Optimizing the Post-implementation Monitoring of a LC-MS/MS Method
|
|
|
- Scott McKinney
- 9 years ago
- Views:
Transcription
1 Optimizing the Post-implementation Monitoring of a LC-MS/MS Method Julianne Cook Botelho, PhD Centers for Disease Control and Prevention Atlanta, GA AACC- Mass Spectrometry in the Clinical Lab: Best Practice and Current Applications Tuesday, September 17, 2103 National Center for Environmental Health Division of Laboratory Sciences
2 Nothing to disclose Disclosures Any mention or allusion to specific vendors or vendor parameters should not be seen as endorsement Some aspects of CLSI C60(to be renamed C62) Liquid Chromatography-Mass Spectrometry Methods; Draft Guideline are included in this presentation CLSI Guideline C60 is still under development and may not represent the final guidelines
3 Learning Objectives After this presentation, you should be able to: 1. Identify what post analytic monitoring is and why it is necessary 2. Describe how post analytic monitoring can be used to evaluate the performance of your LC-MS Instrument 3. Explain how QC materials and other parameters can be used to confirm sample analysis performance 4. Identify how proficiency testing (PT) programs can be used in combination with assessments to monitor assay performance over time
4 What is post-implementation monitoring? Pre-Validation Validation Post-Validation Approach to verify test method performance specification during routine analysis
5 Goals and Objectives of Post-implementation Monitoring Assure validity of measurements Confidence in reported values Identify risk of potential problems Minimize, identify, and correct analytical errors in real time Ensure optimum LC-MS/MS and method performance Less downtime, increased productivity Compliance CLIA, FDA
6 Parameters to monitor Parameter Instrument Performance Sample Analysis Performance Tools System Suitability Sample (SSS) Quality Control Materials (QC) and additional parameters Assay Performance Over Time Proficiency Testing (PT), Accuracy Assessments, Lot to Lot, and Instrument Comparisons
7 Parameters to monitor Parameter Instrument Performance Sample Analysis Performance Tool System Suitability Sample (SSS) Quality Control Materials (QC) and additional parameters Assay Performance Over Time Proficiency Testing (PT), Accuracy Assessments, Lot to Lot, and Instrument Comparisons
8 Parameter Parameters of SSS to monitor for instrument performance Monitor in real time and over time Shift in Retention Time (RT) gradual or sudden Acceptable criteria set during validation Peak Shape Tailing, Fronting, Asymmetry Additional Peaks Drop in peak intensity overall drop (gradual/sudden) Increase in background noise Change in Analyte:IS Ratio
9 SSS- Sample Characteristics Type of Sample non-extracted sample (analyte and IS) Does NOT go through sample preparation Concentration Suggestions- near LLOQ to evaluate instrument sensitivity middle of analytical measurement range near a medical decision point Prepare a large batch and store according to stability standards
10 SSS- Application for Instrument Performance Evaluation Use of SSS Purpose Frequency Before Analysis Confirm instrument performance Before the start of a run During a run Confirm and monitor performance of instrument during the entire run Depends on stability of instrument; if changes in parameters are common consider more injections After instrument maintenance, power outage, break in vacuum, tuning/calibration Confirm instrument is within performance specifications and stable prior to sample analysis Multiple injections Approx. 5-10
11 SSS Shift in RT Parameter to Monitor Source Could Indicate Shift in Retention Time (RT) gradual or sudden Column HPLC Buffers HPLC Pumps degradation; not equilibrated properly; temperature change preparation; evaporation; ph change; bubbles; temperature performance flow rate; mixing; leaks; bubbles; change in length of tubing SSS - Retention Time Over Time Run Number
12 SSS- Peak Shape and Additional Peaks Parameter to Monitor Source Could Indicate Peak Shape Tailing, Fronting, Asymmetry Column degradation, not equilibrated properly, carry over Additional Peaks Column carry over from previous runs/experiments HPLC Buffers contamination Mass Spectrometer curtain plate needs to be cleaned SSS-Chromatogram SSS-Chromatogram Contamination? Area Count Area Count Minutes Minutes
13 SSS- Change in Intensity and Ratio Parameter to Monitor Source Could Indicate Drop in peak intensity Overall (gradual/sudden) Increase in background noise Mass spectrometer HPLC Buffers Injector Column Change in Analyte: IS Ratio Mass spectrometer front end of MS needs to be cleaned; probe height; generator filter evaporation; ph change which could affect ionization; poor quality partial injection; needle depth; clogged injector or loop degradation calibration/tuning of MS needed; charging issue SSS- Area Count of Analyte Over Time Area Count Run Number
14 Parameters to monitor Parameter Instrument Performance Sample Analysis Performance Tool System Suitability Sample (SSS) Quality Control Materials (QC) and additional parameters Assay Performance Over Time Proficiency Testing (PT), Accuracy Assessments, Lot to Lot, and Instrument Comparisons
15 QCs Run Evaluation: Basic parameters to monitor Parameter to Monitor Precision Monitor in real time and over time Acceptable criteria set during validation Concentration Trend Peak Characteristicsshape, RT, intensity, ratio
16 QC Materials- Sample Characteristics Characteristics Requirement Application Type Matrix match MUST undergo ALL aspects of sample preparation Number of Levels Minimum 2 (CLIA) Recommend 3+ Concentrations- near medical decision points, over range of method Replicates in a run 5% of total patient sample analysis 120 sample batch= 6 QCs (3 levels in replicate) Placement in a run Optimized to detect shifts or errors Beginning, middle, and end Acceptability Established during validation (even for commercial QC materials) Use trend rules and multirule, not just +/-SD
17 QC consideration for multiplexed assays and multiple instruments for a validated method Instrumentation Analytes Requirements 1 System 1 Injector 1 Column 1 MS Multiple Systems (even same vendor/model) Multiplex Approach 1 2+ Injectors 2+ Columns 1 MS Multiplex Approach 2 1 injector 2+ Columns 1 MS Multi 1 set of QCs that contains all analytes During review analytes treated as separate runs Single or Multi QCs and Calibrators (CC) shared for method Own set of QCs and CC for each run Scheduled comparisons Single or Multi QCs and CC shared for the method Own set of QCs and CC for each run Schedule comparisons Single or Multi QCs and CC shared for method Split QCs and CC for each run 1 level QC each column
18 Other tools: Basic parameters to monitor Parameter to Monitor IS Area Count Stability Monitor in a run and over time Blanks Calibration Curve Slope R 2 Area counts between runs Acceptable criteria set during validation Evaluate trends within and between runs QI/CI Ratio Peak Characteristicsshape, RT, intensity, ratio
19 Other tools: IS Area Count Stability Parameter to Monitor IS Area Count Stability Could Indicate Deterioration of IS Contamination Pipetting Error Sample preparation recovery Loss of detector sensitivity Ion suppression Charging IS Area Count in a Run Area Count Sample Number
20 Other tools: Blanks and Calibration Curve Parameters Parameter to Monitor Blanks Double Blanks (extracted sample no IS) Blank (extracted sample with IS only) Could Indicate Interference introduced during sample preparation Carryover Degradation/Contamination of IS Solvent Blank (solvent only- no sample prep) Calibration Curve Slope R 2 Area counts between runs Deterioration of calibrators Contamination Loss of detector sensitivity
21 Other tools: QI/CI of Samples to Report Parameter to Monitor QI/CI Ratio Could Indicate Interference Monitor 2 transitions for your analyte and IS QI- quantitation Ion (used to determine your concentration; most abundant) CI- confirmation ion Should not change by +/ % from that of the mean ratio of the standards Monitored within and across all samples and compared to criteria established during validation
22 If a run or sample is out of control, now what? Investigate Use the whole picture SSS, QCs, CC, IS, Blanks, Samples Individual Sample failed Can you repeat it? Is it worthwhile? Document all corrective actions Track parameters over time to monitor the performance of you method catch issues before they become problems
23 Determine if reporting is acceptable Examples Shift in RT of both IS and analyte throughout run in SSS, CC, QC, Samples- data reportable 2 of the 3 QCs for run out of control- data NOT reportable QI/CI ratio for individual sample outside specs- reviewed peak it was an integration problem- corrected- data reportable Low QC fails- determined it was a loss in sensitivity, cleaned front end of the instrument check performance again with SSSreinjection of run- run meets all criteria- data reportable
24 Parameters to monitor Parameter Instrument Performance Sample Analysis Performance Tools System Suitability Sample (SSS) Quality Control Materials (QC) and additional parameters Assay Performance Over Time Proficiency Testing (PT), Accuracy Assessments, Lot to Lot, and Instrument Comparisons
25 Assay Performance Over Time- Accuracy Accuracy assessment strongly recommended every 6 months or when new lot of calibrators are implemented. Evaluation Tiered Approach (same as during validation) 1) Best Practice- Method Comparison to a JCTLM acknowledged Reference Method JCTLM database Recommend EP9 A2 40 samples 2) Evaluation of available JCTLM acknowledged reference materials limited concentrations available, typically 1-4 levels material should be commutable 3) Spike and recovery analysis native matrix, minimum 3 concentration 5 replicates
26 Assay Performance Over Time- Accuracy Consider standardization and accuracy based PT programs to evaluate accuracy Tier 1 Accuracy Evaluation with 40 sample comparison CDC HoSt- Testosterone, Estradiol CDC VDSCP- Vitamin D CDC CRLMN- Lipids NGSP- hemoglobin A1c Tier 2 Accuracy Evaluation with limited number of samples NIST reference standards and other CRM Accuracy Based PT Programs- CAP, DEQAS NIST QAP- Vitamin D CDC LSP- Lipid Standardization Program
27 Assay Performance Over Time- Peer Group PT Programs Examples- CAP, NYDOH Provided matrix like material to evaluate performance 2-3 times a year, ~5 concentration levels. Tend to be multi-analyte panels of material As a participate you are asked to measure this material like a real sample Performance evaluation based on peer group performance Criteria +/- 3 SD of all method mean or peer group mean Monitor for trends/shifts (i.e. are results always on one side of the mean?) and use CRM to investigate further CLSI GP27
28 Assay Performance Over Time- Accuracy Based PT Programs Material used in testing needs to be commutable No Peer Groups All methods evaluated against each other OR to a true value (typically assigned by JCTLM acknowledged method) If reference method is available evaluation can be based on accuracy Results can be used as accuracy monitoring as well as monitoring trends/shifts as with peer group PT
29 Assay Performance Over Time- Instrument Comparison Need with the use of multiple LC-MS/MS Systems (same/different manufactures) Requires instrument comparison every 6 months to evaluate performance Recommend following CLSI EP9A2- method comparison 40 samples Performance Criteria established based on performance criteria and clinical utilization of the results
30 Parameters to monitor Parameter Instrument Performance Sample Analysis Performance Tools System Suitability Sample (SSS) Quality Control Materials (QC) and additional parameters Assay Performance Over Time Proficiency Testing (PT), Accuracy Assessments, Lot to Lot, and Instrument Comparisons
31 Words to the Wise Think about post implementation during development/validation What should I track? What is acceptable? Use technology (excel, LIMs system etc.) Create flags (for run and samples) Track long term performance Constantly evaluate your parameters, are they appropriate? Use calendar to schedule required post analytic monitoring Train your staff What does a flag mean, why would there be a flag, is there something I can do? Document, document, document
32 Acknowledgement C60 Document Development Committee Dr. Bill Clark Dr. Ross Molinaro Dr. Lorin Bachmann CDC Team Dr. Hubert Vesper Ashley Ribera Hans Cooper, MPH
33
34 Thank You Contact Information: For more information please contact Agency for Toxic Substances and Disease Registry 4770 Buford Hwy. NE, Chamblee, GA Telephone: CDC-INFO ( )/TTY: Web: The findings and conclusions in this report are those of the authors and do not necessarily represent the official position of the Centers for Disease Control and Prevention.
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
LC-MS/MS, the new reference method for mycotoxin analysis
LC-MS/MS, the new reference method for mycotoxin analysis What is necessary for the implementation of the multi-residue mycotoxin method in an analytical laboratory? Azel Swemmer [email protected] AFMA
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
Electrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
Ecology Quality Assurance Glossary
Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability
Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories
Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
Performance and advantages of qnmr measurements
Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,
Why System Suitability Tests Are Not a Substitute for Analytical Instrument Qualification
White paper System Suitability Tests and AIQ R.D.McDowall PhD Principal, McDowall Consulting Paul Smith European Validation Program Manager, PerkinElmer Nicola Vosloo PhD European Market Development Leader,
The Theory of HPLC. Gradient HPLC
The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims
TEST METHOD VERIFICATION AND VALIDATION
1 TEST METHOD VERIFICATION AND VALIDATION SWACM 2014 MICHAEL LOEFFELHOLZ, PH.D., ABMM DEPT. PATHOLOGY UNIV TEXAS MEDICAL BRANCH GALVESTON, TX 2 OBJECTIVES Define validation and verification Describe components
Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity
For the benefit of VaporFi, Inc. Samples RMN070714 Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity GC/MS Analysis (QG02) Anabasine, Anatabine, Myosmine, Nornicotine Enthalpy Analytical,
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
Mass Spectrometry Signal Calibration for Protein Quantitation
Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of
STANDARD OPERATING PROCEDURES
PAGE: 1 of 33 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
GC-xt. The most flexible GC /GC-MS liquid sample injector
GC-xt Prep and Load Platform The most flexible GC /GC-MS liquid sample injector Environmental Foods/Beverages Consumer Products Forensics Petrochemicals/Polymers Pharmaceuticals GC-xt Liquid mode Prep
Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ)
Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) Closing Date and Time: Wednesday, April 13, 2016 at 5:00 p.m. (Pacific Time) Contact: Aaron Galloway Fax: 541-888-3250
Assay Qualification Template for Host Cell Protein ELISA
Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
Introduction. The following definitions may help you better understand the components of the data report.
Introduction This document is designed to help our clients understand the quality control requirements and limitations of data reporting. There are three sections to this document. The first section will
Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
AutoSpec Premier. Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements?
AutoSpec Premier Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements? Looking for sensitivity, selectivity and dynamic range? The AutoSpec Premier is the latest
Signal, Noise, and Detection Limits in Mass Spectrometry
Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
SCION SQ Series GC-MS
SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection. LGC/R/2011/181 LGC Limited 2011
A feasibility study of the determination of B-group vitamins in food by HPLC with Mass Spectrometric detection LGC/R/2011/181 LGC Limited 2011 GC11/0005A A feasibility study of the determination of B-group
A commitment to quality and continuous improvement
Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. [email protected] Overview GMP requirements for Quality Control laboratories
Implementing New USP Chapters for Analytical Method Validation
Implementing New USP Chapters for Analytical Method Validation March 2010 Ludwig Huber Fax.: +49 7243 602 501 E-mail: [email protected] Today s Agenda Handling Method Changes vs. Adjustments
Medical Cannabis Laboratory Approval Program
Medical Cannabis Laboratory Approval Program Application Process and Required Documentation After the publication of the Medical Cannabis Laboratory Application, currently expected by February 16, 2015,
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
Air Sampling for Gases and Vapors. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike License. Your use of this material constitutes acceptance of that license and the conditions of use of materials on this
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
Therapeutic Drug Monitoring of Antiretroviral Drugs with HPLC-MS
Therapeutic Drug Monitoring of Antiretroviral Drugs with PLC-M Ursula Gutteck-Amsler, Katharina M. Rentsch Abstract Prospective and retrospective studies have provided some evidence of the clinical and
Overview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.
Halogen Free: What, Why and How Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. We will discuss What are Halogens and how we interact with them. Why Halogens, within consumer
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry
Limits to Detection/Measurement Ionization is fundamentally a Probabilistic Process Just like radioactive decay So is transmission through the analyzer There is an intrinsic statistical uncertainty Proportional
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
S olid Ph ase Extraction (SP E) Method:
Solid Phase Extraction of US EPA 535 for Chloroacetanilide and Acetamide Degradates in Drinking Water Samples Prior to Liquid Chromatography/ Tandem Mass Spectrometry Toni Hofhine, Horizon Technology,
High-Throughput 3-D Chromatography Through Ion Exchange SPE
High-Throughput 3-D Chromatography Through Ion Exchange SPE Application Note 205 Luke Roenneburg and Alan Hamstra (Gilson, Inc.) Introduction 2-dimensional (2-D) separation is the separation of a sample
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
Waters Core Chromatography Training (2 Days)
2015 Page 2 Waters Core Chromatography Training (2 Days) The learning objective of this two day course is to teach the user core chromatography, system and software fundamentals in an outcomes based approach.
Analysis of Liquid Samples on the Agilent GC-MS
Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
and Drug Administration/Office of Regulatory Affairs (FDA/ORA) laboratories.
Page 1 of 13 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6.C.4. added or electronically ) 7. Definitions 8. Records
Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate
Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in
Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation
IJPAR Vol.4 Issue 4 Oct Dec - 2015 Journal Home page: ISSN: 2320-2831 Research Article Open Access Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation V.
Determination of Caffeine in Beverages HPLC-1
Determination of in s HPLC-1 Introduction This experiment provides an introduction to the application of High Performance Liquid Chromatography (HPLC) to the solution of complex analytical problems. Cola
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005
CAC/GL 56-2005 Page 1 of 6 GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005 CONFIRMATORY TESTS When analyses are
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
ORGANIC SAMPLE PREPARATION
ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,
Factors Influencing LC/MS/MS Moving into Clinical and Research Laboratories
Factors Influencing LC/MS/MS Moving into Clinical and Research Laboratories Matthew Clabaugh, Market Development AACC Workshop St. Louis, MO September 17,18 2013 Factors Influencing LCMSMS Moving into
HRMS in Clinical Research: from Targeted Quantification to Metabolomics
A sponsored whitepaper. HRMS in Clinical Research: from Targeted Quantification to Metabolomics By: Bertrand Rochat Ph. D., Research Project Leader, Faculté de Biologie et de Médecine of the Centre Hospitalier
Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO
Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i
Introduction. Methods. Sample Processing
PO-CON1656 High-sensitivity, high-throughput quantitation of catecholamines and metanephrine in plasma by automated WCX-SP coupled to LC/MS/MS for clinical research ASMS 2016 MP-080 Ichiro HIRANO 1, Atsuhiko
Oasis HLB Cartridges and 96-Well Plates
CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION
HPLC Basics, Equipment, and Troubleshooting
HPLC Basics, Equipment, and Troubleshooting HPLC Basics, Equipment, and Troubleshooting explains chromatography in practical terms from the ground up HPLC Basics, Equipment, and Troubleshooting HPLC Basics,
Quality Assurance Program Quality Assurance Plan Periodic Audits Standard Operating Procedures Preventive Maintenance Outline Quality Assurance Plan (QAP) What is a QAP Why is a QAP required? What does
STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080
Training on the ZQ Open access MS Questions? Contact Dr. Allis Chien [email protected] 650-723-0710 0710 STANFORD UNIVERSITY MASS SPECTROMETRY STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas
aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS
aurora M90 A new era in ICP-MS Innovation with Integrity ICP-MS Your Partner in ICP-MS Solutions Bruker continues to find new and novel ways to meet your changing needs. As a leader in elemental analysis
Amino Acid Analyzer L-8900
Amino Acid Analyzer L-8900 VWR - Hitachi Your Partner in Amino Acid Analysis Hitachi has manufactured more than 1800 Amino Acid Analysers for over 40 years. The new Amino Acid Analyser Model L-8900 is
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
Supplementry Discussion Topic: Advantages and disadvantages of current analytical approaches and the best carbon pair for you
Supplementry Discussion Topic: Advantages and disadvantages of current analytical approaches and the best carbon pair for you ph Measurements Advantages Hardware/Chemical /Standard When to use Spectrophotometric
Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
