Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy
|
|
|
- Jordan Neal
- 9 years ago
- Views:
Transcription
1 Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy This set of 117 slides is based on Chapter 11 authored by M.D.C. Evans of the IAEA publication (ISBN ): Radiation Oncology Physics: A Handbook for Teachers and Students Objective: To familiarize students with general principles, particular procedures and quality assurance of computerized treatment planning systems including hardware and software. IAEA International Atomic Energy Agency Slide set prepared in 2006 (updated Aug2007) by G.H. Hartmann (DKFZ, Heidelberg) Comments to S. Vatnitsky: [email protected] CHAPTER 11. TABLE OF CONTENTS 11.1 Introduction 11.2 System hardware 11.3 System software and calculations algorithms 11.4 Commissioning and quality assurance 11.5 Special considerations IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students - 11.Slide 1 (2/117) 1
2 11.1 INTRODUCTION Radiation treatment planning represents a major part of the overall treatment process. It consists of many steps including: Patient diagnostic Tumour staging Image acquisition for treatment planning Localization of tumor and healthy tissue volumes Optimal beam placement Treatment simulation and optimization. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (3/117) 11.1 INTRODUCTION Steps of the treatment planning process, the professionals involved in each step and the QA activities associated with these steps (IAEA TRS 430) TPS related activity IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (4/117) 2
3 11.1 INTRODUCTION This chapter deals explicitly with that component of the treatment planning process that makes use of the computer. Computerized Treatment Planning Systems (TPS) are used in external beam radiation therapy to generate beam shapes and dose distributions with the intent to maximize tumor control and minimize normal tissue complications. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (5/117) 11.1 INTRODUCTION Treatment planning prior to the 1970s was generally carried out through the manual manipulation of standard isodose charts onto patient body contours that were generated by direct tracing or lead-wire representation, and relied heavily on the judicious choice of beam weight and wedging by an experienced dosimetrist. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (6/117) 3
4 11.1 INTRODUCTION Simultaneous development of computerized tomography, along with the advent of readily accessible computing power from the 1970s on, lead to the development of CT-based computerized treatment planning, providing the ability to view dose distributions directly superimposed upon patient s axial anatomy. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (7/117) 11.1 INTRODUCTION Advanced treatment planning systems (TPSs) are now able to represent patient anatomy, tumor targets and even dose distributions as three dimensional models. Clinical target volume, both lungs, and spinal chord, as seen from behind (ICRU 50). IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 5 (8/117) 4
5 11.1 INTRODUCTION Successive improvements in treatment planning hardware and software have been most notable in the graphics, calculation and optimization aspects of current systems. Systems encompassing the virtual patient are able to display: Beams-Eye Views (BEV) of the patient's anatomy eyes optic nerves tumor brain stem Digitally Reconstructed Radiographs (DRR) IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 6 (9/117) 11.1 INTRODUCTION Dose calculations have evolved from simple 2D models through 3D models to 3D Monte-Carlo techniques, and increased computing power continues to increase the calculation speed. Monte Carlo simulation of an electron beam produced in the accelerator head. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 7 (10/117) 5
6 11.1 INTRODUCTION Computerized treatment planning is a rapidly evolving modality, relying heavily on hardware and software. As such, it is necessary for related professionals to develop a workable Quality Assurance (QA) program that reflects the use of the TP system in the clinic, and is sufficiently broad in scope to ensure proper treatment delivery. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 8 (11/117) 11.2 SYSTEM HARDWARE In the 1970s and 1980s treatment planning computers became readily available to individual radiation therapy centers. As computer hardware technology evolved and became more compact, so did Treatment Planning Systems (TPS). IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (12/117) 6
7 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a Treatment Planning System (TPS): Central Processing Unit (CPU) Graphics display Memory Digitizing devices Output devices Archiving and network communication devices IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (13/117) 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (1) Central Processing Unit The Central Processing Unit must have: Sufficient memory. Sufficiently high processor speed. as required by the operating system and the treatment planning software to run the software efficiently. In the purchase phase, specifications for system speed, Random Access Memory (RAM) and free memory, as well as networking capabilities must be carefully considered. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (14/117) 7
8 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (2) Graphics display Graphics display should be capable of rapidly displaying high resolution images. Graphics speed can be enhanced with video cards and hardware drivers (graphics processor). Resolution should be sub-millimeter or better so as not to distort the input. Graphics display should be sufficient for accommodating the patient transverse anatomy on a 1:1 scale, typically 17 to 21 inches (43 to 53 cm) or larger. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (15/117) 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (2) Graphics display IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (16/117) 8
9 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (3) Memory Memory and archiving functions are carried through: a) Removable media: Re-writable hard-disks Optical disks DVDs DAT tape Attention: These devices have been reported to suffer from long term instability. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 5 (17/117) 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (3) Memory Memory and archiving functions are carried out through: b) network on: Remote computer Server Linac with its record-and-verify system Archiving operations may be carried out automatically during low use periods of the day. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 6 (18/117) 9
10 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (4) Digitizing devices Digitizing devices are used to acquire manually entered patient data, such as transverse contours and beams-eye-views of irregular field shapes. Methods: Backlit tablets with stylus for manually tracing shapes Scanners to digitize images from hardcopy such as paper or radiographic film Video frame grabbers IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 7 (19/117) 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (4) Digitizing devices IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 8 (20/117) 10
11 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (5) Output devices Output devices include color laser printers and plotters for text and graphics. Printers and plotters can be networked for shared access. Hardcopy can be to paper or to film via a laser camera. Uninterruptible Power Supplies (UPS). IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 9 (21/117) 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (5) Output devices (cont.) Uninterruptible Power Supplies (UPS) are recommended for the CPU, data servers, and other critical devices such as those used for storage and archiving. UPSs can provide back-up power so that a proper shut-down of the computer can be accomplished during power failures from the regular power distribution grid, and they also act as surge suppressors for the power. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 10 (22/117) 11
12 11.2 SYSTEM HARDWARE Treatment planning system hardware Principal hardware components of a TP system: (6) Communications hardware Communications hardware includes modem or ethernet cards on the local workstations and multiple hubs for linking various peripheral devices and workstations. Large networks require fast switches running at least 100 Mb/s for file transfer associated with images. Physical connections on both small and large networks are run through coaxial cable, twisted pair or optical fiber depending upon speed requirements. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 11 (23/117) 11.2 SYSTEM HARDWARE Treatment planning system configurations TP hardware systems can be classified into two branches: Small TP system configurations for only a few users Stand-alone lay-out and archiving One central CPU for most functions and communication requests. Requiring network switches to communicate with digital imaging devices such as CT-scanners. Laser printer Colour Plotter Ethernet Link Network Switch CPU Keyboard Optical drive Digitizer Mouse Graphics Monitor CT Scanner CT Console IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 12 (24/112) 12
13 11.2 SYSTEM HARDWARE Treatment planning system configurations TP hardware systems can be classified into two branches: Large TP system configurations for many users Often operate on remote workstations within a hospital network. May make use of Internet-based communication systems. May require the services of an administrator to maintain security, user rights, networking, back-up and archiving. Physician Workstation File Server System Manager CT Scanner CT Console DICOM Server Network Switch Optical drive Workstation Workstation Workstation Workstation Film Scanner Digitizer Digitizer Digitizer Digitizer Laser printer Colour Plotter Laser printer Colour Plotter IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 13 (25/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS The software of a TP system includes components for: The computer operating system (plus drivers, etc.). Utilities to enter treatment units and associated dose data. Utilities to handle patient data files. Contouring structures such as anatomical structures, target volumes, etc. Dose calculation. TP evaluation. Hardcopy devices. Archiving. Backup to protect operating system and application programs. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (26/117) 13
14 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Whereas the software modules to handle digital images, contours, beams, dose distributions, etc. are mostly very similar, the dose algorithm is the most unique, critical and complex piece of the TP software: These modules are responsible for the correct representation of dose in the patient. Results of dose calculations are frequently linked to beam-time or monitor unit (MU) calculations. Many clinical decisions are taken on the basis of the calculated dose distributions. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (27/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms It must be emphasized that prior to understanding the sophisticated computerized treatment planning algorithms, a proper understanding of manual dose calculations is essential. For more details of manual dose calculations see Chapter 7. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (28/117) 14
15 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Beam model Because absorbed dose distributions cannot be measured directly in a patient, they must be calculated. Formalism for the mathematical manipulation of dosimetric data is sometimes referred to as the beam model. The following slides are providing an overview of the development of beam models as required when calculation methods have evolved from simple 2 D calculations to 3D calculations. The ICRU Report 42 provides useful examples. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (29/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Early methods First beam models simply consist of a 2D-matrix of numbers representing the dose distribution in a plane. Cartesian coordinates are the most straightforward used coordinate system. An isodose chart for a 10x10 cm beam of 60 Co radiation superimposed on a Cartesian grid of points. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (30/117) 15
16 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms The disadvantages of matrix representation (in the early days of computers) are the large amount of data and the number of different tables of data required. To reduce the number of data, beam generating functions have been introduced. The dose distribution in the central plane D(x,z) was usually expressed by the product of two generating functions D(x,z) = P(z,z ref )g z (x) P(z,z ref ) is the depth dose along central axis relative to the dose at z ref g z (x) is the off axis ratio at depth z IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 5 (31/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Example for P(z,z max ) introduced by van degeijn as a quite precise generating function: where SSD + z P(z,A,SSD,h ) = 100 max SSD + z c c = A field size at center μ(c) = μ 0 a 1 exp( bc) 2 e μ(c)(z z max ) IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 6 (32/117) 16
17 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Example for g(x) introduced by Sterling: g(x) = 1 x/ X 1 exp ( 1) d where = x/ X is the off axis distance x expressed as a fraction of the half geometrical beam width X is an empirical quantity IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 7 (33/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms There are many other formulas available for the generating function for the depth dose along the beam central ray. There are also many dosimetric quantities used for this purpose such as: PDD = percentage depth dose TAR = tissue air ratio TPR = tissue phantom ratio TMR = tissue maximum ratio For more details, please see Chapter 6. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 8 (34/117) 17
18 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms The approach to use two generating functions for the 2D dose distribution in the central plane: D(x,z) = P(z,z ref ) g z (x) can be easily extended to three dimensions: Dxz (, ) = Pzz (, ) g( xy, ) ref It was again van der Geijn, who introduced the factorization: z g ( x, y) = g ( x) g ( y) z 1,z 2,z IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 9 (35/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Another approach is the separation of the dose into its two components and to describe them separately: D = D prim + D scat Primary radiation D prim is taken to be the radiation incident on the surface and includes photons coming directly from the source as well as radiation scattered from structures near the source and the collimator system. Scattered radiation D scat results from interactions of the primary radiation with the phantom (patient). source D prim D scat water phantom IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 10 (36/117) 18
19 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Johns and Cunningham based the separation of primary and scattered radiation dose on a separation of the tissue air ratio TAR into two components: TAR(z,r) = TAR 0 (z,r = 0) + SAR(z,r) TAR 0 (z,r = 0) SAR(z,r) is the TAR at depth z for a field of zero area (= primary radiation) is the term representing the scattered radiation in a circular beam with radius r IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 11 (37/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Accordingly, the dose D at a point x,y,z is given by: D(x,y,z) = D a TAR 0 (z) f(x,y) + i SAR(z,r i ) i 2 D a is the dose in water, free in air at the central axis in depth z f(x,y) is analog to the position factor g(x,y), however free in air The summation is over sectors of circular beams (Clarkson method). IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 12 (38/117) 19
20 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Beams-Eye View of a rectangular field Calculation of radiation scattered to various points using the Clarkson Method: O: at the beam axis P: off axis within the beam Q: outside the beam IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 13 (39/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms The method of decomposition of radiation into a primary and a scattered component is also used in current beam calculation algorithms. The convolution superposition method is a model for that. With this method the description of primary photon interactions ( ) is separated from the transport of energy via scattered photons and charged particles produced through photoelectric effect, Compton scattering and pair production. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 14 (40/117) 20
21 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Scatter components may come from regions in the form of a point, pencil beam, or slab. The pattern of spread of energy from such entities are frequently called "scatter kernels". Point kernel Pencil kernel Slab kernel IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 15 (41/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms In this manner, changes in scattering due to changes in the beam shape, beam intensity, patient geometry and tissue inhomogeneities can be incorporated more easily into the dose distribution. Pencil beam algorithms are common for electron beam dose calculations. In these techniques the energy spread or dose kernel at a point is summed along a line in phantom to obtain a pencil-type beam or dose distribution. By integrating the pencil beam over the patient s surface to account for the changes in primary intensity and by modifying the shape of the pencil beam with depth and tissue density, a dose distribution can be generated. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 16 (42/117) 21
22 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Monte Carlo or random sampling techniques are another currently applied calculation method used to generate dose distributions. Results are obtained by following the histories of a large number of particles as they emerge from the source of radiation and undergo multiple scattering interactions both inside and outside the patient. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 17 (43/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Monte Carlo techniques are able to model accurately the physics of particle interactions by accounting for the geometry of individual linear accelerators, beam shaping devices such as blocks and multileaf collimators (MLCs), and patient surface and density irregularities. Monte Carlo techniques for computing dose spread arrays or kernels used in convolution superposition methods are described by numerous authors, including Mackie, and in the review chapters in Khan and Potish. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 18 (44/117) 22
23 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Calculation algorithms Although Monte Carlo techniques require a large number of particle histories to achieve statistically acceptable results, they are now becoming more and more practical for routine treatment planning. A detailed summary of treatment planning algorithms in general is provided in: The Modern Technology for Radiation Oncology: A Compendium for Medical Physicists and Radiation Oncologist (editor: Van Dyk) IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 19 (45/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers Treatment planning software for photon beams and electron beams must be capable of handling the many diverse beam modifying devices found on linac models. Photon beam modifiers: Jaws Blocks Compensators MLCs Wedges Electron beam modifiers Cones Blocks Bolus IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (46/117) 23
24 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Photon beam modifiers Jaws The field size is defined by motorized collimating jaws. Jaws can move independently or in pairs and are usually located as an upper and lower set. Jaws may over-travel the central axis of the field by varying amounts. The travel motion will determine the junction produced by two abutting fields. The TPS should account for the penumbra and differences in radial and transverse open beam symmetry. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (47/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Photon beam modifiers Shielding blocks Blocks are used for individual field shielding. The TPS must take into account the effective attenuation of the block. The dose through a partially shielded calculation volume, or voxel, is calculated as a partial sum of the attenuation proportional to the region of the voxel shielded. TPSs are able to generate files for blocked fields that can be exported to commercial block cutting machines. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (48/117) 24
25 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Photon beam modifiers Multileaf collimator (MLC) MLC is a beam shaping device that can replace almost all conventional mounted blocks, with the exception of island blocking and excessively curved field shapes. MLCs with a leaf width of the order of 0.5 cm 1.0 cm at the isocenter are typical. MLCs providing smaller leaf widths are referred to as micro MLCs. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (49/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Photon beam modifiers Multileaf collimator The MLC may be able to cover all or part of the entire field opening, and the leaf design may be incorporated into the TPS to model transmission and penumbra. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 5 (50/117) 25
26 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Photon beam modifiers Static Wedges Static wedges remain the principal devices for modifying dose distributions. Isodose lines The TPS can model the effect of the patient dose both along and across the principal axes of the physical wedge, as well as account for any PDD change due to beam hardening and/or softening along the central axis ray line. The clinical use of wedges may be limited to field sizes smaller than the maximum collimator setting. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 6 (51/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Photon beam modifiers Mechanical wedges and Dynamic Wedges More recently, wedging may be accomplished by the use of universal or sliding wedges incorporated into the linac head (Mechanical wedges). Even more elegantly, by dynamic wedging accomplished by the motion of a single jaw while the beam is on. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 7 (52/117) 26
27 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Electron beam modifiers Custom compensators Custom compensators may be designed by TPSs to account for missing tissue or to modify dose distributions to conform to irregular target shapes. TPSs are able to generate files for compensators that can be read by commercial compensator cutting machines. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 8 (53/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Electron beam modifiers Cones or applicators Electron beams are used with external collimating devices known as cones or applicators that reduce the spread of theelectron beam in the air. Design of these cones depends on the manufacturer and affects the dosimetric properties of the beam. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 9 (54/117) 27
28 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Electron beam modifiers Shielding for irregular fields Electron shielding for irregular fields may be accomplished with the use of thin lead or low melting point alloy inserts. Shielding inserts can have significant effects on the dosimetry that should be modeled by the TPS. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 10 (55/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Electron beam modifiers Scattering foil The design of the linac head is important for electron dosimetry, especially for Monte Carlo type calculations. In these conditions particular attention is paid to the scattering foil. The effective or virtual SSD will appear to be shorter than the nominal SSD, and should be taken into consideration by the TPS. Electron Beam Scattering Foil Patient Primary Collimator Ion Chamber Secondary Collimator Electron applicator IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 11 (56/117) 28
29 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Beam modifiers: Electron beam modifiers Bolus Bolus may be used to increase the surface dose for both photon and electron treatments. Bolus routines incorporated into TPS software will usually permit manual or automatic bolus insertion in a manner that does not modify the original patient CT data. It is important that the TPS can distinguish between the bolus and the patient so that bolus modifications and removal can be achieved with ease. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 12 (57/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Heterogeneity corrections Heterogeneity or inhomogeneity corrections generally account for the differences between the standard beam geometry of a radiation field incident upon a large uniform water phantom and the beam geometry encountered by the beam incident upon the patient s surface. In particular, beam obliquity and regions where the beam does not intersect the patient s surface will affect the dose distribution. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (58/117) 29
30 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Heterogeneity corrections Inside the patient, the relative electron density of the irradiated medium can be determined from the patient CT data set. CT-numbers (HU) relative electron density IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (59/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Heterogeneity corrections Most TPS algorithms apply either a correction factor approach or a model based approach. Fast methods: Generalized correction factors Power law method Equivalent TAR method Longer calculation times: Model based approaches Differential SAR approach Monte Carlo based algorithms consider Most methods are still having difficulties with dose calculations at tissue interfaces. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (60/117) 30
31 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Image display and dose volume histograms Beam s Eye View Room s Eye View Beam s eye views (BEVs) and room s eye views (REVs) are used by modern TPSs. The BEV is often used in conjunction with DRRs to aid in assessing tumor coverage and for beam shaping with blocks or an MLC. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (61/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Image display and dose volume histograms Without collision between gantry and table With collision between gantry and table The Room s Eye View gives the user a perception of the relationship of the gantry and table to each other and may help in avoiding potential collisions when moving from the virtual plan to the actual patient set-up. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (62/117) 31
32 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Image display and dose volume histograms DRR treatment fields DRR EPID fields EPID images Virtual portal images with the treatment field superimposed can be used for comparison with the portal images obtained with the patient in the treatment position on the treatment machine. Portal image generation can be accomplished by TPSs by substituting energy shifted attenuation coefficients for CT data sets. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (63/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Image display and dose volume histograms CT and PET image before fusion Matched images Image registration routines can help match simulator, MR, positron emission tomography (PET), single photon emission computed tomography (SPECT), ultrasound and other image sources to planning CT and treatment acquired portal images. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (64/117) 32
33 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Image display and dose volume histograms Volume (%) Dose (Gy) DVHs are calculated by the TPS with respect to the target and critical structure volumes in order to establish the adequacy of a particular treatment plan and to compare competing treatment plans. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 5 (65/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Image display and dose volume histograms Two types of DVHs are in use: Direct (or differential) DVH Cumulative (or integral) DVH Definition: The volume that receives at least the given dose and plotted versus dose. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 6 (66/117) 33
34 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Optimization and monitor unit calculation The possibility of simulating radiation therapy with a computer and predicting the resulting dose distribution with high accuracy allows an optimization of the treatment. Optimization routines including inverse planning are provided by TPSs with varying degrees of complexity. Algorithms can modify beam weights and geometry or calculate beams with a modulated beam intensity to satisfy the user criteria. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (67/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Optimization and monitor unit calculation Imaging (CT, MR, PET) Definition of target volume(s) and critical structures Definition of treatment parameters Simulation of patient irradiation Optimization loop Dose calculation Evaluation of dose distribution Optimization tries to determine Treatment delivery the parameters of the treatment in an iterative loop in such a way that the best possible treatment will be delivered for an individual patient. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (68/117) 34
35 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Optimization and monitor unit calculation Beam time and MU calculation by TPSs is frequently optional. The associated calculation process is directly related to the normalization method. Required input data: Absolute output at a reference point. Decay data for cobalt units. Output factors. Wedge factors. Tray factors and other machine specific. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (69/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Record and verify systems A computer-aided record-and-verify system aims to compare the set-up parameters with the prescribed values. Patient identification data, machine parameters and dose prescription data are entered into the computer beforehand. At the time of treatment, these parameters are identi-fied at the treatment machine and, if there is no difference, the treatment can start. If discrepancies are present this is indicated and the parameters concerned are highlighted. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (70/117) 35
36 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Record and verify systems Networked TPSs allow for interface between linac record and verify systems, either through a direct connection or through a remote server using fast switches. Communication between the TPS and the linac avoids the errors associated with manual transcription of paper printouts to the linac and can help in the treatment of complex cases involving asymmetric jaws and custom MLC shaped fields. Record and verify systems may be provided by: TPS manufacturer. Linac manufacturer. Third party software. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (71/117) 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Biological modeling Distributions modeled on biological effects may in the future become more clinically relevant than those based upon dose alone. Such distributions will aid in predicting both the tumour control probability (TCP) and the normal tissue complication probability (NTCP). TCP and NTCP are usually illustrated by plotting two sigmoid curves, one for the TCP (curve A) and the other for NTCP (curve B). Dose (Gy) IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (72/117) 36
37 11.3 SYSTEM SOFTWARE AND CALCULATION ALGORITHMS Biological modeling These algorithms can account for specific organ dose response and aid in assessing the dose fractionation and volume effects. Patient specific data can be incorporated in biological model to help predict individual dose response to a proposed treatment regime. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (73/117) 11.4 DATA ACQUISITION AND ENTRY Data acquisition refer to all data to establish: Machine model Beam model Patient model An important aspect of the configuration of a TPS is the creation of a machine database that contains descriptions of the treatment machines, i.e., the machine model. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (74/117) 37
38 11.4 DATA ACQUISITION AND ENTRY Machine data Each TPS requires the entry of a particular set of parameters, names and other information, which is used to create the geometrical and mechanical descriptions of the treatment machines for which treatment planning will be performed. It must be ensured that any machine, modality, energy or accessory that has not been tested and accepted be made unusable or otherwise made inaccessible to the routine clinical users of the system. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (75/117) 11.4 DATA ACQUISITION AND ENTRY Machine data The following are examples of machine entry data: Identification (code name) of machines, modalities, beams (energies) and accessories. Geometrical distances: SAD, collimator, accessory, etc. Allowed mechanical movements and limitations: upper and lower jaw limits, asymmetry, MLC, table, etc. Display co-ordinate system gantry, collimator and table angles, table x, y, z position, etc. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (76/117) 38
39 11.4 DATA ACQUISITION AND ENTRY Machine data Caution Issues such as co-ordinates, names and device codes require verification, since any mislabeling or incorrect values could cause systematic misuse of all the plans generated within the TPS. In particular, the scaling conventions for the gantry, table and collimator rotation, etc. used in a particular institution must be fully understood and described accurately. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (77/117) 11.4 DATA ACQUISITION AND ENTRY Beam data acquisition and entry The requirements on the set of beam entry data may be different depending on a specific TPS. Requirements on the beam data set must be well understood. Data sets are mainly obtained by scanning in a water phantom. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (78/117) 39
40 11.4 DATA ACQUISITION AND ENTRY Beam data acquisition and entry Typical photon beam data sets include: Central axis PDDs Off Axis Ratios (profiles) Output factors For a range of square fields For open fields For wedged fields Diagonal field profiles to account for radial and transverse open beam asymmetry; (it may only be possible to acquire half-field scans, depending upon the size of the water tank) IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (79/117) 11.4 DATA ACQUISITION AND ENTRY Beam data acquisition and entry Caution Special consideration must be given to the geometry of the radiation detector (typically ionization chambers or diodes) and to any correction factors that must be applied to the data. Beam data are often smoothed and renormalized both following measurement and prior to data entry into the treatment planning computer. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (80/117) 40
41 11.4 DATA ACQUISITION AND ENTRY Beam data acquisition and entry The penumbra may be fitted to, or extracted from, measured data. In either case it is important that scan lengths be of sufficient length, especially for profiles at large depths, where field divergence can become considerable. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (81/117) 11.4 DATA ACQUISITION AND ENTRY Beam data acquisition and entry Calculation of dose at any point is usually directly linked to the dose under reference conditions (field size, reference depth and nominal SSD etc.). Particular care must therefore be taken with respect to the determination of absolute dose under reference conditions, as these will have a global effect on time and MU calculations. A ref water phantom z ref f = SSD IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 5 (82/117) 41
42 11.4 DATA ACQUISITION AND ENTRY Beam data acquisition and entry Measured beam data relevant to the MLC can include: Transmission through the leaf. Inter-leaf transmission between adjacent leaves. Intra-leaf transmission occurring when leaves from opposite carriage banks meet end on. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 6 (83/117) 11.4 DATA ACQUISITION AND ENTRY Beam data acquisition and entry Beam measurement for electrons is more difficult than for photons because of the continuously decreasing energy of the beam with depth. Electron beam data measured with ionization chambers must be first converted to dose with an appropriate method: Typically using a look-up table of stopping power ratios. Measurements with silicon diodes are often considered to be tissue equivalent and give a reading directly proportional to dose. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 7 (84/117) 42
43 11.4 DATA ACQUISITION AND ENTRY Beam data acquisition and entry Beam data acquired can be entered: Manually using a digitizer tablet and tracing stylus A hard copy of beam data is used, and it is important that both the beam data printout and the digitizer be routinely checked for calibration. Via a keyboard Keyboard data entry is inherently prone to operator error and requires independent verification. Via file transfer from the beam acquisition computer Careful attention must be paid to the file format. File headers contain formatting data concerning the direction of measurement, SSD, energy, field size, wedge type and orientation, detector type and other relevant parameters. Special attention must be paid to these labels to ensure that they are properly passed to the TPS. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 8 (85/117) 11.4 DATA ACQUISITION AND ENTRY Patient data The patients anatomical information may be entered via the digitizer using one or more contours obtained manually or it may come from a series of transverse slices obtained via a CT scan. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (86/117) 43
44 11.4 DATA ACQUISITION AND ENTRY Patient data 3-D information data required to localize the tumor volume and normal tissues may be obtained from various imaging modalities such as: Multi-slice CT or MR scanning. Image registration and fusion techniques in which the volume described in one data set (MRI, PET, SPECT, ultrasound, digital subtraction angiography (DSA) is translated or registered with another data set, typically CT. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (87/117) 11.4 DATA ACQUISITION AND ENTRY Patient data The patient image data may be transferred to the TPS via DICOM formats (Digital Imaging and Communications in Medicine) DICOM 3 format DICOM RT (radiotherapy) format Both formats were adopted in 1993 by the: American College of Radiology (ACR). National Electrical Manufacturers Association (NEMA). IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (88/117) 44
45 11.4 DATA ACQUISITION AND ENTRY Patient data To ensure accurate dose calculation, the CT numbers must be converted to electron densities and scattering powers. The conversion of CT numbers to electron density and scattering power is usually performed with a user defined look-up table. CT-numbers (HU) relative electron density IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (89/117) 11.4 DATA ACQUISITION AND ENTRY Patient data Such tables can be generated using a phantom containing various inserts of known densities simulating normal body tissues such as bone and lung. CIRS torso phantom Gammex RMI CT test tool IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 5 (90/117) 45
46 11.4 DATA ACQUISITION AND ENTRY Patient data The rendering of patient anatomy from the point of view of the radiation source (BEV) is useful in viewing the path of the beam, the structures included in the beam and the shape of the blocks or MLC defined fields. tumor MLC defined field eyes optic nerves brain stem IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 6 (91/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Commissioning is the process of preparing a specific equipment for clinical service. Commissioning is one of the most important parts of the entire QA program for both the TPS and the planning process. Commissioning involves testing of system functions, documentation of the different capabilities and verification of the ability of the dose calculation algorithms to reproduce measured dose calculations. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (92/117) 46
47 11.5 COMMISSIONING AND QUALITY ASSURANCE RTPS USER Commissioning procedures Commissioning results Periodic QA program IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (93/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE IAEA TRS the most complete reference work in the field of QA of radiotherapy TP systems. Provides a general framework on how to design a QA programme for all kinds of radiotherapy TP systems. Describes a large number of tests and procedures that should be considered and should in principle fulfil the needs for all radiotherapy TP system users. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (94/117) 47
48 11.5 COMMISSIONING AND QUALITY ASSURANCE Errors Uncertainty: When reporting the result of a measurement, it is obligatory that some quantitative indication of the quality of the result be given. Otherwise the receiver of this information cannot really asses its reliability. The term "Uncertainty" has been introduced for that. Uncertainty is a parameter associated with the result of a measurement of a quantity that characterizes the dispersion of the values that could be reasonably be attributed to the quantity. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (95/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Errors Errors: In contrast to uncertainty, an error is the deviation of a given quantity following an incorrect procedure. Errors can be made even if the result is within tolerance. However, the significance of the error will be dependent on the proximity of the result to tolerance. Sometimes the user knows that a systematic error exists but may not have control over the elimination of the error. This is typical for a TPS for which the dose calculation algorithm may have a reproducible deviation from the measured value at certain points within the beam. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (96/117) 48
49 11.5 COMMISSIONING AND QUALITY ASSURANCE Errors Tolerance Level: Term tolerance level is used to indicate that the result of a quantity has been measured with acceptable accuracy. Tolerances values should be set with the aim of achieving the overall uncertainties desired. However, if the measurement uncertainty is greater than the tolerance level set, then random variations in the measurement will lead to unnecessary intervention. Therefore, it is practical to set a tolerance level for the measurement uncertainty at the 95% confidence level. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (97/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Errors Action Level: A result outside the action level is unacceptable and demands action. It is useful to set action levels higher than tolerance levels thus providing flexibility in monitoring and adjustment. Action levels are often set at approximately twice the tolerance level However, some critical parameters may require tolerance and action levels to be set much closer to each other or even at the same value. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (98/117) 49
50 11.5 COMMISSIONING AND QUALITY ASSURANCE Errors Illustration of a possible relation between uncertainty, tolerance level and action level tolerance level equivalent to 95% confidence interval of uncertainty standard uncertainty 2 sd 4 sd action level = 2 x tolerance level 1 sd mean value action level = 2 x tolerance level IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 5 (99/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Errors System of actions: If a measurement result is within the tolerance level, no action is required. If the measurement result exceeds the action level, immediate action is necessary and the equipment must not be clinically used until the problem is corrected. If the measurement falls between tolerance and action levels, this may be considered as currently acceptable. Inspection and repair can be performed later, for example after patient irradiations. If repeated measurements remain consistently between tolerance and action levels, adjustment is required. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 6 (100/117) 50
51 11.5 COMMISSIONING AND QUALITY ASSURANCE Errors Typical tolerance levels from AAPM TG53 (examples) Square field CAX: 1% MLC penumbra: 3% Wedge outer beam: 5% Buildup-region: 30% 3D inhomogeneity CAX: 5% For analysis of agreement between calculations and measurements, the dose distribution due to a beam is broken up into several regions. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 7 (101/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Verification Data verification entails a rigorous comparison between measured or input data and data produced by the TPS. Standard test data sets such as the AAPM TG 23 data set can be used to assess TPS performance. Detailed description of tests are provided by: Fraas et al, AAPM Radiation Therapy Committee TG 53: Quality assurance program for radiotherapy treatment planning", Med Phys 25, (1998). IAEA, "Commissioning and quality assurance of computerized planning systems for radiation treatment of cancer", TRS 430. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (102/117) 51
52 11.5 COMMISSIONING AND QUALITY ASSURANCE Verification Typical issues of calculation and verification (TRS 430) 1-D 2-D 3-D Comparison techniques Comparison of one or more depth dose and profile curves. Table of differences of depth dose curves for several field sizes. Isodose line (IDL) comparison: plotted IDLs for calculated and measured data. Dose difference display: subtract the calculated dose distribution from the measured distribution; highlight regions of under- and overexposure, if available. Distance to agreement: plot the distance required for measured and calculated isodose lines to be in agreement, if available. Generate a 3-D measured dose distribution by interpolation of 2-D coronal dose distributions and a depth dose curve, if available. DVH comparison of 3-D calculated and measured distributions, if available DVH of 3-D dose difference distribution, if available. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (103/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Verification Typical commissioning tests Item Digitizer and plotter Geometry Beam junction Rotational beams File compatibility between CT & TPS Image transfer Test Enter a known contour and compare it with final hard copy Oblique fields, fields using asymmetric jaws Test cases measured with film or detector arrays Measured or published data May require separate test software for the transfer Analysis of the input data for a known configuration and density (phantom) to detect any error in magnification and in spatial coordinates IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (104/117) 52
53 11.5 COMMISSIONING AND QUALITY ASSURANCE Spot checks Spot checks of measured data versus those obtained from the TPS are required; these spot checks can involve calculations of fields shielded by blocks or MLCs. Spot checks of static and dynamic wedged fields with respect to measured data points are also recommended. A detector array may be used to verify wedged and, even more importantly, dynamically wedged dose distributions produced by the TPS. Wedge distributions produced by the TPS must be verified for identification, orientation, beam hardening effects and field size limitations. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (105/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Normalization and beam weighting Dose normalization and beam weighting options vary from one TPS to another and have a direct impact on the representation of patient dose distributions. Normalization methods refer to: A specific point such as the isocenter. Intersection of several beam axes. A minimum or maximum value in a slice or entire volume. Arbitrary isodose line in a volume. A minimum or maximum iso-surface. Specific point in a target or organ. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (106/117) 53
54 11.5 COMMISSIONING AND QUALITY ASSURANCE Normalization and beam weighting Beam weighting Various approaches are possible: 1) Weighting of beams as to how much they contribute to the dose at the target 25% 30% 25% 40% 25% 10% 2) Weighting of beams as to how much dose is incident on the patient The two approaches are NOT the same. 20% 25% IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (107/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Normalization and beam weighting Manual checks of beam time or monitor units must be well familiar with the type of normalization and beam weighting method of a specific TPS. Examples are given in more detail in Chapter 7. Since many treatment plans involve complex beam delivery, these manual checks do not need to be precise, yet they serve as a method of detecting gross errors on the part of the TPS. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (108/117) 54
55 11.5 COMMISSIONING AND QUALITY ASSURANCE Dose volume histograms and optimization Current state of the art TPSs use DVHs to summarize the distribution of the dose to particular organs or other structures of interest. According to TRS 430, tests for DVHs must refer to: Type (direct, cumulative and differential) Plan normalization Relative and absolute dose Volume determination Histogram dose bin size Structures Consistency Calculation of grid size and points distribution DVH comparison guidelines Dose statistics IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (109/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Dose volume histograms and optimization Optimization routines are provided by many TPSs, and intensity modulated beams having complex dose distributions may be produced. As these set-ups involve partial or fully dynamic treatment delivery, spot checks of absolute dose to a point, as well as a verification of the spatial and temporal aspects of the dose distributions using either film or detector arrays, are a useful method of evaluating the TPS beam calculations. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (110/117) 55
56 11.5 COMMISSIONING AND QUALITY ASSURANCE Training and documentation Training and a reasonable amount of documentation for both the hardware and software are essential. Typically the training is given on the site and at the manufacturer s facility. Ongoing refresher courses are available to familiarize dosimetrists and physicists with bug fixes and system upgrades. Documentation regarding software improvements and fixes is kept for reference by users at the clinic. TPS manufacturers have lists of other users and resource personnel to refer to. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (111/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Training and documentation Most manufacturers of TP systems organize users meetings, either as stand-alone meetings or in conjunction with national or international scientific meetings of radiation oncologists or radiation oncology physicists. During these meetings special seminars are given by invited speakers and users describing the particular software systems, new developments in hardware and software as well as problems and solutions to specific software problems. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (112/117) 56
57 11.5 COMMISSIONING AND QUALITY ASSURANCE Scheduled Quality Assurance Following acceptance and commissioning of a radiotherapy computerized TP system, a scheduled quality assurance program must be established to verify the output of the TPS. Such scheduled quality assurance program is frequently also referred to as "Periodic Quality Assurance". A recommended structure is given in: IAEA, "Commissioning and quality assurance of computerized planning systems for radiation treatment of cancer", TRS 430. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (113/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Scheduled Quality Assurance Example of a periodic quality assurance program (TRS 430) Hardware Patient specific Weekly Monthly Quarterly Annually Digitizer CPU Digitizer Plotter Backup After upgrade CPU Digitizer Plotter Backup Anatomical information CT transfer Anatomy CT image CT transfer CT image Anatomy External beam software Beam MU check Plan details Pl. transfer Plan details Pl. transfer Beam Pl. transfer Beam Pl. transfer IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 2 (114/117) 57
58 11.5 COMMISSIONING AND QUALITY ASSURANCE Scheduled Quality Assurance In addition, care must be given to in-house built treatment planning systems that are undocumented and undergo routine development. These TP systems may require quality assurance tests at a higher frequency. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 3 (115/117) 11.5 COMMISSIONING AND QUALITY ASSURANCE Scheduled Quality Assurance There is a common thread of continuity: Acceptance Commissioning: Data acquisition Data entry Patient specific dosimetry Treatment delivery The medical physicist must be able to link all these steps together. A well planned and scheduled set of quality assurance tests for the TP system is an important link in the safe delivery of therapeutic radiation. IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 4 (116/117) 58
59 11.6 SPECIAL CONSIDERATIONS Many TP systems are dedicated to special techniques (requiring a dedicated TP system) that require careful consideration, owing to their inherent complexity. Brachytherapy Orthovoltage radiotherapy IMRT Dynamic MLC Total body irradiation (TBI) Micro MLC Stereotactic radiosurgery Tomotherapy Intraoperative radiotherapy D shaped beams for choroidal melanoma Electron beam arc therapy Total skin electron irradiation IAEA Review of Radiation Oncology Physics: A Handbook for Teachers and Students Slide 1 (117/117) 59
Key words: treatment planning, quality assurance, 3D treatment planning
American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical
CHAPTER 11. COMPUTERIZED TREATMENT PLANNING SYSTEMS
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 11. COMPUTERIZED TREATMENT PLANNING SYSTEMS FOR EXTERNAL PHOTON BEAM RADIOTHERAPY MICHAEL D.C. EVANS Department of Medical
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review
NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
Performance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
1. Provide clinical training in radiation oncology physics within a structured clinical environment.
Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible
Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY
Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY W. PARKER, H. PATROCINIO Department of Medical Physics, McGill University Health Centre, Montreal, Quebec, Canada 7.1. INTRODUCTION
Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS W. STRYDOM Department of Medical Physics, Medical University of Southern Africa, Pretoria, South Africa W. PARKER, M. OLIVARES Department of Medical
THE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline
Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50
Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY
Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY D.I. THWAITES Department of Oncology Physics, Edinburgh Cancer Centre, University of Edinburgh, Western General Hospital, Edinburgh, United Kingdom
CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY WILLIAM PARKER Department of Medical Physics McGill
Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care
Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require
Manual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com
Serial Tomotherapy IGRT Total Solutions Treatment Planning Brachytherapy Imaging Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA BMI.v.08.2010 Best NOMOS Your Single Source Oncology Solutions Provider
RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
Lecture 14. Point Spread Function (PSF)
Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect
Introduction to the Monte Carlo method
Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw
Independent corroboration of monitor unit calculations performed by a 3D computerized planning system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and
DICOM: Definitions and Testing
DICOM: Definitions and Testing Bruce Curran University of Michigan (AAPM Representative, DICOM WG-7 Walter R. Bosch Advanced Technology Consortium Washington University of St. Louis Acknowledgements Sam
Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
IAEA-TECDOC-1540. Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems
IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems April 2007 IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems
Accreditation Is Coming
Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt
QA of intensity-modulated beams using dynamic MLC log files
36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,
Clinical Photon Beams
Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips
Intensity-Modulated Radiation Therapy (IMRT)
Scan for mobile link. Intensity-Modulated Radiation Therapy (IMRT) Intensity-modulated radiotherapy (IMRT) uses linear accelerators to safely and painlessly deliver precise radiation doses to a tumor while
Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y
Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How
INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS
INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS HOW DOES RADIATION THERAPY WORK? Cancer is the unrestrained growth of abnormal cells within the body. Radiation therapy kills both
CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS JOHN L. HORTON Department of Radiation Physics University of Texas
IBA Proton Therapy. Biomed days 2015. Vincent Bossier. System Architect [email protected]. Protect, Enhance and Save Lives
Vincent Bossier System Architect [email protected] IBA Proton Therapy Biomed days 2015 Protect, Enhance and Save Lives 1 Agenda AN INTRODUCTION TO IBA WHY PROTON THERAPY CLINICAL WORKFLOW TREATMENT
Radiation Oncology Centers Participating in MassHealth. Daniel Tsai, Assistant Secretary and Director of MassHealth
Executive Office of Health and Human s Office of Medicaid www.mass.gov/masshealth May 2015 TO: FROM: RE: Radiation Oncology Centers Participating in Daniel Tsai, Assistant Secretary and Director of (2015
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der
1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range
1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range The radiation from orthovoltage units is referred to as x-rays, generated by bombarding
Robot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
Our Department: structure and organization
EORTC meeting for Radiation Therapy Technologists: RTT s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled
Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients
대한방사선종양학회지 2005;23(3):169~175 Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients Department of Radiation Oncology, College of Medicine, Ewha Womans University
IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
Challenges in small field MV photon dosimetry
Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland [email protected] Why are
Quality Assurance for Treatment Planning Systems
Quality Assurance for Treatment Planning Systems PTCOG Educational Session May 19 th 2010 Oliver Jäkel Heidelberg Ion Beam Therapy Center and German Cancer Research Center, Heidelberg Introduction Outline
3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.
Efficacy of IMRT with dot decimal compensators for radiotherapy of malignant pleural mesothelioma post extrapleural pneumonectomy Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015
INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design
CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW
RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular
Prepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
Corporate Medical Policy
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Head and Neck File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_head_and_neck
Proton Therapy. What is proton therapy and how is it used?
Scan for mobile link. Proton Therapy Proton therapy delivers radiation to tumor tissue in a much more confined way than conventional photon therapy thus allowing the radiation oncologist to use a greater
Advanced variance reduction techniques applied to Monte Carlo simulation of linacs
MAESTRO Advanced variance reduction techniques applied to Monte Carlo simulation of linacs Llorenç Brualla, Francesc Salvat, Eric Franchisseur, Salvador García-Pareja, Antonio Lallena Institut Gustave
Quality Reports With. PlanIQ. *Export only. 510(k) pending. Patent pending. Your Most Valuable QA and Dosimetry Tools
Quality Reports With PlanIQ * *Export only. 510(k) pending. Patent pending. Your Most Valuable QA and Dosimetry Tools Objective Plan Assessment? How good are my treatment plans? Quality assurance of radiation
Common Core Unit Summary Grades 6 to 8
Common Core Unit Summary Grades 6 to 8 Grade 8: Unit 1: Congruence and Similarity- 8G1-8G5 rotations reflections and translations,( RRT=congruence) understand congruence of 2 d figures after RRT Dilations
ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS. Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic
ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic Disclosure I have received research grants from Siemens
CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.
1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has
Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer
Evaluation of complexity and deliverability of IMRT- treatment plans Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer June 11 Abstract Background: Intensity modulated beams are used
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
DICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes
First Three Years After Project Proton Therapy Facility:
39. Project Related Utilization and Financial Information. Proton Therapy Facility: Unit of Service Delivery Projected Utilization FY2017 FY2018 FY2019 Estimated Patients 448 670 808 Average Revenue Per
MEDICAL DOSIMETRY. COLLEGE OF APPLIED SCIENCES AND ARTS Graduate Faculty: therapy.
Graduate Catalog 2012-2013 Medical Dosimetry / 339 MEDICAL DOSIMETRY COLLEGE OF APPLIED SCIENCES AND ARTS Graduate Faculty: Collins, Kevin Scott, Associate Professor, Ph.D., Jensen, Steve, Emeritus Professor,
intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis 11/2009 5/2016 5/2017 5/2016
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Abdomen File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis
Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector
International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector
World-first Proton Pencil Beam Scanning System with FDA Clearance
Hitachi Review Vol. 58 (29), No.5 225 World-first Proton Pencil Beam Scanning System with FDA Clearance Completion of Proton Therapy System for MDACC Koji Matsuda Hiroyuki Itami Daishun Chiba Kazuyoshi
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
Page: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
Worldwide Quality Assurance networks for radiotherapy dosimetry
Worldwide Quality Assurance networks for radiotherapy dosimetry Joanna Izewska Dosimetry and Medical Radiation Physics Section [email protected] Audit in radiotherapy dosimetry Objectives to enhance confidence
Automotive Applications of 3D Laser Scanning Introduction
Automotive Applications of 3D Laser Scanning Kyle Johnston, Ph.D., Metron Systems, Inc. 34935 SE Douglas Street, Suite 110, Snoqualmie, WA 98065 425-396-5577, www.metronsys.com 2002 Metron Systems, Inc
Manual VIRTUAL RADIATION ONCOLOGY CLINIC (VROC) (1.1) Radiation Oncology Training. Director User Manual (1.1)
Manual 1 VIRTUAL RADIATION ONCOLOGY CLINIC (VROC) (1.1) Radiation Oncology Training Director User Manual (1.1) Virtual Radiation Oncology Clinic Resident User manual Chapter 1 Creating the Virtual Patient
Measurement with Ratios
Grade 6 Mathematics, Quarter 2, Unit 2.1 Measurement with Ratios Overview Number of instructional days: 15 (1 day = 45 minutes) Content to be learned Use ratio reasoning to solve real-world and mathematical
Albany Medical Center Chief Medical Physicist Radiation Oncology Imaging and Related Services
Albany Medical Center Chief Medical Physicist Radiation Oncology Imaging and Related Services POSITION SUMMARY: The Medical Physicist is responsible for assuring the high quality radiation therapy through
HITACHI Proton Beam Therapy System
HITACHI Proton Beam Therapy System Masumi Umezawa, M.S. Hitachi, Ltd., Hitachi Research Laboratory HITACHI Proton Beam Therapy System Contents 1. Overview of HITACHI 2. HITACHI Proton Beam Therapy and
QUALITY ASSURANCE OF TREATMENT PLANNING SYSTEMS PRACTICAL EXAMPLES FOR NON-IMRT PHOTON BEAMS
QUALITY ASSURANCE OF TREATMENT PLANNING SYSTEMS PRACTICAL EXAMPLES FOR NON-IMRT PHOTON BEAMS QUALITY ASSURANCE OF TREATMENT PLANNING SYSTEMS PRACTICAL EXAMPLES FOR NON-IMRT PHOTON BEAMS Ben Mijnheer Agnieszka
Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
The Canadian National System for Incident Reporting in Radiation Treatment (NSIR-RT) Taxonomy March 2, 2015 V2
The Canadian National System for Incident Reporting in Radiation Treatment (NSIR-RT) Taxonomy March 2, 2015 V2 Taxonomy Data Category Number Description Data Fields and Menu Choices 1. Impact 1.1 Incident
Current Standard: Mathematical Concepts and Applications Shape, Space, and Measurement- Primary
Shape, Space, and Measurement- Primary A student shall apply concepts of shape, space, and measurement to solve problems involving two- and three-dimensional shapes by demonstrating an understanding of:
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
Making Better Medical Devices with Multisensor Metrology
Making Better Medical Devices with Multisensor Metrology by Nate J. Rose, Chief Applications Engineer, Optical Gaging Products (OGP) Multisensor metrology is becoming a preferred quality control technology
Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate
Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_the_prostate
4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.
4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology [email protected] http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?
Oncentra Brachy Comprehensive treatment planning for brachytherapy. Time to focus on what counts
Oncentra Brachy Comprehensive treatment planning for brachytherapy Time to focus on what counts Time to focus on what counts In contemporary brachytherapy, the medical physicist needs to process an increasing
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
Alignment Laser System.
- O T 6 0 0 0 Alignment Laser System. The OT-6000. Multi -Target,Two Dimensional Alignment. Introducing the most powerful way to measure alignment at distances up to 300 feet. The OT-6000 Alignment Laser
Accreditation a tool to help reduce medical errors. Professor Arthur T Porter PC MD FACR FRCPC FACRO
Accreditation a tool to help reduce medical errors Professor Arthur T Porter PC MD FACR FRCPC FACRO Errors in Radiotherapy Radiation therapy is a highly regulated medical practice with historically low
Radiation Protection in Radiotherapy
Radiation Protection in Radiotherapy Albert Lisbona Medical Physics Department CLCC Nantes Atlantique 44805 Saint-Herblain France [email protected] Radiation therapy The lecture is oriented to
Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center
Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster Jesse Garant, JG&A Metrology Center Traditional Metrology and Inspection Tactile Devices (Touch Probe)
ENGINEERING METROLOGY
ENGINEERING METROLOGY ACADEMIC YEAR 92-93, SEMESTER ONE COORDINATE MEASURING MACHINES OPTICAL MEASUREMENT SYSTEMS; DEPARTMENT OF MECHANICAL ENGINEERING ISFAHAN UNIVERSITY OF TECHNOLOGY Coordinate Measuring
Failure Modes and Effects Analysis (FMEA)
Failure Modes and Effects Analysis (FMEA) Sasa Mutic Washington University School of Medicine St. Louis Missouri Failure Modes and Effects Analysis Objectives: To motivate the use of FMEA and to provide
Industrial Robotics. Training Objective
Training Objective After watching the program and reviewing this printed material, the viewer will learn the basics of industrial robot technology and how robots are used in a variety of manufacturing
In mathematics, there are four attainment targets: using and applying mathematics; number and algebra; shape, space and measures, and handling data.
MATHEMATICS: THE LEVEL DESCRIPTIONS In mathematics, there are four attainment targets: using and applying mathematics; number and algebra; shape, space and measures, and handling data. Attainment target
How to calibrate an RTD or Platinum Resistance Thermometer (PRT)
How to calibrate an RTD or Platinum Resistance Thermometer (PRT) Application Note Introduction There are two types of calibrations applicable to PRTs characterization and tolerance testing. The type of
Gamma Knife and Axesse Radiosurgery
Gamma Knife and Axesse Radiosurgery John C Flickinger MD Departments of Radiation Oncology & Neurological Surgery University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Origin of Radiosurgery
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images
multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images InterView FUSION InterView FUSION is a new visualization and evaluation software developed by Mediso built on state of
Big Ideas in Mathematics
Big Ideas in Mathematics which are important to all mathematics learning. (Adapted from the NCTM Curriculum Focal Points, 2006) The Mathematics Big Ideas are organized using the PA Mathematics Standards
Variance reduction techniques used in BEAMnrc
Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,