Using Automatic Exposure Control in Digital Radiography
|
|
|
- Regina O’Brien’
- 9 years ago
- Views:
Transcription
1 Using Automatic Exposure Control in Digital Radiography A. Kyle Jones, Ph.D. U.T. M. D. Anderson Cancer Center Department of Imaging Physics 2008 AAPM Meeting, Houston, TX Automatic Exposure Control The purpose of AEC is to deliver consistent, reproducible exposures across a wide range of anatomical thicknesses, tube potentials, and users Detectors used in AEC systems include fluorescent screens with PMTs/photodiodes, ionization chambers, and possibly solid state detectors AEC System Diagram Generator Backup Reference Voltage AEC Controller Terminate? kvp Amplifier Comparator Loosely based on Bushberg, Seibert, Leidholdt, and Boone, The Essential Physics of Medical Imaging. Fundamental AEC performance characteristics Initial acceptance testing Sensor selector/location Density correction Screen sensitivity adjustment Sensitivity vs. speed setting Reproducibility AEC balance/field sensitivity matching AEC sensitivity Patient thickness tracking kvp tracking Beam quality correction curve Backup timer Cell mapping Minimal exposure duration Field size tracking* Ongoing QC testing AEC sensitivity Density correction kvp tracking Patient thickness tracking Reproducibility AEC balance Backup timer Minimum exposure duration Field size tracking* Remember this Test your system at the SID for which the grid is focused and your system calibrated AEC detectors themselves have an inherent energy dependence Slight variations do exist between cells and should be evaluated upon acceptance testing True vignette: Siemens Sireskop system with reciprocating grid. AEC balance consistently failed when tested at an SID of 100 cm. Moving the tube to an SID of 115 cm resulted in passing test. Cause: Grid reciprocation delay was set to result in appropriately exposed images at an SID of 115 cm, where Siemens calibrates. We were able to adjust the delay to balance the AEC cells within 10% and provide acceptable image quality at 100 cm SID. 1
2 Inter-cell differences exist Density control Unit L+R C kvp Unit kvp Right Left Center Adjusts mas upward or downward in increments of 25-30% per step Some newer DR systems do not incorporate this feature Published guidelines/recommendations Tested for proper operation and similar step size (NCRP 99) 0.15 to 0.30 OD per step (AAPM 74) 4-step controller should adjust by 20-25% 25% per step (AAPM 14) Fundamental differences from S/F: Absence of feature on some systems Curious that some systems still include this feature Air kerma vs. sensitivity selection Many DR systems feature selectable sensitivity settings for imaging protocols Refers to AEC, not digital detector These settings adjust the sensitivity of the AEC cells resulting in higher or lower image receptor exposures Akin to using different speed screen/film combinations for different body regions e.g. 200 speed for chest and 400 speed for general While no regulations or guidelines exist, it is prudent to verify that exposure varies logically and predictably with selection Inversely proportional Air kerma vs. sensitivity Fundamental differences from S/F: Similar to screen selector setting Philips DR system Phil Rauch Resolution of EI is not sufficient to identify changes in sensitivity setting mas vs. Sensitivity setting mas PMMA Aluminum Sensitivity setting mas S = mas S Sensitivity 2
3 AEC Balance Common AEC balance schemes Field sensitivity matching To achieve consistent exposures, AEC cells should be balanced Manufacturers of x-ray x systems use several different schemes for AEC balancing Thus, a one-size size-fits-all test will not be valid if you work with a variety of systems Fundamental difference from S/F: Some manufacturers have changed schemes AEC Balance Test AEC balance during acceptance testing to set a baseline standard for balance Ask your service engineer about the calibration/balancing procedure Many manufacturers today use AEC systems that are not serviceable e for balance they can only be replaced In this case you are stuck with manufacturer s s balance scheme Other manufacturers still have tunable AEC cells Pots in generator Software interface Pots in detector housing Published guidelines/recommendations ± 5% across all combinations (AAPM 14) AEC Balance Cells must also be matched when used in combinations AEC systems may terminate the exposure when the most sensitive cell reaches the required current level, or may average the signal between the cells being used Same criteria apply Unit 1 Unit 2 Unit 3 C L R L+R L+C R+C See also X-ray generator and automatic exposure control device acceptance testing by Raymond P. Rossi, M.S., in Specification, Acceptance Testing and Quality Control of Diagnostic X-ray Imaging Equipment, Proceedings of the 1991 AAPM Summer School All In the case of a system malfunction or technical error, the exposure must be terminated after a certain period of time or delivered mas 600 mas 51 kvp 2,000 mas < 51 kvp 21CFR1020 Many digital radiography systems have preset time limits based on kvp and ma settings, and some will terminate if no signal is detected Backup timer Reproducibility K a delivered by the AEC system should be reproducible Published guidelines/recommendations COV < 0.05 (AAPM 14) ± 5% of average (NCRP 99) I ll leave it as an exercise to prove that these guidelines do not say the same thing 3
4 Minimum exposure duration Thin or less dense body parts can lead to very short exposure times for AEC-controlled radiography E.g. chest imaging, small patients AEC systems should be capable of delivering appropriate exposures at these exposure times Published guidelines/recommendations T min < 1/60 sec or 5 mas, whichever is greater (21CFR1020) Minimum exposure duration - DR However, for DR systems, some of which require less exposure than film/screen systems, this limit is insufficient Typical PA chest x-ray x at 125 kvp and 320 ma: ms Most measurement equipment is incapable of measuring these extremely short times kvp divider Spinning top Oscilloscopes Acquire a series of images with a phantom that yields clinically relevant exposure times (3-10 msec) Examine exposure time in DICOM header Measure in a uniform portion of the image COV < 0.05 Also, it is very important to find the minimum response time (MRT) of your AEC/generator and ensure that your AEC exposure times do not fall l below this time AEC sensitivity Typically referenced to the center cell Manufacturer has calibrated the AEC system to the air kerma they believe their detector needs You may need to calibrate the AEC system to the air kerma you and your radiologists know the detector needs AEC Sensitivity AEC system must be calibrated to deliver the necessary but sufficient K a to the image receptor Relatively simple task with screen/film imaging achieve O.D. in linear portion of H+D curve (~ 1.4) Digital imaging is not contrast-limited, but noise limited How can you set up your AEC system to deliver the necessary amount of noise in an image? Fundamental difference from S/F: Wide range of K a will yield usable images, must decide on acceptable noise level in images Rong XJ, Shaw CC, Liu X, Lemacks MR, Thompson SK, Comparison of an amorphous silicon/cesium iodide flat-panel digital chest radiography system with screen/film and computed radiography systems A contrast-detail phantom study, Med. Phys. 28(11): , Use of scored images of contrast-detail phantom (CDRAD) under clinical image processing conditions to attempt to determine detector exposure required to achieve similar detectability for small low-contrast targets Huda W, Slone RM, Belden CJ, Williams JL, Cumming WA, Palmer CK, Mottle on computed radiographs of the chest in pediatric patients, Med. Phys. 199: , Use of a rating system based on degrees of mottle to determine what sensitivity number of a Fuji AC series would be needed to match the mottle in a 600-speed s/f system, then use the actual sensitivity numbers measured to compare CR to s/f. Conclusion: 0.43 mr required to achieve similar LCD for 0.5 mm object compared to 2.09 mr for the same screen-film image (21%) Conclusion: To achieve comparable mottle to 600-speed s/f system a twofold increase in exposure is needed as compared to mean sensitivity number. An exposure consistent with a 200-speed s/f system would be needed to achieve negligible mottle. 4
5 Liu X and Shaw, CC, a-si:h/csi(tl) flat-panel versus computed radiography for chest imaging applications: image quality metrics measurement, Med. Phys. 31(1):98-110, Compare indirect digital radiography and cassette-based digital radiography systems in terms of fundamental image quality metrics. Give radiologists a visual idea of what the impact of dose to the image receptor is on image quality CDRAD phantom exposed under PMMA Conclusion: Indirect digital radiography detector has significantly higher DQE and lower NPS than cassette-based digital radiography detector. The magnitude of the differences is illustrated in the graphs from the manuscript. Can also be done with other phantoms ACR R/F phantom, anthropomorphic phantom, or even add noise to a patient image for comparison Caveats Image processing has a large impact on noise (low-contrast detectability) and high-contrast resolution, and thus AEC sensitivity should be configured with this in mind. Also, pixel size has an impact on noise in images this is especially important for digital receptors where pixel size is variable, such as PSP systems I have a confession to make I have been using a very simplistic method of calibrating our AEC systems for sensitivity (cassette- based) Eight inch PMMA phantom imaged using 80 kvp Processed with Sensitivity S number used as indicator of proper calibration Properly calibrated reader DR compare with acceptance testing data It ain t that easy Methods for calibrating sensitivity Noise-based method K a -based methods Use a CR cassette with a cutout for a detector Solid-state detector behind grid Pre-detector K a and primary transmission through grid * Use an exposure indicator (EI) A word on exposure indicators Sometimes the relationship between EI and detector exposure is well understood or intuitive Cassette-based digital radiography Sometimes the relationship between EI and detector exposure is not obvious/proprietary Exposure indicator must be verified Goldman AAPM Summer School Doyle, P and Martin, CJ, Calibrating automatic exposure control devices for digital radiography, Phys. Med. Biol. 51: ,
6 Noise-based AEC calibration For each kvp, an S number can be associated with the standard deviation of pixel values that produce the desired noise characteristics STDEV is proportional to the square root of the x-ray x fluence absorbed in the IP (after correcting for structured noise) Recommend use of an -based threshold for choosing noise level Example: threshold of 5 Signal difference of 50 can be seen with σ = 10 Still some determination to be done Setting up AEC in Semi-automatic EDR mode yields clinically valid results in Automatic EDR mode Christodoulou, EG, Goodsitt, MG, Chan, H, and Hepburn, T, Phototimer setup for CR imaging, Med. Phys. 27: , K a -based methods CR cassette with cutout S = 200/E, E = exposure (mr( mr) ) for Fuji CR Cutout machined into IP cassette to accept 15 cm pancake ionization chamber Introduces realistic scatter from cassette, overcomes cassette sensing Doyle P, Gentle D, Martin CJ, Optimising automatic exopsure control in computed radiography and the impact on patient dose, Rad. Prot. Dosim. 114:236-39, K a -based methods Solid-state dosimeter with lead backing should be used to eliminate the effect of backscatter Not possible for all DR systems ATION%20DETECTORS_files/MOSFET.jpg Pre-detector K a Simple measurement to make Scatter avoidable Necessitates measurement of primary transmission of anti-scatter grid Some inaccuracy introduced due to beam hardening by grid which is not accounted for in this measurement Doyle P, and Martin CJ, Calibrating automatic exopsure control devices for digital radiography, Phys. Med. Biol. 51: , Measuring grid characteristics 180 cm focused grid measurement Anti-scatter grids are labeled with information such as: Line rate Grid ratio Interspacer material Focus distance They are not, however, labeled with other information Contrast improvement factor Primary attenuation Fetterly/Schueler poster SU-GG GG-I-153 Chamber 1 Grid Patient-equivalent phantom Detector assembly Chamber 2 6
7 Measuring the Bucky factor Rear chamber behind grid Entire setup Typical exposure readings Measuring for clinical grid A level grid is imperative Measurement w/o grid 100 cm focused grid measurement Table Chamber 1 Support for grid Chamber centered in field Patient-equivalent phantom Grid Foam support Chamber 2 Wood support Table Detector assembly Centering and leveling of grid is imperative Complete setup EI-based methods Once the EI has been verified to be accurate across the range of kvps and patient thicknesses used and seen clinically, it should be used to perform these tests EI verification involves some of the same skills and measurements discussed here More from Jeff Shepard coming up You re not done yet 7
8 The Problem Many of our AEC systems have been calibrated for use with screen-film systems The energy response of gadolinium oxysulfide (Gd 2 O 2 S) screens is substantially different from that of image receptors used in digital radiography Thus, to properly expose digital radiographs, we must recalculate the kvp correction curve for our AEC systems to respond correctly considering the image receptor characteristics Absorption cross section (cm 2 /g) 1.00E E E E+01 CsI Gd2O2S BaFBrI The Solution Beam-quality dependent calibration curves for AEC systems used for digital radiography 1.00E E Photon energy (kev) Doyle and Martin have calculated theoretical kvp correction curves for both CR and IDR detectors Also note that the addition of small amounts of Cu filtration does not significantly affect the calibration Pixel value kvp Tracking Unit # Noise/ Pixel value Noise kvp Doyle, P and Martin, CJ, Calibrating automatic exposure control devices for digital radiography, Phys. Med. Biol. 51: ,
9 Thickness tracking kvp Tracking - Unit #2 Patient thickness tracking Unit 1 Unit Pixel value Noise/ Pixel value Noise Unit Unit kvp PMMA (in.) Patient thickness tracking not only adjusts exposure time but also involves beam quality dependencies of detector absorption efficiencies and AEC sensitivity as well as scatter absorption by the AEC system and the contrast improvement factor of the grid Thickness tracking The overall variation I have seen when kvp tracking and thickness tracking are combined is 25-30% for one vendor (60 kvp) and 15% for another vendor When considering the combined impact of kvp and thickness variation, there is likely to be less variation for digital radiography systems () than in S/F (O.D.) due to the fact that the variation in Bucky factor over a range of kvp s no longer matters. Methods to attempt to improve on patient thickness tracking might include Further tightening kvp correction curve Calibrating with and using thin metal filters What about phantoms? Simple, uniform phantoms are desired for AEC testing Also, consulting physicists probably don t want to lug around 65 lbs. of PMMA and engineers certainly won t either But, what about realistic patient scatter Phantoms for AEC calibration Moral of the story: PMMA, water, and aluminum, in the appropriate amounts, deliver similar transmitted spectra to the AEC system and digital receptor So did 0.2 mm of Cu, but fluence was too high to achieve reasonable exposure times. 2 mm of Cu significantly altered the transmitted spectrum Introduction of scatter is a different situation Scatter and AEC calibration Scatter does influence absolute calibration of AEC sensitivity, however, but does not have a large impact on the kvp correction curve Many manufacturers do adjust the sensitivity of the AEC system based on the presence/absence of the anti-scatter grid Doyle, P and Martin, CJ, Calibrating automatic exposure control devices for digital radiography, Phys. Med. Biol. 51: ,
10 Future of Automatic Exposure Control in Radiography Fluoroscopy/angiography has already altered the way ADRC is performed Monitoring signal value in a region of pixels Detector is read many times per second These methods with TFT arrays do not work in radiography Detector readout is destructive Fischer stereo unit CCD array (Tony Siebert) CMOS (Tony Siebert) Pixel architecture, with multiple transistors, allows for sampling without destroying the contained information US Patent Thevenin B, Glasser F, Martin J-L, J Method and device for the taking of digital images with control and optimization of the exposure time of the object to X or g radiation. Sobol W CCD array with two classes of pixels Integrator/comparator Image corrected later Other resources Goldman LW, Yester MV, Specifications, Performance Evaluations, and Quality Assurance of Radiographic and Fluoroscopic Systems in the Digital Era, 2004 Summer School Proceedings Sobol WT, Advances in and specifications for radiographic X-ray systems, pg Goldman LW, Speed values, AEC performance evaluation, and quality control with digital receptors, pg
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
Physics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
Page: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
Digital Mammography Update: Design and Characteristics of Current Systems
: Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries
2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,
Quality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM [email protected] 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
Quality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination
Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography
Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography
The Exposure Index and Its Standardization
The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany [email protected] Content The Exposure Index (EI) What is it? What is it good
Computed Radiography: Acceptance Testing and Quality Control
Computed Radiography: Acceptance Testing and Quality Control J. Anthony Seibert, Ph.D. University of California Davis Department of Radiology Sacramento, California Introduction CR is the primary means
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department
Digital Radiography Image Quality: Image Acquisition
Digital Radiography Image Quality: Image Acquisition Mark B. Williams, PhD a, Elizabeth A. Krupinski, PhD b, Keith J. Strauss, MS c, William K. Breeden, III, MS d, Mark S. Rzeszotarski, PhD e, Kimberly
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)
IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization
Moving Forward What does this mean for the Medical Physicist and the Imaging Community?
Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1
Running head: COMPARISON OF AMORPHOUS SELENIUM AND AMORPHOUS 1 Comparison of Amorphous Selenium and Amorphous Silicon Flat-Panel Detectors November 20, 2012 COMPARISON OF FLAT-PANEL DETECTORS 2 Abstract
Radiography: 2D and 3D Metrics of Performance Towards Quality Index
AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements..
Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.
Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D. Introduction Screen-film image receptors have been the standard detector used in conventional
Standard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
Update on ACR Digital Mammography QC Manual
Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual
EPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams
EPEX Direct-to-Digital System for General Radiography Featuring... Versatility for a complete range of general radiographic exams The Hologic EPEX General Radiography System is a direct-to-digital, fully
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial
Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY [email protected]
Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology
Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment
PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X
DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General
Radiation Safety Characteristics of the NOMAD Portable X-ray System
Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950
HIGH PERFORMANCE MOBILE SURGICAL C-ARM 1K x 1k CCD Digital Camera System H.F. GENERATOR & ROTATING ANODE TUBE WITH DIGITAL WORKSTATION DESCRIPTION: Mobile Surgical C-arm systems are integrated with a triple
X-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
Performance evaluation of computed radiography systems
Performance evaluation of computed radiography systems Ehsan Samei a) Department of Radiology, Duke University Medical Center, DUMC Box 3302, Durham, North Carolina 27710 J. Anthony Seibert Department
Digital Breast Tomosynthesis QC Requirements
Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist
Recommendations for a technical quality control program for diagnostic X-ray equipment
Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer
QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY
AAPM REPORT NO. 74 QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY Report of Task Group #12 Diagnostic X-ray Imaging Committee Members S. Jeff Shepard, Chairman Pei-Jan Paul Lin, Co-Chairman John M. Boone Dianna
Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes
Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Radiographic Image Production X-ray photons emitted from the x-ray tube interact with the body, exit the patient (exit beam) and interact
Learning Objectives: Quality Assurance (QA) is. Quality Assurance Procedures for Digital Radiography. Alternate definition of Quality Assurance (QA)
Quality Assurance Procedures for Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas
Aspire FCRm/CRm Fujifilm CR for Mammography
: May 2013 Fujifilm Medical Systems U.S.A., Inc. 2 CONTENTS FOREWORD... 5 Introduction to Measurements... 6 S value... 7 Overview of Testing... 8 Section A Getting Started... 9 0 Set-up & Baseline Images...10
Best Practices in Digital Radiography
WHITE PAPER Best Practices in Digital Radiography Tracy L. Herrmann, M.Ed., R.T.(R); Terri L. Fauber, Ed.D., R.T.(R)(M); Julie Gill, Ph.D., R.T.(R)(QM); Colleen Hoffman, R.T.(R)(M)(CT); Denise K. Orth,
Radiation safety in dental radiography
Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum
Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations
Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Alzen,, G., Beiderwellen,, K., Volz,, M., Berthold,, L.D.,
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
Characterizing Digital Cameras with the Photon Transfer Curve
Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently
Modern Fluoroscopy Imaging Systems
Modern Fluoroscopy Imaging Systems Eric Gingold, PhD Thomas Jefferson University, Philadelphia, PA SUMMARY Fluoroscopy, or real-time projection X-ray imaging, has been in clinical use since shortly after
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
Annual MQSA Inspection Questions NMQAAC April 19, 2004
1 1 1 1 1 1 1 1 0 1 0 1 0 1 0 1 Annual MQSA Inspection Questions NMQAAC April 1, 00 1.0 Inspection Information 1.1 Name and Address 1. Equipment Registration.0 Facility Inspection Download.0 Facility Inspections.0
Understanding Digital Modalities: System Integration and Use
Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI [email protected] Dicom Digital Imaging and Communications in Medicine Part 3: Information
Fundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.
Slide 1 Digital Mammography Detector QC Slide 2 Introduction What is FFDM QC and why is it important? Eric A. Berns, Ph.D. What to know before you start [email protected] Overview and compare QC tests
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
Diagnostic X-ray facilities as per quality control performances in Tanzania
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 4, FALL 2006 Diagnostic X-ray facilities as per quality control performances in Tanzania Yesaya Y. Sungita, 1 Simon S.L. Mdoe, 1 and Peter
Overview of Digital Detector Technology
Overview of Digital Detector Technology J. Anthony Seibert, Ph.D. Department of Radiology University of California, Davis Disclosure Member (uncompensated) Barco-Voxar Medical Advisory Board ALARA (CR
Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
Prepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
Digital radiography has become the norm in U.S.
Expanded content at www.asrt.org/publications. Image Gently: Using Exposure Indicators To Improve Pediatric Digital Radiography Quentin T Moore, MPH, R.T.(R)(T)(QM) Steven Don, MD Marilyn J Goske, MD,
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
Atomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
X-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
Equipment. Lin muyan. Presented by: CocticaUwititonze Email: [email protected] Cell:+250 (0)7888 44764
Digital Medical Radiography Diagnostic (DR) system X-ray Equipment Presented by: CocticaUwititonze Email: [email protected] Cell:+250 (0)7888 44764 Lin muyan 1. Goals 2. Introduction 3. X-ray machine
ELECTRON SPIN RESONANCE Last Revised: July 2007
QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational
Production of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
ACR AAPM SIIM PRACTICE PARAMETER FOR DETERMINANTS OF IMAGE QUALITY IN DIGITAL MAMMOGRAPHY
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose
AAPM 2005-Continuing Education Course in Radiographic and Fluoroscopy Physics and Technology A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose Ralph Schaetzing,
Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL
Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp
Digital radiography conquers the veterinary world
Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing
COMPUTED RADIOGRAPHY (CR) AND DIGITAL RADIOGRAPHY (DR) STATE X-RAY INSPECTION PROTOCOL
Publication No. E-10-2 Available online only COMPUTED RADIOGRAPHY (CR) AND DIGITAL RADIOGRAPHY (DR) STATE X-RAY INSPECTION PROTOCOL January 2010 Published by Conference of Radiation Control Program Directors,
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It is intended to replace film/screen
Performance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
EDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY
BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY N. GEERAERT a,b,c, R. KLAUSZ a, S. MULLER a, I. BLOCH c, H. BOSMANS b a GE Healthcare, Buc, France b Department
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric
MATRIX TECHNICAL NOTES
200 WOOD AVENUE, MIDDLESEX, NJ 08846 PHONE (732) 469-9510 FAX (732) 469-0418 MATRIX TECHNICAL NOTES MTN-107 TEST SETUP FOR THE MEASUREMENT OF X-MOD, CTB, AND CSO USING A MEAN SQUARE CIRCUIT AS A DETECTOR
Breast MRI Quality Control
Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Discuss the importance of breast MRI quality control (QC). Provide an overview of the new ACR Breast MRI Accreditation
Improving digital image quality for larger patient sizes without compromises
Improving digital image quality for larger patient sizes without compromises Christiaan Kok, Technology Physicist Grids Smit Röntgen, Best, The Netherlands 7.5 7 6.5 fiber grids: 25 lp/cm with variable
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation
APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS
APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS Selection and use of Ultrasonic Ceramic Transducers The purpose of this application note is to aid the user in the selection and application of the Ultrasonic
