Dose Distribution/Scatter Analysis/Dose calculations. The Physics of Radiation Therapy By Faiz Khan (Chapter 9 )

Similar documents
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Chapter XIII. PDD TAR TMR Dose Free Space

1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: kv range

Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline

ACCELERATORS AND MEDICAL PHYSICS 2

Clinical Photon Beams

Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Production of X-rays and Interactions of X-rays with Matter

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Introduction to the Monte Carlo method

Advanced variance reduction techniques applied to Monte Carlo simulation of linacs

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS

Quality Assurance of Radiotherapy Equipment

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY

Principles of radiation therapy

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector

Lecture 2 Macroscopic Interactions Neutron Interactions and Applications Spring 2010

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

Describing Sound Waves. Period. Frequency. Parameters used to completely characterize a sound wave. Chapter 3. Period Frequency Amplitude Power

QA of intensity-modulated beams using dynamic MLC log files

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Challenges in small field MV photon dosimetry

Atomic and Nuclear Physics Laboratory (Physics 4780)

Variance reduction techniques used in BEAMnrc

Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers

Chapter 9 CALIBRATION OF PHOTON AND ELECTRON BEAMS

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

External dosimetry Dosimetry in new radiotherapeutic techniques

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.

Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY

ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

1. Provide clinical training in radiation oncology physics within a structured clinical environment.

TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS

Measurement of Length, Mass, Volume and Density

Chapter 5 TREATMENT MACHINES FOR EXTERNAL BEAM RADIOTHERAPY

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y

Key words: treatment planning, quality assurance, 3D treatment planning

Assessing Radiation Dose: How to Do It Right

Introduction. Chapter 15 Radiation Protection. Regulatory bodies. Dose Equivalent. Regulatory bodies. Main Principles of Radiation Protection

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

TEACHING RADIOTHERAPY PHYSICS CONCEPTS USING SIMULATION: EXPERIENCE WITH STUDENT RADIOGRAPHERS IN LIVERPOOL, UK

INTRODUCTION. A. Purpose

What is the Radiotherapy Quality Control Program (PQRT) of the National Cancer Institute Rio de Janeiro/Brazil?

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools

Quality Assurance of accelerators; the technologists responsibility

RICHARD T ELL ASSOCIATES,INC.

Worldwide Quality Assurance networks for radiotherapy dosimetry

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

THE DOSIMETRIC EFFECTS OF

THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode

Human Exposure Limits

The feasibility of a QA program for ISIORT Trials

2 Absorbing Solar Energy

Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients

A. Rate of Energy Loss

Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015

Physics testing of image detectors

Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the

Manual for simulation of EB processing. Software ModeRTL

BIOMEDICAL ULTRASOUND

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

SURFACE AREA AND VOLUME

[2] At the time of purchase of a Strontium-90 source, the activity is Bq.

Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy

World-first Proton Pencil Beam Scanning System with FDA Clearance

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

8.1 Radio Emission from Solar System objects

Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee Task Group 63

Synthetic Sensing: Proximity / Distance Sensors

MONTE CARLO MODELING OF A VARIAN 2100C 18 MV MEGAVOLTAGE PHOTON BEAM AND SUBSEQUENT DOSE DELIVERY USING MCNP5

Quality Assurance in Stereotactic. Radiotherapy

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Diagnostic X-ray facilities as per quality control performances in Tanzania

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong

X-ray Imaging Systems

Project Management Triangle

FXA UNIT G485 Module Ultrasound. Candidates should be able to :

where h = J s

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

X-ray Imaging Systems

Transcription:

Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy By Faiz Khan (Chapter 9 )

INTRODUCTION It is not possible to measure dose distribution directly in patients during their treatment Data on dose distribution are almost entirely derived from measurements in phantoms

PHANTOMS Basic dose distribution data are usually measured in a water phantom, which closely approximates the radiation absorption and scattering properties of muscle and other soft tissue (Wikipedia.org) for infant water is 75% of body-weight, for adults it is about 60% in males and 55% in females. Another reason for the choice of water as a phantom material is that it is universally available with reproducible radiation properties.

PHANTOMS Solid dry phantoms tissue or water equivalent, it must have the same effective atomic number number of electrons per gram mass density Since Compton effect is the main interaction for megavoltage photon beams used in hospital, the necessary condition for water equivalence is same electron density (number of electrons per cubic centimeter)

PHANTOMS The electron density ( e ) N A is Avogadro's number and a i is the fraction by weight of the ith element of atomic number Z, and atomic weight A i.

PHANTOMS From Faiz Khan

PHANTOMS The most important radiation properties in this regard are the mass attenuation coefficient the mass energy absorption coefficient electron mass stopping power angular scattering power ratios

PHANTOMS anthropomorphic phantom Frequently used for clinical dosimetry Incorporates materials to simulate various body tissues, muscle, bone, lung, and air cavities

DEPTH DOSE DISTRIBUTION PDD (percentage depth dose) is the ratio of dose at any depth to the dose at some reference depth along central axis of the beam and its variation depends on Beam energy Depth Field size Distance from source Beam collimation system

PERCENTAGE DEPTH DOSE For orthovoltage (up to about 400 kvp) and lower-energy x-rays, the reference depth is usually the surface (d o = 0). For higher energies, the reference depth is taken at the position of the peak absorbed dose (d o = dm).

PDD - Dependence on Beam Quality and Depth The PDD (beyond the dmax) increases with beam energy. Fig. 9.3 from Faiz Khan

PDD - Dependence on Beam Quality and Depth the dose build-up region. the skin-sparing effect.

PDD - Dependence on Beam Quality and Depth the dose build-up region. Fig. 9.4 from Faiz Khan

A comparison of relative surface dose (relative to d max ) as a function of field size for several linear accelerator models. Nominal Beam energy/ Field size accelerator model/ 5x5 10x10 20x20 ioniz. ratio open tray open tray open tray 6-MV Siemens KDS-2 8.25 8.44 13.27 14.53 22.73 27.02 IR=0.675 10-MV Siemens KDS-2 5.28 5.78 10.04 11.95 19.82 25.27 IR=0.748 6-MV Varian 2500 ----- 8.9 ----- 15.2 ----- 25.8 IR=0.677 10-MV Philips Sl75-20 ----- 5.7 ----- 11 ----- 21.3 IR=0.735

PDD - Effect of Field Size and Shape Field size Geometrical (Defined by mirror light at surface) Dosimetrical or physical (Defined as the size at 50% of isodose level) SAD FS

PDD - Effect of Field Size and Shape PDD increases with increasing field size Because increase in scattered dose is greater at larger depths than at the depth of D m The field size dependence of percent depth dose is less pronounced for the higher-energy than for the lower-energy beams.

PDD - Effect of Field Size and Shape Equivalent square fields for central axis depth-dose distribution Table 9.2 from Faiz Khan

PDD - Effect of Field Size and Shape Equivalent square fields for central axis depth-dose distribution A simple rule of thumb the A/P parameter (area to perimeter ratio), as such, does not apply to circular or irregularly shaped fields

PDD - Dependence on Source-Surface Distance Dose rate in free space from a point source varies inversely as the square of the distance. (IVSL) scattering material in the beam may cause deviation from the inverse square law. PDD increases with SSD IVSL SSD SSD dm d dm d

PDD - Dependence on Source-Surface Distance PDD increases with SSD

PDD - Dependence on Source-Surface Distance PDD increases with SSD the Mayneord F Factor ( without considering changes in scattering )

PDD - Dependence on Source-Surface Distance PDD increases with SSD

PDD - Dependence on Source-Surface Distance PDD increases with SSD the Mayneord F Factor works reasonably well for small fields since the scattering is minimal under these conditions. However, the method can give rise to significant errors under extreme conditions such as lower energy, large field, large depth, and large SSD change.

Source Distance Source to Skin Distance (SSD): the distance from the source (tungsten target) to the surface of the patient. Source to Axis Distance (SAD): the distance from the source of photons to the isocenter (the point around which the linac head can rotate in 360 o ).

SAD SSD SAD setup is used for rotational/ dynamic therapy SSD setup is mostly used for static therapy

TISSUE-AIR RATIO was first introduced by Johns et al. in 1953 the SSD may vary depending on the shape of the surface contour, the SAD remains constant TAR has been defined to facilitate calculations not only for rotation therapy but also for stationary isocentric techniques as well as irregular fields

TISSUE-AIR RATIO

TISSUE-AIR RATIO Effect of Distance Independent of the distance from the source. Variation with Energy, Depth, and Field Size Backscatter Factor As the beam energy is increased, the scatter is further reduced and so is the backscatter factor

TISSUE-AIR RATIO Effect of Distance Independent of the distance from the source. Variation with Energy, Depth, and Field Size Backscatter Factor increases with field size, its maximum value occurs for beams having a half-value layer (term used to specify the beam quality) between 0.6 and 0.8 mm Cu, depending on field size. Thus, for the orthovoltage beams with usual filtration, the backscatter factor can be as high as 1.5 for large field sizes

TISSUE-AIR RATIO Backscatter Factor For megavoltage beams (Co-60 and higher energies), the backscatter factor is much smaller. BSF for a 10 x 10-cm field for Co-60 is 1.036. This means that the D max, will be 3.6% higher than the dose in free space at the same point Above about 8 MV, the scatter at the depth of D max, becomes negligibly small and the backscatter factor approaches its minimum value of unity

Relationship between TAR and PDD r is field size at depth d and f is SSD

Relationship between TAR and PDD Conversion of Percent Depth Dose from One SSD to Another-the TAR Method

Example A patient is to be treated with an orthovoltage beam having a half-value layer of 3 mm Cu. Supposing that the machine is calibrated in terms of exposure rate in air, find the time required to deliver 200 cgy (rad) at 5 cm depth, given the following data: exposure rate = 100 R/min at 50 cm, field size = 8 x 8 cm, SSD = 50 cm, percent depth dose = 64.8, backscatter factor = 1.20, and rad/r = 0.95

Example

Example A patient is to be treated with Co-60 radiation. Supposing that the machine is calibrated in air in terms of dose rate free space, find the treatment time to deliver 200 cgy (rad) at a depth of 8 cm, given the following data: dose rate free space = 150 cgy/min at 80.5 cm for a field size of 10 x 10 cm, SSD = 80 cm, percent depth dose = 64.1, and backscatter factor = 1.036

Example Determine the time required to deliver 200 cgy (rad) with a Co- 60 ray beam at the isocenter which is placed at a 10 cm depth in a patient, given the following data: SAD = 80 cm, field size = 6 x 12 cm (at the isocenter), dose rate in free space at the SAD for this field = 120 cgy/min and TAR = 0.68 1

SCATTER-AIR RATIO Useful in calculating scattered dose in the medium The computation of the primary and the scattered dose separately is particularly useful in the dosimetry of irregular fields Defined as the ratio of the scattered dose at a given point in the phantom to the dose in free space at the same point depends on the beam energy, depth, and field size. Here d is the depth and r d is field size at that depth

SCATTER-AIR RATIO Dose Calculation in Irregular Fields-Clarkson's Method Any field other than square, rectangular and circular field is considered as irregular The scattered component of the depth dose, which depends on the field size and shape, can be calculated separately from the primary component which is independent of the field size and shape

Irregular Fields-Clarkson's Method This method is used to calculate the scatter contribution from all parts of irregular field The irregular treatment field is divided into sectors (say each of sector is 10 o, then its contribution is 1/36) Each sector is characterized by its radius and can be considered as a circular part of that radius. Scatter contribution from all these sectors is summed using tables of SAR to get the average scatter in the treatment field

SCATTER-AIR RATIO Clarkson's Method Net (SAR) QC = (SAR) QC - (SAR) QB + (SAR) QA TAR = TAR(0) + SAR Treatment area Blocked area