Practical LC-MS Troubleshooting and Maintenance: A Primer

Similar documents
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Electrospray Ion Trap Mass Spectrometry. Introduction

Appendix 5 Overview of requirements in English

Chapter 28: High-Performance Liquid Chromatography (HPLC)

Micromass LCT User s Guide

AMD Analysis & Technology AG

Successful HPLC Operation. A Troubleshooting Guide Version 1.1

EQuan MAX Getting Connected Field Service Guide

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

High Sensitivity MS Determination of Carryover in a New Autosampler Design

Rheodyne Manual Sample Injector Valves

QUANTITATIVE AMINO ACID ANALYSIS. Aurélie Lolia Applications Manager, Biochrom Ltd

SCION SQ Series GC-MS

ChromQuest 5.0 Chromatography Data System

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

Application Note # LCMS-81 Introducing New Proteomics Acquisiton Strategies with the compact Towards the Universal Proteomics Acquisition Method

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

How To Test For Contamination In Large Volume Water

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

Optimizing the Post-implementation Monitoring of a LC-MS/MS Method

Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# RFQ)

Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.

Analysis of Liquid Samples on the Agilent GC-MS

Background Information

23 The Thermal Conductivity Detector

ÄKTA system Training Guide

LC-MS/MS for Chromatographers

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

Chromatograph Maintenance and Troubleshooting Class # 5050

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Trouble Shooting. Pump

Amino Acid Analyzer L-8900

The Theory of HPLC. Gradient HPLC

Tuning & Mass Calibration

HPLC Basics, Equipment, and Troubleshooting

amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS

Overview. Purpose. Methods. Results

Analyst 1.6 Software. Software Reference Guide

Calibrating Differential Pressure Sensors

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

Flushing and Cleaning the A/C System

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Fundamental Principles of Rotary Displacement Meters

After the reactor has been running and the air in the chamber has been purged the co2 can be introduced and the dial in process can begin.

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Analyst 1.6 Software. Getting Started Guide

INSTRUCTIONS Edition AC

Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)

IFC. A new perspective i l ifi ti. p p g p. Single softw FractPAL software plug-in for Agilent ChemStation. Medicinal Chemistry.

Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software

Diesel: Troubleshooting

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Class 5 to 7 Truck and Bus Hydraulic Brake System

EKSIGENT EKSPERT NANOLC 400. Unmatched flexibility for low flow LC/MS

Hardware Manual. AB SCIEX API 4000 LC/MS/MS System

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

[ Care and Use Manual ]

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

TROUBLESHOOTING FLOW CHART

Diesel Fuel Systems. Injection Nozzles

How To Build An Amd Quas 3 Ar Mass Spectrometer

Tutorial 9: SWATH data analysis in Skyline

A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water

HVAC Preventative Maintenance Basics. Presented by Mike Crawley

GC-xt. The most flexible GC /GC-MS liquid sample injector

Waters 2695 Separations Module

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

CDS TROUBLESHOOTING SECTION I. VACUUM Weak vacuum at wand. Gauge reads normal (10hg to 14hg)

Split / Splitless Injection for Capillary GC

Oregon Fuel Injection

Residual Gas Analyzers

UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research

Pesticide Analysis by Mass Spectrometry

ORGANIC SAMPLE PREPARATION

TOF FUNDAMENTALS TUTORIAL

OEM-EP Pressure Controllers

USER S, MAINTENANCE and SERVICE INFORMATION MANUAL

GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Monoclonal Antibody Fragment Separation and Characterization Using Size Exclusion Chromatography Coupled with Mass Spectrometry

A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011

How To Use Gc-Ms

Expectations for GC-MS Lab

Elements Elements describe the essential outcomes. 1. Prepare to diagnose and repair air conditioning and HVAC system

EnAlgae SOP: Biogas analysis

Transcription:

Practical LC-MS Troubleshooting and Maintenance: A Primer Sue Abbatiello, The Broad Institute Tom Blau, Ion Technology Support, Inc. Will Thompson, Duke University

General Troubleshooting Guidelines Visual Inspection is Key! Start By Breaking the System Down Having a BASELINE for all metrics is critical to proper troubleshooting -Logbooks to track problems and solutions -Reference Values for Instrument Readbacks -System Suitability

Break the System Down Review All Available Data Chromatogram, Mass Spectra, Pressure Traces, Error Logs Isolate Major System Components Mass Spectrometer (Infusion Analysis) LC (Standard Mixture) LC System Critical Tool: Standard Mixture Separation! Critical Readbacks: Pressure Traces Flow measurement Injection Check Mass Spectrometer Critical Tool : Infusion! Critical Readbacks: Voltages Vacuum Gauges Detector Signal Isolate Problem To: Column(s) Autosampler Valves Tubing Injector Pump Isolate Problem To: Ionization/Source Vacuum Calibration Mass Analyzer Detector

What to Learn From Your Engineer The time to gather key information to enable troubleshooting is at the time your instrument is installed Compare to Installation Values (CIV) Vacuum Settings (all regions) Voltage Readbacks (screenshots of good values) Source Boards Power Supplies Tune Mixes to Use (and good values) Sensitivity, Resolution, Stability Mass Calibration, S/N AutoTune/Calibration Best Practices Chromatographic Performance Pressure Range for X Column Peak Width/Resolution of critical pr RT Stability Error Log Message Descriptions PM Schedule for Your Use-Case How Best to Find Spare Parts Restrictions on Solvents, ph, etc with your LC Also, your engineer probably has troubleshooting documents that are not widely distributed. Ask for them! (It can t hurt)

What is System Suitability? "System suitability testing is an integral part of many analytical procedures. The tests are based on the concept that the equipment, electronics, analytical operations and samples to be analyzed constitute an integral system that can be evaluated as such. System suitability test parameters to be established for a particular procedure depend on the type of procedure being validated". (FDA) The checking of a system, before or during analysis of unknowns, to ensure system performance (International Conference on Harmonization of Technical Requirements for Registration of Pharmaceutical for Human Use [ICH]) GO No GO Simply put: Analysis of a known sample to assess system performance Helps to identify when the system is not working Define Pass/Fail Criteria

What are metrics of a system suitability protocol? MS LC MS response, sensitivity Mass accuracy Precision Retention time Peak shape, FWHM Chromatographic resolution

How can you assess normal system performance? Tune/calibrate LC and MS components Monitor variability of metrics over several injections Always use the same method for LC and MS, even if it is different than your sample method Ask the vendor for performance specifications for both LC and MS parameters Note when system performance deviates Compare to other systems running the same system suitability protocol

How can you evaluate system suitability data (for free)? RawMeat.exe*: vastscientific.com/rawmeat/ NIST metrics: peptide.nist.gov/metrics/ Skyline: proteome.gs.washington.edu/software/skyline Panorama: www.panoramaweb.org Retention Time Viewer: gibsonproteomics.org/resources/rt-viewer Database searches for peptide ID Excel Any longitudinal data tracking system Check with your instrument vendor * Thermo instruments only

Summary Set up a system suitability protocol that works for you and use it Don t waste precious sample without making sure your LC-MS system is working Use system suitability to monitor changes in hardware, software, any changes at all Keep examples of poor system suitability data to help trouble shoot future issues

General LC Troubleshooting Tips, Tricks & Maintenance

Autosampler Is there air in the syringe? N Is sample being picked up from vial? Y Does the pressure trace during injection look normal? Y Is the flow rate through the autosampler normal? Y Is the chiller on? Y N N N N Check volume of needle wash solvent and sample. Make sure syringe is attached correctly. Check syringe seal and plunger. Check the AS valve and rotor Drop in pressure indicates leak, air, or solvent with lower viscosity in system. Pressure spike indicates blockage or solvent of higher viscosity in system Problem may be with HPLC and flow rate. Check flow rate directly from pump. Check flow rate after autosampler. Loss of flow may indicate leak in AS. Check temperature with thermometer. Check temp settings.

HPLC Pumps Is there solvent in the reservoirs? Y Does the pressure trace look normal? Y Is the flow rate normal? Y Are peaks eluting with reproducible retention times Y Are peak shapes symmetrical? N N N N N Replace solvents and purge. Degas solvents if necessary. See next slides for examples Check flow rate directly from pumps, then through autosampler, look for leaks May have air in the system or degasser not working. Check solvent mixer, pump seals or check valves. Check all connections post-column for voids. Go to Columns & Hardware section

Columns & Hardware Are peak shapes symmetrical? Y Are peaks the usual width? Y Is ESI stable? Y Is the system clogging frequently? N N N Y Check all connections post-column. Assess column age. Run standard sample. Check gradient delivery and flow rate from HPLC. Check high voltage connection. Is ESI tip clean? Is ESI tip leaking? Check tubing cuts (fused silica, PEEKsil, PEEK or stainless steel). Try in-line filter. Check rotors for wear. Centrifuge samples prior to analysis.

MS System: A Troubleshooting Workflow and Common Problems

MS System: A Troubleshooting Workflow Infusion (ESI +/-) (e.g. Vendor-recommended tune Mix. NaCsI, PPG, PolyTyrosine, GluFib, LeuEnk, etc) 1. Sensitivity Check (compare to installation values, CIV) Key System Checks Most Common Causes for Failure Fail Tune Source Parameters ( Fresh /Recovered Tune Page) ESI Emitter Position/Leaks/Stability Source and Cone/Transfer Tube and Skimmer Cleanliness Detector Gain or Voltage Fail ESI Sprayer Stability (Arcing?) Transfer Optics (Entrance Hex/Quad, Q1) Charging Poor Vacuum Pass 2. Signal Stability Check (Signal should not show significant decrease in ~minutes) Fail Pass 3. Mass Resolution Check (CIV) Poor Analyzer Vacuum MS Analyzer Tune Parameters (Recovered Tune Page) Pass 4. MS/MS Fragmentation Check (CIV) Poor Vacuum Syringe Pump High V Power Supply Fresh Tune Solution! Pass Pass Pass Go to #2 (Stability) Go to #3 (Resolution) Go to #4 (Fragmentation) Continued next page

MS Troubleshooting Workflow, continued Fail 4. MS/MS Fragmentation Check (CIV) Pass Key System Checks Most Common Causes for Failure No Collision Gas 5. Calibration Improper Collision Energy MS1 Transmission Window or Calibration Collision Cell Power Supply Fail Nominal Mass Calibration off Corrupt Tune Page Lab/Instrument Temperature Fail Pass 6. System Suitability Troubleshoot LC OR Coupling from LC to MS Pass READY TO RUN! Pass Go to #5 (Calibration) Pass Go to #6 (System Suitability) Remember VISUAL INSPECTION is key at all steps! Can you think of other, less common failures for these system checks?

MS System: A Brief Word On Communications Infusion (ESI +/-) (e.g. Vendor-recommended tune Mix. NaCsI, PPG, PolyTyrosine, GluFib, LeuEnk, etc) But what if I don t have PC to Instrument Communication? Troubleshooting workflow assumes basic PC-to-Embedded PC Communication is Intact Methods to Check for Communications 1. Basic Readbacks (should update regularly and have slight changes) 1. Turbo Speed 2. Vacuum Readings 3. Power Supplies 2. Toggle Controls (watch for response) 1. ESI Voltage on/off 2. Other source parameters 3. Collision Energy low/high if infusing What to Do if No Response 1. If in doubt, reboot! (PC and Instrument embedded PC) 2. Check cable connections 3. Communication (ethernet) cards 4. Hard reboot of instrument electronics

Troubleshooting Typial Causes of Failure (Sensitivity) Tune Source Parameters ( Fresh /Recovered Tune Page) ESI Emitter Position/Leaks/Stability Source and Cone/Transfer Tube and Skimmer Cleanliness Detector Gain or Voltage Ion Transmission is drastically effected by source tuning parameters. Check values against installation values. If same, consider starting a fresh tune page since tune files can become corrupt. Emitter position and spray stability is the both the easiest way to have a problem and to fix one. If possible, record the micrometer settings on your ESI position to make sure it is in the right place. If spray is sputtering, adjust voltage, gases. Consider cleaning or replacing sprayer components. Check for leaks. VISUAL inspection is key. Dirty skimmer and transfer tubes can kill sensitivity. Clean stainless steel components with sonication in 50/50 MeOH/water with 1% Formic acid (gentle) or 30% nitric acid (aggressive). Wash thoroughly. Voltage or gain effects all types of detectors, from EMT on QqQ, to MCP/ADC on Tof, to Electron Multiplier for automatic gain control (ion gating) on trapping instruments. Run manual or automatic gain checks. Poor Vacuum Fresh Tune Solution Mean free path dictates that MS systems need vacuum to operate, and incomplete vacuum will kill ion transmission. Check the vacuum gauges against the installation target values to be sure vacuum levels are appropriate for your mass analyzer. Probably the easiest way to fail a sensitivity check is to have an old tune solution, have the wrong organic composition, or have the ph wrong (add the acid!). Make it fresh!

Troubleshooting Typical Causes of Failure (Signal Instability) ESI Sprayer Stability Transfer Optics (Entrance Hex/Quad/Q1 Charging) Emitter position and spray stability is the both the easiest way to have a problem and to fix one. If possible, record the micrometer settings on your ESI position to make sure it is in the right place. If spray is sputtering, adjust voltage, gases. Consider cleaning or replacing sprayer components. Check for leaks. Check for arcing due to incorrect voltage or erosion of sprayer. Previous samples, mobile phase additives, and even vacuum oil mist can sometimes coat the optics just inside the source. These can be easily removed on most instruments (after venting). Clean according to manufacturers recommendations. Poor Vacuum Syringe Pump Mean free path dictates that MS systems need vacuum to operate, and incomplete vacuum can cause signal instability. Check the vacuum gauges against the installation target values to be sure vacuum levels are appropriate and stable. An incorrect readback can be poor pumping, a leak, or a bad gauge. Pulsation in the flow rate can cause signal instability. This is usually (but not always) at a slower frequency than the instability caused by an unstable electrospray. Check that the flow from the syringe (or other pump) is stable,

Troubleshooting Typical Causes of Failure (Resolution) Poor Analyzer Vacuum Inappropriate vacuum can cause poor resolution in both ToF, LIT, and OrbitTrap analyzers. Check the vacuum gauges against the installation target values to be sure vacuum levels are appropriate and stable. MS Analyzer Tune Parameters (Recovered Tune Page) Small changes in tune parameters can cause resolution issues. Load the backup tune page to make sure nothing odd has happened. If this fails, fall back to your hard copy records. Resolution is highly instrument-type specific, but are most often related to pusher and reflectron tuning for ToFs. Automatic gain control/space charging effects is a common cause of decreased resolution in trapping instruments. High Voltage Power Supplie(s) Check voltage readbacks on analyzer power supplies. If you have not recorded the appropriate values, call your vendor technical support and they typically can give you good values over the phone.

Troubleshooting Typical Causes of Failure (MS/MS Fragmentation) No Collision Gas Check gas inlet pressure, and make sure collision gas is on in tune page. Check collision cell pressure is comparable to installation values (and should increase when turning collision gas on). Improper Collision Energy (CE) MS1 Transmission Window or Calibration Collision Cell Power Supply Nominal Mass Calibration off Corrupt Tune Page Check CE or normalized CE setting to be sure it is correct. On instruments with a quadrupole as the first mass analyzer (QqQ, QToF, Q- Orbitrap), the MS1 transmission window and calibration is often ignored but is critical for precursor ion selection. Check MS1 calibration and transmission window using sliding window around infusion of known analyte. Check voltage readbacks on analyzer power supplies. If you have not recorded the appropriate values, call your vendor technical support and they typically can give you good values over the phone. (Calibration) Remove the calibration (default cal) and check the nominal mass of a known standard. Follow instrument-specific instructions for correction of nominal mass value. Corrupt tune pages can cause calibration failures, manual and automated. Load a backup tune page or start a new tune page.

Preventive Maintenance Simple things to keep your LC/MS running.

Typical PM steps Verify system performance prior to venting. Record vacuum readings. Vent the system following the recommended steps for your system. Always refer to the associated hardware manual. Service the vacuum pump(s). Replace the sprayer / probe electrode. Remove and clean the front interface assembly. Clean or replace air filters. Pump system down to base vacuum ~ 4 hrs. Tune and calibrate the system.

Venting a Mass Spectrometer It is very important to understand the procedures for venting your system. This is either accomplished through the software or by manually turning off switches. The turbo pump(s) should always be allowed to gradually spin down before turning off the rough pumps. Never break vacuum connections to force the venting cycle.

Vacuum Pump maintenance Oil change every 6 months or sooner depending on application and usage. Always follow manufacturers recommendation for type of oil. Scroll pump or Dry pump Tip seal replacement every 18-24 months depending on pump model and manufacturer. Always inspect and clean the pump motor air intake to remove all build up of dust and dirt.

Ion Optics Cleaning Clean orifice, skimmer, interface region and associated optics. Recommended cleaning solutions 10% formic acid, IPA, MeOH & H20 or Alconox & H20 Sonicate 15-20 min. Blow off with dry N2 and inspect.

Ion Source - Probe / Sprayer Maintenance Clean / flush housing. Remove probe Assy. and replace capillary or electrode. Always refer to the manufacturers documentation for recommended adjustment positioning or spacing. Replace associated ferrules and fittings as necessary Inspect associated heaters for integrity and performance. Verify spray performance with standard tuning mix.