User Guide. Model PH210. ph / ORP / Temperature Meter

Similar documents
User Guide. Heavy Duty Dissolved Oxygen Meter. Model

ExStik TM ph Waterproof Meters

User's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

Hygro-Thermometer + InfraRed Thermometer Model RH101

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

4-Channel Thermometer / Datalogger

User's Guide. Pinless Moisture Meter. Model MO257

Care and use guide. Contents. a handy solution made easy. page

User's Guide. Digital Sound Level Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

Ground Resistance Clamp On Tester

Milwaukee USER MANUAL

ph Measurement and Control

User Guide. Model Insulation Tester / Megohmmeter

Wide Range InfraRed (IR) Thermometer with Type K input and Laser Pointer

Integrating Sound Level Meter and Datalogger

User Guide. Temperature and Humidity Datalogger. Model 42280

Instruction Manual Temp 4/5/6 Temperature Meter Series

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL

Care and use guide. Contents. page

Care and use guide. Contents. a handy solution made easy. page

Portable Waterproof ph/dissolved Oxygen Meter

Professional Thermohygrometer

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Mini InfraRed Thermometer with Laser Pointer

AutoRanging Digital MultiMeter

Bluelab Soil ph Meter Instruction Manual

USER GUIDE. LED Light Meter. Model LT45

ph 200 ph-/redox-/temperature Meter Operating instructions

UltraBasic Benchtop ph Meter Operation Manual

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

SenTix 950 SenTix 980

AIRMoN, ph and Conductivity Field Measurement

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen

Manual Ranging MultiMeter

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

User's Guide Integrating Sound Level Datalogger Model

User Manual. Humidity-Temperature Chart Recorder. Model RH520

Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso

RS DIGITAL THERMOMETER

Operating Instructions

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

Care and use guide. soil ph meter. Contents. page

User's Guide. Integrating Sound Level Datalogger. Model Introduction

Dual Laser InfraRed (IR) Thermometer

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:

High Temperature InfraRed Thermometer with Laser Pointer

RS Stock No Instruction Manual RS-1340 Hot Wire Anemometer

Thermistor Thermometer MODEL NO

User's Guide. True RMS Industrial Multimeter

Instruction Manual HI Refractometer for Refractive Index and Brix Measurements

Microwave Meter. Instruction Manual

Relative Humidity Calibration Kit

Digital Air Pressure Gauge Users Manual

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN In Use Date: 2/1/2011 Discontinued Date:

ph 300 and 310 Portable Waterproof ph/mv/ C Meter

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

DISSOLVED OXYGEN PROBE

Instruction Manual HI Portable Waterproof Microprocessor Dissolved Oxygen Meter

Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented

Ground Resistance Clamp-On Tester

Temperature Humidity Meter

Compact ph Measuring Instruments

Appendix C. Vernier Tutorial

Model 4510 Conductivity/TDS Meter Operating Manual

Light Meter (Foot Candles)

USEPA Electrode Method Method 8156

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL

INSTRUCTION MANUAL DTH Fax: (503)

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Measurement by Ion Selective Electrodes

User Guide. 4 wire Earth Resistance Tester Model GRT300

Instruction Manual DO 110. Hand-held Dissolved Oxygen Meter with RS 232C. 68X Rev. 0-11/05. Technology Made Easy...

User Guide Dissolved Oxygen Probes

Dual Laser InfraRed (IR) Thermometer with Color Alert

Dual Laser InfraRed (IR) Thermometer

UT202A Operating Manual. Contents

EM90 INFRA-RED THERMOMETER & THERMOCOUPLE METER

SOUND LEVEL METER 94405

SINGLE RANGE DS METERS Operation Instructions

CONDUCTIVITY SENSOR. Description D0382. Figure 1. The Conductivity Sensor

Auto-ranging Digital Multimeter INSTRUCTION MANUAL

Thermocouple Thermometer Models

PART 1 - INTRODUCTION...

Refer to page 6 to page 19 for detailed instructions.

Pro1030 USER MANUAL. English

Thermocouple Thermometer

Chem 1B Saddleback College Dr. White 1. Experiment 8 Titration Curve for a Monoprotic Acid

MAKING MODERN LIVING POSSIBLE. living connect. Installation and User Guide. Danfoss Heating Solutions

Nokia Wireless Keyboard (SU-8W) User Guide

Your Simple Guide to Battery. Replacement. Customer Care:

ELECTROCHEMISTRY. Theory and Practice

Compact Water Quality Meter B-700 series

Dual Laser InfraRed (IR) Thermometer

Instruction Manual HI Refractometer for Sucrose Measurements

6 H2O + 6 CO 2 (g) + energy

Transcription:

User Guide Model PH210 ph / ORP / Temperature Meter

Introduction Thank you for selecting the Extech PH210 ph/orp/temperature meter. This microprocessorbased device with tactile push buttons is battery operated and ideal for field use. The PH210 simultaneously displays ph or mv (ORP) and solution temperature measurements. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Features LCD display shows ph or mv (ORP) and solution temperature simultaneously ph measurement range 0 to 14.00pH mv (ORP) measurement range ± 1999mV BNC input socket accommodates most ph or ORP electrodes Detachable probe for ease of use in a variety of measurement environments Automatic (via optional temperature probe) or Manual temperature compensation High impedance ph measurements User ph calibration ability DATA HOLD for freezing displayed reading Records MIN MAX readings Selectable function (ph or mv), Temperature measurement units (C/F), Auto power off disable/enable, and Manual solution temperature compensation Meter Description 1. Measurement reading 2. Solution temperature 3. Unit of measure (ph or mv) 4. Power/Escape button 5. Function/Hold/Up arrow button 6. Record (MIN/MAX)/Enter button 7. Setting/down arrow button 8. Battery compartment (rear of meter) 9. ph or mv Electrode BNC connector input 10. Optional ATC probe input jack 1 2 3 4 5 6 7 8 9 10 2

Getting Started 1. Calibrate the instrument often, as described in a later section. Note that only a ph calibration is required; mv (ORP) is calibrated as a by product of the ph calibration. 2. Set the unit of measure for solution temperature ( o C or o F) as described in the Temperature Units Selection section of this guide. 3. Determine if Automatic or Manual temperature compensation will be employed. Refer to the Temperature Compensation section of this guide for details. 4. Follow the instructions and electrode cleaning/rinsing recommendations closely as slight deviations in test procedures can have a significant effect on test results. Calibration (ph) Calibration Considerations This ph meter was factory calibrated by mv signals simulating an ideal ph Electrode in a 25 C (77 F) environment. An ideal electrode will produce 0.0mV at ph of 7.00 in ideal conditions. Electrodes are imperfect however and the temperature at which tests are taken is not always ideal. To maintain a high level of accuracy, and to accommodate any given application, choose a calibration method from the selections below: Basic Calibration (single point calibration): ph 7.00 Complete Calibration (two points): ph 7.00 then ph 4.00 or ph 7.00 then ph 10.00 The meter should be calibrated once per testing session. Calibration is performed using standard buffer solutions at room temperature 25 C (77 F). Calibration should always be performed using the ph 7.00 buffer first, followed by a second calibration buffer if desired. Select the ph 4.00 buffer for the second calibration if typical measurements will be lower than ph 7.00 or use a ph 10.00 buffer for the second calibration if typical measurements will be greater than ph 7.00. If the display shows ERR, clear the existing calibration data as discussed in the ph Calibration Procedure below, and perform the calibration again. ph Calibration Procedure 1. Place the electrode in a buffer solution. Power the meter ON and simultaneously press and release the REC and HOLD buttons; CAL appears in the lower left corner of the display. 2. Use the up and down arrow buttons to scroll through the 4, 7, and 10 calibration points and the Clear function (CLr). Once the correct point is selected press the ENTER button. The reading will flash for several seconds and then END will appear. Calibrate to the other buffers in the same way. Note: If there is a deviation of more than 1 ph unit from the 4, 7, or 10 ph buffer, or if the electrode slope is low, the instrument will assume an error and abort the calibration (ERR will be displayed). In this case, clear the calibration data (see step 2 above) and perform the calibration again. If the problem persists, the electrode may require replacing. 3. For multi point calibration, use the arrow buttons to scroll through the calibration points as described in step 2. Always calibrate with a ph 7 buffer first, and then follow with ph 4 or ph 10. Rinse electrode and probe with distilled water between calibration buffers. 4. Press ESC at any time to abort the calibration and return to the normal operating mode. 3

Measuring ph Important: Electrodes and temperature probes should always be rinsed before and between samples with distilled water to remove all traces of storage solution, process media, or previous test solution. 1. Connect the ph electrode to the meter (BNC connection at bottom of meter). 2. Optionally connect the temperature probe (Extech model 850188 only) to the jack at the bottom of the meter. This temperature probe is used for automatic temperature compensation (ATC). Manual temperature compensation is possible as explained below. 3. Turn the meter on using the POWER/ESC button. 4. Press and hold the FUNCTION button to select ph units if necessary. The unit of measure appears on the LCD (ph). 5. To change the solution temperature units ( o C / o F) refer to the Temperature Units Selection section of this guide. 6. Rinse the ph electrode and temperature probe with distilled water. 7. Place the electrode and temperature probe (if ATC is to be used) in the solution. Allow 30 seconds for the electrode and ATC probe to reach thermal equilibrium with the buffer solution. Refer to the Temperature Compensation section for more information on Automatic and Manual Temperature Compensation. 8. After the measurement stabilizes, view the ph reading and temperature solution temperature on the display. 9. When the test is complete rinse the electrode and probe with distilled water. 10. Store electrodes and probes when not in use according to manufacturer s requirements. Measuring mv (ORP: Oxidation Reduction Potential) Important: Electrodes and temperature probes should always be rinsed before and between samples with distilled water to remove all traces of storage solution, process media, or previous test solution. 1. Connect the ORP electrode to the meter (BNC connection at bottom of meter). 2. Optionally connect the temperature probe (Extech model 850188 only) to the jack at the bottom of the meter. This temperature probe is used for automatic temperature compensation (ATC) purposes. Manual compensation is possible as explained below. 3. Turn the meter on using the POWER/ESC button. 4. Press and hold the FUNCTION button to select mv units if necessary. The unit of measure appears on the LCD (mv or ph). 5. To change the solution temperature units ( o C / o F) refer to the Temperature Units Selection section of this guide. 6. Rinse the ORP electrode and ATC probe (if being used) with distilled or deionized water. 7. Place the electrode and ATC probe (if being used) in the solution. Allow 30 seconds for the ORP electrode and ATC probe to reach thermal equilibrium with the buffer solution. Refer to the Temperature Compensation section for more information on Automatic and Manual Temperature Compensation. 8. After the measurement stabilizes, view the ORP reading in mv on the main display. 9. When the test is complete rinse the electrode and probe with distilled water. 10. Store electrodes and probes when not in use according to manufacturer s requirements. 4

Automatic (ATC) and Manual Temperature Compensation The PH210 automatically compensates (ATC) for solution temperature deviations when the Model 850188 temperature probe (inserted at bottom of meter) is placed in the measurement solution. The temperature of a solution is displayed on the meter s lower LCD line. To manually compensate for solution temperature deviations a separate thermometer must be placed in the solution to monitor the solution s temperature; then proceed as follows: 1. Press and hold the SETTING button for 5 seconds. The display will show noatc. 2. Press ENTER and noatc will flash (indicating the automatic compensation is switched OFF and Manual Temperature Compensation is active). Note that when noatc is not flashing Automatic Temperature Compensation mode is active. 3. Use the up and down buttons to manually adjust the temperature value to match the solution temperature (as determined by the solution thermometer). Once the temperature is programmed, press ENTER to save the data and then press ESC to return to the normal operating mode. 4. If the solution temperature changes, the manual temperature compensation procedure must be repeated. Temperature Units (C/F) Selection To change the temperature units ( o C / o F): 1. Power on the device 2. Press and hold the Setting button for 5 seconds, until the Setting mode is accessed 3. Press the Setting button again; either an F or a C will be displayed 4. Use the up arrow button to select the desired unit of measure for temperature solution 5. Press ENTER to save the setting 6. Press ESC to return to the normal mode of operation Automatic Power OFF Enable/Disable To conserve battery power, the instrument automatically shuts off after approximately 10 minutes of inactivity. To disable or enable this feature, follow the steps below: 1. Power on the device 2. Press and hold the SETTING button for 5 seconds until the Setting mode is accessed 3. Use the SETTING button to step to the screen that shows either OFF YES or OFF NO 4. Use the up arrow button to select YES (Auto Power OFF active) or NO (Auto Power OFF not active) 5. Press the ENTER key to save the setting and to return to the normal operating mode 5

MIN/MAX Data Recording (REC) mode This instrument has the ability record the highest (MAX) and the lowest (MIN) readings recorded. Readings are retained even after the instrument is switched OFF. 1. Momentarily press the REC button (REC will be displayed at the upper right). 2. Press REC again. REC/MAX will be displayed along with the maximum reading. 3. Press REC again to display the MIN reading. REC/MIN will display along with the minimum reading. 4. To delete either the MAX or the MIN reading, press the HOLD button while either the REC/MAX or the REC/MIN icons are displayed. 5. Press and hold REC for 2 seconds to exit and to return to the normal operating mode. The REC and MAX/MIN icons will switch OFF. Specifications Measurements/Parameters Range (Resolution) Accuracy Solution Temperature measurement 0 to 100 o C (32 to 212 o F) 0.8 o C (±1.5 o F) ph measurement 0.00 to 14.00pH (0.01pH) ± 0.04pH mv (ORP) measurement ± 1999mV ±0.5mV Note: Accuracy does not include ph buffer temperature coefficient Automatic Temperature Compensation (ATC) range Manual Temperature Compensation Adjustment range Power Display Operating Temperature and Relative Humidity ph Calibration points Auto Power OFF Low Battery Indication Dimensions / Weight 0 to 65 o C (32 to 149 o F) 0 to 100 o C (32 to 212 o F) 9 volt alkaline battery LCD display for simultaneous ph or mv and temperature readings plus programming menu parameters and icons 0 to 50 o C (32 to 122 o F); <80% RH 4.00, 7.00, 10.00 (1, 2, or 3 point calibration can be performed) 10 minutes from last key press (can be disabled) Battery display icon will flash when battery requires replacing 135 x 60 x 33mm (5.3 x 2.4 x 1.3 ); 200g (7.1 oz.) 6

Battery Replacement The 9V battery that powers the instrument requires replacing when the battery icon flashes on the display. To replace the battery follow the steps below: 1. Disconnect the electrode/probe from the instrument. 2. Open the rear battery compartment using a screwdriver to remove the two Phillips head screws that secure the compartment. 3. Carefully remove the old battery from the compartment lid chamber and gently unsnap the battery s terminals from the wired (red/black) connectors. 4. Install a new 9V battery by snapping the battery terminals onto the wired connectors (observing correct polarity) and inserting the battery into the compartment lid chamber. 5. Install the compartment lid chamber into the meter housing and secure with the two screws. Never dispose of used batteries or rechargeable batteries in household waste. As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold. Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end of life devices to a designated collection point for the disposal of electrical and electronic equipment. Electrode Maintenance and Troubleshooting Most measurement difficulties are caused by the electrode. Careful handling and proper storage of the electrode will increase measurement accuracy and electrode life. Electrodes should be rinsed often to remove all traces of storage solution, process medium, or previous test solution. Electrode Storage The electrode should be stored in a soaker bottle or wetting cap filled with ph4 buffer. Never store an electrode in distilled or deionized water as this will cause migration of the fill solution from the electrode. Electrode Cleaning The type of electrode coating that occurs over time will determine the cleaning method. Soft coatings can be removed by vigorous stirring or by using a squirt bottle. Organic chemical coating (hard coatings) should be chemically removed. Only in extreme cases should the bulb be cleaned mechanically. 7

Troubleshooting an Electrode Symptom Cause Recommendation Long response time or reading drift Clogged Junction Soak in 4.07 M Potassium Chloride (KCl) @ 60 o C for 30 minutes. Strong Alkaline Soak in 0.1 M HCL overnight. measurement Deteriorated Gel layer Replace electrode. Protein coating on electrode surface Soak in 1 gm Pepsin dissolved in 100ml of 0.1 m HCL for 30 minutes or as needed. Oil, paint, dyes, suspended solids etc. are on sensor Rinse electrode alternately with a materials solvent and a buffer 7.00. Organic solvents are coating the sensor Organic mole fraction has to be less than 50% to assure reasonable readings. Limit time of measurement. Keep probe in the 7.0 buffer between readings. Read solution for dry bulb below: Soak electrode tip in wetting cap filled with 1ml 7.00 buffer for 24 to 48 hours Dehydrated membrane Dry Bulb Long term storage without wetting Static Charge Wiping electrodes Rinse electrode in 7.0 buffer and blot. Do no wipe electrode. Same readings in Cracked or broken bulb Replace electrode. Use bulb guard. different buffers and Avoid plunging electrode to bottom of samples container and spinning bars. Wetting Erratic LCD display Samples have low ionic strength (lacks salt); e.g. distilled, de ionized, boiled, lake water (high pressure) cap will protect bulb between tests. For each 50 ml of sample add 1 drop (50uL) of SAT. KCl (No alteration in ph will occur using inert KCl). Copyright 2014 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form www.extech.com 8