Permanent Pressure Loss Comparison Among Various Flowmeter Technologies White Paper

Similar documents
Custody Transfer Measurement. with the V-Cone Flowmeter

FLOW MEASUREMENT 2001 INTERNATIONAL CONFERENCE DERIVATION OF AN EXPANSIBILITY FACTOR FOR THE V-CONE METER

V-Cone, An Alternative to Orifice Meter in Wet Gas Applications

Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF

Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology. HWEA 2011 Conference Craig S. Johnson

Using meters to measure steam flow

Grant Agreement No SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

The Unique Accelabar Flow Meter

measurement, but almost any pipe elbow can be calibrated Elbow meters are not as potentially accurate as venturi,

In semiconductor applications, the required mass flows

Energy and Flow Measurement for Hydronic Systems

Flow Measurement in Pipes and Ducts. Flow Measurement in Pipes and Ducts, Course #503. Presented by:

FLUID FLOW Introduction General Description

RITTER Multiple Sonic Nozzle Calibration System

4.What is the appropriate dimensionless parameter to use in comparing flow types? YOUR ANSWER: The Reynolds Number, Re.

Testing Protocol for Differential Pressure Measurement Devices API MPMS Chapter 22.2

Thermal Dispersion Mass Flow

Gas Custody Transfer Calibration

ABB Flowmeters Review of industrial processing flowmeters

Chapter 10. Flow Rate. Flow Rate. Flow Measurements. The velocity of the flow is described at any

Theory overview of flow measurement using differential pressure devices based on ISO-5167 standard.

Advances in Thermal Dispersion Mass Flow Meter Accuracy

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

INDIAN STANDARDS FOR NATURAL GAS PIPELINE SYSTEM

Flow Measurement Options for Pipeline and Open Channel Flow

V-CONE METERING PROGRAM User Manual

FLOW COMPUTERS METERING, MONITORING, & DATA ACQUISITIONS. Sami Halilah Dynamic Flow Computers Houston, Texas

Gas Flow Measurement Software A Report for

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

Energy Systems Engineering Technology

THEORETICAL UNCERTAINTY OF ORIFICE FLOW MEASUREMENT page 1 THEORETICAL UNCERTAINTY OF ORIFICE FLOW MEASUREMENT INTRODUCTION COEFFICIENT OF DISCHARGE

A basic introduction to steam

THE USE OF A CORIOLIS FLOW METER FOR MEASURING MOLASSES PRODUCTION IN A SUGAR MILL. ABSTRACT

For Multi-Parameter Meters see mvx

Laminar Flow Elements

TANKJKT. Heat Transfer Calculations for Jacketed Tanks SCREEN SHOTS. Copyright By chemengsoftware.com

Flow Measurement. Reference - Flow Measurement FUNDAMENTALS. ρ = fluid density (lb/ft3 or kg/m3)

Flow Sensors. - mass flow rate - volume flow rate - velocity. - stream line parabolic velocity profile - turbulent vortices. Methods of measurement

Correcting a Coriolis Meter for Two Phase Oil & Gas Flow

ME 538: Experiment # 1 Flow measurement and particle generation instruments

Heat Transfer Prof. Dr. Ale Kumar Ghosal Department of Chemical Engineering Indian Institute of Technology, Guwahati

T U R B I N E G A S M E T E R

Chapter 3.5: Fans and Blowers

Rosemount Orifice Plate in Wet Gas

HEAVY OIL FLOW MEASUREMENT CHALLENGES

Experiment # 3: Pipe Flow

F ox W hi t e Paper. Reduce Energy Costs and Enhance Emissions Monitoring Systems

DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING

OPTIMASS 1400 Technical Datasheet

Experiment 3 Pipe Friction

Pipe Flow-Friction Factor Calculations with Excel

Entrained Gas Diagnostic with Intelligent Differential Pressure Transmitter

VAD. Variable Area Desuperheaters

TABLE OF CONTENT

OPTIMASS 6000 Technical Datasheet

APPLICATION OF DENSITOMETERS TO LIQUID MEASUREMENT Class # 2010

FLUID MECHANICS. TUTORIAL No.7 FLUID FORCES. When you have completed this tutorial you should be able to. Solve forces due to pressure difference.

COMBIMASS. Technical Data COMBIMASS eco-bio +

Take total control of your HVAC system with cost-saving flow and thermal energy meters

THEORY AND APPLICATION OF KURZ THERMAL CONVECTION MASS FLOW METERS INSTRUMENTS

Design and Analysis of Various Bluff Objects for Vortex flow meters

19th International North Sea Flow Measurement Workshop October 2001 Kristiansand, Norway

MODEL TEG-F20. Specification Sheet. TEG-F20 Specification Sheet Ver Copyright(C) 2011 Tsukasa Sokken Co., Ltd.

Daniel Venturi Tubes and Flow Nozzles Brochure

Industry faces unrelenting pressure to increase production

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers

AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS

Application of the Orifice Meter for Accurate Gas Flow Measurement page 1. Application of the Orifice Meter for Accurate Gas Flow Measurement.

Minor losses include head losses through/past hydrants, couplers, valves,

Flow Sensors. Convenient, Reliable and Economical Flow Measurement. THORNTON Leading Pure Water Analytics

Experimental Evaluation of the Discharge Coefficient of a Centre-Pivot Roof Window

191: Calibration and Standards in Flow Measurement. Richard Paton National Engineering Laboratory, Scotland, UK 1 GENERAL PRINCIPLES

Fluid Flow Instrumentation

A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW ASME Fluids Engineering Division Summer Meeting

Thermal Mass Flow Meter (Series MST)

INTRODUCTION SOME USES OF SPRAY NOZZLES INTRODUCTION TYPES OF SPRAY NOZZLES

Chapter 8: Flow in Pipes

Measurement & Analytics. VIS Multiphase Flow Meter High gas content applications solved with no radioactive source - Measurement made easy

Gas Well Deliverability Testing

Implementing New Technology at TransCanada Meter Stations -- Better, Faster, Cheaper

Theory, Application, and Sizing of Air Valves

CFD Analysis Of Multi-Phase Flow And Its Measurements

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

Principals of Condensing Boiler System Design

UPM Figure 1 MPFM installed on well pad

Model 43AP Pneumatic Controller, Style B

FLOW METER SPECIFICATION AND APPLICATION ANALYSIS Class #41

Technical & Sizing Bulletin for Meriam LFEs. Laminar Flow Elements

Air Flow Measurements

VAD Variable Area Desuperheaters

Physics 1114: Unit 6 Homework: Answers

SIZING AND CAPACITIES OF GAS PIPING

Redefining Spray Technology

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

Overview of the state of the art heat measurement technologies for larger solar thermal systems

C. starting positive displacement pumps with the discharge valve closed.

Smart Electromagnetic Flowmeter Open channel Flowmeter Detector

Transcription:

Among Various Flowmeter Technologies White Paper Page 1 of 9

White Paper By Stephen A. Ifft Mechanical Engineer McCrometer, Inc. Among Various Flowmeter Technologies Abstract This paper explores the issue of permanent pressure loss through various types of flow metering technologies. Of particular interest is the. The flowmeter is a differential pressure device produced by McCrometer, Hemet, California, USA. By design, the measures differential pressure created by a cone positioned in the center of the pipe. The open area through which the fluid passes is stretched over the edge of the cone. This annular space appears to constrict flow dramatically compared to other metering technologies. This paper compares the permanent pressure loss among meters in a typical application. Introduction The is a patented flowmeter made by McCrometer. With its unique geometry, the creates a differential pressure signal that is proportional to the square of the flowrate. This differential pressure signal is then measured by a secondary device that converts the analog pressure signal to an electronic signal, usually a standard 4-20 ma current signal. The electronic signal can then be used to calculate flowrate, mass rate, etc. using tertiary devices. One typical question regarding the is the amount of permanent pressure loss to be expected through the meter in the application. On first look, the cone appears to occupy an inordinate amount of space in the pipeline, and thus constrict the flow greatly. In reality, the cone restricts the flow very little. Several factors contribute to allowing the to measure flowrate very accurately without significant permanent pressure loss. One method to compare differential pressure meters is by the beta ratio. The beta ratio is the ratio of the smallest open area to the largest open area in the meter. In traditional differential pressure meters, this is usually the ratio of the diameter of the constriction to the diameter of the pipe. This is identical to the ratio of the areas mentioned earlier. s also have beta ratios that share the same area definition. Since the does not have a central opening, the beta ratio is calculated in a slightly different manner. See. figure 1. Page 2 of 9

If a meter has a beta ratio of, for example, 0.65, this means that 42% of the pipe area is still open at the largest constriction (0.65 x 0.65 = 0.42). This applies to all differential pressure flowmeters, including the. A typical method to measure the permanent pressure loss in a differential pressure flowmeter is to state a percentage of the differential pressure created at a given flowrate. For example, a differential pressure meter may state a 50% permanent pressure loss of differential pressure created. If the maximum differential pressure created by the meter is 10 kpa, the permanent pressure loss at that flowrate is 5 kpa. If the flow decreased and the differential pressure becomes smaller, the permanent pressure loss will become smaller. The percentage of differential pressure permanently lost can be graphed as a relationship of beta ratio. See figure 2. The is positioned between the orifice and the nozzle curves. This, however, is not a true comparison between the permanent pressure loss expected through the meters because the requires less differential pressure to remain accurate and repeatable. This is possible because of the stable and controlled differential pressure signal created by the cone. A true comparison would be to compare the actual permanent pressure loss expected through various meters for a given application. Application Comparison An application was chosen to compare the permanent pressure loss through the following meters: 1. based on information from McCrometer 2. Plate 3. Flow 4. Venturi with 15 o exit angle 5. Venturi with 7 o exit angle 6. Turbine 7. Vortex based on information from Miller 1 8. Coriolis 9. Vortex based on information from Foxboro Co. 10. Coriolis based on information from MicroMotion Co. Page 3 of 9

The application parameters were as follows: Application Information: Nominal Line Size: 3 inch Fluid: Water Temperature: 15 o C Pressure: 5 barg Density: 999.387 kg/m 3 Viscosity: 1.059 cp Flowrate: 1145 LPM Meter Sizing Each differential pressure meter was sized according to standard specifications from McCrometer, Miller 1 and ISO 5167 2. The following chart shows specifics for each differential pressure meter: Venturi Beta ratio: 0.85 0.75 0.7 0.699 Max. DP (kpa): 12.3 45.8 24.6 24.6 Cf (Cd): 0.754 0.6 0.99 0.995 The has a recommended differential pressure (12.3 kpa) that is half what is recommended for all other differential pressure meters (24.6 kpa). The orifice was sized with the largest suggested beta ratio to keep the differential pressure as low as possible. At the largest beta ratio, the differential pressure created exceeded the recommended by almost twice. All other meters were selected according to line size alone. Standard meter specifications were used for the turbine and vortex. Other specific meters were selected based on manufacturer s recommendations and the following model numbers were used: MicroMotion Coriolis I/A Series E83 CMF300M In all cases the line size remained 3 diameter nominal. Results By comparing permanent pressure loss for a given application, an accurate comparison can be drawn between metering technologies. See figure 3 for a graphical representation of permanent pressure loss through these meters over a 10:1 flow turndown. See figure 4 for a graph of maximum loss through each meter at maximum flowrate. MicroMotion data appears only in figure 4 due to lack of turndown information from the manufacturer. Page 4 of 9

To summarize, the meters ranked in the following order according to permanent pressure loss: 1. Venturi with 7 o exit angle 2. 3. Venturi with 15 o exit angle 4. Flow 5. Vortex 6. Turbine 7. 8. Plate 9. MicroMotion Coriolis 10. Further Results A similar comparison was completed on a 2 line size to compare the effect of line size under generally similar conditions. The ranking changed slightly: 1. Venturi with 7 o exit angle 2. 3. Venturi with 15 o exit angle 4. Flow 5. Vortex 6. Turbine 7. 8. Plate 9. 10. MicroMotion Coriolis See figures 5 and 6 for graphical comparisons. Page 5 of 9

Figure 1 Permanent Pressure Loss Differential Pressure Meters 1 0.8 % of DP 0.6 0.4 0.2 0 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 Beta Ratio Venturi 15 Venturi 7 Data from R. W. Miller "Flow Measurement Engineering Handbook" Second Edition, McGraw-Hill, 1989. Figure 2 Page 6 of 9

3" 1145 LPM Permanent Pressure Loss (kpa) 10.0 1.0 0.1 Turbine Vortex Venturi 15 o Venturi 7 o 100 1000 Flowrate (LPM) Data from R.W. Miller, Flow Measurement Engineering Handbook, Second Edition, McGraw-Hill, 1989, unless otherwise noted by company name. Figure 3 35 Maximum Loss: 3" 1145 LPM Permanent Pressure Loss (kpa) 30 25 20 15 10 5 Venturi 15 Venturi 7 Turbine Vortex MicroMotion Coriolis 0 Meter Types Data from R. W. Miller "Flow Measurement Engineering Handbook" Second Edition, McGraw-Hill, 1989, unless otherwise noted by company name. Figure 4 Page 7 of 9

Permanent Pressure Loss (kpa) 100.00 10.00 1.00 0.10 2" 527 LPM Turbine Vortex Venturi 15 o Venturi 7 o 0.01 100 Flowrate (LPM) Data from R.W. Miller, Flow Measurement Engineering Handbook, Second Edition, McGraw-Hill, 1989, unless otherwise noted by company name. Figure 5 Maximum Loss: 2" 527 LPM 50 MicroMotion Coriolis Permanent Pressure Loss (kpa) 40 30 20 10 Turbine Vortex Venturi 15 Venturi 7 0 Meter Types Data from R. W. Miller "Flow Measurement Engineering Handbook" Second Edition, McGraw-Hill, 1989, unless otherwise noted by company name. Figure 6 Page 8 of 9

References 1. Miller, R. W., Flow Measurement Engineering Handbook, Second Edition, McGraw-Hill, 1989. 2. International Standard ISO 5167, Measurement of fluid flow by means of orifice plates, nozzles and venturi tubes inserted in circular cross-section conduits running full., Ref. No. ISO-5167-1980 (E). Page 9 of 9