Performance evaluation of computed radiography systems
|
|
|
- Colin Lamb
- 10 years ago
- Views:
Transcription
1 Performance evaluation of computed radiography systems Ehsan Samei a) Department of Radiology, Duke University Medical Center, DUMC Box 3302, Durham, North Carolina J. Anthony Seibert Department of Radiology, UC Davis Medical Center, Sacramento, California Charles E. Willis Department of Radiology, Baylor College of Medicine and Edward B. Singleton Diagnostic Imaging Service, Texas Children s Hospital, Houston, Texas Michael J. Flynn Department of Radiology, Henry Ford Health System, Detroit, Michigan Eugene Mah Department of Radiology, Medical University of South Carolina, Charleston, South Carolina Kevin L. Junck Department of Radiology, University of Alabama Medical Center, Birmingham, Alabama Received 19 September 2000; accepted for publication 15 December 2000 Recommended methods to test the performance of computed radiography CR digital radiographic systems have been recently developed by the AAPM Task Group No. 10. Included are tests for dark noise, uniformity, exposure response, laser beam function, spatial resolution, low-contrast resolution, spatial accuracy, erasure thoroughness, and throughput. The recommendations may be used for acceptance testing of new CR devices as well as routine performance evaluation checks of devices in clinical use. The purpose of this short communication is to provide a tabular summary of the tests recommended by the AAPM Task Group, delineate the technical aspects of the tests, suggest quantitative measures of the performance results, and recommend uniform quantitative criteria for the satisfactory performance of CR devices. The applicability of the acceptance criteria is verified by tests performed on CR systems in clinical use at five different institutions. This paper further clarifies the recommendations with respect to the beam filtration to be used for exposure calibration of the system, and the calibration of automatic exposure control systems American Association of Physicists in Medicine. DOI: / Key words: computed radiography, photostimulable phosphor radiography, acceptance testing, quality control, automatic exposure control I. INTRODUCTION Computed radiography CR, scientifically known as photostimulable phosphor radiography, is a digital technology for the acquisition of radiographic images. 1,2 CR is the most common digital radiography modality in radiology departments today, with an estimated 7000 systems in use worldwide. The technology uses a conventional radiographic acquisition geometry to deposit x-ray energy in a photostimulable phosphor screen with delayed luminescence properties. After irradiation, the screen is stimulated by a scanning laser beam, to release the deposited energy in the form of visible light. The released photostimulated light is captured by a light detector, converted to digital signals, and registered with the location on the screen from which it has been released. The digital data are then postprocessed for appropriate presentation, and are sent to a hard-copy printer or a soft-copy display monitor for medical evaluation. Upon installation and prior to clinical use, CR devices should be evaluated for satisfactory performance. 3,4 As of September 2000, there are five manufacturers of CR imaging devices, Agfa Medical Systems Ridgefield Park, NJ, Fuji Medical Systems Stamford, CT, Eastman Kodak Health Imaging Rochester, NY, Konica Imaging Systems Wayne, NJ, and Lumisys, Inc Sunnyvale, CA. There are currently no industry standards for specifying the performance of these TABLE I. CR systems evaluated in this study. Manufacturer CR device Phosphor screen Agfa ADC-70 MD-10 ADC-Compact ADC-Solo Fuji FCR-9501 ST-VA and ST-VN FCR-9501-HQ AC3-CS FCR-5000 ST-VN Kodak CR-400 GP-25 and HR Lumisys ACR-2000 MD Med. Phys. 28 3, March Õ2001Õ28 3 Õ361Õ11Õ$ Am. Assoc. Phys. Med. 361
2 362 Samei et al.: Performance evaluation 362 TABLE II. Testing devices required to perform the acceptance testing of a CR imaging device. Testing device Calibrated x-ray source Calibrated hard/soft-copy display devices Densitometer if a hard-copy display is to be used Copper and aluminum filters Calibrated ion chamber Stand for the ion chamber Screen cleaning solution and cloths Two metric 30 cm steel rulers for laser-beam function and spatial accuracy tests Three sector-type 0.4 line-pair phantoms of up to 5 lp/mm frequency 0.05 mm lead thickness Low-contrast phantom e.g., Leeds TO.12 Screen-contact wire-mesh pattern Screen-contact fine wire-mesh pattern e.g., mammography screen-film contact tool Small lead block 3 mm thick Antiscatter grid 10:1 or 12:1, 103 ln/in. if the x-ray system does not have one Anthropomorphic phantoms foot, hand, pelvis, chest, etc. Timer Measuring tape Flashlight Role of masking tape devices. The lack of uniformity in measurement procedures among different manufacturers has introduced ambiguity in the meaning of the system specifications. For example, different manufacturers calibrate the response of the system to a given exposure value using different beam qualities and report the response using indices which have different dependences on exposure. In a large medical institution in which CR devices of different kinds might be employed, it is important to assure that the patient images are acquired within a certain exposure range to prevent over- and underexposures. However, the lack of calibration uniformity makes the definition of the acceptable exposure ranges from the CR response values cumbersome. In general, in order to achieve a consistent level of clinical performance, acceptance testing should utilize a uniform cross-platform methodology and uniform criteria so that the results of the tests can be correlated with clinical performance standards. Currently, Task Group No. 10 of the American Association of Physicists in Medicine AAPM TG10 5 is making an effort to provide a comprehensive standardized testing protocol testing and quality control of CR systems. In this work, we have used the preliminary guidelines established by the AAPM Task Group to evaluate the performance of CR systems currently in use at different institutions represented by the co-authors. The paper provides a summary of the tests recommended by the AAPM Task Group, delineates the specific technical aspects of the tests, suggests quantitative measures of the performance results, and recommends uniform quantitative criteria for satisfactory performance. The recommendations provided in this paper are a first step toward meeting a need perceived by practicing clinical medical physicists for quantitative guidelines to be used in conjunction with AAPM TG10 recommended testing procedures. TABLE III. Testing protocol and acceptance criteria for the dark noise test. No exposures. Erase a single screen and read it without exposing it. Screen processing System diagnostics/flat field, Test/sensitivity L 1, Pattern Standard fixed EDR S Image postprocessing Linear Raw data and no edge musica parameters 0.0 GA 1.0, GT A, RE 0.0 enhancement settings, Sensitometry linear window 512, level exposure index Measurements to be IgM, average pixel value PV Avg. pixel value PV Exposure index EI, average Average pixel value PV made and its standard deviation PVSD, and its standard deviation pixel value PV, and its standard and standard deviation and scan average level SAL PVSD within 80% deviation PVSD within 80% PVSD within 80% within 80% of the image of the image area of the image area of the image area Uniform image without any artifacts Uniform without any artifacts except Uniform image for collector profile bands in the without any artifacts screen-movement direction IgM 0.28 PV 280 a EI GP 80, EI HR 380 PV 3425 SAL 130 PVSD 4 PV GP 80, PV HR 80 PVSD 4 PV 350 PVSD 4 PVSD 5 a For those systems in which there is a direct relationship between PV and log E. In the case of an inverse relationship, PV should be greater than 744.
3 363 Samei et al.: Performance evaluation 363 TABLE IV. Testing protocol and acceptance criteria for uniformity CR screen test. This test is applied to all the screens. Visually inspect the screens for physical defects. Verify that the cassette label matches the type of screen inside. Expose the screen to 10 mr ( C/kg) a entrance exposure using 80 kvp, 0.5 mm Cu and 1 mm Al filtration, and 180 cm source-to-image distance SID. If significant heel effect is present, test can be performed with two sequential half-exposures between which the orientation of the cassette is reversed. Screen processing System diagnosis/flat field, Test/sensitivity L 1, Semi EDR Pattern Standard Image postprocessing, Musica parameters 0.0 Sensitometry linear Linear GA 1.0, GT A, RE 0.0 Raw data and no edge enhancement settings, window 512, level exposure index Measurements to be made Average pixel value PV and its standard deviation PVSD within 80% of the image area Average pixel value PV and its standard deviation PVSD within 80% of the image area Average pixel value PV and its standard deviation PVSD within 80% of the image area Average pixel value PV and its standard deviation PVSD within 80% of the image area Screen-to-screen variations: Standard deviation of IgM LMSDs, and mean and standard deviation of PV among screens PVs and PVSDs Screen-to-screen variations: Standard deviation/mean sensitivity SD/Ss and standard deviation of average PV among screens PVSDs Screen-to-screen variations: Standard deviation of exposure index among screens EISDs Screen-to-screen variations: Standard deviation of average PV among screens PVSDs Uniform image without any artifacts PVSD 25 single screen LMSDs 0.02 PVSDs 25 PVSD 20 single screen SD/Ss 5% PVSDs 20 PVSD 20 single screen EISDs 20 PVSD 20 single screen PVSDs 20 a Throughout these tables, for convenience, all exposures are expressed in units of mr 1 mr C/kg. II. METHODS AND RECOMMENDATIONS As listed in Table I, CR devices in use at five different institutions from four major CR manufacturers were evaluated. The inventory of equipment used for testing is listed in Table II. Each system was evaluated for dark noise, screen uniformity, exposure indicator calibration, linearity and autoranging response, laser beam function, limiting resolution, noise and low-contrast resolution, spatial accuracy, erasure thoroughness, aliasing and grid response, and throughput. 6 Special attention was paid to applying a uniform testing protocol for different CR systems, following the recommendations of the AAPM TG10 as closely as practicable. The data from different institutions were collected and processed in a single database. Prior to or shortly after the evaluations, each system s performance was judged clinically acceptable by attending radiologists based on image quality of clinical images acquired with the system. Tables III XIII tabulate the testing protocol and the acceptance criteria derived from the results. For a full description of the tests and the rationale for performing each test, the reader is advised to consult the AAPM TG10 report. The quantitative acceptance criteria were established based on the results of the tests performed on the clinical systems and a uniform level of tolerance in system response across different systems. Table XIV tabulates the response tolerance levels based upon which the acceptance criteria were established. These levels were translated to systemspecific parameters, as reported in Tables III XIII, using the response relationships of the systems tabulated in Table XV. of the clinically acceptable systems tested in this collaborative effort generated results beyond the established criteria. In most instances, the acceptance criteria were at least 20% beyond the extremes of the evaluation results, a reasonable margin considering that the evaluated systems were not operating at the borderline of clinical acceptability. Several experimental precautions were observed in the evaluation of the systems. All the phosphor screens were cleaned and erased prior to executing the testing procedures. Consistent delay times between 1 to 15 min were observed between exposing and reading the screens. Care was taken to reduce backscattered radiation by utilizing cross-table exposures and significant interspace behind the screens. A large source-to-image distance SID 180 cm was used to minimize the heel effect. The raw signal values which were proportional to the log of the incident exposure without any postprocessing were used in the evaluations. All exposures were measured in a consistent fashion: The collimators were set to expose the whole cassette with additional 7 cm margins on each side in the direction perpendicular to the anode cathode axis. The ion chamber was then placed at the center of the beam at 2/3 of the SID. The exposure was measured in five consecutive exposures and the values averaged, E 1. Keeping the ion chamber at 2/3 SID, the chamber was shifted on the central axis perpendicu-
4 364 Samei et al.: Performance evaluation 364 TABLE V. Testing protocol and acceptance criteria for exposure indicator calibration. Agfa Fuji Kodak Lumisgys b Recommended exposure condition a Use multiple screens at least three of a given size/type. Expose the screens to approximately 1 mr ( C/kg enhance exposure using 80 kvp and 0.5 mm Cu/1 mm Al filtration. Screens should be read with a precise 10 min delay. Expose a screen to Expose a screen Expose a screen Expose a screen to approximately 8 manufacturer approximately 1 mr to approximately to approximately mr ( C/kg entrance specified a ( C/kg entrance 1 mr ( C/kg 1 mr ( C/kg exposure using 80 kvp with 1 mm exposure using 75 kvp and 1.5 mm Cu filtration. Screen should be read promptly. entrance exposure using 80 kvp without filtration. Screen should be read with a precise 10 min delay. entrance exposure using 80 kvp and 0.5 mm Cu/1 mm Al filtration. Screen should be read with a precise 15 min delay. Cu filtration. Screen should be read promptly. Screen processing System diagnosis/flat field, Test/sensitivity L 1, Pattern Standard semi-edr Image postprocessing, Irrelevant musica parameters 0.0 Measurements to be IgM and IgM normalized Sensitivity and sensitivity Exposure index EI Mean pixel value PV within 80% made to exactly 1 mr exposure normalized to exactly and exposure index of the image area normalized to to the screen (IgM 1mR ) 1 mr exposure normalized to exactly 1 mr exactly 1 mr (PV 1mR ) using IgM 1mR IgM log(exposure), to the screen (S 1mR ) exposure to the screen or 8 mr (PV 8mR ) exposure SAL and SAL normalized using S 1mR S exposure (EI 1mR ) using to the screen using to exactly 1 mr exposure to the screen (SAL 1mR ) using SAL 1mR SAL/ exposure 0.5 EI 1mR EI 1000 log exposure PV 1mR PV 1000 log exposure) PV 8mR PV 1000 log exposure/8) IgM 1mR single screen S 1mR EI 1mR Pv 8mR single screen IgM 1mR for all single screen single screen PV 1mR single screen screens averaged S 1mR EI 1mR PV 1mR for all screens SAL 1mR single screen for all screens averaged for all screens averaged averaged SAL 1mR for all screens averaged a There is currently a strong consensus that CR systems should be calibrated with a standard filtered beam. Until such time as manufacturers change their recommendations, the calibration procedure can be performed both with the manufacturer-defined technique, to verify conformance with the manufacturer s specifications, and with 0.5 mm Cu/1 mm Al filtration and 10 min delay time, for benchmarking and constancy checks. b The Lumisys ACR-2000 software did not make use of an exposure index at the time of testing. The system is calibrated to produce a pixel value of 600 in response to an 8 mr ( C/kg exposure to the screen. lar to the anode cathode axis toward the edge of the field just outside the useful beam area the shadow of the ion chamber was still fully within the beam without projecting over the cassette area. The exposure was measured in five consecutive exposures again and the values were averaged, E 2. The chamber was kept at the second location during the tests for verification of the exposure values. The average exposure to the cassette in each single exposure was calculated as (E 1 /E 2 )(2/3) 2 measured exposure. III. DISCUSSION To achieve a consistent level of clinical performance from CR systems, acceptance testing procedures should be performed according to a uniform cross-platform methodology. As in any medical physics survey, the performance evaluation of a CR system is also more definitive and objective when the evaluation is quantitative and the results are compared against specific quantitative acceptance criteria. In this work, an attempt was made to outline a cross-platform uniform methodology based on the guidelines being developed by the American Association of Physicists in Medicine Task Group 10. Furthermore, a first attempt was made to recommend quantitative acceptance criteria for satisfactory performance of a CR system based on the current state of practice. The criteria were established using uniform tolerance levels and test results acquired from CR systems in clinical use at five different institutions. The user specificity as opposed to the conventional manufacturer specificity of the acceptance criteria suggested in this paper was necessitated by the desired uniformity of the testing procedures. The criteria, however, do not guarantee optimal clinical performance, which may not be ascertained without comprehensive clinical trials.
5 365 Samei et al.: Performance evaluation 365 TABLE VI. Testing protocol and acceptance criteria for linearity and autoranging response. a Use a single screen multiple screens may also be used if the screen-to-screen variations in the previous test were found minimal. Expose the screen to approximately 0.1, 1, and 10 mr ( , , C/kg entrance exposures in a sequence of three exposure-reading cycles using 80 kvp, 0.5 mm Cu and 1 mm Al filtration, and 180 cm SID. Each time read the screen with a consistent delay time. Screen processing System diagnosis/flat field, Test/ave 4.0 Semi-EDR and fixed EDR 200 repeat also with Test/contrast semi-edr and fixed EDR 200 Pattern Standard Image postprocessing, musica parameters 0.0 Linear GA 1.0, GT A, RE 0.0 Raw data and no edge enhancement settings Measurements to be made IgM, average pixel value PV, and scan average level SAL within 80% of the image area. Slopes and correlation coefficients CCs of linear fits to log SAL vs log E, PVvslog E, and IgM vs log E For Semi EDR, correlation coefficient CC of a linear fit to log S vs log E plot. For fixed EDR, avg. pixel value PV within 80% of the image area, slope and correlation coefficient CC of a linear fit to PV vs log E Exposure index EI and avg. pixel value PV within 80% of the image area. Slope and correlation coefficient CC of a linear fit to EI vs log E and PV vs log E plots Mean pixel value PV within 80% of the image area. Slope, intercept, and correlation coefficient CC of a linear fit topvslog E SAL vs exposure on a linear-log plot should result in a straight line For semi-edr, slope and correlation, sensitivity vs exposure on a log log plot should result in a straight line. For fixed EDR, to PV vs exposure on a linear-log plot should result in a straight line The plot of EI and PV vs exposure on a linear-log scale should result in straight lines The plot of PV vs exposure on a linear-log scale should result in a straight line Slope IgM Slope s Slope EI / Slopes/ Slope SAL / Slope PV / Ave 4 b Slope PV / CCs Slope PV / Slope PV / Con. b CCs 0.95 CCs 0.95 CCs 0.95 a If this test is performed with hard copy prints, the relationship between the pixel value PV and optical density OD should be established beforehand using an electronic test pattern. The relationship between OD and PV should then be incorporated as a transformation in the quantitative analysis of the results. b Note that in some Fuji systems, there is an inverse relationship between PV and log E. For those systems, the polarity of the slope in these equations should be reversed. TABLE VII. Testing protocol and acceptance criteria for the laser beam function. Place a steel ruler roughly perpendicular to the laser-scan direction on a screen. Expose the screen to about 5 mr ( C/kg entrance exposure using a 60 kvp beam without any filtration SID 180 cm. Examine the edges of the ruler on the image for laser beam jitters using magnification. Screen processing System diagnosis/flat field, Test/sensitivity Pattern Standard Semi-EDR Image postprocessing, Linear Raw data and no edge musica parameters 0.0 GA 1.0, GT A, RE 0.0 enhancement settings, sensitometry linear window 512, level exposure index Measurements to be made If any jitter is present, jitter dimension using workstation s measurement or ROI tool. Ruler edges should be straight and continuous without any under- or overshoot of the scan lines in light to dark transitions. There should not be more than occasional 1 jitters.
6 366 Samei et al.: Performance evaluation 366 TABLE VIII. Testing protocol and acceptance criteria for the limiting resolution and resolution uniformity. a This test should be done for each type and size of the screens. Use a 60 kvp, unfiltered x-ray beam SID 180 cm. Place three line-pair pattern devices on the cassette, two in orthogonal directions and one at 45. Expose the screen with an exposure of about 5 mr ( C/kg. Also acquire an image of a fine wire mesh e.g., mammography screen film contact test tool in contact with the cassette to examine the consistency of the resolution response across the image. Screen processing System diagnosis/flat field, Test/sensitivity Pattern Standard semi-edr Image postprocessing, Linear GA 1.0, GT A, RE 0.0 Raw data and no edge musica parameters 0.0 enhancement settings, sensitometry linear window 512, level exposure index Measurements to be made Maximum discernible spatial frequencies in the three directions (R hor, R ver, R 45 ) using a magnified 10, narrowly windowed presentation of the images The image of the wire mesh should be uniform without any blurring across the image R hor / f Nyquist 0.9 R ver / f Nyquist 0.9 R 45 /1.41 f Nyquist 0.9 a Note that the spatial resolution response of a CR system can be more comprehensively evaluated by measuring the modulation transfer function MTF of the system Refs. 7 9, TABLE IX. Testing protocol and acceptance criteria for noise and low-contrast resolution. a This test should be done for each type and size of the screens. A low-contrast resolution pattern is used e.g., Leeds TO.12, 75 kvp beam with 1 mm of Cu filtration. For each screen type/size, acquire three images of the low-contrast phantom using 0.1, 1, and 10 mr ( , , C/kg exposures to the screens. Use a constant delay time of 10 min in reading each of the screens. Screen processing System diagnosis/flat field, Test/contrast Pattern Standard Semi-EDR Image postprocessing, Linear GA 1.0, GT A, RE 0.0 Raw data and no edge musica parameters 0.0 enhancement settings, Sensitometry linear window 512, level 4096 EI for GP screens or level 3796 EI for HR screens Measurements to be made Minimum discernible contrast for each object size contrast detail threshold, Standard deviation of pixel value PVSD within a fixed size and location small region of the images, correlation coefficient CC of the linear fit to log PVSD vs log E. b Contrast-detail threshold should be proportionately lower at Contrast-detail threshold Contrast-detail threshold higher exposures. should be proportionately should be proportionately lower at higher exposures, lower at higher exposures. with higher contrast thresholds for standard-resolution screens. CC 0.95 b a Note that the noise response of a CR system can be more comprehensively evaluated by measuring the noise power spectrum NPS and the detective quantum efficiency DQE of the system at different exposure levels Refs. 8 and 9, b The quantitative evaluation is more valid with uniform images acquired for the linearity test Table VI because of the absence of scattering material in the beam. The expected quantitative response is based on the assumption of a logarithmic relationship between pixel value and exposure Table XV.
7 367 Samei et al.: Performance evaluation 367 TABLE X. Testing protocol and acceptance criteria for spatial accuracy. Place a regular wire-mesh screen film contact test tool over cassette. Expose the cassette to about 5 mr ( C/kg entrance exposure using a 60 kvp beam without any filtration SID 180 cm. Repeat the acquisition with two steel rulers in the vertical and the horizontal directions. Screen processing System diagnosis/flat field, Test/contrast Pattern Standard Semi-EDR Image postprocessing Linear GA 1.0, GT A, RE 0.0 Raw data and no edge musica parameters 0.0 enhancement settings, window 512, level EI Measurements to be made acceptance Distances in the orthogonal directions 15 cm minimum length measured using the measurement tool of the workstation. a Grid pattern spacing should be uniform without any distortion across the image. Measured distance should be within 2% of the actual values. a Alternatively, length measurements can be made on a hard-copy film printed in true-size. TABLE XI. Testing protocol and acceptance criteria for erasure thoroughness. Place a thick lead block at the center of a cassette and expose the screen to about 50 mr ( C/kg using a 60 kvp x-ray beam without any filtration SID 180 cm. Read the screen, and expose it a second time to 1 mr ( C/kg entrance exposure without the lead object using the same beam quality collimated in by about 5 cm on each side of the screen. For a quantitative test re-read the screen after the second exposure without exposing it. Screen processing System diagnosis/flat field, Test/sensitivity Pattern Standard semi-edr Image postprocessing, Linear Raw data and No edge Window setting default musica parameters 0.0 GA 1.0, GT A, RE 0.0 enhancement settings, or equivalent Sensitometry linear Window setting default level EI, window setting to 1 log exposure unit Window setting default or equivalent to default or equivalent or equivalent to 1 log exposure unit 1log exposure unit to 1 log exposure unit Measurements to be IgM, average pixel value PV Avg. pixel value PV Exposure index EI, Average pixel value PV made and its standard deviation PVSD, and its standard deviation average pixel Value PV, and standard deviation PVSD and scan average level SAL PVSD within 80% and its standard deviation within 80% of the within 80% of the reread/unexposed image of the reread/unexposed image PVSD within 80% of the reread/unexposed image reread/unexposed image Absence of a ghost image of the lead block from the first exposure in the reexposed image. a,b IgM 0.28 PV 280 c EI GP 80, EI HR 380 PV 3425 SAL 130 PVSD 4 PV GP 80, PV HR 80 PVSD 4 PV 630 PVSD 4 PVSD 5 a In our tests on the ACR-2000 system, the length of the standard erasure cycle was sufficient for exposures up to 32 mr ( C/kg. Higher exposures to the screen required an additional erasure cycle for complete screen erasure. b Note that erasure time in some systems e.g., Agfa is configurable on an exam-by-exam basis. c For those systems in which there is an direct relationship between PV and log E. In the case of inverse relationship, PV should be greater than 744.
8 368 Samei et al.: Performance evaluation 368 TABLE XII. Testing protocol and acceptance criteria for the aliasing/grid response. This test should be performed for each type and size of screens that will be commonly used. Place the screen in a bucky that contains an antiscatter grid so that the grid lines are parallel to the laser-scan direction. Alternatively, a grid may be placed directly on the screen. Make sure the grid movement is disabled. Expose the screen to 1 mr ( C/kg using an 80 kvp beam filtered with 0.5 mm Cu/1 mm Al filter and a SID according to the specification of the grid. Repeat, placing the screen perpendicular to the laser-scan direction. Repeat the exposures with a moving grid. Screen processing System diagnosis/flat field, Test/contrast Pattern Standard Speed class 200 semi-edr Image postprocessing, Linear GA 1.0, GT A, RE 0.0 Raw data and no edge musica parameters 0.0 A narrow window setting enhancement settings, sensitometry linear level EI, a narrow window setting A narrow window setting Measurements to be made Moiré pattern should not be present when the grid lines are perpendicular to the laser-scan direction. For moving grids, no moiré pattern should be apparent when the screen is placed in either direction. a a Moiré patterns caused by display sampling not addressed in this protocol can be distinguished by their changing behavior with changing the magnification of the image on the soft-copy display device. In light of this limitation, the recommended quantitative criteria should only be considered as helpful suggestions that require further clinical validation in the future. Another limitation of the current work is the fact that many of the evaluation procedures were not fully quantitative or can be influenced by the subjectivity of the examiner. The evaluations of limiting resolution and noise performance Tables VIII and IX are two important examples. The resolution tests used do not evaluate the system transfer characteristics but only establish that some modulation can be detected at the limiting frequency. The noise tests subjectively evaluate the contrast-detail characteristics of the system, and the proposed quantitative test does not evaluate the spatial characteristics of image noise. Ideally, the resolution and noise characteristics of a CR system should be more objectively evaluated by measuring the frequency-dependent modulation transfer function, the noise power spectrum, and the detective quantum efficiency of these systems. A number of investigators have been able to successfully and reproducibly characterize the resolution and noise performance of CR systems using these indices, and more recently reproducible measurements have been made in the field. 7,14 However, a routine implementation of these measurements awaits further standardization of measurement methods, and the de- TABLE XIII. Testing protocol and acceptance criteria for the throughput. Expose 4 screens to 80 kvp, 2 mr ( C/kg. Process the screens sequentially without delay. a Screen processing System diagnosis/flat field, Test/contrast Pattern Standard semi-edr Image postprocessing musica parameters typical of those Irrelevant in clinical usage Measurements to be made Time interval t, in minutes between putting the first screen in and the last image appearing on the CR viewing station b Throughput screens/h 60 4/t Throughput should be within 10% of the system s specifications. a The test can be performed multiple times with different size cassettes. b Contribution of the network configuration is not considered.
9 369 Samei et al.: Performance evaluation 369 TABLE XIV. The CR response tolerance levels based upon which the uniform quantitative acceptance criteria were derived using the equations tabulated in Table XV. All signal levels and standard deviations are expressed in terms of corresponding exposure E values deduced from those quantities. Characteristics Quantity of interest Acceptable tolerance Dark noise Average signal and its standard deviation within 80% of the image area E mr (E C/kg E /E 1% Uniformity Signal standard deviation within 80% of the image area, and the standard E 5% deviation of the average screen signal among screens Exposure calibration The exposure indicator response expressed in terms of exposure to1mr E measured 1 10% ( C/kg entrance exposure Linearity and autoranging The slope of the system response expressed in terms of logarithm of Slope 1 10% exposure vs logarithm of actual exposure Correlation coefficient 0.95 Laser beam function Jitter dimension in pixels Occasional jitters 1 pixel Limiting resolution Maximum discernible spatial frequencies of a high-contrast line-pair R hor / f Nyquist 0.9 pattern in two orthogonal and 45 angle directions R ver / f Nyquist 0.9 R 45 /1.41f Nyquist 0.9 Noise and low-contrast A linear fit of system noise expressed in terms of logarithm of Correlation coefficient 0.95 resolution corresponding E /E) to logarithm of actual exposure Spatial accuracy The difference between the measured (d m ) and actual distances (d 0 )in (d m d 0 )/d 0 2% the orthogonal directions Erasure thoroughness Average signal and its standard deviation within 80% of the reread/ E mr unexposed image (E C/kg E /E 1% Aliasing/grid response No quantitative tolerance levels Throughput Measured throughput in screens per hours (T m ) and the specified (T 0 T m )/T 0 10% throughput (T 0 ) velopment of automated commercial QC products. In this study, the exposures for quantitative measurements were made with 0.5 mm copper and 1 mm additive aluminum filtration in the beam. The use of filtration was based on prior studies 10,15,16 indicating that the use of 0.5 mm Cu filter minimizes the dependency of the results on the kvp inaccuracy and on the variations in the x-ray generator type, as the filter attenuates the soft portion of the spectrum, predominantly responsible for tube-to-tube variations Fig. 1. The use of this filtration also makes the spectrum a more accurate representative of primary x rays incident on the detector in clinical situations Fig. 2. The additional post-cu, 1-mmthick Al filter is used to attenuate any potential secondary low-energy x rays generated in the Cu filter. The use of 0.5 mm Cu/1 mm Al filtration, therefore, is advised for checking the consistency of the response in the acceptance testing and annual compliance inspections of CR systems. This paper outlines the steps for only the physical evaluation of CR systems. In a newly installed system, after completion of the physical acceptance testing and prior to a full clinical utilization, the system should also be evaluated for its clinical performance. The appearance of CR images may vary as a function of radiographic technique factors, the specific recipe of image processing parameters applied to the images, and the type and calibration of the display media. The default image processing parameters of the system for various anatomical sites and views e.g., chest PA, chest lateral, chest portable, knee, etc. should be tested and customized by the application specialists of the manufacturer with assistance of the diagnostic medical physicist and under the direction of the radiologist who is ultimately responsible for the clinical acceptability of the images. Using radiographic techniques provided by the manufacturer, images of various anthropomorphic phantoms should be acquired with various combinations of collimation and positioning, utilizing the appropriate prescribed anatomical menus of the system. In each case, the proper processing of the image and the absence of unexpected positioning and collimation errors should be verified. Attending radiologists should be consulted for acceptability of the image processing parameters for each anatomical menu. Since standard anthropomorphic phantoms have a limited ability to represent human anatomy and patient-to-patient variations, the clinical evaluation and cus-
10 370 Samei et al.: Performance evaluation 370 TABLE XV. The relationship between exposure and pixel value/exposure indicator responses of various CR systems. The relationships which were provided by the manufacturers or derived from their literature, were verified against experimental measurements at 80 kvp with 0.5 mm Cu/1 mm Al filtration. In these relationships, PV is the pixel value, E is the exposure in mr, B is the speed class, and L is the latitude of the system. Exposure indicator IgM and scan Sensitivity S Expsoure index EI quantities average level SAL Exposure indicator relationship SAL cBE IgM 2log SAL log cbe c 1.0 for MD10 screens S 200/E EI 1000 log(e) 2000 Pixel value relationships PV 2499 log SAL log cbe 121 a PV (1024/L) (log E log(s/200)) PV 1000 log(e) c 0 c for GP screens c 1.0 for MD10 screens 511 b c for HR screens PV 1000 log 32/E Exposure/reading 75 kvp and 1.5 mm Cu filtration, 80 kvp without filtration, 80 kvp and 0.5 mm Cu/1 mm Al 80 kvp with 1 mm Cu filtration, condition no reading delay 10 min reading delay filtration, 15 min reading delay no reading delay a Using a 12 bit, linear log E data transfer from Agfa QC workstation. b Assuming a direct relationship between exposure and pixel value. tomization of the image processing parameters should include actual clinical images. Care should be taken that in the validation of the system settings, all examinations performed at the facility are represented. The final customized image processing parameters and system settings for different anatomical menus should be loaded into all units from the same manufacturer in place at the institution or associated medical facilities, where the same exam may be performed on different machines, to assure consistency of image presentations. They should also be documented in a list for future reference. Patient dose is one of the important implementation considerations in the use of CR in a traditional film-based radiology department. 17 In screen film radiography, film density is a direct indicator of patient dose. In CR, however, because of the dissociation of the detection and the display functions of the imaging system, optical density can no longer be used as an indicator of the patient dose. In reading a CR screen, almost all CR systems provide an index that reflects the average exposure received by the screen during the image acquisition Table XV. This exposure indicator can be used to define and monitor patient exposures. Based on the manufacturer s recommendations regarding the intrinsic speed of the system and on the applicable standards of practice, the user should establish, monitor, and enforce the acceptable range of exposure indicator values for the clinical operation in the facility. Note, however, that if a filtration other than that suggested by the manufacturer is used for the exposure calibration of the CR system, as suggested previously, the accepted range of exposure indicator values should be derived based on the comparative results of the two filtration conditions. Automatic exposure control AEC is the primary means for controlling patient exposure in general radiography practice. For screen film systems, the AEC is calibrated for consistency in optical density resultant from varying exposure techniques. Because of the dissimilarity between x-ray absorption characteristics and radiographic speed of CR and conventional screen film radiography systems, an AEC calibrated for screen film radiography is unlikely to be suitable for CR usage. 18 For CR usage, the AEC can be calibrated using an approach similar to that for screen film imaging using the exposure indicator value of the system as the target variable to be controlled. The AEC should be adjusted to result an exposure indicator value within a narrow acceptable range 10% 15% when the kvp or phantom thickness is varied within clinical operational limits. It may also be set to provide a constant change in the exposure indicator value FIG. 1. The relative variation in the response of a CR system signal per unit exposure, where the energy of the beam is varied within 80 kvp 10% range, as a function of Cu filtration in the beam for both single phase and high-frequency/constant-potential generator x-ray systems 12 anode angle, 2.6 mm intrinsic Al filtration. The data were generated by a computational model for simulation of the x-ray spectra, filter attenuation, and absorption characteristics of BaFBr 0.85 I 0.15 :Eu phosphor screens 98 mg/cm 2 phosphor coating weight. The model accuracy has been previously verified against experimental measurements Refs. 8, 10, 14. Note that Agfa CR systems use a slightly different phosphor material (Ba 0.86 Sr 0.14 F 1.1 Br 0.84 I 0.06 ) than the one modeled here.
11 371 Samei et al.: Performance evaluation 371 IV. CONCLUSIONS The methods and acceptance criteria for the performance evaluation of CR systems were presented in a comprehensive tabular form for imaging systems from four major CR manufacturers. The materials can be used as a handbook testing and quality control inspection of CR systems to assure the consistency and reliability of their clinical operation. FIG. 2. a The model-calculated primary x-ray spectra emerging from a 0.5 mm Cu filter and 24 cm tissue-equivalent material. The spectra were normalized to have the same total area. b The model-calculated equivalency of the CR signal per unit exposure for various Cu and tissue-equivalent material see Fig. 1 caption. when plus or minus density steps are applied. Because the CR exposure indicator is a quantity derived from analysis of the image histogram, care must be exercised in the selection of phantoms and processing menus. The phantoms should produce image histograms representative of clinical images, not a very trivial requirement. Otherwise, inaccurate exposure indicator values may result, leading to faulty AEC calibration. Further work on AEC calibration methodology for CR is warranted. a Electronic mail: [email protected] 1 R. Schaetzing, B. R. Whiting, A. R. Lubinsky, and J. F. Owen, Digital radiography using storage phosphors, in Digital Imaging in Diagnostic Radiology, edited by J. D. Newall and C. A. Kelsey Churchill Living Stone, 1990, pp M. Sonoda, M. Takano, J. Miyahara, and H. Kato, Computed radiography utilizing scanning laser stimulated luminescence, Radiology 148, J. A. Seibert, Photostimulable phosphor system acceptance testing, in Specification, Acceptance Testing and Quality Control of Diagnostic X-ray Imaging Equipment, edited by J. A. Seibert, G. T. Barnes, and R. G. Gould AIP, New York, 1994, pp C. E. Willis, R. G. Leckie, J. Carter, M. P. Williamson, S. D. Scotti, and G. Norton, Objective measures of quality assurance in a computed radiography-based radiology department, SPIE Med. Imaging 2432, J. A. Seibert et al., Acceptance testing and quality control of photostimulable phosphor imaging systems, Report of the American Association of Physicists in Medicine AAPM Task Group No. 10 unpublished, in the final review process. 6 A. R. Cowen, A. Workman, and J. S. Price, Physical aspects of photostimulable phosphor computed radiography, Br. J. Radiol. 66, E. Samei, M. J. Flynn, and D. A. Reimann, A method for measuring the presampled MTF of digital radiographic systems using an edge test device, Med. Phys. 25, E. Samei and M. J. Flynn, Physical measures of image quality in photostimulable phosphor radiographic systems, SPIE Med. Imaging 3032, J. T. Dobbins III, D. L. Ergun, L. Rutz, D. A. hinshaw, H. Blume, and D. C. Clark, DQE f of four generations of computed radiography acquisition devices, Med. Phys. 22, E. Samei, D. J. Peck, P. L. Rauch, E. Mah, and M. J. Flynn, Exposure calibration of computed radiography imaging systems abstract, Med. Phys. 25, A C. D. Bradford, W. W. Peppler, and J. T. Dobbins III, Performance characteristics of a Kodak computed radiography system, Med. Phys. 26, W. Hillen, U. Schiebel, and T. Zaengel, Imaging performance of a digital storage phosphor system, Med. Phys. 14, C. E. Floyd, H. G. Chotas, J. T. Dobbins III, and C. E. Ravin, Quantitative radiographic imaging using a photostimulable phosphor system, Med. Phys. 17, M. J. Flynn and E. Samei, Experimental comparison of noise and resolution for 2k and 4k storage phosphor radiography systems, Med. Phys. 26, C. E. Willis, J. C. Weiser, R. G. Leckie, J. Romlein, and G. Norton, Optimization and quality control of computed radiography, SPIE Med. Imaging 2164, D. M. Tucker and P. S. Rezentes, The relationship between pixel value and beam quality in photostimulable phosphor imaging, Med. Phys. 24, M. Freedman, E. Pe, S. K. Mun, S. C. B. La, and M. Nelson, The potential for unnecessary patient exposure from the use of storage phosphor imaging systems, SPIE Med. Imaging 1897, C. E. Willis, Computed Radiography: QA/QC, in Practical Digital Imaging and PACS, Medical Physics Monograph No. 28 Medical Physics Publishing, Madison, 1999, pp
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
Physics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
Computed Radiography: Acceptance Testing and Quality Control
Computed Radiography: Acceptance Testing and Quality Control J. Anthony Seibert, Ph.D. University of California Davis Department of Radiology Sacramento, California Introduction CR is the primary means
Quality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
Quality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM [email protected] 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
Page: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries
2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,
Radiography: 2D and 3D Metrics of Performance Towards Quality Index
AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements..
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
The Exposure Index and Its Standardization
The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany [email protected] Content The Exposure Index (EI) What is it? What is it good
Learning Objectives: Quality Assurance (QA) is. Quality Assurance Procedures for Digital Radiography. Alternate definition of Quality Assurance (QA)
Quality Assurance Procedures for Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed
Standard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
Understanding Digital Modalities: System Integration and Use
Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI [email protected] Dicom Digital Imaging and Communications in Medicine Part 3: Information
Digital Mammography Update: Design and Characteristics of Current Systems
: Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
Aspire FCRm/CRm Fujifilm CR for Mammography
: May 2013 Fujifilm Medical Systems U.S.A., Inc. 2 CONTENTS FOREWORD... 5 Introduction to Measurements... 6 S value... 7 Overview of Testing... 8 Section A Getting Started... 9 0 Set-up & Baseline Images...10
Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology
Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.
Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography
Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,
CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
A PACS-Aware DICOM Image Object
Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors
Health NYC Quality Assurance Guidelines for Primary Diagnostic Monitors June 2015 Content I. Overview... 1 II. Definitions... 1 III. Acceptance Testing Requirements... 4 IV. Bi-weekly PDM Testing... 4
Recommendations for a technical quality control program for diagnostic X-ray equipment
Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer
CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
Medical imaging monitors specification guidelines
Medical imaging monitors specification guidelines Document details Contact for enquiries and proposed changes If you have any questions regarding this document or if you have a suggestion for improvements,
Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.
Slide 1 Digital Mammography Detector QC Slide 2 Introduction What is FFDM QC and why is it important? Eric A. Berns, Ph.D. What to know before you start [email protected] Overview and compare QC tests
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
Using Automatic Exposure Control in Digital Radiography
Using Automatic Exposure Control in Digital Radiography A. Kyle Jones, Ph.D. U.T. M. D. Anderson Cancer Center Department of Imaging Physics 2008 AAPM Meeting, Houston, TX Automatic Exposure Control The
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
Moving Forward What does this mean for the Medical Physicist and the Imaging Community?
Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical
Prepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
Digital Radiography Image Quality: Image Acquisition
Digital Radiography Image Quality: Image Acquisition Mark B. Williams, PhD a, Elizabeth A. Krupinski, PhD b, Keith J. Strauss, MS c, William K. Breeden, III, MS d, Mark S. Rzeszotarski, PhD e, Kimberly
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture
Best Practices in Digital Radiography
WHITE PAPER Best Practices in Digital Radiography Tracy L. Herrmann, M.Ed., R.T.(R); Terri L. Fauber, Ed.D., R.T.(R)(M); Julie Gill, Ph.D., R.T.(R)(QM); Colleen Hoffman, R.T.(R)(M)(CT); Denise K. Orth,
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2
Version 3.1 Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 MQSA Quality Control Manual for Monochrome Monitors for Mammography Foreword Purpose of this Document This document is for
Annual MQSA Inspection Questions NMQAAC April 19, 2004
1 1 1 1 1 1 1 1 0 1 0 1 0 1 0 1 Annual MQSA Inspection Questions NMQAAC April 1, 00 1.0 Inspection Information 1.1 Name and Address 1. Equipment Registration.0 Facility Inspection Download.0 Facility Inspections.0
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography
Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It is intended to replace film/screen
Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices
IT Guidance Documents Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists
WHERE IN THE WORLD JILL LIPOTI?
WHERE IN THE WORLD IS JILL LIPOTI? HELLO FROM NEW JERSEY CRCPD - National Symposium on Fusion Imaging and Multimodalities February 18-20, 2004 Kansas City, Missouri New Jersey s Requirements As They Pertain
Diagnostic Exposure Tracking in the Medical Record
Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY
AAPM REPORT NO. 74 QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY Report of Task Group #12 Diagnostic X-ray Imaging Committee Members S. Jeff Shepard, Chairman Pei-Jan Paul Lin, Co-Chairman John M. Boone Dianna
Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
Digital radiography conquers the veterinary world
Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing
Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18046 Computed Tomography Resolution
MQSA Quality Control Manual for Monochrome Displays for Mammography
MQSA Quality Control Manual for Monochrome Displays for Mammography Ver1.0 Applicable Models LMD-DM50, LMD-DM30 Sony Corporation - 1 - MQSA Quality Control Manual for Monochrome Displays for Mammography
EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography
Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information
Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology
Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology Guidelines on education and training for digital radiology Author:
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X
DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General
Checklist for the Qualification of Digital Detector Array Systems
Paper No. 010-12 Checklist for the Qualification of Digital Detector Array Systems 1 February 2012 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction
ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations
Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Alzen,, G., Beiderwellen,, K., Volz,, M., Berthold,, L.D.,
Bergen Community College Division of Health Professions/ Radiography Program. Course Syllabus and Course of Study for Lecture
Course Information: Course title: Advanced Imaging Equipment and Patient Care Course Code: RAD 184 Credits: 2.0 (30 lecture hours) Pre-requisites: RAD 276 - Imaging Equipment RAD 281 - Radiography II RAD
Digital Breast Tomosynthesis QC Requirements
Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,
Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules
High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules Abstract J.L. Crozier, E.E. van Dyk, F.J. Vorster Nelson Mandela Metropolitan University Electroluminescence (EL) is a useful
Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition)
Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition) New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp
Overview of Digital Detector Technology
Overview of Digital Detector Technology J. Anthony Seibert, Ph.D. Department of Radiology University of California, Davis Disclosure Member (uncompensated) Barco-Voxar Medical Advisory Board ALARA (CR
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
portable x-ray survey report
department of HealtH and Human services centers for medicare & medicaid services form approved omb no. 0938-0027 portable x-ray survey report provider number H1 date surveyed (1) initial (2) survey W H2
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
Routine Ultrasound Equipment Tests for Quality Assurance. Contents
Routine Ultrasound Equipment Tests for Quality Assurance James A. Zagzebski, Ph.D. Nick Rupert, M.S. Ryan DeWall, M.S. Tomy Varghese, Ph.D. Dept. of Medical Physics University of Wisconsin, Madison, WI,
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction
Agency for Health Care Administration
Page 1 of 20 FED - H0000 - INITIAL COMMENTS Title INITIAL COMMENTS CFR Type Memo Tag FED - H0005 - COMPLIANCE WITH FED/STATE/LOCAL LAWS Title COMPLIANCE WITH FED/STATE/LOCAL LAWS CFR 486.100 Type Condition
Update on ACR Digital Mammography QC Manual
Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology
Training needs for professionals in conventional radiology (radiology technicians, physicists, radiologists) joining digital radiology S Peer 1, R Peer 1, P Torbica 1, W Jaschke 1, E Vaño 2 Innsbruck University
Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014
Multi-Channel Radiochromic Film Dosimetry Adapted from A.Micke Spain, April 2014 Single Channel Film Dosimetry Calibration Curve X=RR ave = R ave (D) D R = D R (R ave ) Color channels X=RGB D X = D( X
Standards and Recommendations for Color Medical Displays
Summit on Color in Medical Imaging 2013 Standards and Recommendations for Color Medical Displays A Review of Current Activities at AAPM and IEC Aldo Badano Division of Imaging and Applied Mathematics Office
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
DICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes
EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams
EPEX Direct-to-Digital System for General Radiography Featuring... Versatility for a complete range of general radiographic exams The Hologic EPEX General Radiography System is a direct-to-digital, fully
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination
IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.
IAC Ch 41, p.1 641 41.7 (136C) X-ray machines used for stereotactically guided breast biopsy. 41.7(1) Definitions. In addition to the definitions provided in rules 641 38.2(136C), 641 40.2(136C), and 641
Module 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
