Gatan 682 Precision Etching and Coating System (PECS) Simplified coating instructions

Similar documents
VCR Ion Beam Sputter Coater

Balzers Sputter Coater SCD 050

SALES SPECIFICATION. SC7640 Auto/Manual High Resolution Sputter Coater

Ion Beam Sputtering: Practical Applications to Electron Microscopy

DIAMOND Retractable Rodding Robot Model SPRAYROD-R

RI-215A Operator s Manual. Part Number: RK Revision 0 Released: 10/3/05

FLOWMETER INSTALLATION INSTRUCTIONS

DC LOAD TESTING Gian Trento June 18 th, 2004

BAL-TEC SCD 005. Scientist in charge: Lhoussaine Belkoura 07/14/06. softcomp soft matter composites 1

SET-UP AND INSTALLATION FOR LEAD SCREW CARTRIDGE ASSEMBLY

8 Channel Status Input Panel model SIP-8

Volkswagen Jetta, Golf, GTI 1999, 2000 Brake System 47 Brakes - Hydraulic Components (Page GR-47)

GLOLAB Two Wire Stepper Motor Positioner

Auto-belay Cable Replacement Process

Electron Beam and Sputter Deposition Choosing Process Parameters

1998½-2007 Dodge Ram 5.9L Cummins

Rain+Birdt. Simple To Set Timer (SST) Setup & Operation Instructions. English RAIN BIRD ( ) or visit

SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE

IMMS-CCC. IMMS-CCC Hardwire Central Interface. Installation Instructions

Troubleshooting Guide for Jacks Down LED Lights

BB-18 Black Body High Vacuum System Technical Description

GAERTNER SCIENTIFIC CORPORATION 3650 Jarvis Ave. Skokie, Illinois U.S.A. tel: fax: @gaertnerscientific.

Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P Thermodynamic cycle Heat pump

SIMPLE ELECTRONIC IGNITION WIRE DIAGRAM THERMOSTAT (3) PV PILOT BURNER/IGNITOR - SENSOR Q345A (9) SPARK (4) GND (BURNER) CHASSIS GND

OHIO MEDICAL CORPORATION. Medical Gas Panel. Installation and Maintenance Manual

Cathode Ray Tube. Introduction. Functional principle

Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2. Instruction Manual

TOYOTA TECHNICAL. An example of a tire pressure monitor valve sub-assembly. This mounts to a drop area in the wheel.

RL HW / RL HW+ / RL HGW / RL HV / RL HVPW/RL HVPW-G

1 DESCRIPTION OF THE APPLIANCE

G670 Intermittent Pilot Ignition Controls


INSTALLATION INSTRUCTIONS Air Spring Kit Ford F250/F-350 Single Wheel 2WD Ford F350 Dually 2WD IMPORTANT NOTES

OPERATING INSTRUCTIONS

* DISCLAIMER: Contents. How to Use This Guide: COMMERCIAL INSTALL GUIDE 2

Research RF Sputtering Packages

SPY-BATT Battery Tutor Device Installation Manual Rev /04/2016

ecomax Instructions for use Wall hung room sealed fan assisted condensing boilers For the user

Munchkin Heat Exchanger Maintenance

Your Simple Guide to Battery. Replacement. Customer Care:

WARNING CANADIAN D.O.C. WARNING

Reactive Sputtering Using a Dual-Anode Magnetron System

TOF FUNDAMENTALS TUTORIAL

AC-115 Compact Networked Single Door Controller. Installation and User Manual

Coating Technology: Evaporation Vs Sputtering

PPM Users Manual Signature Software

Installation Instructions S

Thermo Top - Troubleshooting Tree

Service Guide 12/27/03 TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models)

Quick Start Guide See Inside for Use and Safety Information

PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: AW 120 / AW 122 / AW 125 AWD 120 / AWD 121 / AWD 129

PUSH BUTTON START INSTALLATION MANUAL

Gas Oven Repair Guide

PD 100A. Printing data system

Operator Quick Guide EC SENSOR

INSTALLATION INSTRUCTIONS Air Spring Kit Ford F250 Single Wheel 4WD Ford F350 Dually 4WD (2011 F350 Single Wheel 4WD use p/n 6113)

Fuel Cell Characterization Lab


Example of SOP for IR Laser

Model 43AP Pneumatic Controller, Style B

Before installation it is important to know what parts you have and what the capabilities of these parts are.

Volkswagen Jetta, Golf, GTI 1999, 2000 Brake System 46 Brakes - Mechanical Components (Page GR-46)

Wireless 7 Day Programmable Room Thermostat

Manual for Fire Suppression & Methane Detection System

Easy Machining Center Setup

The purposes of this experiment are to test Faraday's Law qualitatively and to test Lenz's Law.

Troubleshooting Salt Generators

LCD Readout Digital Pressure Switch

Oil - kerosene burners

Quick Reference Guide VT55 & OBDII. Customer Service TPMS Version: QSG /12

Handheld USB Digital Endoscope/Microscope

Owner s Guide Guide du propriétaire Guía para el usuario TH115-AF-GB-10. Programmable thermostat Thermostat programmable Termostato programable

Models SFP33 & SFP32 Pressure Transmitters

INSTALLATION AND OPERATING INSTRUCTIONS For Model GL1 Gate Locks

Baxi Platinum Combi HE Range. User s Operating Instructions & Important Warranty Information. Gas Fired Wall Mounted Condensing Combination Boiler

SEMTech Solutions. Leaders in Refurbished SEMs. SEMTech Solutions Windows 7 SOFTWARE CONTROL SYSTEM

Magneto Timing. The selected wire(s) from the magneto(s) distributor must be connected to this cylinder. And the crankshaft/magneto must be spinning.

MTX-D, Oil Temperature/Pressure Gauge

Scan2CRM for ACT! User Guide

Introducing Your Cisco IP Phone

Thank you for purchasing the Huawei E5332 Mobile WiFi. This quick start guide will help you to get started right away.

INSTRUCTION MANUAL CIRRO MK 3/TYPE 2 MIST GENERATOR

LED LEONARDO 120 W. LED Fresnel SPOTLIGHT White light, either Tungsten or Daylight balanced Correlated Color Temperature

OxySense Gen III Operations Manual

INSTALLATION AND SERVICE MANUAL

PC1130 Electric Air Compressor

WHAT YOU DON T KNOW ABOUT ACCUMULATORS CAN KILL YOU!

omega.com ΩOMEGA RS232 Multi-Drop FMA1600-MDB Multi-Drop Box

QUANTIFYING SOLAR ENERGY

Designed for multi-injector endurance testing, the

Rigaku XRD-System Instruction Manual v4/19/03. The Krishnan Group/Wilcox 132, University of Washington

Hygro-Thermometer + InfraRed Thermometer Model RH101

EVANS ELECTRONIC TEMPERATURE CONTROL TROUBLESHOOTING GUIDE for systems equipped with electric coolant valve and external PC board.

FG MOISTURE MONITOR Installation & Operation Manual

R&S ZVA-Z75, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500 Converters Quick Start Guide

Product Description Full Voltage Starting Electric Fire Pump Controllers FTA1000

HAZUBA. HAZUBA AG Rainstrasse Zuchwil T F info@hazuba.ch 06/14

Impact Uni-Vent 754 Portable Ventilator

Quik Lining Systems Heater Owners Guide

Tube Control Measurement, Sorting Modular System for Glass Tube

Transcription:

Gatan 682 Precision Etching and Coating System (PECS) Simplified coating instructions The PECS has the ability to sputter coat material on to a surface. This brief guide will cover using argon ions to sputter coat a thin film on to a sample. Information about etching and more detailed sputtering information can be found in the user manual. 0. Start the use log on the PC. SETUP AND PURGING THE GUNS: Before the PECS can be used, it must be properly setup. 1. Open the valve on the wall (Located in the left back corner) to start the argon flow (see Figure 1). One of the tools has a small leak and the tank will be depleted quickly if not closed after each use. 2. The default configuration of the tool is Figure 1. Argon valve in closed position

A) To have both targets extracted (copper rods visible) Left Target Right Target Figure 2a. Left and right target both in default extracted position B) The sample holder is in the IN position, and the rotation cable disconnected Sample Holder Rotation Cable Figure 2b. Sample holder is IN Rotational Cable is disconnected C) The LCD display will show the fore line pressure of less than 3 Torr and the molecular diffusion pump MDR should read 100%. Figure 2c. Foreline is less than 3 Torr, MDP RPM is 100% D) The analog gauge on the right will be around 10-6 Torr. Confirm that this is how

the tool is set. Figure 2d. Cathode Gauge in default position 3. Purge the guns. There are two toggle switches that say left gun and right gun (see Figure 3a). that when enabled have an indicator light. Note that either the sputtering guns can be enabled or the etch gun, but not both. Purge for 10 minutes. Figure 3a. Turn on left gun and right gun toggle switches Figure 3b. Set timer for 10 minutes and push start to purge the guns

Lights show guns are enabled Figure 3c. Lights will indicate if guns are enabled OPTIMIZING THE GUNS ION CURRENT: VERIFY WITH STAFF (GENERALLY THIS STEP CAN BE SKIPPED IF THE ION CURRENT HAS BEEN OPTIMIZED RECENTLY) 4. Adjust the timer to 5 minutes (5:00) and adjust the beam energy (8kv is a good starting value). Holding the stop button for about 5 seconds resets the timer. Figure 4a. Beam Energy Enable gas flow with the toggle switches. Pressing the start button engages the high voltage and starts the timer counting down. The guns will heat slightly over about 4 minutes, allow the ion currents to stabilize before continuing. Enable gas flow with the toggle switches Figure 4b. Adjust the flow for each gun individually

Observe the ion current for each gun individually Figure 4c. Slightly adjust the flow rate (1/8 turn) and observe the ion current. Each gun is adjusted independently, although the total pressure will affect both guns. When the gun is working, the light on the outside of the gun will illuminate. The current is more stable in this region Figure 4d. As the flow is adjusted the ion current will peak quickly and start to fall. The current will rise with rising pressure, saturate, and then fall slightly. The current is more stable in the pressure region greater than the maximum value. Maximum ion current happens at pressures less than 2x10-5 torr. Turn off the high voltage by pressing the stop button. INSERTING AND SPUTTERING THE SAMPLE: After the PECS has been setup and purged 5. On the left side of the chamber is a large knob (50 mm diameter) and also a black toggle switch on the top of the chamber. The large knob is a shield that can cover the sample and is also a current meter for the etch gun. Shielding the sample allows adjustment of the sputter guns without contaminating the sample. The toggle switch

opens a pneumatically driven shutter for the view port and also turns on a sample illumination light. The toggle switch on the top right opens and closes the shutter only. THE SHIELD MUST BE OPEN WHEN SPUTTERING YOUR SAMPLE Figure 5. Shield knob 6. On the front panel in the air lock control section, move the air lock toggle to the out position. This is driven pneumatically by the argon and takes about 5 seconds. Figure 6. Close-up of Airlock Control 7. Briefly press the vent button. The venting pressure may be enough to make the holder jump out of the guide, be prepared to catch it.

Figure 7a. Holder in normal IN position Figure 7b. Holder in Out Position 8. Remove the holder and place in the cradle. The holder has 4 holes which a SEM stub can be placed. The center hole works better for a single sample on a 12.5mm diameter stub (1 2 inch) and the outer holes can fit several 12.5mm stubs at the same time. Figure 8. Holder in the Cradle 9. Insert the holder completely into the guide and then press and hold vac button. The fore line pressure gauge will briefly spike and then return to normal in about 10 seconds. Once the fore line pressure has returned to less than 3 Torr, move the toggle switch in the airlock control to IN and guide the holder to the inward position noting that the guide pin needs to be seated properly.

Press vac button and wait for Foreline pressure to return to < 3 Torr Figure 9a. Close-up of Airlock Control Figure 9b. Align the guide pin so it seats properly 10. Insert target and move the thickness monitor into place. There are two targets mounted on each stick. Insert copper rod and then rotate knob to select desired target, there is a text label on the knob. The targets on the left rod are Au/Pd, and the second target is platinum. The right rod has carbon and chromium. Note if using Cr alert the lab staff to aide in additional setup. Figure 10a. The Au/Pd Target is inserted

Thickness monitor is attached via BNC cable Figure 10b. Thickness monitor is connected and set to Sensor 11. Prepare the thickness monitor. Rotate the thickness sensor/ shield knob to sensor. Attach BNC cable to external thickness monitor and turn it on. The display will show fail, this is normal. Press stop button to clear. Select film parameters on thickness monitor. Film 1 has been set to Au/Pd, impedance 23.50 density 14.81 ;film 2 Pt impedance 36.04 density 21.40; film 3 Carbon with impedance 40.14, density 3.52; and film 4 Cr Impedance 28.95, 7.20. Tooling factor of 180 for all four films. Press stop to clear Select Film Parameters to specify which target is being used Figure 11. Thickness Monitor after turning it on 12. Configure rock and rotate. Rotate spins the sample in the plane of the holder, rock moves it perpendicular to that plane. Rock is useful for rough samples to enable coating the sides. These can both be used to create a more even coating. A rotation rate 20 RPM is a good start and the toggle should be on. Viewing the sample through the view port can see how these are used.

Figure 12. Rock/Rotate control 13. Start sputtering. Set timer for total deposition time, enable gas flow, enable high voltage by pressing start. The thickness monitor should indicate approximate thickness and should get larger with time. For a current of 200 ua from each gun, typical sputtering rates for Au/Pd is 1.1 A/s, C is 0.1 A/s, Pt is 0.7 A/s. SHUTDOWN AND SAMPLE REMOVAL: 14. When desired thickness is reached, Turn off high voltage by pushing stop on the timer. Turn off gas flow toggle switches. Stop rotation if enabled. Extract target, rotate knob with target and copper rod then pull. 15. Remove sample. Take off rotation cable Set airlock control position toggle to out Press vent Remove sample Insert holder into tube Press vac When fore line pressure is stable move toggle to in. 16. Turn off argon valve at wall. 17. Stop the log PC timer.