1 Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform Charles Yang, Leo Wang, Dipankar Ghosh, Jonathan Beck, Jamie Humphries; Thermo Fisher Scientific, San Jose, CA, USA
2 Introduction The demand for quick and simple analysis of large numbers of samples in agriculture analysis is growing year by year. Throughout the world pesticides are used to control pests that are harmful to crops, humans and animals. These substances can pose a significant health threat and therefore, need to be accurately detected at the lowest levels requested by the governmental authorities typically at low part per billion (ppb) or low part per trillion (ppt) levels. Traditionally, triple stage quadrupole mass spectrometers have been used by the food industry to identify and quantitate these residues. The Thermo Scientific Exactive Plus Orbitrap mass spectrometer provides high-resolution, accurate mass (HR/AM) to unequivocally identify compounds without time-consuming analysis. The results of this unique solution are improved sensitivity and precision. Mass spectrometric detection for screening experiments with HR/AM technology using full scan experiments can detect as many analytes as necessary using only one chromatographic run. An Exactive Plus mass spectrometer with the proven power of the Thermo Scientific Orbitrap mass analyzer and Thermo Scientific TraceFinder software for unified quantitative and targeted screening data processing fulfills these demands with higher confidence and precision. Methods Sample Preparation Onion matrix were prepared for analysis by using a modified quick, easy, cheap, effective, rugged, and safe (QuEChERS) method, which is a sample preparation procedure used to extract pesticides from food. The QuEChERS extracts were obtained from the California Department of Food and Agriculture (CDFA). For the QuEChERS extraction, 15 g of homogenized sample and 15 ml of acetonitrile were used. Then, 200 L of final QuEChERS extract, 300 L of acetonitrile, and 500 L of water were transferred into an autosampler vial, spiked with 20 L of the pesticides standard, and mixed well. A mixture of 60 pesticides with different starting concentrations was also provided by CDFA to make the standard calibration curve in neat matrix and onion matrix. Calibration of the Exactive Plus mass spectrometer was done with external calibration. There is no need to do lock masses because of the proven stability of the Orbitrap mass analyzer. Hardware The new Exactive Plus mass spectrometer has the proven S-Lens configuration for more ion transmission for more sensitivity. The neutral ions are removed with the bent flatapoles prior to entering the C-Trap. Then the ions are transported into the Orbitrap mass analyzer for HR/AM detection. The system also has an HCD collision cell past the C-trap where all ion fragmentation (AIF) happens. All the fragmented ions are then transported back into the C-Trap and sent into the Orbitrap mass analyzer for detection in accurate mass. (Figure 1) FIGURE 1. Exactive Plus schematics Liquid Chromatography Conditions System: Thermo Scientific UltiMa Column: Thermo Scientific Hypers (50 x 2.1 mm, 1.9 μm pa Mobile Phase A: Water with 0.1% formic a Mobile Phase B: Methanol with 0.1% form Flow Rate: 300 μl/min Column Temp: 40 C Sample Injection: 5 μl Gradient Time A B min 0.00 min 0.25 min 70% 30% min 0% 100% min min Mass Spectrometry Conditions Method: full MS scan Mass Range: -120 to 1000 Ion Mode: positive/negative Resolution: 70,000 Ion Source: heated electrospray ioniz Spray Voltage: 2200v Capillary Temp: 295 ºC Vaporizer Temp: 280 ºC Sheath Gas: 32 Auxiliary Gas: 7 The ease-of-use of HR/AM instrumentation requ transitions and collision energies unlike that of tr HR/AM method development is focused only on Results Data processing was performed with TraceFinde was achieved by applying a mass window of 5 p analytes. All analytes gave very good linear res 0.10 ng/ml depending on starting concentration TraceFinder software comes with many features show how the yellow flags next to the name of th was not found in the sample. The red flag indica with the calibration curve and it did not meet the sign indicates there is an issue with one or more to be reviewed. FIGURE 1. TraceFinder software displaying B matrix R^2, list of compounds, and chromat 2 Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform
3 ple analysis of large numbers of samples in agriculture r. Throughout the world pesticides are used to control, humans and animals. These substances can pose a erefore, need to be accurately detected at the lowest mental authorities typically at low part per billion (ppb) or. Traditionally, triple stage quadrupole mass by the food industry to identify and quantitate these ic Exactive Plus Orbitrap mass spectrometer provides (HR/AM) to unequivocally identify compounds without results of this unique solution are improved sensitivity etric detection for screening experiments with HR/AM eriments can detect as many analytes as necessary ic run. An Exactive Plus mass spectrometer with the cientific Orbitrap mass analyzer and Thermo Scientific d quantitative and targeted screening data processing her confidence and precision. r analysis by using a modified quick, easy, cheap, EChERS) method, which is a sample preparation icides from food. The QuEChERS extracts were epartment of Food and Agriculture (CDFA). For the homogenized sample and 15 ml of acetonitrile were EChERS extract, 300 L of acetonitrile, and 500 L of autosampler vial, spiked with 20 L of the pesticides ixture of 60 pesticides with different starting ed by CDFA to make the standard calibration curve in Calibration of the Exactive Plus mass spectrometer was There is no need to do lock masses because of the mass analyzer. pectrometer has the proven S-Lens configuration for sensitivity. The neutral ions are removed with the bent C-Trap. Then the ions are transported into the Orbitrap ction. The system also has an HCD collision cell past ntation (AIF) happens. All the fragmented ions are then p and sent into the Orbitrap mass analyzer for detection ematics Liquid Chromatography Conditions System: Thermo Scientific UltiMate 3000 UHPLC system Column: Thermo Scientific Hypersil GOLD aq C18 column (50 x 2.1 mm, 1.9 μm particle size) Mobile Phase A: Water with 0.1% formic acid and 10 mm ammonium formate Mobile Phase B: Methanol with 0.1% formic acid and 10 mm ammonium formate Flow Rate: 300 μl/min Column Temp: 40 C Sample Injection: 5 μl Gradient Time A B min 0.00 min 0.25 min 70% 30% min 0% 100% min min Mass Spectrometry Conditions Method: full MS scan Mass Range: -120 to 1000 Ion Mode: positive/negative Resolution: 70,000 Ion Source: heated electrospray ionization (H-ESI) Spray Voltage: 2200v Capillary Temp: 295 ºC Vaporizer Temp: 280 ºC Sheath Gas: 32 Auxiliary Gas: 7 The ease-of-use of HR/AM instrumentation requires no optimization of mass transitions and collision energies unlike that of triple stage quadrupole methods. HR/AM method development is focused only on the HPLC. Results Data processing was performed with TraceFinder software. Specificity of analysis was achieved by applying a mass window of 5 ppm to the theoretical mass of the analytes. All analytes gave very good linear response in the calibration range (0.01 to 0.10 ng/ml depending on starting concentration in mixture). TraceFinder software comes with many features including flagging. Figures 2 and 3 show how the yellow flags next to the name of the compound indicate the compound was not found in the sample. The red flag indicates that the compound has an issue with the calibration curve and it did not meet the flagging requirement. The caution sign indicates there is an issue with one or more compounds in the sample that needs to be reviewed. FIGURE 1. TraceFinder software displaying Boscalid calibration curve plot of matrix R^2, list of compounds, and chromatogram FIGURE 2. TrceFinder displaying Imidaclopr R^2, list of compounds, and chromatogram In addition to the quantitation, TraceFinder soft page where you can see all the reports prior to of reports are available with the ability to perfor tool you can search for a particular compound i every page (Figure 3). Figure 4 shows a popular feature of TraceFinde step, last minute changes can be accomplished to return to the method development step. Table 1 shows the list of >90 targeted pesticide shows the retention time, R 2 and LOQ for the c FIGURE 3. A simple high-density calibratio up to 9 calibration curves on one sheet. In a particular compound by clicking on the bi Thermo Scientific Poster Note PN63685_e 08/12S 3
4 nditions o Scientific UltiMate 3000 UHPLC system o Scientific Hypersil GOLD aq C18 column 2.1 mm, 1.9 μm particle size) r with 0.1% formic acid and 10 mm ammonium formate anol with 0.1% formic acid and 10 mm ammonium formate L/min FIGURE 2. TrceFinder displaying Imidacloprid calibration curve plot of matrix, R^2, list of compounds, and chromatogram FIGURE 4. The Data Review screen allows th without going back to Method Development. A B 70% 30% 0% 100% ions S scan to 1000 ve/negative 0 d electrospray ionization (H-ESI) v C C strumentation requires no optimization of mass gies unlike that of triple stage quadrupole methods. t is focused only on the HPLC. ed with TraceFinder software. Specificity of analysis mass window of 5 ppm to the theoretical mass of the ery good linear response in the calibration range (0.01 to rting concentration in mixture). with many features including flagging. Figures 2 and 3 xt to the name of the compound indicate the compound The red flag indicates that the compound has an issue it did not meet the flagging requirement. The caution ue with one or more compounds in the sample that needs In addition to the quantitation, TraceFinder software has a very easy-to-view report page where you can see all the reports prior to print out. More than 35 different types of reports are available with the ability to perform customization. With a simple find tool you can search for a particular compound in the report without having to review every page (Figure 3). Figure 4 shows a popular feature of TraceFinder software. During the Data Review step, last minute changes can be accomplished prior to report printing without having to return to the method development step. Table 1 shows the list of >90 targeted pesticides analyzed in this experiment. Table 2 shows the retention time, R 2 and LOQ for the components analyzed in onion matrix. FIGURE 3. A simple high-density calibration report is shown below allowing up to 9 calibration curves on one sheet. In this window you can also search for a particular compound by clicking on the binocular to open the search window. Table 1. List of 90 pesticides and associate theoretical m/z with calculated mass errors ftware displaying Boscalid calibration curve plot of nds, and chromatogram 4 Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform
5 playing Imidacloprid calibration curve plot of matrix, d chromatogram FIGURE 4. The Data Review screen allows the user to perform quick changes without going back to Method Development. Table 2. Tabulated results of LOQ s per co Table 1. List of 90 pesticides and associated retention times with actual vs. theoretical m/z with calculated mass errors. n, TraceFinder software has a very easy-to-view report l the reports prior to print out. More than 35 different types the ability to perform customization. With a simple find rticular compound in the report without having to review ature of TraceFinder software. During the Data Review an be accomplished prior to report printing without having elopment step. 0 targeted pesticides analyzed in this experiment. Table 2 2 and LOQ for the components analyzed in onion matrix. -density calibration report is shown below allowing on one sheet. In this window you can also search for y clicking on the binocular to open the search window. Conclusion TraceFinder software coupled with the spectrometer powered by Orbitrap tech quantitative and screening data in one p The results show very linear R 2 with ver pesticides in difficult matrix. The power of resolution at 70,000 (FWH interferences which can cause lower-en The mass error also shows confidence Acknowledgements The authors wish to thank to Roland Carlson extracts, pesticide mixture and methodology f All trademarks are the property of Thermo Fisher Scientific a encourage use of these products in any manners that might Thermo Scientific Poster Note PN63685_e 08/12S 5
6 creen allows the user to perform quick changes Development. Table 2. Tabulated results of LOQ s per compound, retention times and curve fit R 2 s and associated retention times with actual vs. ted mass errors. Conclusion TraceFinder software coupled with the Exactive Plus benchtop mass spectrometer powered by Orbitrap technology provides easy access to full quantitative and screening data in one package. The results show very linear R 2 with very low LOQs which will assist in detecting pesticides in difficult matrix. The power of resolution at 70,000 (FWHM) at m/z 200 helps resolve matrix interferences which can cause lower-end detection issues. The mass error also shows confidence in the results of the data acquired. Acknowledgements The authors wish to thank to Roland Carlson at CDFA for providing the QuEChERS extracts, pesticide mixture and methodology for this study. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. 6 Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform
7 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Scientific Dionex products are designed, developed, and manufactured under an ISO 9001 Quality System. Australia Austria Benelux Brazil China Denmark France Germany India Ireland Italy Japan Korea Singapore Sweden Switzerland Taiwan UK USA and Canada PN63685_E 08/12S
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System Olaf Scheibner, Maciej Bromirski, Thermo Fisher Scientific, Bremen, Germany
More informationLC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
More informationSimultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
More informationSPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
More informationUHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
More informationFast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
More informationThermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
More informationSelective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
More informationRapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection
Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection Derek Hillbeck, Thermo Fisher Scientific, Runcorn, UK Application Note 2999 Key
More informationA Complete Solution for Method Linearity in HPLC and UHPLC
Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,
More informationOverview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
More informationUsing Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
More informationThermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance
Thermo Scientific SOLA SPE cartridges and plates Technical Guide Join the revolution unparalleled performance Join the revolution next-generation SPE Thermo Scientific SOLA products revolutionize Solid
More informationDetermination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
More informationQuantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
More informationExtraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
More informationThermo Scientific Dionex Chromeleon 7 Chromatography Data System Software
Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software Frank Tontala, Thermo Fisher Scientific, Germering, Germany Technical Note 708 Key Words Chromeleon Chromatography Data System,
More information# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
More informationHow To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
More informationAnalysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection Application Note Food Testing & Agriculture Authors Edgar Naegele and Thomas
More informationTargeted and Nontargeted MS Analysis of Contaminants in Storm Water Retention Ponds
Targeted and Nontargeted MS Analysis of Contaminants in Storm Water Retention Ponds Gordon Getzinger 1, P. Lee Ferguson 1, Jonathan Beck 2, Charles Yang 2, Frans Schoutsen 3 1 Duke University, Durham,
More informationDaniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
More informationSimultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
More informationDetermination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
More informationThe Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
More informationDiscovery of Pesticide Protomers Using Routine Ion Mobility Screening
Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,
More informationCONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
More informationAnalysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
More informationPesticide residue analysis: New Trends. Michal Godula, Ph.D. EU Marketing Manager Food Safety & Environmental Thermo Fisher Scientific
Pesticide residue analysis: New Trends Michal Godula, Ph.D. EU Marketing Manager Food Safety & Environmental Thermo Fisher Scientific Why pesticide analysis? Extensively used in agriculture to control
More informationThermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity
Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use Confidence in Results With Data Integrity 2 Make the World Healthier With the LC-MS Tests You Run Confidence in Test Results With Data
More informationThird Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software
Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software White Paper 70885 Executive Summary Over the last few decades chromatographic instrumentation and separations have
More informationWhat is Eluent Generation?
What is Eluent Generation? Monika Verma Thermo Fisher Scientific, Sunnyvale, CA, USA White Paper 70607 Executive Summary Eluent generation is an electrolytic process that allows the production of high
More informationAccelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform
Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform Remco Swart Pittcon March 8-12, 2015 1 The world leader in serving science Outline Thermo Scientific Vanquish UHPLC
More informationThermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
More informationWATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
More informationUltra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
More informationThermo Scientific PepFinder Software A New Paradigm for Peptide Mapping
Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping For Conclusive Characterization of Biologics Deep Protein Characterization Is Crucial Pharmaceuticals have historically been small
More informationSupplementary Materials and Methods (Metabolomics analysis)
Supplementary Materials and Methods (Metabolomics analysis) Metabolite extraction and mass spectrometry analysis. 12 Vldlr-/- knock out and 12 wild type (WT) retinas were separately frozen at -80 ºC in
More informationLC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
More informationOverview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System
Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS
More informationA fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics,
More informationA Navigation through the Tracefinder Software Structure and Workflow Options. Frans Schoutsen Pesticide Symposium Prague 27 April 2015
A Navigation through the Tracefinder Software Structure and Workflow Options Frans Schoutsen Pesticide Symposium Prague 27 April 2015 Kings day in The Netherlands 1 Index Introduction Acquisition, Method
More informationAnalysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
More informationAutomated Method Development Utilizing Software-Based Optimization and Direct Instrument Control
Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control Dr. Frank Steiner, 1 Andreas Brunner, 1 Fraser McLeod, 1 Dr. Sergey Galushko 2 1 Dionex Corporation, Germering,
More informationTargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet
TargetQuan 3 Software Leading the way in regulatory POPs quantification Bullet Bullet Bullet Leading the way in regulatory POPs quantification Analyse Samples Open Sequence Process Sequence Analyse Samples
More informationGENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
More informationThermo Scientific SIEVE Software for Differential Expression Analysis
m a s s s p e c t r o m e t r y Thermo Scientific SIEVE Software for Differential Expression Analysis Automated, label-free, semi-quantitative analysis of proteins, peptides, and metabolites based on comparisons
More informationAppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT
AppNote 1/2014 Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine Fred D. Foster, John R. Stuff, Edward A. Pfannkoch GERSTEL, Inc., 701 Digital Dr. Suite J,
More informationChemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
More informationAdvantages of the LTQ Orbitrap for Protein Identification in Complex Digests
Application Note: 386 Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests Rosa Viner, Terry Zhang, Scott Peterman, and Vlad Zabrouskov, Thermo Fisher Scientific, San Jose, CA,
More informationSimultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs
Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan
More informationAnalysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D.
Gas Chromatograph Mass Spectrometer No. GCMS-1603 Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS Shilpi Chopra, Ph.D. Introduction Organophosphorus (OP) pesticides are a class
More informationThermo Scientific ExactFinder Software
mass spectrometry Thermo Scientific ExactFinder Software Workflow software for routine targeted and general unknown screening Unifies qualitative and quantitative high-resolution accurate mass workflows
More informationPesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
More informationA High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
More informationNew drugs of abuse? A case study on Phenazepam & Methoxetamine
New drugs of abuse? A case study on Phenazepam & Methoxetamine Presenter: Nadia Wong Co authors: Dr Yao Yi Ju & Alex Low Xuan Kai Analytical Toxicology Laboratory Clinical & Forensic Toxicology Unit Applied
More informationHRMS in Clinical Research: from Targeted Quantification to Metabolomics
A sponsored whitepaper. HRMS in Clinical Research: from Targeted Quantification to Metabolomics By: Bertrand Rochat Ph. D., Research Project Leader, Faculté de Biologie et de Médecine of the Centre Hospitalier
More informationMass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers. Rob Grothe Thermo Fisher Scientific, San Jose, CA
Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers Rob Grothe Thermo Fisher Scientific, San Jose, CA Overview Purpose: Robust and accurate calibration of mass positions
More informationAppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS
AppNote 1/2012 Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS Oscar G. Cabrices, Fred D. Foster, John R. Stuff, Edward
More informationOverview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
More informationHigher Resolution LC-MS and MS-MS Analysis of Lipid Extracts Using Benchtop Orbitrap-based Mass Spectrometers and LipidSearch Software
Higher Resolution LC-MS and MS-MS Analysis of Lipid Extracts Using Benchtop Orbitrap-based Mass Spectrometers and LipidSearch Software David A Peake, 1 Yasuto Yokoi, 2 Reiko Kiyonami 1 and Yingying Huang
More informationThermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID
Thermo Scientific Compound Discoverer Software A New Generation of integrated solutions for small molecule structure ID Transforming Small Molecule Research Solutions for Small Molecule Structural ID Thermo
More informationPosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
More informationIntroduction. Methods. Sample Processing
PO-CON1656 High-sensitivity, high-throughput quantitation of catecholamines and metanephrine in plasma by automated WCX-SP coupled to LC/MS/MS for clinical research ASMS 2016 MP-080 Ichiro HIRANO 1, Atsuhiko
More informationEQuan MAX Getting Connected Field Service Guide
EQuan MAX Getting Connected Field Service Guide This guide provides procedures to connect the various components for the Thermo Scientific EQuan MAX automated high-throughput LC-MS system. Contents Hardware
More informationCliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.
Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine
More informationJennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
More informationAnalysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
More informationA feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection LGC/R/2011/181 LGC Limited 2011 GC11/0005A A feasibility study of the determination of B-group
More informationMethod development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report
of melamine-ware by GC-MS and LC-MS/MS Page 1 of 15 Method development for analysis of formaldehyde in foodsimulant extracts of melamine-ware by GC-MS and LC-MS/MS December 2012 Contact Point: Chris Hopley
More informationTherapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS
Therapeutic Drug Monitoring of Antiretroviral Drugs with PLC-M Ursula Gutteck-Amsler, Katharina M. Rentsch Abstract Prospective and retrospective studies have provided some evidence of the clinical and
More informationAccurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview
More informationGas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
More informationfor mass spectrometry calibration tools Thermo Scientific Pierce Controls and Standards for Mass Spectrometry
Thermo Scientific Pierce Controls and Standards for Mass Spectrometry calibration tools for mass spectrometry Ensure confidence in instrument performance with Thermo Scientific Pierce Calibration Solutions
More informationNANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published
More informationDevelopment and Validation of Methods for Chemotherapy Drugs on the New Prelude SPLC LC-MS/MS System
Development and Validation of Methods for Chemotherapy Drugs on the New Prelude SPLC LC-/ System Kerry Hassell, Dayana Argoti, Sarah J. Fair and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA
More informationMultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
More informationThermo Scientific Vanquish UHPLC System. Revolutionize Your. UHPLC Experience
Thermo Scientific Vanquish UHPLC System Revolutionize Your UHPLC Experience We feel that achieving better performance is not the same as spending more effort. We know that every breakthrough starts with
More informationAnalyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
More informationTechnical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
More informationSTANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080
Training on the ZQ Open access MS Questions? Contact Dr. Allis Chien allis@stanford.edu 650-723-0710 0710 STANFORD UNIVERSITY MASS SPECTROMETRY STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD
More informationDetermination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)
Application Note 340 Determination of Fat in Dried Milk Products INTRODUCTION Many extraction techniques for the determination of fat in food are labor-intensive or require long extraction times. The Roese-Gottlieb
More informationEarning Customer Loyalty
Earning Customer Loyalty Loyalty cannot be purchased. It s something we earn. When it comes to attracting and retaining dedicated customers, we rely on excellent service to build solid relationships. A
More informationImproving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification
Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Overview LightSight Software for Metabolite Identification is a complete environment
More informationMethod Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.
More informationOptimize Your Process Operations by Improving Process Monitoring
Integral Process Ana Optimize Your Process Operations by Improving Process Monitoring All industrial processes have a common need reliable, accurate, and timely methods for analyzing process chemistry.
More informationStep-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
More informationThermo Scientific ConFlo IV Universal Interface
m a s s s p e c t r o m e t r y Thermo Scientific ConFlo IV Universal Interface The Universal Interface Part of Thermo Fisher Scientific ConFlo IV Universal Interface The Universal Interface The development
More informationStandard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS
Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLCDAD and HPLCMS/MS Jongheon Shin College of Pharmacy Seoul National University Summary The standard analytical
More informationThermo Scientific HyperSep Solid Phase Extraction Method Development Guide
chromatography Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide The following guide provides considerations, tips and general guidelines for developing SPE methods using the Thermo
More informationCORESTA RECOMMENDED METHOD N 72
CORESTA RECOMMENDED METHOD N 72 DETERMINATION OF TOBACCO-SPECIFIC NITROSAMINES IN SMOKELESS TOBACCO PRODUCTS BY LC-MS/MS (February 216). INTRODUCTION The CORESTA Smokeless Tobacco Sub-Group studied various
More informationIncreasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative
More information[ Care and Use Manual ]
PREP Calibration Mix DIOS Low i. Introduction Pre-packaged PREP Calibration Mixtures eliminate the need to purchase and store large quantities of the component calibration reagents, simplifying sample
More informationAMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
More informationApplication Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow
Application Note # LCMS-92 Interlaboratory Tests Demonstrate the Robustness and Transferability of the Toxtyper Workflow Abstract There is high demand in clinical research and forensic toxicology for comprehensive,
More information[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS. Simplifed
[ DISQUE DISPERSIVE SAMPLE PREPARATION EXTRACTION PRODUCTS ] QuEChERS Simplifed The QuEChERS method (an acronym for Quick, Easy, Cheap, Effective, Rugged, and Safe) followed by dispersive solid-phase extraction
More informationSimultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
More information