Digital and Computed Radiography: Acceptance Testing and Quality Control

Similar documents
INTRODUCTION. A. Purpose

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

Physics testing of image detectors

Quality Control of Full Field Digital Mammography Units

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

Quality Control and Maintenance Programs

Page: 1 of 6 Page: 1 of 6

Digital Mammography Update: Design and Characteristics of Current Systems

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Checklist for the Qualification of Digital Detector Array Systems

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Computed Radiography: Acceptance Testing and Quality Control

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

The Exposure Index and Its Standardization

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

Performance testing for Precision 500D Classical R/F System

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Update on ACR Digital Mammography QC Manual

Digital Breast Tomosynthesis QC Requirements

Dental Radiography Core Subject. Digital Radiography

X-RAY IMAGING Emerging Digital Technology - CMOS Detectors

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Recommendations for a technical quality control program for diagnostic X-ray equipment

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

DICOM Correction Item

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Performance evaluation of computed radiography systems

Overview of Digital Detector Technology

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Quality Assurance of Radiotherapy Equipment

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Bergen Community College Division of Health Professions/ Radiography Program. Course Syllabus and Course of Study for Lecture

Understanding Digital Modalities: System Integration and Use

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

Equipment. Lin muyan. Presented by: CocticaUwititonze Cell:+250 (0)

Digital radiography conquers the veterinary world

AN INNOVATIVE & CUTTING-EDGE DIGITAL RADIOGRAPHY SYSTEM

Using Automatic Exposure Control in Digital Radiography

Fundamentals of Cone-Beam CT Imaging

COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Prepublication Requirements

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Breast MRI Quality Control

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

Accreditation Is Coming

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

Digital Radiography Image Quality: Image Acquisition

Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology

EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

Learning Objectives: Quality Assurance (QA) is. Quality Assurance Procedures for Digital Radiography. Alternate definition of Quality Assurance (QA)

Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition)

QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis

Aspire FCRm/CRm Fujifilm CR for Mammography

MQSA Quality Control Manual for Monochrome Displays for Mammography

Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1

SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration

THE ROYAL AUSTRALIAN AND NEW ZEALAND COLLEGE OF RADIOLOGISTS

Radiation safety in dental radiography

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

Best Practices in Digital Radiography

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Interfacing Fuji CR in a DICOM Network Environment

Chapter 15. Creating the Digital Image. Objectives. Key Terms

XRS-3 X-RAY SOURCE OPERATOR S MANUAL

Analog and Digital Systems of Imaging in Roentgenodiagnostics

XR200 X-RAY SOURCE OPERATOR S MANUAL

Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014

Transcription:

MAHIDOL UNIVERSITY Wisdom of the Land Digital and Computed Radiography: Acceptance Testing and Quality Control Napapong Pongnapang, Ph.D. Department of Radiological Technology Faculty of Medical Technology Mahidol University Bangkok, Thailand

Content Digital Radiography Photostimulable Phosphor (CR) Flat Panel Detectors (DR) Acceptance Testing Guidelines Setting up a Quality Control Program

Computed Radiography

Photostimulable Luminescence

X-Rays Flat Panel Detector

Flat Panel Detectors Direct DR (DDR): Amorphous Selenium Detector matrix of transistors, without photon conversion layer Indirect DR (IDR): Amorphous Silicon TFT or CCD with CsI conversion layer

Flat Panel Detector Configurations Indirect DR Direct DR

Acceptance Testing - Purpose To ensure the machine acquired meets detail specifications under the commercial contract To ensure clinical quality of the machine for optimal patient safety and image quality To set up baseline values for Quality Control (QC) purpose

Acceptance Testing - Process Get list of equipment package (inventory) Check standard acceptance testing procedures specific to each equipment American Association of Physicist in Medicine (AAPM) Institute of Physics and Engineering in Medicine (IPEM) Others Organizations provide standards (IAEA soon) Get Test Tools Phantoms Dosimeter Other measurement equipments

Acceptance Testing - Components X-Ray Tube kvp, Timer Tube Leakage HVL, Linearity, reproducibility CR Plate/Scanner Detectors Patient Dose

Acceptance Testing : X-Ray Machine Use similar methods for general x-ray machines Concerned Points Don t direct expose the detector with high dose! Use Lead apron to block the detector

Acceptance Testing CR Protocols

Acceptance Testing (AAPM) Phosphor Plate Dark Noise System Linearity, Auto-Ranging and Exposure Response Receptor Reproducibility, Density Uniformity & Artifact Analysis Phosphor Plate/Cassette Throughput Laser Beam Function Spatial Resolution Wire Mesh Test - Resolution Uniformity Across Receptor Low Contrast Sensitivity/ Detectability Distance Accuracy Measurements and Aspect Ratio Test Accuracy/Thoroughness of Erasure Cycle

CR Acceptance Test/QC Consideration Image Acquisition Electro-mechanical readout Image processing PACS/RIS interface Image handling

Recommended Acceptance Tests Physical inspection/inventory/pacs interface Image plate uniformity and dark noise Signal response: Linearity and Slope Signal response: Exposure calibration and beam quality Laser beam function High contrast resolution

Recommended Acceptance Tests Noise-Low contrast response Distortion Erasure thoroughness Artifacts analysis Positioning and collimation robustness Image plate throughput

Acceptance Test Tools Required Exposure / dose meters Spatial resolution phantom Low contrast phantom SMPTE pattern Anthropomorphic phantom Document log/spreadsheet/instructions

Phosphor plate dark noise The resultant film (or soft-copy image) for each plate should demonstrate a clear, uniform, artifact-free image. Exposure indicators for automatic processing should indicate no incident exposure. Obvious artifacts, density shading, or uniformities present on any output image should be evaluated further.

System Linearity, Auto-ranging and Exposure Response Make sure linearity response of CR system Test by calibrating 1 mr exposure at specific vendor condition Fuji use S = 200 Semi-Auto mode exposed at 80 kvp @ specified distance Other vendors; refer totg10

S Number - Sensitivity It reflects the center of the usable portion of the histogram Calibration is based on a 1 mr exposure at 80 kvp to the IP. Using a 72 distance through air to achieve an S number of 200 with a fixed latitude of 1.

Receptor Reproducibility, Density Uniformity & Artifact Analysis Make sure uniform image intensity across irradiated field At same exposure and image processing condition, same uniformity is expected No artifacts should be detected.

Phosphor Plate/Cassette Throughput Make sure throughput meets with vendor s specifications Fuji FCR 5000+ = 109 plates/hr

Spatial Resolution Make sure spatial resolution meets specification provided by vendors This test requires same test conditions specified by vendors There are many factors affecting resolution

Wire Mesh Test - Resolution Uniformity Across Receptor Similar to screen-film contact test done in conventional radiography To make sure uniformity across receptor

Wire Mesh Test - Resolution Uniformity Across Receptor

Low Contrast Sensitivity/ Detectability To test low contrast sensitivity of CR to different levels of radiation dose To see effects of SNR levels

Low Contrast UAB Phantom Low contrast at 5 mr 2.9% Low contrast at 1 mr 3.28%

Distance Accuracy Measurements and Aspect Ratio Test To test accuracy of distance measurement Test by imaging a known size object

Distance Accuracy Measurements and Aspect Ratio Test

Accuracy/Thoroughness of Erasure Cycle Make sure that erasure cycle functions well Need very high radiation exposure and dense material

Accuracy/Thoroughness of Erasure Cycle BEFORE AFTER

Acceptance Testing DR Protocols Currently, no published dedicated protocol available IPEM No.91describes extended methods to test Flat panel on top of general x-ray equipment testing AAPM is likely to release TG 151 in 1-2 years

DR Test Protocols What are now available? IPEM AAPM TG 116 for Exposure Index Standards AAPM TG 18 for Display Devices Vendor Specific Test Protocols GE Siemens Philips Cannon Others

Review: Available International/National Standards Deutsches Institut Fur Normung, Germany DIN 6868-58 (2001) Part 58 The Institute of Physics and Engineering in Medicine, UK IPEM Report #91 (2005) KCARE, UK Provides details on how to do acceptance testing for DR

What are recommended? DIN 6868-58 Dynamic Range Limited Spatial resolution Low Contrast Homogeneity Artifacts KCARE DR IPEM Dosimetry Dark Noise Detector dose indicator consistency Uniformity Blurring and stitching artefacts Limiting spatial resolution Threshold Contrast at different dose levels Detector dose indicator monitoring Image uniformity Low contrast sensitivity Limiting spatial resolution Detector dose indicator reproducibility Measured uniformity Uniformity of resolution Threshold contrast detail detectability Scaling error Dark noise

Detector dose indicator monitoring IPEM refers the method to the British Institute of Radiology Basicly, you check if the Detector Dose Indicator corresponds to the dose exposed to the detector AAPM TG 116 provides details on Exposure Index for different vendors

Exposure Index www.philips.com

Exposure Index www.philips.com

Image Uniformity Expose blank detector to see any lines or rectangular areas If any seen, detector needs to be recalibrated

Low contrast sensitivity Test contrast level at different dose to detector Needs test objects such as Leeds TOR and Cu attenuator (or equivalent) Establish baseline for future comparison

Low Contrast UAB Phantom Low contrast at 5 mr 2.9% Low contrast at 1 mr 3.28%

Limiting Spatial Resolution Lead grating resolution bar pattern is recommended Use low tube voltage about 50 kvp without filtration Angle the test pattern at 45 degree Establish baseline for future comparison

Line pairs test pattern

Detector dose indicator repeatability Expose the detector with same dose for 3 times Record the Exposure Index 70 kvp with 1.0 mm Cu attenuator is recommended for dose of 10 ugy

Detector dose indicator reproducibility Similar to previous test but do at different dose levels Can be used as monitoring tool for long term decline of the system

Measured Uniformity Use image from previous test Draw 5 ROIs about 100x100 pixels Calculate SD from five ROIs divide by the mean value

Uniformity of resolution Use fine wire mesh Expose at low tube potential at 50 kvp, SID = 1 meter No filtration added Check for blurred areas

Scaling Error Use highly attenuated objects such as ruler or test pattern with grids of known distance Check for accuracy of distance measurement Repeat with film, if laser printer is used.

Dark Noise Obtain image with no or very low exposure Purpose is to check electronic noise Can also expose the detector using highly radiation attenuated block

Setting up a QC Program for DR

Quality Control Modern equipments have good maintenance program, why need QC? QC = maintain standard For image quality, dose and safety Results of QC can be used to dictate level of action For further tests needed, call engineer, stop scanning Reassurance of ongoing quality, solving clinical image problems, provide evidence to manufacturer

How do we get start? Medical physicist performs acceptance testing Medical physicist sets up QC activities run by QC Technologist/Physicist

IPEM Recommendation: Routine Testing for DR

What is needed? Computer friendly phantom Objective quantitative analysis method System performance tracking and database logs Exposure monitoring tools and database tracking

CR/DR QC Phantom CR/DR EZ - Daily QC Phantom One shot image quality evaluation

CR/DR QC Phantom ACR R/F Accreditation Phantom

Daily QC

Hard copy/display QC SMPTE/TG-18 Test Pattern

Laser Film QC Weekly: View SMPTE pattern Verify gray levels 0/5% & 95/100% patches Film 6 on 1 4 on 1 if necessary Plot OD of 10%, 40% & 90& patches Observe film for artifacts SMPTE Test Pattern

Action Limits SMPTE patch OD Control Limit 0 2.45 0.15 10% 2.10 0.15 40% 1.15 0.15 90% 0.30 0.08

Conclusion Digital Radiography includes PSP, IDR and DDR Acceptance Testing and QC for each type of the technology are different Numbers of standard test protocols are available but not harmonized Acceptance test procedures and set up QC program must be done to ensure the quality of the radiological services provided