External dosimetry Dosimetry in new radiotherapeutic techniques
|
|
|
- Maryann Cooper
- 9 years ago
- Views:
Transcription
1 External dosimetry Dosimetry in new radiotherapeutic techniques Albert Lisbona Medical Physics Department CLCC René Gauducheau Saint-Herblain France
2 Objective : To describe external dosimetric methods used in new radiotherapeutic techniques.
3 Course objectives What are "new "radiotherapy techniques those using xray beams for : «Standard» - IMRT, helical tomotherapy, Linac based stereotactic radiotherapy (radiosurgery, ).
4 Radiotherapy objective Centre de Lutte Contre le Cancer To deliver the absorbed dose prescribed in the totality of the target volume at better than ± 5% while limiting to the maximum the irradiation of surrounding healthy tissues, organs Delivered dose =>Prescribed dose ±5%
5 External Radiotherapy Medical prescription : Dose variations from 5 to 10% to the target volume can lead to a significant change in the local tumor control and in the toxicity Random errors Human factor Systematic errors Network, TPS, beam calibration,...
6 Objective ( ) Imply : A precise knowledge of the physical and dosimetric properties of the beams used for radiotherapy (in house calibration, relative dose distributions, ). Perticular attention to pay during the different steps : Treatment preparation (anatomic patient data acquisition, choice of the radiotherapy technique, medical prescription ), Treatment delivery (patient positionning, in vivo dosimetry, controls).
7 Detectors
8 To know the detectors used in radiotherapy beams (limitations eg : detector size vs field size) To use in connection with radiotherapy technique.
9 Some air cavity ionisation chambers chambres used in radiotherapy beams Ionisation chambers 0,1cm 3 < Vair < 1cm 3 7 mm < internal diameter < 25 mm - low Z for thin wall thickness - wall material air or water equivalent RX or γ Co 60 E- HE Capuchon d équilibre électronique d épaisseur adaptée au Co 60 dans l air
10
11 Reference detector for radiotherapy = ionisation chamber calibrated in a Second Standard Dosimetry Lab Centre de Lutte Contre le Cancer application = «absolute dosimetry» and relative dosimetry Main advantages and characteristics of the ionisation chamber - well known geometric and physics properties - good response reproducibility - low response dependance with energy, dose, dose rate - calibration in a SSDL every 3 ans (France) "Absolute" Dosimétry : Reference dose rate measurement
12 Relative dosimetry : Check of symetry and homogeneity beam 1D, 2D ou 3D measurements
13 DIODES : PRINCIPLE Centre de Lutte Contre le Cancer irradiation electron-hole creation current absorbed dose
14 MOSFET : Metal Oxyde Semiconductor Field Effect Transistor Centre de Lutte Contre le Cancer Characteristics : Si small transistor (2 p zones inserted in n zone) isolant métallique Principle : A polarization voltage is applied during irradiation Th irradiation induces an increase of the transistor threshold voltage The measurement of the threshold voltage is a linear function of absorbed dose
15 RADIOTHERMOLUMINESCENCE DOSIMETRY PRINCIPE
16
17 Photographic film Films Kodak Dose sensitivity Conversion curve OD vs dose depending on film characteristics Dose rate sensitivity Film response is independant versus dose rate (for the dose rate range used for radiotherapy)
18 Radiochromic Film Characteristics New type of film for dosimetry in radiotherapy - one of the more used : Gafchromic film Gafchromic film composition - 9,0 % H - 60,6 % C -11,2 % Zn -19,2 % O > composition closed to human tissue, better tissue equivalent (Z between 6 et 6.5, closer than radiographic film) Non irradied Gafchromic film is incolore > blue coloration when irradiated
19 Principle Centre de Lutte Contre le Cancer Under irradiation radiochromic film polymerises due to a special included colourant Self-processsing : Definitive film coloration < 24 h after irradiation The polymer on the support absorbes the light so transmission light through the film can be measured with a flat scanner (or a bed scanner) Optical density measured for a narrow wavelength (choice of a light source adapted to the film in order to get the best sensitivity, response) OD vs dose linear relation can be established under calibration
20 IMRT
21 IMRT Today several technical solutions to deliver IMRT treatments : «Standard» IMRT (static mlc beams), Rotational IMRT (VMAT, IMAT), Helical radiotherapy (Tomotherapy).
22 3D CRT To increase target dose with sparing of healthy tissue. Use of adapted equipments : 3D dose calculation software (convolutionsuperposition algorithms), Linac, MLC, Real-time imaging devices. IMRT To increase target dose with sparing of healthy tissue by 3D CRT optimization. Same equipments + : Inverse planning software using doses-volumes constraints (knowledge data bases).
23 The linac
24 The multi leaves collimator (from 3 mm to 10 mm width size at isocenter)
25 Treatment Planning System A 3D software application using : A calculation software supporting convolution superposition algorithms, An inverse planning software planification for IMRT. SV PTV = Σ i=1 N [ IW i PTV ( Di PTV -GD PTV )2 ] SV OAR = Σ j=1 M [ IW j OAR ( D j OAR -GD OAR )2 ]
26 Example : Head & Neck and IMRT
27 Ballistic 4 MV X-ray 5 beams with following gantry angulations : 0, 72, 144, 216, 288
28 Example of doses-volumes Centre de Lutte Contre le Cancer constraints in H&N Target volumes objectives 70 Gy to the therapeutic volume at least 95 % of the dose delivered to 95 % of the target volume. 50 Gy to the prophylactic volumes (delineation as Amsterdam-Brusells medical consensus) at least 95 % of the dose delivered to 95 % of the target volumes. Organs At Risk (OAR) constraints Spinal cord : 100 % vol < 46 Gy Larynx : 50 % vol < 40 Gy Left and Right parotids : 50 % vol < 30 Gy
29
30
31 Patient setup
32 Centre de Lutte Contre le Cancer MLC Segmentation
33 Qualitative evaluation 3D CRT IMRT
34 Qualitative evaluation 3D CRT IMRT
35 Qualitative evaluation 3D CRT IMRT
36 Qualitative evaluation 3D CRT IMRT
37 Quantitative evaluation 3D CRT IMRT
38 IMRT Pre treatment verification
39 Objectives To validate the calculation by a measurement To "treat" one fraction of a phantom (anthropomorphic shape, plastic slabs) with the fluences defined and validated by the physician and the physicist for the patient treatment. In the phantom we can introduce different detectors : Ionization chambers, Radiographic, radiochromic films, 2D arrays detectors
40 Equipment pre-treatment verification Centre de Lutte Contre le Cancer Parallelepipedic Phantom (RW3) DSP 95 cm Each treatment beam in a simple, reproducible geometry 5 cm An ionisation chamber can be inserted and/or a radiographic film between 2 slabs of tissue equivalent material Beam per beam Verification
41 H&N rao IMRT Field A dosimetric film (photographic or radiochromic) is calibrated (OD vs dose) and directly compared to the calculated TPS dose distribution.
42 H&N rao IMRT Field
43 H&N ant IMRT Field
44 H&N ant IMRT Field Comparison of two dose matrices (calculated, measured).here the criterions are dose difference (3 %) and distance to agreement (3 mm). Values exceding 1 fail the criteria.
45 Patient specific QA - Equipment Centre de Lutte Contre le Cancer H&N phantom from PTW (RW3 water equivalent material) insertion of a small volume detection ionisation chamber insertion of a radiographic film between 2 slabs Global verification using the patient treatment ballistic validated
46 Contrôle qualité Plan de traitement Analyse qualitative Centre de Lutte Contre le Cancer Dose profiles comparison Calculated and measured isodoses
47 External audits or comparisons between delivered and calculated Centre de Lutte Contre le Cancer doses Results from studies of the accuracy of dose determinations of IMRT treatments. (GUIDELINES FOR THE VERIFICATION OF IMRT- ESTRO Booklet 9)
48 Linac based Stereotactic Radiosurgery Stereotactic Radiotherapy
49
50 Stereotactic RadioSurgery - SRS BrainLab frame
51 Stereotactic RadioTherapy - SRT intracranial fractionné Bivalb mask BrainLab with mouth support
52 Dedicated Linac - NOVALIS Centre de Lutte Contre le Cancer Novalis Technical Data Integrated M3 micro-mlc 10x10 cm² 26 pairs of tungsten leaves 60mm thickness 14 pairs 3mm width at isocenter (42x42mm²) 6 pairs 4.5mm width at isocenter 6 pairs 5.5mm width at isocenter Leaves Transmission < 4% Leaves speed : 1.5 cm/s
53 Radiosurgical circular cones Centre de Lutte Contre le Cancer Radiosurgical circular cones ranging from Ø 4 mm to 15 mm Patient frame (or mask) support with 5 degrees of freedom
54 Acoustic neuroma : 1 single fraction % Centre de Lutte Contre le Cancer
55 Radiosurgery Trigeminal Neuralgia : 1 single fraction 80 Gy Centre de Lutte Contre le Cancer
56 Cerebral ArterioVenous Malformations Centre de Lutte Contre le Cancer AVM 22 80% - 1 fraction 6 DCA
57
58 Dedicated Linac - NOVALIS Centre de Lutte Contre le Cancer Caractéristiques techniques : Novalis Integrated BrainLab M3 micro-mlc 10x10 cm²
59 Dosimetry measurements Centre de Lutte Contre le Cancer Percentage Depth Dose (PDD) C McKerracher and D I Thwaites - Phys. Med. Biol. 44 (1999) Ø 40 mm Ø 12.5 mm
60 Dosimetry measurements Centre de Lutte Contre le Cancer Off Axis Ratio - OAR C McKerracher and D I Thwaites - Phys. Med. Biol. 44 (1999) Ø40 mm Ø12.5 mm
61 Off Axis Ratio - OAR Dosimetry measurements George X Ding - Phys. Med. Biol. 51 (2006) MV Various detectors Field size 6x6 mm²
62 Off Axis Ratio - OAR Dosimetry measurements Detector choice spatial resolution Diode - diamond for very narrow beams Pin Point ionisation chamber for other field sizes Alternative : photographic or radiochromic film Jaws settings (Primary collimation) influence minimlc and circular cones Jaws settings values too small (vs mlc or cone size) can lead to a large degradation of the small field size OAR
63 Dosimetry measurements Centre de Lutte Contre le Cancer Off Axis Ratio - OAR George X Ding - Phys. Med. Biol. 51 (2006) MV Field size 6x6 mm² Jaws settings leaves
64 Output Factor (Scatter Factor) As for PDD and OAR this measurement becomes more difficult to perform as field size decreases The delivered dose to the patient is directly proportionnal to the OF measured value. Drawbacks : Dosimetry measurements Size detector vs field size Lateral Electronic Equilibrium Problem Detectors Ionization chamber Pin Point type cm³ Diamond detector Diode Radiographic or radiochromic film
65 Dosimetry measurements Centre de Lutte Contre le Cancer Output Factor George X Ding - Phys. Med. Biol. 51 (2006) For a minimlc BrainLab m3
66 Output Factor (Scatter Factor) MiniMLC : SFD Diode vs PinPoint 0.015cm³ Dosimetry measurements Novalis 6MV Xray beam 1,30 1,20 1,10 Output Factor Diode/Pin Point Normalized to 100x100 1,00 0,90 0,80 0,70 0,60 Pin Point 0,015cm3 Diode SFD Ratio Diode/PP Dose difference = 25 % 0,50 0, MLC Field Size [mm] Most important differences below 12x12mm², for clinical situations this difference is reduced by field size shaped X-jaws and Y-jaws settings 2mm more than leaves settings An under response is reported for diode (2%) between 18x18 et 60x60 mm²
67 Dosimetry measurements Centre de Lutte Contre le Cancer Output Factor (Scatter Factor) Circular cones : measurement comparisons between differents detectors Novalis 6MV Xray beam Scatter Factors circular cones - CRG Diamant Nantes PinPoint Nantes Solberg Diode SFD Nantes Film EDR2 1 0,9 Dimensions 4 to 15 mm Dode relative 0,8 0,7 0,6 0,5 0, Jaws settings 50x50mm² Important differences below 10mm Diametre des collimateurs en mm
68 There is no international dosimetry code of practice (IAEA 2010?) for the narrow beams in radiotherapy (IMRT, radiosurgery, )
69 End to End checks
70 End to End checks Jaws settings X mm Y mm Mes/Cal 0,918 0,979 0,995 0,984
71 Conclusion Centre de Lutte Contre le Cancer Mesures dans les minifaisceaux sont délicates Instruments de mesures spécifiques adaptés aux mesures à réaliser investissements Moyens humains et temps machine Minutie Contraintes techniques liées aux mesures Connaître les limites : matériel, détecteurs
72 Conclusion Adequate staff, and sufficient training for all the members of the team involved in new techniques. Guidance, recommendation on quality assurance for new techniques (using new concepts based on risk analysis). Dosimetric audits for comparison and validation.
73 Gracias Thank you
Radiation Protection in Radiotherapy
Radiation Protection in Radiotherapy Albert Lisbona Medical Physics Department CLCC Nantes Atlantique 44805 Saint-Herblain France [email protected] Radiation therapy The lecture is oriented to
THE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
1. Provide clinical training in radiation oncology physics within a structured clinical environment.
Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible
QA of intensity-modulated beams using dynamic MLC log files
36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,
Challenges in small field MV photon dosimetry
Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland [email protected] Why are
Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y
Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How
RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer
Evaluation of complexity and deliverability of IMRT- treatment plans Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer June 11 Abstract Background: Intensity modulated beams are used
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW
RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular
MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline
Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50
Quality Assurance in Stereotactic. Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
Gamma Knife and Axesse Radiosurgery
Gamma Knife and Axesse Radiosurgery John C Flickinger MD Departments of Radiation Oncology & Neurological Surgery University of Pittsburgh School of Medicine Pittsburgh, Pennsylvania Origin of Radiosurgery
Worldwide Quality Assurance networks for radiotherapy dosimetry
Worldwide Quality Assurance networks for radiotherapy dosimetry Joanna Izewska Dosimetry and Medical Radiation Physics Section [email protected] Audit in radiotherapy dosimetry Objectives to enhance confidence
NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
Independent corroboration of monitor unit calculations performed by a 3D computerized planning system
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and
Quality Assurance of accelerators; the technologists responsibility
Quality Assurance of accelerators; the technologists responsibility Christa Timmermans Radiation Technologist, Erasmus MC- Daniel den Hoed, Rotterdam, The Netherlands EORTC-ROG RT technologists Section
IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD
Good Morning Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD Barbara Agrimson, BS RT(T)(R), CMD Steve Rhodes, BS RT(T), CMD Disclaimer This presentation
Radiosurgery 8/16/2011. Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago
Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago Radiosurgery The use of radiation as a surgical tool Small volumes of tissues
Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014
Multi-Channel Radiochromic Film Dosimetry Adapted from A.Micke Spain, April 2014 Single Channel Film Dosimetry Calibration Curve X=RR ave = R ave (D) D R = D R (R ave ) Color channels X=RGB D X = D( X
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of
Dosimetric impact of the 160 MLC on head and neck IMRT treatments
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 6, 2014 Dosimetric impact of the 160 MLC on head and neck IMRT treatments Prema Rassiah-Szegedi, a Martin Szegedi, Vikren Sarkar, Seth Streitmatter,
Our Department: structure and organization
EORTC meeting for Radiation Therapy Technologists: RTT s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled
First Three Years After Project Proton Therapy Facility:
39. Project Related Utilization and Financial Information. Proton Therapy Facility: Unit of Service Delivery Projected Utilization FY2017 FY2018 FY2019 Estimated Patients 448 670 808 Average Revenue Per
ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS. Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic
ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic Disclosure I have received research grants from Siemens
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
Total Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com
Serial Tomotherapy IGRT Total Solutions Treatment Planning Brachytherapy Imaging Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA BMI.v.08.2010 Best NOMOS Your Single Source Oncology Solutions Provider
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care
Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require
Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
$ " %#&&%"'( &%")$#*((+,-.##/ $01&0112 344 5*/(&%"2&0116 '(! (## " *# )#! &%") 78 ' ** #* ( 9! '(' :! 9* ## '$ %# & &$%%&%!
78/0
Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY
Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY D.I. THWAITES Department of Oncology Physics, Edinburgh Cancer Centre, University of Edinburgh, Western General Hospital, Edinburgh, United Kingdom
Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.
Efficacy of IMRT with dot decimal compensators for radiotherapy of malignant pleural mesothelioma post extrapleural pneumonectomy Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan
The feasibility of a QA program for ISIORT Trials
Baveno, Italy June 22-24, 2012 The feasibility of a QA program for ISIORT Trials Frank W. Hensley 1, Don A. Goer 2, Sebastian Adamczyk 3, Falk Roeder 1, Felix Sedlmayer 4, Peter Kopp 4 1 University Clinics
Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021
Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021 Introduction Prostate is one of the treatment sites that is well suited for IMRT. For
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
Key words: treatment planning, quality assurance, 3D treatment planning
American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical
Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
Clinical Study Monte Carlo-Based Dose Calculation in Postprostatectomy Image-Guided Intensity Modulated Radiotherapy: A Pilot Study
Radiotherapy Volume 215, Article ID 682463, 6 pages http://dx.doi.org/1.1155/215/682463 Clinical Study Monte Carlo-Based Dose Calculation in Postprostatectomy Image-Guided Intensity Modulated Radiotherapy:
Clinical Photon Beams
Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips
Radiation Oncology Centers Participating in MassHealth. Daniel Tsai, Assistant Secretary and Director of MassHealth
Executive Office of Health and Human s Office of Medicaid www.mass.gov/masshealth May 2015 TO: FROM: RE: Radiation Oncology Centers Participating in Daniel Tsai, Assistant Secretary and Director of (2015
What is the Radiotherapy Quality Control Program (PQRT) of the National Cancer Institute Rio de Janeiro/Brazil?
Congreso Internacional Conjunto Cancún 2004 LAS/ANS-SNM-SMSR/International Joint Meeting Cancun 2004 LAS/ANS-SNM-SMSR XV Congreso Anual de la SNM y XXII Reunión Anual de la SMSR/XV SNM Annual Meeting and
Analysis of Trajectory Log Files of TrueBeam Medical Electron Linear Accelerator for Patient Specific IMRT QA
Cancer Research and Oncology: Open Access Received: Nov 16, 2015, Accepted: Jan 12, 2016, Published: Jan 15, 2016 CancerResOncol, Volume 1, Issue 1 http://crescopublications.org/pdf/crooa/crooa-1-001.pdf
Radiation Therapy Coverage, Coding, and Reimbursement for New Technologies
Radiation Therapy Coverage, Coding, and Reimbursement for New Technologies Presented by Susan Granucci, Healthcare Reimbursement Specialist Topics 1. Intensity Modulated Radiation Therapy (IMRT) 2. Tomotherapy
Corporate Medical Policy
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Head and Neck File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_head_and_neck
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der
Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015
INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design
Failure Modes and Effects Analysis (FMEA)
Failure Modes and Effects Analysis (FMEA) Sasa Mutic Washington University School of Medicine St. Louis Missouri Failure Modes and Effects Analysis Objectives: To motivate the use of FMEA and to provide
A SAFE, NON-INVASIVE TREATMENT OPTION: GAMMA KNIFE PERFEXION
A SAFE, NON-INVASIVE TREATMENT OPTION: GAMMA KNIFE PERFEXION Not actually a knife, the Gamma Knife Perfexion is an advanced radiosurgery device which uses extremely precise beams of radiation to treat
Measuring Patient Exposure in Interventional Radiology
Measuring Patient Exposure in Interventional Radiology GAFCHROMIC XR-R, a Wide-Area Dosimetry Film to Assess Peak Skin Dose (PSD) Xiang Yu, Ph.D. Director, Advanced Materials International Specialty Products
at a critical moment Physician Suggestion Line...
Radiation Oncology Exceptional care at a critical moment When your patients require radiation therapy, they deserve the very best care available to them. The Department of Radiation Oncology provides exceptional
A Thesis. entitled. Retrofitted Micro-MLC SRS System. George H. Hancock
A Thesis entitled Development of a Comprehensive Linac-based Quality Assurance Program for a Retrofitted Micro-MLC SRS System. by George H. Hancock Submitted to the Graduate Faculty as partial fulfillment
REPORT ORGANIZATION ABSTRACT I. INTRODUCTION CALCULATION OF EFFECTIVE DOSE PER IMAGE ACQUISITION CALCULATION OF EFFECTIVE DOSE PER TREATMENT COURSE
E S T I M A T I O N O F T H E I M A G I N G D O S E F O R T H E C Y B E R K N I F E R O B O T I C R A D I O S U R G E R Y S Y S T E M REPORT ORGANIZATION ABSTRACT I. INTRODUCTION II. III. IV. CALCULATION
LINAC-based stereotactic radiosurgery for treatment of trigeminal neuralgia
LINAC-based stereotactic radiosurgery for treatment of trigeminal neuralgia Bruce J. Gerbi, Patrick D. Higgins, 1 Kwan H. Cho, 2 and Walter A. Hall 3 Department of Therapeutic Radiology-Radiation Oncology,
Transit Dosimetry for Patient Treatment Verification with an Electronic Portal Imaging Device. Sean L Berry
Transit Dosimetry for Patient Treatment Verification with an Electronic Portal Imaging Device Sean L Berry Submitted in partial fulfillment of the requirements for the degree of Doctor of Philosophy in
Performance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
An Integer Programming Approach to Conversion from Static to Continuous Delivery of Intensity Modulated Radiation Therapy
Elin Hynning An Integer Programming Approach to Conversion from Static to Continuous Delivery of Intensity Modulated Radiation Therapy Master Thesis November, 2011 Division of Optimization and Systems
Basic Radiation Therapy Terms
Basic Radiation Therapy Terms accelerated radiation: radiation schedule in which the total dose is given over a shorter period of time. (Compare to hyperfractionated radiation.) adjuvant therapy (add-joo-vunt):
Department of Radiation Oncology
Department of Radiation Oncology Welcome to Radiation Oncology at Emory Clinic Every member of Emory Clinic Department of Radiation Oncology strives to provide the highest quality of care for you as our
World-first Proton Pencil Beam Scanning System with FDA Clearance
Hitachi Review Vol. 58 (29), No.5 225 World-first Proton Pencil Beam Scanning System with FDA Clearance Completion of Proton Therapy System for MDACC Koji Matsuda Hiroyuki Itami Daishun Chiba Kazuyoshi
Baylor Radiosurgery Center
Radiosurgery Center Baylor Radiosurgery Center Sophisticated Radiosurgery for both Brain and Body University Medical Center at Dallas Radiosurgery Center 3500 Gaston Avenue Hoblitzelle Hospital, First
MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015)
MVP/Care Core National 2015 Radiation Therapy Prior Authorization List (Effective January 1, 2015) CODE CODE TYPE Description 0330 REVCODE RADIOLOGY/THERAPEUTIC - GENERAL CLASSIFICATION 0333 REVCODE RADIOLOGY/THERAPEUTIC
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review
intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis 11/2009 5/2016 5/2017 5/2016
Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Abdomen File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis
CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS JOHN L. HORTON Department of Radiation Physics University of Texas
Evolution of Head and Neck Treatment Using Protons. Mayank Amin, M.Sc,CMD
Evolution of Head and Neck Treatment Using Protons Mayank Amin, M.Sc,CMD Facility Layout Gantry Room 3 Fixed Beams Room 4 HEBT Gantry Room 2 Gantry Room 1 Synchrotron Linac Treatment Planning Imaging Area
High-Energy Electron Beam Therapy Dosimetry with Ionisation Chambers
Schweizerische Gesellschaft für Strahlenbiologie und Medizinische Physik Société Suisse de Radiobiologie et de Physique Médicale Società Svizzera di Radiobiologia e di Fisica Medica Swiss Society of Radiobiology
GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2
GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures
Dose Tracking System:
Dose Tracking System: A Paradigm Shift in Patient Dose Monitoring Andrew Kuhls-Gilcrist, Ph.D. XRVL Clinical Science Manager Toshiba America Medical Systems, Inc. 2 Dose Tracking System: A Paradigm Shift
IBA Proton Therapy. Biomed days 2015. Vincent Bossier. System Architect [email protected]. Protect, Enhance and Save Lives
Vincent Bossier System Architect [email protected] IBA Proton Therapy Biomed days 2015 Protect, Enhance and Save Lives 1 Agenda AN INTRODUCTION TO IBA WHY PROTON THERAPY CLINICAL WORKFLOW TREATMENT
Chapter XIII. PDD TAR TMR Dose Free Space
Ver 2.11 Chapter XIII PDD TAR TMR Dose Free Space Joseph F. Buono RTT Allied Health Science Nassau Community College 1 Education Drive Garden City, NY 11530-6793 phone: 516-572 - 7536 oice - 9460 Secretary
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.
4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology [email protected] http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?
Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT
Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT Wen Li, Andrew Vassil, Lama Mossolly, Qingyang Shang, Ping Xia Department of Radiation Oncology Why
Advanced variance reduction techniques applied to Monte Carlo simulation of linacs
MAESTRO Advanced variance reduction techniques applied to Monte Carlo simulation of linacs Llorenç Brualla, Francesc Salvat, Eric Franchisseur, Salvador García-Pareja, Antonio Lallena Institut Gustave
ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY
TECHNICAL REPORTS SERIES No. 398 Absorbed Dose Determination in External Beam Radiotherapy An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water Sponsored by the
Accreditation Is Coming
Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt
Acknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers
Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers H. Helen Liu, PhD Department of Radiation Physics, U.T. MD Anderson Cancer Center, Houston, TX AAPM, Seattle, 2005 Acknowledgement
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction
Horizon Blue Cross Blue Shield of New Jersey 2012 Radiation Therapy Payment Rules
55875 55876 55876 Description Transperineal placement of needles or catheters into prostate for interstitial radioelement application, with or wout cytosocopy 55920 Placement of interstitial device(s)
What is the CyberKnife System?
CYBERKNIFE Robotic Radiosurgery System PATIENT BROChURE 1 What is the CyberKnife System? have stabilizing frames bolted to their head or limit their breathing during treatment to minimize movement of the
GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3
GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures
Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study
Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu
VOLUMETRIC INTENSITY-MODULATED ARC THERAPY VS. CONVENTIONAL IMRT IN HEAD-AND-NECK CANCER: A COMPARATIVE PLANNING AND DOSIMETRIC STUDY
doi:10.1016/j.ijrobp.2008.12.033 Int. J. Radiation Oncology Biol. Phys., Vol. 74, No. 1, pp. 252 259, 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/09/$ see front
02 CyberKnife: Treatment Delivery
TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options
Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy
Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy This set of 117 slides is based on Chapter 11 authored by M.D.C. Evans of the IAEA publication (ISBN 92-0-107304-6):
Comparison of Epson scanner quality for radiochromic film evaluation
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Comparison of Epson scanner quality for radiochromic film evaluation Hani Alnawaf, 1,3 Peter K.N. Yu, 2 Martin Butson 1,2,3a South
SEAAPM Symposium The Practice of Quality Assurance in an Era of Change
SEAAPM Symposium The Practice of Quality Assurance in an Era of Change Directors: Dan Bourland and Ingrid Marshall Vendors, breaks, and food next door Thank you to SEAAPM EXCOM, Commercial Partners, Faculty,
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS
Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS W. STRYDOM Department of Medical Physics, Medical University of Southern Africa, Pretoria, South Africa W. PARKER, M. OLIVARES Department of Medical
Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate
Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_the_prostate
IMRT: dal presente al futuro L esperienza italiana
IMRT: dal presente al futuro L esperienza italiana Giuseppe Scielzo U.O.A. Fisica Sanitaria Ordine Mauriziano Reggio Emilia, 14-15 Ottobre 2003 Contents IMRT in Italy - AIFM BEV -AIRO BEV - OEV Breaking
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed
Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 2, 2013 Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification Valeria Casanova Borca, Massimo Pasquino, a