Tips & Tricks GPC/SEC: Analysis and Calibration For Low Molar Mass Macromolecules



Similar documents
Chemical Structure of the polymer showing end groups: CH 3 CH 2 CH H

Tips & Tricks GPC/SEC: Select the Right Columns for Your Molar Mass Range

Pesticide Analysis by Mass Spectrometry

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Correlation of the Mass Spectrometric Analysis of Heat-Treated Glutaraldehyde Preparations to Their 235nm / 280 nm UV Absorbance Ratio

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

How To Use An Acquity Qda Detector

An Introduction to Gel Permeation Chromatography and Size Exclusion Chromatography PRIMER

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software

Separation of Amino Acids by Paper Chromatography

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Waters Core Chromatography Training (2 Days)

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

Size Exclusion Chromatography

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Monoclonal Antibody Fragment Separation and Characterization Using Size Exclusion Chromatography Coupled with Mass Spectrometry

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

MASCOT Search Results Interpretation

Smartline HPLC by design

consultancy services

How To Test For Contamination In Large Volume Water

Background Information

Quick Start Guide Chromeleon 7.2

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Size-exclusion Chromatography of Polymers. Bernd Trathnigg

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

for mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Comparison of the Reversed-Phase Selectivity of Solid Core HPLC Columns

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

The world's largest elearning website for analytical scientists

Selectivity and Detectability Optimizations in HPLC

AppsLab Library. A new online platform for Thermo Fisher Scientific applications and methods. Susanne Kramer, Ph.D. Product Manager AppsLab Library

Mass Spectrometry Signal Calibration for Protein Quantitation

Process Instrumentation

DO MORE AND DO IT WITH CONFIDENCE. Empower 3 Chromatography Data Software

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

Training Courses 2014 HPLC GC SPE

The Theory of HPLC. Gradient HPLC

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Analyzing Small Molecules by EI and GC-MS. July 2014

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA

Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary

EZ-Run Protein Gel Solution. EZ-Run Protein Standards. EZ-Run Gel Staining Solution. Traditional SDS-Page Reagents. Protein Electrophoresis

Separation of Peptides from Enzymatic Digestion on Different Acclaim Columns: A Comparative Study

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.

Application Note. Separation of three monoclonal antibody variants using MCSGP. Summary

Chapter 21 Principles of Chromatography and Mass Spectrometry Problems 1,2,3,4,8,9,11,13

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

Size Exclusion Chromatography

LC-MS/MS for Chromatographers

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: )

Aspects of industrial purification of peptides using large-scale chromatography. Lars Andersson and Jonas Persson

Laboratory procedures in polymer chemistry

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites

molecules ISSN

DiscovIR TM Deposition and Detection System

Diffusion and Fluid Flow

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

A Complete Solution for Method Linearity in HPLC and UHPLC

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Lab 2 Biochemistry. Learning Objectives. Introduction. Lipid Structure and Role in Food. The lab has the following learning objectives.

Turning HTE Agilent LC/MS Data to Knowledge in 1hr

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

PURIFICATION. Preparative Chromatography Modular Solutions from Manual Systems to Full Automation. Scale Up. Reverse. Phase. Impurity Isolation

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Carbohydrate Analysis: Column Chemistries and Detection

EXAMINATION IN COURSE KJ 2053 CHROMATOGRAPHY

designed to never let you down PerkinElmer Series 200 HPLC Systems

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

HPLC Basics, Equipment, and Troubleshooting

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS

IFC. A new perspective i l ifi ti. p p g p. Single softw FractPAL software plug-in for Agilent ChemStation. Medicinal Chemistry.

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Transcription:

GPC/SEC: Analysis and Calibration For Low Molar Mass Macromolecules Daniela Held, Silvia Fugmann, and Peter Kilz, PSS Polymer Standards Service GmbH, Mainz, Germany. Macromolecules can range from several hundred daltons to several million daltons in molar mass. There are many applications where only products with a low molecular weight are used. Gel permeation chromatography/size exclusion chromatography (GPC/SEC) analysis can be easily optimized to yield improved results when analysing oligomers and low molar mass macromolecules. Column selection and system calibration are key to successful characterization. New detection techniques also allow an additional insight. Column Selection Oligomers and low molar mass macromolecules should be analysed on dedicated columns or column sets. Only small pores are required for the analysis of small molecules, therefore columns filled with a low porosity material are best for this kind of analysis. Shear degradation is not an issue that is encountered and so columns for this molar mass region often consist of stationary phase materials of smaller particle sizes. This, therefore, results in a better resolution so that often the resolution power of a dedicated column bank is high enough to resolve oligomeric samples into single peaks. Figure 1 shows an overlay of an oligomeric polymethyl methacrylate (PMMA, Mw =, PDI = 1.) on a dedicated column set for oligomers; and on a column set for medium molar masses with 5-µm and 10-µm particle sizes, respectively. It is clear that a low molar mass sample can look very different depending on the resolution of the column in this molar mass range. The combination for low molar masses clearly shows several single maxima, whereas the 5-µm medium molar mass combination shows only some shoulders, and the 10-µm medium molar mass combination shows only an unimodal peak. Calibration and Peak Identification When a good separation is achieved with the low molar mass column set (as shown in Figure 1), it is possible to identify the single oligomers and add them as additional calibrants to the calibration curve. A true gel permeation chromatography/size exclusion chromatography (GPC/SEC) molar mass calibration standard should come with molar mass averages (M n and M w ) at least, in addition to the molar mass at the peak maximum (M p ). The Mp is the value that should be entered into the calibration curve. 1 In the case of the PMMA oligomer (as shown in Figures 1 and ) the M p is 0 Da. The PMMA has been synthesized using group transfer polymerization (GTP), resulting in a PMMA with two hydrogen end groups. As the molar mass of the repetition unit MMA (methyl methacrylate) is 100 Da, the molar masses for the other peaks can be easily determined. They increase from the highest peak at 0 Da stepwise with 100 for every peak with lower elution volume; and decrease stepwise by 100 for every peak with higher elution volume. Figure illustrates the PMMA oligomer with the single peaks labelled. Table 1 summarizes the peak information and shows the molar mass that 1 0

Figure 1: Overlay of the chromatograms of an oligomeric polymethyl methacrylate (PMMA) on a column set for low molar masses (green) and column sets for medium molar masses with 5 µm particle size (blue) and 10 µm particle size (black). VISCOTEK GPC/SEC: SETTING THE STANDARD With 30 years of chromatography and light scattering experience, the Viscotek GPC/SEC systems from Malvern provide solutions GPSBMMBQQMJDBUJPOTJOUIFQSPUFJOBOEQPMZNFSÞFMET CHARACTERIZE: Polymers, Proteins, Polysaccharides DETERMINE: Molecular Weight, Size, Structure DETECTORS: Light Scattering, Viscometer, Refractive Index, UV PDA, DLS The Viscotek range includes detectors, software and complete systems, BMMCBDLFECZUIFHMPCBM.BMWFSOTVQQPSUBOEUSBJOJOHOFUXPSL can be entered into the calibration curve with the corresponding elution volume. The resulting PMMA calibration curve is shown in Figure 3. It consists of several entries from higher molar masses (unimodal peaks, not shown) and several entries of the oligomeric PMMA, which will positively influence the results of the sample analysis. This is because with an increased number of calibration points, the precision of the calibration fit function increases. Additional results without even looking at the mass distribution are available for the analysis of unknown PMMA samples. This is shown in Figure, where the PMMA oligomer from above is overlaid with degree of polymerizations 3 and 4, respectively. This overlay allows direct identification of the degree of polymerization, as well as products from backbiting, a typical side reaction in anionic polymerization of MMA. Backbiting by-products elute between two degrees of polymerization. Both products, P3 Material relationships For the latest GPC/SEC applications information Click Here To learn more visit: www.malvern.com/settingthestandard 10 1 0

Figure : Overlay of the oligomeric polymethyl methacrylate (PMMA) with peak labelling (compare Table 1) and two samples, P3 and P4. and P4, show backbiting peaks for P3 (compare peak B), and P4 respectively. In addition, P3, P5, and P can be identified in sample P4 because they co-elute with the peaks C, E, and F of the reference material. Absolute Molar Masses Although the terms absolute or true molar masses are often associated with light scattering, viscometry, or triple detection, these are often not the best solutions for low molar mass samples. Especially for samples with a low refractive index increment (dn/dc), detection is impossible or only possible when extremely high (untypical) concentrations are injected. This often results in a loss of resolution and, even worse, of distribution information. If absolute molar masses are required and matching calibration standards are not available, two scenarios can be used. Case Study 1: A separation with resolution into single peaks is possible. The resulting picture is similar to the green curve in Figure 1. If the molar masses of initiator, repetition unit, and end group 11 1 0

Table 1: Molar mass, elution volume, and area for the resolved peaks of the oligomeric polymethyl methacrylate (PMMA) sample. Peak Identifier Molar Mass (Da)* are available, a conventional calibration curve can be established as described above. With this calibration curve true molar mass averages of unknown samples can be determined. Case Study : If separation into single peaks is not achievable, or if additional information is required, a hyphenation of GPC/SEC with mass spectrometry (MS) can help. MS has been proven to be a powerful method for macromolecules with lower molar masses. On-line hyphenation of MS with GPC/SEC allows calibration curves to be constructed that can yield true molar masses for oligomers (hompolymers and copolymers alike). For every elution volume the matching mass spectrum is Degree of Polymerization Elution Volume (ml)* Area (%) A 0 30.7 0.3 B ** Backbiting 3 30.1 1.40 C 30 3.7 4.40 D 40 4..74 E 50 5.41 1.3 F Reference from Certificate: 0 7.1 15. G 70 7 7.4.33 H 0 7.13 1.5 I 0.0. J 100 10.5 10.1 K 110 11.7.34 available. If the molar mass of the repetition unit and the ionization agent are known, a direct and automatic determination of the molar mass at the elution volume is possible. 3 Therefore a matching calibration curve can be constructed without using any calibration standards. This allows the determination of true molar masses. In addition, the MS detector can identify side products and end groups. Summary The analysis of oligomeric samples with GPC/SEC has high optimization potential. The availability of new stationary phases with small particle sizes and hyphenation with modern detection move easily move reliably move quickly move ahead α-, β-, γ-amino Acids Peptides Ensure faster discovery and faster separations with our growing selection of immobilised chiral stationary phases. move ahead with confidence with ZWIX Zwitterionic Stationary Phases CHIRALPAK ZWIX(+) CHIRALPAK ZWIX(-) Chiral Analysis of Underivatised Amino Acids Reversal of Elution Order Compatible with MS Detection Chiral Technologies Europe cte@chiral.fr www.chiral.fr Tel: +33 (0)3 7 5 00 1 1 0

Figure 3: Resulting calibration curve with multiple entries from the calibration sample polymethyl methacrylate (PMMA)-Mp 0 and other molar mass reference materials. Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials Using the Advanced Vial Closure System (AVCS) ON-DEMAND WEBCAST Register Free at /avcs EVENT OVERVIEW: Just one more turn that has been the chromatographers answer to the age old question of How tight should I seal my autosampler chromatography vial closure? Unfortunately, for a lot of chromatographers, this answer has resulted in a myriad of chromatographic problems. Truth be told, just one more turn is more than likely the absolute worst thing that one can do to seal an autosampler vial. Key Learning Objectives: The impact of analyst subjectivity on vial sealing Optimal vial/closure sealing and how it can be reliably achieved The Advance Vial Closure System design techniques offers new pathways for investigation and the determination of true molar masses. Reference 1. D. Held, The Column 4(), 1 (00).. T. Gruendling, M. Guilhaus, and C. Barner-Kowollik, Macromolecules 4, 3 374 (00). 3. D. Held, The Column 7(), 11 (011). Daniela Held studied polymer chemistry in Mainz, Germany, and works in the PSS software and instrument department. She is also responsible for education and customer training. Silvia Fugmann is a laboratory technician at PSS Mainz, Germany, and is responsible for the characterization of reference materials. Peter Kilz studied polymer chemistry in Mainz, Germany, and Liverpool, Great Britain, and is head of the software and instrumentation department. He is also responsible for the development and addition of new detection techniques to PSS WinGPC software. E-mail: Website: Dheld@pss-polymer.com www.pss-polymer.com The Advance Vial Closure System (AVCS) allows bench chemists and sample prep technicians to remove the subjectiveness out of achieving the optimal compression when sealing a vial. The benefit of using a vial and closure designed as a complete system, allows the closure, septum, and PRESENTER: Dave Edwards Sample Handling Product Manager Thermo Fisher Scientific Detlev Lennartz Product Manager Vials and Closures Chromatography Consumables and Specialty Products Thermo Fisher Scientific MODERATOR: Laura Bush Editorial Director LCGC vial to work together to prevent compression from exceeding the optimal range. This web seminar will review the impact an innovative vial & closure design, like AVCS, can have on improving data quality by providing data that supports how an AVCSdesigned vial & closure eliminates under and over compression of vial seals while improving robotic autosampler pick-up of vials, ensuring more centered vial closure positioning and greater optical recognition of vials placed in autosamplers. For questions, contact Kristen Farrell at kfarrell@advanstar.com Who Should Attend: Any chromatographer or mass spectroscopist concerned improving data quality & performance QA/QC looking for ways to improve data quality Laboratory managers focusing on sample throughput and data integrity Sponsored by Presented by 13 1 0