER:
IEEE 802.11b WLAN PC Card
IEEE 802.11b WLAN PC Card User s Guide Version: 1.3 August 2001 Please install the Utility/Driver first before inserting the PCMCIA Card. FCC Class B Radio Frequency Interference Statement The manufacturer
Analyzer 2.0. Installation Guide. Contents
Analyzer 2.0 Analyzer 2.0 is a comprehensive suite of tools designed to help you analyze complex aqueous and multisolvent electrolyte solution chemistry, thermophysical properties, phase behavior and corrosion
IMPORTANT! CPD 8424R1
IMPORTANT! You need to reinstall EPSON Image Authentication System (IAS) on your camera if it was installed before your camera was repaired. Please see your EPSON Image Authentication System User s Guide
5-Bay Raid Sub-System Smart Removable 3.5" SATA Multiple Bay Data Storage Device User's Manual
5-Bay Raid Sub-System Smart Removable 3.5" SATA Multiple Bay Data Storage Device User's Manual www.vipower.com Table of Contents 1. How the SteelVine (VPMP-75511R/VPMA-75511R) Operates... 1 1-1 SteelVine
Database Binary Backup PC Utility
Database Binary Backup PC Utility Reference Manual (Version 3.xx) DRAFT 2 The flexible way to communicate The information contained in this document is proprietary and is subject to all relevant copyright,
CashFlow Programming Module CPM
CashFlow Programming Module CPM Installation & User Guide GB REV G4 08/08/2006 Table of Content Table of Content... 2 CashFlow Programming Module (CPM)... 4 What is a CPM?... 4 CPM Identification... 4
Legal Notes. Regarding Trademarks. Model supported by the KX printer driver. 2010 KYOCERA MITA Corporation
Legal Notes Unauthorized reproduction of all or part of this guide is prohibited. The information in this guide is subject to change for improvement without notice. We cannot be held liable for any problems
Copyright 1999 by EPSON America, Inc. 3/99 Torrance, California, USA
EPSON is a registered trademark of SEIKO EPSON Corporation. PhotoPC is a registered trademark, and HyPict, Image Authentication System, and FaxAdvice are trademarks of EPSON America, Inc. SoundAdvice is
Installing Windows XP Professional
CHAPTER 3 Installing Windows XP Professional After completing this chapter, you will be able to: Plan for an installation of Windows XP Professional. Use a CD to perform an attended installation of Windows
GX-2009 Data Logger Management Program Operator s Manual
GX-2009 Data Logger Management Program Operator s Manual Part Number: 71-0163RK Revision: C Released: 3/19/12 www.rkiinstruments.com Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by
Instruction manual. testo easyheat Configuration and Analysis software
Instruction manual testo easyheat Configuration and Analysis software en 2 General Information General Information This documentation includes important information about the features and application of
Table of Contents. 1. Overview... 3. 1.1 Materials Required. 3 1.2 System Requirements. 3 1.3 User Mode 3. 2. Installation Instructions..
Table of Contents 1. Overview..... 3 1.1 Materials Required. 3 1.2 System Requirements. 3 1.3 User Mode 3 2. Installation Instructions.. 4 2.1 Installing the On Call Diabetes Management Software. 4 2.2
DriveRight. Fleet Management Software. Getting Started Guide. CarChip. DriveRight. Drivers. Vehicles. Product #8186
DriveRight Fleet Management Software Getting Started Guide CarChip DriveRight Drivers Vehicles Product #8186 DriveRight Fleet Management Software Getting Started Guide; P/N 8186 Davis Instruments Part
Datacolor TOOLS. Datacolor TOOLS QCX. Datacolor TOOLS WORKSTATION
Datacolor TOOLS Datacolor TOOLS QCX Datacolor TOOLS WORKSTATION Installation Instructions Datacolor TOOLS Installation Guide (June, 2009) All efforts have been made to ensure the accuracy of the information
PC Program User s Guide (01.01.05) 1. Connecting the PC. 2. Installation and Start-up. 3. Programming
Work Off-Line Set up a new database or modify a previously saved database (without being connected to the telephone system). This helps minimize on-site programming time. Work On-Line (connected directly
Exeba -ATS. User Guide. Escan Technologies Corporation
Escan Technologies Corporation Exeba -ATS User Guide Escan Technologies Corp. 12140 Severn Way Riverside, CA 92503 Phone (909) 270-0043 Fax (909) 270-0920 1 ESCAN TECHNOLOGIES CORPORATION Exeba -ATS User
USB 2.0 VGA ADAPTER USER MANUAL
USB 2.0 VGA ADAPTER USER MANUAL CONTENTS INTRODUCTION... 3 FEATURES... 3 SYSTEM REQUIREMENTS... 3 PACKAGE CONTENTS... 3 SUPPORTED COMMON DISPLAY RESOLUTION... 4 TECHNICAL SPECIFICATIONS... 4 INSTALLATION
EPSON Stylus C84N. Supplement. Includes information about networking utilities and installation on Windows 98 and Me
EPSON Stylus C84N Supplement Includes information about networking utilities and installation on Windows 98 and Me Copyright Notice All rights reserved. No part of this publication may be reproduced, stored
Network FAX Driver. Operation Guide
Network FAX Driver Operation Guide About this Operation Guide This Operation Guide explains the settings for the Network FAX driver as well as the procedures that are required in order to use the Network
JEROME COMMUNICATIONS SOFTWARE (JCS) INSTALLATION and OPERATION MANUAL V1.02 for Windows XP
AZI P/N 700-0042-B Last update May 2009 USER MANUAL JEROME COMMUNICATIONS SOFTWARE (JCS) INSTALLATION and OPERATION MANUAL V1.02 for Windows XP May 2009 ARIZONA INSTRUMENT LLC 3375 N Delaware Street Chandler,
Centran Version 4 Getting Started Guide KABA MAS. Table Of Contents
Page 1 Centran Version 4 Getting Started Guide KABA MAS Kaba Mas Welcome Kaba Mas, part of the world-wide Kaba group, is the world's leading manufacturer and supplier of high security, electronic safe
OPTAC Fleet Viewer. Instruction Manual
OPTAC Fleet Viewer Instruction Manual Stoneridge Limited Claverhouse Industrial Park Dundee DD4 9UB Help-line Telephone Number: 0870 887 9256 E-Mail: [email protected] Document version 4.0 Part Number:
Aquadyne TechTip TITLE: TROUBLESHOOTING PC COM PORT AND MODEM PRODUCTS AFFECTED SYMPTOMS POSSIBLE CAUSES
Aquadyne TechTip TITLE: TROUBLESHOOTING PC COM PORT AND MODEM COMMUNICATIONS WITH AN OCTOPUS. Article # 072297-1 Last reviewed: 03/25/98 Keywords: Serial Port, Modem, Polling, AquaWeb, Node Not Responding
Guide to Installing BBL Crystal MIND on Windows 7
Guide to Installing BBL Crystal MIND on Windows 7 Introduction The BBL Crystal MIND software can not be directly installed on the Microsoft Windows 7 platform, however it can be installed and run via XP
Batch Eligibility Long Term Care claims
Hewlett Packard Enterprise Provider Electronic Solutions software lets Connecticut Medical Assistance Program providers verify patient s eligibility and submit and correct claims for services all electronically.
Manual. Danfoss Central Station Communications Software Version 3
Manual Danfoss Central Station Communications Software Version 3 RS.8A.Q1.02 1-2001 Contents Introduction 4 System Overview 4 Minimum and Recommended Hardware Requirements 4 Using the Keyboard to Select
VERITAS Backup Exec 9.1 for Windows Servers Quick Installation Guide
VERITAS Backup Exec 9.1 for Windows Servers Quick Installation Guide N109548 Disclaimer The information contained in this publication is subject to change without notice. VERITAS Software Corporation makes
ScanShell.Net Install Guide
ScanShell.Net Install Guide Please install the software first - DO NOT PLUG IN THE SCANNER The scanner has been carefully packaged to avoid damage during transportation. Before operating the scanner, please
Vicon Flash Upgrade Software
INSTRUCTION MANUAL Notes Refer to XX134 NO. XX134-13-01 REV. 1212 Vicon Flash Upgrade Software It is possible to upgrade software for the SurveyorVFT and Surveyor -Mini Camera Domes through a PC using
Config software for D2 systems USER S MANUAL
DT-CONFIG SOFTWARE Config software for D2 systems USER S MANUAL CONTENTS 1. Introductions ------------------------------------------------------------------- 3 2. System Requirement and Connection ----------------------------------------
S&G Audit Lock. Audit Trail Software Manual
S&G Audit Lock Audit Trail Software Manual The Sargent & Greenleaf Audit Lock is designed to provide a high level of security. One of the ways they accomplish this is by offering detailed audit trails.
Operating Manual QUESTOR
QUESTOR AS 273 Management Software Document: KSW3s527.0004 / en 2010.08 Edition: August 2010 QUESTOR TABLE OF CONTENT 1 Product description EN-4 1.1 Purpose... EN-4 1.2 System components... EN-4 1.2.1
Getting Started with Vision 6
Getting Started with Vision 6 Version 6.9 Notice Copyright 1981-2009 Netop Business Solutions A/S. All Rights Reserved. Portions used under license from third parties. Please send any comments to: Netop
User s Manual. Management Software for ATS
ATS Monitor User s Manual Management Software for ATS Table of Contents 1. ATS Monitor Overview... 2 2. ATS Monitor Install and Uninstall... 2 2.1. System Requirement... 2 2.2. Software Install... 2 2.3.
Partner. Sage Pastel. Accounting. Installation Guide
Sage Pastel Accounting Partner Installation Guide Sage Pastel: +27 11 304 3000 Sage Pastel Intl: +27 11 304 3400 www.pastel.co.za www.sagepastel.com [email protected] [email protected] Sage Pastel Accounting
PORTABLE DATA COLLECTION TERMINAL
PORTABLE DATA COLLECTION TERMINAL WINDOWS TRANSFER UTILITY HELP FILES BETA EDITION Compsee, Inc. A Subsidiary of McRae Industries, Inc. 2 Welcome Thank you for using the Apex III Windows Transfer Utility
ViewPower. User s Manual. Management Software for Uninterruptible Power Supply Systems
ViewPower User s Manual Management Software for Uninterruptible Power Supply Systems Table of Contents 1. ViewPower Overview... 2 1.1. Introduction...2 1.2. Structure...2 1.3. Applications...2 1.4. Features...2
Combi B PC software 8213X. Installation and operating instructions
Combi B PC software 8213X Installation and operating instructions EN Table of Contents 1 Important notes... 3 2 Glossary of abbreviations and terms... 4 3 Symbols and special font types used... 5 4 Scope
MGC WebCommander Web Server Manager
MGC WebCommander Web Server Manager Installation and Configuration Guide Version 9.0 Copyright 2007 Polycom, Inc. All Rights Reserved Catalog No. DOC2138C Version 9.0 Proprietary and Confidential The information
NDA-30141 ISSUE 1 STOCK # 200893. CallCenterWorX-Enterprise IMX MAT Quick Reference Guide MAY, 2000. NEC America, Inc.
NDA-30141 ISSUE 1 STOCK # 200893 CallCenterWorX-Enterprise IMX MAT Quick Reference Guide MAY, 2000 NEC America, Inc. LIABILITY DISCLAIMER NEC America, Inc. reserves the right to change the specifications,
Page 1 S5.5 Monitoring Software Manual Rev 1.2 for Software Ver 1.0
Trantec S5.5 Monitoring Software Manual Initial Setup Minimum PC Requirements 1.6 GHz CPU, 256 MB Memory, 1 available USB port, Windows XP Recommended PC >2.0 GHz CPU, 512 MB (XP) 1 GB (Vista) memory,
Motorola Phone Tools. Quick Start
Motorola Phone Tools Quick Start Contents Minimum Requirements...2 Before Installing Motorola Phone Tools...3 Installing Motorola Phone Tools...4 Mobile Device Installation and Configuration...5 Online
1 Minimum system requirements
Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 [email protected] www.metrohm.com Installation MagIC Net 2.x 1 Minimum system requirements Operating system RAM Memory
VRC 7900/8900 Avalanche Enabler User s Manual
VRC 7900/8900 Avalanche Enabler User s Manual WLE-VRC-20030702-02 Revised 7/2/03 ii Copyright 2003 by Wavelink Corporation All rights reserved. Wavelink Corporation 6985 South Union Park Avenue, Suite
SYNERGY CONFIGURATION SOFTWARE AND HARDWARE SETUP V 2.2 11/02
SYNERGY CONFIGURATION SOFTWARE AND HARDWARE SETUP V 2.2 11/02 CONTENTS: SECTION 1: WIN 95/98...2 INSTALL SYNERGY CONFIGURATION SOFTWARE...2 INSTALL NIC PCI CARD AND DRIVER FOR WIN 95/98/ME...2 RECORD SETTINGS
Introduction and Overview
Inmagic Content Server Workgroup 10.00 Microsoft SQL Server 2005 Express Edition Installation Notes Introduction and Overview These installation notes are intended for the following scenarios: 1) New installations
FAS Asset Accounting FAS CIP Accounting FAS Asset Inventory SQL Server Installation & Administration Guide Version 2011.1
FAS Asset Accounting FAS CIP Accounting FAS Asset Inventory SQL Server Installation & Administration Guide Version 2011.1 2011 Sage Software, Inc. All rights reserved. Published by Sage 2325 Dulles Corner
Meridian 1 Meridian 1 Attendant PC LAN Interface Installation Guide
Meridian 1 Meridian 1 Attendant PC LAN Interface Installation Guide Document Number: P0874391 Document Release: Standard 1.00 Date: November 1997 Year Publish FCC TM 1997 All rights reserved Printed in
2-Bay Raid Sub-System Smart Removable 3.5" SATA Multiple Bay Data Storage Device User's Manual
2-Bay Raid Sub-System Smart Removable 3.5" SATA Multiple Bay Data Storage Device User's Manual www.vipower.com Table of Contents 1. How the SteelVine (VPMP-75211R/VPMA-75211R) Operates... 1 1-1 SteelVine
Virtual CD v10. Network Management Server Manual. H+H Software GmbH
Virtual CD v10 Network Management Server Manual H+H Software GmbH Table of Contents Table of Contents Introduction 1 Legal Notices... 2 What Virtual CD NMS can do for you... 3 New Features in Virtual
Quickstart Tutorial. Bradford Technologies, Inc. 302 Piercy Road, San Jose, California 95138 800-622-8727 fax 408-360-8529 www.bradfordsoftware.
Quickstart Tutorial A ClickFORMS Tutorial Page 2 Bradford Technologies. All Rights Reserved. No part of this document may be reproduced in any form or by any means without the written permission of Bradford
Model 288B Charge Plate Graphing Software Operators Guide
Monroe Electronics, Inc. Model 288B Charge Plate Graphing Software Operators Guide P/N 0340175 288BGraph (80207) Software V2.01 100 Housel Ave PO Box 535 Lyndonville NY 14098 1-800-821-6001 585-765-2254
Network Scanner Tool R3.1. User s Guide Version 3.0.04
Network Scanner Tool R3.1 User s Guide Version 3.0.04 Copyright 2000-2004 by Sharp Corporation. All rights reserved. Reproduction, adaptation or translation without prior written permission is prohibited,
TECHNICAL BULLETIN. Configuring Wireless Settings in an i-stat 1 Wireless Analyzer
i-stat TECHNICAL BULLETIN Configuring Wireless Settings in an i-stat 1 Wireless Analyzer Before configuring wireless settings, please enable the wireless functionality by referring to the Technical Bulletin
Power Log PC Application Software
Power Log PC Application Software Users Manual May 2007 Rev. 1, 5/08 2007-2008 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks
Diamond II v2.3 Service Pack 4 Installation Manual
Diamond II v2.3 Service Pack 4 Installation Manual P/N 460987001B ISS 26APR11 Copyright Disclaimer Trademarks and patents Intended use Software license agreement FCC compliance Certification and compliance
SECTION 3 - INSTALLATION
SECTION 3 - INTRODUCTION This section explains how to install the Strategic Enterprise Management Application Programming Interface (semapi) software and how to use the ICICONF.EXE program to configure
Getting Started with IntelleView POS Administrator Software
Getting Started with IntelleView POS Administrator Software Administrator s Guide for Software Version 1.2 About this Guide This administrator s guide explains how to start using your IntelleView POS (IntelleView)
User Manual. Thermo Scientific Orion
User Manual Thermo Scientific Orion Orion Star Com Software Program 68X637901 Revision A April 2013 Contents Chapter 1... 4 Introduction... 4 Star Com Functions... 5 Chapter 2... 6 Software Installation
Getting Started. rp5800, rp5700 and rp3000 Models
Getting Started rp5800, rp5700 and rp3000 Models Copyright 2011 Hewlett-Packard Development Company, L.P. The information contained herein is subject to change without notice. Microsoft, Windows, and Windows
GE Sensing. SiCalPro. Monitor and control software. User manual - K0452
GE Sensing SiCalPro Monitor and control software User manual - K0452 Safety Before you use the SiCalPro software program, make sure that you read and understand all the related data. This includes: the
Operating Instructions. Windows. Logo Transfer Software 2.0. Thank you for purchasing this Panasonic product.
Operating Instructions Logo Transfer Software 2.0 Windows Thank you for purchasing this Panasonic product. Before using this software, please read the instructions carefully. TQBH9059-5 Table of Contents
Cyberex BCM-SFCM Configurator/Monitor User's Manual
BRANCH/SUBFEED CIRCUIT MANAGEMENT Cyberex BCM-SFCM Configurator/Monitor User's Manual Updated: 04/01/13 Thomas & Betts Power Solutions 5900 Eastport Boulevard Richmond, VA 23231-4453 Tel: 800-238-5000
Introducing Windows CE
3 Introducing Windows CE Windows CE Operating System Windows CE Services Software Applications Online Help Introducing Windows CE 3-1 WINDOWS CE OPERATING SYSTEM Windows CE is the smallest member of the
Section 5: Installing the Print Driver
Section 5: Installing the Print Driver In This Section Print Driver Installation for USB Connection Using the Ethernet Connection and a Crossover Cable Print Driver: The Epilog print driver is the software
DataSuite Installation and Activation Guide
DataSuite Installation and Activation Guide This document provides a detailed overview of the installation and activation procedure for the DataSuite v2.4.0.0 software. DataSuite allows you to configure
Administration guide. Host software WinCCU Installation. Complete gas volume and energy data management
Administration guide Host software WinCCU Installation Complete gas volume and energy data management Contents 1 Introduction... 1 Safety first... 1 Warning... 1 Typographic conventions... 1 Product versioning...
DIGITAL MULTIFUNCTIONAL SYSTEM
MODEL: MX-M283N MX-M363N MX-M453N MX-M503N DIGITAL MULTIFUNCTIONAL SYSTEM Software Setup Guide BEFORE INSTALLING THE SOFTWARE SETUP IN A WINDOWS ENVIRONMENT SETUP IN A MACINTOSH ENVIRONMENT TROUBLESHOOTING
3 Setting up Databases on a Microsoft SQL 7.0 Server
3 Setting up Databases on a Microsoft SQL 7.0 Server Overview of the Installation Process To set up GoldMine properly, you must follow a sequence of steps to install GoldMine s program files, and the other
HP Point of Sale (POS) Peripherals Configuration Guide ap5000 VFD Windows (non-opos)
HP Point of Sale (POS) Peripherals Configuration Guide ap5000 VFD Windows (non-opos) Document Version 3.23 November 2010 1 Copyright 2007-2010 Hewlett-Packard Development Company, L.P. The information
SOFTWARE INSTALLATION INSTRUCTIONS
AUDITGARD LGA Electronic Combination Lock SOFTWARE INSTALLATION INSTRUCTIONS Implementation Package The AuditGard Software implementation package includes: 707013 AuditGard Software Install CD 42145 Programming/Audit
Enterprise Vault Installing and Configuring
Enterprise Vault Installing and Configuring Enterprise Vault 6.0 Legal Notice Copyright 2005 Symantec Corporation. All rights reserved. Symantec, the Symantec Logo, VERITAS, the VERITAS Logo, and Enterprise
NOTE: You may want to choose the Word Wrap option from Notepad Edit menu to make this text more readable.
------------------ Interactions/Gas Exchange Version 1.2, CD Version Copyright 2003 by John Wiley & Sons, Inc. All rights reserved. ------------------ NOTE: You may want to choose the Word Wrap option
Tapi Link For GoldMine - Installation Guide
Version No: 2.2.1.668 (for GoldMine 5.x and 6.x) Version No: 3.5.1.668 (for GoldMine 7.x) Version No: 4.5.1.668 (for GoldMine 8.x) Version No: 5.5.1.668 (for GoldMine 9.x) Papillon Technology 1997-2011
ABB solar inverters. Installation and setup instruction Aurora Manager TL software and ABB PVI-USB-RS232_485 adapter
ABB solar inverters Installation and setup instruction Aurora Manager TL software and ABB PVI-USB-RS232_485 adapter Introduction This document details the procedure for the installation on a user s computer
LPR for Windows 95/98/Me/2000/XP TCP/IP Printing User s Guide. Rev. 03 (November, 2001)
LPR for Windows 95/98/Me/2000/XP TCP/IP Printing User s Guide Rev. 03 (November, 2001) Copyright Statement Trademarks Copyright 1997 No part of this publication may be reproduced in any form or by any
