QUALITATIVE AND QUANTITATIVE ANALYSIS BY INFRARED SPECTROSCOPY AND MOLECULAR FLUORESCENCE (VERSION 1.0)
|
|
- Nathan Craig
- 7 years ago
- Views:
Transcription
1 REVISED Infrared Spectroscopy/Molecular Fluorescence, Page 1 QUALITATIVE AND QUANTITATIVE ANALYSIS BY INFRARED SPECTROSCOPY AND MOLECULAR FLUORESCENCE I. BACKGROUND. (VERSION 1.0) In this experiment, a solid unknown will be analyzed by use of infrared spectroscopy and molecular fluorescence. The unknown species will be identified from a list of possibilities by collecting the characteristic infrared fingerprint of the compound. A quantitative analysis will then be performed by use of fluorescence spectroscopy. Infrared Spectroscopy Infrared spectroscopy is based on the absorption of infrared energy by molecular species and the corresponding excitation of vibrational modes of the molecule. The mechanism of the absorption transition is based on the interaction of a bond dipole in the molecule with the electric field of the incident photon. The presence of multiple bond dipoles within a typical organic molecule, the possibility for several vibrational modes being associated with each dipole, and the coupling of vibrational and rotational transitions gives rise to a rich pattern of absorptions and a corresponding characteristic infrared fingerprint for a compound. Because a bond dipole requires a difference in electronegativity between the bonded atoms, bond dipoles are naturally associated with organic functional groups. The infrared spectral pattern is thus rich in information regarding the functional groups present in a molecule. The spectra of solid samples can be acquired in several ways. The solid material can be formed into a transparent disk by mixing it with an infrared-inactive material such as KBr and applying pressure. The resulting disk is analyzed with a conventional transmission measurement. The solid can be dissolved in a solvent and also analyzed with a transmission measurement. The drawback to this approach is the strong infrared absorbance of most common solvents. The solvent absorbance may effectively blank out entire regions of the spectrum. Those solvents with relatively good infrared transparency unfortunately also tend to be hazardous compounds (e.g., CS 2, CHCl 3, CCl 4, etc.). Handling and disposal issues thus arise when these solvents are used. Another possibility is to make a mull with a mineral oil (Nujol) and place the resulting material between two salt plates. A transmission measurement is then performed. The mineral oil obliterates the C-H stretching region between cm -1, but the fingerprint region is preserved. A third sampling approach, termed attenuated total reflectance (ATR), is depicted in Figure 1. In the figure, the infrared beam is propagated through a diamond waveguide by use of the principle of total internal reflection. Light launched into the waveguide reaches the interface between the waveguide material and the analyte in contact with it. The refractive indices are such that the light is reflected at the interface and returns to the waveguide. The light thus propagates through the waveguide as a series of reflective bounces back and forth at the interface. This is illustrated in Figure 1. Light exiting the waveguide is detected. When undergoing reflection, the electric field associated with the propagating photon penetrates a short distance into the analyte. This is termed the evanescence field of the photon. This penetration of the evanescence field allows interaction between the photon and analyte and can result in absorption if the vibrational and photon frequencies match. Thus, the spectral signature of the analyte is superimposed on the detected light. When a solid material is analyzed by ATR, an important requirement is good contact
2 REVISED Infrared Spectroscopy/Molecular Fluorescence, Page 2 Figure 1. Schematic of attenuated total reflectance with a diamond waveguide. between the sample and the waveguide. The short penetration depth (µm) of the evanescence field into the sample dictates that the sample must be pressed against the waveguide. The ATR accessory for use with solids often has a mechanism for applying pressure to the sample, thereby assuring good contact with the waveguide. Fluorescence Spectroscopy Fluorescence spectroscopy is a sensitive optical emission technique in which sample molecules are excited with a photon source. Those molecules that relax by radiant emission can be subsequently detected by measuring the intensity of that emission. The mechanism of this excitation and emission is illustrated in Figure 2. In the diagram, electronic excitation can occur from the ground state (S 0 ) to either of two excited states (S 1 and S 2 ). Each electronic state has as associated set of vibrational and rotational states. Following excitation, a variety of relaxation processes can occur to return the molecule to the preferred lower energy state. These processes can either be radiative (i.e., light-emitting) or non-radiative (i.e., non-lightemitting). Fluorescence is defined as radiative emission arising from a transition between a singlet excited electronic state and the ground state. The non-radiative processes are more efficient and are typically favored. Often, these processes are driven by collisions between the analyte molecule and the surrounding molecules of the solvent. For this reason, most molecules do not emit light upon relaxation from the excited electronic state. Instead, the energy is returned to the environment, typically causing the sample to increase in temperature. However, certain compounds under certain conditions (e.g., temperature, ph, etc.) do relax radiatively with enough probability that the photon emission can serve as the basis for an analytical measurement. In these cases, the fluorescence measurement is very sensitive because it is a zero-background method (i.e., one is observing the emission against a zero background as opposed to measuring a ratio of light intensities such as that required in a transmission experiment). Fluorescence can also be more selective than an absorption measurement because other potentially interfering constituents of the sample may not relax radiatively.
3 REVISED Infrared Spectroscopy/Molecular Fluorescence, Page 3 Figure 2. Schematic for radiative and non-radiative relaxation processes following absorption of a photon. Transmission experiments produce measurements of absorbance, which is linearly related to concentration through the Beer-Lambert law: A = -log 10 (P / P 0 ) = εbc (1) In Eq. 1, A is absorbance, P is the light power transmitted through the sample, P 0 is the light power incident on the sample, ε is the molar absorptivity, b is the optical path length, and C is analyte concentration. In the fluorescence experiment, F, the light power emitted by the sample is proportional to the light power absorbed: F = kφ F (P 0 P) (2) In Eq. 2, Φ F is termed the fluorescence quantum yield, defined as the fraction of excited molecules that undergo fluorescence. The constant, k, is the optical collection efficiency, defined as the fraction of emitted photons that are detected by the spectrometer. Solving Eq. 1 for P yields: P = P εbc (3) Combining Eqs. 2 and 3 yields:
4 REVISED Infrared Spectroscopy/Molecular Fluorescence, Page 4 F = kφ F P 0 (1-10 -εbc ) (4) The expression in Eq. 4 defines the relationship between measured fluorescence intensity and analyte concentration. Clearly, this relationship is nonlinear. At dilute concentrations, however, the functional relationship between F and C becomes approximately linear. One component of the experimental design in fluorescence is to define a working range of concentration that yields a response that is linear enough to allow use of standard linear calibration methods based on least-squares analysis. Another issue in implementing a successful analysis based on fluorescence measurements is the control of experimental parameters such as ph. Changing the ph can change the probability of radiative emission through the creation of different resonance structures of the molecule. Consequently, optimization of ph is an important part of the experimental design. II. REAGENTS. The following solutions will be provided: ph 7 buffer (0.1 M NaH 2 PO 4 + NaOH to ph 7) ph 10 buffer (0.1 M H 3 BO 3 + NaOH to ph 10) ph 2 buffer (0.05 M H 2 SO 4 + NH 4 OH to ph 2) III. PROCEDURE. First Laboratory Period The goal of the work in the first laboratory period is to identify the unknown and to establish optimal conditions for the fluorescence analysis of each compound. Groups 1 and 2 should begin with infrared spectroscopy, while Groups 3 and 4 should begin with fluorescence spectroscopy. Infrared Spectroscopy Your group will be given an unknown and will be assigned one standard for which you will acquire infrared spectra. The instructor will demonstrate the operation of the infrared spectrometer and provide instruction in the sample preparation methods available. Collect spectra of the unknown and the standard. Provide copies of the spectra of the standard to the other groups. Fluorescence Spectroscopy A. Prepare 10 ppm solutions of your assigned standard in 100 ml volumetric flasks. Make three separate solutions by diluting with the ph 2, ph 7, and ph 10 buffers. For each solution, determine the peak excitation wavelength and record an emission spectrum based on the selected excitation wavelength. Save the recorded spectra in text files. In each case, use the corresponding buffer as the spectroscopic blank (i.e., the measurement of the solution diluted with ph 2 buffer uses ph 2 buffer as the blank). It is possible that your standard will fluoresce strongly enough for the emission intensity to be off-scale at 10 ppm. If this is the case, proceed to part B. B. With each solution, perform serial dilutions by a factor of 4 by taking a 25 ml aliquot of the
5 REVISED Infrared Spectroscopy/Molecular Fluorescence, Page 5 solution with a volumetric pipette and diluting to 100 ml in a volumetric flask. Using the same excitation wavelength determined previously, record emission spectra of each solution. Continue this 1:4 serial dilution until the emission spectra are no longer distinguishable from noise. Second Laboratory Period -- End of First Laboratory Period -- Table 1. Optimal Fluorescence Conditions Compound Optimal ph Excitation λ (nm) Emission λ (nm) Working Range (ppm) Quinine sulfate Salicylic acid Salicylamide A. On the basis of the collected infrared spectra of the standards, identify your unknown. B. Using a pure sample of your unknown compound, prepare a 10 ppm solution in a 1 L volumetric flask. Dilute with water. If the compound will not dissolve, the ph may need to be adjusted. Using 100 ml volumetric flasks, perform serial dilutions of this solution as needed in order to produce four standards that span the working range of your unknown as defined in Table 1 above. Perform all dilutions with water until the final solution. Dilute that solution with the appropriate ph buffer as defined in Table 1. C. Set the excitation monochromator to the optimal excitation wavelength for your compound. Establish a 20 nm scanning window centered on the optimal emission wavelength. Scan the emission spectral window. Make two replicate measurements of each standard without refilling the sample cell. Save the recorded spectra in text files so that they can be exported into Excel. Note the maximum and minimum fluorescence intensities across the four standards. D. Prepare a solution of your unknown that produces a fluorescence signal that falls in the middle of the range spanned by your standards. Begin by using the same procedure used to produce the high concentration standard. Start with 0.01 g of the unknown diluted to 1 L in a volumetric flask. As before, perform dilutions with water until the final solution that is to be measured. Dilute that solution with the appropriate ph buffer. Measure the fluorescence and adjust the dilution scheme as needed. When the optimized solution is found, make two replicate measurements. Save the recorded spectra in text files IV. CALCULATIONS. -- End of Second Laboratory Period -- A. Draw the chemical structure of your unknown. Use the data in the provided table of infrared group frequencies to assign at least three major bands in the infrared spectrum of your unknown to a specific vibrational mode (e.g., C-O stretch, etc.).
6 REVISED Infrared Spectroscopy/Molecular Fluorescence, Page 6 B. Compute the exact concentration for each of your calibration standards. C. For the calibration data, your measured emission spectra consist of recorded emission intensities across the 20 nm scanning window. Import these traces into Excel and compute the maximum intensity within the scanning window (use the max function in Excel). This value will be used as fluorescence intensity in subsequent calculations and plots. Note that fluorescence is measured in arbitrary units. D. For the measurement of the calibration standards, use your replicate measurements to tabulate the mean, standard deviation, and 95% confidence interval for the fluorescence intensities. E. Construct a plot (with error bars) of fluorescence intensity vs. analyte concentration (mg/l) for the calibration data. Use least-squares procedures to determine the calibration model. Report the slope, intercept, the value of r 2 and the standard error of estimate for your calibration model. F. Using your calibration model, compute the concentration of your unknown as measured and work backward through your dilution procedure to calculate the weight percent of the analyte in the original solid sample. Predict each replicate measurement separately and then compute an average, standard deviation, and 95% confidence interval.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
More informationAustin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law
Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine
More informationProblem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:
Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T
More informationEXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare
More informationKINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)
Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:
More informationMolecular Spectroscopy
Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended
More informationEXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline
EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution
More informationQUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.
QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition
More informationSpectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
More informationFundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
More informationIntroduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
More information2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY
2nd/3rd Year Physical Chemistry Practical Course, Oxford University 2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY (4 points) Outline Spectrometry is widely used to monitor the progress
More informationSpectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
More informationRaman spectroscopy Lecture
Raman spectroscopy Lecture Licentiate course in measurement science and technology Spring 2008 10.04.2008 Antti Kivioja Contents - Introduction - What is Raman spectroscopy? - The theory of Raman spectroscopy
More informationExperiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
More informationUniversity of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical
More information0 10 20 30 40 50 60 70 m/z
Mass spectrum for the ionization of acetone MS of Acetone + Relative Abundance CH 3 H 3 C O + M 15 (loss of methyl) + O H 3 C CH 3 43 58 0 10 20 30 40 50 60 70 m/z It is difficult to identify the ions
More informationThe Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD
TN-100 The Fundamentals of Infrared Spectroscopy The Principles of Infrared Spectroscopy Joe Van Gompel, PhD Spectroscopy is the study of the interaction of electromagnetic radiation with matter. The electromagnetic
More informationApplication Note: Absorbance
Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light
More informationColorimetric Determination of Iron in Vitamin Tablets
Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered
More informationATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
More information5.33 Lecture Notes: Introduction to Spectroscopy
5.33 Lecture Notes: ntroduction to Spectroscopy What is spectroscopy? Studying the properties of matter through its interaction with different frequency components of the electromagnetic spectrum. Latin:
More informationInfrared Spectroscopy: Theory
u Chapter 15 Infrared Spectroscopy: Theory An important tool of the organic chemist is Infrared Spectroscopy, or IR. IR spectra are acquired on a special instrument, called an IR spectrometer. IR is used
More informationwhere h = 6.62 10-34 J s
Electromagnetic Spectrum: Refer to Figure 12.1 Molecular Spectroscopy: Absorption of electromagnetic radiation: The absorptions and emissions of electromagnetic radiation are related molecular-level phenomena
More information13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
More informationPesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
More informationRaman Spectroscopy. 1. Introduction. 2. More on Raman Scattering. " scattered. " incident
February 15, 2006 Advanced Physics Laboratory Raman Spectroscopy 1. Introduction When light is scattered from a molecule or crystal, most photons are elastically scattered. The scattered photons have the
More informationA Beer s Law Experiment
A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining
More informationChemistry 111 Lab: Intro to Spectrophotometry Page E-1
Chemistry 111 Lab: Intro to Spectrophotometry Page E-1 SPECTROPHOTOMETRY Absorption Measurements & their Application to Quantitative Analysis study of the interaction of light (or other electromagnetic
More informationSymmetric Stretch: allows molecule to move through space
BACKGROUND INFORMATION Infrared Spectroscopy Before introducing the subject of IR spectroscopy, we must first review some aspects of the electromagnetic spectrum. The electromagnetic spectrum is composed
More informationReaction Stoichiometry and the Formation of a Metal Ion Complex
Reaction Stoichiometry and the Formation of a Metal Ion Complex Objectives The objectives of this laboratory are as follows: To use the method of continuous variation to determine the reaction stoichiometry
More informationAbsorption by atmospheric gases in the IR, visible and UV spectral regions.
Lecture 6. Absorption by atmospheric gases in the IR, visible and UV spectral regions. Objectives: 1. Gaseous absorption in thermal IR. 2. Gaseous absorption in the visible and near infrared. 3. Gaseous
More informationInfrared Spectroscopy 紅 外 線 光 譜 儀
Infrared Spectroscopy 紅 外 線 光 譜 儀 Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample (nondestructive method). The amount of light absorbed
More informationLab #11: Determination of a Chemical Equilibrium Constant
Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a
More informationUltraviolet-Visible (UV-Vis) Spectroscopy Background Information
1 Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information Instructions for the Operation of the Cary 300 Bio UV-Visible Spectrophotometer See the Thermo OMNIC Help reference on page 49. Ultraviolet-Visible
More informationUpon completion of this lab, the student will be able to:
1 Learning Outcomes EXPERIMENT B4: CHEMICAL EQUILIBRIUM Upon completion of this lab, the student will be able to: 1) Analyze the absorbance spectrum of a sample. 2) Calculate the equilibrium constant for
More informationChemistry 111 Laboratory Experiment 7: Determination of Reaction Stoichiometry and Chemical Equilibrium
Chemistry 111 Laboratory Experiment 7: Determination of Reaction Stoichiometry and Chemical Equilibrium Introduction The word equilibrium suggests balance or stability. The fact that a chemical reaction
More informationOrganic Chemistry Tenth Edition
Organic Chemistry Tenth Edition T. W. Graham Solomons Craig B. Fryhle Welcome to CHM 22 Organic Chemisty II Chapters 2 (IR), 9, 3-20. Chapter 2 and Chapter 9 Spectroscopy (interaction of molecule with
More information2 Spectrophotometry and the Analysis of Riboflavin
2 Spectrophotometry and the Analysis of Riboflavin Objectives: A) To become familiar with operating the Platereader; B) to learn how to use the Platereader in determining the absorption spectrum of a compound
More informationINFRARED SPECTROSCOPY (IR)
INFRARED SPECTROSCOPY (IR) Theory and Interpretation of IR spectra ASSIGNED READINGS Introduction to technique 25 (p. 833-834 in lab textbook) Uses of the Infrared Spectrum (p. 847-853) Look over pages
More informationUltraviolet Spectroscopy
Ultraviolet Spectroscopy The wavelength of UV and visible light are substantially shorter than the wavelength of infrared radiation. The UV spectrum ranges from 100 to 400 nm. A UV-Vis spectrophotometer
More informationANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION
Chem 306 Section (Circle) M Tu W Th Name Partners Date ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION Materials: prepared acetylsalicylic acid (aspirin), stockroom samples
More informationTime out states and transitions
Time out states and transitions Spectroscopy transitions between energy states of a molecule excited by absorption or emission of a photon hn = DE = E i - E f Energy levels due to interactions between
More informationSpectrophotometric Determination of the pka of Bromothymol Blue
Spectrophotometric Determination of the pka of Bromothymol Blue INRODUCION cidbase indicators are compounds that are simply weak acids (or bases) that exhibit different colors depending on whether they
More informationPhenolphthalein-NaOH Kinetics
Phenolphthalein-NaOH Kinetics Phenolphthalein is one of the most common acid-base indicators used to determine the end point in acid-base titrations. It is also the active ingredient in some laxatives.
More informationDetermination of Molecular Structure by MOLECULAR SPECTROSCOPY
Determination of Molecular Structure by MOLEULAR SPETROSOPY hemistry 3 B.Z. Shakhashiri Fall 29 Much of what we know about molecular structure has been learned by observing and analyzing how electromagnetic
More informationLab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student. LBS 171L Section 9 TA: Dana October 27, 2005
How uch Phosphate is the Body Being Exposed to During a Lifetime by Showering? Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student LBS 171L Section 9 TA: Dana October 7, 005 [Note:
More informationPROTON NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY (H-NMR)
PROTON NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY (H-NMR) WHAT IS H-NMR SPECTROSCOPY? References: Bruice 14.1, 14.2 Introduction NMR or nuclear magnetic resonance spectroscopy is a technique used to determine
More informationCopyright 1999 2010 by Mark Brandt, Ph.D. 12
Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits
More informationDETERMINACIÓN DE ESTRUCTURAS ORGÁNICAS (ORGANIC SPECTROSCOPY) IR SPECTROSCOPY
DETERMINACIÓN DE ESTRUCTURAS ORGÁNICAS (ORGANIC SPECTROSCOPY) IR SPECTROSCOPY Hermenegildo García Gómez Departamento de Química Instituto de Tecnología Química Universidad Politécnica de Valencia 46022
More informationProton Nuclear Magnetic Resonance Spectroscopy
Proton Nuclear Magnetic Resonance Spectroscopy Introduction: The NMR Spectrum serves as a great resource in determining the structure of an organic compound by revealing the hydrogen and carbon skeleton.
More informationSimple Laser-Induced Fluorescence Setup to Explore Molecular Spectroscopy. Abstract
Simple Laser-Induced Fluorescence Setup to Explore Molecular Spectroscopy S. B. Bayram and M.D. Freamat Miami University, Department of Physics, Oxford, OH 45056 (Dated: July 23, 2012) Abstract We will
More informationCHE334 Identification of an Unknown Compound By NMR/IR/MS
CHE334 Identification of an Unknown Compound By NMR/IR/MS Purpose The object of this experiment is to determine the structure of an unknown compound using IR, 1 H-NMR, 13 C-NMR and Mass spectroscopy. Infrared
More informationUsed to determine relative location of atoms within a molecule Most helpful spectroscopic technique in organic chemistry Related to MRI in medicine
Structure Determination: Nuclear Magnetic Resonance CHEM 241 UNIT 5C 1 The Use of NMR Spectroscopy Used to determine relative location of atoms within a molecule Most helpful spectroscopic technique in
More informationUV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this
More informationCHEM 51LB: EXPERIMENT 5 SPECTROSCOPIC METHODS: INFRARED AND NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY
CHEM 51LB: EXPERIMENT 5 SPECTROSCOPIC METHODS: INFRARED AND NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY REACTIONS: None TECHNIQUES: IR, NMR Infrared (IR) and nuclear magnetic resonance (NMR) spectroscopy are
More informationFTIR Instrumentation
FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation
More information2. Molecular stucture/basic
2. Molecular stucture/basic spectroscopy The electromagnetic spectrum Spectral region for atomic and molecular spectroscopy E. Hecht (2nd Ed.) Optics, Addison-Wesley Publishing Company,1987 Spectral regions
More informationNuclear Magnetic Resonance Spectroscopy
Nuclear Magnetic Resonance Spectroscopy Nuclear magnetic resonance spectroscopy is a powerful analytical technique used to characterize organic molecules by identifying carbonhydrogen frameworks within
More informationChapter 5 -- The Spectrophotometric Determination of the ph of a Buffer. NAME: Lab Section: Date: Sign-Off:
Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer NAME: Lab Section: Date: Sign-Off: Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer Introduction Weak acids,
More informationRaman Scattering Theory David W. Hahn Department of Mechanical and Aerospace Engineering University of Florida (dwhahn@ufl.edu)
Introduction Raman Scattering Theory David W. Hahn Department of Mechanical and Aerospace Engineering University of Florida (dwhahn@ufl.edu) The scattering of light may be thought of as the redirection
More informationDetermination of the Mass Percentage of Copper in a Penny. Introduction
Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny
More informationFor example: (Example is from page 50 of the Thinkbook)
SOLVING COMBINED SPECTROSCOPY PROBLEMS: Lecture Supplement: page 50-53 in Thinkbook CFQ s and PP s: page 216 241 in Thinkbook Introduction: The structure of an unknown molecule can be determined using
More informationChem 131A: Absorbance of Riboflavin
Chem 131A: Absorbance of Riboflavin Purpose: The purpose of this experiment is to: 1) Familiarize the student with the use of the HP 8452 diode array spectrophotometer, 2) examine the limitations of the
More informationCalculation of Molar Masses. Molar Mass. Solutions. Solutions
Molar Mass Molar mass = Mass in grams of one mole of any element, numerically equal to its atomic weight Molar mass of molecules can be determined from the chemical formula and molar masses of elements
More informationDetermination of Equilibrium Constants using NMR Spectrscopy
CHEM 331L Physical Chemistry Laboratory Revision 1.0 Determination of Equilibrium Constants using NMR Spectrscopy In this laboratory exercise we will measure a chemical equilibrium constant using key proton
More informationUsing the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
More informationCary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V.
1234 Hach Hall 515-294-5805 www.cif.iastate.edu Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V. Location: Contact: 1240 Hach Hall Steve Veysey, 1234 Hach Hall Safety All researchers working
More informationUV-Visible Spectroscopy
UV-Visible Spectroscopy UV-Visible Spectroscopy What is UV-Visible Spectroscopy? Molecular spectroscopy that involves study of the interaction of Ultra violet (UV)-Visible radiation with molecules What
More informationOrganic Spectroscopy. UV - Ultraviolet-Visible Spectroscopy. !! 200-800 nm. Methods for structure determination of organic compounds:
Organic Spectroscopy Methods for structure determination of organic compounds: X-ray rystallography rystall structures Mass spectroscopy Molecular formula -----------------------------------------------------------------------------
More informationApplication Note AN4
TAKING INVENTIVE STEPS IN INFRARED. MINIATURE INFRARED GAS SENSORS GOLD SERIES UK Patent App. No. 2372099A USA Patent App. No. 09/783,711 World Patents Pending INFRARED SPECTROSCOPY Application Note AN4
More informationDetermining the Structure of an Organic Compound
Determining the Structure of an Organic Compound The analysis of the outcome of a reaction requires that we know the full structure of the products as well as the reactants In the 19 th and early 20 th
More informationBlackbody Radiation References INTRODUCTION
Blackbody Radiation References 1) R.A. Serway, R.J. Beichner: Physics for Scientists and Engineers with Modern Physics, 5 th Edition, Vol. 2, Ch.40, Saunders College Publishing (A Division of Harcourt
More informationLecture 1: Basic Concepts on Absorption and Fluorescence
Lecture 1: Basic Concepts on Absorption and Fluorescence Nicholas G. James Cell and Molecular Biology University of Hawaii at Manoa, Honolulu The Goal The emission of light after absorption of an outside
More informationMeasuring Protein Concentration through Absorption Spectrophotometry
Measuring Protein Concentration through Absorption Spectrophotometry In this lab exercise you will learn how to homogenize a tissue to extract the protein, and then how to use a protein assay reagent to
More informationTreasure Hunt. Lecture 2 How does Light Interact with the Environment? EMR Principles and Properties. EMR and Remote Sensing
Lecture 2 How does Light Interact with the Environment? Treasure Hunt Find and scan all 11 QR codes Choose one to watch / read in detail Post the key points as a reaction to http://www.scoop.it/t/env202-502-w2
More informationQuantitation of Peptides and Amino Acids with a Synergy HT using UV Fluorescence
Quantitation of Peptides and Amino Acids with a Synergy HT using UV Fluorescence Introduction Eukaryotic and prokaryotic cells contain a number of compounds that are fluorescent with UV light excitation.
More informationBack to Basics Fundamentals of Polymer Analysis
Back to Basics Fundamentals of Polymer Analysis Using Infrared & Raman Spectroscopy Molecular Spectroscopy in the Polymer Manufacturing Process Process NIR NIR Production Receiving Shipping QC R&D Routine
More informationNMR and other Instrumental Techniques in Chemistry and the proposed National Curriculum.
NMR and other Instrumental Techniques in Chemistry and the proposed National Curriculum. Dr. John Jackowski Chair of Science, Head of Chemistry Scotch College Melbourne john.jackowski@scotch.vic.edu.au
More information3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
More informationRaman Spectroscopy Basics
Raman Spectroscopy Basics Introduction Raman spectroscopy is a spectroscopic technique based on inelastic scattering of monochromatic light, usually from a laser source. Inelastic scattering means that
More informationNuclear Magnetic Resonance Spectroscopy
Nuclear Magnetic Resonance Spectroscopy Introduction NMR is the most powerful tool available for organic structure determination. It is used to study a wide variety of nuclei: 1 H 13 C 15 N 19 F 31 P 2
More informationBeer's Law: Colorimetry of Copper(II) Solutions
Exercise 11 Page 1 Illinois Central College CHEMISTRY 130 Name: Beer's Law: Colorimetry of Copper(II) Solutions Objectives In this experiment, we will use Beer's Law to determine the unknown concentrations
More informationAbsorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40
Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Note: there is a second document that goes with this one! 2046 - Absorbance Spectrophotometry - Calibration Curve Procedure. The second document
More informationColorimetry Extinction coefficient (ε) Lambda max (λ max ) Qualitative vs. quantitative analysis
Lab Week 2 - Spectrophotometry Purpose: Introduce students to the use of spectrophotometry for qualitative (what is it) and quantitative (how much is there of it) analysis of biological samples and molecules.
More informationMolar Mass of Polyvinyl Alcohol by Viscosity
Molar Mass of Polyvinyl Alcohol by Viscosity Introduction Polyvinyl Alcohol (PVOH) is a linear polymer (i. e., it has little branching) of Ethanol monomer units: -CH 2 -CHOH- Unlike most high molar mass
More informationImproved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs
Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs TracePro Opto-Mechanical Design Software s Fluorescence Property Utility TracePro s Fluorescence Property
More informationChemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
More informationThe Measurement of Sensitivity in Fluorescence Spectroscopy
The Measurement of Sensitivity in Fluorescence Spectroscopy Among instrumental techniques, fluorescence spectroscopy is recognized as one of the more sensitive. In fluorescence, the intensity of the emission
More informationINTERNATIONAL OLIVE COUNCIL
INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 19/Rev. 3 February 2015 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: iooc@internationaloliveoil.org
More informationA More Efficient Way to De-shelve 137 Ba +
A More Efficient Way to De-shelve 137 Ba + Abstract: Andrea Katz Trinity University UW REU 2010 In order to increase the efficiency and reliability of de-shelving barium ions, an infrared laser beam was
More informationDETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY
DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY Eduard Gilli 1,2 and Robert Schennach 1, 2 1 Graz University of Technology, 8010 Graz, Austria 2 CD-Laboratory for Surface Chemical and Physical
More informationph: Measurement and Uses
ph: Measurement and Uses One of the most important properties of aqueous solutions is the concentration of hydrogen ion. The concentration of H + (or H 3 O + ) affects the solubility of inorganic and organic
More informationOrganic Spectroscopy
1 Organic Spectroscopy Second Year, Michaelmas term, 8 lectures: Dr TDW Claridge & Prof BG Davis Lectures 1 4 highlight the importance of spectroscopic methods in the structural elucidation of organic
More informationExperiment 2 Kinetics II Concentration-Time Relationships and Activation Energy
2-1 Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy Introduction: The kinetics of a decomposition reaction involving hydroxide ion and crystal violet, an organic dye used
More informationPUMPED Nd:YAG LASER. Last Revision: August 21, 2007
PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow
More informationOLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
More informationUnit title: Chemical Laboratory Techniques
Unit title: Chemical Laboratory Techniques Unit code: H/601/0355 QCF level: 4 Credit value: 15 Aim This unit gives learners the opportunity to practise and become proficient in a range of practical skills
More informationChemistry 2351: Inorganic Chemistry I Laboratory Manual
Spectroscopic Determination of a Complex Ion's Stoichiometry by Job's Method ABSTRACT This experiment is adapted from Angelici's classic experiment, but uses Fe(H 2 O) 6-n (SCN) n 3-n as the complex ion.
More informationOptical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus
Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,
More information