Finnigan ELEMENT2. High Performance High Resolution ICP-MS



Similar documents
Thermo Scientific ConFlo IV Universal Interface

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

Thermo Scientific ExactFinder Software

aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS

Thermo Scientific SIEVE Software for Differential Expression Analysis

Sample Analysis Design Isotope Dilution

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry

aurora ICP-MS Innovation with Integrity The Science of Sensitivity ICP-MS

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

How To Use The Nexion 300

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES

AMD Analysis & Technology AG

The 30-Minute Guide to ICP-MS

Getting Answers at the Source Requires a New Way of Thinking About Software

Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

How To Use An Ionsonic Microscope

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?

Pesticide Analysis by Mass Spectrometry

Specifications. MassHunter WorkStation for Agilent 7700 Series ICP-MS. Specifications: MassHunter Workstation (G7211A) for the Agilent 7700 Series

Elemental Analyses by ICP-AES

WATERS QUANTITATIVE ANALYSIS solutions

OLIVÉR BÁNHIDI 1. Introduction

AutoSpec Premier. Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements?

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Simplicity. Reliability. Performance

VALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

Thermo Scientific Qtegra Intelligent Scientific Data Solution Software. Delivering quality. Driving productivity

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

MODERN ATOMIC THEORY AND THE PERIODIC TABLE

h e l p s y o u C O N T R O L

[ Care and Use Manual ]

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID

Appendix 5 Overview of requirements in English

Universal Data Acquisition (UDA)

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

13C NMR Spectroscopy

Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and

Thermo Scientific irm-gc/ms Technology

Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Standard Operation Procedure. Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS

Determination of elements by ICP-AES and ICP-MS

LC-MS/MS Solutions for the Pharmaceutical and CRO Industries TSQ Series -based Solutions for Drug Discovery and Development

Elemental analyzer multi EA 5000

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

LC-MS/MS for Chromatographers

Agilent Cary 60 UV-Vis

Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES. FIMS 100 and 400. Flow Injection Mercury Systems

SUS. Company Profile. Ulrich Nell, Feldstr.23, D Oberhausen, Tel. 0049(0)208/ Fax 0049(0)208/658536

APPLICATION OF ICP-MS TECHNOLOGY FOR TRACE METALS ANALYSIS

NexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight.

Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist

QGA Quantitative Gas Analyser

DDX 7000 & Digital Partial Discharge Detectors FEATURES APPLICATIONS

G4 ICARUS HF. Innovation with Integrity. Simultaneous Determination of Carbon and Sulfur in Solids. Combustion Method

High Sensitivity ICP-MS: Overcome the problem of complex samples

Start of introducing ICP MS method for analyzing water samples for Strontium. November 27, 2014 Tokyo Electric Power Company

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

CLASS TEST GRADE 11. PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change

SCION SQ Series GC-MS

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances

FT-NIR for Online Analysis in Polyol Production

Agilent Television Power Consumption Testing. Application Note

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

How To Build An Amd Quas 3 Ar Mass Spectrometer

Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests

Amount of Substance.

Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers. Rob Grothe Thermo Fisher Scientific, San Jose, CA

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics

Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers

Wipe Analysis to Determine Metal Contamination on Critical Surfaces

Flow Injection Analysis

Bruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

CHEM 1411 Chapter 5 Homework Answers

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

GC-xt. The most flexible GC /GC-MS liquid sample injector

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

G8 GALILEO. Innovation with Integrity. High-End Melt-extraction Analyzer. Inert Gas Method

Transcription:

m a s s s p e c t r o m e t r y Finnigan ELEMENT2 High Performance High Resolution ICP-MS Analyze Detect Measure Control

Finnigan ELEMENT2 High Performance High Resolution ICP-MS Finnigan ELEMENT2 High Performance High Resolution ICP-MS An Instrument for every Application A success story around the world with over 3 instruments installed, for example: Belgium: OCAS. Research centre for the application of steel with New Wave laser ablation system. www.ocas.be Sweden: Analytica AB. Scandinavia s leading contract lab with Emma Engström. www.analytica.se Canada: Seastar Chemicals Inc. A manufacturer of high purity chemicals with Dr. Brad McKelvey. www.seastarchemicals.com USA: Micron Technology Inc. Advanced semiconductor solutions with Kevin Coyle. www.micron.com Germany: Institute for Transuranium Elements. Shown with a glovebox for radioactive sample measurements. http://itu.jrc.cec.eu.int USA: Desert Research Institute. Hydrologic Sciences Division with Steve Lambert. www.dri.edu

Multielement analysis across the periodic table covering a mg/l to sub pg/l concentration range Compatible with inorganic and organic solution matrices and solids High mass resolution to access spectrally interfered isotopes Produces unambiguous elemental spectra A multielemental detector for transient signals For example, CE, HPLC, GC, FFF and laser ablation High precision isotope ratios Independent of interferences or interfered isotopes Fully automated tuning and analysis In conjunction with a comprehensive, customizable quality control system Reliability and robustness to serve as a 24 / 7 production control tool Highest sample throughput Highest flexibility and accessibility to serve as an advanced research tool

Finnigan ELEMENT2 High Performance High Resolution ICP-MS High Mass Resolution The unequivocal separation of analyte ions from spectral interferences is a prerequisite of accurate and precise analysis. High mass resolution is the universal means for this separation. Spectral interferences are the main limitation of ICP-MS. The argon plasma gas, water, acid and the sample matrix itself can combine to introduce a wide range of polyatomic ion species. The resultant interfering species may have the same nominal mass as an analyte ion and thus return a falsely high value for the analyte. Numerous strategies have been used in order to try to minimize or circumvent the formation of these spectral interferences. These include mathematical corrections, special sample introduction systems, special plasma parameters and collision/dynamic reaction cells to neutralize part of the interferences. High resolution with a sector-field mass spectrometer simply distinguishes the analyte from interference by difference in mass. The capability of high mass resolution is a feature unique to the Finnigan ELEMENT2. This ability can be used for quantification and isotope ratio analysis for nearly the whole periodic table, and in almost all matrices. Even in a sample matrix as simple as ultra pure water (UPW), interferences do exist especially at low analyte concentration levels: 3.94 3.95 3.96 3.97 3.98 3.99 31. 31.1 Phosphorus in UPW, Medium Resolution Zoom in 28 Si 31 P 12 C 16 O 15 N 16 O 14 N 16 OH 3.94 3.95 3.96 3.97 3.98 3.99 31. 31.1 14 N2 27.94 27.95 27.96 27.97 27.98 27.99 28. 28.1 Silicon in UPW, Medium Resolution 12 C 16 O2 28 Si 16 O 44 Ca 43.9 43.92 43.94 43.96 43.98 44. Calcium in UPW, Medium Resolution

32 S 32 S 63.89 63.9 63.91 63.92 63.93 63.94 63.95 63.96 63.97 Zoom in 64 Zn Zinc in H 2 SO 4 (1 % w/w), High Resolution 32 S 16 O2 63.89 63.9 63.91 63.92 63.93 63.94 63.95 63.96 63.97 The more complex the sample matrix, the wider the range of interferences that will occur. The main advantage of high mass resolution as a technique to remove spectral interferences is that it is not just limited to a particular type of interference. With high mass resolution, iron at m/z 56 is easily separated in a simple matrix from 4 Ar 16 O as well as in more complex matrices from, for example, 4 Ca 16 O. Arsenic at m/z 75 can be determined in HCl, separated from the 4 Ar 35 Cl interference, or, in a matrix containing Ca and chloride, from 4 Ca 35 Cl. Matrices such as mineral acids and organic solvents, which form a plethora of spectral interferences, can be easily analyzed using the high resolution mode. High resolution results in simple clear spectra and does not create new interferences. 4 Ar 16 O 12 C2 16 O2 56 Fe 55.8 55.85 55.9 55.95 56. 56.5 56.1 Iron in Liquid Crystal Matrix, Cold Plasma, Medium Resolution 44 Ca 16 O 44 Ca 16 OH 59.88 6 Ni 43 Ca 16 OH 32 S 16 O 12 C 6.88 61.88 59.9 59.92 59.94 59.96 59.98 6.9 6.92 6.94 6.96 6.98 61.92 61.92 61.94 61.96 61.98 Nickel Isotopes in Groundwater, Medium Resolution 61 Ni 62 Ni 46 Ca 16 O Resolution Specification (1 % peak valley definition) 3 fixed resolutions: Low Resolution > 3 Medium Resolution > 4 High Resolution > 1

Finnigan ELEMENT2 High Performance High Resolution ICP-MS Principle The Finnigan ELEMENT2 is a double focusing magnetic sector field ICP-MS. Plasma and Interface The argon plasma ion source and sampling interface of the Finnigan ELEMENT2 are at ground potential 1. This enables the straightforward coupling of peripherals like HPLC, CE, GC and laser ablation. The Finnigan ELEMENT2 interface reduces the initial kinetic energy spread from ~2 to ~5 ev by capacitively decoupling the plasma from the load coil, using a grounded guard electrode. This reduced energy spread increases the ion transmission and delivers superior sensitivities at all resolutions. The ion transfer optics focus the ions from the plasma interface on the entrance slit of the double focusing analyzer. The Finnigan ELEMENT2 ion transfer optics are designed for low background, highest sensitivity and minimum mass bias at maximum stability 2. Highly laminated, water cooled magnet, m/z: 7-24 - 7 in < 15 ms 25 ppm/8 hour mass stability High Resolution The Finnigan ELEMENT2 is able to fully automatically change between three fixed resolutions by switching the positions of the entrance and exit slits in < 1 s. The patented design of the fixed slit mechanism 3 offers maximum stability and reproducibility of resolutions. MR LR HR 1 Patent issued: US 5552599, GB 2282479 2 Patent issued: US 5625185 3 Patent issued: US 545178, GB 2281438 3 fixed resolutions: R = 3 R = 4 R = 1 Ion Transfer Optics Guarantees a flat response curve, low background and high sensitivity

Simultaneous measurement in analog and counting modes > 1 9 linear dynamic range <.2 cps dark noise Fully automatic cross calibration Interface at ground potential Easily changeable cones Resistant against chemicals Mass Separation The magnetic field disperses ions according to their mass and energy. The magnet used in the Finnigan ELEMENT2 is specifically designed for use in ICP-MS applications. It is relatively small (sufficient for the mass range 26 u), highly laminated and efficiently water-cooled for the highest mass stability. Changing of the magnetic field is controlled by a magnetic field regulator with a new high power stage, which delivers the fastest scan speed ever possible with a magnetic sector field instrument 4. After passing through the magnetic field the ions enter the electrostatic analyzer for energy focusing. The combination of the magnetic and electrostatic fields results in the double focusing, high resolution properties of the Finnigan ELEMENT2. Detection System The Finnigan ELEMENT2 is equipped with a discrete dynode detector system. Rather than having the ion beam directly strike the detector to initiate an electron cascade, the secondary electron multiplier implemented in the Finnigan ELEMENT2 uses a conversion dynode at -8 kv, producing a uniform response across the mass range. The detector is linear over nine orders of magnitude from ppq to ppm concentrations. The quantification of trace and major elements is therefore possible in a single analysis. 4 Thermo Electron Application Note AN311_E12/3C Torch with Guard Electrode (GE) The GE decreases the ion energy spread, thus increasing ion transmission. This and the high acceleration voltage used in a sector field ICP-MS results in increased sensitivity. The GE is also required for Cold Plasma measurements

Finnigan ELEMENT2 High Performance High Resolution ICP-MS Resolution Low Resolution Low resolution (R = 3) is used for the analysis of non-interfered isotopes. In this mode the Finnigan ELEMENT2 s sensitivity is the highest of all commercially available ICP-MS instruments. Additionally, the flat top peak shape is an advantage for high precision isotope ratio measurements. Medium Resolution Medium resolution (R = 4) guarantees interference-free analysis for most elements in the majority of sample matrices. For example, transition elements are routinely measured in medium resolution due to the formation of many interfering polyatomic species in the mass range 24 7 u. High Resolution High resolution (R = 1) is used for the analysis of elements in the most challenging sample matrices. For example, high resolution is used to separate As and Se from argon dimer interferences and argon chloride interferences in chlorine matrices, heavy rare earth elements from light rare earth element oxides in geological matrices, platinum group elements from argon-transition metal molecular species, and/or the oxides of Hf, Ta and W. Intensity [cps] 35 3 25 2 15 1 5 114.6 114.7 114.8 1 ng/l 115 In, Low Resolution Intensity [cps] 35 3 25 2 15 1 114.9 115. 115.1 115.2 114.88 114.89 114.9 114.91 114.92 1 ng/l 115 In, Medium Resolution 6 5 Intensity [cps] 4 3 2 1 114.895 114.9 114.95 114.91 1 ng/l 115 In, High Resolution

Intensity [cps] 7 6 5 4 3 179 Hf 16 O Since the change in mass resolution is achieved by changing the width of the entrance and exit slits of the mass spectrometer, the instrumental sensitivity of a high resolution ICP-MS is dependent on the resolution mode used. Therefore, the Finnigan ELEMENT2 with three fixed resolutions has three sensitivities: the wider the slit, the higher the sensitivity. 2 1 195 Pt Fixed sensitivity ratio between resolutions: independent of mass and matrix. 194.85 194.9 194.95 195. 195.5 195.1 1 µg/l Hafnium, 1 ng/l Platinum, High Resolution Even in high resolution mode, the intrinsic sensitivity of the Finnigan ELEMENT2 provides sub ppt detection limits. 6 5 4 Ar 12 C Intensity [cps] 4 3 2 1 52 Cr 37 Cl 16 O 52. 53 Cr 52.2 52.4 52.6 52.8 53. Chromium in Blood, Medium Resolution 25 2 153 Eu 16 O Intensity [cps] 15 1 5 168.86 169 Tm 168.88 168.9 168.92 168.94 168.96 168.98 169. Thulium in Europium Matrix, High Resolution Sensitivity Specification Low Resolution (R = 3) 115 In > 1x1 6 cps/ppb Medium Resolution (R = 4) 115 In > 1x1 5 cps/ppb High Resolution (R = 1) 115 In > 1.5x1 4 cps/ppb Background in all 3 resolutions <.2 cps

Finnigan ELEMENT2 High Performance High Resolution ICP-MS Sensitivity and Stability Sensitivity and Low Background It is obvious that the highest instrumental sensitivity is essential to achieve the lowest detection limits. However, it is the signal to noise ratio that dictates the detection limit. The Finnigan ELEMENT2 guarantees an off-peak background of <.2 cps for all three resolutions. Detection limits in the fg/l range are possible. Detection Limits in Complex Matrices Even in cases where the lowest detection limits are not the goal, the exceptionally high sensitivity of the Finnigan ELEMENT2 offers an important advantage. With the higher sensitivity, higher dilution factors for complex matrices can be used without sacrificing detection limits. Intensity [cps].8.7.6.5.4.3.2.1. bkg<.2 cps 226 Ra bkg<.2 cps 225.4 225.6 225.8 226. 226.2 226.4 Mineral Water, undiluted (5 % HNO 3 ), ~2 fg/l Radium, Low Resolution Sensitivity and Dilute and Shoot The requirement for matrix separation for samples such as seawater, mineral acids or organic solvents is replaced by a simple dilution. Using this dilute and shoot approach, the matrix load on the sample introduction system, plasma and interface is reduced. This is also particularly important for the analysis of nuclear sample matrices where the minimization of waste is of paramount importance. 23 Na 9 Zr 17 Ag 151 Eu 47 Ti 52 Cr 63 Cu 66 Zn 79 Br 75 As 155 Gd Resolution Low Low Low Low Medium Medium Medium Medium Medium High High LoD [ng/l] in solution 1.3.6.3.4.3.3.6 1.5 5. 2..2 Detection Limits in 5 mg/l Uranium 6 Normalized Intensity [ppq] 5 4 3 2 1 1 2 3 4 5 6 Concentration [ppq] 226 Ra, External Calibration, 1, 4, 6 pg/l, R 2 = 1.

Intensity [cps] 12 1 8 6 4 2 5.1 ng/l As -5 5 1 15 Concentration [ppt] Signal Stability and Detection Limits Sector field ICP-MS offers superior ion transmission stability due to the high acceleration voltage (-8 kv) and excellent focusing properties. In combination with the high sensitivity, low background, interference-free measurements and an advanced sample introduction system, the Finnigan ELEMENT2 delivers the lowest detection limits independent of the sample matrix. The resulting low detection limits combined with highest stability at single digit ppt level enable quantification at the lowest concentrations. Standard Addition Analysis of Arsenic in HCl (1 % v/v) 11 1 9 Recovery [%] 8 7 6 5 Cu Zr Ag Ba Tl Cr Mn Li Na Fe element 1 ng/l Spike Recovery UPW concentration [ppt] 6. 5.8 5.6 5.4 5.2 5. 4.8 4.6 4.4 Sn12(LR) 2.5% Fe56(MR) 2.2% As75(HR) 4.1% Signal Stability 2 Hours Low, Medium and High Resolution: 5 ng/l Stability Specifications < 1 % RSD in 1 minutes < 2 % RSD in 1 hour 4.2 4. 1 2 3 4 5 6 7 8 9 1 11 12 13 14 15 16 analysis #

Finnigan ELEMENT2 High Performance High Resolution ICP-MS Mass Stability The mature design of the Finnigan ELEMENT2 delivers the fastest scan speed ever for a magnetic sector field ICP-MS. With the Finnigan ELEMENT2 a jump from 7 to 24 to 7 is realized in < 15 ms. Scanning Routine scanning is performed by a combination of magnetic and electric jumps. The Finnigan ELEMENT2 has the unique capability to scan +3% from the magnet mass by decreasing the acceleration voltage. The magnetic field is kept constant while the acceleration voltage is varied. This ingenious combination of scan techniques delivers the fastest scanning ever realized by a sector field ICP-MS. Coil Current (A) 1 9 8 7 6 5 4 3 2 1 m/z 7-238 75 ms 5 ms 5 ms m/z 238-7 5 ms m/z 7-238 75 ms m/z 238-7 5 ms 5 ms 5 ms 5 ms 5 1 15 2 25 Time (ms) Timing for a Full Magnet Cycle Mass (m/z) 238 7 Mass Stability The Finnigan ELEMENT2 guarantees the highest mass stability of any ICP-MS. This enables fast peak top jumping analyses rather than scanning across the whole peak, thus significantly decreasing analysis times. A combination of mature hardware and intelligent software guarantees mass stability in high resolution, which makes mass calibration a rare event. The synthesis of these characteristics combined with high sensitivity, allowing short integration times, makes the Finnigan ELEMENT2 the fastest sector field ICP-MS ever. Centroid Mass 77.921 77.92 77.919 77.918 77.917 77.916 77.915 77.914 77.913 Analysis y-scale is equivalent to peak width in high resolutiom The scan speed of the Finnigan ELEMENT2 in combination with with the high sensitivity opens the dimension of its use as a detector for fast transient signals. 12-Hour Mass Stability for 78 Se in high resolution Specifications Mass stability: 25 ppm / 8 hours Magnetic scan speed: m/z 7 to 24 to 7 < 15 ms

Ion Detection System analog signal The goal of real life analysis is the simultaneous determination of major and trace elements. Conversion Dynode Electrons e - Ions m + analog counting SEM 12345 counting signal The Finnigan ELEMENT2 is equipped with a discrete dynode detection system that enables the quantification of both trace and major elements in a single analysis across a dynamic range of 1 9. The Finnigan ELEMENT2 detector system incorporates a conversion dynode at -8 kv. The high acceleration to -8 kv results in a mass independent detector response, enabling fully automatic cross calibration between the counting and analog modes. No user interaction is necessary. Finnigan ELEMENT2 Ion Detection System 2 1 The automatic cross calibration ensures a constantly updated detector response at all times ready for the unexpected sample! Additionally the detector enables the measurement of large isotope ratios. With the minor isotope in counting mode and the abundant isotope in analog mode the highest precision and accuracy can be obtained. 8 Intensity [cps] 15 1 1ng/L 25 TI Analog: red line Counting: blue line 6 4 Rel. Intensity [%] 5 2 24.4 24.6 24.8 25. 25.2 25.4 25.6 Simultaneous Measurement of the Counting and Analog Signals Isotopes of Uranium Determination of large Isotope ratios using the both mode blue signals = counting, red signals = analog 234 235 236 238 Specifications Dynamic Range > 1 9 with automatic gain calibration

Finnigan ELEMENT2 High Performance High Resolution ICP-MS The Software Suite The Finnigan ELEMENT2 Software The Finnigan ELEMENT2 software controls and monitors all instrument functions for ICP-MS analysis. This includes data acquisition and auto-tuning of the Finnigan ELEMENT2. The software also provides the full range of quantification procedures required in elemental analysis (qualitative, quantitative, semi-quantitative, isotope dilution) as well as isotope ratio and time resolved analysis modes. Controlling and Tuning Autotuning of all parameters, including ICP parameters, torch position, lenses and multiplier voltage Fully automated and configurable plasma start and stop sequence Easy autosampler setup using a graphical display of the autosampler Setup of Methods Intuitive and easy selection of target isotopes in periodic table or spreadsheet display modes Automatic, customized isobaric interference correction Spreadsheet style click-and-drag cell fill down Creating and Running Sequences Intuitive and easy creation of sample analysis sequences in graphical or spreadsheet display modes Integrated powerful QA/QC package that meets internationally regulated requirements including US EPA 2.8, 62. The flexible editor included can be used to define specific QA/QC criteria in any laboratory. Remote Control & Diagnostic Displaying Results and Creation Displaying Results and Creation of Reports Real time display of spectra, calibration curves, fully quantitative results and time resolved analyses On-line export of time resolved data in several formats (ASCII, GRAMS, Spectacle, GLITTER, ANDI Xcalibur for further analysis in third party programs Setting & Read-Back of All Instrument Parameters Instrument Startup and Tune The Finnigan ELEMENT2 software is a state-of-the-art, simple to use software suite taking advantage of the reliability and stability of the modern operating system Microsoft Windows XP Professional. The software package is optimized for the needs of the routine analyst, providing stability and ease of use for basic operation of the Finnigan ELEMENT2, and yet retains the flexibility for advanced operation. Due to the use of the Microsoft Windows XP Professional operating system and standardized programming, the data system computer can be easily connected to a network, enabling data transfer and allowing remote control of the Finnigan ELEMENT2.

Product Specification Specifications and Installation Requirements Sensitivity (Concentric Nebulizer) > 1 x 1 9 counts per second (cps)/ppm In Detection Power Dark Noise Dynamic Range Mass Resolution Signal Stability Scan Speed (magnetic) Scan Speed (electric) < 1 ppq for non-interfered nuclides <.2 cps > 1 9 linear with automatic gain calibration 3, 4, 1, (1 % valley, equivalent to 5 % height); 6, 8, 2, (FWHM) < 1 % RSD over 1 minutes < 2 % RSD over 1 hour m/z 7 to 24 to 7 < 15 ms 1 ms/jump, independent of mass range Oxide and Doubly Charged Ions ratio measured BaO + /Ba + <.2 Ba 2+ / Ba + <.3 Power 3-phase, 23/4 V ±1 %, 5/6 Hz fused 32 A per phase Power consumption: ~9 kva Environment Temperature 18 24 C (64 75 F) Humidity 5 6 %, noncondensing, non-corrosive Cooling Water ~ 2 l/h Temperature 1 2 C 4 6 bar (43 65 psi) Argon Purity 99.996 min. 18 L/min Regulated pressure 8 1 bar (116 145 psi) Uninterrupted argon supply recommended Plasma Exhaust 1 x 6 cm Ø; 9 m 3 /h (Argon + suspended sample) Electronics Exhaust 2 x 15 cm Ø; 8 m 3 /h Finnigan ELEMENT2, Footprint and Dimensions (all dimensions in cm) 37 125 13 67 36 141 79 88 63 174 16 141 33

Plasma Capabilities from Thermo Electron In addition to these offices, Thermo Electron Corporation maintains a network of representative organizations throughout the world. The use of an Inductively Coupled Plasma source (ICP) is the accepted and most powerful technique for the analysis and quantification of trace elements in both solid and liquid samples. Its applications range from routine environmental analyses to the materials industry, geological applications to clinical research and from the food industry to the semiconductor industry. Thermo Electron Corporation is the only instrument manufacturer to offer the full range of Inductively Coupled Plasma Spectrometers (ICP-OES, Quadrupole and Sector ICP-MS) to satisfy every aspect of plasma spectrometry from routine to highly demanding research applications. IRIS Intrepid II Develop your lab from the easy-to-use IRIS Intrepid II ICP-OES to the high performance X Series Quadrupole ICP-MS and up to the ultra-sophisticated Finnigan ELEMENT2 and Finnigan NEPTUNE Sector ICP-MS instruments. Each instrument combines leading-edge technology, fit for purpose and affordability with a tradition of quality, longevity, accuracy and ease of use. X Series ICP-MS Australia +61 2 9898 1244 analyze.au@thermo.com Austria +43 1 333 534 analyze.at@thermo.com Belgium +32 2 482 3 3 analyze.be@thermo.com Canada +1 8 532 4752 analyze.ca@thermo.com China +86 1 585 3588 analyze.cn@thermo.com France +33 1 6 92 48 analyze.fr@thermo.com Germany +49 613 48 analyze.de@thermo.com Italy +39 2 95 591 analyze.it@thermo.com Japan +81 45 453 91 analyze.jp@thermo.com Netherlands +31 76 587 98 88 analyze.nl@thermo.com Nordic +46 8 556 468 analyze.se@thermo.com South Africa +27 11 57 184 analyze.sa@thermo.com Spain +34 91 657 493 analyze.es@thermo.com Switzerland +41 61 48784 analyze.ch@thermo.com UK +44 1442 233555 analyze.uk@thermo.com USA +1 8 532 4752 analyze.us@thermo.com www.thermo.com Finnigan ELEMENT2 Finnigan NEPTUNE DIN EN ISO 91:2 Thermo Electron (Bremen) GmbH is certified DIN EN ISO 91:2 24 Thermo Electron Corporation. All rights reserved. Microsoft and Windows are registered trade marks of the Microsoft Corporation. All other trademarks are the property of Thermo Electron Corporation and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. BR327_E 5/4C