MRI Artifacts: Mechanism and Control

Size: px
Start display at page:

Download "MRI Artifacts: Mechanism and Control"

Transcription

1 MRI Artifacts: Mechanism and Control Chun Ruan Abstract A wide variety of artifacts is routinely encountered on MR images. This article presents the cause, appearance, diagnostic effect, and available remedies for the artifacts that are most frequently observed on MR images and are of greatest clinical significance. Combined with routine preventive maintenance of imaging equipment, consistent quality control, and appropriate selection of imaging parameters, awareness of the manifestations of these artifacts will allow image quality and diagnostic interpretation to be optimized. I. Introduction Magnetic resonance imaging (MRI) is widely used in medical diagnosis for its various advantageous features, such as high-resolution capability, the ability to produce an arbitrary anatomic cross-sectional image, and high tissue contrast. Unfortunately, there are many potential sources of image artifacts associated with the technology of MRI. They can potentially degrade images sufficiently to cause inaccurate diagnosis. Many MR artifacts are neither obvious nor understandable from previous experience with conventional types of imaging. While some MR artifacts are machine specific, the majority are inherent in the imaging method itself. MR imaging artifacts can be grouped into two general categories. First, there are artifacts that are hardware related. These artifacts are relatively uncommon fortunately, because they are often difficult to diagnose and usually require service personnel to correct. The second category consists of artifacts related to the patient or under operator control. This category is encountered much more commonly and may often be easily prevented or corrected once they are recognized. II. Technique and Methods A. Motion Artifacts Motion is the most prevalent source of MR imaging artifacts. As the name implies, motion artifacts are caused by motion of the imaged object or a part of the imaged object during the imaging sequence. Motion results in two effects on MR images. View-to-view effects are caused by motion that occurs between the acquisitions of successive phase-encoding steps. The inconsistent location and signal intensity of spins that move as phase-encoding data are acquired result in phase errors. When motion is periodic that is, occurs in a regular pattern the result is complete or incomplete replication of the moving tissue, commonly referred to as ghosting artifacts. These artifacts are observed along the phase-encoding direction of the image, regardless of the direction in which the motion actually occurred. Periodic physiologic motions that commonly result in ghosting artifacts include cardiac motion, respiratory motion, vascular pulsation, and cerebrospinal fluid (CSF) pulsation. 1 Motion occurring between the time of radiofrequency (RF) excitation and echo collection (i.e., within-view) results in a lack of coherent phase among the population 1

2 of moving spins at the time of echo formation. 2 This incoherence manifests as blurring and increased image noise. Unlike phase errors encountered in view-to-view motion effects, this within-view effect is expressed throughout the image. It is most frequently associated with random motion, as can occur with gastrointestinal peristalsis, swallowing, coughing, eye motion, and gross patient movement Respiratory motion Respiratory motion results in ghosting artifacts and blurring that can obscure or simulate lesions. A variety of methods have been used to reduce the effect of respiratory motion artifacts. Mechanical methods, such as use of an abdominal or thoracic binder or taking images with the patient in a prone position, are intended to restrict the amplitude of respiratory motion. However, these maneuvers often produce patient discomfort and may therefore have counterproductive effects; Signal averaging is the use of multiple data acquisitions to improve the signal-to-noise ratio (SNR) of the image ( Fig.1). In this process, the prominence of ghosting artifacts is reduced by approximately the square root of the number of signal averages obtained. Of course, imaging time increases linearly with the number of signal averages; With respiratory triggering, data are collected only during a limited portion of the respiratory cycle, usually near end-expiration, when respiratory movement is minimal. The major drawback of this technique is its marked prolongation of imaging time, because so much of the time is not being used productively for data acquisition; Respiratory ordered phase encoding involves monitoring the patient s respiratory cycle during imaging using a bellows device. Unlike respiratory triggering, however, with this method data are collected in a continuous fashion. This method does not restrict the operator s selection of TR, and it does not significantly extend imaging time; Gradient moment nulling involves the application of additional gradient pulses to correct for phase shifts among a population of moving protons at the time of echo collection. This method corrects for constant-velocity motion and helps reduce the signal loss and ghosting associated with such movement. Use of gradient moment nulling requires a nominal TE, which may preclude its use on T1-weighted pulse sequences. This method does not prolong image acquisition time; Another method to reducing the signal intensity of moving tissue is the use of fat suppression methods. With this method, not only subcutaneous fat but also mediastinal, mesenteric, retroperitoneal, and other stores of internal fat are suppressed and are thus less capable of generating ghosting artifacts. 2

3 2. Cardiac motion Figure 1. The image on the left exhibits respiratory motion as blurring of the structures as well as motion-induced ghosting. By using multiple averages motion can be reduced in the same way that multiple averages increase the signal to noise ratio. The image on the right was obtained with 16 averages. Cardiac motion produces a series of ghost artifacts along the phase-encoding direction of the image, in addition to blurring and signal loss of cardiac and juxtacardiac structures. 4 The major approach for reducing cardiac motion artifacts is electrocardiographic triggering, in which data collection is synchronized with cardiac phase ( Fig. 2). This synchronization enables cardiac tissue to be located in a consistent position as each successive phase-encoding step is acquired, resulting in increased tissue signal intensity and decreased phase errors. Other approaches include the use of fast imaging sequences that reduce the opportunity for motion during data acquisition, gradient moment nulling, and spatial RF presaturation pulses. 3. Vascular Pulsation Figure 2. The image on the left was acquired without any form of motion compensation technique for cardiac motion. The image on the right was obtained using cardiac gating. Vascular pulsation artifacts are recognized by their alignment with the responsible vessel along the phase-encoding direction of the image. These artifacts reproduce the cross-sectional size and shape of the responsible vessel, but not necessarily its signal intensity. Spatial RF presaturation pulses applied outside the field of view help reduce the signal intensity of inflowing blood and, hence reduce the resultant pulsation artifact. Other practical methods for reducing the prominence of vascular pulsation artifacts include positioning the section of interest in the middle of a multisection acquisition, thus reducing any potential entry phenomenon, and maximizing the saturation of flowing spins. B. Susceptibility Artifacts Susceptibility artifacts occur as the result of microscopic gradients or variations in the magnetic field strength that occurs near the interfaces of substance of different magnetic susceptibility ( Fig. 3). Large susceptibility artifacts are commonly seen surrounding ferromagnetic objects inside of diamagnetic materials (such as the human body). These gradients cause dephasing of spins and frequency shifts of the surrounding tissues. The net results are bright and dark areas with spatial distortion of surrounding anatomy. These artifacts are worst with long echo times and with 3

4 gradient echo sequences. Susceptibility artifacts can be made less prominent by performing imaging at low magnetic field strength, using smaller voxels, decreasing echo time, and increasing receiver bandwidth. Gradient-echo and echo-planar sequences should be avoided, because they accentuate susceptibility artifacts. The use of spin-echo and particularly fast spin-echo sequences should be considered. C. Chemical Shift Artifacts Figure 3. An axial MRI of the head in a patient with mascara on her eyelids. Susceptibility artifacts from the mascara obscure the front half of the globes. A chemical shift artifact is caused by the difference in chemical shift (Larmor frequency) of fat and water. The artifact manifests itself as a misregistration between the fat and water pixels in an image ( Fig. 4). The effect being that fat and water spins in the same voxel are encoded as being located in different voxels. The magnitude of the effect is proportional on the magnitude of the Bo field and inversely proportional to the sampling rate in the frequency encoding direction. For a constant sampling rate, the larger Bo, the greater the effect. 5 Chemical shift artifacts are typically observed along the frequency-encoding direction but can also occur along the slice-selection direction of the image. Chemical shift artifact can be reduced by performing imaging at low magnetic field strength, by increasing receiver bandwidth, or by decreasing voxel size. The artifacts tend to be more prominent on T2-weighted than on T1-weighted images. Fat suppression methods often eliminate visible artifacts, and gradient reorientation can redirect chemical shift artifacts to another portion of the image. 4

5 D. Wrap Around Artifacts Figure 4. This artifact is shown in an axial image of a kidney where the bright border along the top of the kidney and the dark border along the bottom of the kidney represent the artifact. A wrap around artifact is the occurrence of a part of the imaged anatomy, which is located outside of the field of view, inside of the field of view ( Fig. 5). This artifact is caused by the selected field of view being smaller than the size of the imaged object. Or more specifically the digitization rate is less than the range of frequencies in the FID or echo. 6 The solution to a wrap around artifact is to choose a larger field of view, adjust the position of the image center, or select an imaging coil which will not excite or detect spins from tissues outside of the desired field of view. Figure 5. The first image shows wrap-around of the back of the head on to the front of the head, where the phase-encoded direction is anterior-posterior. The second image has the phase and frequency directions reversed resulting in absence of the aliasing artifact. Oversampling was used in the frequency direction to eliminate the aliasing. E. Partial Volume Artifacts A partial volume artifact is any artifact which is caused by the size of the image voxel. It occurs when multiple tissue types are encompassed within a single voxel. For example, if a small voxel contains only fat or water signal, and a larger voxel might contain a combination of the two, the large voxel possess a signal intensity equal to the weighted average of the quantity of water and fat present in the voxel. Volume averaging is most likely to occur in the slice-selection direction of the image, which has the largest voxel dimension. It also occurs when structures are oriented obliquely to the imaging plane and when structures move in and out of a given section during image acquisition. Volume averaging can simulate abnormalities, decrease the visualization of low-contrast abnormalities, and blur or distort affected structures. Partial volumeaveraging is usually recognized by careful analysis of adjacent images. Decreasing voxel size, particularly reducing section thickness, can be useful if further confirmation is required ( Fig.6). Three-dimensional Fourier transform imaging is particularly useful, because it provides thin sections with no intervening gaps and is conductive to reformatting in alternate imaging planes; Simple acquisition of additional two-dimensional images in alternate imaging planes is very helpful for resolving issues relating to partial volume averaging. Retrospective reformatting of two-dimensional data can also be performed using an interpolation algorithm and can be of further assistance. 5

6 Figur e 6. These two axial T1-weighted images of the head were obtained at exactly the same location, yet the second image shows the VII and VIII cranial nerves while the first does not. The reason for the vanishing nerve is explained by partial volume averaging. The first slice was obtained with a thickness of 10 mm while the second was at a thickness of 3 mm. F. Gibbs Ringing Artifacts Gibbs ringing artifacts are bright or dark lines that are seen parallel and adjacent to borders of abrupt intensity change. The ringing is caused by incomplete digitization of the echo. This means the signal has not decayed to zero by the end of the acquisition window, and the echo is not fully digitized. This artifact is seen in images when a small acquisition matrix is used. Solutions include use of a higher resolution imaging matrix and filtration methods ( Fig.7). Gradient reorientation will displace the artifacts to another portion of the image. Figure 7. The fine lines visible in the image on the left are due to undersampling of the high spatial frequencies. This results in a "ringing" type of artifact following these borders in the phase direction (R to L in this image). This problem can be easily fixed by taking more samples such as the image on the right with 256 phase encodes. G. Zebra Stripes Zebra stripes can be observed along the periphery of gradient-echo images where there is an abrupt transition in magnetization at the air-tissue interface. They are accentuated by aliasing that results from the use of a relatively small field of view. Solutions include expanding the field of view, using spin-echo pulse sequences, or using oversampling techniques to reduce aliasing. H. Slice-overlap Artifacts 6

7 The slice-overlap artifact is the loss of signal seen in an image from a multi-angle, multi-slice acquisition, as is obtained commonly in the lumbar spine. If the slices obtained at different disk spaces are not parallel, then the slices may overlap. If two levels are done at the same time, e.g., L4-5 and L5-S1, then the level acquired second will include spins that have already been saturated. This causes a band of signal loss crossing horizontally in image, usually worst posteriorly. Therefore, overlap of sections within areas of diagnostic interest should be carefully avoided. I. RF Overflow Artifacts RF overflow artifacts cause a non-uniform, washed-out appearance to an image. This artifact occurs when the signal received by the scanner from the patient is too intense to be accurately digitized by the analog-to-digital converter. Autoprescanning usually adjusts the receiver gain to prevent this from occurring but if the artifact still occurs, the receiver gain can be decreased manually. J. Entry Slice Phenomenon Entry slice phenomenon occurs when unsaturated spins in blood first enter into a slice or slices. It is characterized by bright signal in a blood vessel (artery or vein) at the first slice that the vessel enters. Usually the signal is seen on more than one slice, fading with distance. This artifact has been confused with thrombosis with disastrous results. The characteristic location and if necessary, the use of gradient echo flow techniques can be used to differentiate entry slice artifacts from occlusions. K. Zipper Artifacts There are various causes for zipper artifacts in images. Most of them are related to hardware or software problems beyond the physicist immediate control. The zipper artifacts that can be controlled easily are those due to RF entering the scanning room when the door is open during acquisition of images. 7 RF from some radio transmitters will cause zipper artifacts that are oriented perpendicular to the frequency axis of image. Broad-band noise degrades the entire image, whereas narrow frequency noise produces linear bands that transverse the phase-encoding direction of the image. Solutions include identifying and removing external RF sources, ensuring that the door to the imaging room remains closed, and verifying the integrity of the magnet room enclosure and associated seals. L. Cross-Excitation Cross-excitation is caused by the imperfect shape of RF slice profiles, which leads to the unintended excitation of adjacent tissue. This excitation results in the saturation of such tissue, manifest as decreased signal intensity and decreased contrast that can hinder lesion detection. One way to avoid this artifact is to introduce an intersection gap that is 10% to 50% of the prescribed section thickness. Another method is interleaved image acquisition, in which odd-numbered sections are initially acquired, followed by acquisition of even-numbered sections. 8 Also, optimized RF pulses that have a more rectangular slice profile can be implemented. 7

8 Figure 8.The dark bands visible on the T1-weighted Spin Echo axial image are due to the intersection of other axial images through the image. These are acquired in a multi-planar fashion and thus cause pre-excitation (saturation) of the protons in the area where the slices intersect. The sagittal image illustrates where the slices were obtained and how they intersect posteriorly. M. Shading Shading artifacts manifest as foci of relatively reduced signal intensity involving a portion of the image. Abnormalities contained in the shaded portion of the MR image may be obscured. 9 There are many potential causes for this artifact, including partial volume averaging malfunction of the RF transmitter, amplifier, or receiver, excessive RF absorption, etc. 10, 11 To minimize shading artifacts, the anatomy of interest should be centered within the magnet, within the coil, and within the group of sections to be acquired. III. Conclusions Artifacts are common in magnetic resonance imaging. Most occur as a result of interactions of multiple factors, especially motion. MR artifacts primarily cause image degradation, although they can occasionally mimic pathological lesions. Some MR artifacts, such as those caused by periodic respiratory or vascular motion, may obscure important diagnostic findings. These and other MR artifacts may be minimized in some instances by proper selection of the directions of the phase and frequency encoding gradients. The ability to select and vary the direction of these gradients is a useful option in MR imaging. Investigations in the origins of MRI artifacts can not only lead to further understanding of the imaging process itself, but can also improve the quality and diagnostic accuracy of MR examination. 1 Saloner D, Flow and motion, Magn Reson Imaging Clin N Am 1999 Nov;7(4): Barish MA and Jara H, Motion artifact control in body MR imaging, Magn Reson Imaging Clin N Am 1999 May;7(2): Hedley M, Yan H, Motion artifact suppression: a review of post-processing techniques, Magn Reson Imaging 1992;10(4): Huber ME, Hengesbach D and Botnar RM, Motion artifact reduction and vessel enhancement for free-breathing navigator-gated coronary MRA using 3D k-space reordering, Magn Reson Med 2001 Apr;45(4): Altbach MI, Trouard TP and Van de Walle R, Chemical-shift imaging utilizing the positional shifts along the readout gradient direction, IEEE Trans Med Imaging 2001 Nov;20(11): Tsai CM and Nishimura DG, Reduced aliasing artifacts using variable-density k-space sampling trajectories, Magn Reson Med 2000 Mar;43(3): Mugler JP 3rd, Overview of MR imaging pulse sequences, Magn Reson Imaging Clin N Am 1999 Nov;7(4):

9 8 Clark JA 2nd and Kelly WM, Common artifacts encountered in magnetic resonance imaging, Radiol Clin North Am 1988 Sep;26(5): Jones RW and Witte RJ, Signal intensity artifacts in clinical MR imaging, Radiographics 2000 May- Jun;20(3): Mirowitz SA, MR imaging artifacts. Challenges and solutions, Magn Reson Imaging Clin N Am 1999 Nov;7(4): Mirowitz SA, MR imaging artifacts. Challenges and solutions, Magn Reson Imaging Clin N Am 1999 Nov;7(4):

5 Factors Affecting the Signal-to-Noise Ratio

5 Factors Affecting the Signal-to-Noise Ratio 5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

MRI SEQUENCES. 1 Gradient Echo Sequence

MRI SEQUENCES. 1 Gradient Echo Sequence 5 An MRI sequence is an ordered combination of RF and gradient pulses designed to acquire the data to form the image. In this chapter I will describe the basic gradient echo, spin echo and inversion recovery

More information

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI Douglas C. Noll* and Walter Schneider Departments of *Radiology, *Electrical Engineering, and Psychology University

More information

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R GE 3.0T 3.0T WRIST Invivo 8CH Wrist Coil Sequence Ax T2 Cor PD Cor PDFS Cor T1 Cor PD (Small FOV) FOV (mm) 80 80 80 80 40 Matrix 384x224 384x256 320x256 384x320 320x192 Phase Direction RL RL RL RL RL #

More information

Musculoskeletal MRI Technical Considerations

Musculoskeletal MRI Technical Considerations Musculoskeletal MRI Technical Considerations Garry E. Gold, M.D. Professor of Radiology, Bioengineering and Orthopaedic Surgery Stanford University Outline Joint Structure Image Contrast Protocols: 3.0T

More information

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications Safety Issue 3.0 T MRI Pro 3.0 T MRI Con Immediate fringe field surrounding magnet A ferromagnetic object inadvertently brought into the scan room will experience a sharp increase in attraction toward

More information

MRI Physics for Radiologists

MRI Physics for Radiologists Alfred L. Horowitz MRI Physics for Radiologists A Visual Approach Second Edition With 94 Illustrations Springer-Verlag New York Berlin Heidelberg London Paris Tokyo Hong Kong Barcelona Budapest Alfred

More information

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi Diffusione e perfusione in risonanza magnetica E. Pagani, M. Filippi DW-MRI DIFFUSION-WEIGHTED MRI Principles Diffusion results from a microspic random motion known as Brownian motion THE RANDOM WALK How

More information

GE Medical Systems Training in Partnership. Module 12: Spin Echo

GE Medical Systems Training in Partnership. Module 12: Spin Echo Module : Spin Echo Spin Echo Objectives Review the SE PSD. Review the concepts of T, T, and T*. Spin Echo PSD RF Gz Gy 90 80 Gx Spin Echo - SE Spin echo is a standard pulse sequence on Signa MRi/LX and

More information

Overview. Optimizing MRI Protocols. Image Contrast. Morphology & Physiology. User Selectable Parameters. Tissue Parameters

Overview. Optimizing MRI Protocols. Image Contrast. Morphology & Physiology. User Selectable Parameters. Tissue Parameters Overview Optimizing MRI Protocols Clinical Practice & Compromises Geoffrey D. Clarke, Radiology Department University of Texas Health Science Center at San Antonio Pulse timing parameters for adjusting

More information

Basic Principles of Magnetic Resonance

Basic Principles of Magnetic Resonance Basic Principles of Magnetic Resonance Contents: Jorge Jovicich jovicich@mit.edu I) Historical Background II) An MR experiment - Overview - Can we scan the subject? - The subject goes into the magnet -

More information

The MRI Study Guide for Technologists

The MRI Study Guide for Technologists The MRI Study Guide for Technologists Kenneth S. Meacham The MRI Study Guide for Technologists With 51 Illustrations Springer-Verlag New York Berlin Heidelberg London Paris Tokyo Hong Kong Barcelona Budapest

More information

SITE IMAGING MANUAL ACRIN 6698

SITE IMAGING MANUAL ACRIN 6698 SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012

More information

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine G16.4426/EL5823/BE6203 Medical Imaging MRI Instrumentation, Data Acquisition, Image Reconstruction Riccardo Lattanzi, Ph.D. Assistant Professor Department of Radiology, NYU School of Medicine Department

More information

7/16/2010. Pulse Sequences and Acquisition Techniques for Breast MRI. Objectives. ACR Breast MRI Accreditation Program Launched May 2010

7/16/2010. Pulse Sequences and Acquisition Techniques for Breast MRI. Objectives. ACR Breast MRI Accreditation Program Launched May 2010 Pulse Sequences and Acquisition Techniques for Breast MRI ACR Breast MRI Accreditation Program Launched May 2010 Ron Price Vanderbilt University Medical Center Nashville, TN 37232 Information available:

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging What are the uses of MRI? To begin, not only are there a variety of scanning methodologies available, but there are also a variety of MRI methodologies available which provide

More information

MRI for Paediatric Surgeons

MRI for Paediatric Surgeons MRI for Paediatric Surgeons Starship David Perry Paediatric Radiologist Starship Children s Hospital CHILDREN S HEALTH What determines the brightness of a pixel in MRI? i.e. What determines the strength

More information

32-Channel Head Coil Imaging at 3T

32-Channel Head Coil Imaging at 3T 32-Channel Head Coil Imaging at 3T Thomas Benner Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA

More information

Phantom Test Guidance

Phantom Test Guidance Phantom Test Guidance for the ACR MRI Accreditation Program The American College of Radiology 1891 Preston White Dr Reston, VA 20191-4397 www.acr.org Contents 0.0 INTRODUCTION............................................................................

More information

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter

More information

Lecture 14. Point Spread Function (PSF)

Lecture 14. Point Spread Function (PSF) Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect

More information

Kap 8 Image quality, signal, contrast and noise

Kap 8 Image quality, signal, contrast and noise 4/5/ FYS-KJM 474 contrast SNR MR-teori og medisinsk diagnostikk Kap 8 Image qualit, signal, contrast and noise resolution vailable MRparameters speed Main source of noise in MRI: Noise generated within

More information

Super-Resolution Reconstruction in MRI: Better Images Faster?

Super-Resolution Reconstruction in MRI: Better Images Faster? Super-Resolution Reconstruction in MRI: Better Images Faster? Esben Plenge 1,, Dirk H. J. Poot 1, Monique Bernsen 2, Gyula Kotek 2, Gavin Houston 2, Piotr Wielopolski 2, Louise van der Weerd 3, Wiro J.

More information

MGH Adult Diffusion Data Scanning Protocols

MGH Adult Diffusion Data Scanning Protocols MGH Adult Diffusion Data Scanning Protocols Structural scans SIEMENS MAGNETOM ConnectomA syngo MR D11 \\USER\INVESTIGATORS\Default\AAHScout_64 TA:0:14 PAT:3 Voxel size:1.6 1.6 1.6 mm Rel. SNR:1.00 :fl

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research. ParaVision 6 The Next Generation of MR Acquisition and Processing for Preclinical and Material Research Innovation with Integrity Preclinical MRI A new standard in Preclinical Imaging ParaVision sets a

More information

Introduction to MRI. Basics of MRI. Welcome to the exciting world of MRI! This is a tutorial that will cover the basic aspects of MRI.

Introduction to MRI. Basics of MRI. Welcome to the exciting world of MRI! This is a tutorial that will cover the basic aspects of MRI. Introduction to MRI Welcome to the exciting world of MRI! This is a tutorial that will cover the basic aspects of MRI. Copyright 1994-1998 Ray Ballinger, All Rights Reserved. Send questions and comments

More information

Advanced MRI methods in diagnostics of spinal cord pathology

Advanced MRI methods in diagnostics of spinal cord pathology Advanced MRI methods in diagnostics of spinal cord pathology Stanisław Kwieciński Department of Magnetic Resonance MR IMAGING LAB MRI /MRS IN BIOMEDICAL RESEARCH ON HUMANS AND ANIMAL MODELS IN VIVO Equipment:

More information

Unwarping Echo Planar Images Using CMTK 1 Release 1.2

Unwarping Echo Planar Images Using CMTK 1 Release 1.2 Unwarping Echo Planar Images Using CMTK 1 Release 1.2 Torsten Rohlfing October 1, 2012 Neuroscience Program, SRI International, Menlo Park, CA Abstract This document describes the workflow for unwarping

More information

Twelve. Figure 12.1: 3D Curved MPR Viewer Window

Twelve. Figure 12.1: 3D Curved MPR Viewer Window Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from

More information

High Spatial Resolution EPI Using an Odd Number of Interleaves

High Spatial Resolution EPI Using an Odd Number of Interleaves Magnetic Resonance in Medicine 41:1199 1205 (1999) High Spatial Resolution EPI Using an Odd Number of Interleaves Michael H. Buonocore* and David C. Zhu Ghost artifacts in echoplanar imaging (EPI) arise

More information

runl I IUI%I/\L Magnetic Resonance Imaging

runl I IUI%I/\L Magnetic Resonance Imaging runl I IUI%I/\L Magnetic Resonance Imaging SECOND EDITION Scott A. HuetteS Brain Imaging and Analysis Center, Duke University Allen W. Song Brain Imaging and Analysis Center, Duke University Gregory McCarthy

More information

Principles of functional Magnetic Resonance Imaging

Principles of functional Magnetic Resonance Imaging 1 Principles of functional Magnetic Resonance Imaging Martin A. Lindquist Department of Biostatistics; Johns Hopkins University Tor D. Wager Department of Psychology & Neuroscience; University of Colorado

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

ENG4BF3 Medical Image Processing. Image Visualization

ENG4BF3 Medical Image Processing. Image Visualization ENG4BF3 Medical Image Processing Image Visualization Visualization Methods Visualization of medical images is for the determination of the quantitative information about the properties of anatomic tissues

More information

3/30/2013. Disclosure. Advanced Neuro MRI: Imaging Techniques and Protocol Optimization. MRI, 35 year ago. MRI Today. Outlines

3/30/2013. Disclosure. Advanced Neuro MRI: Imaging Techniques and Protocol Optimization. MRI, 35 year ago. MRI Today. Outlines http://www.magnet.fsu.edu Disclosure Advanced Neuro MRI: Imaging Techniques and Protocol Optimization Research funding provided by Siemens Healthcare. Chen Lin, PhD DABR Indiana University School of Medicine

More information

Characterization of small renal lesions: Problem solving with MRI Gary Israel, MD

Characterization of small renal lesions: Problem solving with MRI Gary Israel, MD Characterization of small renal lesions: Problem solving with MRI Gary Israel, MD With the widespread use of cross-sectional imaging, many renal masses are incidentally found. These need to be accurately

More information

Nuclear Magnetic Resonance

Nuclear Magnetic Resonance Nuclear Magnetic Resonance Practical Course M I. Physikalisches Institut Universität zu Köln May 15, 2014 Abstract Nuclear magnetic resonance (NMR) techniques are widely used in physics, chemistry, and

More information

REVIEW. Magnetic Resonance: An Introduction to Ultrashort TE (UTE) Imaging

REVIEW. Magnetic Resonance: An Introduction to Ultrashort TE (UTE) Imaging REVIEW Magnetic Resonance: An Introduction to Ultrashort TE (UTE) Imaging Matthew D. Robson, PhD, Peter D. Gatehouse, DPhil, Mark Bydder, PhD, and Graeme M. Bydder, MB, ChB Abstract: The background underpinning

More information

The File-Card-Browser View for Breast DCE-MRI Data

The File-Card-Browser View for Breast DCE-MRI Data The File-Card-Browser View for Breast DCE-MRI Data Sylvia Glaßer 1, Kathrin Scheil 1, Uta Preim 2, Bernhard Preim 1 1 Department of Simulation and Graphics, University of Magdeburg 2 Department of Radiology,

More information

Proton magnetic resonance spectroscopy in the brain: Report of AAPM MR Task Group #9

Proton magnetic resonance spectroscopy in the brain: Report of AAPM MR Task Group #9 Proton magnetic resonance spectroscopy in the brain: Report of AAPM MR Task Group #9 Dick J. Drost a) Nuclear Medicine and MRI Department, St. Joseph s Health Centre, London, Ontario N6A 4L6, Canada William

More information

Generation and Detection of NMR Signals

Generation and Detection of NMR Signals Generation and Detection of NMR Signals Hanudatta S. Atreya NMR Research Centre Indian Institute of Science NMR Spectroscopy Spin (I)=1/2h B 0 Energy 0 = B 0 Classical picture (B 0 ) Quantum Mechanical

More information

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami

More information

RF Coils... They ve Come a Long, Long Way

RF Coils... They ve Come a Long, Long Way GE Healthcare MR Field Notes The sensitivity area relates to the diameter of the coil. By Increasing coil size, the area of sensitivity also increases. But as with most MR princicloser a coil s position

More information

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents Preface Symbols and fundamental constants 1. What is spectroscopy? A semiclassical description of spectroscopy Damped harmonics Quantum oscillators The spectroscopic experiment Ensembles and coherence

More information

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference.

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference. Pulsed Fourier Transform NR The rotating frame of reference The NR Eperiment. The Rotating Frame of Reference. When we perform a NR eperiment we disturb the equilibrium state of the sstem and then monitor

More information

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager Technique and Safety of performing EEG/fMRI measurements by Pierluigi Castellone, Electronics Engineer Brain Products General Manager Contents of the presentation Why recording simultaneous EEG and fmri?

More information

7/22/2011. Breast MRI: Pulse Sequences, Acquisition Protocols, and Analysis. Objectives. Challenges in DCE Breast Imaging

7/22/2011. Breast MRI: Pulse Sequences, Acquisition Protocols, and Analysis. Objectives. Challenges in DCE Breast Imaging Breast MRI: Pulse Sequences, Acquisition Protocols, and Analysis Objectives Ron Price Vanderbilt University Medical Center Nashville, TN 37232 1. Review background of MRI breast cancer imaging 2. Present

More information

Physiological Basis of the BOLD Signal. Kerstin Preuschoff Social and Neural systems Lab University of Zurich

Physiological Basis of the BOLD Signal. Kerstin Preuschoff Social and Neural systems Lab University of Zurich Physiological Basis of the BOLD Signal Kerstin Preuschoff Social and Neural systems Lab University of Zurich Source: Arthurs & Boniface, 2002 From Stimulus to Bold Overview Physics of BOLD signal - Magnetic

More information

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim- QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound

More information

NMR Techniques Applied to Mineral Oil, Water, and Ethanol

NMR Techniques Applied to Mineral Oil, Water, and Ethanol NMR Techniques Applied to Mineral Oil, Water, and Ethanol L. Bianchini and L. Coffey Physics Department, Brandeis University, MA, 02453 (Dated: February 24, 2010) Using a TeachSpin PS1-A pulsed NMR device,

More information

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Excelart Vantage Magnet Configuration: Ultra-short-bore Strength (or W x H): 1.5 T Homogeneity, ppm V-RMS: Dimensions of maximum

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Material (ESI) for Physical Chemistry Chemical Physics. This journal is the Owner Societies 2016 Electronic Supplementary Information Achieving High Resolution and Controlling

More information

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT GUIDE TO SETTING UP AN MRI RESEARCH PROJECT Formal requirements and procedures OVERVIEW This document is intended to help a principle investigator set up a research project using magnetic resonance imaging

More information

FDA Guidelines for Magnetic Resonance Equipment Safety

FDA Guidelines for Magnetic Resonance Equipment Safety FDA Guidelines for Magnetic Resonance Equipment Safety Loren A. Zaremba, Ph.D. Center for Devices and Radiological Health Food and Drug Administration Outline I. Introduction II. Static Magnetic Field

More information

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS Selection and use of Ultrasonic Ceramic Transducers The purpose of this application note is to aid the user in the selection and application of the Ultrasonic

More information

Whitepaper. Image stabilization improving camera usability

Whitepaper. Image stabilization improving camera usability Whitepaper Image stabilization improving camera usability Table of contents 1. Introduction 3 2. Vibration Impact on Video Output 3 3. Image Stabilization Techniques 3 3.1 Optical Image Stabilization 3

More information

RICHARD T ELL ASSOCIATES,INC.

RICHARD T ELL ASSOCIATES,INC. RICHARD T ELL ASSOCIATES,INC. An Evaluation of Test Measurement Data Obtained on the KW-Gard RF Protective Suit March 30, 1998 Prepared for Euclid Garment Manufacturing Company 333 Martinel Drive Kent,

More information

MRI Accreditation Program Clinical Image Quality Guide

MRI Accreditation Program Clinical Image Quality Guide MRI Accreditation Program Clinical Image Quality Guide Introduction Read this entire document carefully. This document provides guidance for interpreting physicians and technologists on the criteria evaluated

More information

510(k) Summary. This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.

510(k) Summary. This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807. MAR, 2009 / \MLECH MRI-TECH Sp. z o.o., ul. Zielihska 3, 31-227 Krak6w, Poland MRI-TECH Canada, Inc. 206-3820 Cambie Street Vancouver BC V5Z 2X7 510(k) Summary This summary of 510(k) safety and effectiveness

More information

Interference. Physics 102 Workshop #3. General Instructions

Interference. Physics 102 Workshop #3. General Instructions Interference Physics 102 Workshop #3 Name: Lab Partner(s): Instructor: Time of Workshop: General Instructions Workshop exercises are to be carried out in groups of three. One report per group is due by

More information

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information

Imaging Heart Motion Using Harmonic Phase MRI

Imaging Heart Motion Using Harmonic Phase MRI 186 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 19, NO. 3, MARCH 2000 Imaging Heart Motion Using Harmonic Phase MRI Nael F. Osman, Elliot R. McVeigh, and Jerry L. Prince* Abstract This paper describes a

More information

In Practice Whole Body MR for Visualizing Metastatic Prostate Cancer

In Practice Whole Body MR for Visualizing Metastatic Prostate Cancer In Practice Whole Body MR for Visualizing Metastatic Prostate Cancer Prostate cancer is the second most common cancer in men worldwide, accounting for 15% of all new cancer cases. 1 Great strides have

More information

The Magnetic Resonance Imaging Examination

The Magnetic Resonance Imaging Examination The Magnetic Resonance Imaging Examination The purpose of The American Registry of Radiologic Technologists (ARRT ) Examination in Magnetic Resonance Imaging (MRI) is to assess the knowledge and cognitive

More information

Diffusion-Weighted MR Imaging of the Liver at 3.0 Tesla Using Tracking Only Navigator Echo (TRON): a Feasibility Study

Diffusion-Weighted MR Imaging of the Liver at 3.0 Tesla Using Tracking Only Navigator Echo (TRON): a Feasibility Study JOURNAL OF MAGNETIC RESONANCE IMAGING 30:1027 1033 (2009) Original Research Diffusion-Weighted MR Imaging of the Liver at 3.0 Tesla Using Tracking Only Navigator Echo (TRON): a Feasibility Study Marko

More information

13C NMR Spectroscopy

13C NMR Spectroscopy 13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

Quantitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs

Quantitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Quantitative Comparison of Conventional and Oblique MRI for Detection of Herniated Spinal Discs Doug Dean ENGN 2500: Medical Image Analysis Final Project Outline Introduction to the problem Based on paper:

More information

Single-scan longitudinal relaxation measurements in high-resolution NMR spectroscopy

Single-scan longitudinal relaxation measurements in high-resolution NMR spectroscopy Journal of Magnetic Resonance 164 (2003) 321 328 www.elsevier.com/locate/jmr Single-scan longitudinal relaxation measurements in high-resolution NMR spectroscopy Nikolaus M. Loening, a, * Michael J. Thrippleton,

More information

LONI De-Identification Policy

LONI De-Identification Policy The following defines how different file formats are de-identified with LONI tools. Each metadata attribute of each file can have the following operations performed: Operation keep Remove Replace Description

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals

More information

Magnetic Resonance Imaging Level 1

Magnetic Resonance Imaging Level 1 MIAP1 Revised July 2015 MEDICAL IMAGING ADVISORY PANEL 1 Course Syllabus Magnetic Resonance Imaging Level 1 Page 1 of 13 MIAP1 MRI Course Syllabus Guide Under Revision 07 July 2015 Please refer to Medical

More information

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture

More information

ELECTRON SPIN RESONANCE Last Revised: July 2007

ELECTRON SPIN RESONANCE Last Revised: July 2007 QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron

More information

Pulsed Nuclear Magnetic Resonance An Experiment for UCSB s Advanced Laboratory

Pulsed Nuclear Magnetic Resonance An Experiment for UCSB s Advanced Laboratory 1 Pulsed Nuclear Magnetic Resonance An Experiment for UCSB s Advanced Laboratory Foreword Pulsed nuclear magnetic resonance is fascinating in its own right, and is also an incredibly important tool for

More information

DSAM Digital Quality Index (DQI) A New Technique for Assessing Downstream Digital Services

DSAM Digital Quality Index (DQI) A New Technique for Assessing Downstream Digital Services Application Note DSAM Digital Quality Index (DQI) A New Technique for Assessing Downstream Digital Services Overview As cable operators move to digital simulcast and all digital networks, the majority

More information

2. MATERIALS AND METHODS

2. MATERIALS AND METHODS Difficulties of T1 brain MRI segmentation techniques M S. Atkins *a, K. Siu a, B. Law a, J. Orchard a, W. Rosenbaum a a School of Computing Science, Simon Fraser University ABSTRACT This paper looks at

More information

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a BASIC PRINCIPLES INTRODUCTION TO NUCLEAR MAGNETIC RESONANCE (NMR) 1. The nuclei of certain atoms with odd atomic number, and/or odd mass behave as spinning charges. The nucleus is the center of positive

More information

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D. 4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology jiang.steve@mgh.harvard.edu http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?

More information

DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA.

DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA. DEVELOPMENT OF AN IMAGING SYSTEM FOR THE CHARACTERIZATION OF THE THORACIC AORTA. Juan Antonio Martínez Mera Centro Singular de Investigación en Tecnoloxías da Información Universidade de Santiago de Compostela

More information

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: kturek@mri-tech.pl

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: kturek@mri-tech.pl Cirrus 0.2T MRI for Everyone North America, Asia, Europe contact: kturek@mri-tech.pl MRI-TECH inc. Cirrus MRI system for all your needs: Low costs Low maintenance High quality Open geometry Imaging of

More information

MRI SCANNERS: A BUYER S GUIDE

MRI SCANNERS: A BUYER S GUIDE 09 MRI SCANNERS: A BUYER S GUIDE MRI SCANNERS: A BUYER S GUIDE D. Price, I. Delakis, C. Renaud and R. Dickinson Correspondence Centre for Evidence-based Purchasing 152C Skipton House 80 London Road London

More information

Accelerating MRI Data Acquisition Using Parallel Imaging and Compressed Sensing

Accelerating MRI Data Acquisition Using Parallel Imaging and Compressed Sensing University of Wisconsin Milwaukee UWM Digital Commons Theses and Dissertations December 2012 Accelerating MRI Data Acquisition Using Parallel Imaging and Compressed Sensing Haifeng Wang University of Wisconsin-Milwaukee

More information

Understanding astigmatism Spring 2003

Understanding astigmatism Spring 2003 MAS450/854 Understanding astigmatism Spring 2003 March 9th 2003 Introduction Spherical lens with no astigmatism Crossed cylindrical lenses with astigmatism Horizontal focus Vertical focus Plane of sharpest

More information

Boom Influence on Yagi Antenna Dragoslav Dobričić, YU1AW (Serbia) dragan@antennex.com

Boom Influence on Yagi Antenna Dragoslav Dobričić, YU1AW (Serbia) dragan@antennex.com Boom Influence on Yagi Antenna Dragoslav Dobričić, YU1AW (Serbia) dragan@antennex.com Introduction T he boom of the Yagi antenna is an inevitable part of its construction. Theoretically, Yagi antennas

More information

Uses of Derivative Spectroscopy

Uses of Derivative Spectroscopy Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative

More information

Ultrasonic Wave Propagation Review

Ultrasonic Wave Propagation Review Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic

More information

Head Position and Impact Direction in Whiplash Injuries: Associations with MRI-Verified Lesions of Ligaments and Membranes in the Upper Cervical Spine

Head Position and Impact Direction in Whiplash Injuries: Associations with MRI-Verified Lesions of Ligaments and Membranes in the Upper Cervical Spine Head Position and Impact Direction in Whiplash Injuries: Associations with MRI-Verified Lesions of Ligaments and Membranes in the Upper Cervical Spine 1 Journal of Neurotrauma Volume 22, Number 11, November

More information

REVIEW. Magnetic Resonance in Medicine 00:00 00 (2014)

REVIEW. Magnetic Resonance in Medicine 00:00 00 (2014) REVIEW Magnetic Resonance in Medicine 00:00 00 (2014) Recommended Implementation of Arterial Spin-Labeled Perfusion MRI for Clinical Applications: A Consensus of the ISMRM Perfusion Study Group and the

More information

C1 Medical Imaging Modalities & Characteristics. 4005-759 Linwei Wang

C1 Medical Imaging Modalities & Characteristics. 4005-759 Linwei Wang C1 Medical Imaging Modalities & Characteristics 4005-759 Linwei Wang Major Types of Medical Imaging Modalities X-ray Imaging Computed Tomography (CT) Magnetic Resonance Imaging (MRI) Nuclear Imaging Positron

More information

BARE PCB INSPECTION BY MEAN OF ECT TECHNIQUE WITH SPIN-VALVE GMR SENSOR

BARE PCB INSPECTION BY MEAN OF ECT TECHNIQUE WITH SPIN-VALVE GMR SENSOR BARE PCB INSPECTION BY MEAN OF ECT TECHNIQUE WITH SPIN-VALVE GMR SENSOR K. Chomsuwan 1, S. Yamada 1, M. Iwahara 1, H. Wakiwaka 2, T. Taniguchi 3, and S. Shoji 4 1 Kanazawa University, Kanazawa, Japan;

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes

More information

Nuclear Magnetic Resonance Spectroscopy

Nuclear Magnetic Resonance Spectroscopy Nuclear Magnetic Resonance Spectroscopy Introduction NMR is the most powerful tool available for organic structure determination. It is used to study a wide variety of nuclei: 1 H 13 C 15 N 19 F 31 P 2

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging North American Spine Society Public Education Series What Is Magnetic Resonance Imaging (MRI)? Magnetic resonance imaging (MRI) is a valuable diagnostic study that has been used

More information

Clinical applications of MRI in radiation therapy. Jatta Berberat, PhD Kantonsspital Aarau jatta.berberat@ksa.ch

Clinical applications of MRI in radiation therapy. Jatta Berberat, PhD Kantonsspital Aarau jatta.berberat@ksa.ch Clinical applications of MRI in radiation therapy Jatta Berberat, PhD Kantonsspital Aarau jatta.berberat@ksa.ch Background and introduction Magnetic Resonance Imaging Relaxation mechanisms Imaging gradients

More information

Maximizing Receiver Dynamic Range for Spectrum Monitoring

Maximizing Receiver Dynamic Range for Spectrum Monitoring Home Maximizing Receiver Dynamic Range for Spectrum Monitoring Brian Avenell, National Instruments Corp., Austin, TX October 15, 2012 As consumers continue to demand more data wirelessly through mobile

More information