The accurate calibration of all detectors is crucial for the subsequent data



Similar documents
Calorimetry in particle physics experiments

Electron-Muon Ranger (EMR)

Measurement of Muon Lifetime and Mass Using Cosmic Ray Showers

CMS Tracking Performance Results from early LHC Running

Tutorial 4.6 Gamma Spectrum Analysis

Electron-Muon Ranger (EMR)

AMS DAQ, TRD DSP Software

Copyright 2007 Casa Software Ltd. ToF Mass Calibration

ATLAS Test Beam Analysis in Stockholm: An Overview

Energy Deposition in MICE Absorbers and Coils

Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission

Introduction to Geiger Counters

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Experiment 10. Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado

07 - Cherenkov and transition radiation detectors

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Information about the T9 beam line and experimental facilities

Evaluation Tools for the Performance of a NESTOR Test Detector

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)

The OPERA Emulsions. Jan Lenkeit. Hamburg Student Seminar, 12 June Institut für Experimentalphysik Forschungsgruppe Neutrinophysik

Physics 111 Homework Solutions Week #9 - Tuesday

13C NMR Spectroscopy

Interstellar Cosmic-Ray Spectrum from Gamma Rays and Synchrotron

Cross section, Flux, Luminosity, Scattering Rates

Module 13 : Measurements on Fiber Optic Systems

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Functional Data Analysis of MALDI TOF Protein Spectra

How To Use An Ionsonic Microscope

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

MSc in Autonomous Robotics Engineering University of York

AP Physics 1 and 2 Lab Investigations

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

A Guide to Detectors Particle Physics Masterclass. M. van Dijk

High Energy Physics. Lecture 4 More kinematics and a picture show of particle collisions

Solid State Detectors = Semi-Conductor based Detectors

Electrical Resonance

Testing thermo-acoustic sound generation in water with proton and laser beams

Clock Jitter Definitions and Measurement Methods

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1

MICE detectors and first results. M. Bonesini Sezione INFN Milano Bicocca

ISTITUTO NAZIONALE DI FISICA NUCLEARE

Application Note Noise Frequently Asked Questions

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

POSSIBL-E EXPERIMENTS ON THE 200-GeV ACCELERATOR. A. D. Krisch University of Michigan. R. Serber Columbia University.

The data acquisition system of the XMASS experiment

EXPERIMENTAL ERROR AND DATA ANALYSIS

Does Quantum Mechanics Make Sense? Size

T(CR)3IC Testbed for Coherent Radio Cherenkov Radiation from Cosmic-Ray Induced Cascades

x 1 ' = x 1 vt 1 x 1 ' = 4.0 m t 1 = 1.0 s x 2 vt 2 ' = 4.0 m t 2 ' = x 2 = 3.0 s x 1 = x 2 x 1 ' + vt 1 ' + vt 2 v (t 1 t 2 ) = x 2 ' x 1 ' = x 2

Industry Environment and Concepts for Forecasting 1

π 0 γγ calibration Savrina Daria, ITEP, Moscow SINP MSU, Moscow Victor Egorychev, ITEP, Moscow Ivan Belyaev

Symmetric Stretch: allows molecule to move through space

Synthetic Sensing: Proximity / Distance Sensors

AS COMPETITION PAPER 2008

ORTEC. Fast-Timing Discriminator Introduction. Choosing the Right Timing Discriminator for the Application. Counting

Uses of Derivative Spectroscopy

ENERGY LOSS OF ALPHA PARTICLES IN GASES

THE STATISTICAL TREATMENT OF EXPERIMENTAL DATA 1

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION

CHAPTER 6 PHOTON COUNTING

E190Q Lecture 5 Autonomous Robot Navigation

Introduction to X-Ray Powder Diffraction Data Analysis

Charged Particle in a Magnetic Field

Timing Errors and Jitter

World-first Proton Pencil Beam Scanning System with FDA Clearance

THE DØ Luminosity Monitor (LM) consists of two array of

Cathode Ray Tube. Introduction. Functional principle

Chapter 6: The Information Function 129. CHAPTER 7 Test Calibration

electron does not become part of the compound; one electron goes in but two electrons come out.

Modern Construction Materials Prof. Ravindra Gettu Department of Civil Engineering Indian Institute of Technology, Madras

How To Understand Light And Color

The interaction of Cu(100)-Fe surfaces with oxygen studied with photoelectron spectroscopy. I

The Effect of Network Cabling on Bit Error Rate Performance. By Paul Kish NORDX/CDT

AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE

Three-dimensional figure showing the operation of the CRT. The dotted line shows the path traversed by an example electron.

An option for the SHiP Muon Detector: Scintillator bars with WLS fibers and SiPMs readout

Extended Resolution TOA Measurement in an IFM Receiver

Power Electronics. Prof. K. Gopakumar. Centre for Electronics Design and Technology. Indian Institute of Science, Bangalore.

Experiment 5. Lasers and laser mode structure

KNOWLEDGE BASE ARTICLE Zero Voltage Transmission (ZVT) Technology. Basics of the GPZ 7000 Technology: Zero Voltage Transmission (ZVT)

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Experiment #1, Analyze Data using Excel, Calculator and Graphs.

8.2 Elastic Strain Energy

arxiv:astro-ph/ v1 15 Sep 2005

Agilent AN 1315 Optimizing RF and Microwave Spectrum Analyzer Dynamic Range. Application Note

Objectives. Electric Current

INTERFERENCE OF SOUND WAVES

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference.

Ionosphere Properties and Behaviors - Part 2 By Marcel H. De Canck, ON5AU

Grazing incidence wavefront sensing and verification of X-ray optics performance

1. Units of a magnetic field might be: A. C m/s B. C s/m C. C/kg D. kg/c s E. N/C m ans: D

A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS

Transcription:

Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved in the measurements has to be monitored. In this chapter the calibration of TAPS, Crystal Barrel and the Tagger are presented in detail. 4.1 TAPS The TAPS detector needs four kinds of calibrations, i.e. pulse-shape calibration, energy calibration, time calibration and trigger threshold calibrations. These calibrations are roughly adjusted using cosmic-ray data obtained before each data taking period. All these calibrations are discussed in the following sections. 4.1.1 Energy calibration The TAPS energy calibration is done in three stages. First, before data taking, a rough calibration is performed with cosmic muons providing a relative gain adjustment. During the data taking the energy calibration is monitored and the π 0 peak position is used to further improve the energy calibration of all detectors of TAPS. As a last check, the η invariant-mass peak is analyzed for determining the energy dependence of the calibration. 43

CHAPTER 4. CALIBRATION 44 Figure 4.1: Example of a cosmic-ray energy spectrum for a single TAPS module. Calibration based on cosmic muons At the beginning of each data taking period cosmic-ray data are taken with the TAPS detector. Since all the TAPS modules have the same shape and orientation, the average energy deposition by minimum-ionizing particles (cosmic muons) is always the same i.e. 38.5 MeV. An example of a raw cosmic-ray spectrum for one of the TAPS detector modules is shown in figure 4.1. In this figure one can clearly identify the pedestal produced by a pulser (see section 3.2.4), the CFD threshold and the minimum ionizing peak. Using the pedestal value, which correspond to E γ = 0 MeV, and the minimum-ionizing peak, which correspond to E γ = 38.5 MeV, the energy offset and gains are calculated using the following formula: gain = 38.5 MeV cosmic peak [ch nr] pedestal [ch nr] (4.1) and offset = gain pedestal [ch nr] (4.2) This calibration is used to monitor the TAPS detector during the measurement. This is done for the narrow- and wide-gate integration.

45 4.1. TAPS Calibration based on the π 0 invariant-mass peak Even though the cosmic calibration is good enough to monitor the detector behavior online, for the actual data analysis a more precise energy calibration is needed, which is obtained by using the π 0 invariant-mass peak position. Because of the decay π 0 γγ, the events with at least two neutral hits in the TAPS detector are selected. The 2γ invariant-mass spectrum is produced for all possible pairs of neutral hits. The invariant mass is given by: M γ1 γ 2 = 2E γ1 E γ2 (1 cosθ 1,2 ) (4.3) where, E γ1 and E γ2 are the measured energies of photon 1 and 2, respectively, and θ 1,2 is the opening angle between them. From this invariant-mass spectrum an overall correction factor for the gain of all detectors was determined that shifts the neutral pion peak to the correct mass of 138.98 MeV. For a more precise fine tuning, an invariant mass for each possible pair of detectors is calculated and that mass is entered into two histograms, corresponding to two detectors of two neutral hits. The exact position of the π 0 peak position is determined by plotting the invariant-mass squared spectrum and fitting the resulting peak with a Gaussian function. Since the pedestals are adjusted accurately during the cosmic calibration and are continuously monitored, only the gain has newly been determined for each detector module. This procedure is repeated iteratively until the 2γ invariant-mass peak converges to the π 0 mass for each of the 528 detector modules. It takes about 3 to 4 iterations. The final result is depicted in figure 4.2 (left) showing the 2γ invariant-mass squared spectrum and (right) the alignment of the different detectors. This calibration is repeated for each day of data taking, in order to control the effect induced by day-night temperature variations and long-term energy shifts. The final result for overall π 0 peak position is shown in figure 4.3 with a resolution of 8 MeV (σ). Calibration based on η π 0 mass After the calibration using the π 0 peak position for each individual detector module is performed, a second calibration point is used for extra fine tuning and correcting possible non-linearities. This point is the invariant mass of the η γγ decay. The η peak is observed consistently 1.5% higher than the true η mass would allow. This effect occurs due to leakage on to detectors outside the cluster

CHAPTER 4. CALIBRATION 46 Figure 4.2: The 2γ invariant-mass spectrum for TAPS after the π 0 calibration. Left: the 2γ invariant-mass squared spectrum for one of the TAPS detector modules fitted by a Gaussian function plus a polynomial. The line shown is the correct value for π 0 -mass squared. Right : The π 0 peak position in individual TAPS detector modules. Figure 4.3: The 2γ invariant-mass spectrum for TAPS after the π 0 calibration. Left: The overall π 0 spectrum (all detector modules together). The correct π 0 mass is shown by a line at 134.9 MeV. Right: The η peak position is at higher mass (dashed curve) than the correct value (vertical line) after the π 0 calibration. After the energy dependent correction the η peak shows the correct mass (solid curve).

47 4.1. TAPS Beam time b c Jan 2003 0.99482-1.7567e-5 Feb 2003 1.01482-1.0200e-5 Jul 2003 1.00071-3.4097e-5 Aug 2003 1.01008-4.1680e-5 Table 4.1: Energy dependent correction factors for different data taking periods. as stated by [59] and [15]. This non-linearity in the peak position is corrected by measuring the position of two invariant-mass peaks (π 0 and η) and determining the new values for the energy of the photons using the following equation: where, E new = a + b E old + c E 2 old (4.4) c = 1 547.3 M η 134.98 547.3 (4.5) b = 1 (134.98 c) (4.6) M η is the η peak position in the 2γ invariant-mass spectrum. The parameter a = 0 because the zero-point of the energy calibration function is determined by the pedestal and may not move. The parameters b and c are fixed by using the π 0 γγ and η γγ invariant-mass peaks and the formula for the invariant mass of two particles (equation 4.3). These correction factors (b and c) have been determined for each data taking period and are shown in table 4.1. 4.1.2 Pulse-shape calibration The BaF 2 scintillators of the TAPS detector have the intrinsic property that the shape of the pulse differs for protons and photons. This feature allows to distinguish between these two particles. The signals from photons and protons have two scintillation light components, narrow and wide. These two components are extracted by integrating the signal twice, one for a narrow time window E short (50 ns) and another for a wide time window E long (2 µs). The integration for the narrow time window contains the information about the fast light component and that for the wide time window contains the information about both the fast and slow components. For protons and photons the ratio of fast and slow component is markedly different. The pulse shape is usually plotted in terms of

CHAPTER 4. CALIBRATION 48 Figure 4.4: The pulse-shape spectrum in polar coordinates after calibration. The line shows the cut used to differentiate between photons and protons. the variables R P S and θ P S, defined in polar coordinates as: R P S = E 2 long + E2 short (4.7) θ P S = arctan( E long E short ) (4.8) To identify the protons and photons from pulse-shape spectra, the photon band has to be separated from the proton band. As a first step the cosmic calibrations are used for both wide-gate energy (E long ) and narrow-gate energy (E short ). As E long measures the complete energy and E short measures only part of the energy, a calibration for E short is needed. The narrow-gate energy has to be adjusted such that for photons θ P S = 45. From the above equations, this condition will be true if E long and E short are equal. The pulse-shape spectrum is produced for each detector module. If the photon band deviates from the desired θ P S = 45 line at low values of R P S, the narrow-gate pedestal is adjusted and at high values for R P S, the narrow-gate gain is corrected. The procedure is repeated until the position of the photon band converges to 45. The pulse-shape spectrum after calibration is shown in figure 4.4. The photon band around 45

49 4.1. TAPS Figure 4.5: The projection of figure 4.4 on the x-axis for different values of R P S. The line at θ P S = 45 shows the photon peak position. The line left side of 45 is the cut applied to separate protons from the photons.

CHAPTER 4. CALIBRATION 50 and proton band left to the photon band are clearly visible. To define a separation between protons and photons, the y-axis in figure 4.4 was divided in several R P S bins. Each bin was then projected onto the x-axis. The projections are shown in figure 4.5. In all these graphs the photon peak is at θ P S = 45 and the proton peak is on the left-side of the photon peak. The photon peak was fitted by a Gaussian function. The cut that separates photons from protons is placed at 3σ away from the mean of the photon peak. The cut is shown by the line next to the line at 45 in figure 4.5. The corresponding cuts applied for different values of R P S are also shown in figure 4.4 by a smooth line. This procedure has been followed for each of the 528 detector modules separately. 4.1.3 Time calibration The time calibration for the TAPS detector involves the determination of gain and offset for TDC s. A series of pulses of different but known frequencies has been fed to the TDC s. This gives a pattern of sharp peaks in each time spectrum from which the gain is calculated. To determine offsets, the events with two or more neutral hits have been selected by using the cut on π 0 mass (for π 0 calibration refer sec. 4.1.1). The time difference between these two hits has been analyzed. Corresponding spectra have been made for all the possible combinations for each detector module. The offset is adjusted so that the peak in these spectra aligns at zero, which means that photons arrive at these detector modules at the same time. The new constants for the offset have been determined for each day of data taking. Figure 4.6 (left) shows the time difference between two photons after the calibration. The peak is fitted by a Gaussian function. A width (F W HM) of 0.65 ns is observed, while the mean is located at 0 ns. Figure 4.6 (right) shows the alignment of each detector module after the calibration. About 10% of the detectors are found with bad time information. These detectors can not be calibrated because they have either very broad time spectra or more than one peak in the spectra. Such detectors are marked as bad in a defect table. If any of these detectors occurs at the center of the cluster, its TDC information is ignored. Instead, the TDC information from its neighbouring detector (which is a part of the cluster) has been used. If this neighbouring detector is also marked as bad, another neighbouring detector has been checked for a valid TDC information. This check has been carried out until a good TDC is found in a cluster. During the measurement there is the possibility that the coincident signals might cross the threshold at different times if their amplitudes differ. Because

51 4.1. TAPS Figure 4.6: The difference in the timing signals between two photons in TAPS, after calibration. Left: All detectors together. Right: For individual detector modules. Beam time a b c Jan 2003-0.1593 0.11160e-2-0.11082e-5 Feb 2003-0.10249 0.07818e-2-0.63905e-6 Jul 2003-0.48668 0.88268e-3-0.75603e-6 Aug 2003-0.44120 0.69529e-3-0.50863e-6 Table 4.2: Time-walk correction factors for different data taking periods. of this effect a small deviation in the time is seen for photons of low energies; this is called the time-walk. To take care of this time-walk a correction function is determined from histograms of the photon-photon time in TAPS versus the photon energy. These histograms are fitted with the energy-dependent formula: E start = b 2c (4.9) t corr = a b2 4c fore E start (4.10) t corr = a + be + ce 2 fore E start (4.11) This calibration is performed as an overall correction for all modules and is done for every data taking period separately. In table 4.2, the time-walk correction factors are given for different data taking periods.

CHAPTER 4. CALIBRATION 52 Figure 4.7: TAPS LED threshold determination. Left: Example of an energy spectrum without any conditions is superimposed by the energy spectrum after the LED threshold has been applied. Right: The ratio of the two spectra from the left side of the graph. The vertical line shown in this picture is a cut used to determine the LED threshold. 4.2 LED threshold calibration The LED s are used to discriminate the events below a certain energy. They decide whether the event will be recorded or not. Therefore a proper LED calibration has to be done at the beginning of data taking. This calibration is based on cosmic rays. Three measurements were taken with three different threshold settings (mv) of LED s. These three points in combination with the cosmic calibration give the calibration function for the LED thresholds. It determines the correlation between the threshold voltage setting (mv) and the threshold energy (MeV) of the LED module. The LED-high and LED-low thresholds have been applied to each detector unit. During the offline analysis more precise LED thresholds have been determined. For this calibration, the energy spectrum without LED thresholds and with LED thresholds has been produced for each TAPS detector module. The ratio of both spectra indicates the threshold value (figure 4.7). The thresholds are determined using the point where the ratio = 0.9. This procedure is applied for both LED values, high and low. Since during the data taking only one type of LED s are used in the bit pattern, only one type of LED value is determined at one time. This procedure is repeated for each beam time, once to determine the LED-high and once for the LED-low thresholds for each detector module.

53 4.3. CRYSTAL BARREL Figure 4.8: The 2γ invariant-mass squared spectrum for one of the Crystal Barrel modules at different stpdf of the iteration procedure for gain adjustment. 4.3 Crystal Barrel The procedure to calibrate the Crystal Barrel is very similar to the procedure to calibrate TAPS. The main difference is that the cosmic calibration for the Crystal Barrel is not practical. This is because the Crystal Barrel modules have different sizes and orientations and the energy deposited in each module is not the same. Since QDC s of the Crystal Barrel have two ranges, two types of calibration have to be done. The calibration for the low-energy part has been done using the π 0 calibration method as discussed to calibrate TAPS modules. Because the pedestals are determined and subtracted online, only the gains need to be determined. For the calibration of the high-energy range a certain amount of laser light corresponding to different known energies has been injected into each crystal. Then the gains have been calculated for each crystal. Similar to TAPS, for the final analysis a more precise energy calibration has been done based on the π 0 invariant mass. The π 0 invariant mass has been studied for each crystal module. The π 0 peak position of a cluster of modules has been adjusted to 134.9 MeV by correcting the gain of the individual module. This procedure is repeated until the module reaches a stable value. The position of the π 0 peak after different iterations is shown in figure 4.8. In this picture the 2γ invariant-mass squared is plotted. More details on the Crystal Barrel calibration can be found in [60].

CHAPTER 4. CALIBRATION 54 4 (ns) Time Taps-Tagger 2 0-2 -4 0 60 120 180 240 300 360 420 480 Tagger fiber Figure 4.9: TAPS-Tagger time difference for all Tagger fibers. 4.4 Tagger The energy of the photons created by the bremsstrahlung process was measured by the tagger. The electron beam deflected by the dipole magnet hits scintillating fibers in the tagger. Two type of calibrations are necessary for the tagger. One is an energy calibration in which the photon energy corresponding to the position of each fiber is determined. The other is a time calibration. The time information from the tagger is very useful to reduce the random background. 4.4.1 Time Calibration The TDC s of each scintillating fiber are calibrated using the events with at least one neutral hit occurring in TAPS. Prior to this calibration the TAPS TDC s are calibrated. The time difference between each fiber and TAPS is plotted. The new offset for the fiber TDC is determined so that the prompt peak shifts to zero. In figure (4.9) the TAPS-Tagger time difference for each tagger fiber is shown. The time resolution for all fibers together is found to be 2 ns (FWHM), which is worse than the TAPS resolution.

4.4.2 Energy Calibration The energy of the photon is determined by tracking the position of the corresponding electron on the fiber. The radius of the circular path traveled by the electron is given by: ρ = p Bq (4.12) where, ρ = bending radius, p = electron momentum, B = magnetic field, q= charge. From this formula the radius is directly proportional to the momentum of the electron if the magnetic field and charge are constant. Since before the bremsstrahlung process all the electrons have the same energy, it is easy to calculate the energy of the photon. A fourth order polynomial function was calculated by using the measured field map of the bending magnet and the known positions of the fibers. The electron beam of 3.2 GeV is used to calculate the corresponding energy of each fiber. The check of this calibration was carried out by keeping the magnetic field constant and changing the electron beam energy. Such a procedure takes a lot of time, since it involves the fine tuning of the beam at the accelerator. Therefore the measurements have been taken only at four points for the magnetic field for a beam energy of 3.2 GeV. The four points are shown in figure 4.10. A discrepancy up to 30 MeV for lower photon beam energies and a few MeV for higher photon beam energies has been observed. So one more calibration procedure was carried out. It makes use of the polarization of the beam. The linearly polarized beam was obtained by replacing the amorphous radiator by a crystalline radiator. When the beam is properly aligned and the angle of the crystal is known, the exact energy of the coherent peak can be calculated [61]. The second order polynomial is calculated from the last two methods. This polynomial is applied to the fourth-order polynomial calculated from the first method. More details about the Tagger calibration can be found in [62].

CHAPTER 4. CALIBRATION 56 3 3 10 2.5 Energy value (Mev) 2 1.5 1 0.5 0 0 50 100 150 200 250 300 350 400 450 Fiber index Figure 4.10: Tagger energy correction.

57 4.4. TAGGER